8.2Auxiliary air valve ................................................................................................................. 38
9Pneumatics 39
9.1PEEP/Pmax valve control .................................................................................................... 39
9.2APL bypass valve control ..................................................................................................... 40
All rights reserved. Copyright reserved.
K5330500IECIVZ.fm 10.11.05
I
Contents
10Electrical block diagram 41
11Function description: Control PCB 41
12Control panel assembly 42
13FiO2 Measurement 44
14Respiratory Flow Measurement 45
15Gas flow rate measurement 46
16Anesthetic vaporizer(s) 47
17Leak test 49
17.1 System leak test ................................................................................................................... 49
17.2 Patient leak test .................................................................................................................... 49
Annex
Parts catalog
Technical Information
All rights reserved. Copyright reserved.
K5330500IECIVZ.fm 10.11.05
II
General
1
2
Fabius GSGeneral
1NotesThis Technical Documentation conforms to the IEC 60601-1 standard.
Read each step in every procedure thoroughly before beginning any test.
Always use the proper tools and specified test equipment. If you deviate from
the instructions and/or recommendations in this Technical Documentation,
the equipment may operate improperly or unsafely, or the equipment could
be damaged.
Use only original Dräger parts and supplies.
The maintenance procedures described in this Technical Documentation may
be performed by qualified service personnel only. These maintenance
procedures do not replace inspections and servicing by the manufacturer.
The information in this manual is confidential and may not be disclosed to
third parties without the prior written consent of the manufacturer.
Strictly follow the Instructions for Use manual / Operating
Instructions! This Technical Documentation does not replace the
Instructions for Use manual / Operating Instructions. Any use of the
product requires full understanding and strict observation of the
product-specific Instructions for Use manual/ Operating Instructions.
Reference is hereby made to the observance of the relevant safety
provisions, for example in Germany, the Medical Product Law (MPG), the
Medical Device Operator Ordinance (MPBetreibV), the Pressure Container
Ordinance (Druckbehälterverordnung), the Technical Rules for Pressurized
Gases (Technische Regeln Druckgase), or the Occupational Health and
Safety Provisions (Unfallverhütungsvorschriften).
Unless otherwise stated, reference is made to laws, regulations or
standards (as amended) applicable in the Federal Republic of Germany.
Follow the laws and regulations applicable in your country.
Observe protection mark DIN 34. Copyright reserved.
Version 2.0_ Released_Printed on_10.11.05_K5330500_General.fm
5330.500
3
GeneralFabius GS
1.1Symbols and
Definitions
This symbol indicates a warning.
This symbol indicates tips and useful information.
This symbol is used to alert against unsafe practices when handling
electrostatic sensitive devices (ESD).
Definitions according to German standard DIN 31051:
Inspection= examination of actual condition
Maintenance= measures to maintain specified condition
Repair= measures to restore specified condition
Servicing= inspection, maintenance, and repair
Version 2.0_ Released_Printed on_10.11.05_K5330500_General.fm
Observe protection mark DIN 34. Copyright reserved.
4
5330.500
Function Description
5
6
Fabius GSFunction description
1General Information
about the Fabius GS
The Fabius GS comprises the following assemblies:
–Display and Control Panel
–Flowmeter assembly
–Gas box: Gas Inlet Assembly and related items
–Breathing system
–Pneumatic assembly
–Ventilator
–Vaporizers
–Trolley
Monitoring, electrical connections and gas connections as shown in Figure 1,
Figure 2, Figure 3, and Figure 4.
Figure 1 Front view of Fabius GS anesthesia system, for legend see Table
1
All rights reserved. Copyright reserved.
