DMQ QS5 Series, QS5 B, QS5 DL, QS5 E, QS5 DLE User Manual

Page 1
I
Page 2
P/N: QSD 501 – ENG – Rev 3 – 2012
Page 3
Chapter 1
1.1
Know the QS5
1
1.1.1
Front panel
1
1.1.2
Connectors
2 1.2
Install or replace batteries
2 1.3
Connecting the transducer
5 1.4
The “Q” key
6 1.5
Display illumination and contrast
6
1.5.1
Display backlight illumination
6
1.5.2
Display contrast
7 1.6
Locking and unlocking the keypad
7
Chapter 2
Measuring with the QS5
2.1
Numerical measure screen (Normal)
9
2.1.1
Screen mode-1 (Min., Max.)
9
2.1.2
Screen mode-2 (Percentage Value)
10
2.1.3
Screen mode-3 (Datalogger)
11
2.2
Keys in numerical measure modes
11 2.3
BScan graphic measure screen
14 2.4
Keys in BScan graphic measure sceen mode
15 2.5
Zero calibration (Auto)
15 2.6
1-Point and 2-Point calibration
17 2.7
Changing the material velocity
19
2.7.1
Change to a known velocity
19
2.7.2
Change velocity using a known thickness value
19
2.8
Adjusting trigger and gain
20
2.9
Understanding the coupling icons and the energy level indicator
21
2.9.1
Coupling icons
21
2.9.2
Coupling energy level indicator
22 2.10
Measuring thru-coating or thru-paint
25
I
Page 4
Chapter 3
3.1
Instructions on using the menu system
28
3.1.1
Text editor
29 3.2
Main menu
31
3.2.1
Change measuring unit
31
3.2.2
Alarm settings
31
3.2.3
Set BScan range
33
3.2.4
Select language
33
3.2.5
Unit information
34 3.3
General configuration options
34
3.3.1
Set time and date
35
3.3.2
Set time and date format
35
3.3.3
Set keypad sensitivity
36
3.3.4
Set auto-off time
37
3.3.5
Adjust display contrast
37
3.3.6
Beep activation
38
3.3.7
Introduction screen
38
3.3.8
Owner information
38
3.3.9
Lock configuration options
40
3.3.10
Model upgrade licenses
41
3.3.11
Return to factory default settings
42 3.4
Measuring configuration options (Ultrasonic)
43
3.4.1
Set material velocity
43
3.4.2
Select measure mode
44
3.4.3
Adjust gain
45
3.4.4
Adjust trigger
45
3.4.5
Select refresh rate
45
3.4.6
Select transducer frequency
46
3.4.7
Select coating type
46
3.4.8
Select calibration mode
47
3.4.9
Hold last value
48
3.4.10
Energy level compensation (DAC)
48
3.4.11
Material condition analyzer
49
II
Page 5
3.4.12
Nums Zoom measuring screen mode
50
3.4.13
Continuous measuring monitor
50
3.4.14
Adjust pulser excitation voltage
51
3.4.15
Return to factory default settings
52
Chapter 4
4.1
Understanding how data is organized
53 4.2
Memory menu
53 4.3
Creating a file
54 4.4
Actions over single files
55
4.4.1
View data in a single file
56
4.4.2
The “Q” key in a grid
56
4.4.3
The “Q” key in a histogram
57
4.4.4
The “Q” key in a SCAN graphic
59
4.4.5
Send data from a single file
60
4.4.6
Rename a file
60
4.4.7
View file size
60 4.5
Actions over all files
60
4.5.1
Send all files
61
4.5.2
Erase all files
61 4.6
The Datalogger in the measuring screen
62
4.6.1
Datalogger information
62
4.6.2
Quick memory menu (Mem key)
63 4.7
Connecting to a PC with DataCenter
64
4.8
Datalogger configuration
65
4.8.1
Configure communications
65
4.8.2
Capture modes
66
4.8.3
Set number of values
67
4.8.4
Selecting fixed batch
67 4.9
Advanced Datalogger configuration
68
4.9.1
Store every (N) number of values
68
4.9.2
Record operator actions
68
4.9.3
Record date and time
69
III
Page 6
Tips on how to measure correctly
Technical specifications
Additional information
IV
Page 7
And thank you for purchasing a QS5 ultrasonic corrosion and wall thickness gauge.
At dmq we develop, manufacture and distribute software and quality control instruments offering innovation and solutions that come as a direct result of listening to your needs as a user. We apply some of the latest technology available in the industry to build instruments that are robust, precise, and easy to operate.
We are convinced that our products would not be complete without permanent technical and after sales support. So in addition to a great product we offer:
Quick answers to your inquiries. Unlimited access to technical information as well as
application notes.
Special offers for registered customers. Firmware and software upgrades at no charge. Attention to your inquiries and suggestions.
We hope that the QS5 will meet and exceed your application needs.
V
Page 8
The information included in this manual applies to all QS5 series ultrasonic corrosion and wall thickness gauges including models B, DL, E and DLE.
dmq is a registered trademark of demeq S.R.L and its affiliate companies.
The information contained in this manual is intended to educate users on the operation of the QS5 gauges. Failure to read and understand this manual can lead to measurement errors. Decisions based on measurements and or results that are erroneous can lead to property damage, personal injury or even death. Demeq S.R.L assumes no responsibility as a result of the improper use of our instruments.
ASTM E797
The user must be trained in ultrasonic testing procedures and on how to properly configure the unit for testing. All of the following should be taken into consideration:
Understand the theory on sound wave propagation. Select the instrument as well as the transducer that is best
suited for your application.
Know the specific requirements for the test you will be
conducting.
This manual provides all of the information needed to configure and operate the QS5 gauge. However there are additional factors that can
VI
Page 9
affect tests done with this instrument. Specific information on those factors is outside the scope of this manual. When in doubt you should always seek expert advice or refer to specific textbooks on ultrasonic testing. Additional information can also be found on the internet and through local government agencies as well as in technical institutes.
The QS5 operates with dual crystal transducers that use the “pitch­catch” principal. Dual crystal transducers use two piezoelectric crystals
whereby one crystal is excited with short electrical pulses from the unit sending short acoustic waves into the test piece. The second crystal then receives the acoustic waves reflected at the end of the test piece. The frequency of the acoustic waves changes between 2 and 10 MHz depending on the probe being used. All QS5 gauges operate in Echo-1 or normal (no coat) mode where a measurement is the result of the sum of the time delay from the transducer and the time it takes for the acoustic wave to rebound back to the transducer after it reaches the end of the test piece. In Echo-2 or thru-paint mode (QS5 E and QS5 DLE only), the unit measures the time interval between the first and second echoes and does not take into consideration the time delay from the transducer.
VII
Page 10
Figure 1: Echo-1 and Echo-2 mode acoustic wave representation
In ultrasonic testing information is obtained from measuring sound waves. Users must be very careful when making assumptions in regards to the entire test piece condition when there are areas that have not been inspected. In large and massive test pieces it would be practically impossible to test the entire piece and therefore conclusions for the whole piece based solely on the areas that have actually been inspected, should only be done by experienced operators. Only trained users should evaluate test pieces that could present erosion and or corrosion.
