SAFE OPERATING PROCEDURES AND PROPER USE OF THE EQUIPMENT
ARE THE RESPONSIBILITY OF THE USER OF THIS SYSTEM.
Directed Energy, Inc. (DEI) provides information on its products and associated
hazards, but it assumes no responsibility for the after sale operation and safety
practices.
Users are responsible for reading and understanding all safety precautions listed
in this manual prior to operation of the equipment. There are no user serviceable
parts in this product, please contact the factory for service information.
This quick-start guide provides a step-by-step guide to manually operating the
PDG-2515. Please refer to the appropriate sections of the manual for additional
instructions and guidelines. The SAFETY warnings in the manual must be read
and understood prior to operating the PDG-2515.
What’s In The Box?
1ea. PDG-2515 Digital Delay & Pulse Generator
1ea. Power Cord
2ea. 50 Ohm Coaxial Output Cable
1ea. Operating Manual
Unpacking
Carefully remove each component listed above. Your PDG-2515 and
accessories were inspected at the factory prior to shipment. Inspect the PDG2515, power cord and output cables for signs of shipping damage. If any damage
is noted contact the factory prior to operating the equipment.
Setup
Connect OUT1 and/or OUT2 to an appropriate load using 50 Ohm coaxial cable
(included).
Plug the AC power cord into the rear power entry module of the PDG-2515, and
into an appropriate AC socket.
Power Up
Turn on the PDG-2515 by pressing the power button located above the power
entry module in the rear of the unit.
The front panel display will read initialize and display the software and hardware
versions internal to the unit on the front panel. After a few moments the unit will
finish initialization and the first screen/menu that will appear is the menu for
CHANNEL 1. Note: The buttons and encoder on the front panel will have no
effect as the unit comes up from initialization.
Front Panel Controls & Connections
Encoder
This is for varying the parameters such as Frequency, Pulse Width, Delay,
Trigger Source, etc.
The output button enables/disables the pulse output.
Menu Button
This button is for navigating the PDG-2515 menus.
Enter Button
Used for variables that will be permanently stored in the unit or require
setup times internal to the PDG-2515. When a question mark appears
after the variable, the enter button will “confirm” or in some cases save
that particular setting.
Up Arrow Button
The UP ARROW button allows the user to go up in any particular menu to
change selections. If the user is at the top selection it will wrap back
around to the bottom selection.
Down Arrow Button
The DOWN ARROW button allows the user to go down in any particular
menu to change selections. If the user is at the bottom selection it will
wrap back around to the top selection.
X10 Button
The X10 button is used when varying parameters that have exponents
that can change by a power of 10. By pressing the X10 button the user
can increment the exponent within its acceptable range. This allows the
user to quickly set the parameters, without having to excessively scroll the
encoder wheel. The X10 button functions with the following commands:
Set Pulse Width 1 and 2
Set Delay 1 and 2
Set Frequency (*Note: if depressed while dialing in frequency it will
increase the frequency at a much higher rate)
Set Counters/Bursts/Packets
The X10 button is effectively a 10X multiplier. For example, when
setting frequency, the resolution of the encoder wheel is three digits
(i.e. 1.00Hz, 10.0Hz, 100Hz, 1.00KHz). Therefore without adjusting
the range, the encoder wheel must be turned many times to
increase the frequency from 1Hz to 100Hz. However if the
frequency is set to 1.00Hz, if the X10 button is pressed once, it
changes to 10.0Hz. If it is pressed twice, it changes to 100Hz.
Setting the delay and pulse width is similar to setting frequency. For
example, to set the delay to 300ns, set the delay to 25ns, press the
X10 button once to increment it to 250ns, then rotate the encoder
wheel to increase it to 300ns. At this point, pressing the X10 button
again would increase the delay to 3.00us, pressing it again
increments it to 30.0us, etc.
An additional function is that if it is held while changing ANY
frequency value, after multiplying the value by a factor of ten, if the
encoder is moved it will increase/decrease at a much faster rate.
This makes the switching frequency much quicker and easier.
PW/PRF Button
This is for varying the parameters Frequency and Pulse Width and
toggling between the two settings quickly. If you are in any menu that is
not a CHANNEL MENU, it will flip you immediately to the CHANNEL 1
menu and the FREQUENCY setting.
Out 1 and Out 2
These are the pulse output connectors. The outputs of the PDG-2515
should be connected to the inputs of the receiving instrument or system
using 50 Ohm coaxial cable. For optimal pulse fidelity, these outputs
should be terminated into 50 Ohms.
Sync Out
This is a 5V signal into 1 meg-Ohm or 2.5V signal into 50 Ohms that is
synchronous with the frequency generator trigger. This is designed to
assist the user in triggering when monitoring the output signals with an
oscilloscope, or for use in triggering an external instrument.
Trigger In
This is an input signal into 1KOhms used for external Pulse Repetition
Frequency (PRF) triggering. It is designed to allow the user to trigger
multiple PDG-2515 units from a single pulse generator. This allows the
user to create synchronous waveforms from multiple units.
Power LED
Indicates that the power switch on the rear of the unit is in the ‘on’ position
and that AC power is applied to the unit.
Front Panel Controls & Connections
AC Power Entry Module
The driver is grounded through the grounding conductor of the AC power
cord. To avoid electrical shock, plug the driver into a properly wired
receptacle before making connection to any input or output connectors.
Use only a power cord that is in good condition.
The PDG-2515 accepts input voltages of 90-264VAC, 47-63Hz.
Example Setup
The default setup is 1 kHz, with zero delay and 100 us pulse on both channels.
The pulse in not inverted and the pulse mode is in continuous pulse generation
mode which means that pulse count is not available. By pressing the OUTPUT
button on the front panel you will see these pulses on the respective outputs.
Turning the encoder with the pulses enabled and the unit will sweep the
frequency on the fly, hitting the X10 button will bring you to 10 times the
frequency. Pressing the down arrow twice will highlight the pulse width selection.
Turn the encoder and the width will sweep. Try this with the polarity setting; it will
change on the fly. Turn off the pulses.
Press MENU to continue to the settings for CHANNEL 2. Its default is the same
as CHANNEL 1.
Hitting MENU again will bring up the OUTPUT MENU. This menu allows the unit
to change pulse modes, disable and enable the outputs (depending on the pulse
mode), choose internal or external trigger (depending on the pulse mode) and
choose whether the sync is based off CHANNEL 1 or 2.
Hitting the arrow keys allows the user to decide which settings on the current
menu they would like to adjust. Press the up arrow once to scroll around to the
bottom of the menu or press the down arrow 5 times. The contrast setting will be
highlighted.
Turn the encoder to adjust the contrast. The contrast will adjust “on the fly” and a
question mark will appear at the end of the contrast. Hit ENTER and save the
contrast saving into the non-volatile memory of the unit, this setting will saved
into the unit and keep this in memory even after power-down.
The PDG-2515 is a two channel digital delay/pulse generator. Each channel
offers independently adjustable delay and width, with 25 nanosecond resolution.
