Dräger drager X-dock 5300, X-dock 6600, X-dock 6300 Instructions For Use Manual

Dräger X-dock 5300 Dräger X-dock 6300/6600
Gebrauchsanweisung
de
4 Instructions for Use
en
16 Notice d’utilisation
fr
27 Instrucciones de uso
es
39 Instruções de uso
pt
51 Istruzioni per l'uso
it
63 Gebruiksaanwijzing
nl
75 Brugsanvisning
da
87 Käyttöohjeet
fi
98 Bruksanvisning
no
109 Bruksanvisning
sv
120 Instrukcja obsługi
pl
131
ru
143 Upute za uporabu
hr
156 Navodilo za uporabo
sl
167 Návod na použitie
sk
179 Návod k použití
cs
191
Инструкция за употреба
bg
203 Instrucţuni de utilizare
ro
215 Használati útmutató
hu
227
Οδηγίες Χρήσης
el
239 Kullanma talimatları
tr
251
使用说明
zh
262
0
X-dock 5300 (X-am)
X-dock 6300/6600
X-dock 5300 (Pac) X-dock 6300/6600
X-dock 5300/6300 X-dock 6600
max. 10
00133286.eps
1 2
3
4
5
8
6 7
13
16
17
1918
15
14
12
11
9 10
13
Zu Ihrer Sicherheit
!
!
i
i

1 Zu Ihrer Sicherheit

1.1 Allgemeine Sicherheitshinweise

Vor Gebrauch des Produkts die Gebrauchsanweisung auf-
merksam lesen.
Gebrauchsanweisung genau beachten. Der Anwender
muss die Anweisungen vollständig verstehen und den An­weisungen genau Folge leisten. Das Produkt darf nur ent­sprechend dem Verwendungszweck verwendet werden.
Gebrauchsanweisung nicht entsorgen. Aufbewahrung und
ordnungsgemäße Verwendung durch die Nutzer sicher­stellen.
Nur entsprechend geschultes und fachkundiges Personal
darf dieses Produkt verwenden.
Lokale und nationale Richtlinien, die dieses Produkt betref-
fen, befolgen.
Nur entsprechend geschultes und fachkundiges Personal
darf das Produkt überprüfen, reparieren und instand hal­ten. Dräger empfiehlt, einen Service-Vertrag mit Dräger ab­zuschließen und alle Instandhaltungsarbeiten durch Dräger durchführen zu lassen.
Ausreichend geschultes Servicepersonal muss das Pro-
dukt entsprechend den Anweisungen in diesem Dokument prüfen und instandhalten.
Für Instandhaltungsarbeiten nur Original-Dräger-Teile und
-Zubehör verwenden. Sonst könnte die korrekte Funktion des Produkts beeinträchtigt werden.
Fehlerhafte oder unvollständige Produkte nicht verwen-
den. Keine Änderungen am Produkt vornehmen.
Dräger bei Fehlern oder Ausfällen vom/von Produkt(teilen)
informieren.

1.2 Bedeutung der Warnzeichen

Die folgenden Warnzeichen werden in diesem Dokument ver­wendet, um die zugehörigen Warntexte zu kennzeichnen und hervorzuheben, die eine erhöhte Aufmerksamkeit seitens des Anwenders erfordern. Die Bedeutungen der Warnzeichen sind wie folgt definiert:
WARNUNG
Hinweis auf eine potenzielle Gefahrensituation. Wenn diese nicht vermieden wird, können Tod oder schwere Verletzungen eintreten.
VORSICHT
Hinweis auf eine potenzielle Gefahrensituation. Wenn diese nicht vermieden wird, können Verletzungen oder Schädigungen am Produkt oder der Umwelt eintreten. Kann auch als Warnung vor unsachgemäßem Ge­brauch verwendet werden.
HINWEIS
Zusätzliche Information zum Einsatz des Produkts.

2 Beschreibung

2.1 Produktübersicht (siehe Ausklapper)

1Master 2 Modul 3 Status-LED 4 Touchscreen-Display 5 Funktionstaste 6 Frischlufteinlass mit Frischluftfilter 7 Diebstahlsicherungs-Slot 8 Stromversorgung 9 USB-Anschlüsse 10 Ethernet-Anschluss 11 Mini-USB-Anschluss 12 Gasauslass 13 Gaseinlässe 14 Drucklufteinlass 15 Typenschild 16 X-am 125 Modul 17 Ladezustand-LED 18 X-am 125+ Modul (mit Ladefunktion) 19 Pac-Modul

2.2 Funktionsbeschreibung

2.2.1 Master

Der Master übernimmt für die Wartungsstation die Ablaufsteu­erung zum Funktionstesten, Kalibrieren, Justieren sowie Funk­tionen zur Benutzerverwaltung, Geräteverwaltung, zum Drucken von Standardberichten und Standardzertifikaten (nur mit PostScript-Druckern) sowie die Schnittstelle zum Nutzer.

2.2.2 Module

In die Module sind die gerätspezifischen Schnittstellen, wie z. B. IR-Kommunikation, Begasungseinheit und Ladekontakt integriert. Zusätzlich beinhalten die Module Sensorik für die Detektion des optischen, akustischen Alarms und des Vibrati­onsalarms der Geräte.

2.3 Verwendungszweck

Dräger X-dock 5300/6300/6600 ist eine modular aufgebaute Wartungsstation. Mit der X-dock können automatisierte Kali­brierungen, Justierungen und Begasungstests von tragbaren Gasmessgeräten parallel und unabhängig voneinander durch­geführt werden. Ein System besteht aus einem Master für 3 (X-dock 5300/6300) oder 6 (X-dock 6600) Prüfgase. Die X-dock 5300 umfasst einen Master inklusive einem Modul und ist nicht erweiterbar. An den Master der X-dock 6300 und 6600 können bis zu 10 Module angeschlossen werden. Die Module erkennen automatisch wann ein Gerät eingelegt wurde und re­geln die Gaszufuhr, so dass jederzeit eine entsprechende Gasversorgung des Geräts gewährleistet ist.
4 Dräger X-dock 5300/6300/6600
Installation
i
i
i
i
i
i
Folgende Gasmessgeräte können mit der X-dock und den ent­sprechenden Modulen verwendet werden:
X-dock 5300/6300/6600
mit Pac-Modul: mit X-am 125 (+) Modul:
Dräger Pac 3500 Dräger Pac 5500 Dräger Pac 7000
Dräger X-am 1700 Dräger X-am 2000 Dräger X-am 2500 Dräger X-am 5000 Dräger X-am 5600

2.4 GPL (General Public License)

Teile der Gerätesoftware nutzen Open-Source Software, die unter GPL, LGPL oder einer anderen Open Source Lizenz ver­öffentlicht wurden. Es handelt sich im Einzelnen um GPL GPLv2, LGPL, MIT, PostgreSQL, Apache, Apache 2, zlib. Die Quelltexte der verwendeten Software können für einen Zeit­raum von mindestens drei Jahren nach Erwerb des die Soft­ware beinhaltenden Produkts auf Datenträger CD unter Angabe der Materialnummer 83 21 874 über Dräger bezogen werden. Die jeweiligen Lizenzbestimmungen der benannten Software liegen auf CD bei.

3 Installation

HINWEIS
Auf ausreichenden Platz für den Gesamtaufbau achten.
Der Master und alle Module müssen die gleiche Firm­ware-Version haben. Wenn dies nicht der Fall ist, muss ein Firmware-Update durchgeführt werden (sie­he Kapitel 6.2 auf Seite 12).
1. Ggf. Module entsprechend der dazugehörigen Montagean­weisung an den Master montieren (nur bei X-dock 6300/
6600).
Maximal 10 Module können an einen Master montiert
werden.
Die verfügbaren Module können beliebig kombiniert
werden.
2. Ggf. Wand- oder Flaschenhalterung entsprechend der da­zugehörigen Montageanweisung montieren.
3. Tüllen von den vorgesehenen Gaseinlässen und vom Gasauslass entfernen.
HINWEIS
Wenn die Tülle vom Gasauslass nicht entfernt wird, kann die Station den Selbsttest nicht fehlerfrei durch­führen.
4. Gaszufuhrschläuche auf die Gaseinlässe des Master ste­cken und mit dem Druckregelventil der Prüfgasflasche ver­binden.
HINWEIS
Dräger empfiehlt, eine Schlauchlänge von 10 m für die Gaszufuhrschläuche nicht zu überschreiten.
5. Ggf. Abgasschlauch (max. 10 m Länge) an Gasauslass anschließen.
6. Druckluft- oder Frischluftzufuhr sicherstellen:
Druckluftschlauch an Druckluftanschluss anschließen
(Ausgangsdruck des Druckregelventils 0,5 bar, Volu­menstrom >3 L/min).
ODER
Ggf. Frischluftschlauch an Frischluftfilter anschließen.
7. Netzteil anschließen.
Station mit bis zu 3 Modulen: Netzteil 24 V / 1,33 AStation mit 4 bis 10 Modulen: Netzteil 24 V / 6,25 A
Das gesamte System wird über den Master mit Strom ver­sorgt.
HINWEIS
i
i
Dräger empfiehlt Dräger-Prüfgasflaschen und Dräger­Druckregelventile (siehe Bestellliste) zu verwenden. Alternativ besteht die Möglichkeit ein passendes Druckregelventil mit 0,5 bar Ausgangsdruck und >3 L/min Volumenstrom zu verwenden.
Dräger empfiehlt einen Abgasschlauch (max. 10 m Länge) an den Gasauslass anzuschließen und mit die­sem das Prüfgas ins Freie zu leiten.

