Drawell SMART Series User Manual

USER’S MANUAL
SMART Series
Biological Microscope
Drawell International Technology Limited
Shanghai Drawell Scientific Instrument Co.,Ltd
Add : Suite 1506,Lane581 XiuChuan Rd., Greeland
Max-Mall,PuDong,Shanghai,China
Tel: 0086 21 50966080 Fax: 0086 21 50966080 Web : www.drawell.com.cn Email : sales01@drawell.com.cn
SMART
To ensure the safety and obtain satisfactory performance, please
study this operation instruction thoroughly before your operation.
SMART
Series
Series Biological Microscope
System Microscope
Operation Instruction
--
SMART
CONTENTS
Purpose
2
………………………………………………………………
1. Parts Name
3
………………………………………………………………
2. Specifications
4
………………………………………………………………
3. Installation
5
………………………………………………………………
4. Operation
6
………………………………………………………………
5. Image Collection
6
………………………………………………………………
6. Maintenance
7
………………………………………………………………
7. Troubleshooting
8
………………………………………………………………
8. Outfits
9-10
………………………………………………………………
SMART series biological microscope
is designed for college teaching, clinical
verify. The series microscope with
original style, steady structure,
convenient operation and clear image is
suitable for observing various biological
specimens, and mostly applied in
college and hospital.
Series Biological Microscope
--
SMART
B
A
①②C_③
D_
E_
FHJKG
Eyepiece
Objective
Clamp
Mechanical Stage
Condenser
Light Collector
Main Body
Seidentopf Binocular Head (Trinocular Head)
{
Power Input Fuse
F
L
M
N
A
Diopter Adjustment Ring
B
Nosepiece
C
Clamp Handle
D
Handle of Iris Aperture Diaphragm
E
Field Diaphragm Ring
F
Body Handle
G
Head Thrumb Screw
H
Mechanical Stage Moving Knob
J
Right Coarse & Fine Focusing Knobs
K
Potentiometer
L
Power Switch
M
Condenser Focusing Knob
N{
Tension Adjustment Ring
Left Coarse & Fine Focusing Knobs
Series Biological Microscope
1. Parts Name
--
SMART
Objective
Total Magnifications
Eyepiece
4X
10X
20X
(Optional)
40X
100X
10X
40X
100X
200X
400X
1000X
16X
64X
160X
320X
640X
1600X
Objective
Numerical
Aperture (N.A.)
Objective
Field
Resolving
Power
Working
Distance
4X
0.10
5mm
2.8µm
6.73mm
10X
0.25
2mm
1.1µm
4.19mm
20X (S)
0.40
1mm
0.69µm
2.14mm
40X (S)
0.65
0.5mm
0.42µm
0.45mm
100X (Oil) (S)
1.25
0.2mm
0.22µm
0.12mm
Series Biological Microscope
2. Specifications
2.1 Total magnifications
2.2 Objectives
2.3 The other specification
2.3.1 Mechanical tube length: 160mm
2.3.2 Conjugate distance: Infinity (Non- Infinity: 195mm)
2.3.3 Head: Seidentopf Binocular or Trinocualr, Inclined 30°, Rotable 360°,
Anti-fungal systems. Interpupillary Adjustable Distance Is 50-75mm,
Diopter adjustable range ±5.
2.3.4 Nosepiece: Quadplex nosepiece
2.3.5 Mechanical stage: Size 145mm×140mm
X-Y travel 76mm×52mm
2.3.6 Focusing systems: Coaxial Coarse And Fine Focusing Knobs, Coarse stroke 26mm,
Fine division 2µm, Condenser up-down range 22mm
2.3.7 Condenser: Abbe condenser, N.A. 1.25, Adjustable aperture, Aperture center can be adjustable.
2.3.8 Illumination: Non-spherical System
--
SMART
Condenser
screw
Condenser
installing ring
Condenser
3. Installation
Please clean the operation desk before installation.
Put out the microscope of the carton and put it on
the desk.
Make sure the supply voltage meets the instrument’s
requirement and the power switch is off.
Installation Instruction Fig.:
1.Turn the binocular (trinocular) head to working
position;
2.Put out the dust cover of the eyepiece tube;
3.Insert into the eyepiece;
4.Install Abbe Condenser
3.1 Eyepiece Tube:
Loose the Head Thrumb Screw G, turn the
tube to observing position, then tighten the
Screw G.