Up to five user configurations may be stored in internal non-volatile memory, and
recalled through the easy to understand front panel.
Functions & Features
Delay And Width and
Duty Cycle.
Frequency
Resolution
Trigger Sources
Pulse Generator
Mode
Pulse Count Mode
Divide By N Mode
Burst Mode
Each channel provides independently adjustable
delay and width, with 25ns resolution. The duty
cycle is limited based on the frequency. The duty
cycle is limited approximately by the equation:
MAX DC = 97-[(1E-5)*FREQ].
Pulse repetition frequency from single shot to
continuous pulses from 100 mHz to 3.0 MHz.
100mHZ to 1.0 MHz -> 3 digits
1.0 MHz + -> 2 digits
Internal, External, Single Shot.
Generates pulses at the rate set by the internal
frequency generator, or by an external trigger.
Outputs pulses at the rate set by the internal
frequency generator or external trigger and stops
after N pulses.
Channel 1 operates in continuous mode, channel 2
pulses every Nth pulse on channel 1.
Generates a burst of pulses, repeating at a
frequency.
Counted Burst Mode
Store And Recall
Generates a burst of pulses, and stops after N
bursts.
Stores up to 5 user configurations in non-volatile
memory.
Output Channels Two (2) outputs, each with independent digitally controlled delay
and pulse width
Delay Range 0 to 99.9 Seconds
Pulse Width Range 25 Nanoseconds to 99.9 Seconds
Delay And Pulse Width Resolution 25 Nanoseconds below 1 Microsecond
50 Nanoseconds from 1 Microsecond to 10 Microseconds
3 Digits Above 10 Microseconds
ccuracy (Delay and Pulse Width) ±2%
Time Base 120MHz Frequency Synthesizer
Delay Jitter (1st Sigma) TBD
Trigger Delay (External Trigger to SYNC (T0)
Output
Internal Throughput Delay (SYNC (T0) To Output 1
Or Output 2 Rising Edge)
Output Rise & Fall Times (Into 50Ω)
Over/undershoot <5%
mplitude +4V into 50W
Output Connectors BNC, Front Panel
INTERNAL RATE GENERATOR
Modes Continuous, Single-Shot, Burst, External Trigger, Divide By N,
Rate Single-Shot or 100 millihertz to 3MHz (Internal)
Resolution 3 Digits (<1MHz), 2 Digits (>1MHz)
Modes), 1
Rate Jitter (Burst Mode), 1st Sigma TBD
Burst and Pulse Count Mode 2 to 65,535 Pulses per Burst or Count
EXTERNAL TRIGGER INPUT
Type Positive Edge Trigger
Input Amplitude
Input Impedance
Minimum Trigger Pulse Width 25ns
Input Trigger Connector BNC, Front Panel
T0 (SYNC) OUTPUT
T0 (Sync) Output
Connector BNC, Front Panel
COMPUTER INTERFACE
Interface Type N/A
Interface Connector N/A
GENERAL
Internal Non-Volatile Storage 5 Configurations
Operating Temperature Range 0°C to +40°C
Cooling Requirements
Input AC Power 90-264VAC, 47-63Hz, 28W Maximum
Dimensions (H X W X D) 3 ½” H x 8” W x 6 1/4” D (8.9cm H x 20.3cm W x 15.9cm D)
Weight 2 lbs. (.9kg) Approximate
st
Sigma
75 Nanoseconds
<25 Nanoseconds
<5 Nanoseconds
Pulse Count
Single-Shot to 5MHz (External Trigger)
The versatile functionality of the PDG-2515 makes it ideal for the needs of
numerous applications as a small bench top instrument. The delay resolution and
width resolution of each channel allow a unique ability to offset control signals,
coupled with ability to logically manipulate the outputs together to form gate
signals or specific waveforms make automation of testing and control an easier
process.
Safety
Normal use of test equipment exposes the user to a certain amount of danger
from electrical shock because testing may be performed with exposed voltages
present, dictating the use of caution when operating or servicing this equipment.
The following is a summary of general safety precautions that must be observed
during all phases of operation and repair of the PDG-2515.
Operating Safety Summary
The safety information contained in this summary is for both operating and
servicing personnel. Specific warnings may be found throughout this manual, but
may not appear in this summary.
Power Source
The PDG-2515 is designed to operate from a regulated power source. To
assure proper system operation the input voltage should not vary by more
than 10% from the recommended specification. The PDG-2515 accepts
input voltages from 90VAC to 264VAC at 47Hz to 63Hz.
A protective grounding connection by way of the grounding conductor in
the AC power cord is essential.
Grounding
The PDG-2515 is grounded through the grounding conductor of the AC
power cord. To avoid electrical shock, plug the PDG-2515 into a
properly wired receptacle before making connection to any input or
output connectors. Use only a power cord that is in good condition.
Cover Removal
To avoid personal injury, do not remove the covers. Do not operate the
PDG-2515 while the covers are removed. Operating the unit with the top
cover removed voids any and all warranties associated with the PDG-
2515.
General Operating Precautions
Do not open the unit while the system is in operation. Use caution when
handling or removing the output cables while the unit is operating. Turn off
equipment when making connections in high voltage or high current
circuits. Discharge high voltage capacitors after removing power.
Servicing Safety Summary
DEI strongly recommends that all repairs and adjustments be performed by
factory qualified personnel. DEI will not be responsible for personal injury or
damage to the pulse generator that occurs during repair by any party other than
the factory. Any repairs, adjustments or modifications made by anyone other than
authorized DEI personnel voids the factory warranty.
PREPARATION FOR USE
General
After unpacking, initial inspection and electrical installation procedures
should be performed to assure that the unit is in good working order. If it is
determined that the unit is damaged, the carrier should be notified
immediately. Repair problems should be directed to the service
department,
Directed Energy, Inc. (DEI),
Fort Collins, Colorado.
Telephone: (970) 493-1901
FAX: (970) 493-1903
EMAIL: deiinfo@directedenergy.com
Initial Inspection
• Inspect unit for exterior mechanical damage.
• Inspect power input cord and input power module for obvious signs of
damage.
Electrical Installation
Standard units are shipped ready for use with a nominal 90 - 264 VAC input.
Operation
This section provides basic operating instructions for the PDG-2515 hardware.
Operating Tips
1. The two output channels have synchronized frequencies. Pulse width and
pulse delay settings of one channel are dependent on the frequency of the
other channel. When changing frequency be aware that both pulse width
and delay are within range on both channels.
2. When using the 10X button to increase the rate at which the frequency
changes, when it is first depressed the value will change by 10X, further
change is accomplished by using the encoder while holding the 10X
button down.
Power-Up Procedures
The unit should be powered up using the following procedures:
Connect OUT 1 and/or OUT 2 to an appropriate load using 50 Ohm coaxial cable
(See Section 6.0).
Plug the AC power cord into the PDG-2515, and into an appropriate AC socket.
Turn on the PDG-2515 by pressing the power button on the rear of the unit
located above the power entry .
Configure the unit from the front panel. See the quick-start guide in Section 1.0 of
this manual for specific operating instructions, and Sections 8 and 9 for detailed
operation instructions.
