EN 12100-1, EN 12100-2, EN 60204-1, EN 61000-6-2, EN 61000-6-4
Date: 12 / 3 / 2007 At: 10:00 am
Daniel ODonnell
Daniel ODonnell
Daniel ODonnellDaniel ODonnell
Daniel O'Donnell
Page 3
ChemLogic 1 User’s Manual
Default Passwords
Setup Menu 0
Service Menu 1234
DOD Technologies, INC
www.dodtec.com
740 McArdle Dr. • Unit C
Crystal Lake, IL 60014
Phone 815.788.5200 • Fax 815.788.5300
Page 4
TABLE OF CONTENTS
IMPORTANT ............................................................................................................................................... 1
CL1 Menu Overview.............................................................................................................................. 13
5.3 Power on Initialization .................................................................................................................... 13
5.4 Main Menu ...................................................................................................................................... 13
APPENDIX G – CHEMLOGIC CASSETTES........................................................................................ 40
APPENDIX H – OPTIC BLOCK CLEANING PROCEDURE ............................................................ 41
Page 6
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
IMPORTANT
: Read Appendix G before handling ChemLogic Cassett
es.
IMPORTANT
Only trained operators must be allowed to work on the Cl1.
ELECTRICAL HAZARD
The CL1 may only be serviced by trained service personnel.
Contact the manufacturer for more information
.
Be sure to read this manual in its entirety before installing and operating the CL1. Contact the
manufacturer with any questions regarding the installation or use of the CL1.
Rev 10.07
Page 1
Page 7
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
Chapter 1 - Overview
1.1 Introduction
The DOD Technologies ChemLogic 1 (CL1) continuously monitors a single location
(called a point) for toxic and corrosive gas. It responds to gas that exceed a programmed
alarm level by:
• Triggering visual alarms on the display that warn of high or low concentrations
• Triggering relays or activating analog outputs to external devices
• Displaying the gas type and gas concentration
• Recording the alarm information and storing it to removable storage.
The CL1 triggers relays for two levels of gas concentrations. These programmable limits
are factory-set at 1 TLV and 2 TLV for their respective gases.
The point may be up to 150feet (45 m) from the CL1 location depending on the type of
gas being monitored. This allows operators to monitor the gas concentration in an area
removed from the location where gas may actually be leaking.
The CL1 provides a fast response to a wide range of gases. It was designed for maximum
uptime, so routine maintenance and service can be performed quickly and easily.
The CL1 uses DOD Technologies ChemLogic paper tape technology for fast and
accurate gas detection.
1.2 Sampling and Monitoring
The system draws sample flow through the inlet on the bottom of the unit and across the
ChemLogic tape. The gas is then exhausted through a port on the side of the CL1.
1.3 Flow Connections
Flow connections consist of “quick-connect” ports on the bottom and side of the CL1.
There is one inlet and one exhaust outlet.
1.4 Electrical Connections
The unit is powered with a standard AC Power plug. A single 14 pin connector on the side of
the CL1 provides all electrical connections for the outputs and remote reset.
1.5 Theory of operation
The sample flow is diverted across the ChemLogic Tape. The CHEMLOGIC 1 uses an
advanced optical detection system to measure the light level reflected from the ChemLogic
tape. As the target gas is detected, the color of the of the ChemLogic tape changes. This
color change results in a loss of reflected light across the ChemLogic tape. This loss of
reflected light is detected by the advanced optics system in the CHEMLOGIC 1. The
CHEMLOGIC 1 will then report an appropriate gas concentration reading and/or a gas alarm.
Rev 10.07
Page 2
Page 8
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
Chapter 2 - Features
2.1 External Layout
Maintenance Door
Exhaust Outlet
14 pin
I/O Connector
A/C Power & Switch
Gas tubing inlet
(underside)
Keypad & Display
Take-up reel
Door latches
(side of unit)
ChemLogic Tape
Figure 2.1
2.1.1 Maintenance door
The maintenance door allows easy access to the ChemLogic tape and the display/keypad. The
latches on the side of the unit are used to open & close/secure the door.
IMPORTANT: The maintenance door should remain closed and latched except when
changing the ChemLogic tape.
2.1.2 Keypad & Display
The CL1 uses a two color LCD display with a 20 button keypad including 4 programmable
“soft keys”. See chapter 5 for a complete description on the use of the keypad and display.
2.1.3 ChemLogic tape and Take-up Reel
ChemLogic paper tapes are accessed by opening the maintenance door. Refer to chapter 6
regarding tape installation/replacement.
2.1.4 Gas Inlet/Exhaust
The gas being monitored flows through the inlet on the bottom of the unit, across the paper
tape and out the exhaust port on the side of the CL1. Sample tubing and exhaust use a quick
connection system for simple installation. See section 3.3 for information on connecting the
sample and exhaust tubing.
Rev 10.07
Page 3
Page 9
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
. IMPORTANT:
DANGER:
End of line filters are required. See section 6.4
2.1.5 A/C Power & Switch
A/C power is connected on the left side panel with a standard power cord. The on/off power
switch is located adjacent to the power cord connection.
2.1.6 14 Pin I/O Connector
The connector on the side of the unit connects to alarm relays, fault relays, 4-20ma output, and
the remote reset input.
2.2 Maintenance area
The maintenance area allows easy access for changing ChemLogic tape in the CL1. Figure
2.2 shows the internal layout with the access panel open. See section 6.3 for tape installation.
Figure 2.2
2.3 Internal Layout – Service area
Internal access to the CL1 for service uses the latches located on the right side of the front
panel. Figure 2.3 shows the internal layout of the CL1 with the service door open.
Turn off the unit and disconnect A/C power to the unit before
opening the Service Door.
The door should be opened by trained service personnel (See section 6.2)
Rev 10.07
Page 4
Page 10
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
Inside w/Serv
ice Door
Back of Service Door
Output Connection Block
SD Card access door
Easy access
Power Supply
long life pumps
Figure 2.3
2.5 Micro Secure Digital (SD) card
The CL1 uses a SD card to store historical information including concentration logging, event
history, configuration information, and TWA data. SD cards may be purchased through DOD
Technologies – see Appendix A.
Rev 10.07
Page 5
Page 11
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
Chapter 3 - Installation
3.1 Selecting a location
The CL1 should be placed in a location as central as possible to the locations being monitored
while considering the following restrictions:
♦ The maximum sample line length is 150 ft. for all gases other than Diisocyanates.
