Dräger Evita 4 Instructions For Use Manual

Evita 4
Intensive Care Ventilator
Instructions for Use
D
MEDICAL
Software 4.n
1-159-99
Working with these Instructions for Use
Header line – the title... of the main chapter
The title of the specific sub-section is printed underneath the main header – to help you find your way quickly from subject to subject.
Page body... the Instructions for Use
in combined text/illustrations. The information is expressed in the form of practical actions, giving the user direct hands-on experience in learning how to use the machine.
Left-hand column – the text...
provides explanations and instructs the user step-by-step in the practical use of the product, with short, clear instructions in easy-to-follow sequence. Bullet points indicate separate actions. Where several actions are described, numbers are used both to refer to the relevant details in the illustrations and to specify the sequence of actions.
Operation Calibrating
Calibrating the CO2 sensor
The CO2 sensor must be calibrated: – if the check values are not met on checking
calibration with filter or calibration gas.
– as part of the half-yearly inspection of Evita 4.
Switch on Evita 4. Wait about for 3 minutes for
the machine to complete its warm-up phase.
Press the »Calibration« key.
Display (example):
Touch »CO2« screen key.
Display (example):
Carry out CO2 zero calibration, page 74.
After the CO2 zero calibration: Touch the »Cal.« screen key.
Right-hand column – the illustrations...
provide the visual reference for the text and make it easier to locate the various parts of the equipment. Elements mentioned in the text are highlighted. Unnecessary details are avoided. Screen displays prompt the user to proceed and confirm correct actions.
Connect the calibration gas supply.
Use the cuvette from the calibration set!
1 Connect the calibration gas cylinder and the
cuvette of the calibration set to the hose.
2 Remove the CO2 sensor from its park bracket
and fit it to the cuvette of the calibration set.
Read the CO2, O2 and N2O concentrations
(vol.%) of the calibration gas from the test cylinder.
2
1
3
2
What's new in Evita 4 software 4.n*
What's new in Evita 4 software 4.n
Specification of the humidifier used
– »Active humidifier«
or – »HME/Filter« (artificial nose) – for more accurate measurement of the volume
parameters
Apnoea ventilation On/Off
– can be selected as starting configuration
Extended range of settings for the alarm time TApnoea >>
– from 5 to 60 seconds
Frequency can be reduced to 0
– for BIPAP and SIMV, for weaning without transitions
Ventilation mode BIPAPAssist
– for pressure-controlled assisted ventilation
>>
(formerly 15 to 60 seconds)
Additional weaning parameters
available as software version 4.n plus upgrade in addition to the parameter occlusion pressure P 0.1 Evita 4.4n also determines the parameters – RSB Rapid Shallow Breathing index and – NIF Negative Inspiratory Force index
External flow source available as software version 4.n plus upgrade – The amount of external flow is calculated by Evita 4.4n
(e.g. for additional tracheal gas insufflation) and adjusts the volume monitoring tolerances in order to avoid inadvertent alarms
Extended use of loop presentations
available as software version 4.n plus upgrade – Loops can be zoomed and frozen – Loops can be displayed permanently in the upper part
of the screen
Patient mode »prev. patient« can be selected
– to adopt the settings, including alarms, which were
effective before switching off the equipment
Leakage compensation On/Off
– for activation and deactivation of the automatic
leakage compensation function
Extended logbook entries
– Evita 4.4n identifies alarms which are active but not
displayed with an asterisk
Monitoring of tube blockages
– New alarm message »Tube blocked !!!«
Evita Remote (Remote Pad) optionally available – Remote control pad for parallel remote operation of
function keys on Evita 4
NIV
optionally available – Application mode to support non-invasive ventilation
therapies
Nurse call
optionally available – Connection for transmitting alarm signals to a central
hospital alarm station
* See pages 187 and 188 for new features in software
versions 2.n and 3.n
3
4
Contents
Contents
For Your Safety and that of Your Patients ..............................................11
Intended Medical Application.................................................................. 13
Operating Concept...................................................................................15
Structure of the Control Unit................................................................... 16
On-Screen Controls................................................................................. 17
Screen Keys for Function Selection
without Confirmation..................................................................................18
Screen keys for Function Selection,
Adjustment and Confirmation.....................................................................18
On-Screen Parameter Setting Knobs.........................................................20
Screen Pages...........................................................................................21
Standard page...........................................................................................22
»Adjustment« Screen Page........................................................................ 22
»Alarm Limits« Screen Page...................................................................... 24
»Measured values« Screen Page............................................................... 25
»Measurement Manoeuvre« Screen Page.................................................. 26
»Calibration« Screen Page.........................................................................26
»Configuration« Screen Page.....................................................................27
Positioning the Control Unit.....................................................................27
Ergonomic Positioning...............................................................................27
Preparation...............................................................................................29
Attaching components.............................................................................30
Fitting expiration valve................................................................................30
Fitting flow sensor..................................................................................... 30
Fitting O2 sensor capsule.......................................................................... 31
Note on Use of Heat and Moisture Exchanger (HME)............................. 31
Ventilation Adults and Children............................................................... 32
Connecting Aquapor humidifier..................................................................32
Connecting ventilation hoses..................................................................... 32
Fitting temperature sensor.........................................................................33
Fitting CO2 cuvette and CO2 sensor......................................................... 34
Ventilating Infants.................................................................................... 34
Fitting bacterial filter.................................................................................. 34
Fitting humidifier and ventilation hose.........................................................35
5
Contents
Contents
If using bacterial filters............................................................................ 35
Supply and Connections..........................................................................36
Electrical power supply..............................................................................36
Note on use of a socket strip for ancillary equipment..................................36
Temporary interruption of power supply..................................................... 36
Gas supply................................................................................................37
Evita Remote (optional).............................................................................38
