Walk-In Temp / Door /Alarm / Light Module with Mounting Box and Wiring
1. GENERAL DESCRIPTION
Model XWA11V-KIT, 100x64 mm format, is a microprocessor based light and alarm management controller, suitable for
temperature monitoring in a walk-in cooler or freezer. It is provided with two (2) relay outputs that supply 120 Volts to
control lights and signal an alarm. It is also provided with one (1) NTC probe input for temperature measurement. To
ease installation on existing walk-in coolers the unit has been installed in a VBOX pre-wired with an outboard buzzer,
magnetic door switch (MDS), a 9V battery harness, a pigtail harness for the unit power and the light. There is also a five
pin port that allows the user to program the parameter list with a “Hot Key”.
Note: The default settings are listed in the back of this manual. They are set for Coolers (Medium
Temp). For Freezers (Low Temp) you MUST Change the ALL and ALU settings.
We have responded to many requests to make a change in the way the light key operates. It now
overrides the Digital Input to allow a manual OFF prior to the “Lht” being exhausted.
See Section 4.0
2. GENERAL WARNINGS
2.1 PLEASE READ BEFORE USING THIS PRODUCT
This manu
reference.
The instrument shall not be used for purposes different from those described hereunder. It cannot be used
as a safety device.
Check the applicatio n limits before proceeding
2.2 SAFETY PRECAUTIONS **READ ME FIRST!**
Check if the supply voltage is correct before connecting the instrument.
Do not expose the back of the instrument to w ater or moisture: use the controll er only within the operating limits
avoiding sudden temperature changes with high atmospheric humidity to prevent the form ation of condensation.
Be sure to seal any J-Box with RTV sealant to prevent cold and moisture intrusion.
Warning: disconnect all electrical connections before performing any maintenance operation.
Fit the probe where it is not damaged by the end-user. The instrument must not be opened.
In case of failure or faulty operation send the instrument back to the distributor (see address) with a detailed
description of the fault.
Consider the maximum current that can be applied to each relay (see Technic al Data).
Ensure that the wires for probes, loads and the power supply are separated and far enough from each other,
without crossing or intertwining or you may get bad temp readings.
The box supplied is not suitable for exterior use. If a water tight box is required co ntact the distributor.
al is part of the product and should be kept close to the instrument for easy and quick
Press and hold this key for more than 5 seconds to turn the controller OFF. (onF=y)
Press and hold this key for more than 1 second to turn the controller back ON.
mming Mode press to select a parameter or to confirm an operation.
dIXELV8.8-KIT
HIGH
Press to se
Press to se
LOW
In Progra
e the HIGH Temp ALARM (ALU parameter)
e the LOW Temp ALARM (ALL parameter)
mming Mode press to browse parameter codes.
Press to increase the displayed value.
Press to mute the buzzer (+ relay) when an ALARM is happening.
Hot keyprogramming: with the instrument on, insert the hot key and then press
the UP button.
In Progra
mming Mode press to browse parameter codes.
Press to decreases the displayed value.
Each LED function is described in the following table:
LED MODE Function
N
O
ON
ARM signalin
AL
The light is on
°CONCelsius operation
°FONFahrenheit operation
4. TEMP ALARMS SETTING
4.1 HOW TO SET THE MIN TEMPERATURE ALARM
To modify the minimum (LOW) Temp ALARM: hold the
the minimum Temp alarm is displayed.
Change the value using the UP and DOWN keys.
Press
the SET key to confirm the new value an
4.2 HOW TO SET THE MAX TEMPERATURE ALARM
To modify the max (HIGH) Temp ALARM: hold the+
max Temp alarm is displa
Change the value using the UP and DOWN keys.
Press
the SET key to confirm the new value an
yed.
d exit.
d exit.
dIXELV8.8-KIT
g
LOW
keys pressed for 3 seconds until
+
HIGH
keys pressed for 3 seconds until the
5. PROGRAMMING
5.1 HOW TO CHANGE A PARAMETER VALUE MAIN MENU
1.Enter the Programming Mode by pressing the SET and DOWN key for 3 seconds ( and will start
blinking).
Select the required parameter. By using the UP or DOWN
2.
3.
Press the “SET” key to display its value (now only the
4.Use
Press
“UP” or “DOWN” to change its valu
“SET” to store the new value and move to the following parameter.
e.