♦ Diisocyanates have a 6” maximum sample line
♦ Using the shortest possible sample line length will reduce transport times and increase
the response time of the CL1.
♦ A/C power is required to the unit.
♦ Locate near proper ventilation keeping in mind the maximum length of the exhaust
tubing is 25ft.
♦The CL1 requires stable temperature and humidity levels within range to operate
properly.
3.2 Mounting
Refer to figure 3.1 for dimensional requirements for wall mount. Be sure the CL1 is properly
secured to the wall.
Rev 10.07
Page 6
Page 12
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
Figure 3.1
Rev 10.07
Page 7
Page 13
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
IMPORTANT:
IMPORTANT:
Gas inlet
3.3 Sample Tubing
Sample tubing connects to the CL1 through the inlet on the bottom of the unit. (see Figure
3.2). Sample tubing must be FEP 1/4 OD x 1/8 ID (150 ft max length – 6” for diisocyanates)
which may be purchased from DOD Technologies, INC (See Appendix A).
on bottom
of CL1
Figure 3.2
Fully depress the sample tube into fitting past the O-Ring when attaching. To detach the tube,
push on the collet and pull the tubing out.
Sample tubing used with the CL1 must be 1/4”OD x 1/8” ID FEP.
Use of any other tubing may damage the CL1 and/or cause inaccurate gas
concentration readings.
3.3.2 End of line particulate filters
An end of line particulate filter must be used on all gasses except Diisocyanates to prevent loss
of concentration. Filters require regular maintenance – see chapter 6.
The CL1 requires filtration to prevent dust accumulation in tubing
and internal damage to the unit. Dust that collects in the tubing or the internal system
may cause sample loss and inaccurate gas concentration readings.
End of line particulate filters may be purchased from DOD technologies (see appendix A)
3.4 Exhaust tubing
The exhaust line must be ¼”OD x 3/16” ID tubing with a maximum length of 25ft.
Polyethylene is recommended although polypropylene or Teflon may also be used. Exhaust
tubing may be purchased from DOD technologies (see appendix A)
3.5 A/C Power
A 6ft power cord is included – DO NOT use extension cords with the CL1. Longer cords are
available from most electrical supply stores.
3.6 Output Wiring
The output relays for Alarm Level 1 and System Fault are rated at 6 amp maximum.
The 4-20ma output is adjustable within the CL1 to increase precision. When not in analysis
the CL1 will output the 2ma calibrated value. There is a maximum of 500 ohm load
impedance on the 4-20ma circuit.
Rev 10.07
Page 8
Page 14
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
DANGER:
See Appendix B for a listing of output module connections.
Before doing any wiring modifications be sure that power to
the CL18 is disconnected.
Rev 10.07
Page 9
Page 15
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
Chapter 4 – Setup & Configuration
Follow this checklist to configure the CL1 for first time operation.
1. ___ Set passwords (section 5.6.c)
2. ___ Set system date and time (section 5.5.g)
3. ___ Gas selection (section 5.5.i)
Appendix E contains the specifications for each gas.
4. ___ Alarm level settings (section 5.4.e)
5. ___ Latching relays (section 5.5.c)
6. ___ Energized Relays (section 5.5.e)
7. ___ Idle Timeout (section 5.5.f)
8. ___ Testing Contacts (section 5.4.g)
9. ___ Adjustment of 4-20 ma (section 5.5.h)
10. ___ Install new ChemLogic tape (section 6.3)
11. ___ Install removable media (section 6.6)
12. ___ Format removable media (Appendix F.3)
13. ___ Tape Saver Mode (TSM section 5.5.b)
14. ___ Grease the Gate Cam (section 6.7)
Setup complete - Ready for monitoring.
Rev 10.07
Page 10
Page 16
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
Note 2
Note
1
Note
3
Chapter 5 – Basic Operation
5.1 Using the keypad
The CL1 is controlled using the keypad located around the display. All of the display screens
are accessed through a simple menu system.
Figure 5.1
1) The arrow keys located under the screen are used for two purposes.
a. To navigate through the menus using the <ENTER> key to select the highlighted
item on the menu. In most cases the <ESC> key will exit the current screen and
return to the top of the previous menu.
b. To select among multiple data fields on some of the screens for editing.
2) The arrow keys on the sides of the display referred to as “soft keys” are used only when
associated with a button on a screen. The use of these keys varies from screen to screen – see
the detailed descriptions later in this chapter.
3) When numeric entry is required on any screen the following procedure is used.
1. Use the arrow keys below the screen to select which field to edit. The currently
selected box will have a dashed line around it.
2. Press the <ENTER> key first
3. Use the numeric/function keys to enter the value
4. Also note the up arrow may be used for +/- and the down arrow may be used for the
decimal point when necessary.
5. Press the <ENTER> key again to complete the entry
NOTE: The <F1> key is used to exit analysis and return to the main menu.
Rev 10.07
Page 11
Page 17
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
5.2 Alarm & Fault screens
Three types of messages used in the CL1:
1. Alarm messages (level 1 and level 2)
2. System faults (critical faults that terminate and/or prevent analysis)
3. Event messages (warnings, non-critical faults, and various events)
See Appendix D for a description of each message. The history log keeps the most recent 128
messages of any type in memory at all times. All three types of messages are also logged to
the Micro SD card if present.
Various screens are used in the CL1 to show some or all of the messages (figure 5.2). Each
of these screens work in a similar manner.
Figure 5.2
The arrow next to the upper left soft key indicates to press that key to view the full log. When
the soft key is pressed a screen similar to figure 5.3 will be displayed. Pressing the <ESC>
key will return to the previous menu or screen.
Figure 5.3
From the screen shown in figure 5.3 the arrow keys below the screen allow scrolling of the
messages in the log. On some of the fault/event screens the <F4> key will acknowledge all
the current fault/events. NOTE: The history log does not allow messages to be cleared. The
details for each screen in section 5.4 – 5.6 will describe which keys may be used.
Pressing the <ESC> key once will return to the screen in figure 5.2. Pressing the <ESC> key
again from that screen will return to the previous menu or screen.