Connection................................................................................................38
Note automatic self-test............................................................................. 39
Nurse call (optional)..................................................................................40
Technical Data...........................................................................................40
Before Using for the First Time............................................................... 41
Selecting the language of the display texts.................................................41
Device Check...........................................................................................42
Before use on patient................................................................................ 42
Performing device check........................................................................... 43
Checking the hose system for leaks...........................................................46
Positioning the control unit......................................................................47
To position the control unit on the wall rail..................................................47
To position the control unit on the device................................................... 47
Operation................................................................................................. 49
Starting up............................................................................................... 50
Switching on............................................................................................. 50
Patient mode............................................................................................50
Selecting the patient mode........................................................................ 51
Entering the ideal body weight...................................................................51
Select the previous settings.......................................................................51
Starting ventilation..................................................................................... 52
Setting Ventilation Modes........................................................................53
IPPV..........................................................................................................53
SIMV, SIMV/ASB......................................................................................57
BIPAP, BIPAP/ASB.................................................................................. 60
BIPAPAssist................................................................................................62
CPAP, CPAP/ASB....................................................................................63
MMV, MMV/ASB...................................................................................... 65
APRV........................................................................................................ 67
Independent Lung Ventilation ILV............................................................69
Preparation................................................................................................69
Setting the master and slave device...........................................................71
6
Contents
Contents
Apnoea ventilation................................................................................... 76
Setting Alarm Limits.................................................................................77
In the Event of an Alarm.......................................................................... 78
Cancel alarm tone..................................................................................... 79
Information J............................................................................................ 79
Displaying Curves and Measured Values................................................ 80
Displaying measured values.......................................................................81
Trends.......................................................................................................82
Loops........................................................................................................83
Reference curve displays...........................................................................83
Single stroke displays................................................................................83
Zoom loops (optional)................................................................................83
Display loops in the upper graphic area..................................................... 84
Logbook....................................................................................................85
Screen freeze............................................................................................86
Special Functions.....................................................................................87
Manual inspiration......................................................................................87
Exspiration Hold........................................................................................ 87
Medicament nebulisation........................................................................... 88
Oxygen enrichment for bronchial suction................................................... 91
Special measurement procedure: intrinsic PEEP........................................93
Special measurement procedure: occlusion pressure P 0.1....................... 94
Shut-down.................................................................................................95
Selecting Standby Mode..........................................................................96
Quitting Standby Mode..............................................................................96
Calibration................................................................................................97
Calibrating O2 sensor................................................................................97
Calibrating flow sensor..............................................................................98
External flow source.................................................................................. 99
Checking/calibrating CO2 sensor............................................................100
CO2 zero checking................................................................................. 100
Testing CO2 calibration with test filter..................................................... 102
Testing CO2 calibration with test gas...................................................... 103
Calibrating CO2 sensor...........................................................................105
Resetting CO2 calibration........................................................................106
Configuration......................................................................................... 107
Sound.....................................................................................................108
Adjusting the volume of the alarm tone.....................................................108
7
Contents
Contents
Screen....................................................................................................109
Selecting displayed measured values.......................................................109
Selecting displayed curves...................................................................... 111
Selecting displayed trends.......................................................................112
Ventilation.............................................................................................. 113
Selecting ventilation modes..................................................................... 113
Selecting Pmax pressure limit..................................................................115
Selecting AutoFlow as start-up ventilation mode....................................116
Apnoea ventilation On/Off....................................................................... 117
Selecting patient mode............................................................................118