4 Grundlagen

4.1 Station ein- oder ausschalten

HINWEIS
i
i
Wenn für 10 Minuten keine Aktion erfolgt, wechselt die Station automatisch in den Standby-Modus.
Um die Station einzuschalten:
Taste am Master für ca. 1 Sekunde gedrückt halten.
Während des Einschaltprozesses werden folgende Infor­mationen angezeigt:
Software-Versionsnummer
Um die Station auszuschalten:
Taste am Master für ca. 3 Sekunden gedrückt halten.
Die Station schaltet aus.
Standby-Modus:
Der Standby-Modus wird nach ca. 10 Minuten ohne Aktivi-
täten an der Station (Eingabe über Touchscreen-Bildschirm oder Öffnen/Schließen einer Modulklappe) aktiviert.
Wenn die Station in den Standby-Modus wechselt, wird ein
ggf. angemeldeter Benutzer automatisch abgemeldet. Bei Wechsel in den Betriebsmodus muss sich der Benutzer neu anmelden.
Der Touchscreen-Bildschirm wird während des Standby-
Modus ausgeschaltet.
Das Ladeverhalten der X-am 125+ Module mit Ladefunkti-
on wird durch den Standby-Modus nicht betroffen. Ladun­gen werden weiter durchgeführt.
Um in den Betriebsmodus zu wechseln:
Funktionstaste kurz betätigen oderTouchscreen berühren oder
Dräger X-dock 5300/6300/6600 5
Grundlagen
i
i
eine Modulklappe öffnen oder schließen.

4.2 Ersteinrichtung der Station

1. Station einschalten, siehe Kapitel 4.1 auf Seite 5.
2. Mit dem vorkonfigurierten Benutzer “admin” anmelden (Benutzername: admin, Kennwort: 123456), siehe Kapitel 4.5 auf Seite 6.
3. Prüfgaseinlass konfigurieren, siehe Kapitel 4.6 auf Seite 7.
4. Ggf. Sprache ändern: a. > Systemkonfiguration > Sprache wählen. b. Gewünschte Sprache auswählen. c. Auswahl mit OK bestätigen.
5. Ggf. Datum und Uhrzeit einstellen: a. > Systemkonfiguration > Datum & Uhrzeit wäh-
len.
b. Gewünschte Einstellungen vornehmen. c. Einstellungen mit OK bestätigen.

4.3 Touchscreen-Bildschirm

Die Schaltflächen des Touchscreen-Bildschirms verändern sich dynamisch abhängig von der gerade ausgeführten Aufga­be. Zum Ausführen einer Aktion das entsprechende Symbol auf dem Display wählen.
Es kann jederzeit die Taste am Master gedrückt werden, um zum Startbildschirm zu gelangen.

4.4 Start- und Testbildschirme

4.5 Benutzer an- oder abmelden

HINWEIS
i
i
Zum Anmelden wird eine Benutzer-ID benötigt. Diese muss durch den Administrator vorher angelegt werden.
Standardmäßig ist ein Benutzer mit Administratorrechten an­gelegt:
Benutzername: admin Kennwort: 123456
HINWEIS
Dräger empfiehlt nach der Erstinbetriebnahme das Kennwort des voreingestellten Benutzers “admin” zu ändern.
Um einen Benutzer anzumelden:
1. wählen. a. auswählen. b. Gewünschten Benutzernamen aus der Liste auswäh-
len.
oder a. Benutzer wählen auswählen. b. Namen des gewünschten Benutzers eingeben.
2. Kennwort eingeben und mit bestätigen.
Die Schaltflächen der Start- und Testbildschirme verändern sich dynamisch abhängig vom Anmelde-, Einzelmodus-Zu­stand und der Anzahl der verwendeten Module. Für weitere In­formationen siehe Technisches Handbuch X-dock 5300/6300/
6600.

4.4.1 Symbole

Menü Diese Schaltfläche wählen, um
Bestätigen Diese Schaltfläche wählen, um
Abbrechen Diese Schaltfläche wählen, um
Zurück Diese Schaltfläche wählen, um
Benutzer an- oder abmelden
in das Menü zu gelangen.
eine Eingabe oder Funktion zu bestätigen.
eine Eingabe oder Funktion abzubrechen.
zum vorherigen Bildschirm zu gelangen.
Diese Schaltfläche wählen, um Benutzer an- oder abzumelden. Die Zahl im Symbol gibt die jeweilige Berechtigungsstufe an (siehe Kapitel 6 auf Seite 11).
HINWEIS
i
i
Während der Eingabe des Benutzernamens werden automatisch 3 Suchvorschläge von gespeicherten Benutzernamen angezeigt. Zur Schnellauswahl den gewünschten Benutzernamen auswählen.
Um den aktuellen Benutzer abzumelden:
1. wählen. Informationen zum aktuellen Benutzer werden angezeigt.
2. wählen. Der aktuelle Benutzer wird abgemeldet.
6 Dräger X-dock 5300/6300/6600
Grundlagen
!
i
i

4.6 Prüfgaseinlass konfigurieren

WARNUNG
Die eingetragenen Prüfgaskonzentrationen müssen identisch mit den Angaben auf der verwendeten Prüf­gasflasche sein. Bei falschen Angaben kommt es zu fehlerhaften Messergebnissen.
Um einen Prüfgaseinlass zu konfigurieren:
1. > Prüfgaskonfig. d. Station wählen.
Die Übersicht der Prüfgasanschlüsse wird angezeigt.
Prüfgaseinlass auswählen
Alle notwendigen Angaben für die Konfiguration werden automatisch ausgefüllt. Die Losnummer und das Verfalls­datum können zusätzlich manuell eingetragen werden.
HINWEIS
i
i
Die automatisch eingetragenen Werte müssen mit den Angaben auf der Prüfgasflasche abgeglichen werden. Wenn sich die Werte unterscheiden, gelten die Anga­ben auf der Prüfgasflasche und die Werte müssen ma­nuell korrigiert werden.
2. Ggf. Weitere Optionen wählen und wählen, um die Fla-
schenfüllstandskontrolle zurückzusetzen.
3. Ggf. weitere Prüfgaseinlässe auf gleiche Weise konfigurie­ren.
Bei Verwendung einer Prüfgasflasche eines anderen Herstel­lers:
1. Prüfgaskomponente anlegen oder löschen.
Mit eine neue Prüfgaskomponente anlegen.Mit die aktuelle Prüfgaskomponente löschen.
HINWEIS
i
i
Durch Löschung aller Prüfgaskomponenten werden alle Angaben des Prüfgaseinlasses gelöscht.
2. Gewünschten Prüfgaseinlass auswählen. Das Konfigurationsmenü erscheint.
Prüfgaseinlass auswählen
Sachnr. eingeben ►◄Losnr. eingeben Verfallsdatum
eingeben
Prüfgas auswählen
Konzentration eingeben
Weitere Optionen
Prüfgaskomponente
►◄Konfiguration
►◄Einheit
anlegen
löschen
auswählen
Bei Verwendung einer Dräger-Prüfgasflasche:
HINWEIS
Bei Eingabe einer Sachnummer einer Dräger-Prüfgas­flasche wird automatisch die Flaschenfüllstandskont­rolle angezeigt, sofern diese nicht deaktiviert wurde (siehe Kapitel 4.6.1 auf Seite 7).
2. Prüfgas auswählen.
3. Prüfgaskonzentration eingeben.
4. Prüfgaseinheit auswählen.
5. Ggf. weitere Prüfgaskomponenten anlegen.
6. Folgende Informationen können optional angegeben wer­den:
Sachnummer der PrüfgasflascheLosnummer der PrüfgasflascheVerfallsdatum der Prüfgasflasche
7. Ggf. Weitere Optionen angeben.
8. Für weitere Informationen siehe Technisches Handbuch X-dock 5300/6300/6600.

4.6.1 Einstellungen

1. > Prüfgaskonfig. d. Station > Einstellungen wählen.
Prüfgas für X-am 2000 auswählen
Einstellung des Frischlufteingangs Experten-Einstellungen
Prüfgas für Pac 7000 OV auswählen Flaschenfüllstandskontrolle
1. Sachnummer der Dräger-Prüfgasflasche eingeben.
Dräger X-dock 5300/6300/6600 7
Grundlagen
i
i
Für das Pac 7000 OV kann für den OV-Sensor zwischen 2 ver­schiedenen Prüfgasen gewählt werden, die auch für die Jus­tierung und den Test eingesetzt werden. Es stehen Kohlenstoffmonoxid (CO) und Ethylenoxyde (EO) zur Aus­wahl.
Für das X-am 2000 kann zwischen 3 verschiedenen Prüfga­sen gewählt werden, die auch für die Justierung und den Test eingesetzt werden. Es stehen Methan (CH und Pentan (C ist der Sensor unterschiedlich empfindlich justiert. Nähere In­formationen hierzu sind den jeweiligen Sensordatenblättern zu entnehmen.
HINWEIS
Das entsprechende Gas muss an einem der Ga­seinlässe angeschlossen und in der Gaskonfiguration eingestellt sein.
Zusätzlich besteht die Möglichkeit bei Propan und Pentan eine Option für „erhöhte Empfindlichkeit“ zu setzen. Durch diese Option wird künstlich die Empfindlichkeit erhöht, um Sensoren so zu justieren, dass diese ungefähr eine Nonan-Empfindlich­keit besitzen (also eine Empfindlichkeit als wären sie mit No­nan justiert worden). Nähere Informationen zum Thema Querempfindlichkeitsjustierung sind den jeweiligen Sensorda­tenblättern zu entnehmen.
Um das Prüfgas für X-am 2000 auszuwählen:
1. X-am 2000 wählen.
2. Gewünschtes Prüfgas aus der Liste auswählen. Zur Auswahl stehen:
Methan - CHPropan - CPentan - PENT
Bei Propan und Pentan kann zusätzlich die Option “Erhöh­te Empfindlichkeit” (Dampfempfindlichkeit) aktiviert wer­den.
3. Auswahl mit OK bestätigen.
) zur Auswahl. Je nach ausgewähltem Gas
5H12
(Standardeinstellung)
4
3H8
), Propan (C3H8)
4
WARNUNG
!
Diese Funktion darf nur durch geschultes und fachkun­diges Personal aktiviert werden, da eine falsch ge­wählte Prüfgaskonzentration zu einem positiven Testergebnis führen kann, obwohl das Gasmessgerät zu spät alarmiert.
Um das Testverhalten bei fehlenden Prüfgasen einzustellen:
1. Fehlendes Gas wählen.
2. Checkbox aktivieren (Standardeinstellung: aktiviert).
3. Auswahl mit OK bestätigen.
Mit dieser Funktion kann eingestellt werden, ob ein Test oder eine Justierung auch dann durchgeführt werden soll, wenn ein benötigtes Prüfgas nicht angeschlossen ist.
WARNUNG
!
Wenn diese Funktion deaktiviert ist, wird der entspre­chende Kanal nicht geprüft bzw. justiert.
Um das Prüfgas für Pac 7000 OV auszuwählen:
1. Pac 7000 OV wählen.
2. Gewünschtes Prüfgas aus der Liste auswählen. Zur Auswahl stehen:
Ethylenoxid - EO (Standardeinstellung)Kohlenstoffmonoxid - CO
3. Auswahl mit OK bestätigen.
Um die Flaschenfüllstandskontrolle einzustellen:
HINWEIS
i
i
Die Flaschenfüllstandskontrolle steht nur für Flaschen, die über eine Dräger Sachnummer konfiguriert werden zur Verfügung.
Um den Frischlufteingang einzustellen:
1. Frischluft wählen.
2. Pumpe (Frischlufteinlass; Standardeinstellung) oder Drucklufteinlass wählen.
3. Auswahl mit OK bestätigen.
In den Experten-Einstellung können folgende Einstellungen vorgenommen werden:
Max. Konzentration für schnellen Begasungstest ignorie-
ren
Testverhalten bei fehlenden Prüfgasen einstellen
Um die von Dräger empfohlene max. zulässige Konzentration für den schnellen Begasungstest zu ignorieren:
1. Max. Konz. für BTQ ignorieren wählen.
2. Checkbox aktivieren (Standardeinstellung: deaktiviert).
3. Auswahl mit OK bestätigen.
Wird diese Funktion aktiviert, können für den schnellen Bega­sungstest höhere Prüfgaskonzentrationen als von Dräger empfohlen verwendet werden.
8 Dräger X-dock 5300/6300/6600
1. Gasfüllstandsüberwachung wählen.
2. Kontrollkästchen Gasfüllstandsüberwachung ein akti-
vieren oder deaktivieren.
3. Auswahl mit OK bestätigen.
Um die Flaschenfüllstandskontrolle für eine neue Prüfgasfla­sche zurückzusetzen:
1. Neue Prüfgasflasche an Prüfgasanschluss anschließen.
2. > Prüfgaskonfig. d. Station wählen.
3. Gewünschten Prüfgaseinlass auswählen.
4. Weitere Optionen wählen und wählen, um die Fla-
schenfüllstandskontrolle zurück zu setzen.
Gebrauch
!
i
i
i
i
00233286.eps
1
2
!