To exit: Press SET + UP or wait 15 seconds without pressing a ke
NOTE: the se
t value is stored even when the procedure is exited, by waiting the time-out to expire.
5.2 THE HIDDEN MENU (PR2)
The hidden menu includes all the parameters of the instrument.
5.2.1 HOW TO ENTER THE HIDDEN MENU (PR2)
Enter the Programming Mode by pressing the Set + down key for 3seconds (
1.
When a parameter is displayed, release and re-press the SET + down for more than 7seco
2.
3. The
4. Select the re
5.
6. Use
7. Press
Pr2 label will be displayed immediately followed from the HY parameter. NOW YOU ARE IN THE
HID
DEN MENU.
quired parameter as above
Press the “SET” key to display its value (Now only the
“UP” or “down” to change its value.
“SET” to store the new value and move to the following parameter.
To exit: Press SET + up or wait 15seconds without pressing a key.
NOTE: the set value is stored even when the procedure is exited by waiting the time-out to expire.
5.2.2 HOW TO MOVE A PARAMETER FROM THE HIDDEN MENU TO THE FIRST LEVEL AND
VICEVERSA
Each parameter present in the HIDDEN MENU can be removed or put into “THE FIRST LEVEL” (user
level) by pressing “SET + down”.
In HIDDEN MENU when a parameter is present in First Level the decimal point LED is on.
1. Keep the UP and DOWN key
The “POF” message will be displayed and the keyboard will be locked
2.
only to see th
3. If a key is presse
5.4 TO UNLOCK THE KEYBOARD
Keep the UP and DOWN keys pressed together for more than 3seconds. The “PON” message is displayed
e Set Point or the MAX o MIN Temp stored
d more than 3seconds the “POF” message will be displayed.
s pressed for more than 3 second
s.
. At this point it will be possible
6. LIGHT MANAGEMENT
6.1 TIMED REGULATION: I1L = Y
With i1L = y the light remains on at least for the LHt parameter.
The LHt timer is re-initialized every time the light button is pushed or the door is opened.
With LHt=0 the light remains on until the light button is pushed again or when the door is closed.
The light is switched on every time one of the following conditions happens:
the door is open (i1F = dor)*DOOR SWITCH PRESENT
the presence sensor is activated (i2F = LHt)*OCCUPANCY SENSO
the light button is pushed
The light is switched off when all the following conditions happen:
the LHt time
the door is closed (i1F = dor)*DOOR SWITCH PRE
the presence sensor is de-activated (i2F = LHt)*OCCUPANCY SENSOR PR
Light button regulation: i1L
r is exhausted
SENT
= n
R PRESENT
ESENT
FLH (opera
end of the LHt time as a warning that the lights are about to turn off. (This should only be used for
incandescent or LED lighting only!)
6.2 MANUAL OVERRIDE VIA LIGHT BUTTON OF XWA11V VERSION: 8.8 OR HIGHER ONLY
With the intorduction of V8.8 it is now possible to overrided the light on time with a press of the light button on
the face of the XWA11V. Regardless of mode in which the light had been turned on the Light button will always
overide a door switch activation or occupancy sensor input.
mode) The lights will flash (for 2 minutes) every 20 seconds for the FLH time (0-5 min) before the
7. INSTALLATION AND MOUNTING
7.1 MOUNTING OF XWA11V
The XWA11V must be mounted on a vertical
panel using the VBOX wall mount box. The
back of the box has two screw mounting
holes and a larger hole for the electrical
connections to pass through.
The Temp range allowed for correct
operation is 32 – 140°F. Avoid installation in
places subject to strong vibrations, corrosive
gases, excessive dirt or humidity. The same
recommendations apply to probes.
The Door Switch Magnet should be
mounted no more than ¾” of an inch from
the switch.
The instrument is provided with ¼” (.250”) fast-on terminal blocks to connect the digital and analog inputs.
Relay
s and power supply have a Fast-on connection ¼” (.250”). For supply connections use 14
r copper or CU wire only, rated to at least 90°C (194°F). Before connecting cables make sure the po
large
ply complies with the instrument requirements. Separate the probe cables from the power supply an
sup
output po
load
Not
8.1 PROBE CONNECTIONS
The probe shall be mounted with the bulb upwards to prevent damage due to casual liquid infiltration. It is
recommended to place the thermostat probe away from air streams to correctly measure the average room
temperature. The probe can be extended up to 300 ft. Check calibration when running long lengths over
100ft.