Rev 10.07
Page 12
Page 18
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
Main Me
nu (5.4)
Power Up (5.3)
CL1 Menu Overview
Setup Menu (5.5)
Setup Mode
Auto Analysis Timer
Start Analysis
System Faults
Gas Alarms
Gate & Tape
Alarm Levels
Event/Alarm History
Test Alarms
Setup Menu
Service Menu
See Note 1
Main Menu (return to)
Tape Saver
Latching Relays
Enable Pumps
Enable Points
Alarm Relays
Idle Timeout
Date/Time
Test 4-20ma
Select Gas
Conc. Logging
Service Menu (5.6)
See Note 2
Main Menu
Optic Calib.
Passwords
Status
Optic Config
Clear History
Note 1: The setup menu is not password protected by default.
Note 2: The service menu is password protected by default.
Passwords can be changed/cleared in the Password screen on the Service Menu (5.6.c)
5.3 Power on Initialization
When the CL1 is powered on it will begin with an initialization screen (figure 5.4) which is
followed by the restart screen (figure 5.5).
Figure 5.4 Figure 5.5
If the operator touches the <MENU> soft key before the timer reaches 0 the Main
Menu(section 5.4.) appears otherwise after a timeout the system will automatically start
analysis (section 5.4.a)
5.4 Main Menu
Once the CL1 is configured the main menu contains all the screens necessary for normal
operation. The main menu is not password protected.
Rev 10.07
Page 13
Page 19
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
Current gas being monitored
Blinking
Tape Saver Mode
Press to display current
Current gas concentration and units
Press to display
current
Ga
s concentration bar
Text blinks during alarm
Current alarm levels
P1 or P2 flashing display
Indicates new
When Tape Saver Mode
Figure 5.6
Each selection on the main menu is detailed below
5.4.a Start Analysis (Figure 5.7)
Selecting this menu item will bring up the analysis screen below and start gas analysis.
Pressing <F1> during analysis will end analysis and return to the main menu.
indicator and timer.
is NOT active, displays
# of days remaining on
Chemlogic Tape.
See Sec 5.4.b
graph displays % of full
scale.
“Full Scale” text
displayed when
appropriate
and scale used
when pump disabled.
gas alarms
warnings & events
warning/event message(s)
Figure 5.7
The screen continuously displays the current gas concentration as long as the system remains in analysis mode.
WARNING – Tape Saver Mode may affect the display of actual gas concentration. See
section 5.5.b for a complete explanation of Tape Saver Mode (TSM).
When the concentration exceeds one or both of the alarm levels the text “Alarm 1” and “Alarm 2” will blink
appropriately. Anytime Alarm 1 is active the concentration display will also blink. When the concentration
passes the alarm levels the corresponding alarm output relay is triggered. The trigger will remain set depending on
the setting of Latching Relays – see section 5.4.c .
The bar graph on the right side of the display reflects the % of full scale for the current concentration reading. At
the lower left of the screen the display will show ‘P1’ or ‘P2’ if a pump is disabled either manually or
automatically. See section 5.4.d
During analysis the two keys on the right side of the display are active. When pressed the upper right key next to
the alarm levels will display a screen showing the current alarm messages. Analysis remains active while this
screen is displayed. If alarms are active they may be cleared by pressing <F4> (see alarm screen operation in
section 5.3.f). NOTE : If an alarm is cleared but gas is still present above the present alarm level the alarm will
immediately trigger once again. Pressing <ESC> will return to the analysis screen.
An asterisk next to the lower right soft key indicates that there are new warning/fault messages. Pressing the key
on the lower right next to the asterisk will display the messages while remaining in analysis mode. Any active
warning/fault messages may be cleared by pressing <F4> (see alarm screen operation in section 5.3.f). Pressing
<ESC> will return to the analysis screen.
Rev 10.07
Page 14
Page 20
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
Status Area
5.4.b System Faults (Figure 5.8)
See section 5.2 for the use of the fault/alarm screens. The <F4> key is available to clear
messages from the detail screen. The <ESC> key will return to the main menu.
Figure 5.8
5.4.c Gas Alarms (Figure 5.9)
See section 5.2 for the use of the fault/alarm screens. The <F4> key is available to clear
messages from the detail screen. The <ESC> key will return to the main menu.
Figure 5.9
5.4.d Gate & Tape (Figure 5.10)
Touch the lower left <GATE> soft key to open or close the gate. When a new Chemlogic
tape is loaded press the <RESET> soft key to reset the counter for a new tape. The tape
windows remaining displays the number of tape advances expected before the tape is empty.
This number is used in the calculation of the warning message “Chemlogic Tape Low”.
Press the <DONE> soft key to return to the main menu.
Figure 5.10
5.4.e Alarm Levels (Figure 5.11)
Use the arrow keys below the screen to select between Alarm 1 and Alarm 2. Notice the
dotted line around the currently selected field. To enter a new alarm level follow the
procedure outlined in section 5.1There are several errors that can occur and will be displayed
in the “Status Area” of the screen. . Press the <DONE> soft key to return to the main menu.
Page 15
Rev 10.07
Figure 5.11
Page 21
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
5.4.f Event/Alarm History (Figure 5.12)
See section 5.2 for the use of the fault/alarm screens. The messages shown in the history log
cannot be cleared or acknowledged. The most recent 128 messages are contained in the
history log. The <ESC> key will return to the main menu.
Figure 5.12
5.4.g Test Alarms (Figure 5.13)
Use the <ALARM 1>, <ALARM 2>, and <FAULT> soft keys to turn the output relays on
and off. When the <Done> soft key is pressed all of the faults are automatically reset.
Figure 5.13
Rev 10.07
Page 16
Page 22
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
Current Mode
Current state of Tape
# of seconds to wait before
“Trigger” Concentration
Use this soft key to select
Use this soft key to toggle
5.5 Setup Menu (Figure 5.14)
The Setup Menu contains the screens necessary to configure the CL1 for operation. The setup
menu is not password protected by default but may be password protected (see service menu).
Figure 5.14
Each selection on the setup menu is detailed below
5.5.a Main Menu
Selecting this menu item will return to the main menu (Section 5.4)
DANGER: Make sure the Tape Saver Modes are completely understood before
enabling. Tape Saver modes allow the operator to bypass continuous gas monitoring.
Tape Saver Mode is used to reduce tape advancement during a gas release. Three tape saver
modes are available on the CL1 : Fixed Cycle mode(Default), Single Reading, and
Continuous. Use the upper left soft key to select among the three modes.
among 3 Modes
tape advance
TSM on and off.