5Gebrauch

WARNUNG
Ein defekter Druckminderer an der Prüfgasflasche kann zu einem erhöhten Druck in der Station führen. Dadurch können sich die Prüfgasschläuche lösen und Prüfgas austreten.
Gesundheitsgefahr! Prüfgas nicht einatmen. Gefah­renhinweise der entsprechenden Sicherheits-Daten­blätter beachten. Für Abführung in einen Abzug oder nach außen sorgen.
HINWEIS
Dräger empfiehlt um Prüfgasverlust zu vermeiden, die Prüfgasflaschen zu schließen, wenn die Station längere Zeit unbeaufsichtigt ist.
Geräte- und Kanalfehler können dazu führen, dass ei­ne Justierung nicht möglich ist.

5.1 Sichtprüfung durchführen

Eine Sichtprüfung der Gasmessgeräte vor jedem Einsetzen in die Station durchführen.
1. Unversehrtheit des Gehäuses, der äußeren Filter und der Typenschilder prüfen.
2. Batteriekontakte und Sensoreingänge auf Verschmutzun­gen prüfen.
HINWEIS
Geräte, die die Sichtprüfung nicht bestanden haben, dürfen nicht in die Station eingelegt werden. Die Gesamtbewertung des Tests kann sonst nicht korrekt durchgeführt werden.
5.2 Gasmessgerät in Modul einsetzen oder
herausnehmen
1. Ggf. Verriegelung leicht nach oben drücken und Modul­klappe nach oben öffnen.
2. Gasmessgerät in das entsprechende Modul legen.
3. Modulklappe schließen. Das Gasmessgerät wird automatisch erkannt.
Nur bei X-am-125+ Modulen mit Ladefunktion:
Nach dem Einlegen des Gasmessgeräts wird der Lade-
zustand für ca. 5 Sekunden durch die Ladezustand­LED angezeigt.
Die Ladefunktion startet automatisch ca. 15 Minuten
nach dem letzten Test.
Um das Gasmessgerät aus dem Modul herauszunehmen:
1. Verriegelung leicht nach oben drücken und Modulklappe nach oben öffnen.
2. Gasmessgerät herausnehmen.
5.3 X-am 125+ Modul mit Ladefunktion
(optional)
WARNUNG
Explosionsgefahr! Nicht unter Tage oder in explosi­onsgefährdeten Bereichen laden. Die X-am 125+ Mo­dule mit Ladefunktion sind nicht nach den Richtlinien für Schlagwetter und Explosionsschutz gebaut.
Ein Laden der Akkus der Gasmessgeräte ist nur mit dem
X-am 125+ Modul mit Ladefunktion möglich.
Die Ladezeit beträgt bei vollständig entladenem Akku ca. 4
Stunden.
Eine neue NiMH-Versorgungseinheit erreicht nach 3 vollen
Lade-/Entladezyklen seine volle Kapazität. Gerät nie lange (max. 2 Monate) ohne Energieversorgung lagern, da sich die interne Pufferbatterie verbraucht.
Wenn eine Störung vorliegt:
Gerät aus dem Modul herausnehmen und wieder einlegen.Wenn die Störung dann nicht behoben ist, Modul reparie-
ren lassen.
VORSICHT
!
Ein Kurzschließen der Ladekontakte in den Modulen, z. B. durch hineingefallene metallische Gegenstände, führt nicht zu Schäden an der Station, sollte jedoch wegen möglicher Erhitzungsgefahr und Fehlanzeigen am Modul vermieden werden.
Übersicht Ladezustand-LED
Farbe Zustand Bedeutung
grün dauerhaft leuchtend Ladezustand 100 % grün blinkend Akku wird geladen. rot blinkend Ladefehler
Um das Gasmessgerät in das Modul einzusetzen:
Dräger X-dock 5300/6300/6600 9
Gebrauch
!
i
i
X-am 5000
X-dock 5300
0
0
01033286.eps

5.4 Selbsttest der Station

Ein Selbsttest wird durchgeführt:
Beim Start der Station.Wenn der letzte erfolgreiche Selbsttest länger als 24 Stun-
den her ist und ein Test durchgeführt wird.
Getestet wird die Dichtigkeit der Station, Funktion der Pumpe, SW-Version der einzelnen Module und des Masters.

5.5 Test durchführen

WARNUNG
Bei Begasung mit Methan, Propan oder Butan im Be­reich >100 %UEG muss ein Abgasschlauch (max. 10 m Länge) an dem Gasauslass angeschlossen sein, um die Absaugung von überschüssigen explosionsfä­higem Gas zu gewährleisten.
HINWEIS
Standardmäßig ist der Einzelmodus aktiviert.
Mehrere Tests können parallel im Einzelmodus ge­startet und ausgeführt werden.
Ein Fehlschlagen eines LED-, Hupen- oder Vibrations­tests führt zu einer negativen Bewertung des Gesamt­tests und somit zu einer Sperrung des entsprechenden Gasmessgeräts.
Wenn der Favoritenmodus aktiviert ist:
Gewünschten Test aus der Favoritenleiste wählen.
Der Test wird automatisch gestartet. Die Status-LED blinkt blau. Die einzelnen Testphasen werden angezeigt.
Wenn der Testplaner-Modus aktiviert ist:
Ggf. Benutzer an der Station abmelden (siehe Kapitel 4.5
auf Seite 6).
Der voreingestellte Test wird nach dem konfigurierten Zeit-
plan durchgeführt.
Wenn der Log-In Modus aktiviert ist:
Benutzer an der Station anmelden (siehe Kapitel 4.5 auf
Seite 6).
Gewünschten Test aus der Favoritenleiste wählen.
Der Test wird automatisch gestartet. Die Status-LED blinkt blau. Die einzelnen Testphasen werden angezeigt.
Test bestanden:
Eine Überprüfung der Sensor-Reserve wird nur bei Sensoren durchgeführt, die diese Funktion unterstüt­zen. Das Ergebnis wird bei den Testdetails angezeigt und gibt eine Information über den Zustand des Sen­sors.
Folgende Tests sind vorkonfiguriert:
Test 1: QUI
Test 2: EXT
Test 3: CAL
Schneller Begasungstest inkl. Alarmüberprüfung. Erweiterter Begasungstest inkl. Nullpunktprüfung und Alarmüberprüfung. Justierung, Alarmtest, Frischluftspülung und Zertifikat.
1. Ggf. Prüfgasflaschen öffnen.
2. Ggf. X-dock einschalten.
3. Sichtprüfung der Gasmessgeräte durchführen (siehe Kapitel 5.1 auf Seite 9).
4. Gasmessgeräte in Module einsetzen (siehe Kapitel 5.2 auf Seite 9).
Wenn der Einzelmodus aktiviert ist:
Der voreingestellte Test wird durch das Schließen der Mo-
dulklappe automatisch gestartet. Die Status-LED blinkt blau. Die einzelnen Testphasen werden angezeigt.
Eine Bestätigung wird auf dem Display angezeigt.Die Status-LED blinkt grün.Ggf. gewünschtes Gerätefeld wählen, um weitere Informa-
tionen zu erhalten.
Gasmessgerät aus dem Modul herausnehmen.
10 Dräger X-dock 5300/6300/6600
Wartung
X-am 5000
X-dock 5300
0
0
01133286.eps
X-am 5000
X-dock 5300
0
0
01133286.eps
Test bestanden, mit Einschränkungen:
Der Zustand bedeutet, dass Teiltests des Favoriten aufgrund von speziellen Einstellungen nicht durchgeführt werden konn­ten.
Eine Bestätigung wird auf dem Display angezeigt.Die Status-LED blinkt gelb.Ggf. gewünschtes Gerätefeld wählen, um weitere Informa-
tionen zu erhalten.
Gasmessgerät aus dem Modul herausnehmen.
Test nicht bestanden:

5.6 Nach dem Gebrauch

1. Ggf. Gasmessgeräte aus Modulen herausnehmen.
2. Prüfgasflaschen schließen.
HINWEIS
i
i
Um den Energieverbrauch niedrig zu halten, empfiehlt Dräger das Gerät nach Gebrauch entsprechend der Gebrauchsanweisung auszuschalten.