Start-up values for ventilation parameters and alarm limits........................119
Setting start-up values for ventilation parameters »VT, f«...........................119
Setting start-up values for ventilation parameters »Pressure, O2, I:E«....... 121
Leakage compensation On/Off................................................................122
Setting start-up values for alarm limits......................................................123
System Defaults.....................................................................................124
Setting external interface......................................................................... 124
Setting time and date.............................................................................. 125
Setting language and units...................................................................... 125
Service diagnosis....................................................................................126
Troubleshooting.....................................................................................127
Preparing............................................................................................... 135
Dismantling............................................................................................136
CO2 sensor (optional)............................................................................. 136
Temperature sensor................................................................................ 136
Medicament nebuliser............................................................................. 137
Ventilation hoses..................................................................................... 137
Flow sensor............................................................................................ 137
Expiration valve........................................................................................138
Humidifier................................................................................................138
Disinfecting/Cleaning............................................................................ 139
Assembling............................................................................................ 142
Fitting expiration valve............................................................................. 142
Before Reusing on Patient.....................................................................143
Maintenance Intervals ...........................................................................143
Clean or replace cooling air filter............................................................. 144
Disposing of batteries and O2 sensors.................................................... 144
Removing/fitting ambient air filter.............................................................144
Correct disposal of apparatus................................................................. 145
8
Contents
Contents
What's what........................................................................................... 147
Control unit............................................................................................. 148
Front connections................................................................................... 149
Back panel..............................................................................................150
Technical Data....................................................................................... 151
Environmental conditions.........................................................................152
Settings.................................................................................................. 152
Performance data....................................................................................153
Measured value displays..........................................................................153
Monitoring...............................................................................................155
Operating data........................................................................................ 156
Machine outputs......................................................................................157
Electromagnetic compatibility (EMC)....................................................... 158
Classification...........................................................................................158
UMDNS-Code........................................................................................ 158
Materials used.........................................................................................158
Description.............................................................................................159
Ventilation Modes.................................................................................. 160
Volume-controlled ventilation with PLV and AutoFlow............................ 160
Classic volume constant mandatory ventilation stroke.............................. 160
Manual pressure limiting with Pmax ........................................................ 160
AutoFlow.............................................................................................. 161
Start-up procedure with AutoFlow.........................................................162
Sigh........................................................................................................ 163
SIMV.......................................................................................................164
ASB........................................................................................................165
BIPAP.....................................................................................................166
BIPAPAssist............................................................................................. 167
APRV......................................................................................................168
MMV.......................................................................................................168
Flow measurement..................................................................................170
Compensation of the effect of hose system compliance...........................170
Conversion according to ambient conditions............................................170
Automatic leakage compensation.............................................................172
Weaning parameters............................................................................... 174
Occlusion pressure P 0.1........................................................................174
Rapid Shallow Breathing RSB.................................................................175
Negative Inspiratory Force NIF................................................................ 175
Intrinsic PEEP......................................................................................... 176
Insp. O2 concentration during medicament nebulisation...........................177
9
Contents
Contents
Abbreviations......................................................................................... 178
Symbols................................................................................................. 181
Bibliography...........................................................................................182
Parts List................................................................................................ 183
Order List............................................................................................... 185
What was new in Evita 4 software 2.n................................................... 187
What was new in Evita 4 software 3.n................................................... 188
Index...................................................................................................... 189
10
For Your Safety and that of Your Patients
For Your Safety and that of Your Patients
Strictly follow the Instructions for Use
Any use of the apparatus requires full understanding and strict observation of these instructions. The apparatus is only to be used for purposes specified here.