3.2 Put out the eyepiece dust cover.
3.3 Eyepiece
Put out the eyepiece of the carton, and insert it
into the tube. Please don’t touch the lens of the
eyepiece by hand.
installing ring. Loose the
condenser screw, then install the
condenser, and make the
condenser graduation face to the
front, Tighten the condenser
screw, and higher the condenser
1
2
4
2.3.9 Filter: Built in blue filer
2.3.10 Electric components: Input voltage AC85-265V/LED AC170-270V, 50/60Hz
Series Biological Microscope
Output voltage DC1.2-6V/LED DC0.4-3.5V
6V/20W halogen lamp
Rotation potentiometer with power switch
Fuse 2A ф5×20
3.4 Condenser
Put out the condenser of the carton, then turn Condenser Focusing Knob M to lower condenser
--
SMART
5) Adjust condenser aperture diaphragm
Adjust condenser (condenser focusing screw M) and aperture (Aperture diaphragm lever D)
Observing
1) Turn power switch, adjust light
Turn on Switch L, and turn Potentiometer K.
2) Put the sample
Put the sample on stage , (open the clamp
C, and put the sample into it), turn stage knob
H to put the sample into optical path.
3) Put 10X objective into
optical path and focus
Turn Nosepiece B, put 10 × objective into
optical path. Turn Coarse & Fine Focusing
Knobs I and N to focus.
4) Adjust interpupillary distance and diopter
Adjust head to make two fields into one;
adjust diopter adjuster A to make diopter for
two eyes same.
6) Adjust and center field
diaphragm
Adjust field diaphragm (field diaphragm ring
E)
Put the objective into
light path, and focus
installing ring to top.
Please don’t touch the lens by hand.
3.5 Power
Put in power, open Switch L, and turn Potentiometer J.
Series Biological Microscope
4. Operation
4.1 Bright field operation process instruction
--
SMART
6.3.1 Turn off power, and pull out plug.
6.3.2 Wait the bulb become cool. Please be sure that the bulb is cool, then follow
by the nest operations.
6.3.3 Lay aside the microscope reliably, unscrew the
knurled thumb screw of the lamp housing cover on
the underside of base.
6.3.4 Pull over the lamp housing cover.
6.3.5 Pull out the bulb should be replaced, hold a
Series Biological Microscope
5.Image Collection
5.1 Installing
Connect the C-mount with CCD camera or connect camera with camera adaptor, then connect it with
c-mount, finally put it into microscope.
5.2 Using
First get a clear image from eyepiece, then put out lever on the side of trinocular head, and collect
image with camera. Clear image should be in screen. Adjust B14 fine focusing knobs to get it clear if
image isn’t clear.
6. Maintenance
6.1 Clean microscope
6.1.1 Don’t touch the lens with hand, Dust on lens should be cleaned by soft brush or absorbent
cotton or cleaned by absorbent cotton, lens paper with the mixture of alcohol and ether (proportion 1:4).
6.1.2 Alcohol and ether all are burnt early, please take them away from fire. Be careful for turn on
and off power.
6.1.3 Don’t clean painted metal and galvanizing metal with organic solvent such as alcohol, ether or
the mixture of the both. Silicon cloth or soft cleaning preparation is suggested to clean it.
6.1.4 Plastic should be cleaned by soft cloth with clear water.
6.2 Environment of using and placing
6.2.1 Microscope should be used and placed in a cool, dry, non-dust, non-shake and non-corrosive
gases environment.
6.2.2 Microscope should be used in environment of indoor temperature 0°-40°C and maximum
relative humidity 85%.
6.2.3 Removing equipment is suggested to be installed when microscope used in heavy humidity
area to avoid fungus and mist damage instrument.
6.2.4 Please pay attention to prevent microscope from violent shake and vibration in application and
in carrying. Don’t drag it on the surface of worktable to avoid damage to microscope and worktable.
6.3 Replacement of bulb
--
SMART
After working for above 10 hours continuously,
better cut off the microscope about 30 minutes.
6.4 Replacement of fuse
6.4.1 Cut off power of microscope, and pull out the
plug.
6.4.2 Unscrew fuse cap in the back of base, take out
old fuse.
6.4.3 Replace a new fuse, then screw the fuse cap.
Trouble
Causation
Remedy
Switch on but bulb dark
Plug is unreliable
Plug in again
Bulb is broken
Change bulb
Fuse is broken
Change fuse
Bulb is flickering or brightness is unsteady
Bulb is unstable
Insert it again
Bulb is broken
Replacing bulb
Brightness of view field isn’t enough or is Uneven
Bulb specification doesn’t meet the requirement
Replacing bulb
Brightness isn’t adjusted correctly
Adjust rotation potentiometer
Objective isn’t in correct position
Make the objective in correct position
The size of iris aperture is too small
Adjust the size of iris aperture
Brightness of view field isn’t enough or is Uneven
Lens (objective, eyepiece, condenser, light collector)
has
dust
Clean it Position of condenser is too low
Higher condenser
Image isn’t clear
(contrast or definition
isn’t enough)
Cover glass of specimen doesn’t meet the requirement
Use required thickness cover glass (0.17mm)
Cover glass of specimen isn’t in up direction
Place specimen correctly
Surface of objective lens isdirty (especially it is easy for the front lens of 40X objective to dip in immersion oil)
Clean it Immersion oil isn’t used for
100X objective (oil)
Use immersion oil
Immersion oil doesn’t meet the requirement
Use immersion oil supplied by us
Series Biological Microscope
new bulb with silk cloth to avoid fingerprint and dust affect bulb brightness and service life, and
insert fully the contact pins into the bulb socket.