FRONT PANEL OPERATING INSTRUCTIONS
THE CHANNEL MENU
The CHANNEL MENU will display which channel menu the unit is in on
the top of the display. Changing the settings is done by simply rotating the
encoder or pressing the X10 button. When the setting has reached it’s
maximum value on the encoder it will stop increasing the value as the
encoder turns. The X10 button on the other hand will set the value to the
nearest minimum exponent value when the maximum value has been
reached (example: 500 kHz will go to 500 mHz). This maximum value is
set by the internal limitations of the unit and also the pulse settings (ie:
pulse delay and duty cycle). In order the change which setting to adjust
the up and down arrows will highlight the current setting. In the CHANNEL
MENU all settings can be done while the unit is enabled, all settings can
“sweep”. Changing the pulse width and delay settings while enabled will
not effect the pulse, the change will take place on the next available clock
cycle internally. The frequency needs to be stopped and adjusted and
then turned back on internally. This means that there could be some delay
while sweeping frequency before the next frequency takes hold, this is
more noticeable as the frequency is near to slower limits while the unit
syncs up again to the internal oscillator. The polarity sweep will take hold
immediately, as will the disable and enable functions. The polarity function
will flip the polarity of the output. Pulse count will only be available in pulse
count and burst modes.
THE OUTPUT MENU
The OUTPUT MENU is where the unit can change its mode, disable a
channel, change the trigger, change the base of the sync, and also adjust
the contrast settings of the unit. The operating modes will be discussed in
more detail in section 9.0. Enabling and disabling a channel can also
happen on the fly in pulse generation and Divide By N modes. In pulse
count mode this needs to be set before hand so the unit does not sit in an
unknown state if it has not finished it’s count set, therefore, the unit will not
allow it to be changed while pulsing. In Burst mode, disabling the channel
is not available and both are enabled. When the mode is switched from
Burst mode, the enable/disable settings will return to where they were
previous to Burst mode. The trigger can be set to internal or external in
pulse generation and pulse count modes. Note that the unit will “lock up”
until it has finished it’s pulse count in external mode so the external
triggering device must give it AT LEAST as many pulses as is entered. If
this does not occur, the device can be “released” by switching modes. The
sync can also be run off channel 1 or channel 2. This is especially useful
in Divide By N mode and pulse count mode. Not that in Burst mode, the
sync is a packet period limited by at most 95% duty cycle regardless of
this setting.
The contrast setting can also be changed in this menu. If the contrast is
adjusted by using the encoder knob or using the x10 button, a ‘?’ will
appear at the end of the setting. This means that while the unit is powered
this is the setting that the unit will retain but when powered down and
powered up again the unit will revert to the last saved contrast. By
depressing the ENTER button the contrast will be saved in non-volatile
memory and the unit will come up under this contrast setting. The contrast
is has numerous settings, and a coarse and a fine adjust. The letter is a
course adjustment and the number is the fine adjust. The course adjust is
manipulated with the x10 button and the fine adjust is set by the encoder
knob.
THE LOGIC MENU
The LOGIC MENU allows the user to logically AND or OR outputs OUT1
and OUT2. LOGIC1 will output on the OUT1 channel (note: the invert
function will invert the channel AFTER the logic has been applied – this is
shown in the LOGIC AND POLARITY FUNCTIONAL DIAGRAM below).
LOGIC2 will output on the OUT2 channel.
The CONFIGS MENU allows the user to save or load up to 5 different
configurations. After selecting the appropriate configuration to either load
or save (1 through 5) hit the ENTER button and the configuration of the
unit will be saved. If the user tries to load a configuration that has not been
saved yet, the unit will send an error message of “Not Valid Config” and
nothing will be loaded. If the load or save is successful the active
configuration (ACTIVE CON: X) will reflect this with the correct setting.
*Note: if any values are changed, the active configuration will no longer be
valid and will reflect either -- or a ‘0’.
Operating Modes
Timing Diagram
The timing diagram below defines the frequency, delay and width
variables, and their temporal relationship to the T0 frequency generator
signal.
Please note that in external trigger mode, all outputs are delayed from the
leading edge of the external trigger signal.
Pulse Generator Mode
In Pulse Generator Mode, the PDG-2515 generates pulses at the
specified frequency denoted in the CHANNEL MENU (changing the
frequency on either channel will change the frequency on both channels
simultaneously – the frequencies are not independent in this mode). If
externally triggered, the rate is set by the external trigger (i.e. one pulse is
output each time a rising edge occurs on the TRIGGER input.
The delay of each of the outputs is relative to the leading edge of the
SYNC output in internal frequency generator and single shot modes. In
external trigger mode, the delay is relative to the leading edge of the
external trigger input.
The pulse width and delay for each channel are independently variable.
In Pulse Count Mode, the PDG-2515 generates a series of pulses, the
number of which is set by the PULSE CT variables on the respective
channel menu. When the number of pulses output on an output channel
equals the number set, the output of that channel is automatically disabled
and the output LED will reflect this
The PULSE CT variable for each channel can be set to any value between
1 and 65,535. In setting the PULSE CT, the x10 button will increase this
number by a factor of ten and the encoder wheel will increase/decrease
the number of pulses incrementally by one. For example, to set the count
to 1,500, first set the count to 15 using the encoder wheel, then press the
X10 button two times. The first press will increment the count to 150, and
the second to 1,500.
Pressing the OUTPUT button enables the output pulse series. The Pulse
On LED will illuminate until the pulse series is completed, at which time
the LED will extinguish. Each time the PULSE ON button is pressed, the
output pulse series will be generated.
In Internal PRF mode, the frequency (repetition rate) of the pulses are
determined by the internal frequency generator. If externally triggered, the
frequency of the pulses is determined by the frequency of the input trigger.
The delay of each of the outputs is relative to the leading edge of the T0
SYNC output.
Figure 5 Example: Pulse Count Mode
Divide By N Mode
In Divide by N Mode, OUT1 generates continuous output pulses (as in
Pulse Generator Mode), and OUT2 generates an output pulse every “Nth”
Output 1 pulse. The divisor is set through the Div Out1 By variable in the
FREQ DIV in the CHANNEL2 menu. The X10 button will multiply this
Be aware that coming out of this mode, if any variables are out of range by
going into a mode that conflicts with these settings, the channel/channels
will be set back to the default values.
Figure 6 Example 1: Divide by N Mode
Figure 7 Example 2: Divide by N Mode
Burst Mode
In Burst Mode, a set number of pulses (the burst packet) are output from
OUT1. Each burst packet is sent at the packet frequency. The burst
packet will start ONE period of the burst packet frequency after the rising
edge of the packet frequency (OUT 2 will ALWAYS output the packet
frequency at a limited duty cycle and cutoff any length of burst after that
length of time). The software will not check to check timing constraints in
this mode, meaning that if the timing if violated, the output may be
erroneous and not predictable (only while the constraints are violated – for
example if the number of bursts exceeds the time period of the packet
frequency this would violate the hardware of the unit). Violating the timing
can be a number of things such as a burst frequency larger than the
packet frequency, more pulses/delays than allowed in a single packet so
that timing overlaps, etc. The timing constraints are left up to the user’s
configurations so that the unit may be used to it’s full potential, however
the user may want to output a signal that does not violate any timing
constraints to receive a correct burst and frequency.