In order to minimize test result errors the following must be observed:
Before taking actual measurements the unit and transducer have to be calibrated. Calibration of the transducer is also known as zero calibration or delay calibration (Page 15). If the unit and transducer are not calibrated, or calibrated incorrectly, the resulting measurements will be unreliable. Each and every time a transducer is changed it must be calibrated to the unit.
Because the sound velocities in materials will often vary significantly from published values, best results are obtained when the instrument is calibrated on a reference test block made from the same material as that of the actual test piece. The block should be flat, of a smooth finish, and be as thick as the maximum thickness of the test piece. Careful consideration by the user must be made when evaluating the accuracy of a thickness value because considerable changes in the sound velocity of a material may have occurred as a result of a treatment. So for example a test piece made of steel that has been thermally treated, will look the same as a regular test piece made of
VIII
Page 11
untreated steel, but the sound velocities vary significantly. Instruments should be calibrated before and after testing to minimize errors.
Transducers must be in good condition and show no visible signs of excessive wear or defects on their surface. Badly worn transducers can provide erroneous measurements. The thickness of the material being tested must be within the specified thickness range of the transducer and the temperature of the material must also be within the specified temperature range of the transducer.
Calibration as well as actual testing should be performed under similar coupling conditions. Couplant must be applied in an even and consistent manner to minimize variations in couplant layers that may result in testing errors. The amount of couplant should be kept to a minimum and consistent pressure should be applied on the transducer when measuring.
Doubling refers to a thickness measurement that displays a reading that is twice or even three times the real test piece thickness. Doubling often may happen when trying to measure below and above the specified range for the transducer and when using worn transducers. When in doubt, the test piece thickness must be measured using other methods and or instruments and the unit must then be calibrated using reference test blocks of known thickness values that are made of the same material as the actual test piece. This is very important when measuring unknown test pieces for the first time.
When possible calibrate your instrument on-site using a test block that is at the same temperature as the test piece in order to minimize errors.
IX
Page 12
All QS5 series corrosion and thickness gauges are for industrial use only and cannot be used in medical applications. The QS5 operates on two AA size batteries. We strongly recommend that you use only top brand name alkaline batteries. Disposal of your QS5 gauge and its components must be done in compliance with all applicable regulations.
Because of its complexity level, software is never really completely error free. For this reason in software controlled instruments always make sure that the operations required for your application are in correct working order.
Demeq S.R.L provides a limited warranty for a period of 2 (two) years on electronic units and for 6 (six) months on transducers from the date of purchase and may be extended for up to 5 years. Every instrument undergoes thorough testing during manufacturing as well as before shipping. In the event warranty service where to become necessary, dmq and or your local distributor or representative will make a reasonable effort to replace your defective unit with another new or used unit, while your instrument undergoes warranty repair.
X
Page 13
Figure 1.1: Front of the unit
1. Graphic LCD display with LED backlight illumination
2. Move Left key / Switch measuring modes: normal
measurements or Echo-1 / Thru-Paint or Echo-2 (Echo)
3. Move Up key / Manually store a value (Store)
4. Move Right key / Switch to graphical measure mode screen
(BScan)
5. Menu key / Enter and exit measure screen / Exit and return to
menus (Home)
Chapter 1 1
Page 14
6. Move Down key / Quick access to memory menu screen
options (Mem)
7. Change backlight illumination key (On, Off, Auto)
8. Enter key / Edit values on the measure screen (Edit)
9. The key: Power On and Shutdown (touch and hold for 2
seconds) / Make quick and short touches to activate special functions and features
10. Calibration disk
11. Horizontal scrolling center point (lock and unlock keypad on
measure screen)
12. Vertical scrolling center point (adjust LCD contrast)
Figure 1.2: Unit connectors
1. Receiving transducer connector (Type Lemo 00)
2. Emitting transducer connector (Type Lemo 00)
3. USB mini connector to connect to PC using a USB cable
(supplied with the unit)
2 Chapter 1
Page 15
The QS5 is powered by 2 (two) AA batteries that are placed in the battery compartment located in the back of the unit. To access to the battery compartment slide the cover as shown in figure 1.3-1 and gently push the extraction ribbon upward and slightly towards the right to release the batteries (figure 1.3-2).
When you install new batteries, first insert the positive end of each battery so that it coincides with the positive pole inside the battery compartment as you see in figure 1.4-1. Always leave the extraction ribbon underneath the batteries.
Figure 1.3: Removing batteries
Chapter 1 3
Page 16
Notes
Use new alkaline top brand batteries only.
Do not mix new and old batteries. Always replace both batteries.
Rechargeable batteries type NiMH can be used but will result in less time of continuous operation.
Figure 1.4: Replace / Insert batteries
4 Chapter 1
Page 17
Important
Do not remove batteries while the unit is powered as this may affect the Datalogger (See Appendix: “Additional information, Error Messages”)
The QS5 has two type Lemo “00” connectors located on top of the unit. Because this is the standard for most dual crystal probes, you could basically connect any probe to your QS5. Keep in mind that the gain may need to be adjusted when changing probes.
To connect the probe, simply align any of the male connectors located at the end of the transducer cable to any of the female connectors located on the gauge and press gently until connected.
To release the probe hold the knurled section on the male connectors and pull gently away from gauge until released.
Figure 1.5: Connecting the transducer
Chapter 1 5
Page 18
The has three functions:
When the unit is off, touch for 2 seconds to power on
the unit.
When the unit is on, touch for 2 seconds to shutdown
the unit.
With the unit on, making short touches to the will
activate special functions described in later chapters of this manual.
Backlight illumination and contrast options can be changed from any screen in the unit.
Touch to change the backlight illumination settings.
Figure 1.6: Backlight illumination options
6 Chapter 1
Page 19
The display contrast on all dmq units is digital. Touch the white dot located in the center of
the vertical scrolling bar between the -
keys and a contrast window will open. Move your finger towards the top and or bottom of the dotted line to adjust the contrast on your display.
Figure 1.7: Display contrast adjustment
To lock the keypad place your finger on the white dot located in the
center of the horizontal scrolling bar between the - keys. Move your finger to the right following the dotted line and a window on the unit display will open with the word Lock. Continue moving your finger in the same direction until you enter locked mode. The window on the display will close and the locked keypad indicator will show on the top right of the unit screen.
Figure 1.8: Locking the keypad
Chapter 1 7
Page 20
\
Important
The keypad can only be locked and unlocked in the measuring screens.
Sliding the finger to the left will unlock the keypad.
Figure 1.9: Unlocking the keypad
8 Chapter 1
Page 21
The QS5 can display measurements in 3 distinct numerical measuring modes or screens each one showing specific information. To switch
numerical measure screens touch the key.