6 Wartung

6.1 Instandhaltungsintervalle

HINWEIS
i
i
Je nach sicherheitstechnischen Erwägungen, verfahrenstechnischen Gegebenheiten und gerätetechnischen Erfordernissen ist die Länge der Instandhaltungsintervalle auf den Einzelfall abzustimmen und ggf. zu verkürzen. Dräger empfiehlt für den Abschluss eines Service-Vertrags sowie für Instandsetzungen den DrägerService.

6.1.1 Vor jeder Inbetriebnahme

Folgende Arbeiten sind vor jeder Inbetriebnahme des Geräts durchzuführen:
Verschlauchung auf Verschmutzung, Versprödung und Be-
schädigung prüfen und ggf. auswechseln.
Festen Sitz der Schläuche prüfen, um Gasaustritt zu ver-
meiden.
Anschlüsse aller Kabel auf festen Sitz prüfen.Sichtkontrolle der Module und Sensorabdichtungen. Bei
starker Verschmutzung oder sichtbaren Defekten muss die Sensorabdichtung ausgetauscht werden.

6.1.2 Jährlich

Inspektion der gesamten X-dock-Station durch fachkundiges Personal.
Eine Fehlermeldung wird auf dem Display angezeigt.Die Status-LED blinkt rot.Ggf. gewünschtes Gerätefeld wählen, um weitere Informa-
tionen zu erhalten.
Fehler identifizieren und beheben.Ggf. Test wiederholen.
Übersicht Status-LED
Farbe Zustand Bedeutung
blau blinkend Prozess in Bearbeitung grün blinkend Test erfolgreich bestanden
gelb blinkend
rot blinkend
Test bestanden, mit Einschränkungen Test nicht bestanden/ abgebrochen
Dräger X-dock 5300/6300/6600 11
Wartung
!
i
i
!

6.2 Firmware-Update durchführen

VORSICHT
Während des Installationsvorgangs darf die Span­nungsversorgung der Station nicht getrennt werden. Die Station kann sonst beschädigt werden.
HINWEIS
Die Station unterstützt keine USB-Datenspeicher mit NTFS-Dateisystem.
1. Firmware-Update aus dem Netz herunterladen: a. www.draeger.com aufrufen. b. X-dock Produktseite aufrufen und das Firmware-Up-
date auf einen leeren USB-Datenspeicher in das Wur­zelverzeichnis (Root-Verzeichnis) entpacken.
VORSICHT
Auf dem USB-Datenspeicher dürfen keine älteren Firmware-Dateien vorhanden sein!
2. USB-Datenspeicher mit Firmware-Update an den USB-An­schluss der Station anschließen. Das USB-Symbol erscheint in der Statuszeile.
3. > Systemkonfiguration > Up- date wählen. Eine Liste mit allen auf dem USB­Datenspeicher verfügbaren Firm­ware-Updates wird angezeigt.
4. Aus der Liste das gewünschte Firmware-Update auswählen. Das ausgewählte Firmware-Update wird blau markiert.
6. Nach erfolgreicher Übertragung auf die Station wird auto­matisch ein Neustart der Station mit anschließender Instal­lation des Firmware-Updates durchgeführt. Während des Installationsvorgangs leuchten die Status-LEDs der Modu­le weiß.
7. Nach erfolgreicher Installation wechselt die Station in den Betriebsmodus. Die Station ist betriebsbereit.
5. Firmware-Update mit OK starten.
Der Fortschritt der Installation wird angezeigt..
12 Dräger X-dock 5300/6300/6600
Entsorgung
i
i
00633286.eps
00733286.eps

6.3 Dichtungseinsatz wechseln

HINWEIS
Die Dichtungseinsätze müssen in regelmäßigen Abständen (z. B. bei jeder Inspektion) gewechselt werden oder je nach Bedarf früher.

6.4 Frischluftfilter wechseln

HINWEIS
i
i
Der Frischluftfilter muss bei regelmäßiger Nutzung und abhängig von den Einsatzbedingungen typisch alle 2 Monate gewechselt werden.
1. Alten Frischluftfilter abschrauben.
2. Neuen Frischluftfilter aufschrauben.

6.5 Touchscreen kalibrieren

1. Beim Starten der Anlage die Funktionstaste gedrückt hal­ten bis die Kalibrieranzeige angezeigt wird.
2. Jeweils auf die 5 nacheinander angezeigten Positionsmar­kierungen drücken.

6.6 Reinigung

VORSICHT
!
Raue Reinigungsgegenstände (Bürsten usw.), Reini­gungsmittel und Lösungsmittel können den Frischluft­filter zerstören.
Das Gerät bedarf keiner besonderen Pflege.
Bei starker Verschmutzung kann das Gerät vorsichtig mit
einem feuchten Tuch abgewischt werden.
Gerät mit einem Tuch abtrocknen.
1. Modulklappe öffnen.
2. Beide äußeren Verriege­lungsnasen nach innen drücken und Dichtungsein­satz nach unten heraus­ziehen.
3. Schläuche vom Dichtungs­einsatz lösen.
4. Dichtungseinsatz austau­schen.
5. Schläuche an neuen Dich­tungseinsatz stecken (Pfeile auf Dichtungseins­atz und Schlauch beach­ten).
6. Äußere Verriegelungsna­sen nach innen drücken und Dichtungseinsatz so in die Modulklappe einsetz­ten, bis die Verriegelungs­nasen einrasten.
7. Dichtungseinsatz auf kor­rekten Sitz in der Modul­klappe prüfen.

7Entsorgung

Produkt gemäß den geltenden Vorschriften entsorgen.
Entsorgung von Elektro- und Elektronikgeräten:
Gemäß Richtlinie 2002/96/EG darf dieses Produkt nicht als Siedlungsabfall entsorgt werden. Es ist daher mit dem nebenstehenden Symbol gekennzeichnet. Dräger nimmt dieses Produkt kostenlos zurück. In­formationen dazu geben die nationalen Vertriebsor­ganisationen und Dräger.
Dräger X-dock 5300/6300/6600 13

Technische Daten

8 Technische Daten 9 Bestellliste

Maße (H x B x T):
Master ca. 120 x 130 x 250 mm Modul ca. 90 x 145 x 250 mm
Gewicht:
Master ca. 1500 g Modul ca. 960 g
Umweltbedingungen:
bei Betrieb bei Lagerung
Gasanschlüsse: 1x Frischluftanschluss
X-dock 5300/6300 3x Gaseinlass X-dock 6600 6x Gaseinlass
Eingangsdruck:
für Messgas 0,5 bar ±20 % für Druckluft 0,5 bar ±20 %
Stromversorgung: 11 V - 28 V Gleichspannung,
Anschlüsse: 3x USB 2.0 Standard-A-
Serien-Nr. (Baujahr): Das Baujahr ergibt sich aus dem
CE Kennzeichnung: Elektromagnetische
0°C bis +40°C
-20 °C bis +50 °C 700 bis 1300 hPa max. 95% relative Feuchtigkeit
1x Drucklufteinlass 1x Gasauslass
6,25 A
Anschluss, (Host, Kabel <3 m) 1x USB 2.0 Mini-B-Anschluss,
(Device, Kabel <3 m) 1x Ethernet-Anschluss RJ45
Datenübertragungsrate 10/100 Mbit
3. Buchstaben der auf dem Typenschild befindlichen Fabriknummer: B=2010, C=2011, D=2012, E=2013, F=2014, G=2015, H=2016, usw. Beispiel: Seriennummer ARFH­0054, der 3. Buchstabe ist F, also Baujahr 2014.
Verträglichkeit (Richtlinie 2004/ 108/EG)
Benennung und Beschreibung
Dräger X-dock 5300 X-am 125 83 21 880 Dräger X-dock 5300 Pac 83 21 881 Dräger X-dock 6300 Master 83 21 900 Dräger X-dock 6600 Master 83 21 901
Dräger X-dock Modul X-am 125 83 21 890 Dräger X-dock Modul X-am 125+
(mit Ladefunktion) Dräger X-dock Modul Pac 83 21 892 Dräger X-dock Modul X-am 125,
AA-Version Dräger X-dock Modul X-am 125+,
AA-Version (mit Ladefunktion) Dräger X-dock Modul Pac,
AA-Version
Wandhalterung einfach 83 21 922 Wandhalterung komfort 83 21 910 Flaschenhalterung (Tischvariante) 83 21 918 Flaschenhalterung für Hutschiene 83 21 928 Netzteil 24 V / 1,33 A (bis zu 3 Module) 83 21 849 Netzteil 24 V / 6,25 A (bis zu 10 Module) 83 21 850 KFZ-Adapter X-dock 83 21 855 Druckregelventil 0,5 bar 83 24 250 Set Pumpenfilter (besteht aus Filter und
einer Schlauchanschlusstülle) Fluorkautschukschlauch 12 03 150 Dichtungseinsatz (X-am) 83 21 986 Dichtungseinsatz (Pac) 83 21 987 Displayschutzfolie X-dock Master 83 21 804 Aufkleber für Modulnummerierung 83 21 839
Barcode-Etikette außen (22 x 8 mm, 500 Stk.)
Barcode-Scanner 83 18 792
Dräger X-dock Manager Basic 83 21 860 Dräger X-dock Manager Professional 83 21 870 Dräger X-dock Manager Lizenz
(1x, beide Versionen) Dräger X-dock Manager Lizenz
(5x, beide Versionen)
HINWEIS
i
i
Dräger empfiehlt Dräger-Prüfgasflaschen zu verwenden.
Bestellnr.
83 21 891
83 24 260
83 24 261
83 24 262
83 19 364
AG02551
83 21 857
83 21 858
14 Dräger X-dock 5300/6300/6600

10 Glossar

Abkürzung Erklärung
ALARM Alarmelementetest BTQ Schneller Begasungstest BTX Erweiterter Begasungstest CAL Justierung DB Datenbank DBMS Datenbankmanagementsystem DL Datenlogger FAV Favorit HORN Hupe LED Leuchtdiode LEL Untere Explosionsgrenze MSD Massenspeichergerät MST Master SPAN Empfindlichkeitsjustierung SW Software T90 Ansprechzeitentest TWA Schichtmittelwert UNDEF Unbekannt UNK Unbekannte Angabe VIB Vibration ZCHECK Nullpunktprüfung ZERO Nullpunktjustierung
Glossar
Dräger X-dock 5300/6300/6600 15
For your safety
!
!
i
i

1 For your safety

1.1 General safety notes

Before using this product, carefully read the Instructions for
Use.
Strictly follow the Instructions for Use. The user must fully
understand and strictly observe the instructions. Use the product only for the purposes specified in the Intended use section of this document.
Do not dispose of the Instructions for Use. Ensure that they
are retained and appropriately used by the product user.
Only trained and competent users are permitted to use this
product.
Comply with all local and national rules and regulations
associated with this product.
Only trained and competent personnel are permitted to
inspect, repair and service the product. Dräger recommend a Dräger service contract for all maintenance activities and that all repairs are carried out by Dräger.
Only trained and competent personnel are permitted to
inspect, repair and service the product as detailed in these Instructions for Use.
Use only genuine Dräger spare parts and accessories, or
the proper functioning of the product may be impaired.
Do not use a faulty or incomplete product. Do not modify
the product.
Notify Dräger in the event of any component fault or failure.