Figure 8 Example: Burst Mode
Single Shot Mode
In this mode, the pulse widths and the pulse delays can be set to their
maximum values without any timing constraints. After the OUTPUT button
is depressed, OUT1 and OUT2 will generate a single pulse according to
the individual channel settings.
No power LED or screen
does not come up.
Pulse exhibits excessive
ringing.
Frequency limits. Check duty cycle of EACH channel, both channels
Pulse Width limits. Check the frequency and duty cycle. Note that duty
Pulse Delay limits. Check the frequency and duty cycle. Note that duty
In external trigger pulse is
not stable or consistent
with settings.
In Burst Mode pulse is
not stable or consistent
with settings.
Configuration will not
load.
Sync output not a square
wave or has excessive
ringing.
Check power source, power switch (located on rear
of unit).
Check load impedance.
must be within the specifications of the unit to
increase the frequency. Note that duty cycle of a
channel is the addition of the pulse width and the
delay.
cycle of a channel is the addition of the pulse width
and the delay settings.
cycle of a channel is the addition of the pulse width
and the delay settings.
Check duty cycle of input trigger signal, check that
the signal frequency and trigger signal duty cycle are
within specification.
Check to make sure timing constraints are not
exceeded; duty cycle of frequency is not exceeded
with the addition of extra pulses.
Make sure a valid configuration had been previously
saved in configuration slot.
Check sync lines termination; note that sync does
not have same impedance as output lines.
This quick-start guide provides a step-by-step guide to manually operating the
PDG-2520. Please refer to the appropriate sections of the manual for additional
instructions and guidelines. The SAFETY warnings in the manual must be read
and understood prior to operating the PDG-2520.
What’s In The Box?
1ea. PDG-2520 Digital Delay & Pulse Generator
1ea. Power Cord
2ea. 50 Ohm Coaxial Output Cable
1ea. Operating Manual
Unpacking
Carefully remove each component listed above. Your PDG-2520 and
accessories were inspected at the factory prior to shipment. Inspect the PDG2520, power cord and output cables for signs of shipping damage. If any damage
is noted contact the factory prior to operating the equipment.
Setup
Connect OUTputs to an appropriate load using 50 Ohm coaxial cable (included).
Plug the AC power cord into the rear power entry module of the PDG-2520, and
into an appropriate AC socket.
Power Up
Turn on the PDG-2520 by pressing the power button located above the power
entry module in the rear of the unit.
The front panel display will read initialize and display the software and hardware
versions internal to the unit on the front panel. After a few moments the unit will
finish initialization and the first screen/menu that will appear is the menu for
CHANNEL 1. Note: The buttons and encoder on the front panel will have no
effect as the unit comes up from initialization.
Front Panel Controls & Connections
Encoder
This is for varying the parameters such as Frequency, Pulse Width, Delay,
Trigger Source, etc.
The output button enables/disables the pulse output.
Menu Button
This button is for navigating the PDG-2520 menus.
Enter Button
Used for variables that will be permanently stored in the unit or require
setup times internal to the PDG-2520. When a question mark appears
after the variable, the enter button will “confirm” or in some cases save
that particular setting.
Up Arrow Button
The UP ARROW button allows the user to go up in any particular menu to
change selections. If the user is at the top selection it will wrap back
around to the bottom selection.
Down Arrow Button
The DOWN ARROW button allows the user to go down in any particular
menu to change selections. If the user is at the bottom selection it will
wrap back around to the top selection.
X10 Button
The X10 button is used when varying parameters that have exponents
that can change by a power of 10. By pressing the X10 button the user
can increment the exponent within its acceptable range. This allows the
user to quickly set the parameters, without having to excessively scroll the
encoder wheel. The X10 button functions with the following commands:
Set Pulse Width 1, 2, 3, and 4
Set Delay 1, 2, 3, and 4
Set Frequency (*Note: if depressed while dialing in frequency it will
increase the frequency at a much higher rate)
Set Counters/Bursts/Packets
The X10 button is effectively a 10X multiplier. For example, when
setting frequency, the resolution of the encoder wheel is three digits
(i.e. 1.00Hz, 10.0Hz, 100Hz, 1.00KHz). Therefore without adjusting
the range, the encoder wheel must be turned many times to
increase the frequency from 1Hz to 100Hz. However if the
frequency is set to 1.00Hz, if the X10 button is pressed once, it
changes to 10.0Hz. If it is pressed twice, it changes to 100Hz.
Setting the delay and pulse width is similar to setting frequency. For
example, to set the delay to 300ns, set the delay to 25ns, press the
X10 button once to increment it to 250ns, then rotate the encoder
wheel to increase it to 300ns. At this point, pressing the X10 button
again would increase the delay to 3.00us, pressing it again
increments it to 30.0us, etc.
An additional function is that if it is held while changing ANY
frequency value, after multiplying the value by a factor of ten, if the
encoder is moved it will increase/decrease at a much faster rate.
This makes the switching frequency much quicker and easier.
PW/PRF Button
This is for varying the parameters Frequency and Pulse Width and
toggling between the two settings quickly. If you are in any menu that is
not a CHANNEL MENU, it will flip you immediately to the CHANNEL 1
menu and the FREQUENCY setting.
Out 1, Out 2, Out 3, Out 4
These are the pulse output connectors. The outputs of the PDG-2520
should be connected to the inputs of the receiving instrument or system
using 50 Ohm coaxial cable. For optimal pulse fidelity, these outputs
should be terminated into 50 Ohms.
Sync Out
This is a 5V signal into 1 meg-Ohm or 2.5V signal into 50 Ohms that is
synchronous with the frequency generator trigger. This is designed to
assist the user in triggering when monitoring the output signals with an
oscilloscope, or for use in triggering an external instrument.
Trigger In
This is an input signal into 1KOhms used for external Pulse Repetition
Frequency (PRF) triggering. It is designed to allow the user to trigger
multiple PDG-2520 units from a single pulse generator. This allows the
user to create synchronous waveforms from multiple units.
Power LED
Indicates that the power switch on the rear of the unit is in the ‘on’ position
and that AC power is applied to the unit.
Front Panel Controls & Connections
AC Power Entry Module
The driver is grounded through the grounding conductor of the AC power
cord. To avoid electrical shock, plug the driver into a properly wired
receptacle before making connection to any input or output connectors.
Use only a power cord that is in good condition.
The PDG-2520 accepts input voltages of 90-264VAC, 47-63Hz.
Example Setup
The default setup is 1 kHz, with zero delay and 100 us pulse on both channels.
The pulse in not inverted and the pulse mode is in continuous pulse generation
mode which means that pulse count is not available. By pressing the OUTPUT
button on the front panel you will see these pulses on the respective outputs.