Saver Mode
Figure 5.15
In Fixed Cycle mode the Minutes entered on the screen is the MINIMUM time each cycle will
take regardless of whether gas is detected or not. The maximum time of the cycle is the
normal cycle time of the CL1 tape being used(normally approx 54 minutes). Enter the # of
minutes (1-54) which will be the minimum time before tape advance. If gas is detected and
the tape becomes saturated in less than the time specified, the system will wait until the cycle
time is complete before advancing. When this occurs the timer on the Analysis screen will
begin counting down the # of seconds remaining until tape advance. In Fixed Cycle Mode
the conc. Level is not used and will always show a non-zero value.
In Single reading mode the time specified (minutes) is the exact analysis cycle time. At the
start of each cycle the system will take 1 reading and report the concentration(including zero).
The timer will then continue to countdown to zero before advancing to the next cycle. In
Single Reading Mode the conc. Level is not used and will always show 0.
Rev 10.07
Page 17
Page 23
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
‘TSM’ displayed whenever Tape Saver
TSM seconds counter counts down to 0
In Continuous mode both the time (minutes) and the conc. Level are used. If the
concentration should reach the level specified the timer will start and count down from the
time specified to zero before advancing the tape.
Figure 5.15
WARNING – When tape saver mode is active the actual gas concentration may not be
displayed. Anytime the counter (in seconds – see figure 5.16) is counting down to zero
NO ANALYSIS IS BEING PERFORMED and the concentration displayed may not be
accurate.
While tape saver mode is enabled the letters ‘TSM’ will always appear on the analysis screen.
before tape advance. During this time
no analysis is done.
Mode is enabled. Blinking ‘TSM’
indicates TSM counter running.
Figure 5.16
Example using figures 5.15 and 5.16: When the values displayed in figure 5.15 are
entered and TSM is enabled (ON), a gas concentration reading of 100.3 would start the
TSM counter and freeze the concentration reading at 100.3 ppb.(see figure 5.16)
Regardless of the amount of gas present during this time the counter will start at 300
seconds and countdown to 0 before advancing the tape and calculating a new gas
concentration.
After the tape advance the CL1 will again monitor for gas and display the calculated
value. If the new calculated value is below the TSM concentration the concentration
displayed will begin to update appropriately. Once the gas concentration reaches 100.0
the timer will start again and the value will remain at the last calculated concentration.
Press the upper left soft key to toggle the latching relays on and off. The display adjacent to
the soft key always displays the current state of the relays. (ex : in figure 5.17 the relays are
currently non-latching).
Rev 10.07
Page 18
Page 24
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
Figure 5.17
Latching relays will cause a gas alarm relay to remain active even if the gas concentration
drops below the alarm level until it is acknowledged. If the relays are set to non-latching the
output the relay outputs will toggle on and off as the gas concentration goes above and below
the corresponding alarm level.
Press the upper soft keys to toggle between enabled/disabled on the corresponding pump. The
display adjacent to the soft key always displays the current state of the pump. (ex : in figure
5.18 both pumps are currently enabled). When a pump is disabled a flashing ‘P1’ or ‘P2’
corresponding to the disabled pump will appear in the lower left corner of the analysis screen.
(See section 5.4.b) Disabling both pumps will trigger a critical system fault and prevent
analysis mode.
Figure 5.18
5.5.e Enable Points (Default configuration : Only available on CL2)
Similar to the Enable Pumps screen, this screen will allow either point on a CL2 to be
enabled/disabled.
Use the upper left soft key to select Energize/De-Energized alarm relays. The soft key always
shows the CURRENT state of the relays The initial default value is Energized.
Enter the number of minutes before an ‘Idle Timeout Fault’ occurs. When the system is out of
analysis for the specified amount of time the fault will occur. Specifying zero(0) minutes will
disable the fault from occurring. The maximum amount of time allowed is 45 minutes.
Rev 10.07
Page 19
Page 25
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
Figure 5.20
5.5.h Date/Time (Figure 5.21)
Use the arrow keys below the screen to select among the date & time fields and enter the
values using the procedure outlined in section 5.1.3. The <SET> soft key must be pressed
to apply the changes that are entered.
Figure 5.21
5.5.i Test 4-20ma (Figure 5.22)
Use the soft keys to select 2 ma, 4 ma, or 20 ma. which will bring up the screen in figure 5.23
AND set the 4-20ma output to the corresponding level. Using the “-“ or “+” soft keys the
output value can be adjusted as needed for the selected level. The adjusted value is
automatically saved each time the “-“ or “+” soft key is used. The new values are
permanently stored in the CL1 and used in the 4-20ma scaling for gas concentrations.
Figure 5.22 Figure 5.23
5.5.i Select Gas (Figure 5.24)
Selecting this menu item will display the gas selection shown below in figure 5.24. Press any
of the 3 indicated soft keys to select the gas & range. The <ESC> key will return to the setup
menu.
Figure 5.24
5.5.j Conc. Logging (Figure 5.25)
Selecting this menu item will display the screen shown below in figure 5.25. You can select
between “All Conc. > 0” (default) and “Alarm Level 1”. Selecting the first choice will cause
the system to log all concentrations detected. Selecting the latter will only log concentrations
>= Alarm Level 1. The <ESC> key will return to the setup menu.
Rev 10.07
Page 20
Page 26
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
Figure 5.25
Rev 10.07
Page 21
Page 27
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
5.6 Service Menu
The Service menu is intended for client use but should remain password protected to prevent
unauthorized access. The default Service Menu password is included with this manual. The
password can be changed as needed (see section 5.6.c)
5.6.a Main Menu
Selecting this menu item will return to the main menu (Section 5.4)
5.6.b Optic Calib (Figure 5.26)
This screen is used to manually calibrate the optics which normally only needs to be done
when the system is first configured. Touch the <Start> soft key to start the calibration
process. When the process starts the DAC & Ref values will be set to 0. When the process
completes both values will update with non-zero values indicating that the process is
complete.
Figure 5.26
5.6.c Passwords(Figure 5.27)
This screen is used to change the passwords for both the Setup Menu and Service menu. A
password of 0 will disable the password feature for the specified menu. The default
passwords are included with this manual.
Figure 5.27
WARNING – Be sure to keep the Service menu password in a safe place. If the
service password is lost you will no longer be able to access the service menu and
may need to restore default values and completely reconfigure the CL1.