Figure 2.1: Measure screen mode-1
1. Material velocity
2. Datalogger mode indicator (Page 62)
3. Measured value signal in differential measuring mode (Page
44)
4. Battery level indicator
5. Thickness value in the selected unit
6. Icon indicating that a value was stored
7. Locked keypad icon indicator (Page 7)
Chapter 2 9
Page 22
Note
Touching the once will reset maximum and minimum values to zero and touching it again will display the measure screen in mode-2 described next.
8. Inactive calibration disk indicator (Page 16)
9. Coupling level indicator / material condition analyzer (Page
21)
10. Echo energy level indicator (Page 22)
11. Gain level (Page 20)
12. Maximum measured value
13. Trigger level (Page 20)
14. Minimum measured value
15. Percentage value as compared to
the nominal value
16. Nominal value (represents 100% of
the percentage value)
Figure 2.2: Measure screen mode-2
10 Chapter 2
Page 23
Figure 2.3: Measure screen mode-3
17. Data storage modes (Grid / SCAN)
18. Name of open file where data is being stored (Page 53)
19. Position within the grid of the last stored value (Column,
Row)
In all three numerical measure screen modes the following applies in terms of keypad functionality:
: Touch once to switch from Echo-1 to Echo-2. Touch again to switch back to Echo-1 (the Echo key is active only in QS5 E and QS5 DLE models).
: Touch to manually store the measurement in the memory.
: Touch to switch to real time graphic BScan mode.
: Touch to exit the measure screen and enter the main menu.
: Touch to view quick memory access options.
Chapter 2 11
Page 24
: Touch to activate the Select / Edit mode. Four flashing arrows will appear over the fields of measurement, velocity, trigger and gain.
Figure 2.4: Measuring screen in select / edit mode
To select thickness measurement touch To select material velocity touch To select trigger touch To select gain touch
Always use the - keys to edit the values in any given field and
touch to save.
Trigger and gain can be edited while measuring. This allows you to see the new thickness value (resulting from the changes made earlier to the trigger and gain settings) to be displayed immediately.
12 Chapter 2
Page 25
Note
The measurement field can only be edited if an actual measurement was taken. If the measurement is 0.00 this field cannot be edited.
Tips
Touch the when editing velocity and a screen will appear with the most commonly used materials and their velocities. Use the cursor keys to scroll the list and
touch to select or touch to exit and return to the edit velocity field.
When a measurement was taken and the velocity is changed, a new measurement is automatically displayed that reflects the change in velocity.
: Short touches allow you to switch between the three measure screen modes. Touch for 2 or more seconds to shutdown the unit.
: Changes the backlight illumination.
: Adjust display contrast.
Chapter 2 13
: Lock and unlock the keypad.
Page 26
The QS5 allows you to view measured values in real time graphics.
Figure 2.5: BScan graphic
1. Zoom level
2. Battery level indicator
3. Datalogger mode indicator (Page 62)
4. Thickness value in the selected unit
5. Icon indicating that a value was stored
6. Locked keypad icon indicator (Page 7)
7. Inactive calibration disk indicator (Page 15)
8. Coupling level indicator / material condition analyzer
9. Echo energy level indicator
10. Vertical thickness reference axis in the graphic
11. BScan graphic area
12. High alarm setting (Page 31)
13. Low alarm setting (Page 31)
14 Chapter 2
Page 27
Keypad functions in the BScan graphic mode are the same as in the numerical measure screen modes (Page11) except for the following keys:
: Go to the numerical measure screen mode
: Short touches will zoom in / out of the graphic. Touch for 2 or more seconds to shutdown the unit.
The QS5 uses a zero (delay) calibration procedure that is activated by simply coupling the transducer over the calibration disk located in front of the unit.
Figure 2.6: Zero calibration procedure
When this automatic procedure ends, the unit is calibrated to measure steel which is the default material velocity used in this calibration method.
Chapter 2 15
Page 28
Tip
In order to measure the thickness of the calibration disk without activating the automatic calibration procedure simple touch the edge of the disk until an icon appears as shown in (Figure 2.1, 8. Inactive calibration disk
indicator, Page 9) While still touching the edge of the
disk, couple the transducer to obtain its thickness measurement. When the transducer is coupled you no longer need to touch the edge of the disk.
16 Chapter 2
Page 29
In order to obtain precise measurements, especially when exchanging transducers, the QS5 can be calibrated with your own test piece samples. To access the two calibration modes available on the QS5, go to the Ultrasonic menu (page 43) and enter the Calibr. option. Select
2-Points and follow the procedure below for calibration:
Figure 2.7: One point and two point calibration procedure
1. Measure using the calibration disk located in the unit front
panel.
The unit will proceed with the automatic calibration procedure and “Calibrating” will be displayed on the unit
Chapter 2 17
Page 30
Notes
In the 2 point calibration procedure, the thickness of the 1st point (first test piece sample) MUST be lower than the thickness of the second test piece sample. The unit assumes that the material velocities of the test piece samples are the same and that the correct velocity has been set in the unit before the calibration procedure was initiated. Use of the unit calibration disk does not change the velocity set by you.
screen. When the calibration is done, the unit will display
“Calibration… Done!”. At this point the unit will ask that you measure the first test piece sample “Measure 1st Pt.” and
two additional options will be displayed: Edit and Exit.
2. Touch (Exit) to end the calibration using just the unit
calibration disk, or measure the first test piece sample.
3. After a measurement is displayed touch (Edit) to edit
the value using the cursor keys so that it matches the actual test piece thickness.
4. When you are done editing the measurement, touch to
save and the unit will ask “Measure 2nd Pt.”.
5. When a measurement is displayed for the second test piece
simple, touch (Edit) to edit the value using the cursor keys so that it matches the second test piece simple thickness. When you are done editing the measurement,
touch to end the 2-point calibration procedure.
18 Chapter 2
Page 31
Tip
Touch the when editing velocity and a screen will appear with the most commonly used materials and their velocities. Use the cursor keys to scroll the list and
touch to select or touch to exit and return to the edit velocity field.
If you know the velocity of the material you will be measuring you can manually enter the velocity by doing the following:
Go to any numerical measure screen mode. If you are in
graphic BScan mode touch (BScan) to switch to normal mode.
Touch from any numerical measure screen to enter the
Select / Edit mode.
Touch to edit the velocity field. Use the cursors to set the velocity and touch to confirm
or touch to cancel.
If you do not know the velocity of the material but you know the material thickness you can set the velocity using the material thickness value by doing the following:
Measure the material for which you already know the
thickness.
Lift the transducer and touch to enter the Select / Edit
mode.
Chapter 2 19
Page 32
Note
When changing the material velocity using a known thickness value the unit must have already been calibrated for the transducer being used (Page 15).
Touch to select the thickness measurement field. Use the cursors to set the thickness value and touch to
confirm or touch to cancel.
When editing the thickness value observe that the velocity will change accordingly.
The QS5 has two controls (Trigger and Gain) that have a direct impact on the ultrasonic signal process so that any dual crystal transducer can be used and optimized for best performance.
The following image shows you how to interpret changes done to the trigger level represented in higher and lower trigger values.