1.2 Definitions of alert icons

The following alert icons are used in this document to provide and highlight areas of the associated text that require a greater awareness by the user. A definition of the meaning of each icon is as follows:
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, could result in physical injury, or damage to the product or environment. It may also be used to alert against unsafe practices.
NOTICE
Indicates additional information on how to use the product.

2 Description

2.1 Product overview (see fold-out section)

1Master 2 Module 3 Status LED 4 Touchscreen display 5 Function key 6 Fresh air inlet with filter 7 Antitheft slot 8 Power supply 9USB ports 10 Ethernet port 11 Mini USB port 12 Exhaust 13 Gas inlets 14 Compressed air inlet 15 Type plate 16 X-am 125 module 17 Charge status LED 18 X-am 125+ module (with charging function) 19 Pac module

2.2 Feature description

2.2.1 Master

The master assumes control of the sequencing of the maintenance station for functional testing, calibration and adjustment, as well as functions for user management, instrument management, printing of standard reports and standard certificates (using PostScript printers only), and also the user interface.

2.2.2 Modules

The instrument-specific interfaces such as, e. g., IR communication, gas exposure unit and charging contact are integrated into the modules. In addition, the modules contain sensors for detecting the visual, audible and vibration alarms on the instruments.

2.3 Intended use

The Dräger X-dock 5300/6300/6600 is a maintenance station of modular construction. The X-dock can be used to perform automated calibrations, adjustments and bump tests on portable gas monitors in parallel and independently from one another. A system consists of one master for 3 (X-dock 5300/
6300) or 6 (X-dock 6600) test gases. The X-dock 5300 consists of a master station with one module and cannot be expanded. Up to 10 modules can be connected to the X-dock 6300 and 6600 master. The modules automatically detect when an instrument has been inserted and control the gas supply so that an appropriate supply to the instrument is ensured at all times.
16 Dräger X-dock 5300/6300/6600
Installation
i
i
i
i
i
i
The following gas measuring devices can be used with the X-dock and its modules:
X-dock 5300/6300/6600
with Pac module: with X-am 125 (+) module:
Dräger Pac 3500 Dräger Pac 5500 Dräger Pac 7000
Dräger X-am 1700 Dräger X-am 2000 Dräger X-am 2500 Dräger X-am 5000 Dräger X-am 5600

2.4 GPL (General Public Licence)

Some of the device software includes open-source software, which has been published under GPL, LGPL or another open­source licence. They include GPL GPLv2, LGPL, MIT, PostgreSQL, Apache, Apache 2, zlib. The source texts of the software can be obtained from Dräger on a CD under material number 83 21 874 for at least three years after purchase of the software. The licence terms and conditions of the software are included on CD.

3 Installation

NOTICE
Ensure adequate space for the entire assembly.
The master and all modules must have the same firmware version. If this is not the case, a firmware update needs to be carried out (see chapter 6.2 on page 23).
Connect the compressed air hose to the compressed
air port (outlet pressure of pressure regulator valve
0.5 bar, flow rate >3 l/min).
OR
If required, connect the fresh air hose to the fresh air
filter.
7. Connect the power pack.
Station with up to 3 modules: Power pack 24 V / 1.33 AStation with 4 to 10 modules: Power pack 24 V / 6.25 A
The entire system is supplied with power via the master.
NOTICE
i
i
Dräger recommends the use of Dräger gas cylinders and Dräger pressure regulator valves (see order list). Alternatively there is the option of using a suitable pressure regulator valve with 0.5 bar outlet pressure and >3 l/min flow rate.
Dräger recommends connecting an exhaust hose (max. 10 m long) to the exhaust outlet to discharge the test gas into the open air.

4Basics

4.1 Switching the station on or off

NOTICE
i
i
If no action has occurred for 10 minutes, the station switches automatically to standby mode.
1. If applicable, install modules on the master station in accordance with the assembly instructions (X-dock 6300/ 6600 only).
A maximum of 10 modules can be installed on one
master station.
The available modules can be combined in any way
desired.
2. If applicable, install wall mount or cylinder holder in accordance with the assembly instructions.
3. Remove the sleeves from the intended gas inlets and from the gas outlet.
NOTICE
If the sleeve is not removed from the gas outlet, the station will be unable to conduct the self-test without errors.
4. Fit the gas feed hoses to the gas inlets on the master and connect them to the regulator valve on the test gas cylinder.
NOTICE
Dräger recommends not exceeding a hose length of 10 m for the gas feed hoses.
5. If required, connect an exhaust hose (max. 10 m long) to the exhaust outlet.
6. Ensure a supply of compressed air or fresh air:
To switch the station on:
Press and hold the key on the master for approx.
1 second. The following information is displayed during the switch-on
process:
Software version number
To switch the station off:
Press and hold the key on the master for approx.
3 seconds. The station switches off.
Standby mode:
Standby mode is activated after approx. 10 minutes with no
activity on the station (input via touchscreen or opening/ closing of a module cover).
When the station switches over to standby mode, any user logged in will be logged off automatically. The user will need to log back in again when switching back to operating mode.
The touchscreen is switched off in standby mode.The charging function of X-am 125+ modules with charging
function is not affected by standby mode. The charging process is not interrupted.
To switch to operating mode:
Tap the function key ortouch the touchscreen oropen or close a module flap.
Dräger X-dock 5300/6300/6600 17
Basics

4.2 Initial setup of station

1. Switch on the station, see chapter 4.1 on page 17.
2. Log in with the preconfigured "admin" user (user name: admin, password: 123456), see chapter 4.5 on page 18.
3. Configure the test gas inlet, see chapter 4.6 on page 18.
4. Change language if necessary: a. Select > System configuration > Language. b. Selected the required language. c. Confirm selection with OK.
5. Set date and time where necessary: a. Select > System configuration > Date & time. b. Make the desired settings. c. Confirm settings with OK.

4.3 Touchscreen display

The buttons on the touchscreen display change dynamically depending on the task being executed. To execute an action, select the corresponding icon on the display.
Press the key on the master at any time to access the start screen.

4.4 Start and test screens

The keys on the start and test screen change dynamically depending on the log-in status, single-mode status and the number of modules in use. For more information see the X­dock 5300/6300/6600 Technical Manual.

4.4.1 Symbols

4.5 Log user in or out

NOTICE
i
i
A User-ID is required to log in. This must be created beforehand by the Administrator.
A user with administrator rights is created by default:
User name: admin Password: 123456
NOTICE
i
i
Dräger recommends changing the admin password after initial start-up.
To log in a user:
1. Select . a. Select . b. Select the desired user name from the list. or a. Select Select user. b. Enter desired user name.
2. Enter the password and confirm with .
NOTICE
i
i
When entering the user name, 3 already saved user names will automatically be suggested and displayed. For quick selection, please select the desired user name.
Menu Select this button to access the
menu.
Confirm Select this button to confirm an
input or function.
Cancel Select this button to cancel an
input or function.
Back Select this button to access the
previous screen.
Log user in or out Select this button to log users in
or out. The number in the symbol indicates the permission level (see chapter 6 on page 23).
To log out the current user:
1. Select . Information about the current user will be displayed.
2. Select . The current user will be logged out

4.6 Configuring the test gas inlet

WARNING
!
The gas concentrations entered must be identical to the specifications on the gas cylinder used. Incorrect details will result in faulty measurement results.
To configure a gas inlet:
1. Select > Station gas configuration. An overview of the test gas connections is displayed.
18 Dräger X-dock 5300/6300/6600
Select the gas inlet
i
i
i
i
2. Select desired gas inlet. The configuration menu will appear.
Select Test gas inlet
Enter part no. ►◄Enter batch no. Enter Expiry date
Basics
When using a gas cylinder from another manufacturer:
1. Create or delete test gas component.
Create a new test gas component with .Delete the current test gas component with .
NOTICE
i
i
Deleting all test gas components deletes all information of the test gas inlet.
2. Select gas.
3. Enter the gas concentration.
4. Select gas units.
5. If required, create additional test gas components.
6. The following information can optionally be given:
Part No. of the gas cylinderLot number of test gas cylinderUse-by date of the gas cylinder
7. If required, enter Further options.
8. For more information see the X-dock 5300/6300/6600 Technical Manual.