Maintenance
The apparatus must be inspected and serviced regularly by trained service personnel at six monthly intervals (and a record kept). Repair and general overhaul of the apparatus may only be carried out by trained service personnel. We recommend that a service contract be obtained with DrägerService and that all repairs also be carried out by them. Only authentic Dräger spare parts may be used for maintenance. Observe chapter "Maintenance Intervals".
Accessories
Do not use accessory parts other than those in the order list.
Liability for proper function or damage
The liability for the proper function of the apparatus is irrevocably transferred to the owner or operator to the extent that the apparatus is serviced or repaired by personnel not employed or authorized by DrägerService or if the apparatus is used in a manner not conforming to its intended use.
Dräger cannot be held responsible for damage caused by non-compliance with the recommendations given above. The warranty and liability provisions of the terms of sale and delivery of Dräger are likewise not modified by the recommendations given above.
Dräger Medical AG & Co. KGaA
Not for use in areas of explosion hazard
This apparatus is neither approved nor certified for use in areas where combustible or explosive gas mixtures are likely to occur.
Safe connection with other electrical equipment
Electrical connections to equipment which is not listed in these Instructions for Use should only be made following consultations with the respective manufacturers or an expert.
11
For Your Safety and that of Your Patients Safe use of the equipment
Safe use of the equipment
This equipment must only be used under the supervision of qualified medical staff, so that help is available immediately if any faults or malfunctions occur.
This equipment must not be used with flammable gases or anaesthetic agents. Danger of fire!
Do not use mobile telephones within 10 metres of ventilators!
Mobile telephones may impair the functioning of electromedical equipment and endanger the patient1).
Appropriate ventilation monitoring The built-in monitoring facilities of Evita 4 ensure
appropriate monitoring of ventilation therapy and there­fore detect any undesirable changes in the following ventilation parameters:
– Airway pressure, Paw – Expiratory minute volume, MV – Inspiratory O2 concentration, FiO2 – Inspiratory breathing gas temperature, T – Expiratory CO2 concentration, etCO2 (optional) – Inspiratory breathing volume, VTI – Apnoea time – Tachypnoea monitoring
Back-up ventilation with an independent manual ventilation device
If a fault is detected in Evita 4 so that its life-support functions are no longer assured, ventilation using an independent ventilation device must be started without delay – if necessary with PEEP and/or increased inspiratory O2 concentration (e.g. with the Dräger Resutator 2000).
Changes in these parameters may be caused by: – Acute changes in the patient's condition
– Incorrect settings and faulty handling – Equipment malfunctions – Failure of power and gas supplies
If a fault occurs in this equipment, separate measuring instruments should be used.
1) Dräger medical equipment meets the requirements for immunity to interference in accordance with the specific product standards and EN 60601-1-2 (IEC 601-1-2). Depending on the type of mobile telephone used and on the application situation, however, field strengths exceeding the values specified in the applicable standards may develop in the immediate vicinity of the mobile telephone and therefore lead to faults and malfunctions.
12
Intended Medical Application
Contents
Intended Medical Application
Contents
Intended Medical Application.................................................................. 14
13
Intended Medical Application
Intended Medical Application
Long-term ventilator for intensive care. For adults, children and neonates. For premature babies with the "NeoFlow" option.
With the following ventilation modes: IPPV Intermittent Positive Pressure Ventilation,
controlled and assisted constant-volume ventilation. With the options: – CPPV (Continuous Positive Pressure Ventilation) – PLV (Pressure Limited Ventilation) – AutoFlow
for automatic regulation of inspiration flow
– IRV (Inversed Ratio Ventilation) SIMV Synchronized Intermittent Mandatory Ventilation,
procedure for weaning patients off the ventilator after they have started spontaneous breathing. With the options: – PLV (Pressure Limited Ventilation) – AutoFlow
for automatic regulation of inspiration flow.
MMV Mandatory Minute Volume Ventilation, spontaneous breathing with automatic adjustment of mandatory ventilation to the patient's minute volume requirement. With the options: – PLV (Pressure Limited Ventilation) – AutoFlow
for automatic regulation of inspiration flow.