Version 1.0_ Released_Printed on_10.11.05_F5330500_Function_Description.fm
5330.500
7
Function descriptionFabius GS
Table 1Legend to Figure 1
No.Name
1Trolley
2O2 flush
3Cosy 2 breathing system
4Ventilator
5Oxygen flowmeter (auxiliary)
6Display
7Control panel
8Vaporizers
9Cylinder Pressure Gauges
10Pipeline Pressure Gauges
11Flow Control Valves
12Total fresh gas flowmeter
Version 1.0_ Released_Printed on_10.11.05_F5330500_Function_Description.fm
All rights reserved. Copyright reserved.
8
5330.500
Fabius GSFunction description
Figure 2Rear view showing interface panel and power entry, for legend
see Ta b l e 2
Table 2Legend to Figure 2
No.Name
1Serial communication ports (only one is shown)
2Tube connection for PEEP valve
3Tube connection for APL bypass valve
4O2 sensor connection
5Airway pressure connection
6Spirolog sensor connection
7Power entry
8Cover for power fuses (2x 2.5 A)
All rights reserved. Copyright reserved.
Version 1.0_ Released_Printed on_10.11.05_F5330500_Function_Description.fm
5330.500
9
Function descriptionFabius GS
No.Name
9ON/OFF switch
10Battery fuse
10
Figure 3 Rear view showing gas pipeline and PIN index cylinder connec-
tions, for legend see Ta b l e 3
Table 3Legend to Figure 3
No.Name
1Pipeline tube connections
2N2O or AIR PIN index cylinder connections
3O2 PIN index cylinder connection
4O2 PIN index cylinder connection
5330.500
Version 1.0_ Released_Printed on_10.11.05_F5330500_Function_Description.fm
All rights reserved. Copyright reserved.
Fabius GSFunction description
Figure 4Rear view showing gas pipeline and cylinder connections (units
without PIN index cylinder connections), for legend see Table 4
Table 4Legend to Figure 4
No.Name
1Pipeline tube connections
2N2O cylinder connection
3O2 cylinder connection
All rights reserved. Copyright reserved.
Version 1.0_ Released_Printed on_10.11.05_F5330500_Function_Description.fm
5330.500
11
Function descriptionFabius GS
2Function diagram of
Fabius GS
Figure 5Function diagram of Fabius GS - Cosy 2 breathing system
12
5330.500
Version 1.0_ Released_Printed on_10.11.05_F5330500_Function_Description.fm
All rights reserved. Copyright reserved.
Fabius GSFunction description
Figure 6Function diagram of Fabius GS - Cosy 2.5 (2.6) breathing system
All rights reserved. Copyright reserved.
Version 1.0_ Released_Printed on_10.11.05_F5330500_Function_Description.fm
5330.500
13
Function descriptionFabius GS
3Battery backupFabius GS backup power is provided by two series-connected 12 V recharge-
able batteries. These batteries remain on charge as long as the machine is
plugged into an active AC outlet. Should power supply fail while the machine
is in operation, the batteries will allow the machine to continue operating for a
minimum of 45 minutes, provided that the batteries are fully charged.
The batteries are accessible by opening the ventilator compartment. The
3.15A battery fuse is located at the back of the control box.
4Fabius GS piping
diagram
Figure 7Battery backup arrangement
Figure 8Fabius GS piping diagram
14
5330.500
Version 1.0_ Released_Printed on_10.11.05_F5330500_Function_Description.fm
All rights reserved. Copyright reserved.
Fabius GSFunction description
5Function descrip-
tion of gas box
The supply gases flow through the filters and non-return valves in the gas
inlet assembly. Pipeline supply pressures are indicated on pipeline pressure
gauges located on the flowmeter assembly. Cylinder pressure gauges are
located on the trolley assembly. The pressures of O2 and N2O delivered to
the flowmeter assembly are set by regulators on the gas inlet assembly.
Should the O2 supply fail or if its pressure decrease below a certain limit, the
O2 low alarm switch signals an alarm.
Figure 9 Gas box function diagram, part 1
If the O2 flush button is pressed, oxygen is delivered to the fresh-gas outlet.