Use the arrow keys below the screen to select among the password fields and enter the values
using the procedure outlined in section 5.1.3. The passwords are automatically saved and set
to the value displayed.
5.6.d Optic Config
Selecting this menu item will display a screen of values that may aid service personnel in
verifying the optic system. Under normal conditions the screen is not used. Hit <ESC> to
return to the Service Menu.
Rev 10.07
Page 22
Page 28
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
recover the entries after they have been cleared.
5.6.e Status
Selecting this menu item will display a screen of values that may aid service personnel in
finding system faults some of which may be edited in the screen. Under normal conditions the
screen is not used. Hit <ESC> to return to the Service Menu.
5.6.f Clear History (Figure 5.28)
The history screen contains the 128 most recent event, fault, and alarm messages. Each of the
messages logs the date/time occurred and the date/time cleared as separate entries. In the
“Event/Alarm History” on the Main Menu (Section 5.4.f) it is not possible to clear the history.
By using this screen the history may be cleared if necessary. Use the same procedure used in
the alarm and fault screens to clear the entries. Press the soft key next to the arrow then press
<F4> to clear. <ESC> will then bring back the screen in 5.28 and <ESC> again to return to
the Service menu.
Figure 5.28
WARNING – Clearing the history log is not recommended. There is no way to
5.6.g Configuration
The configuration menu is for certified service personnel only.
Rev 10.07
Page 23
Page 29
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
Insert 3/32” Allen
Important:
Important:
Maintenance door latch
Service door latches
Service door lock
the Service Door.
Chapter 6 - Maintenance
6.1 Maintenance Door Access
The maintenance door is used to access the ChemLogic tape and the display/keypad. To open
the maintenance door release the middle latch on the side of the CL1 as shown in Figure 6.1.
When maintenance is complete be sure to close the maintenance door and secure the latch.
The maintenance door should remain securely latched at all times except
when servicing the ChemLogic tape or using the keys on the display.
(tool required)
Figure 6.1
6.2 Service Door Access
DANGER: Service must be performed by trained service personnel
only. Turn off the unit and unplug A/C power to the unit before opening
wrench to remove
service door lock .
Figure 6.2
Before opening the service door make sure the maintenance door is securely latched and
power is removed. The service door safety lock must be removed with a 3/32” Allen wrench
before the service door can be opened.(Figure 6.2) Release the top and bottom latches on the
side of the CL1 as shown in Figure 6.1 to open the service door. When service is complete be
sure to close the service door and secure all latches on the side of the unit. Then reinstall the
service door safety lock and tighten securely.
The service door must remain securely latched with the safety lock installed
at all times when not servicing the unit. Verify that both latches are secure and the
safety lock is installed to prevent unauthorized access.
Rev 10.07
Page 24
Page 30
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
IMPORTANT
: Read Appendix G before handling ChemLogic Cassettes.
6.3 ChemLogic Paper tape
The ChemLogic paper tape has an expiration date printed on the label. Expired tape should be
disposed of and replaced with new tape to assure proper gas concentration readings. Each DOD
ChemLogic cassette will last for 60 days under normal usage. See Appendix A for ordering
information.
ChemLogic Paper Tape Path
Figure 6.2
A. Review the paper tape path shown in figure 6.2
B. Open the maintenance door. (Section 6.1)
C. Select Gate & Tape from the main menu (Section 5.4.d)
D. Scroll down with the arrow keys and select “Gate & Tape”
E. Touch the <Open/Close Gate> soft key to open the gate.
F. Remove the top tape sleeve with the tape spooled around it by gently pulling away
from the machine.
G. Remove the old tape reel from the bottom spindle by lifting over the washer.
H. Discard the old tape & cassette appropriately
Rev 10.07
Page 25
Page 31
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
Warning :
I. Remove the new tape reel from the protective packaging.
J. Pull the sleeve off the new tape reel and place the reel on the bottom spindle.
K. Place the new sleeve on the top reel.
L. Feed the new tape through the system as shown in figure 6.2
M. Verify the tape position on all rollers before proceeding.
N. Remove the cover off 1 of the adhesive strips on the take-up reel cover.
O. Press the tape onto the adhesive.
P. Turn the top spindle a full turn to secure the tape
Q. Press the Gate Button several times to make sure the tape is advancing properly.
R. On the screen touch the ‘Reset’ soft key. (See section 5.4.d)
6.4Endoflineparticulatefilters
Keep fingers clear during tape advance.
Rev 10.07
Page 26
Page 32
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
6.4 End of line particulate filters
End of line (point of detection) particulate filters which protect the CL1 from damage are
required for most gases. Table 6.1 details the type of filter required for each gas – Filters must
be replaced on a regular basis as shown in the table. Filter orientation is not critical in either
application.
Description Suggested Replacement DOD Filter Part #
6 Months 780248
Diisocyanates
NO FILTER N/A
Table 6.1
6.5 Flow adjustment
The pumps in the CL1 automatically adjust to keep a constant flow to the system. No manual
adjustment is necessary.
6.6 Micro Secure Digital (SD) Card replacement
See Appendix F for detailed information on the contents of the SD Card.
To install a Micro SD card: Align its 8-pin gold edge connector down, facing the front of the
CL1 unit as shown in Figure 6.3; then carefully push it all the way into the Memory slot.
Ensure that it clicks into place. To remove the Micro SD card: Push down on the top of the
card gently to release the spring. The card pops up for removal. Make sure to format the SD
card before first use whenever a new card is purchased.
Rev 10.07
Page 27
Page 33
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
Note :
FAT32 file system will not work
Figure 6.3 – Installing Removable Memory Card
When the Micro SD card format was introduced, it was originally called TransFlash. Cards
labeled either Micro SD or TransFlash, with up to 2.0 GB of Flash memory, are compatible.
The CL1 memory slot is equipped with a “push-in, push-out” connector and a Micro SD card
can be safely inserted into the Memory slot whether the CL1 is powered is On or Off.
The CL1 Micro SD Memory slot uses the PC-compatible FAT16 File System. You must
format all SD cards before use with the FAT16 format. – See Appendix F.3
Micro SD disks formatted with the
6.4Endoflineparticulatefilters
in the CL1.
Appendix F details the data stored on the SD card and how to access it on a personal
computer.