BIPAPAssist (Biphasic Positive Airway Pressure Assisted) Pressure-controlled assisted ventilation
APRV Airway Pressure Release Ventilation, Spontaneous breathing on two pressure levels with long time ranges – independently adjustable.
Special modes:
Apnoea Ventilation
For switching over automatically to volume-controlled mandatory ventilation, if breathing stops. If apnoea occurs, Evita 4 emits an alarm after the preset alarm period (TApnoea > ) and starts volume-controlled ventilation.
ILV Independent Lung Ventilation, Separate, differentiated, synchronised ventilation with two Evita units, one for each lung.
Diagnostics: Intrinsic PEEP-measurement
for determining intrinsic PEEP and measuring trapped volume.
Occlusion pressure measurement
for evaluating breathing drive during spontaneous breathing.
SB Spontaneous Breathing, Spontaneous breathing at ambient pressure.
CPAP Continuous Positive Airway Pressure, Spontaneous breathing with positive airway pressure.
ASB Assisted Spontaneous Breathing, pressure-assisted spontaneous breathing.
BIPAP* Biphasic Positive Airway Pressure, Pressure-controlled ventilation combined with free spontaneous breathing during the complete breathing cycle, and adjustable pressure increase to CPAP level.
–––––––––––
* Registered trade mark
With monitoring for:
airway pressure, Paw expiratory minute volume, MV inspiratory O2 concentration, FiO2 inspiratory breathing gas temperature, T expiratory CO2 concentration, etCO2 inspiratory breathing volume, VTI apnoea time tachypnoea monitoring to detect rapid, shallow spontaneous breathing
Automatic gas switch-over. In the event of a gas failure, the change-over to another gas is automatic.
14
Operating Concept
Contents
Operating Concept
Contents
Structure of the Control Unit.................................................................. 16
On-Screen Controls................................................................................ 17
Screen Keys for Function Selection
without Confirmation................................................................................ 18
Screen keys for Function Selection,
Adjustment and Confirmation....................................................................18
On-Screen Parameter Setting Knobs........................................................20
Screen Pages..........................................................................................21
Standard page......................................................................................... 22
»Adjustment« Screen Page....................................................................... 22
»Alarm Limits« Screen Page..................................................................... 24
»Measured values« Screen Page.............................................................. 25
»Measurement Manoeuvre« Screen Page................................................. 26
»Calibration« Screen Page........................................................................26
»Configuration« Screen Page....................................................................27
Positioning the Control Unit....................................................................27
Ergonomic Positioning..............................................................................27
15
Operating Concept Structure of the Control Unit
Structure of the Control Unit
The main components of the control unit are the screen, a set of fixed function keys and the central rotary dial- knob.
The function keys are used to call up the screen pages appropriate to the application.
In addition to curves, measured values and status displays, the screen contains, in a separate field, touch­sensitive keys and touch-sensitive rotary knobs for parameter setting.
The touch-sensitive screen keys and the screen knobs are used in a similar way to ordinary keys and knobs:
D Evita 4
Touching with the fingertip is equivalent to pressing a key or taking hold of a knob.
The display always contains only the screen keys and screen knobs required for function selection and/or adjustment.
Settings and confirmations are made by turning and pressing the central, rotary knob.
The keys for routine functions are placed to the right and left on the outside of the front panel.
Frequently used function keys are placed on the right, e.g.
the key »? « for selecting the standard page
or the »Alarm Reset« key for resetting or confirming messages.
Less frequently used function keys are placed on the left­hand side of the front panel, e.g. the key » « for switching the medicament nebuliser on/off, or the »O2 suction« key for bronchial suctioning.
D Evita 4
16
The power switch for switching the device on/off.
The power switch is located on the back panel and has a pivoting cover to protect against being inadvertently switched off.
Operating Concept
Structure of the Control Unit
On-Screen Controls
On-Screen Controls
The lower half of the screen contains touch-sensitive coloured screen keys and screen knobs.
Touching these controls with the fingertip is equivalent to pressing key or taking hold of a knob.