Turning the encoder with the pulses enabled and the unit will sweep the
frequency on the fly, hitting the X10 button will bring you to 10 times the
frequency. Pressing the down arrow twice will highlight the pulse width selection.
Turn the encoder and the width will sweep. Try this with the polarity setting; it will
change on the fly. Turn off the pulses.
Press MENU to continue to the settings for CHANNEL 2. Its default is the same
as CHANNEL 1. Pressing the MENU button twice more will bring the user
through to the last channel menu.
Hitting MENU again will bring up the first output menu, OUTPUT MENU1. This
menu allows the unit to change pulse modes, disable and enable the outputs
(depending on the pulse mode), and choose whether the sync is based off
CHANNEL 1, 2, 3, or 4.
Hitting MENU again will bring up the second output menu, OUTPUT MENU2.
This menu allows the unit to select the trigger, adjust the contrast, and select the
logic applied to the output of CHANNEL 1, 2, 3, and 4.
Hitting the arrow keys allows the user to decide which settings on the current
menu they would like to adjust. Press the up arrow once to scroll around to the
bottom of the menu or press the down arrow 5 times. The contrast setting will be
highlighted.
Turn the encoder to adjust the contrast. The contrast will adjust “on the fly” and a
question mark will appear at the end of the contrast. Hit ENTER and save the
contrast saving into the non-volatile memory of the unit, this setting will saved
into the unit and keep this in memory even after power-down.
The PDG-2520 is a four channel digital delay/pulse generator. Each channel
offers independently adjustable delay and width, with 25 nanosecond resolution.
Up to five user configurations may be stored in internal non-volatile memory, and
recalled through the easy to understand front panel.
Functions & Features
Delay And Width and
Duty Cycle.
Frequency
Resolution
Trigger Sources
Pulse Generator
Mode
Pulse Count Mode
Divide By N Mode
Each channel provides independently adjustable
delay and width, with 25ns resolution. The duty
cycle is limited based on the frequency. The duty
cycle is limited approximately by the equation:
MAX DC = 97-[(1E-5)*FREQ].
Pulse repetition frequency from single shot to
continuous pulses from 100 mHz to 3.0 MHz.
100mHZ to 1.0 MHz -> 3 digits
1.0 MHz + -> 2 digits
Internal, External, Single Shot.
Generates pulses at the rate set by the internal
frequency generator, or by an external trigger.
Outputs pulses at the rate set by the internal
frequency generator or external trigger and stops
after N pulses.
Channel 1 and Channel 3 operates in continuous
mode. Channel 2 and Channel 4 pulses every Nth
pulse.
Burst Mode
Counted Burst Mode
Store And Recall
Generates a burst of pulses, repeating at a
frequency on Channel 1.
Generates a burst of pulses, and stops after N
bursts.
Stores up to 5 user configurations in non-volatile
memory.
Output Channels Four (4) outputs, each with independent digitally controlled delay
and pulse width
Delay Range 0 to 99.9 Seconds
Pulse Width Range 25 Nanoseconds to 99.9 Seconds
Delay And Pulse Width Resolution 25 Nanoseconds below 1 Microsecond
50 Nanoseconds from 1 Microsecond to 10 Microseconds
3 Digits Above 10 Microseconds
ccuracy (Delay and Pulse Width) ±2%
Time Base 120MHz Frequency Synthesizer
Delay Jitter (1st Sigma) TBD
Trigger Delay (External Trigger to SYNC (T0)
Output
Internal Throughput Delay (SYNC (T0) To Output 1
Or Output 2 Rising Edge)
Output Rise & Fall Times (Into 50Ω)
Over/undershoot <5%
mplitude +4V into 50W
Output Connectors BNC, Front Panel
INTERNAL RATE GENERATOR
Modes Continuous, Single-Shot, Burst, External Trigger, Divide By N,
Rate Single-Shot or 100 millihertz to 3MHz (Internal)
Resolution 3 Digits (<1MHz), 2 Digits (>1MHz)
Modes), 1
Rate Jitter (Burst Mode), 1st Sigma TBD
Burst and Pulse Count Mode 1 to 65,535 Pulses per Burst or Count
EXTERNAL TRIGGER INPUT
Type Positive Edge Trigger
Input Amplitude
Input Impedance
Minimum Trigger Pulse Width 25ns
Input Trigger Connector BNC, Front Panel
T0 (SYNC) OUTPUT
T0 (Sync) Output
Connector BNC, Front Panel
COMPUTER INTERFACE
Interface Type RS232 or GPIB
Interface Connector DB9 or GPIB connector
GENERAL
Internal Non-Volatile Storage 5 Configurations
Operating Temperature Range 0°C to +40°C
Cooling Requirements
Input AC Power 90-264VAC, 47-63Hz, 28W Maximum
Dimensions (H X W X D) 3 ½” H x 8” W x 6 1/4” D (8.9cm H x 20.3cm W x 15.9cm D)
Weight 2 lbs. (.9kg) Approximate
st
Sigma
TBD
<25 Nanoseconds
<5 Nanoseconds
Pulse Count
Single-Shot to 5MHz (External Trigger)
The versatile functionality of the PDG-2520 makes it ideal for the needs of
numerous applications as a small bench top instrument. The delay resolution and
width resolution of each channel allow a unique ability to offset control signals,
coupled with ability to logically manipulate the outputs together to form gate
signals or specific waveforms make automation of testing and control an easier
process.
Safety
Normal use of test equipment exposes the user to a certain amount of danger
from electrical shock because testing may be performed with exposed voltages
present, dictating the use of caution when operating or servicing this equipment.
The following is a summary of general safety precautions that must be observed
during all phases of operation and repair of the PDG-2520.
Operating Safety Summary
The safety information contained in this summary is for both operating and
servicing personnel. Specific warnings may be found throughout this manual, but
may not appear in this summary.
Power Source
The PDG-2520 is designed to operate from a regulated power source. To
assure proper system operation the input voltage should not vary by more
than 10% from the recommended specification. The PDG-2520 accepts
input voltages from 90VAC to 264VAC at 47Hz to 63Hz.
A protective grounding connection by way of the grounding conductor in
the AC power cord is essential.
Grounding
The PDG-2520 is grounded through the grounding conductor of the AC
power cord. To avoid electrical shock, plug the PDG-2520 into a
properly wired receptacle before making connection to any input or
output connectors. Use only a power cord that is in good condition.
Cover Removal
To avoid personal injury, do not remove the covers. Do not operate the
PDG-2520 while the covers are removed. Operating the unit with the top
cover removed voids any and all warranties associated with the PDG-
2520.
General Operating Precautions
Do not open the unit while the system is in operation. Use caution when
handling or removing the output cables while the unit is operating. Turn off
equipment when making connections in high voltage or high current
circuits. Discharge high voltage capacitors after removing power.
Servicing Safety Summary
DEI strongly recommends that all repairs and adjustments be performed by
factory qualified personnel. DEI will not be responsible for personal injury or
damage to the pulse generator that occurs during repair by any party other than
the factory. Any repairs, adjustments or modifications made by anyone other than
authorized DEI personnel voids the factory warranty.