6.3.6 Close the lamp housing cover, and screw the knurled thumb screw.
7. Troubleshooting
In the period of using Smart series microscope, if there is any trouble occurs, please referring to the
following sheet listed some common troubleshooting resolve them.
--
SMART
Trouble
Causation
Remedy
There is bubble in immersion oil
Clear the bubble way
Size of iris aperture isn’t proper
Adjust the size of iris aperture
Position of condenser is too low
Readjust the position of condenser
One side of image is dark or image is moving as focusing
Objective isn’t in correct position
Make the objective in correct position
Specimen isn’t placed correctly
Place specimen levelly on
stage
and clip it with clamp
Objective touches specimen as changing low times objective to high times objective
Cover glass of specimen isn’t in up direction
Place specimen correctly
Cover glass doesn’t meet the requirement
Use required thickness cover glass (0.17mm)
Image observed by two eyes aren’t in superposition entirely.
Interpupilary distance isn’t adjusted correctly
Adjust interpupilary distance according to two eyes
It is easy for eyes to be tired during observing
Diopter isn’t adjusted correctly
Readjust diopter
Items
Specification
SMART-1
SMART-2
SMART-3
SMART-4
Eyepiece WF 10X-18mm
●●
WF 10X- 20mm
●●●●●●
Finity Achromatic Objective
4X
10X
40X(S)
100X/1.25 (Oil) (S)
Finity Plan Objective
4X
10X
40X/0.65 (S)
100X/1.25 (Oil) (S)
Infinity E-Plan Objectives
4X
10X
40X (S)
100X (Oil) (S)
Infinity Plan Objectives
4X
10X
40X (S)
100X (Oil) (S)
Plan 20X○○
Plan 60X (S)
Seidentopf Binocular Head
Inclined 30°,Rotatable 360°, Interpupilary Distance 48-75mm
●●●
Seidentopf Trinocular Head
Inclined 30°,Rotatable 360°, Interpupilary Distance 48-75mm, Light Distribution 20:80
○○○
Nosepiece
Quadplex , Quintuple (Optional)
●●●
Mechanical Stage
Stage Size: 145X140mm Travel: 76mmX52mm, Scale: 0.1mm
Coaxial Coarse and Fine
Focusing Knobs
Travel Range: 26mm, Scale: 2um Two Slide Holder
●●●
Integrated stage, Stage size: 182X140mm
○○○
Series Biological Microscope
8. SMART Series Biological Microscope Outfits
--
SMART
Travel range: 77mm×52mm, Two-slide holder
Condenser
Abbe N.A. 1.25(Iris Diaphragm)
●●●
Illumination 3W-LED Illumination Systems
●●●
6V/20W Halogen Lamp
○○○
6V/30W Halogen Lamp
○○○
Field Diaphragm
○○○
Dark Field Condenser
Dry, apply to 4 X - 40 X objective
○○○
Wet, apply to 100X objective
○○○
Polarizing Attachment
Analyzer/Polarizer
○○○
Independent phase contrast unit
10X /20X /40X /100X
Fluorescence Attachment
Epi fluorescence unit ( six-hole disc media which can be fixed with Uv /V/B/G and another filters) ,100W mercury lamp.
Epi fluorescence unit ( six-hole disc media which can be fixed Uv /V/B/G), LED fluorescence lamp.
Filter
Blue○○○○
Green○○○○
Yellow○○○○
C-Mount
0.47X Focus adjustable
○○○
1X○○○○
Mirror○○○○
Wire winding device
○○○
Series Biological Microscope
Note: “”In Table Is Standard Attachment. “” Is Optional Accessories
Packing Size: 420mm × 275mm × 450mm
Gross Weight: 7.5 kgs Net Weight: 6.5 kgs
-10-
SMART
Series Biological Microscope
Shanghai Drawell Scientific Instrument Co.,Ltd
Drawell International Technology Limited
Add : Suite 1506,Lane581 XiuChuan Rd., Greeland Max-Mall,PuDong,Shanghai Te l: 0086 21 50966080 Web : www.drawell.com.cn Email : sales01@drawell.com.cn
-11-