6.7 Fuse Replacement
The system power is fused with a 2 amp fast-acting (5x20mm) fuse.
6.8 Apply grease periodically
The cam attached to the gate motor should be greased every 6 months to prevent wear. Apply
a small amount of number 2 type petroleum or synthetic grease (such as McMaster Carr
#1378K27) to the rounded portion of the cam. – See Figure 6.4
Top of gate mechanism
Figure 6.4 – Grease may be applied either from the top of the gate mechanism or the bottom.
Rev 10.07
Bottom of gate mechanism
Page 28
Page 34
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
Chapter 7 – Service & Support
For information on service and support for your CL1 contact DOD Technologies, INC. using
the information below.
Phone Support
M-F 8am – 5pm (Central Time Zone)
815.788.5200
Service Center
740 McArdle Dr. Unit C
Crystal Lake, IL 60014
Visit our website
www.dodtec.com
Rev 10.07
Page 29
Page 35
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
1-400-020 CL1ChemLogic Tape – Mineral Acids (60 days)
1-200-020 CL1 ChemLogic Tape – Phosgene (60 days)
780248 Disposable non-corrosive end of line filters – Hydrides & Phosgene
60009 Filter housing for use with Teflon membranes for Mineral Acid units
60010 Teflon Membranes – 100 pieces – 47mm
(replacement membrane for use with Part #60009)
2-500-502 Teflon Membranes – 10 pieces – 47mm
(replacement membrane for use with Part #60009)
2-400-007 Tubing FEP 1/4 OD x 1/8 ID x 100' Sample Tubing
Note:All applications except Diisocyanates – 150’ Maximum Sample
Length Diisocyanates – 6” Maximum - Sold in 100 foot increments
2-400-008 Tubing Polypro 1/4 OD x 3/16 ID Exhaust Tubing (25 feet)
Rev 10.07
Page 30
Page 36
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
Appendix B – I/O Connection Detail
14 pin CL1 I/O connector
Figure B.1
Pin
Desription Usage
A System Fault Normally Open
B System Fault Common
C System Fault Normally Closed
D Gas - Alarm Level 1 Normally Closed
E External Fault&Alarm Reset Momentarily connect to pin N for reset
F Gas - Alarm Level 1 Common
G Gas - Alarm Level 1 Normally Open
H 4-20 ma Positive (500 ohm max impedance)
J Ground
K 4-20 ma Negative (500 ohm max impedance)
L Gas - Alarm Level 2 Normally Closed
M Gas - Alarm Level 2 Common
N External Fault&Alarm Reset Momentarily connect to pin E for reset
P Gas - Alarm Level 2 Normally Open
Note :The System Fault relay is energized at all times except when a fault is detected. The
alarm fault relays are configured for energized or de-energized on the Setup Menu.
Note : If the “Positive Pressure Sample Option” is installed, pins L,M,&P can be used to
control a solenoid valve that will activate whenever the pump is not powered.
Rev 10.07
Page 31
Page 37
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
Appendix C – Technical Specifications
Physical Dimensions
Height 12.5 “ (31.75 cm)
Width 10.25” (26.04 cm)
Depth 9” (22.86 cm))
Weight 29 lbs. (13.2 kg)
Operating Temperature 0 to 40 Degrees Celcius
Flow
Sample Tubing FEP ¼” OD x 1/8” ID (150 ft Max)
Diisocyanates– 6” Maximum FEP ¼” OD
Exhaust Tubing ¼” OD x 3/16” ID Polypropylene
Pumps 50% duty cycle – long life (2)
Electrical
Power 3 Amp Max @ 120v AC (60 Hz)
Output Relays 24 VDC 6 Amp Max – 240VAC 6 Amp Max
4-20ma Output 500 ohm load Max
External Reset Contact closure (Normally Open)
Misc
Removable Media Micro SD up to 2 GB formatted (FAT16 File System)
Noise Level < 15 dB
Rev 10.07
Page 32
Page 38
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
Appendix D - System Event Messages
COLOR CODING
RED Alarm Gas Alarm
ORANGE Fault Service Fault
YELLOW Fault Minor Fault
BLUE Event Information Message
GREEN Event Status Update
CL1 Msg
Code Message Possible Cause Corrective Action
1 No Gas Data Loaded System Restored
2 DAC Calibration
Optics Dirty
Required
4 No Gas Selected No Gas Selected
5 Gate Open Fault Motor did not leave home
6 Gate Close Fault Motor did not return
home
7 High Background Fault No Tape Loaded
Optics Dirty
Contact DOD Technnologies support
(815) 788-5200
See manual appendix H for cleaning
instructions and add a filter to inlet if
possible. Also, re-adjust optics -
make sure fresh tape is inserted and
gate is closed. Then go to service
menu-optics and press start.
Select gas in setup menu-select gas
option
Make sure that all wires coming from
the HMI to the motor are fully
inserted(Grey=Q12 and
black=ground). Verify belts are all
connected and cam is greased. If
these steps don't work, contact DOD
Technnologies support (815) 7885200
Check issue from "Gate and Tape"
screen. If the gate opens and does
not stop, there is a problem with the
micro swich. Make sure all wires are
connected(I8 and V+). Slightly bend
the switch arm toward the motor
collar to ensure the set screw is
closing the switch. If these steps
don't work, contact DOD
Technnologies support (815) 7885200
Go to "Gate & Tape" and load fresh
tape.
See manual appendix H for cleaning
instructions and add a filter to inlet if
possible. Also, re-adjust optics -
make sure fresh tape is inserted and
gate is closed. Then go to service
menu-optics and press start.
Rev 10.07
Page 33
Page 39
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
8 Optic Comm Failure Electrical problem
9 All Pumps Disabled Flow Blocked and/or
Unstable Inlet/Exhaust
10 Flow Fluctuation Fault Flow Blocked and/or
Unstable Inlet/Exhaust
11 Idle Timeout Out of Analysis Beyond
Timer
12 Low Background Fault Wrong tape or optic
problem
13 Purge Option Fault Tubing Blocked
14 Optic Auto Cal Failure Optics Dirty
33 SD Card Fault No SD Card or Card Full
34 Analyzing Analysis mode started
35 Chemlogic Tape Low ChemLogic tape nearing
end
36 Pump 1 Disabled Flow Blocked and/or
Unstable Inlet/Exhaust
37 Pump 2 Disabled Flow Blocked and/or
Unstable Inlet/Exhaust
38 Large Flow Fluctuation Flow Blocked and/or
Unstable Inlet/Exhaust
39 High Flow Fault Optics Orifice Dirty
40 Alarm Level 1 Gas Detected
41 Alarm Level 2 Gas Detected
Remove optic block cover and make
sure all wires going to the optics are
fully inserted. Contact DOD
Technnologies support (815) 7885200 if these steps don't correct the
problem.
