USER’S MANUAL
SMART Series
Biological Microscope
Drawell International Technology Limited
Shanghai Drawell Scientific Instrument Co.,Ltd
Add : Suite 1506,Lane581 XiuChuan Rd., Greeland
Max-Mall,PuDong,Shanghai,China
Tel: 0086 21 50966080
Fax: 0086 21 50966080
Web : www.drawell.com.cn
Email : sales01@drawell.com.cn

SMART
To ensure the safety and obtain satisfactory performance, please
study this operation instruction thoroughly before your operation.
Series Biological Microscope
System Microscope
Operation Instruction
-1-

SMART
SMART series biological microscope
is designed for college teaching, clinical
verify. The series microscope with
original style, steady structure,
convenient operation and clear image is
suitable for observing various biological
specimens, and mostly applied in
college and hospital.
Series Biological Microscope
-2-

SMART
Seidentopf
Binocular Head
(Trinocular Head)
Handle of Iris Aperture Diaphragm
Mechanical Stage Moving Knob
Right Coarse & Fine Focusing Knobs
Left Coarse & Fine Focusing Knobs
Series Biological Microscope
1. Parts Name
-3-

SMART
Objective
Total Magnifications
Eyepiece
Numerical
Aperture (N.A.)
Series Biological Microscope
2. Specifications
2.1 Total magnifications
2.2 Objectives
2.3 The other specification
2.3.1 Mechanical tube length: 160mm
2.3.2 Conjugate distance: Infinity (Non- Infinity: 195mm)
2.3.3 Head: Seidentopf Binocular or Trinocualr, Inclined 30°, Rotable 360°,
Anti-fungal systems. Interpupillary Adjustable Distance Is 50-75mm,
Diopter adjustable range ±5.
2.3.4 Nosepiece: Quadplex nosepiece
2.3.5 Mechanical stage: Size 145mm×140mm
X-Y travel 76mm×52mm
2.3.6 Focusing systems: Coaxial Coarse And Fine Focusing Knobs, Coarse stroke 26mm,
Fine division 2µm, Condenser up-down range 22mm
2.3.7 Condenser: Abbe condenser, N.A. 1.25, Adjustable aperture, Aperture center can be adjustable.
2.3.8 Illumination: Non-spherical System
-4-

SMART
Condenser
installing ring
3. Installation
Please clean the operation desk before installation.
Put out the microscope of the carton and put it on
the desk.
Make sure the supply voltage meets the instrument’s
requirement and the power switch is off.
Installation Instruction Fig.:
1.Turn the binocular (trinocular) head to working
position;
2.Put out the dust cover of the eyepiece tube;
3.Insert into the eyepiece;
4.Install Abbe Condenser
3.1 Eyepiece Tube:
Loose the Head Thrumb Screw G, turn the
tube to observing position, then tighten the
Screw G.
3.2 Put out the eyepiece dust cover.
3.3 Eyepiece
Put out the eyepiece of the carton, and insert it
into the tube. Please don’t touch the lens of the
eyepiece by hand.
installing ring. Loose the
condenser screw, then install the
condenser, and make the
condenser graduation face to the
front, Tighten the condenser
screw, and higher the condenser
2.3.9 Filter: Built in blue filer
2.3.10 Electric components: Input voltage AC85-265V/LED AC170-270V, 50/60Hz
Series Biological Microscope
Output voltage DC1.2-6V/LED DC0.4-3.5V
6V/20W halogen lamp
Rotation potentiometer with power switch
Fuse 2A ф5×20
3.4 Condenser
Put out the condenser of the carton, then turn Condenser Focusing Knob M to lower condenser
-5-

SMART
5) Adjust condenser aperture
diaphragm
Adjust condenser ⑤ (condenser focusing
screw M) and aperture (Aperture diaphragm
lever D)
1) Turn power switch,
adjust light
Turn on Switch L, and turn Potentiometer K.
Put the sample on stage ④ , (open the clamp
C, and put the sample into it), turn stage knob
H to put the sample into optical path.
3) Put 10X objective into
optical path and focus
Turn Nosepiece B, put 10 × objective into
optical path. Turn Coarse & Fine Focusing
Knobs I and N to focus.
4) Adjust interpupillary
distance and diopter
Adjust head ⑧ to make two fields into one;
adjust diopter adjuster A to make diopter for
two eyes same.
6) Adjust and center field
diaphragm
Adjust field diaphragm (field diaphragm ring
E)
Put the objective into
light path, and focus
installing ring to top.
Please don’t touch the lens by hand.
3.5 Power
Put in power, open Switch L, and turn Potentiometer J.
Series Biological Microscope
4. Operation
4.1 Bright field operation process instruction
-6-

