Dräger Infinity Series, Infinity Delta, Infinity Kappa, Infinity Delta XL Quick Reference Manual

Quick
Reference
Guide
Infinity Series Monitor
Delta/Delta XL/Kappa
Lead Placement 2 - 5 Getting Started, Connecting your Patient 6 Admit 6 Understanding Fixed Keys 7 Alarms
Types of Alarms 8 Adjust Arrhythmia Alarms 8
ECG Monitoring
Show All Leads 9 Change Displayed Leads 9
Activate Pacemaker Detection 9 Non-Invasive Blood Pressure (NBP) Monitoring 10 Respiratory Monitoring
Change Waveform Size 10
Change Displayed Lead 10 SpO2Monitoring 10 Temperature Monitoring 11 Fast Access Data Key
Trends 11
Review Stored Events 11 Alarm Limits Fixed Key 12 Recordings 12 Print Functions 12 Menu Functions
Patient Category 13
Standby 13
Privacy Screen 13 Transferring Patient 14 Troubleshooting 16
Infinity Delta/Delta XL/Kappa - Quick Reference Guide Intended for quick reference only. Refer to Instruction For Use Manual.
INFINITY DELTA/DELTA XL/KAPPA - QUICK REFERENCE GUIDE
| 01
CONTENT
© 2010 Draeger Medical, Inc.
Electrode & Lead Placement
DO’S & DO NOT’S
Do Place the electrode on a clean and dry area of skin.
Utilize soap, water and dry.
Do Remove hair if necessary for proper electrode
placement.
Do Change electrodes Q 24-48 hours, or per hospital
policy.
Do Not Utilize alcohol to dry and clean area of skin
for electrode placement.
Do Not Place electrodes over bony prominences or
area of increased muscle movement/activity.
Do Not Assume electrodes are “Good” if the tape
portion is attached to the patient’s skin.
Do Not Braid, tie or knot leads.
Infinity Delta/Delta XL/Kappa - Quick Reference Guide Intended for quick reference only. Refer to Instruction For Use Manual.
INFINITY DELTA/DELTA XL/KAPPA - QUICK REFERENCE GUIDE
© 2010 Draeger Medical, Inc.
| 02
5 Lead Placement
Infinity Delta/Delta XL/Kappa - Quick Reference Guide Intended for quick reference only. Refer to Instruction For Use Manual.
INFINITY DELTA/DELTA XL/KAPPA - QUICK REFERENCE GUIDE | 03
© 2010 Draeger Medical, Inc.
White - Right Arm
Green - Right Leg
Black - Left Arm
Red - Left Leg
Brown - V2 - Left of sternum 4th intercostal space
RA
RL
LA
LL
V2
D-22640-2010
6 Lead Placement
Infinity Delta/Delta XL/Kappa - Quick Reference Guide Intended for quick reference only. Refer to Instruction For Use Manual.
INFINITY DELTA/DELTA XL/KAPPA - QUICK REFERENCE GUIDE | 04
© 2010 Draeger Medical, Inc.
White - Right Arm
Green - Right Leg
Black - Left Arm
Red - Left Leg
Brown - V2 - Left of sternum 4th intercostal space
Grey - V5 - Left at 5th Intercostal space Anterior Axillary line
RA
RL
LA
LL
V2
V5
D-22640-2010
Infinity Delta/Delta XL/Kappa - Quick Reference Guide Intended for quick reference only. Refer to Instruction For Use Manual.
INFINITY DELTA/DELTA XL/KAPPA - QUICK REFERENCE GUIDE | 05
© 2010 Draeger Medical, Inc.
Pacemaker Signal Interfering with Monitor ECG QRS Detection
Pacemaker interference is the # 1 cause of false ASY and other arrhythmia calls.
PATIENT WITH PACEMAKER
Try the following lead configurations:
Five Lead Paced Patients
Six Lead Paced Patients
D-22639-2010
D-22641-2010
Getting Started - Rotary Knob
The rotary trim knob at the bottom right corner of the monitor turns left and right. It highlights selections and requires pressing in to confirm the selections (click).
The monitor does not need to be powered off between patients. Remember to “Discharge” between patients once all documentation has been completed. This erases all trended information and returns the monitor to default settings.