Figure 2.8: Representation of high and low levels of trigger
20 Chapter 2
Page 33
The image below shows how you to interpret changes done to the gain level represented in higher and lower gain values:
Figure 2.9: Representation of high and low levels of gain
The trigger level determines the echo signals detected by the unit. A higher trigger means lower sensitivity where only larger echoes are
detected. The gain level works like a “volume control on an audio device”, the higher the gain setting the more amplified the echoes will
be.
A low trigger and a high gain are ideal in obtaining good sensitivity, but if the trigger is set too low or the gain too high “noises” are generated in the transducer that can result in erroneous or unstable measurements. Transducers that are defective, badly worn or of poor quality will also produce noises.
By default the unit always indicates whether the transducer is coupled (measuring) or if the transducer is not coupled (in the air). But when the analyzer option located in the Ultrasonic menu (page 47) is
Chapter 2 21
Page 34
Important
Material defects that are represented by the coupling icons serve only as a warning and should not be interpreted as definitive results. The specific area where the possibility of a material abnormality is detected should then be further inspected to confirm or discard the warning.
activated the unit now also alerts the operator on the possible condition of the test piece being measured. When the material condition analyzer - available exclusively from dmq - is enabled, the QS5 can display any of the following coupling icons:
Figure 2.10: Coupling icon representations
The analyzer only works in Echo-1 mode. When measuring in Echo-2 or thru-paint mode the coupling icon will simply indicate that you are measuring in that mode (represented with two parallel lines).
The vertical bar located on the far right side of the measuring screen shows the echo signal intensity or energy level received by the transducer giving you valuable information such as:
The coupling condition between the transducer and the test
piece.
The reflective (bottom) test piece surface condition where
22 Chapter 2
low energy levels would typically indicate the possible presence of corrosion or pitting.
Page 35
The material attenuation level that allows the operator to
detect metal treatments that have an effect on the ultrasonic characteristics of a material.
The QS5 has two modes for representing the energy absorbed by the transducer:
No DAC: Shows only the energy level as provided by the transducer
With DAC: Compensates the energy level based on the measured thickness. Use DAC to detect material conditions such as thermal treatments and test piece defects easier regardless of the measured thickness. DAC is activated in the Ultrasonic menu (page 47).
The higher the energy level reflected from the test piece bottom, the higher the energy level will be on the vertical bar. This is an indication of good coupling. Low energy indicates any of the following: poor coupling (not enough coupling gel), high attenuation materials (treated metals and some plastics), an irregular test piece bottom surface (possible corrosion or pitting) or if DAC is disabled it may be reflecting the test piece thickness.
Figure 2.11: Vertical bar energy level representations
Chapter 2 23
Page 36
Tips
Sudden intensity loses usually mean that there are defects in the test piece related to corrosion or pitting.
When measuring thick test pieces with DAC turned off, the energy level decreases because a greater distance (thickness) usually results in more attenuated echoes being received by the transducer.
When a test piece made of steel that will typically return high energy levels, is returning low energy, this usually means that it was subject to thermal treatment. To confirm or discard this possibility, measure the velocity of the test piece. Steel that has been thermally treated will have a different material velocity.
In case of low energy levels always make sure you have good coupling and that you used enough coupling gel to avoid erroneous interpretations.
When measuring in Echo-2 or thru-paint the energy level may be low because the unit uses the energy produced by the second echo which is normally smaller and more attenuated when compared to the first echo used in normal Echo-1 measurements.
24 Chapter 2
Page 37
The QS5 E and QS5 DLE gauges can measure thru-coatingings such as paint or epoxy so you no longer need to scrape and remove the coating in a test piece. The unit now “skips” the coating and measures only the actual test piece thickness.
To enable thru-coating or thru-paint mode (Echo-2) touch the (Echo) key from the measuring screen and the unit will display the mode you are in. (Echo-1 normal, or Echo-2 coating (thru-paint).
Switch between Echo modes touching and wait until the window indicating the mode you are in disappears. This means that the unit is now set for the Echo mode you selected.
Figure 2.12: Change of Echo windows (normal or coating)
Multi-echo units use separate ultrasonic configurations for each echo mode. This means that when switching between Echo-1 and Echo-2 modes the gain, trigger and voltage of the pulse emission are maintained for each mode. In the example below the gain level for Echo-1 is 75 and for Echo-2 it is 63. These gain levels will not change as you switch modes.
Chapter 2 25
Page 38
Figure 2.13: Ultrasonic settings on each echo mode
All unit menu screens display an icon indicating the echo mode that the unit is operating in.
Figure 2.14: Echo mode icon in menu screens
In Echo-2 mode the unit does not need to be calibrated and calibration is actually blocked because when operating with multiple echoes the delay in the transducer does not affect the thickness measurement.
To obtain precise and reliable measurements in Echo-2 mode the following considerations and limitations should be considered:
The test piece coating should not exceed 2mm. This may vary
according to the attenuation of the coating material.
The thickness of the actual test piece should not exceed
25mm.
26 Chapter 2
Page 39
Lose or poor adherence between the coating and the test
piece can result in scattering of ultrasonic waves making it impossible to measure.
Use sufficient coupling gel for stable coupling in order to
avoid erroneous or incorrect readings.
Always use dmq hi-damp transducers to measure in Echo-2
mode. Standard probes will not work correctly in this mode.
Chapter 2 27
Page 40
The instructions explained in this chapter apply to all of the menus in the unit.
To scroll QS5 menu options use the - cursor keys. When you reach the end of the menu and move to the next menu option it becomes circular as shown below.
Figure 3.1: Example of how a circular menu works
To select a menu option touch and to exit and return to the
previous menu touch . To go to the measuring screen touch from the main menu.
Figure 3.2: The “Home” key from the measuring screen
28 Chapter 3
Page 41
The text editor is used to input, change, and delete; letters, numbers and symbols.
Figure 3.3: Alphanumeric editor screens
1. Selected key
2. Cursor
3. Text to be edited
4. Virtual keyboard
Use the cursor keys on the unit to scroll the virtual keyboard until you
find the character you want to use and touch to select.
Touch to move to the upper case virtual keyboard and to the numbers and symbols keyboard as seen on figure 3.3.
There are 4 keys that are common to all virtual keyboard screens:
Figure 3.4: Common virtual keyboard keys
1. Move cursor to the left
2. Move cursor to the right
Chapter 3 29
Page 42
3. Erase the character on which the cursor is on
4. Enter and exit
Figure 3.5: Quick access keys for the virtual keyboard editor
Touch the key to open the quick access keyboard to the most commonly used virtual keyboard keys. Each virtual key corresponds to a key on the front panel of the unit as follows:
: Move the cursor to the left
: Erase the character on which the cursor is on
: Move the cursor to the right
: Undo
: Erase the character to the left of the cursor
: Enter and exit
To close the quick access keyboard and continue using the virtual
keyboard touch .
30 Chapter 3
Page 43
The main menu is the first list of options that appears when you exit the measure screen and it includes the most important settings.