When blockage is fixed, go to Setup
Menu and enable pumps. If problem
persists, contact DOD (815) 788-
5200.
When blockage is fixed, go to Setup
Menu and enable pumps. If problem
persists, contact DOD (815) 788-
5200.
Re-enter Analysis
Re-adjust optics - make sure fresh
tape is inserted and gate is closed.
Then go to service menu-optics and
press start. If these steps don't
work, contact DOD DOD
Technnologies support (815)7885200
Find block in sample line
See manual appendix H for cleaning
instructions and add a filter to inlet if
possible. Also, re-adjust optics -
make sure fresh tape is inserted and
gate is closed. Then go to service
menu-optics and press start.
If a SD card is installed remove and
re-insert
N/A
Replace tape
Go to Setup Menu and enable pump.
If problem persists, contact DOD
(815) 788-5200.
Go to Setup Menu and enable pump.
If problem persists, contact DOD
(815) 788-5200.
When blockage is fixed, go to Setup
Menu and enable pumps. If problem
persists, contact DOD (815) 788-
5200.
See manual appendix H for cleaning
instructions and add a filter to inlet if
possible. If problem persists, contact
DOD (815) 788-5200.
After gas leak is found and
corrected, reset alarms
After gas leak is found and
corrected, reset alarms
Rev 10.07
Page 34
Page 40
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
42 Remote Reset
43 Power On
44 Optic Calibration Operator initiated
45 Pump 1 high voltage Pump drawing high
current or Bad Pump
46 Pump 2 high voltage Pump drawing high
current or Bad Pump
47 Warning : Optics Dirty Dirty Optics
48 Tape Saver Mode Tape Saver Enabled
49 Hour Diagnostic OK Hourly Checkup
50 Flow Failure 1 L Optics Orifice Dirty
51 Flow Failure 2 L Optics Orifice Dirty
52 Flow Disruption Flow Inconsistent
53 Alarm Level 1 Pt 2 Gas Detected
54 Alarm Level 2 Pt 2 Gas Detected
55 Flow Failure 1 H Sample Inlet Blocked or
Internal Leak
56 Flow Failure 2 H Sample Inlet Blocked or
Internal Leak
N/A
N/A
N/A
Leak somewhere between the optics
and pumps.Check that pumps are
connected and all fittings between
the optics and the pumps are fully
inserted. If problem persists, contact
DOD (815) 788-5200.
Leak somewhere between the optics
and pumps.Check that pumps are
connected and all fittings between
the optics and the pumps are fully
inserted. If problem persists, contact
DOD (815) 788-5200.
See manual appendix H for cleaning
instructions and add a filter to inlet if
possible. If problem persists, contact
DOD (815) 788-5200.
N/A
N/A
See manual appendix H for cleaning
instructions and add a filter to inlet if
possible. If problem persists, contact
DOD (815) 788-5200.
See manual appendix H for cleaning
instructions and add a filter to inlet if
possible. If problem persists, contact
DOD (815) 788-5200.
Contact DOD Technnologies
support (815)-788-5200
After gas leak is found and corrected
reset alarms
After gas leak is found and corrected
reset alarms
Leak somewhere between the optics
and pumps.Check that pumps are
connected and all fittings between
optics and pumps are fully inserted.
Also, make sure all three wires are
connected to the transducer. If
problem persists, contact DOD (815)
788-5200.
Leak somewhere between the optics
and pumps.Check that pumps are
connected and all fittings between
optics and pumps are fully inserted.
Also, make sure all three wires are
connected to the transducer. If
problem persists, contact DOD (815)
788-5200.
Rev 10.07
Page 35
Page 41
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
Appendix F – Removable Media
To ensure that historical and performance information is stored properly be sure to insert a
Micro - Secure Digital (SD) card in the proper slot at all times. See section 6.6 for installation
instructions.
The SD card stores all information in standard comma separated values (*.CSV) format for
easy access with any computer. Remove the SD card from the CL1 as described in section 6.6
and transfer the files to your personal computer*. The files are best viewed with programs
that convert CSV format to rows and columns (Microsoft Excel, OpenOffice, etc) but can be
viewed with any text editor.
Figure F.1 shows the folders found on the SD card depending on the configuration or your
CL1 and the use of the SD cards. All files have the format YYMMDD.CSV. A new file is
automatically created each day – be sure that the Date/Time are set correctly (Section 5.5.e)
Figure F.1
F.1 Concentration Log Files (YYMMDD*.CSV)
Whenever the concentration detected exceeds the preset alarm level in the CL1, the system
starts a log file on the SD card (if available). The log file will update approximately every 2
seconds until the gas is no longer detected. A sample alarm file is shown in figure F.2 The
columns show: Date, Time, Gas Concentration, Alarm Level 1, Alarm Level 2, and the
selected gas range.
Date Time rPtGasConc rPtAlarmLevel1Cu rPtAlarmLevel2Cu nPt1GasIdxCur
09/17/07 04:14:20 PM 107.68 100 200 2
09/17/07 04:14:22 PM 114.41 100 200 2
09/17/07 04:14:24 PM 114.41 100 200 2
09/17/07 04:14:26 PM 119.19 100 200 2
Figure F.2
F.2 Event Log Files (YYMMDD*.CSV)
Each time a new event, fault or alarm is added to the History log on the CL1, the message is
written to the current days event log file on the Micro SD card (if available). A sample event
log is shown in figure F.3. The columns show: Date, Time, Event Number, and Action
(unused). The Event Numbers are listed in Appendix D.
Rev 10.07
Page 37
Page 43
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
WARNING
–
Formatting the Micro SD card will erase all information stored on the card.
