He/O2 Analyzer
| 31 |
Defects and removal thereof
Error 5 Device is too warm
• Thetemperatureoftheanalyzeristoohigh(forinstance,ifitwasexposedtosunlightor
stored in a car heated by sunlight); the measurement may be inaccurate. Cool down the
analyzer below 40 degrees Celsius (104°F).
Warning Battery low
• Thebatteryisalmostempty,ithastobereplacedorusetheexternalpowersource.
Error 6 External power overvoltage
• Theallowedvoltageoftheexternalpowersourcewasexceeded.Thismaybecausedby
mains excessive voltage or by the use of an inappropriate or faulty adapter. After sending
this message, the analyzer will immediately shut down to avoid the overheating of voltage
stabilizer circuits. When the defect of the adapter is removed or if the battery is used, the
analyzer can be reactivated.
13.2 Instrument malfunctions
Helium content not displayed
• Theprobemeasuresthespeedofheliuminbothdirections.Ifthesedatadier,measu-
ring is deemed incorrect and the helium content is not displayed. This situation can be
encountered if the measured gas flows through the probe too fast (lower flow speeds are
compensated for by the calculation), or there is significant background noise (for example,
in the vicinity of a running compressor or gas motor).
• Lowerthegasowspeed(thisappliestotheanalyzernotconnectedusingthegenuine
sampler), move farther from the source of the noise.
The instrument shows less than 0 % or more than 100 % helium in the mixture
• Thedisplayedvaluearisesdirectlyfromthemeasuredandcalculatedvalueandhasinno
way been altered in the background. The possible measuring error is symmetrical and, for
example, when measuring a mixture without any helium content, 0.2 % or −0.2 % helium
can be displayed. Accordingly, when measuring pure helium, e.g. 99.8 % or 100.2 % can
be displayed. If the deviation is within the measuring tolerance, this is not the result of
a defect.
The instrument shows less than 0 % or more than 100 % of oxygen in the mixture
• Thedisplayedvaluearisesdirectlyfromthemeasuredandcalculatedvalueandhasinno
way been altered in the background. The possible measuring error is symmetrical and, for
example, when measuring a mixture without any oxygen content, 0.5 % or −0.5 % oxygen
can be displayed. Accordingly, when measuring pure oxygen, e.g. 99.5 % or 100.5 % can
be displayed. If the deviation is within the measuring tolerance, this is not the result of
a defect. In the case of a greater deviation, recalibrate the oxygen sensor. If the problem
persists, replace the sensor.