4.6.1 Settings

1. Select > Station gas configuration > Settings.
Create Test gas
component
Select Test gas
Enter concentration ►◄Select unit
Further options
►◄Delete configuration
When using a Dräger gas cylinder:
NOTICE
Upon entering the part no. of a Dräger gas cylinder, a gas cylinder level indicator is displayed automatically, unless this function was deactivated before (see chapter 4.6.1 on page 19).
1. Enter the Part No. of the Dräger gas cylinder. All the necessary details for the configuration will be
automatically filled in. The batch number and the expiry date can be entered manually in addition.
NOTICE
The values entered automatically must be matched with those indicated on the test gas cylinder. If these values are not identical, the value indicated on the gas cylinder is to be considered and a manual correction of the values must be made.
2. If necessary select Further options and , to reset the gas cylinder level indicator.
3. If required, configure other gas inlets in the same way.
Select test gas for X-am 2000 Fresh air inlet setting
Expert settings
Select test gas for PAC 7000 OV
Gas cylinder level indicator
The OV sensor of the PAC 7000 OV allows to choose between two different test gases that are also used for calibration and testing. These test gases are carbon monoxide (CO) and ethylene oxide (EO).
For the X-am 2000, it is possible to choose between 3 different test gases that are also used for calibration and testing. The three options are methane (CH pentane (C sensitivity levels depending on the gas selected. More
). The sensor is calibrated with different
5H12
), propane (C3H8) and
4
information on this can be found in the relevant sensor data sheets.
NOTICE
i
i
The corresponding gas must be connected to one of the gas inlets and set in the gas configuration.
Dräger X-dock 5300/6300/6600 19
Use
!
There is also the option of setting an "increased sensitivity" option for propane and pentane. This option artificially increases the sensitivity to calibrate the sensors so that these have approximately a nonane sensitivity level (in other words, a sensitivity level as if they were calibrated for nonane). More information on the subject of cross-sensitivity calibration can be found in the relevant sensor data sheets.
To select the test gas for X-am 2000:
1. Select X-am 2000.
2. Select the required test gas from the list. The following selection is available:
Methane - CHPropane - CPentane - PENT
For propane and pentane, the "Increased sensitivity" (vapour sensitivity) option can also be activated.
3. Confirm selection with OK .
To set the fresh air input:
1. Select Fresh air.
2. Select pump (fresh air inlet; default setting) or compressed air inlet.
3. Confirm selection with OK .
The following settings can be made in Expert settings:
Ignore maximum concentration for quick bump testSet test behaviour for missing test gases
To ignore the max. allowable concentration for the quick bump test recommended by Dräger:
1. Select Ignore max. conc. for BTQ.
2. Activate checkbox (default setting: deactivated).
3. Confirm selection with OK.
If this function is activated, higher test gas concentrations can be used for the quick bump test than those recommended by Dräger.
WARNING
Only trained and experienced personnel are permitted to activate this function, because an incorrectly selected test gas concentration may result in a positive test result even though the instrument alarmed too late.
To set the test behaviour with missing test gases:
1. Select Missing gas.
2. Activate checkbox (default setting: activated).
3. Confirm selection with OK.
This function can be used to set whether or not a test or calibration is carried out when a required test gas is not connected.
(default setting)
4
3H8
WARNING
!
If this function is deactivated, the corresponding channel is not tested or calibrated.
To select the test gas for PAC 7000 OV:
1. Select Pac 7000 OV.
2. Select the required test gas from the list. The following selection is available:
Ethylene oxide - EO (default setting)Carbon monoxide - CO
3. Confirm selection with OK.
To set the gas cylinder level indicator:
NOTICE
i
i
The gas cylinder level indicator is only available for cylinders that are configured via a Dräger part no.
1. Select Gas level monitoring.
2. Activate or deactivate check box Gas level monitoring on.
3. Confirm selection with OK.
To reset the gas cylinder level indicator for a new test gas cylinder:
1. Connect a new test gas cylinder to a test gas connection.
2. > Select Station gas configuration.
3. Select desired gas inlet.
Select Further options and select , to reset the gas cylinder level indicator.

5Use

WARNING
!
A defective pressure reducer on the gas cylinder can lead to increased pressure in the station. The gas hoses may loosen as a result and gas may escape.
Health hazard! Test gas must not be inhaled. Observe the hazard warnings in the relevant Safety Data Sheets. Provide venting into a fume cupboard or outside the building.
NOTICE
i
i
To prevent loss of gas, Dräger recommends closing the gas cylinders when the station is left unattended for long periods.
Adjustment may not be possible due to instrument and channel errors.

5.1 Conducting a visual inspection

A visual inspection of the gas measurement systems should be conducted every time before being inserted into the station.
1. Check that the housing, external filters and the nameplates are intact.
20 Dräger X-dock 5300/6300/6600
Use
i
i
00233286.eps
1
2
!
2. Check the battery contacts and sensor inputs for dirt.
NOTICE
Devices that have not passed the visual inspection must not be inserted into the station. Otherwise the test cannot be fully evaluated correctly.

5.2 Inserting or removing the gas measuring device into or from a module

To insert the instrument in the module:
1. If necessary, push the lock up slightly and open the module
cover upwards.
2. Place the instrument in the corresponding module.
3. Close the module cover.
The instrument will be detected automatically.
X-am-125+ module with charging function only:
After the instrument is inserted, the charging status is
displayed for approx. 5 seconds via the charging status LED.
The charging function starts automatically approx. 15
mintues after the last test.
To remove the instrument from the module:
1. Push the lock up slightly and open the module cover
upwards.
2. Remove the instrument.

5.3 X-am 125+ module with charging function (optional)

WARNING
Risk of explosion! Do not charge underground or in explosion hazard areas. The X-am 125+ modules with charging function are constructed in compliance with regulations for fire-damp weather and explosion protection.
A new NiMH supply unit reaches its full capacity after 3 full
charge/discharge cycles. Never store the device for extended periods (max. 2 months) without a power supply as the internal buffer battery will run down.
If an error occurs:
Remove from the module and re-insert.If this does not correct the error, have the module repaired.
CAUTION
!
A short-circuit in the charging contacts in the modules, e. g. due to metal objects falling into the device, will not damage the station but should be avoided due to a potential risk of overheating and error displays on the module.
Overview of charge status LEDs
Colour Status Meaning
green on continuously Charge status 100 % green flashing Battery charging. red flashing Charging error

5.4 Station self-test

A self-test is carried out:
When the station is started up.When the last successful self-test was more than 24 hours
ago and a test is being carried out.
The station is tested for leaks, pump function, and the software version of the individual modules and the master.

5.5 Conducting a test

WARNING
!
When using methane, propane or butane in the range >100 %LEL, an exhaust hose (max. 10 m long) must be connected to the exhaust outlet to ensure the extraction of excess explosive gas.
NOTICE
i
i
Single mode is activated by default.
Several tests can be started and executed in parallel in the Single mode.
A failure of a LED, horn or vibration test results in a negative evaluation of the overall test, and thus in the locking of the respective gas measurement system.
A testing of the sensor reserve is only performed with sensors that support this function. The results are displayed under test details and give information about the state of the sensor.
It is only possible to charge the instrument batteries using
the X-am 125+ module with charging function.
The charging time is approx. 4 hours for a completely
empty battery.
Dräger X-dock 5300/6300/6600 21
Use
X-am 5000
X-dock 5300
0
0
01033286.eps
X-am 5000
X-dock 5300
0
0
01133286.eps
X-am 5000
X-dock 5300
0
0
01133286.eps
The following tests are preconfigured:
Test 1: QT Fast bump test including alarm testing Test 2: EXT
Test 3: CAL
Extended bump test incl. zero-point check and alarm testing. Calibration, alarm test, fresh air flushing and certificate.
1. If necessary, open the test gas cylinders.
2. If necessary, switch on the X-dock.
3. Perform a visual inspection of the gas measurement systems (see chapter 5.1 on page 20).
4. Insert the instruments into the modules (see chapter 5.2 on page 21).
If Single mode is activated:
The preset test is started automatically by closing the
module cover. The Status LED flashes blue. The individual test phases are displayed.
If Favorites mode is activated:
Select required test from favourites bar.
The test will be started automatically. The Status LED flashes blue. The individual test phases are displayed.
If Test scheduler mode is activated:
If necessary, log user out on the station (see chapter 4.5 on
page 18).
The preset test is performed according to the configured
time schedule.
If required, select the desired instrument field for additional
information.
Remove the instrument from the module.
Test passed, with restrictions:
The status means that partial tests of the favourite could not be performed because of special settings.
Confirmation is shown on the display.The Status LED flashes yellow.If required, select the desired instrument field for additional
information.
Remove the instrument from the module.
Test not passed:
If Log-in mode is activated:
Log user in on the station (see chapter 4.5 on page 18).Select the desired test from favourites bar.
Test will be started automatically. The status LED flashes blue. The individual test phases are displayed.
Test passed:
Confirmation is shown on the display.The Status LED flashes green.
An error message is shown on the display.The Status LED flashes red.If required, select the desired instrument field for additional
information.
Identify and rectify the error.Repeat the test if necessary.
Overview of status LEDs
Colour Status Meaning
blue flashing Process in progress green flashing Test passed yellow flashing Test passed, with restrictions red flashing Test failed/cancelled
22 Dräger X-dock 5300/6300/6600
Maintenance
i
i
i
i
!
i
i

5.6 After use

1. If required, remove instruments from modules.
2. Close the test gas cylinders.
NOTICE
To keep energy consumption low, Dräger recommends switching off the equipment after use according to the Instructions for Use.

6 Maintenance

6.1 Maintenance intervals

NOTICE
The maintenance intervals must be established in each individual case and shortened if necessary, depending on safety considerations, process conditions, and the technical requirements of the equipment. Dräger recommend a Dräger service contract for all maintenance activities and that all repairs are carried out by Dräger.

6.1.1 Before every start-up

The following work must be carried out before every start-up of the equipment:
Check the hoses for dirt, brittleness and damage and
replace if necessary.
Check the hoses are secure, to prevent escapes of gas.Check that all cable connections are secure.Visual inspection of the modules and sensor seals. If very
dirty or if there are visible defects, the sensor seal must be replaced.
CAUTION
!
There must not be any older firmware files on the USB data storage device.
2. Connect the USB data storage device with firmware update to the USB port on the station. The USB icon will appear in the status bar.
3. Select > System configuration > Update. A list of all firmware updates available on the USB storage device will be displayed.
4. Select the desired firmware update from the list. The selected firmware update is marked in blue.
5. Start the firmware update with OK. The progress of the installation is displayed:

6.1.2 Annually

Inspection of the entire X-dock station by competent personnel.

6.2 Perform a firmware update

CAUTION
The station power supply must not be disconnected during the installation process. The station may be damaged if this is not observed.
NOTICE
The station does not support any USB data storage device with an NTFS file system.
1. Download the firmware update from the internet: a. Go to www.draeger.com. b. Go to the X-dock product page and unzip the firmware
update to the root directory of an empty USB data storage device.
6. Following successful transfer to the station, the start is re­started automatically and the firmware update is installed immediately afterwards. During the installation process, the status LEDs on the modules will be white.
7. After the installation is complete, the station changes to operating mode. The station is ready for operation.

