OM111 11
P1972-OM-367 Rev 6 Chapter 5
CHAPTER 5 FAULT FINDING
FAULT PROBABLE CAUSE REMEDY
5.1 Mask free flowing on inlet
side.
Free flow control knob
screwed too far in.
Back off free flow control
knob.
Inlet gas supply pressure
set too high.
Reduce gas supply
pressure (6-12 bar over
chamber depth).
Inlet valve diaphragm
levers incorrectly set after
maintenance.
Reset diaphragm inlet
levers.
Damage to inlet valve
seat.
Replace inlet valve stem
assembly.
Damage to inlet valve
nipple.
Replace inlet valve nipple.
5.2 Mask continually sucking
chamber gas on exhaust
side.
Exhaust valve diaphragm
levers incorrectly set after
maintenance.
Reset diaphragm exhaust
levers.
Damage to exhaust valve
seat.
Replace exhaust valve
stem assembly.
Damage to exhaust valve
nipple.
Replace exhaust valve
nipple.
Mask positioned face
down whilst removed from
diver.
Position mask with oral
nasal facing downward.
5.3 Difficulty inhaling. Free flow control knob
incorrectly set.
Adjust free flow control
knob clockwise for easy
breathing.
Inlet valve diaphragm
levers incorrectly set after
maintenance.
Reset diaphragm inlet
levers.
Incorrect supply pressure. Reset supply pressure to
recommended pressures
for depth (6-12 bar over
chamber).