The fresh-gas ejector prevents the fresh gas from flowing back into the anesthetic vaporizer. This avoids an increase in the anesthetic gas concentration.
All rights reserved. Copyright reserved.
Version 1.0_ Released_Printed on_10.11.05_F5330500_Function_Description.fm
5330.500
15
Function descriptionFabius GS
6SORC (Sensitive
Oxygen Ratio Controller)
Figure 10 Gas box function diagram, part 2
The SORC is a control element that functions like an N2O shut-off device and
ensures a vital O2 concentration in the fresh gas. In the event of an O2 shortage, the SORC limits the N2O flow such that the O2 concentration in the
fresh gas does not decrease below 21 vol.%.
If the O2 flow control valve is closed or if the O2 flow is lower than or equal to
200 mL/min, the SORC interrupts the N2O flow.
N2O can be added as of an O2 flow of approx. 300 mL/min. In this case, the
SORC also prevents O2 concentrations below 21 vol.%.
The SORC bypass allows O2 to bypass the restrictor in the SORC when O2
flows above 10 L/min are needed.
16
5330.500
Version 1.0_ Released_Printed on_10.11.05_F5330500_Function_Description.fm
All rights reserved. Copyright reserved.
Fabius GSFunction description
Figure 11 SORC function diagram, part 1
The O2 and N2O flows are adjusted with the flow control valves.
Restrictors located at the outlets of the SORC generate back-pressures.
These back-pressures exert a force on the control diaphragms of the SORC.
The O2 back-pressure opens the SORC. The N2O back-pressure closes the
SORC. The pressure ratio at the control diaphragm affects the N2O flow.
The resistors and the spring force are dimensioned such that a minimum concentration of 21 vol.% of O2 is always ensured. The maximum O2 flow is
approx. 12 L/min.
Figure 12 SORC function diagram, part 2, for legend see Tab le 5
All rights reserved. Copyright reserved.
Version 1.0_ Released_Printed on_10.11.05_F5330500_Function_Description.fm
5330.500
17
Function descriptionFabius GS
Table 5Legend to Figure 12
No.Name
1Control diaphragms
2Restrictors
3N2O non-return valve
4Operating-point adjusting screw
5Flow control valves
7Cosy 2 breathing
system
The Cosy 2 breathing system allows three modes of patient ventilation:
–Manual ventilation and spontaneous breathing
–Volume controlled ventilation
–Pressure controlled ventilation
The APL valve (adjustable pressure limiting valve), lever type, has a selector
switch which can be used to toggle between “MAN” and “SPONT”.
On APL valves with control knob, switching from “IPPV/SPONT” to “MAN” is
carried out by turning the knob.
In the “MAN” position, the breathing system is closed to atmosphere. This
position is used for manual ventilation of the patient. The APL valve opening
pressure can be adjusted from 5 to 70 cmH2O (mbar).
In the “SPONT” position the APL valve is open to atmosphere. This position
is used for spontaneous breathing.
Using the control box and the PEEP Pmax valve, the pressure limit (Pmax)
can also be adjusted during volume control from 15 cmH2O (mbar) to
70 cmH2O (mbar) via the membrane keypad.
18
5330.500
Version 1.0_ Released_Printed on_10.11.05_F5330500_Function_Description.fm
All rights reserved. Copyright reserved.
Fabius GSFunction description
Figure 13 Cosy 2 breathing system, for legend see Tabl e 6
Table 6Legend to Figure 13
No.Name
1Expiratory connection
2Flow sensor (Spirolog) (not shown)
3PEEP/Pmax valve
4MAN/SPONT APL valve
5APL Bypass valve
6Fresh-gas decoupling valve
7Fresh-gas port
8Ventilator port
9Anesthetic gas scavenging port
10Absorber
All rights reserved. Copyright reserved.
Version 1.0_ Released_Printed on_10.11.05_F5330500_Function_Description.fm
5330.500
19
Loading...
+ 53 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.