6.3 Changing the sealing insert

NOTICE
i
i
The sealing inserts must be changed at regular intervals (e. g. at each inspection) or sooner as required.
Dräger X-dock 5300/6300/6600 23
Disposal
00633286.eps
00733286.eps
i
i
1. Open the module cover.
2. Squeeze the two external locking lugs in and withdraw the sealing insert downwards.
3. Release the hoses from the sealing insert.
4. Replace the sealing insert.
5. Fit the hoses to the new sealing insert (note the arrows on the sealing insert and the hose).
6. Squeeze the external locking lugs in and insert the sealing insert into the module cover until the locking lugs engage.
7. Check that the sealing insert is fitted correctly in the module cover.

6.5 Calibrate touchscreen

1. While starting the system, press and hold the function key until the calibration display appears.
2. Press the five consecutively displayed position markers as they appear.

6.6 Cleaning

CAUTION
!
Abrasive cleaning implements (brushes etc.), cleaning agents and cleaning solvents can destroy the fresh-air filter.
The device does not need any special care.
If very dirty, the equipment can be carefully wiped down
with a damp cloth.
Carefully dry the device using a cloth.

7 Disposal

Dispose of the product in accordance with the applicable rules and regulations.
Disposing of electric and electronic equipment:
In accordance with EU Directive 2002/96/EC this product must not be disposed of as household waste. This is indicated by the adjacent icon. You can return this product to Dräger free of charge. For information please contact the national marketing organisations and Dräger.

6.4 Changing the fresh air filter

NOTICE
With regular use and depending on the conditions of use, the fresh air filter should be changed typically every 2 months.
1. Unscrew the old fresh air filter.
2. Screw in the new fresh air filter.
24 Dräger X-dock 5300/6300/6600

8 Technical data 9 Order list

i
i

Technical data

Dimensions (H x W x D):
Master approx. 3.54 x 5.70 x 9.84 inches
(120 x 130 x 250 mm)
Module approx. 3.54 x 5.70 x 9.84 inches
Weight:
Master
Module
Ambient conditions:
During operation During storage
Gas connections: 1x fresh air connection
X-dock 5300/6300 3x gas inlets X-dock 6600 6x gas inlets
Inlet pressure:
for the measured gas 0,5 bar ±20 % for compressed air 0,5 bar ±20 %
Power supply: 11 V - 28 V DC, 6.25 A
Connections: 3x USB 2.0 standard A
Serial no. (year of manufacture):
CE mark: Electromagnetic compatibility
(90x145x250mm)
approx. 4.02 pounds (48.23 ounces; 1500 g)
approx. 2.57 pounds (30.86 ounces; 960 g)
032 °F to +104 °F (0 °C to +40 °C)
-20-4 °F to +122 °F (-20 °C to +50 °C)
700 to 1300 hPa max. 95% relative humidity
1x compressed air inlet 1x exhaust outlet
connection, (host, cable <3 m) 1x USB 2.0 mini B connection,
(device, cable <3 m) 1x Ethernet port RJ45
Data transmission rate 10/100 Mbit
The year of construction is given by the 3rd letter in the factory number located on the nameplate: B=2010, C=2011, D=2012, E=2013, F=2014, G=2015, H=2016, etc. Example: Serial number ARFH­0054, the 3rd letter is F, so the year of construction is 2014.
(Directive 2004/108/EC)
Name and description
Dräger X-dock 5300 X-am 125 83 21 880 Dräger X-dock 5300 Pac 83 21 881 Dräger X-dock 6300 Master 83 21 900 Dräger X-dock 6600 Master 83 21 901
Dräger X-dock Module X-am 125 83 21 890 Dräger X-dock Module X-am 125+
(with charging function) Dräger X-dock Module Pac 83 21 892 Dräger X-dock Module X-am 125,
AA version Dräger X-dock Module X-am 125+,
AA version (with charging function) Dräger X-dock Module Pac, AA version 83 24 262
Single wall mount 83 21 922 Comfort wall mount 83 21 910 Cylinder holder (table-top version) 83 21 918 Cylinder holder for top-hat rail 83 21 928 Power pack 24 V / 1.33 A (up to 3
modules) Power pack 24 V / 6.25 A (up to 10
modules) X-dock car adapter 83 21 855 Pressure regulator valve 0.5 bar 83 24 250 Pump filter set (includes filter and hose
connector) Fluoroelastomer hose 12 03 150 Sealing insert (X-am) 83 21 986 Sealing insert (Pac) 83 21 987 X-dock Master display protector film 83 21 804 Stickers for module numbering 83 21 839
Barcode label, exterior (22 x 8 mm, 500 labels)
Barcode scanner 83 18 792
Dräger X-dock Manager Basic 83 21 860 Dräger X-dock Manager Professional 83 21 870 Dräger X-dock Manager Licence
(1x, both versions) Dräger X-dock Manager Licence
(5x, both versions)
NOTICE
Dräger recommends the use of Dräger test gas cylinders.
Order No.
83 21 891
83 24 260
83 24 261
83 21 849
83 21 850
83 19 364
AG02551
83 21 857
83 21 858
Dräger X-dock 5300/6300/6600 25
Glossary

10 Glossary

Abbreviation Description
ALARM Alarm element test BTQ Quick bump test BTX Extended bump test CAL Calibration DB Database DBMS Database Management System DL Data logger FAV Favorite HORN Horn LED Light-emitting diode LEL Lower Explosive Limit MSD Mass Storage Device MST Master SPAN Span calibration SW Software T90 Response time test TWA Time-weighted average UNDEF Unknown UNK Unknown info VIB Vibration ZCHECK Zero-point check ZERO Zero-point calibration
26 Dräger X-dock 5300/6300/6600
Pour votre sécurité
!
!
i
i

1 Pour votre sécurité

1.1 Consignes générales de sécurité

Veuillez lire attentivement la notice d'utilisation du produit
avant de l'utiliser.
Respecter rigoureusement la notice d'utilisation.
L'utilisateur doit comprendre entièrement les instructions et les suivre scrupuleusement. Respecter rigoureusement le domaine d'application indiqué.
Ne pas jeter la notice d'utilisation. Veillez à ce que les
utilisateurs conservent et utilisent ce produit de manière adéquate.
Seul un personnel suffisamment formé et expérimenté
peut utiliser ce produit.
Respecter les directives locales et nationales relatives à ce
produit.
Seul le personnel compétent et possédant la formation
adéquate est autorisé à contrôler, réparer et entretenir le produit. Dräger recommande de conclure un contrat de service qui pourra se charger de tous les travaux de maintenance.
Le personnel de service suffisamment formé doit contrôler
et maintenir en état le produit selon les indications de ce document.
Pour les travaux d'entretien, n'utiliser que des pièces et
des accessoires orignaux Dräger. Sans quoi, le fonctionnement correct du produit pourrait être compromis.
Ne pas utiliser des produits défectueux ou incomplets. Ne
pas effectuer de modifications sur le produit.
Informer Dräger en cas d'erreurs ou de panne du produit
et/ou des pièces du produit.

2 Description

2.1 Aperçu du produit (voir dépliant)

1 Maître 2 Module 3 LED d'état 4 Écran tactile 5 Touche de fonction 6 Entrée d'air frais avec filtre air frais 7 Logement antivol 8 Alimentation électrique 9Ports USB 10 Port Ethernet 11 P o r t m i n i U S B 12 Sortie de gaz 13 Entrées de gaz 14 Entrée d'air comprimé 15 Plaque signalétique 16 Module X-am 125 17 Etat de charge de la LED 18 Module X-am 125+ (avec fonction de charge) 19 Module Pac

2.2 Description du fonctionnement

2.2.1 Maître

Pour la station de maintenance, la station maître gère les tests de fonctionnement, le calibrage, l'ajustage et les fonctions de gestion des utilisateurs, de gestion des appareils, d'impression des comptes-rendus et certificats standard (uniquement avec imprimantes PostScript) ainsi que l'interface vers l'utilisateur.

1.2 Définition des symboles d'avertissement

Les symboles d'avertissement suivants ont pour fonction de caractériser et souligner les textes d'avertissement qui requièrent l'attention accrue de l'utilisateur. Les symboles d'avertissement sont définis comme suit :
AVERTISSEMENT
Signale une situation potentiellement dangereuse qui, si elle n'est pas évitée, peut constituer un danger de mort ou d'accident grave.
ATT ENTION
Signale une situation potentiellement dangereuse qui, si elle n'est pas évitée, peut constituer des dommages physiques ou matériels sur le produit ou l'environnement. Peut également servir d'avertissement en cas d'utilisation non conforme.
REMARQUE
Informations complémentaires sur l'utilisation du produit.

2.2.2 Modules

Les modules intègrent les interfaces spécifiques aux appareils, notamment la communication IR, l'unité de test au gaz et le contact de charge. Par ailleurs, les modules comprennent les capteurs de détection des alarmes visuelle, sonore et vibratoire des appareils.

2.3 Domaine d'application

Dräger X-dock 5300/6300/6600 est une station de maintenance modulaire. La X-dock permet de réaliser des calibrages, des ajustages et des tests de gaz automatiques des détecteurs de gaz portables, simultanément et de manière indépendante. Un système se compose d'une station maître pour 3 (X-dock 5300/6300) ou 6 (X-dock 6600) gaz étalon. La X-dock 5300 comprend une station maître, y compris un module, et n'est pas extensible. Jusqu'à 10 modules peuvent être raccordés à la station maître de X-dock 6300 et 6600. Les modules détectent automatiquement l'insertion d'un appareil et régulent l'alimentation en gaz afin de garantir à tout moment l'alimentation adéquate de l'appareil en gaz.
Dräger X-dock 5300/6300/6600 27
Installation
i
i
i
i
i
i
i
i
Les détecteurs de gaz suivants peuvent être utilisés avec X­dock et les modules correspondants :
X-dock 5300/6300/6600
avec module Pac : avec module X-am 125 (+) :
Dräger Pac 3500 Dräger Pac 5500 Dräger Pac 7000
Dräger X-am 1700 Dräger X-am 2000 Dräger X-am 2500 Dräger X-am 5000 Dräger X-am 5600

2.4 GPL (General Public License)

Certaines parties du logiciel des appareils utilisent des logiciels open source publiés sous GPL, LGPL ou une autre licence Open Source. Il s'agit dans le détail de GPL GPLv2, LGPL, MIT, PostgreSQL, Apache, Apache 2, zlib. Le texte source des logiciels utilisés peut être obtenu auprès de Dräger pendant au moins trois ans à compter de la date d'achat du produit contenant le logiciel sur CD, en indiquant la référence 83 21 874. Les conditions posées à l'utilisation de la licence figurent sur le CD du logiciel.