The colour displays the status of the "control" and "LEDs":
green = usable white = not usable yellow = adjust/confirm black = effective function/display
IPPV
60
40
20
0
-10
O2 [%]
A
Assist
[mbar]
P
aw
0 2 4 6 8 t [s]
21
.500
VT
60
Flow
m
1.7
Tinsp
21
f
5
PEEP
FiO2 21
P
33 >
peak
P
27
plat
MV 5.3 _
Basic settings
Extra settings
IPPV
BIPAP
SIMV
ASB
1 2 3
50
7.0
4.0
17
Operating Concept On-Screen Controls
Screen Keys for Selecting Functions without Confirmation
e.g. for paging through the system on-screen
for changing the menu for switching over displays
example:
Press the »Table« key = select display.
The key goes black to show that the function is active.
IPPV
60
40
20
0
-10
20
10
0
11:03 12:03
A
Assist
[mbar]
P
aw
0246
MV
11:06 total 1.5
spon 0.0
m
40
20
0
11:03 12:03
Table Table
8 t [s]
f
11:06 total 1.5
spon 0.0
FiO2 21
P
33 >
peak
P
27
plat
MV 5.3 _
Zoom in
Zoom out
Table
Trends
Loops
Logbook
1 2
3
50
7.0
4.0
Screen Keys for Function Selection, Adjustment and Confirmation
Display (example):
IPPV
60
40
20
0
-10
MV 5.3 L/min
f
spn
V
Ti
P
aw
A
P
[mbar]
aw
0 2 4 6 8 t [s]
0.0 bpm
.497 L
33 L/min
m
7.0
4.0
_
_
_
_
30
780
50
T
Apnoea
etCO
s
– – mmHg
2
FiO2 21
P
33 >
peak
P
27
plat
MV 5.3 _
15
_
100
_
0
Monitoring
Logbook
Limits
1 2 3
50
7.0
4.0
18
1 Touch the relevant screen key for the alarm limits,
e.g.:
MV 2.3 L/min
9.3
3.1
The colour changes from green to yellow = setting function is set.
Operating Concept
On-Screen Controls
1
MV 2.3 L/min
2 Turn the rotary knob = adjust the alarm limit.
The value is displayed in the screen key.
3 Press the rotary knob = the colour changes from
yellow to green, and the set alarm limit is confirmed and effective.
To cancel the setting:
Touch the screen key again
or
touch another screen key.
2
3
19
Operating Concept On-Screen Controls
Screen Knobs for Setting Parameters
Display (example):
e.g. »PEEP« screen knob. 1 Touch the »PEEP« screen knob:
It changes colour from green to yellow = setting function selected.
2 Turn rotary knob = Adjust setting. The value is
displayed in the knob.
3 Press rotary knob = Confirm. The knob changes
colour from yellow to green, and the setting is validated and takes effect.
IPPV
60
40
20
0
-10
O2 [%]
A
Assist
[mbar]
P
aw
0 2 4 6 8 t [s]
21
.500
VT
60
Flow
m
1.7
Tinsp
21
f
5
PEEP
1
PEEP
2
FiO2 21
P
33 >
peak
P
27
plat
MV 5.3 _
Basic settings
Extra settings
5
PEEP [mbar]
10
1
2 3
IPPV
BIPAP
SIMV
ASB
50
7.0
4.0
While pressure values, such as Pmax, are being set, they are displayed in the Paw (t) curve as a dashed black line.
To cancel the setting:
Press the screen knob again
or
press another screen knob.
PEEP [mbar]
3
20
Screen Pages
Operating Concept
Screen Pages
All the screen pages have the same structure, i.e. their contents are always arranged in the same positions on the screen:
Messages indicating ventilation modes and alarms, displays of measured values and curves, and help functions, always appear in the same position on the screen:
The active ventilation mode/patient mode is
displayed on the left-hand side of the top line. The ventilation mode is indicated by its abbreviation, e.g. BIPAP.
The patient mode is indicated by a symbol: A m for adults P m for paediatric
In the case of spontaneous breathing activity by the patient, a lung symbol is briefly displayed as indicator.
Curves are displayed in the upper left-hand quarter of
the screen.
The lower half of the screen shows curves and
measured values or screen keys and screen knobs – depending which screen page is selected.
Current ventilation mode/patient mode Alarms
Curves Measured
Curves or screen knobs Measured
Help functions
values
values or screen keys
Alarms are displayed on the right of the top line.Measured values are displayed in the upper right-
hand quarter of the screen.