8.2 BATTERY BACK-UP
Battery Back-up will automatically power the XWA display in case of a power failure for up to 36 hours.
Display will flash when battery runs low. (Battery not included)
wer connections. Do not exceed the maximum current allowed on each relay and in case of heavie
s use a suitable extern
e:Maximum current allowed for all the loads is 15A.
al relay.
AWG or
wer
d
9. USE OF THE PROGRAMMING “HOT KEY”
The XWA11V units can UPLOAD or DOWNLOAD the parameter list from its own E2 internal memory to the
“H
OT KEY” and vice versa.
r
9.1 TO PROGRAM A HOT KEY FROM THE INSTRUMENT (UPLOAD)
1. Program one controller with the front keypad.
2. When the controller is ON
by flashing “End”
3. Push
4. Turn
9.2 HOW TO PROGRAM AN INSTRUMENT USING A HOT KEY (DOWNLO AD )
1. Turn OFF the instrument.
2. Insert a programmed “Hot Key” into the 5 PIN receptacle and then turn the Controlle
3.
4.
5. Remove
“SET” key and the End will stop flashing.
OFF
the instrument and remove the “Hot Key”, then turn it ON again.
NOTE: the “Er
to restart the upload again or remove the “Hot key” to abort the operatio
Automatically the parameter list of the “Hot Key” is downloaded into the Controller memory, the “doL
messag
After 10 seconds the instrument will restart operation
NOTE the message “Err” is displayed for failed programming. In this case turn th
you want to restart the d
r” message is displayed for failed programming. In this case pu
e is blinking followed a by flashing “End”.
the “Hot Key”.
, insert the “Hot key” and push UP key; the "uPL" message appears followed a
sh again
n.
with the new parameters.
e unit off and then on if
ownload again or remove the “Hot key” to abort the operatio
key if you want
r ON.
n.
10. ALARM SIGNALS
Message Mode Cause Outputs
“P1” Flashing Probe failure Alarm output ON
“PoF” Flashing (3s) Keyboard locked Not changed
“Pon” Flashing (3s) Keyboard un-locked Not changed
“HA”Alternated with temp Maximum temperature alarm Alarm output ON
“LA” Alternated with temp Minimum temperature alarm Alarm output ON
“dA” Alternated with temp Door switch alarm Alarm output ON
“EA” Alternated with temp External alarm Alarm output ON
“PAn” Alternated with temp Serious external alarm Alarm output ON
dEFAlternated with temp Defrost is running Not changed
The alarm message is displayed until the alarm condition is reset.
”
10.1 SILENCING BUZZER
Once the alarm signal is detected the buzzer can be silenced by pressing the (UP) key. (tbA=y)
10.2 ALARM RECOVERY
Probe alarms: “P1” (probe1 faulty), they automatically stop 10seconds after the probe restarts normal
operation. Check connections before replacing the probe.
Temperature° alarms “HA” and “LA” automatically stop as soon as the thermometer temperature returns to
normal operating values or when the defrost starts.
Door switch alarm “dA” stops as soon as the door is closed.
External alarms “EAL”, “BAL”, “dEF” stop as soon as the external digital input is disabled.
dIXELV8.8-KIT
11. TECHNICAL DATA
Housing: self extinguishing ABS
Case: face; 100x64 mm (3.94”x2.52”). Depth; 45.5mm
Mounting
Frontal protection: IP65 (CONTROL ONLY) the box is not wa
Conn
Po
wer supply: 120Vac
Power absorption:
Ambient T
Display: 3 digits, red LED, 14.2 mm (.56”) hi
Inputs: 1 NT
Digital inputs: 2 free voltage
Relay outputs: Voltage Supplied Relay Contac
Other output: alarm buzzer
Data storing: on the non-volatile memory (EEPROM).