3 Installation

REMARQUE
S'assurer que l'espace de montage est suffisant.
6. Établir l'alimentation en air comprimé ou en air frais :
raccorder le tuyau d'air comprimé au raccord d'air
comprimé (pression de sortie de la vanne de régulation de la pression 0,5 bar, débit volumétrique >3 l/min).
OU
Au besoin, raccorder le tuyau d'air frais au filtre air frais.
7. Raccorder l'alimentation électrique.
Station comprenant jusqu'à 3 modules : alimentation
électrique 24 V / 1,33 A
Station comprenant de 4 à 10 modules : alimentation
électrique 24 V / 6,25 A
Le système est alimenté électriquement par la station maître.
REMARQUE
Dräger recommande d'utiliser des bouteilles de gaz étalon Dräger et des vannes de régulation de la pression Dräger (voir la liste de commande). Il est également possible d'utiliser une vanne de régulation de la pression adéquate avec une pression de sortie de 0,5 bar et un débit de >3 L/min.
Dräger recommande de raccorder un tuyau d'échappement de gaz (longueur max. 10 m) à la sortie du gaz, ce qui permettra d'évacuer le gaz étalon.
La station maître et les modules doivent avoir la même version de firmware. Dans le cas contraire, effectuer une mise à jour du firmware (voir la section 6.2 à la page 34).
1. Au besoin, monter les modules sur la station maître en suivant les instructions (uniquement pour X-dock 6300/
6600).
10 modules maximum peuvent être installés sur une
station maître.
Les modules disponibles peuvent être combinés selon
besoin.
2. Au besoin, installer le support mural ou de bouteille selon les instructions.
3. Retirer les raccords d'extrémité sur les entrées et la sortie de gaz prévues.
REMARQUE
Si le raccord d'extrémité n'est pas retiré de la sortie de gaz, la station ne pourra pas effectuer proprement l'auto-test.
4. Brancher les tuyaux d'alimentation en gaz aux entrées de gaz de la station maître et les relier à la vanne de régulation de la pression de la bouteille de gaz étalon.
REMARQUE
Dräger recommande de ne pas dépasser une longueur de 10 m pour les tuyaux d'alimentation en gaz.
5. Au besoin, raccorder le tuyau d'échappement du gaz (longueur max. 10 m) à la sortie du gaz.

4 Fonctions de base

4.1 Mise en marche et arrêt de la station

REMARQUE
i
i
En cas d'inactivité pendant 10 minutes, la station bascule automatiquement en mode Veille.
Pour mettre en marche la station :
Appuyer sur la touche de la station maître pendant
1 seconde. Les informations suivantes s'affichent pendant la mise en marche :
Version du logiciel
Pour arrêter la station :
Appuyer sur la touche de la station maître pendant
3 secondes. La station s'arrête.
Mode Veille :
Le mode Veille est activé s'il n'y a pas d'activité enregistrée
sur la station pendant 10 minutes (entrée sur l'écran tactile ou ouverture/fermeture d'un volet du module).
Lorsque la station passe en mode Veille, un utilisateur
éventuellement connecté sera déconnecté automatiquement. Lors du passage au mode d'exploitation, l'utilisateur doit tout d'abord se reconnecter.
L'écran tactile est désactivé pendant le mode Veille.Le comportement en charge des modules X-am 125+ avec
fonction de charge n'est pas concerné par le mode Veille. Les charges sont poursuivies.
28 Dräger X-dock 5300/6300/6600
Fonctions de base
i
i
i
i
Pour passer dans le mode d'exploitation :
Actionner brièvement la touche de fonction outoucher l'écran tactile ououvrir ou fermer un volet du module.

4.2 Configuration initiale de la station

1. Mettre en marche la station, voir la section 4.1 à la page 28.
2. Se connecter avec le nom d'utilisateur prédéfini « admin» (Nom d'utilisateur : admin, mot de passe : 123456), voir la section 4.5 à la page 29.
3. Configurer l'entrée du gaz étalon, voir la section 4.6 à la page 29.
4. Modifier évent. la langue : a. Sélectionner > Configuration du système >
Langue.
b. Sélectionner la langue souhaitée. c. Confirmer la sélection avec OK.
5. Régler éventuellement l'heure et la date : a. Sélectionner > Configuration du système > Date &
Heure.
b. Modifier les paramètres souhaités. c. Confirmer les réglages avec OK.

4.3 Écran tactile

Les touches de l'écran tactile changent dynamiquement en fonction de la tâche exécutée. Pour exécuter une action, sélectionner le symbole correspondant sur l'écran.
Appuyer sur la touche de la station maître pour accéder à l'écran d'accueil.

4.4 Écrans d'accueil et de test

Les touches des écrans d'accueil et de test changent dynamiquement en fonction de l'état de connexion, du mode individuel et du nombre de modules utilisés. Pour en savoir plus, voir le manuel technique X-dock 5300/6300/6600.

4.4.1 Symboles

Menu Appuyer sur ce symbole pour
accéder au menu .
Confirmer Appuyer sur ce symbole pour
confirmer une saisie ou une fonction.
Annuler Appuyer sur ce symbole pour
annuler une saisie ou une fonction.
Connexion ou déconnexion d'un utilisateur
Appuyer sur ce symbole pour connecter ou déconnecter un utilisateur. Le numéro qui s'affiche dans le symbole correspond au niveau d'autorisation (voir la section 6 à la page 34).

4.5 Connexion ou déconnexion d'un utilisateur

REMARQUE
Utiliser l'ID d'utilisateur pour la connexion. L'administrateur le crée au préalable.
Un utilisateur possédant les droits d'administrateur est créé par défaut :
Nom d'utilisateur : admin Mot de passe : 123456
REMARQUE
Après la première mise en service, Dräger recommande de modifier le mot de passe « admin » prédéfini.
Pour connecter un utilisateur :
1. Sélectionner .
a. Sélectionner . b. Sélectionner le nom d'utilisateur voulu dans la liste. ou a. Sélectionner Sélectionner l'utilisateur . b. Entrer le nom de l'utilisateur voulu.
2. Entrer le mot de passe et confirmer avec .
REMARQUE
i
i
Lors de l'entrée du nom d'utilisateur, 3 suggestions de recherche correspondant à des noms d'utilisateur enregistrés s'affichent. Sélectionner le nom d'utilisateur voulu.
Pour déconnecter l'utilisateur actif :
1. Sélectionner .
Les informations sur l'utilisateur actif s'affichent.
2. Sélectionner .
L'utilisateur actif est déconnecté.

4.6 Configuration de l'entrée du gaz étalon

Retour Appuyer sur ce symbole pour
revenir dans l'écran précédent.
Dräger X-dock 5300/6300/6600 29
AVERTISSEMENT
!
Les concentrations en gaz étalon entrées doivent être identiques aux informations fournies sur la bouteille de gaz étalon utilisée. En cas d'erreur, les résultats de la mesure seront incorrects.
Fonctions de base
i
i
Pour configurer une entrée de gaz étalon :
1. > Sélectionner Config. du gaz (station).
L'aperçu des raccords de gaz étalon s'affiche.
Sélectionner l'entrée de gaz étalon
2. Sélectionner l'entrée de gaz étalon souhaitée. Le menu de configuration s'affiche.
Sélectionner l'entrée de gaz étalon
Entrer le code article ►◄
Entrer la date de péremption
Sélectionner le gaz étalon
Entrer la concentration
Autres options
Entrer le numéro de lot
Créer les
composants
►◄Supprimer la
►◄Sélectionner
du gaz étalon
configuration
l'unité
Si une bouteille de gaz étalon Dräger est utilisée :
REMARQUE
i
i
Les valeurs entrées automatiquement doivent être comparées avec les données figurant sur la bouteille de gaz étalon. Si les valeurs diffèrent, ce sont celles figurant sur la bouteille de gaz étalon qui sont valables et celles entrées automatiquement doivent être corrigées manuellement.
2. Le cas échéant, sélectionner Autres options et pour
restaurer le contrôle du niveau de remplissage.
3. Au besoin, configurer les autres entrées de gaz étalon de la même manière.
Si une bouteille de gaz étalon d'un autre fabricant est utilisée :
1. Créer ou supprimer les composants du gaz étalon.
Avec , créer un nouveau composant du gaz étalon.Avec , supprimer le composant actuel du gaz étalon.
REMARQUE
i
i
La suppression de tous les composants du gaz étalon efface toutes les informations de l'entrée de gaz étalon.
2. Sélectionner le gaz étalon.
3. Entrer la concentration du gaz étalon.
4. Sélectionner l'unité du gaz étalon.
5. Au besoin, créer d'autres composants de gaz étalon.
6. Les informations suivantes peuvent être fournies en option :
Code article de la bouteille de gaz étalonNuméro de lot de la bouteille de gaz étalonDate de péremption de la bouteille de gaz étalon
7. Au besoin, indiquer les Autres options.
8. Pour en savoir plus, voir le manuel technique X-dock 5300/ 6300/6600.

4.6.1 Paramètres

1. Sélectionner > Config. du gaz (station) > Paramètres.
REMARQUE
Lorsque vous entrez le numéro de référence d'une bouteille de gaz étalon Dräger, le contrôle du niveau de remplissage de la bouteille s'affiche automatiquement dès lors qu'il n'a pas été désactivé
Sélectionner le gaz étalon pour X-am 2000
(voir la section 4.6.1 à la page 30).
Réglage de l'entrée d'air frais
1. Entrer le code article de la bouteille de gaz étalon Dräger. Les informations nécessaires à la configuration sont enregistrées automatiquement. Elles peuvent être complétées manuellement par le numéro de lot et la date
Réglages expert Sélectionner le gaz étalon pour Pac 7000 OV Contrôle du niveau de remplissage de la
bouteille
de péremption.
Pour le Pac 7000 OV, il est possible pour le capteur OV de choisir entre deux gaz étalons différents qui peuvent aussi
30 Dräger X-dock 5300/6300/6600
Loading...
+ 244 hidden pages