Dräger DrugCheck 3000 Quick Start Manual

Page 1
Dräger DrugCheck®
3000
Simple. Safe. Sound.
Quick start guide
Page 2
The Dräger DrugCheck® 3000 is a compact and quick drug test based on oral fluid specimens for easy and unambiguous drug detection on the spot. Using the disposable kit, you test people for cocaine, opiates, amphetamine, methamphetamine, and cannabis simultaneously.
This short guide will take you through each step of performing the test.
Page 3
Safety tab
Window with test strip
Test opening
Transport opening
Test cassette with
indicator ring
Swab with colour
indicator
FIGURE A – DRÄGER DRUGCHECK® 3000 TEST-KIT
Page 4
ANALYSING CONTROL AND TEST LINES
No control line No test line Invalid test
No control line Test line appears Invalid test
Control line appears Test line appears Negative test
Control line appears No test line Positive test
Control line (C) Test line for drug Result
Page 5
(1) (2)
Page 6
(1) Area showing control lines (C) (2) Area showing test lines (methamphetamine (MET), amphetamine (AMP),
cannabis (THC), opiates (OPI), cocaine (COC))
FIGURE B – EXAMPLES FOR TEST RESULTS
The test is valid and negative.
The test is valid and positive for opiates.
The test is invalid.
C
MET
THC
C
AMP
OPI
COC
C
MET
THC
C
AMP
OPI
COC
C
MET
THC
C
AMP
OPI
COC
(1)
(2)
Page 7
The test kit (fig. A) consists of a swab for obtaining an oral fluid
specimen and a test cassette for the analysis. The test result is
shown on the test strip (fig. B).
Remove the swab from the test cassette and give it to the test person.
1
Page 8
C
MET
THC
C
AMP
OPI
COC
C
MET
THC
C
AMP
OPI
COC
Page 9
Instruct the test person to move the swab around in their mouth,
underneath the tongue and in the cheek pouches for approx.
15 seconds to gather oral fluid. The person must not suck or chew
on the swab!
2
Page 10
Page 11
Insert the swab back into the test cassette. Press it hard into the
funnel-shaped test opening as far as it will go, in order to break the
vial containing the buffer liquid inside the test cassette. There must
not be a gap between the swab and the test cassette.
3
Page 12
Page 13
Shake the test kit until the red indication ring has lost its color. This
process may take approx. 15 – 30 seconds. Small amounts of color
left on the ring do not influence the measurement.
4
Page 14
Page 15
5
Place the test kit on a level surface or hold it upright in your hand.
Depending on the preferred detection threshold for THC, it is
necessary to select one measurement option and obey the respective
waiting period (pre-incubation).
See the packaging of the Dräger DrugCheck 3000 for detailed
information on the thresholds in ng/ml for all substances.
Page 16
Page 17
After the waiting period, break the safety tab and peel it off downwards
completely (fig. 1). Then press the swab in an upright position into
the test kit as far as it will go (fig. 2). The control and test lines will
appear after approx. 2 to 3 minutes. After appearing the results can be
read. If no red line appears after 5 minutes this indicates a positive
test result. The test results must be evaluated within 10 minutes. If
the control lines do not appear within this period the test is invalid.
6
Page 18
90 71 917 |
©
2016 Drägerwerk AG & Co. KGaA
CORPORATE HEADQUARTERS
Drägerwerk AG & Co. KGaA Moislinger Allee 53–55 23558 Lübeck, Germany www.draeger.com