Date Time nEventNumberCF nEventActionCF
09/17/07 11:07:52 AM 43 1
09/17/07 11:07:52 AM 33 1
09/17/07 11:08:18 AM 48 1
09/17/07 11:08:24 AM 34 1
Figure F.3
F.3 Formatting Micro SD Disks
The Micro SD Cards used in the CL1 must be formatted with the original FAT (FAT16)
file system. Most new disks purchased are pre-formatted with FAT32 which will not
work in the CL1. The disks may be formatted either from a PC or in the CL1. It is highly recommended that the disk be formatted in the CL1 using the procedure
outlined in the following section instead of using a Windows based PC.
Formatting the disk in a MS Windows based PC
WARNING – Make sure to have the correct drive letter for the SD Disk drive. Selecting the
incorrect drive and formatting could result in complete loss of data and/or operation of the
computer.
1. Insert the Micro SD card into the SD Card adapter included.
2. Insert the SD card adapter into the correct drive slot on the computer.
3. Open “My Computer”
4. See WARNING above - Right click on the drive letter corresponding to the drive
containing the SD Card adapter. Select ‘Format’ from the drop down menu.
5. Under the heading ‘File System’ select FAT – DO NOT SELECT FAT32
6. Uncheck the “Quick Format” box if it is selected
7. Press Start to format the disk.
8. When complete the disk is ready for use in the CL1.
Rev 10.07
Page 38
Page 44
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
Formatting the disk in the CL1
♦ Insert the Micro SD card into the slot.
♦ Wait 5 seconds for the drive to initialize.
♦ Hold in both the “UP” arrow and ‘DOWN’
arrow at the same time which should bring up the
screen in figure 2.2.
Figure 2.1
Figure 2.2
Figure 2.3
Figure 2.4
Figure 2.5
♦ Use the down arrow key to scroll down to
the menu item “Removable Media” shown in
figure 2.3
♦ Use the <ENTER> key to display the
contents of the Micro SD card as shown in figure
2.4
♦ Press the softkey on the lower right side
which will bring up the screen shown in figure 2.5
♦ Press <F3> to format the disk which will
bring up the screen shown in figure 2.6 .
♦ Press the enter key to begin the format.
♦ When complete press <ESC> several times
to return the CL1 Main Menu.
Figure 2.6
Rev 10.07
Page 39
Page 45
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
Appendix G – ChemLogic Cassettes
DOD ChemLogiccassettes are designed for use in colorimetric based gas
monitoring systems. Since this is a particular end-use function and does not release
or result in exposure to hazardous chemicals under normal use Material Safety Data
Sheets (MSDS) are not required.
ChemLogic cassettes are non-toxic and require no special precautions for
protection. However contact with skin may cause the cassette to react, change
color, and no longer be useful in your gas monitoring system. We therefore
recommend that rubber gloves be worn at all times while handling including removal
and installation. Always wash hands as a precaution after handling DOD Chemlogic
cassettes.
Rev 10.07
Page 40
Page 46
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
Appendix H – Optic Block Cleaning Procedure
The CL1 may require periodic cleaning of the optics block when the instrument is used in
environments with a high level of airborne particulate or a “High Background” Fault has
been reported.
There are two methods for cleaning the optics. One is a basic cleaning using canned air
and the other is a more in-depth cleaning of the optics block that requires the optics block
back plate be removed.
In most cases the basic cleaning will be sufficient to clear an optics fault and return the
DAC to normal operating levels.
Basic Optics Cleaning
1) Exit analysis mode and open the gate and remove the tape.
2) Power down the CL1
3) Use a canned air source to blow into the sample tube inlet as shown in the figure
1.
Figure 1
4) Power on the CL1, enter the “Gate and Tape” mode and reinstall the tape.
5) Exit the tape load screen and enter the “Service Mode”. Use password 1234 to
enter “Service Mode”.
6) Enter “Optic Calib” mode
7) The upper DAC value will be 255 if a “High Background Fault” has been
reported on the CL1.
8) Press the “Start” key and the CL1 will recalibrate the optics. The DAC value
should go down to 140 or below. If it does not, you will need to perform the
Advanced Optics Cleaning”.
9) If the DAC value is 140 or below, press “Done”, go to “Main Menu” and clear the
original “High Background” in the “Fault” screen.
10) Exit the “Fault” screen and re-enter analysis.
Rev 10.07
Page 41
Page 47
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
Optics Back Plate
Cover
Optics
Back Plate
Advanced Optics Cleaning
1) Exit analysis mode and open the gate and remove the tape.
2) Power down the CL1
3) Remove the optics back plate cover as shown in Figures 3 & 4
Figure 2 Figure 3
4) Remove the Optics Back Plate by removing the two Phillips head screws.
5) Use a canned air supply to blow into channel two of the optics. Channel 2 is 2nd
closest opening to the CL1 as shown in Figure 4. Use the canned air tube to direct
flow to both the top and bottom of the optics block.
Figure 4
NOTE: There are two versions of the Optics Back Plate. Rev 2 is displayed in Figure 4.
This version of the Back Plate has two holes that align with the posts.
Rev1 Optics Back Plate’s does not have these alignment holes.
6) Re-Install the Optics Back Plate:
a. If you have a Rev 2 Optics Back Plate slide the Optics Back Plate onto the
posts and re-install the Phillips head screws.
Rev 10.07
Page 42
Page 48
C L 1 U S E R S G U I D E D O D T E C H N O L O G I E S , I N C
b. If you have a Rev 1 Optics Back Plate, re-install the Optics Back plate but
do not tighten the screws.
i. Squeeze the Optics Back Plate and the Optics Block using your
fingers.
ii. Fully tighten the screws while squeezing the blocks together.
7) Power on the CL1 and enter the “Gate and Tape” mode and reinstall the tape.
8) Exit the tape load screen and enter the “Service Mode”. Use password 1234 to
enter “Service Mode”.
Note: Based on the version of software you may be asked to calibrate while
exiting the “Gate and Tape” mode. If this occurs, press the “Cancel or NO”
choice.
9) Enter “Optic Calib” mode
10) The upper DAC value will be 255 if a “High Background Fault” has been
reported on the CL1.
11) Press the “Start” key and the CL1 will recalibrate the optics. The DAC value
should go down to 140 or below. If it does not, you will need to perform the
Advanced Optics Cleaning”.
12) If the DAC value is 140 or below, press “Done”, go to “Main Menu” and clear the
original “High Background” in the “Fault” screen. If not, repeat the process and
insure the both the upper and lower channel of the optics block is blown free of
particulate.
13) Exit the “Fault” screen and re-enter analysis.
Rev 10.07
Page 43
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.