PREPARATION FOR USE
General
After unpacking, initial inspection and electrical installation procedures
should be performed to assure that the unit is in good working order. If it is
determined that the unit is damaged, the carrier should be notified
immediately. Repair problems should be directed to the service
department,
Directed Energy, Inc. (DEI),
Fort Collins, Colorado.
Telephone: (970) 493-1901
FAX: (970) 493-1903
EMAIL: deiinfo@directedenergy.com
Initial Inspection
• Inspect unit for exterior mechanical damage.
• Inspect power input cord and input power module for obvious signs of
damage.
Electrical Installation
Standard units are shipped ready for use with a nominal 90 - 264 VAC input.
Operation
This section provides basic operating instructions for the PDG-2520 hardware.
Operating Tips
1. The four output channels have synchronized frequencies. Pulse width and
pulse delay settings of one channel are dependent on the frequency of the
other channels. When changing frequency be aware that both pulse width
and delay are within range on both channels.
2. When using the 10X button to increase the rate at which the frequency
changes, when it is first depressed the value will change by 10X, further
change is accomplished by using the encoder while holding the 10X
button down.
Power-Up Procedures
The unit should be powered up using the following procedures:
Connect OUTPUTs to an appropriate load using 50 Ohm coaxial cable (See
Section 6.0).
Plug the AC power cord into the PDG-2520, and into an appropriate AC socket.
Turn on the PDG-2520 by pressing the power button on the rear of the unit
located above the power entry .
Configure the unit from the front panel. See the quick-start guide in Section 1.0 of
this manual for specific operating instructions, and Sections 8 and 9 for detailed
operation instructions.
FRONT PANEL OPERATING INSTRUCTIONS
THE CHANNEL MENU
The CHANNEL MENU will display which channel menu the unit is in on
the top of the display. Changing the settings is done by simply rotating the
encoder or pressing the X10 button. When the setting has reached it’s
maximum value on the encoder it will stop increasing the value as the
encoder turns. The X10 button on the other hand will set the value to the
nearest minimum exponent value when the maximum value has been
reached (example: 500 kHz will go to 500 mHz). This maximum value is
set by the internal limitations of the unit and also the pulse settings (ie:
pulse delay and duty cycle). In order the change which setting to adjust
the up and down arrows will highlight the current setting. In the CHANNEL
MENU all settings can be done while the unit is enabled, all settings can
“sweep”. Changing the pulse width and delay settings while enabled will
not effect the pulse, the change will take place on the next available clock
cycle internally. The frequency needs to be stopped and adjusted and
then turned back on internally. This means that there could be some delay
while sweeping frequency before the next frequency takes hold, this is
more noticeable as the frequency is near to slower limits while the unit
syncs up again to the internal oscillator. The polarity sweep will take hold
immediately, as will the disable and enable functions. The polarity function
will flip the polarity of the output. Pulse count will only be available in pulse
count and burst modes.
OUTPUT MENU1
OUTPUT MENU1 is the menu where the unit can change its mode,
disable/enable a channel, and change the base of the sync. The operating
modes will be discussed in more detail in section 9.0. Enabling and
disabling a channel can also happen on the fly in all modes except Burst
Mode. In Burst mode, enabling/disabling the channels is not available and
Channel 1 and 2 are enabled, Channel 3 and 4 are disabled automatically.
When the mode is switched from Burst mode, the enable/disable settings
will return to where they were previous to Burst mode. The trigger can be
set to internal or external in pulse generation and pulse count modes.
Note that the unit will “lock up” until it has finished it’s pulse count in
external mode so the external triggering device must give it AT LEAST as
many pulses as is entered. If this does not occur, the device can be
“released” by switching modes. The sync can also be run off any channel
except under Burst Mode. This is especially useful in Divide By N mode
and pulse count mode. Not that in Burst mode, the sync is a packet period
limited by at most 95% duty cycle regardless of this setting.
OUTPUT MENU2
OUTPUT MENU2 is where the unit can change the trigger, adjust the
contrast settings of the unit, and set the logic applied to the output of all
Channels. The operating modes will be discussed in more detail in section
9.0. The trigger can be set to internal or external in pulse generation and
pulse count modes. Note that the unit will “lock up” until it has finished it’s
pulse count in external mode so the external triggering device must give it
AT LEAST as many pulses as is entered. If this does not occur, the device
can be “released” by switching modes. The sync can also be run off any
channel when the unit is not in Burst mode (Note that in Burst mode, the
sync is a packet period limited by at most 95% duty cycle regardless of
this setting and outputs on Channel2 as well as the SYNC output.
The contrast setting can also be changed in this menu. If the contrast is
adjusted by using the encoder knob or using the x10 button, a ‘?’ will
appear at the end of the setting. This means that while the unit is powered
this is the setting that the unit will retain but when powered down and
powered up again the unit will revert to the last saved contrast. By
depressing the ENTER button the contrast will be saved in non-volatile
memory and the unit will come up under this contrast setting. The contrast
is has numerous settings, and a coarse and a fine adjust. The letter is a
course adjustment and the number is the fine adjust. The course adjust is
manipulated with the x10 button and the fine adjust is set by the encoder
knob. Logic allows the user to logically AND or OR outputs OUT1 and
OUT2 with each other and OUT3 and OUT4 with each other. LOGIC1 will
output on the OUT1 channel (note: the invert function will invert the
channel AFTER the logic has been applied – this is shown in the LOGIC
AND POLARITY FUNCTIONAL DIAGRAM below). LOGIC2 will output on
the OUT2 channel, LOGIC3 will output on the OUT3 channel, and
LOGIC4 will output on the OUT4 channel.
The CONFIGS MENU allows the user to save or load up to 5 different
configurations. After selecting the appropriate configuration to either load
or save (1 through 5) hit the ENTER button and the configuration of the
unit will be saved. If the user tries to load a configuration that has not been
saved yet, the unit will send an error message of “Not Valid Config” and
nothing will be loaded. If the load or save is successful the active
configuration (ACTIVE CON: X) will reflect this with the correct setting.
*Note: if any values are changed, the active configuration will no longer be
valid and will reflect either -- or a ‘0’.
Operating Modes
Timing Diagram
The timing diagram below defines the frequency, delay and width
variables, and their temporal relationship to the T0 frequency generator
signal.
Please note that in external trigger mode, all outputs are delayed from the
leading edge of the external trigger signal.
Pulse Generator Mode
In Pulse Generator Mode, the PDG-2520 generates pulses at the
specified frequency denoted in the CHANNEL MENU (changing the
frequency on either channel will change the frequency on both channels
simultaneously – the frequencies are not independent in this mode). If
externally triggered, the rate is set by the external trigger (i.e. one pulse is
output each time a rising edge occurs on the TRIGGER input.
The delay of each of the outputs is relative to the leading edge of the
SYNC output in internal frequency generator and single shot modes. In
external trigger mode, the delay is relative to the leading edge of the
external trigger input.
The pulse width and delay for each channel are independently variable.