SMART
6.3.1 Turn off power, and pull out plug.
6.3.2 Wait the bulb become cool.
▲ Please be sure that the bulb is cool, then follow
by the nest operations.
6.3.3 Lay aside the microscope reliably, unscrew the
knurled thumb screw of the lamp housing cover on
the underside of base.
6.3.4 Pull over the lamp housing cover.
6.3.5 Pull out the bulb should be replaced, hold a
Series Biological Microscope
5.Image Collection
5.1 Installing
Connect the C-mount with CCD camera or connect camera with camera adaptor, then connect it with
c-mount, finally put it into microscope.
5.2 Using
First get a clear image from eyepiece, then put out lever on the side of trinocular head, and collect
image with camera. Clear image should be in screen. Adjust B14 fine focusing knobs to get it clear if
image isn’t clear.
6. Maintenance
6.1 Clean microscope
6.1.1 Don’t touch the lens with hand, Dust on lens should be cleaned by soft brush or absorbent
cotton or cleaned by absorbent cotton, lens paper with the mixture of alcohol and ether (proportion
1:4).
6.1.2 Alcohol and ether all are burnt early, please take them away from fire. Be careful for turn on
and off power.
6.1.3 Don’t clean painted metal and galvanizing metal with organic solvent such as alcohol, ether or
the mixture of the both. Silicon cloth or soft cleaning preparation is suggested to clean it.
6.1.4 Plastic should be cleaned by soft cloth with clear water.
6.2 Environment of using and placing
6.2.1 Microscope should be used and placed in a cool, dry, non-dust, non-shake and non-corrosive
gases environment.
6.2.2 Microscope should be used in environment of indoor temperature 0°-40°C and maximum
relative humidity 85%.
6.2.3 Removing equipment is suggested to be installed when microscope used in heavy humidity
area to avoid fungus and mist damage instrument.
6.2.4 Please pay attention to prevent microscope from violent shake and vibration in application and
in carrying. Don’t drag it on the surface of worktable to avoid damage to microscope and worktable.
6.3 Replacement of bulb
-7-

SMART
▲After working for above 10 hours continuously,
better cut off the microscope about 30 minutes.
6.4 Replacement of fuse
6.4.1 Cut off power of microscope, and pull out the
plug.
6.4.2 Unscrew fuse cap in the back of base, take out
old fuse.
6.4.3 Replace a new fuse, then screw the fuse cap.
Bulb is flickering or
brightness is
unsteady
Brightness of view field
isn’t enough or is
Uneven
Bulb specification doesn’t
meet the requirement
Brightness isn’t adjusted
correctly
Adjust rotation
potentiometer
Objective isn’t in correct
position
Make the objective in correct
position
The size of iris aperture is too
small
Adjust the size of iris aperture
Brightness of view field
isn’t enough or is
Uneven
Lens (objective, eyepiece,
condenser, light collector)
has
dust
Clean it
Position of condenser is too low
Image isn’t clear
(contrast or definition
isn’t enough)
Cover glass of specimen
doesn’t meet the
requirement
Use required thickness cover
glass (0.17mm)
Cover glass of specimen isn’t in
up direction
Surface of objective lens
isdirty (especially it is easy for
the front lens of 40X objective
to dip in immersion oil)
Clean it
Immersion oil isn’t used for
Immersion oil doesn’t meet the
requirement
Use immersion oil
supplied by us
Series Biological Microscope
new bulb with silk cloth to avoid fingerprint and dust affect bulb brightness and service life, and
insert fully the contact pins into the bulb socket.
6.3.6 Close the lamp housing cover, and screw the knurled thumb screw.
7. Troubleshooting
In the period of using Smart series microscope, if there is any trouble occurs, please referring to the
following sheet listed some common troubleshooting resolve them.
-8-

SMART
There is bubble in immersion oil
Size of iris aperture isn’t
proper
Adjust the size of iris
aperture
Position of condenser is too low
Readjust the position of
condenser
One side of image is
dark or image is
moving as focusing
Objective isn’t in correct
position
Make the objective in
correct position
Specimen isn’t placed correctly
Place specimen levelly on
stage
and clip it with clamp
Objective touches
specimen as changing
low times objective to
high times objective
Cover glass of specimen isn’t in
up direction
Cover glass doesn’t meet the
requirement
Use required thickness
cover glass (0.17mm)
Image observed by two
eyes aren’t in
superposition entirely.
Interpupilary distance isn’t
adjusted correctly
Adjust interpupilary
distance according to two eyes
It is easy for eyes to be
tired during observing
Diopter isn’t adjusted correctly
Finity Achromatic
Objective
Infinity E-Plan
Objectives
Seidentopf Binocular Head
Inclined 30°,Rotatable 360°,
Interpupilary Distance 48-75mm
Seidentopf
Trinocular Head
Inclined 30°,Rotatable 360°,
Interpupilary Distance 48-75mm,
Light Distribution 20:80
Quadplex , Quintuple (Optional)
Stage Size: 145X140mm
Travel: 76mmX52mm, Scale: 0.1mm
Coaxial Coarse and Fine
Focusing Knobs
Travel Range: 26mm, Scale: 2um
Two Slide Holder
Integrated stage, Stage size:
182X140mm
Series Biological Microscope
8. SMART Series Biological Microscope Outfits
-9-

SMART
Travel range: 77mm×52mm,
Two-slide holder
Abbe N.A. 1.25(Iris Diaphragm)
Illumination
3W-LED Illumination Systems
Dry, apply to 4 X - 40 X objective
Wet, apply to 100X objective
Independent phase
contrast unit
Epi fluorescence unit ( six-hole disc
media which can be fixed with Uv /V/B/G
and another filters) ,100W mercury
lamp.
Epi fluorescence unit ( six-hole disc
media which can be fixed Uv /V/B/G),
LED fluorescence lamp.
Series Biological Microscope
Note: “●”In Table Is Standard Attachment. “○” Is Optional Accessories
Packing Size: 420mm × 275mm × 450mm
Gross Weight: 7.5 kgs Net Weight: 6.5 kgs
-10-

SMART
Series Biological Microscope
Shanghai Drawell Scientific Instrument Co.,Ltd
Drawell International Technology Limited
Add : Suite 1506,Lane581 XiuChuan Rd., Greeland Max-Mall,PuDong,Shanghai
Te l: 0086 21 50966080
Web : www.drawell.com.cn
Email : sales01@drawell.com.cn
-11-