Dräger Pac 3000 Instruction Manual

D
Pac
®
3000 CO, H
2
S, O
2
Instructions for Use
ϝΎϤόΘγϻ΍ ΔϘϳήσ
1 For Your Safety
The use of the Dräger Pac 3000 instruments assumes a complete
knowledge and adherence to the users manual.
— The Dräger Pac 3000 may only be used in areas subject to
explosion hazards which are explicitly covered under the Ex Approvals which have been given to the Dräger Pac 3000.
— Dräger Pac 3000 is not for use in oxygen-enriched
atmospheres. — Please check calibration before safety relevant use. — The performing of calibration and bump testing shall be
conducted according to local regulations. — The gas opening is equipped with a dust and water filter.
This filter protects the sensor against dust and water.
Do not destroy the filter. Replace destroyed or clogged filter
immediately. — Dräger Safety cannot be held responsible for damages incurred
when the above guidelines are not explicitly adhered to. The
warrenty and liability provisions of the terms of sale and delivery
of Dräger Safety are not modified by the recommendations given
above.
2 Intended Use
— Personal gas alarm in the workplace.
3 What is what
4 Standard configuration
1)
1) Please be aware of special settings by customer requirements.
1 Alarm LED 6 Gas Opening 2 Horn 7 Screw 3 Remaining Display 8 Clip 4 OK Key On/Off/Alarm Acknowledge 9 Label 5 + Key Off/Bump Test 10 IR interface
CO H2S O
2
Measuring Range 0 to 500
ppm
0 to 100
ppm
0 to 25
vol.-%
Vibrating alarm Yes Yes Yes
1
6
5
2
3
9
4
7
7
1
7
8
10
7
9
St-5056/5053-2004.eps
2) For O2 A1 is the lower alarm threshold, used to indicate oxygen deficiency
5 Operation
5.1 Remaining life of the instrument
— Once activated check the remaining life by pressing
[+]
while instrument is turned off. The remainig time in days will be shown. After another press "d" will be shown. After another press the gas to be measured will be shown, e. g. "750", "d", "CO".
5.2 Activating a new instrument
— Press and hold
[+]
for approx. 3 seconds while "3, 2, 1" appears in the display. The instrument’s usable life is now started. The gas to be measured will be shown. After 10 seconds the display will turn off or if
[+]
is pressed again, "d" will be shown. After 10 seconds the display will turn off or after another press the remaining time in days will be shown. After 10 seconds the display will turn off.
5.3 Turning the instrument on
— Press and hold
[OK]
. The display counts down until startup:
"3, 2, 1".
— All display segments are lit. Next, the LED, Alarm and Vibrating
alarm are activated in sequence. Please check these before each use.
— The instrument will perform a self test. — The software version and the gas name are displayed. — The number of days of remaining operation are shown,
e. g. "750", "d".
— The A1 and A2 alarm limits are displayed. —For the O
2
sensor: after the first activation of the instrument, a sensor warm up time of up to approx. 15 minutes is needed. The gas name flashes until the warm up time has passed.
5.4 Before entering a working place
— After turning the instrument on, the gas name to be measured
will normally be shown in the display.
— Check for the notice icon
[!]
. When lit, it is recommended that
you perform a bump test as described in section 5.5.
CO H2S O
2
Alarm Threshold A1 30 ppm 10 ppm 19 vol.-%
2)
acknowledgeable Yes Yes No latching No No Yes Alarm Threshold A2 60 ppm 20 ppm 23 vol.-% acknowledgeable No No No latching Yes Yes Yes Display Gas Name Gas Name Gas Name Bump Test Mode off off off Life Signal off off off Turning the instru-
ment off
always
allowed
always
allowed
always
allowed
— Clip the instrument to clothing before working in or near
potential gas hazards.
Insure that the gas opening is not covered and that the instrument is also near to your breathing area.
5.5 Performing a “bump test” with gas
— Prepare a Dräger test gas cylinder with 0.5 l/min and a gas
concentration higher than the alarm threshold to be tested.
— Connect Dräger Pac 3000 and the test gas cylinder to the
calibration adapter or to the Dräger Bump Test Station.
— To enter the bump test mode press the
[+]
3 times within 3 seconds. The instrument beeps twice, quickly. The Display begins to flash slowly.
— Open the regulator valve to let test gas flow over the sensor. — If gas concentration exceeds the alarm thresholds A 1 or A 2 the
corresponding alarm will occur.
— To finish the bump test press
[OK]
, the
[!]
icon is removed from
the display and the instrument returns to the measuring mode.
— If during the bump test no alarm occurs within 1 minute, the
instrument alarm mode is entered to indicate failure. The error icon
[X]
and the notice icon
[!]
are flashing; error code 240 is
shown upon acknowledgement.
"– – –"
is shown instead of the
gas name, and the
[X]
and
[!]
icon are lit. In this case the bump
test can be repeated or the instrument can be calibrated.
— The bump test can also be made by an automatic function. This
function can be activated using the PC software Pac Vision or CC Vision (see section 5.8).
— If the bump test mode was entered by mistake, while the notice
icon
[!]
was not lit and no gas flow over the sensor, press
[+]
to cancel the bump test mode and switch over to measurement screen.
5.6 During operation
— If the allowable measurement range is exceeded or a negative
drift occurs, the following will appear in the display:
“”
(too high concentration) or
“”
(negative drift). — Alarms are indicated as described in section 6. — Continuous function of the instrument is indicated by the life
signal, which is a beep every 60 seconds, if configured (see section 4).
5.7 Turning the instrument off
— Simultaneously hold both keys for approximately 2 seconds until
“3” appears in the display. Continue to hold both keys until the countdown is finished. The alarm and LED will be activated momentarily.
5.8 Calibration and configuration
— For calibration or individual configuration connect Dräger
Pac 3000 to a PC using the connecting cradle or the E-Cal System. Calibration and configuration can be done with installed software Pac Vision or CC Vision. A calibration “due date” can be set using the operation timer (in days). Strictly follow the instructions for use of the modules and software in use.
Loading...
+ 1 hidden pages