Touch from the measuring screen to access this menu.
Figure 3.6: Main menu
Note: The options of Configure, Ultrasonic and Memory are explained later on in this manual.
Touch on Unit in the main menu to open the list of available units.
Use the - keys to scroll the menu.
Touch to select the unit and touch to save and exit this menu.
Figure 3.7: Units menu
The QS5 has high and low alarms that alert the operator when the measurement is greater than the value set for the high alarm and or when the measurement falls below the value set for the low alarm.
Touch on Alarms to open the alarm menu options.
Chapter 3 31
Page 44
Note
The material condition analyzer operates only in Echo-1 mode. In Echo-2 or thru-paint mode the flaw alarm is automatically disabled.
Touch on High or Low to open the numbers editor where you can set alarm values using the cursor keys.
Touch to save the alarm value that you entered and to return to the previous menu.
Figure 3.8: Alarms menu options
Any of the following alarm types can be selected:
Flaw: The alarm is activated when the material condition analyzer detects the presence of a possible defect such as corrosion or pitting. Enabling this option will activate the Analyzer (page 49).
Beep: Audible intermittent alarm type.
Screen: Visible alarm that causes measurements to be displayed in
dotted instead of regular numbers.
Light: Visible alarm that activates the display backlight illumination causing it to flash.
32 Chapter 3
Page 45
Here you can set high and low values (thickness range) that will be represented on the vertical axis of the graphic BScan measure screen.
Figure 3.9: BScan range and graphic representation
Touch on BScan Range to open the menu.
Touch on High or Low to open the numbers editor where you
can set a value using the cursor keys. Touch to save and return to the previous menu.
Touch on Language (which is also identified with a flag) to view available language options.
Use the cursor keys to scroll available
language options and touch to select.
Touch to save and exit this menu.
Figure 3.10: Language menu options
Chapter 3 33
Page 46
Note
The information required to obtain model upgrade licenses is included in this option.
Select Unit Info to view information including owner data as well as the software version that your unit is running.
To view different unit information screens touch the - keys.
To return to the main menu touch .
Touch on the Configure option to open the general configuration options menu.
Use the - keys to scroll the menu.
Touch to select any of the menu options.
Touch to exit and to return to the previous menu.
34 Chapter 3
Figure 3.11: General configuration options menu
Page 47
Choose Set Clock to open the time editor and use the cursor keys to
set the time. Touch to save and to enter the date editor screen.
Set the date and touch to save and exit.
Figure 3.12: Time and date editing screens
Choose Clock Format to open the menu that allows you to set the time format (12Hs or 24Hs) and the date format (D/M/Y – M/D/Y).
Touch on the option you wish to select
and touch to save and exit.
Figure 3.13: Date and time format menu options
Chapter 3 35
Page 48
Tips
If the unit will be operated using security gloves we recommend that the sensitivity level be raised.
To make the keypad “harder” simply lower the sensitivity
level.
In applications where the front of the unit may be exposed to water and or vapors the sensitivity should be lowered.
This option allows you to set the keypad sensitivity. The higher the number is that you set here, the more sensitive the keypad will be.
Touch on Key Sens. and use the - keys to change the
keypad sensitivity. Touch to save and the keypad will already be working with the new sensitivity level.
Figure 3.14: Key sensitivity and confirmation screens
To confirm the change in sensitivity touch . If you touch any other key or the counter that appears on your screen reaches 0.0, the sensitivity will return to its previous setting.
The factory default setting is 50. Under special conditions we suggest that the sensitivity level be changed.
36 Chapter 3
Page 49
The unit will shutdown automatically if no key is touched or no measurement is made when you set the auto-off time.
Touch on AutoOff to set the time before the unit automatically shutdown.
Touch the - keys to set the time and
touch to save and exit.
Touch to exit without making changes.
Figure 3.15: Auto-off time setting screen
Use this setting to adjust the screen contrast making it lighter or darker where 1 is the lightest and 32 is the darkest.
Touch on Contrast and use the - keys to change the contrast on your screen.
Touch to save or touch to exit without making changes.
Figure 3.16: Screen contrast settings
Chapter 3 37
Page 50
Tips
Contrast on LCD screens can change with temperature. Use the contrast option to compensate for changes caused by temperature in order to maintain optimal viewing conditions.
Beep refers to the sounds that the unit makes when keys are touched and when the audible alarm is activated.
Touch to enable or disable the beep option.
The introduction screen is the first screen that you would see (if this option is enabled) after the unit is turned on. It includes owner information such as name, telephone number and e-mail.
Touch to enable or disable this option.
This option allows you to enter owner information (the info that would appear on the introduction screen).
Touch on Set ID, enter the password (the factory default
password is 12345) and touch to access user info menu options.
38 Chapter 3
Page 51
Notes
When showing user information, the e-mail address is displayed as "Name@Domain".
Important
The factory default password is 12345. You can change this password after adding your user information.
Figure 3.17: Enter password to access the user info menu
The user information that can be added or changed at anytime includes the following:
Name: Set or change the owner name.
Phone: Set or change the telephone number.
Name@: Set or change the e-mail (before the @).
@Domain: Set or change the domain for the e-mail (after the @).
Edit Pass: Allows you to change the password needed to access this
menu.
Chapter 3 39
Page 52
Certain configuration options on your QS5 unit can be locked in order to avoid unwanted changes. Use of the locking options allow a supervisor to optimize unit configuration settings required for a specific test and then pass the unit on to an operator for him or her to conduct the actual measurements knowing that the unit has been properly configured and that the settings cannot be changed.
Touch on Locks, enter your password and then touch again to view the configuration options that can be locked.
Figure 3.18: Enter password and options lock menu
Each option is followed by a lock indicating whether the feature is locked (closed lock) or unlocked (open lock).
Touch on each of the following options to lock or unlock:
Calibrate: Lock or unlock unit calibration options.
Configure: Lock or unlock ultrasonic configuration options (trigger
and gain).
Datalogger: Lock or unlock Datalogger configuration options.
40 Chapter 3
Page 53
QS5 models can be changed with software licenses that you can purchase from dmq. If you want to purchase an upgrade license we will need to know the following information:
Unit model Unit serial number The type of license that you would like to purchase
Touch on Licenses to view all licenses available for your unit (checkmarks indicate active licenses).
Figure 3.19: Licenses screen
To enter the new license number that you purchased from dmq touch
or to exit and return to the previous menu touch .
Use the cursor keys to enter the license
number and touch to save.
Figure 3.20: Enter new license number screen
After you enter the new license number the unit will respond with one of the following messages:
Chapter 3 41
Page 54
Figure 3.21: Response messages after a license is entered
If the license number that you entered is correct the unit will show an updated license screen where the newly purchased license appears followed by a checkmark.
Choose Set Default to return to the original factory default general configuration options.
Touch and a confirmation screen will appear.
Touch to confirm and return to the
previous menu or touch to exit without making changes.