Figure 12 Example 2: Pulse Generator Mode with logic
Pulse Count Mode
In Pulse Count Mode, the PDG-2520 generates a series of pulses, the
number of which is set by the PULSE CT variables on the respective
channel menu. When the number of pulses output on an output channel
equals the number set, the output of that channel is automatically disabled
and the output LED will reflect this
The PULSE CT variable for each channel can be set to any value between
1 and 65,535. In setting the PULSE CT, the x10 button will increase this
number by a factor of ten and the encoder wheel will increase/decrease
the number of pulses incrementally by one. For example, to set the count
to 1,500, first set the count to 15 using the encoder wheel, then press the
X10 button two times. The first press will increment the count to 150, and
the second to 1,500.
Pressing the OUTPUT button enables the output pulse series. The Pulse
On LED will illuminate until the pulse series is completed, at which time
the LED will extinguish. Each time the PULSE ON button is pressed, the
output pulse series will be generated.
In Internal PRF mode, the frequency (repetition rate) of the pulses are
determined by the internal frequency generator. If externally triggered, the
frequency of the pulses is determined by the frequency of the input trigger.
The delay of each of the outputs is relative to the leading edge of the T0
SYNC output.
Figure 13 Example: Pulse Count Mode
Divide By N Mode
In Divide by N Mode, OUT1 and OUT3 generates continuous output
pulses (as in Pulse Generator Mode), and OUT2 and OUT4 generates an
output pulse every “Nth” Output pulse. The divisor is set through the Div Out1 By variable in the FREQ DIV in the CHANNEL2 menu. The X10
button will multiply this number by a factor of 10.
Be aware that coming out of this mode, if any variables are out of range by
going into a mode that conflicts with these settings, the channel/channels
will be set back to the default values. Also be aware that channels 1 and
3 stay synchronized and channels 2 and 4 must have the same value for
N. The delays and pulse widths on these channels are all independent.
In Burst Mode, a set number of pulses (the burst packet) are output from
OUT1. Each burst packet is sent at the packet frequency. The burst
packet will start ONE period of the burst packet frequency after the rising
edge of the packet frequency (OUT 2 will ALWAYS output the packet
frequency at a limited duty cycle and cutoff any length of burst after that
length of time). The software will not check to check timing constraints in
this mode, meaning that if the timing if violated, the output may be
erroneous and not predictable (only while the constraints are violated – for
example if the number of bursts exceeds the time period of the packet
frequency this would violate the hardware of the unit). Violating the timing
can be a number of things such as a burst frequency larger than the
packet frequency, more pulses/delays than allowed in a single packet so
that timing overlaps, etc. The timing constraints are left up to the user’s
configurations so that the unit may be used to it’s full potential, however
the user may want to output a signal that does not violate any timing
constraints to receive a correct burst and frequency. OUT 3 and OUT 4
are not available in this mode, they will stay at DC levels.
Figure 16 Example: Burst Mode
Single Shot Mode
In this mode, the pulse widths and the pulse delays can be set to their
maximum values without any timing constraints. After the OUTPUT button
is depressed, the OUTputs will generate a single pulse according to the
individual channel settings.
No power LED or screen
does not come up.
Pulse exhibits excessive
ringing.
Frequency limits. Check duty cycle of EACH channel, both channels
Pulse Width limits. Check the frequency and duty cycle. Note that duty
Pulse Delay limits. Check the frequency and duty cycle. Note that duty
In external trigger pulse is
not stable or consistent
with settings.
In Burst Mode pulse is
not stable or consistent
with settings.
Configuration will not
load.
Sync output not a square
wave or has excessive
ringing.
Check power source, power switch (located on rear
of unit).
Check load impedance.
must be within the specifications of the unit to
increase the frequency. Note that duty cycle of a
channel is the addition of the pulse width and the
delay.
cycle of a channel is the addition of the pulse width
and the delay settings.
cycle of a channel is the addition of the pulse width
and the delay settings.
Check duty cycle of input trigger signal, check that
the signal frequency and trigger signal duty cycle are
within specification.
Check to make sure timing constraints are not
exceeded; duty cycle of frequency is not exceeded
with the addition of extra pulses.
Make sure a valid configuration had been previously
saved in configuration slot.
Check sync lines termination; note that sync does
not have same impedance as output lines.
The PDG25xx series of pulse generators have the option of a communications
interface. The options are RS232 or GPIB interface. Both interfaces take the
same ascii based command set, although the strings are terminated differently.
The unit will reply at least one byte on every command with the exception of the
changing of the GPIB address, and the changing of the baud rate. This allows
the user to communicate as quickly and as efficiently as possible. If the PDG unit
has been purchased with communications interface, the front panel will display a
REMOTE MODE menu that displays either the current GPIB address or the
current baud rate depending on the interface installed in the unit. This is provided
so that the user may know how to set up the communications interface from a
remote terminal upon powering up the unit if the communications settings are
unknown.
RS232 Operation:
The default baud rate on the RS232 interface is 115200. Switching the baud
rate is possible, and will be saved in non-volatile memory, but the unit will not
respond in a command to make GUI interfaces easier to implement without
closing out and re-opening a COMM PORT before a reply is issued. The send
and return strings are ASCII based framed after a SCPI protocol. The string is
TERMINATED BY A RETURN (‘\r’) CHARACTER. After the unit receives the
return character the command will be parsed, processed and a reply will be
generated so that the next command may be sent. Be aware that some
commands complete significantly faster than others, the load and save
configuration command could take up to 400 milliseconds to complete due to
the overhead associated with the commands. The response from the unit will
be terminated with a ‘\r’ character as well; this makes the implementation and
design of a GUI much simpler and more efficient than a timeout.
GPIB Operation:
The default address of the GPIB interface is 1. Switching the address is
possible and will be saved in non-volatile memory, but be aware that the unit
will not respond in a command to make GUI interfaces easier to implement
without closing out and re-opening a new address on the remote interface
before a reply is issued. The send and return strings are ASCII based framed
after a SCPI protocol. The string is TERMINATED BY AN EOI line assertion –
there is no termination character, sending one will only end in a
communications error. After the unit receives the EOI the command will be
parsed, processed and a reply will be generated so that the next command
may be sent. Be aware that some commands complete significantly faster
than others, the load and save configuration command could take up to 400
milliseconds to complete due to the high overhead. The response from the
unit will be terminated with a EOI as well, this makes the implementation and
design of a GUI much simpler.
The reply generated will be either response to the data sent to the unit, as in
a query, or an error code. The error code will be prefaced by a ‘#’ and then
the ascii code itself. The codes are explained below. The error code, whether
it is 0 (which indicates NO ERROR) or a valid error is prefaced by the ‘#’
character to distinguish the data from a valid reply. For example, if the pulse
width was queried (CHNx:PULS:WIDT?), a valid response would not have a
‘#’ character as the first character, but if there was an error (such as invalid
channel, parse error, etc.) the response would send a ‘#’ character as it’s first
byte.