Measuring and regulation range:
Resolution: 1 °F
Accuracy: ±1 °F
: J-box or wall-mount in suitable enclosure
ections: ¼” (.250”) fast-on for power, ¼” (.250”) fast-on for probes and Digi
UT Display update 60
OnF Off function enabling y
ALU High temperature alarm setting (med temp / low temp)
ALL Low temperature alarm setting (med temp / low temp)
AFH Temperature alarm differential 2
ALd Temperature alarm delay 30
dAo Delay of temperature alarm at start-up 1.3
EdA Alarm delay at the end of defrost 30
dot Delay of temperature alarm after closing the door
LHt Light timer 15
FLH Light Flashing Timer (time before light goes out that it will flash)
doA Open door alarm delay 15
oA1 First relay configuration
oA2 Second relay configuration LHt
AOP Alarm relay polarity CL
i1P Digital input 1 polarity OP
i1L Door switch to turn light ON y
i1F Digital input 1 operating mode dor
i2P Digital input 2 polarity cL
i2F Digital input 2 operating mode PAn
did Time interval/delay for digital input alarm 0
tbA Alarm relay disabling n
PbC Kind of probe ntc
dP1 Real temperature Probe 1 Pr2
CF T measurement unit: °C = Celsius; °F = Fahrenheit. When the measurement unit is cha nged the Set
Point and the values of some parameters have to be modified.
rES Resolution (for °C): (in = 1°C; dE = 0.1 °C) allows decimal point display.
Ut Display update: The time delay of the Temperature readout (0 to 255 seconds)
onF Off function enabling: n = off function disabled; y = off function enabled;
ALU High Temp alarm setting: (ALL to 150°C or 302°F); when this temperature is reached and after the
ALd delay time the HA alarm is enabled.
ALL Low Temp alarm setting: (- 50°C or -58°F to ALU) when this temperature is reached and after the
ALd delay time, the LA alarm is enabled.
AFH Temp alarm differential: (0.1 to 25.5°C; 1 to 45°F) differential for temperature alarm Set Point and fan
regulation Set Point, always a positive value
ALd Temp alarm delay: (0 to 255 min) time interval between the detection of an alarm condition and the
corresponding alarm signaling.
dAO Delay of Temp alarm at start-up: (0min to 23hrs, 50min) time interval between the detection of the
temperature alarm condition after the instrument power on and the alarm signaling.
EdA Alarm delay at the end of defrosts: (0 to 255 min) Time interval b etween the detection of the
temperature alarm condition at the end of defrost and the alarm signaling.
dot Delay of Temp alarm after closing the door: (0
condition after closing the door.
tat probe calibration: (-12.0 to12.0°C/ -21 to 21°F) allows to adjust possible offset of the
255 min) Time delay to signal the temperature alarm
8
Page 9
Operating Instructions XWA11V
LHt Light timer: (0-255 min) Amount of time the light will stay on after pressing the light switch on the
keyboard or after the door is closed.
FLH Light Flashing: (0-5 min) The light will “double flash” every 20 seconds during the FLH time period before
turning off after the LHt time. This should only be enabled for use with incandesant or LED lighting!
doA Open door alarm delay:(0-255 min) delay between the detection of the open door condition and its
alarm signaling: the flashing message “dA” is displayed.
oA1 First relay configuration: (14-15): ALr = alarm; LHt = light; onF = on/off relay
oA2 Second relay configuration: (14-16): ALr = alarm; LHt = light; onF = on/off relay
AOP Alarm relay polarity: cL = closing contacts (Relay will close on alarm giving voltage output);
oP = opening contacts (Relay will open on alarm no voltage output).
i1P Digital input 1 polarity (1-2): CL : the digital input is activated by closing the contact;
OP: the digital input is activated by opening the contact
i1L Door switch to turn light ON(1-2): (y / n) To turn the light ON automatically when the door is open. The
light will turn off based on LHt . Keyboard switch must be turned ON first.
i1FDigital input 1 operating mode(1-2): EAL = external alarm; dor = door switch; dFr = A defrost is
running; LHt = keep light ON (signal from occupancy sensor) override LHt.;
i2P Digital input 2 polarity (1-3): CL : the digital input is activated by closing the contact;
OP: the digital input is activated by opening the contact
i2F Digital input 2 operating mode: configure the digital input function:
EAL = External alarm;
PAn =Panic alarm;
dFr = A defrost is running; (need external CT’s)
LHt = Keep light ON (signal from occupancy sensor) overrides LHt.
did Time interval/delay for digital input alarm:(0-255 min.) If I2F=EAL or PAn (external alarms), “did”
parameter defines the time delay between the detection and the successive signaling of the alarm.
tbA Alarm relay & Buzzer disabling: (y ; no)
Pbc Type of probe (PTC, NTC)
dP1 Probe 1 Temperature
rEL Software release for internal use.
Ptb Parameter table code: read only.