Figure 3.22: Set factory default settings confirmation screen
42 Chapter 3
Page 55
Select Ultrasonic from the main menu to display measuring configuration options.
Setting these options correctly is critical in attaining reliable measurements.
Figure 3.23: Measuring configuration options menu
Touch on the Veloc. option to change the material velocity so that it matches the material that will be measured.
Use the cursor keys to change the velocity and touch to enter. To
exit without making changes touch .
Touch the when editing velocity and a screen will appear with a list of the most commonly used materials and their velocities. Use the
cursor keys to scroll the list and touch to select or touch to exit and return to the edit velocity field.
Chapter 3 43
Page 56
Figure 3.24: Change material velocity and material velocity list
Touch on Measure to open the measure modes options menu.
Figure 3.25: Measure mode options menu and set nominal value
The modes in which measurements are represented are:
Absolute: the unit displays the real measured value
Different.: the displayed value is the result obtained from calculating:
Where the nominal value is a reference value for the test piece
Touch on Nominal to set this value and touch to save. To
exit without making changes touch .
Differential = Real value – Nominal value
44 Chapter 3
Page 57
Touch on Gain to change this value and touch to save. To
exit without making changes touch . For additional information on setting the gain refer to "Changing trigger and gain" (page 20).
Touch on Trigger to change this value and touch to save.
To exit without making changes touch . For additional information on setting the gain refer to "Changing trigger and gain" (page 20).
Touch on Refresh to open the refresh rate options menu (the speed in which measurements will be taken).
Use the - keys to scroll menu
options and touch to select.
To exit without making changes touch .
Figure 3.26: Refresh rate speed options
The factory default for this setting is 4 Hz (4 measurements per second) which is enough for most applications. When looking for additional details or for scanning a specific test piece section we recommend switching to higher refresh rates.
Chapter 3 45
Page 58
Notes
A higher refresh rate increases energy consumption therefore reducing battery life.
Storing measurements in the Datalogger at high speeds means the size of the memory is reduced a lot quicker. Measuring at 20Hz over one minute is equal on average, to 1200 measurements.
Touch on Freq to open the transducer frequency rate options menu.
Use the - keys to scroll menu options and touch to
select.
To exit without making changes touch .
Figure 3.27: Transducer frequency options
Most standard transducers are 5MHz. When a different frequency transducer is used you should select the correct frequency in this menu.
Touch on Coat Type to open the menu options for the attenuation level of the coating type when measuring in Echo-2 or thru-coatinging.
46 Chapter 3
Page 59
Use the - keys to choose any of the
two available menu options and touch to select. To exit without making changes
touch .
Figure 3.28: Type of coating options
To measure thru-coatingings that are less than 1mm thick, select the Low Loss option. For high attenuation coatings such as epoxy (typically used in shipyards) and for coatings greater than 1mm thick, select the High Loss option.
Even though the coating type option is available for all QS5 models this setting only affects the QS5 E and QS5 DLE models that can operate in Echo-2 mode or thru-coatinging.
Touch on Cal. to open the unit calibration menu options in the measuring screen.
Use the - keys to choose any of the
two available menu options and touch to select. To exit without making changes
touch .
Figure 3.29: Unit calibration menu options
The QS5 has two calibration modes:
Auto: Calibration is automatically done when coupling the transducer using the calibration disk located in the front of the unit (page 15).
Chapter 3 47
Page 60
2-Point: Calibration is done with one or two test piece samples of a known thickness, and made of the same material that you will be measuring (page 17). The 2-Point calibration will usually result in more precise thickness measurements.
When you enable the Hold last option the last measured value will be displayed on the unit screen even when the transducer is not coupled. When hold last is disabled and the transducer is not coupled the unit display will read "--.--".
Touch to enable or disable this option.
Figure 3.30: Measuring screens with hold last enabled or disabled
When the DAC Bar option is enabled the unit compensates the energy level on the vertical energy bar based on measured thickness (page 22). When this option is disabled the energy level on the vertical bar is shown just as provided by the transducer.
Touch to enable or disable this option.
48 Chapter 3
Page 61
Important
Material defects that are represented by the coupling icons serve only as a warning and should not be interpreted as definitive results. The specific area where the possibility of a material abnormality is detected should then be further inspected to confirm or discard the warning.
Note
Activating the flaw alarm (page 32) will automatically enable the analyzer option.
Enable Analyzer so that the unit will alert the operator on the presence of possible defects with icons that appear on the measuring screen (page 21). When the analyzer option is disabled the coupling icon only indicates whether the transducer is coupled (measuring) or not coupled (in the “air”).
Touch to enable or disable this option.
Chapter 3 49
Page 62
Note
Enabling the continuous measuring monitor slightly increases battery consumption. By default this option is disabled.
When Nums Zoom is enabled the numerical measuring screen displays thickness measurements in larger numbers and only the velocity and minimum value are displayed. The minimum value is reset by touching
the key.
Touch to enable the Nums Zoom option and touch again to disable.
Figure 3.31: Normal and Nums Zoom numerical measure screens
When AutoMain is enabled the unit continuously monitors the transducer so that when the transducer is coupled it automatically switches to the measuring screen regardless of the menu you are in.
Touch to enable or disable this option.
50 Chapter 3
Page 63
Touch on Pulser to adjust the transducers pulser excitation level.
Touch the - keys to adjust the
tension and touch to save and exit. To
exit without changes touch .
Figure 3.32: Transducer pulser excitation tension editor screen
Just like with the trigger and gain settings the pulser tension is set separately for Echo-1 and Echo-2 modes.
Considerations when adjusting the pulser excitation tension: Higher excitation or tension levels mean that more energy is applied to the emissions crystal in the transducer making it easier to measure in high attenuation and large or thick materials.
When measuring small or thin materials high tensions produce saturated echoes that can affect measurement results particularly when operating in Echo-2 mode.
Avoid high tensions particularly when using small diameter transducers as this could damage the transducer. For 5MHz transducers the tension should not exceed 150V. Large diameter transducers of a lower frequency can be set to higher tensions for better performance.
Chapter 3 51
Page 64
Choose Set Default to return to the original factory default ultrasonic configuration settings.
Touch and a confirmation screen will appear.
Touch to confirm and return to the
previous menu or touch to exit without making changes.
Figure 3.22: Set factory default settings confirmation screen
52 Chapter 3
Page 65
In order to optimize the use of the Datalogger in your QS5 DL and QS5 DLE it is important that you understand how data is organized.
In the QS5 data is stored in 8 individual files with alphanumeric names. Each file contains a grid with columns and rows. And each grid contains columns (identified with consecutive letters; A, B, …AA, AB…) that store an “N” number of values or measurements (batch). Each value is identified with a column letter and a row number.
Figure 4.1: How data is organized in the Datalogger
Select Memory from the main menu to view all menu options for the Datalogger. The options in this menu will be different depending on your QS5 model. The QS5 B and QS5 E store values that can only be viewed on the unit screen but cannot be sent to a PC or a printer.