Help functions appear in the bottom line of the
screen. On the right, Evita 4 provides setting instructions. On the left, Evita 4 provides information on the current status – this information can be accessed by pressing key » «
The solid function keys to the right of the screen are used to select the screen pages for the following specific application situations:
Settings Alarm limits Measured values Special measurement procedures
Calibration Configuration
D Evita 4
21
Operating Concept Screen Pages
Screen page
For displaying the ventilation status
Press »? « key.
Display (example):
The standard page shows the ventilation situation at a glance – reduced to the most important measurement parameters and curves.
Four measured values are shown on the right, and two curves on the left.
Other measured values and curves can be selected in the standard page and all subsequent screen pages.
To select other measured value combinations:
1
Touch screen key » « repeatedly.
2 3
To select other curves:
Touch key » «, and touch the screen key
corresponding to the desired curve.
»Settings« screen
For displaying the setting parameters. The bottom right-hand side of the screen contains the
screen keys for selecting the ventilation modes. The screen key displayed in black (IPPV in the example)
represents the currently activated ventilation mode. The bottom left-hand side of the screen contains the on-
screen rotary control knobs. The values of the setting parameters are displayed in the
screen knobs relevant to the ventilation mode. The user-definable start-up settings are marked by an
arrow (j) on the scales of the screen knobs. See "Configuration" on page 107 onwards.
012 37 207 01337 207
Changing the settings of an active ventilation mode
Touch the appropriate screen knob, which will change
colour from green to yellow = setting function enabled.
Turn the rotary knob on the control unit = adjustment
of the value of setting in the screen knob.
Press the rotary knob: the screen knob changes
colour from green to yellow = the setting is confirmed (validated) and active.
22
Selecting another ventilation mode and setting its parameters
Touch the appropriate screen key, e.g. »BIPAP«. The
key changes colour from green to yellow, and the parameter setting page for BIPAP is displayed.
To set the parameters for BIPAP:
Touch the screen knob, which changes colour from
green to yellow = adjustment function selected.
Turn rotary knob = adjust value displayed in screen
knob.
Operating Concept
Screen Pages
Press rotary knob: the screen knob changes colour
from yellow to green = setting validated and effective.
If the indicator "LED" next to a screen knob is illuminated white, the knob setting will only be effective after the new ventilation mode has been switched on (example: »PASB« knob).
If the indicator "LED" is illuminated yellow, the relevant knob setting is already active in the existing ventilation mode (example: »O2« knob).
The start-up values effective on switching on the ventilator are marked on the relevant knob-scale with an arrow (j). Example: PASB = 0 mbar
Press the rotary knob: the screen key changes colour
from yellow to black = the ventilation mode is active.
For detailed instructions on setting the ventilation modes, please refer to page 49.
Cancel selection/setting
014 37 207
PEEP +
10
PASB
Press the screen key or screen knob again.
or
Press another screen key or another screen knob.
To quit a screen page:
1 Press »? « key = return to standard page
or 2 press any of the function keys next to the screen on
the right.
D Evita 4
2
1
23
Operating Concept Screen Pages
»Alarm limits« Screen Page
This page is used for: Displaying the measured values and the corresponding
alarm limits. Setting the alarm limits. Setting the monitoring function. Displaying the logbook. The alarm limits are grouped together in a field and
combined with a curve and four measured values.
Limits, monitoring and logbook are selected by the screen keys on the right of the screen. The currently activated screen key is highlighted in black.
017 37 207
Displaying/Setting Alarm Limits
Touch the »Limits« screen key. The screen key will
change to black. The monitored measured values will be displayed, together with their alarm limits:
Example:
MV 5.4 L/min
5.5
7.5
Left-hand screen key = lower alarm limit. Right hand screen key = upper alarm limit.
Set the alarm limit:
Touch the relevant screen key.
The key changes colour to yellow = adjustable.
Turn the rotary knob = adjust value displayed in the
key.
Press the dial-knob. The screen key changes colour
to green = setting confirmed. The alarm limit is now effective.
For detailed operating instructions, please refer to page 77.
24
»Measured values« Screen Page
This page is used to display: – the measured values in table format – the trend curve – loops – logbook.
Tables, trend, loop and logbook are selected by the right­hand block of screen keys.
Example table of measured values »Table 1«
For detailed operating instructions, please refer to page 81.