CONNECTING YOUR PATIENT TO THE MONITOR
- Prep the patient’s skin, place ECG electrodes
appropriately and connect leads.
- Place SpO
2
sensor on the patient.
- Place NIBP cuff on the patient.
- Press Main Screen Key to begin active patient
monitoring.
TO ADMIT AT THE BEDSIDE USING THE ROTARY KNOB
Admission via the Infinity Central Station is easiest, however, you may admit from the bedside monitor:
- Press Menu fixed key.
- Highlight Admit/Discharge, “click”.
- Highlight Admit and enter the patient’s data using the
Rotary Knob.
Infinity Delta/Delta XL/Kappa - Quick Reference Guide Intended for quick reference only. Refer to Instruction For Use Manual.
INFINITY DELTA/DELTA XL/KAPPA - QUICK REFERENCE GUIDE | 06
© 2010 Draeger Medical, Inc.
Understanding Fixed Keys
Alarm Silence Silences the triggered bedside alarm
for 60 seconds Alarm Limits Allows limits to be set on all parameters All Alarms Off Initiates 1-5 minutes suspension of all
alarms and arrhythmia processing.
Alarms remain suspended until you
press the All Alarms Off fixed key again
or the timeout period expires. Code Starts on screen timer Discharge Quick Discharge key, press twice Record Initiates timed recordings to recorder Print Screen Initiates screen copy to laser printer NBP Start/Stop Starts single NBP measurement, starts
interval mode, stops measurement already in progress.
Press twice quickly to suspend interval
Fast Access Fast Access to frequently used
functions
Mark Stores all waveforms on screen to event
recall and places a marked set of vital
signs into the trends Main Screen Returns to main monitor screen Menu Access all monitor functions
Infinity Delta/Delta XL/Kappa - Quick Reference Guide Intended for quick reference only. Refer to Instruction For Use Manual.
INFINITY DELTA/DELTA XL/KAPPA - QUICK REFERENCE GUIDE | 07
© 2010 Draeger Medical, Inc.
Alarms
3 DIFFERENT TYPES OF ALARMS
L-T (Life Threatening): alarm flashes red and sounds two alternating tones. L-T alarm must be manually silenced by pressing the Alarm Silence Fixed Key.
SER (Serious): alarm flashes yellow and sounds the same tone twice. Alarm can be manually silenced using the Alarm Silence Fixed Key or the alarm will automatically reset if the violation ceases.
ADV (Advisory): alarm flashes white and sounds a single tone. Alarm can be manually silenced using the Alarm Silence Fixed Key or the alarm will automatically reset if the violation ceases.
TO ADJUST ARRHYTHMIA ALARMS/SETTINGS:
- Highlight ARR parameter box, click.
- Customize type of alarm, rate, count and archive function as desired.
ARR may also be accessed from the Alarm Limits
Fixed Key by choosing ARR.
Infinity Delta/Delta XL/Kappa - Quick Reference Guide Intended for quick reference only. Refer to Instruction For Use Manual.
INFINITY DELTA/DELTA XL/KAPPA - QUICK REFERENCE GUIDE | 08
© 2010 Draeger Medical, Inc.
1
2
3
Accessing Parameter Functions:
ECG MONITORING
To Display All Monitored Leads
- Highlight HR parameter box, click.
- Select Show All Leads, click.
HINT: For best results choose waveforms that are all positive or all negative and greater than 0.25 mv tall.
To Change Displayed Leads
- Highlight HR parameter box, click.
- Highlight Lead Setup, click.
- Highlight Channel 1, click.
- Highlight desired lead, click.
- Repeat above to change other channels.
To Activate Pacemaker Detection
- Highlight HR parameter box, click.
- Highlight ECG Options, click.
- Highlight Pacer Detection On, click.
NOTE: Pacer Detection is not active when Electrosurgical Unit (ESU) filter suppression is used.
Infinity Delta/Delta XL/Kappa - Quick Reference Guide Intended for quick reference only. Refer to Instruction For Use Manual.
INFINITY DELTA/DELTA XL/KAPPA - QUICK REFERENCE GUIDE | 09
© 2010 Draeger Medical, Inc.