Chapter 4 53
Page 66
Figure 4.2: Memory menu options based on unit model
Touch on Files and use the cursor keys to navigate the list of files in the unit.
Figure 4.3: Creating a file
Select a file that appears as Empty (available file) and touch to name the file with up to 10 alphanumeric characters.
After you enter a name a checkmark will appear next to it meaning that a new file has been created and is ready to be used.
54 Chapter 4
Page 67
Remember that only one file can be open at any given time so whenever a new file is created, if a file was already open, the file that was open will be closed. Once a file has been closed, it cannot be reopened and new values can no longer be stored. In closed files, values can only be viewed.
Whenever you create a new file, and another file is already open, a warning screen will ask if you want to close the last file.
Touch to confirm that you want to close the open file to create (open) a new one or
touch to cancel and return to the menu.
Figure 4.4: Close file confirmation screen
Touch on any closed file and a menu will open that displays options for that file.
Figure 4.5: Single file menu options
Chapter 4 55
Page 68
Touch on View Data to view the
contents of the file. Touch to exit the file.
Figure 4.6: File view in a grid format
To move inside the grid use the cursor keys.
Touch on a value to open a histogram that includes that same value as well as the values included in the foregoing cells within that column.
Touch to exit and return to the grid.
Figure 4.7: Histogram representation
High and low alarm values appear in the histogram as dotted horizontal lines.
Touch to open the quick access menu that allows you to go directly to a position inside the grid.
Select the Row, Column, and Cell using the - - keys.
56 Chapter 4
Page 69
Figure 4.8: Quick access menu options in a grid
Row: Enter a row number using the cursor keys so that when you
touch the grid will position itself directly on that row.
Column: Enter the column letter using the cursor keys so that when
you touch the grid will position itself on that column.
Cell: This is a combination of (column and row) so that you can go to a specific cell after you enter the row number and column letter.
Touch the key in the histogram to open the quick access menu that allows you to obtain statistical information on the batch of values being displayed.
Chapter 4 57
Page 70
Figure 4.9: Quick access menu options in a histogram
Error: Displays the number of errors and error percentage values in the
batch based on the high and low alarm settings.
Stats: View statistical information for the batch including Min., Max., Range and Mean values.
Stats+: View the standard deviation and the percentage (coefficient variation) for the batch.
Distr: Displays the distribution of values for the batch in a graphic format (dynamic batch).
58 Chapter 4
Page 71
Touch on a SCAN cell to open the quick access menu that allows you to easily move inside the scan and to view statistical information on the scan.
Figure 4.10: Quick access menu in a scan graphic
Goto: Access a specific position within the scan by entering its location
(List).
List: Opens the list of measured values in the scan (use the - to scroll)
Stats: View the options of Stats, Error and Distr.
Stats: View statistical information including Max., Min., Mean and
Range of all values within that SCAN cell.
Chapter 4 59
Page 72
Error: Displays the number of errors and error percentage values in the scan based on the high and low alarm settings.
Distr: displays the distribution of values for the scan in a graphic format.
Touch on Send File to send (units with Datalogger only) a single file to a PC using Windows HyperTerminal, or to a printer using an RS232 connection. This option does not work in USB mode.
Touch on Rename to open the text editor that allows you to change the file name.
Touch on Size to view the number of stored values within a single file (and the size of the file as a percentage of total unit memory). You can also view the date and time in which the file was created.
Touch on All Files to open the menu for actions that will affect all files stored in the unit memory.
Figure 4.11: Actions over all files menu
60 Chapter 4
Page 73
Touch on Send All to send (units with Datalogger only) all files stored in the unit memory to a PC using Windows HyperTerminal, or to a printer using an RS232 connection. This option does not work in USB mode.
The Erase All action permanently deletes all files stored in the unit memory and recovers 100% of the memory capacity.
Before files are deleted, a screen will be displayed asking you to confirm or to cancel this action.
Touch to cancel and return to the
previous menu or touch to begin deleting all files.
Figure 4.12: Erase all confirmation screen
When the erase all action has been confirmed the following screens will be displayed:
Figure 4.13: Erase all progress screen
Chapter 4 61
Page 74
X
Datalogger is disabled (no open file)
M
Manual capture mode
U
Single auto capture mode
C
Continuous auto capture mode
S
SCAN block capture mode
The QS5 Datalogger measure screen includes information and options to assist you in storing data.
Figure 4.14: Information in the Datalogger measure screen
1. Datalogger status indicator icon followed by any of the
following letters:
2. Storage type (Grid or SCAN). If measurements are being
stored in a grid, “Grid” will be displayed. When a SCAN is stored the display will read “Scan: N” where N is the position
of the scan block in the grid.
3. Name of the open file
4. Position of the stored value within the grid (Column, Row)
62 Chapter 4
Page 75
Notes
The Datalogger status indicator can be viewed on all QS5 measure screens as well as in the graphic screen. If no file is open the Datalogger will display “No File” and the unit will automatically switch to the Min. Max. numerical measure screen mode.
Touch on the measuring screen to open the quick memory menu.
Figure 4.15: Quick memory menu options
In each quick memory menu screen two options are displayed at the
bottom of each screen. To access these options use the -
keys. The key maintains its normal functionality and the key is used to switch quick screens as seen on figure 4.15.
Options in the first quick memory screen:
Batch: Close the open batch and open a new one.
Delete: Erase the last stored value
Options in the second quick memory screen:
Scan: Opens a SCAN block. If one is already open no action will be made.
Capt: Changes the capture modes (Manual – Single – Continuous). Allows you exit the Scan mode as well.
Chapter 4 63
Page 76
Touch on Connect to enter the “waiting to connect” mode.
Touch to exit and cancel the connection.
Figure 4.16: Connecting to a PC
With the unit in “waiting to connect” make sure that the USB or the
RS232 cable (depending on the type of connecting cable that you are using) is properly connected to both the unit and the PC and click on <Connect> in DataCenter.
When a successful connection is established the files in your unit memory will appear in DataCenter. To view their contents simply double click on each file.
For additional information on dmq DataCenter software refer to the manual included in the CD that you received with your QS5 or download the manual at http://www.demeq.com/Download.html
64 Chapter 4
Page 77
In the memory menu touch on Configure to open the Datalogger configuration options menu.
Figure 4.17: Datalogger configuration options menu
The first two options in the configure Datalogger menu are Mode and Send which allow you to select how the unit will communicate with a
PC or a printer.
Figure 4.18: Configure communications options
Touch on Mode to select the type of connection.
USB: Select USB to connect to a PC using a USB cable (included). You must already have DataCenter installed in your PC.
Chapter 4 65
Page 78
Notes
The print option can only be used with RS232 printers. Files cannot be sent directly to a USB printer.
When you use DataCenter to connect, your selection of PC or Print will not affect communication.
RS232: Select RS232 to connect to a PC or a printer using an RS232 cable (optional).
Touch on Send to select whether you will send files to a PC or a printer.
PC: When using an RS232 cable, the unit sends data in an optimal format for Windows HyperTerminal (38400-8-N-1).