The ASCII commands:
The commands sent are in four letter words separated by a colon. The data is
separated by a single white space character (0x20) and is in the from
x.xxE+xx. For example, to set the frequency to 1.35 KHz you would send the
string CHAN:PULS:FREQ 1.35E+03. Each channel in the unit is distinguished
by a CHNx, where x is the channel the remote is talking to. For example, to
set the pulse width of channel 2 to 150 us, the string sent would be
CHN2:PULS:WIDT 1.50E-04.
The Command Set:
WORD DATA OPERATION COMMAND
CHAN PULS FREQ X.XXE+XX READ,WRITE SET PULSE FREQUENCY
BRST FREQ X.XXE+XX READ,WRITE SET BURST FREQUENCY
DLAY X.XXE+XX READ,WRITE SET BURST DELAY
WIDT X.XXE+XX READ,WRITE SET BURST WIDTH
CNTR integer READ,WRITE SET BURST COUNT
PKET FREQ X.XXE+XX READ,WRITE SET PACKET FREQUENCY
CNTR integer READ,WRITE SET PACKET COUNT
CHNx PULS DLAY X.XXE+XX READ,WRITE SET CHANNEL x's DELAY
WIDT X.XXE+XX READ,WRITE SET CHANNEL x's WIDTH
CNTR integer READ,WRITE SET CHANNEL x's COUNT
LOGC 0 READ,WRITE SET NO LOGIC ON CHANNEL x
1 SET "AND" LOGIC ON CHANNEL x
2 SET "OR" LOGIC ON CHANNEL x
DVAL integer READ,WRITE SET CHANNEL x's DIV N VALUE
DCYC unsigned char READ,WRITE SET CHANNEL x's DUTY CYCLE
INVT 0 READ,WRITE CHANNEL x NOT INVERTED
1 READ,WRITE INVERT CHANNEL x
STAT 0 READ,WRITE DISABLE CHANNEL x
1 READ,WRITE ENABLE CHANNEL x
OUTP MODE 0 READ,WRITE SET MODE TO PULSE GENERATOR
2 READ,WRITE SET MODE TO DIVIDE-BY-N MODE
3 READ,WRITE SET MODE TO BURST MODE
SET MODE TO INTERNAL SINGLE
4 READ,WRITE
TRIG 1 READ,WRITE
2 READ,WRITE
STAT 0 READ,WRITE DISABLE PULSING
1 READ,WRITE ENABLE PULSING
SYNC x READ,WRITE SET SYNC TO CHANNEL x
SYST CONF LOAD x WRITE LOAD SYSTEM CONFIGURATION x
SAVE x WRITE SAVE CONFIGURATION x
FIRM VERS N/A READ GET FIRMWARE VERSION OF UNIT
HDWR VERS N/A READ GET HARDWARE VERSION OF UNIT
COMM GPIB ADDR x READ,WRITE SET GPIB ADDRESS x (x = 1-30)
SERL BAUD 1 READ,WRITE SET RS232 BAUD RATE TO 9600
2 SET RS232 BAUD RATE TO 14400
3 SET RS232 BAUD RATE TO 19200
4 SET RS232 BAUD RATE TO 28800
5 SET RS232 BAUD RATE TO 38400
6 SET RS232 BAUD RATE TO 57600
7 SET RS232 BAUD RATE TO 115200
FIRM VERS N/A READ
SHOT
SET CHANNEL TO INTERNAL
TRIGGER
SET CHANNEL TO EXTERNAL
TRIGGER
# 0 0x00 NO ERROR
# 1-63 0x01-0x3A RESERVED
# 64 0x40 DATA IS OUT OF RANGE
# 65 0x41 FREQUENCY IS DC LIMITED
# 66 0x42 PULSEWIDTH IS FREQUENCY LIMITED
# 67 0x43 DELAY IS FREQUENCY LIMITED
# 68 0x44 INVALID CONFIGURATION
# 69-111 0x45-0x6A RESERVED
# 112 0x70 UNRECOGNIZED WORD
# 113 0x71 BAD STRING FORMAT
# 114 0x72 BAD DATA FORMAT
# 115 0x73 UNRECOGNIZED PARSE ERROR
# 116 0x74 BAD RESOLUTION
# 117-119 0x75-0x77 RESERVED
# 120 0x78 COMMAND NOT AVAILABLE
# 121 0x79 INTERNAL SYSTEM ERROR
# 122-255 0x80-0xFF RESERVED
We want to hear from you! Directed Energy, Inc. provides a standard line of High
Voltage Pulse Generators, Laser & Diode Drivers, and Test & Measurement
Equipment. Additionally, DEI has the capability to provide custom modules and
products to fit your volume or OEM application.
Web
Visit our web site at www.DirectedEnergy.com
Also visit these other IXYS companies
This product has been designed to provide simple trouble-free use by qualified
technical personnel, under normal operating conditions. In the event that you
require technical support and/or product service our technical support staff is
available to assist you. You can contact our team as noted previously. DEI’s
technical support is available as follows.
Product Service
• Warranty Service is provided per the terms and conditions of DEI’s
warranty described in the next section of this manual. To provide accurate
and complete warranty service please contact our technical support team
who will help you determine the nature of the problem and ensure that the
unit must be returned to DEI. You will then be given an RMA number that
our team will use to track your unit. Warranty service is free of repair or
replacement charges.
• Non-Warranty Service is provided at our standard repair rate and is billed
by the hour. There is a minimum evaluation charge for all non-warranty
repairs.
• Shipping charges for both warranty and non-warranty repairs are the
responsibility of the customer.
Technical Support
Technical support is available on our website or via email at
Directed Energy, Inc. (DEI) warrants equipment it manufactures to be free from
defects in materials and factory workmanship under conditions of normal use,
and agrees to repair or replace any standard product that fails to perform as
specified within one year after date of shipment to the original owner. OEM,
modified and custom products are warranted, as stated above, for ninety (90)
days from date of shipment to the original owner. This Warranty shall not apply to
any product that has been:
• Repaired, worked on, or altered by persons unauthorized by DEI in such a
manner as to impair, in DEI's sole judgment, the performance, stability, or
reliability of the product;
• Subjected to misuse, negligence or accident; or
• Connected, installed, adjusted, or used otherwise than in accordance with
instructions furnished by DEI.
DEI reserves the right to make any changes in the design or construction of its
products at any time, without incurring any obligation to make any change
whatever in units previously delivered.
DEI's sole obligation, and buyer's sole remedies, under this agreement shall be
limited to a refund of the purchase price, or at DEI's sole discretion, to the repair
or replacement of products in kind that prove, to DEI's satisfaction, to be
defective, when returned to the DEI factory, transportation prepaid by the buyer,
within the warranty period. DEI shall in no way be liable for damages
consequential or incidental to defects in its products, for failure of delivery in
whole or in part, for injuries resulting from its use, or for any other cause.
Returns must be preauthorized and accompanied by a DEI return authorization
number.
The foregoing states the entire warranty extended by DEI, and is given and
accepted in lieu of 1) any and all other warranties, expressed or implied,
including by not limited to the implied warranties of merchantability and fitness for
any particular purpose and 2) any obligation, liability, right, claim or remedy in
contract or tort.