Operating Concept
Screen Pages
018 37 207019 37 207
Display Logbook
Touch the »Logbook« screen key.
Turn the dial-knob = select alarm events.
For detailed operating instructions, please refer to page 85.
25
Operating Concept Screen Pages
»Special Procedures« Screen Page
This page is used to display and perform the following special measuring procedures:
– Intrinsic PEEP and – Occlusion pressure P 0.1
The desired special procedure is selected by the appropriate screen key on the right. The result of the last special procedure is displayed.
Example: Intrinsic PEEP:
To start the special procedure:
Touch the »Start« screen key.
For detailed operating instructions, please refer to page 93 and 94.
»Calibration« Screen Page
This page is used for calibrating – the O2 sensor – the Flow sensor – the CO2 sensor
Select the desired sensor with the »O2«, »Flow« or
»CO2« screen keys. Calibration starts as soon as the relevant key is pressed.
020 37 207021 37 207
Evita 4 provides the necessary calibration instructions in the Help Function line at the bottom of the screen.
For detailed operating instructions, please refer to page 97 et seq..
26
»Configuration« Screen Page
For selecting/adjusting the following functions:
Sound
Setting the volume of the alarm tone.
Screen
Selecting the displayed measured values. Selecting the displayed curves. Selecting the displayed trends.
Ventilation
Selecting ventilation modes. Selecting the patient mode. Selecting the initial setting.
System Defaults
Setting the external interface. Setting the time and date. Selecting the language and measurement units. Selecting service diagnosis.
Operating Concept
Screen Pages
Control Unit Location
022 37 207
For detailed operating instructions, see page 107.
Control Unit Location
To adapt to the situation of the ventilation location, the control unit can be placed
either directly on the device or separately, on a wall rail.
For detailed instructions on placing, see page 47.
Ergonomic Positioning
To ensure best viewing, free of reflections.
D Evita 4
1 Hold down the blue segments on the right and left
and
2 at the same time, tilt the control unit to the desired
position.
1
2
1
2
27
28
Preparation
Contents
Preparation
Contents
Attaching components............................................................................30
Fitting expiration valve...............................................................................30
Fitting flow sensor.................................................................................... 30
Fitting O2 sensor capsule.........................................................................31
Note on Use of Heat and Moisture Exchanger (HME)............................31
Ventilation Adults and Children.............................................................. 32
Connecting Aquapor humidifier................................................................ 32
Connecting ventilation hoses.................................................................... 32
Fitting temperature sensor........................................................................33
Fitting CO2 cuvette and CO2 sensor........................................................34
Ventilating Infants................................................................................... 34
Fitting bacterial filter................................................................................. 34
Fitting humidifier and ventilation hose........................................................35
If using bacterial filters........................................................................... 35
Supply and Connections.........................................................................36
Electrical power supply.............................................................................36
Note on use of a socket strip for ancillary equipment.................................36
Temporary interruption of power supply....................................................36
Gas supply...............................................................................................37
Evita Remote (optional)........................................................................... 38
Connection.............................................................................................. 38
Note automatic self-test............................................................................ 39
Nurse call (optional).................................................................................40
Technical Data......................................................................................... 40
Before Using for the First Time.............................................................. 41
Selecting the language of the display texts................................................41
Device Check..........................................................................................42
Before use on patient............................................................................... 42
Performing device check.......................................................................... 43
Checking the hose system for leaks..........................................................46
Positioning the control unit.....................................................................47
To position the control unit on the wall rail.................................................47
To position the control unit on the device.................................................. 47
29
Preparation Attaching components
The following instructions include: – Equipment assembly. – Electrical and gas connections. – Setting the language for the display texts. – Automatic device check with sensor calibration.
Attaching components
Always use properly prepared parts, see Preparing,
page 135.
Fitting the expiration valve
Tilt the control unit upwards.
Push the expiration valve as far as it will go into the
mounting. Check that it is properly engaged by gently pulling the port.
Fitting the flow sensor
1 Push socket to left as far as it will go.
1
2 Fit flow sensor – with the probe facing towards the
ventilator – into the mounting and push it into the socket as far as it will go.
Then:
3 Push flow sensor to the right as far as it will go into
the rubber lip of the expiration valve.
30
3
2
Loading...
+ 162 hidden pages