Print: Using an RS232 cable and printer, the unit sends data in an optimal format for mini-printers of 40 columns (9600/8-N-1).
Touch on Capture to select the mode in which values will be stored in the Datalogger.
66 Chapter 4
Figure 4.19: Memory capture modes menu
Page 79
The QS5 has the following capture modes:
Manual: Touch the key to store values.
Single: Each time the transducer is coupled a thickness measurement
is stored. In this mode values can also be stored using the key.
Continuous: When the transducer is coupled thickness measurements are continuously stored at a rate of:
Refresh / N
Where refresh refers to the number of values per second that the unit reads (Page 45) and N refers to the Store Every (N number of values) that can be set on the Advanced Datalogger menu options (Page 67).
Touch on Histogram to set the number of values for each batch when the Fix Batch option is enabled.
When the Fix Batch option is enabled the unit automatically closes a batch and opens a new one when the number of values that you set for the histogram is reached. When the Fix Batch option is disabled values are stored until the batch is manually closed. You can close a batch by entering the “Batch” option by touching the “Mem” key while you are in the measuring screen.
Touch the key to enable or to disable the Fix Batch option.
Chapter 4 67
Page 80
Touch on Advanced to access the advanced configuration options menu for the Datalogger.
Figure 4.20: Advanced Datalogger configuration options menu
Touch on Store Every to set the number of measurements done by the unit before a value (measurement) is actually stored in the Datalogger when Continuous capture mode is enabled (Page 66).
When you enable history the Datalogger will register the unit settings for each measurement that is stored. Information such as gain, trigger, velocity and calibration settings are stored along with the thickness value. This data can only be viewed in DataCenter. When history is enabled the unit memory capacity decreases as more data is stored.
Touch on History to enable or disable this option and touch
to exit.
68 Chapter 4
Page 81
Touch on Clock to set the time and date options that will be recorded by the Datalogger each time a new batch is opened.
Touch - to scroll all menu options
and touch to select and exit.
Touch to exit without making changes.
Figure 4.21: Record date and time in the Datalogger
Off: The Datalogger does not record date and time.
Date: Each time a new batch is opened the Datalogger records the
date.
Time: Each time a new batch is opened the Datalogger records the time.
Both: Each time a new batch is opened the Datalogger records the date and time.
Chapter 4 69
Page 82
Do not measure outside the thickness range specified for the transducer that you are using.
Use just enough coupling gel to ensure stable measurements. Avoid using excessive coupling gel as it may be added to the thickness value of the actual test piece being measured.
Always do the zero calibration using the calibration disk located in the front of the unit particularly when changing transducers.
Have a reference test piece of the same material that you will be measuring and know the velocity of the material or its thickness value so that you can precisely determine the velocity of the material based on its thickness.
Gain and trigger levels are set from factory with the optimal values that correspond to the transducer that was shipped with your unit. When a transducer is changed gain and trigger levels should be adjusted to optimize transducer functionality.
Appendix
Page 83
Measuring principal
Pulse-Echo / Echo-Echo
Measuring range
0.6 mm to 500 mm (Pulse-Echo)
2.5 mm a 20 mm (Echo-Echo)
Transducer frequency
2 to 10 MHz
Capture speeds
4, 8 and 20 Hz
Pulser tension
Adjustable 20 V to 210 V
V-Path
Automatic correction
Units
Millimeters, Inches and Microseconds
Resolution
0.01 mm from 0.60 to 99.99 mm
0.1 mm from 100.0 mm and above
0.001 in from 0.040 to 9.999
0.01 in from 10.00 in and above
Material velocity
100 to 19999 m/S
3.937 to 787.3 in/ms
Calibration
Auto, 1-Point and 2-Point
Alarms
Defects, Min., Max. Audible and visual
Languages
English, Spanish, Portuguese
Datalogger
Capacity of 32000 values organized in up to 8 files with alphanumeric names. Manual, Single, Continuous capture modes. Record Date and Time. View data including statistics in grid or graphic formats.
Connection to PC
USB native or RS232 (optional)
Appendix
Page 84
Display
Graphic LCD 128 x 64 pixels with LED backlight illumination and contrast adjustment.
Keypad
Touch-sense with no mechanical parts and sensitivity adjustment.
Battery life
100 hours with 2 each type AA batteries
Operating temp.
- 10°C to + 50°C
Enclosure
High impact ABS with rubber sides. Size is 78 x 117 x 24 mm.
Weight
200 g with batteries
Appendix
Page 85
The QS5 was developed and manufactured for years of trouble free operation and even though the unit does not require special care the following precautions should be considered:
Avoid contact with corrosive and abrasive substances. Do not clean the unit with solvents. Do not leave the unit display exposed to direct solar light for
prolonged periods of time as this could damage the display.
Remove the batteries if the unit will be stored for an
extended period of time.
Remove the transducer using the connectors and not the
cable.
Do not twist or strangle the transducer cables. Do not expose the unit to temperatures below -10°C / 14°F or
above 50°C / 122°F.
Appendix
Page 86
dmq part No.
Description
QSM 300
Small high impact carrying case
QSS 201
2Mhz transducer – 15 mm diameter
QSS 501
5Mhz standard transducer – 10 mm diameter
QS5 501 HD
5MHz Hi-Damp transducer – Thru-Paint
QSS 701
7Mhz transducer – 5 mm diameter
QSR 161
6 step calibration block (mm)
QSR 141
4 step calibration block (in)
QSG 001
Coupling gel
QSL 102
QS5 B to QS5 DL upgrade license
QSL 301
QS5 E to QS5 DLE upgrade license
QAC 002
RS232 cable to connect to a PC
QAC 003
RS232 cable to connect to a printer
For additional information and accessories for your QS5 visit www.demeq.com/Accessories-QS.html
Appendix
Page 87
Error 1
Internal Error
Cause
Internal Error
Solutions
Shutdown the unit, wait a few seconds, and power back on. Contact dmq.
Error 2
Attempt to store a value over an existing value.
Cause
Improper unit shutdown (Example: Removing batteries) and powering the unit back on to store values in the Datalogger.
Solution
Download Datalogger values to PC or printer and erase memory.
Error messages may eventually open on your unit screen and are informational only. If one of these messages opens on your display follow the instructions described below and if the problem persists please send us a detailed report at www.demeq.com/form_Support.html
Figure A.2: System error message
If a message with a different number appears on your unit display please contact dmq immediately.
Appendix
Page 88
Our website is a powerful customer support tool where you will find the latest information as it relates to your QS5 including:
Application notes Manuals and brochures Software updates Model upgrade license information New accessories
To download software updates to your QS5 you must have dmq DataCenter installed on your PC. To download the latest updates for your unit refer to www.demeq.com/Download.html
Our service department is committed to providing prompt and courteous service. Should you encounter any trouble with your QS5 please send us a detailed description of your problem to www.demeq.com/form_Support.html
Appendix
Page 89
P/N QSD-501-ENG-Rev 3 ©2012 dmq
77 Capitulo 4
Loading...