Dionex MSQ User Manual

MSQ
Document No. 031871
Hardware
Revision 02
October 2003
This publication is protected by federal copyright law. No part of this publication may be copied or distributed, transmitted, transcribed, stored in a retrieval system, or transmitted into any human or computer language, in any form or by any means, electronic, mechanical, magnetic, manual, or otherwise, or disclosed to third parties without the express written permission of Dionex Corporation, 1228 Titan Way, Sunnyvale, California 94088-3603 U.S.A.
DISCLAIMER OF WARRANTY AND LIMITED WARRANTY
THIS PUBLICATION IS PROVIDED “AS IS” WITHOUT WARRANTY OF ANY KIND. DIONEX CORPORATION DOES NOT WARRANT, GUARANTEE, OR MAKE ANY EXPRESS OR IMPLIED REPRESENTATIONS REGARDING THE USE, OR THE RESULTS OF THE USE, OF THIS PUBLICATION IN TERMS OF CORRECTNESS, ACCURACY, RELIABILITY, CURRENTNESS, OR OTHERWISE. FURTHER, DIONEX CORPORATION RESERVES THE RIGHT TO REVISE THIS PUBLICATION AND TO MAKE CHANGES FROM TIME TO TIME IN THE CONTENT HEREINOF WITHOUT OBLIGATION OF DIONEX CORPORATION TO NOTIFY ANY PERSON OR ORGANIZATION OF SUCH REVISION OR CHANGES.
TRADEMARKS
Chromeleon is a registered trademark of Dionex Corporation. Cone Wash, M-Path, MSQ, and Xcalibur are trademarks of Thermo Electron Corporation. Microsoft is a registered trademark of Microsoft Corporation. Q-tip is a trademark of Chesebrough-Pond’s, Inc. Teflon, Vespel, and Viton are registered trademarks of E.I. du Pont de Nemours and Company.
PRINTING HISTORY
Revision 01, July 2002 Revision 02, October 2003
The products of Dionex Corporation are produced under ISO 9001 accredited quality management systems.
Published by Technical Publications, Dionex Corporation, Sunnyvale, CA 94086.
Chapter 1
Introducing the MSQ 1.
Introducing the MSQ.................................................................................................................1-i
Introduction ....................................................................................................................................1-1
System Overview............................................................................................................................1-2
What Is Mass Detection?...................................................................................................1-4
Exterior Features of the MSQ............................................................................................1-5
The Source–An Introduction to API Techniques ...........................................................................1-8
Electrospray.......................................................................................................................1-9
Atmospheric Pressure Chemical Ionization.....................................................................1-13
Source Fragmentation......................................................................................................1-16
Cone Voltage Ramping ...................................................................................................1-18
Polarity Switching ...........................................................................................................1-19
Application of API Techniques .......................................................................................1-20
The Self-Cleaning Source: Cone Wash ........................................................................................1-23
Introduction .....................................................................................................................1-23
Functional Description ....................................................................................................1-24
The Reference Inlet System..........................................................................................................1-25
Introduction .....................................................................................................................1-25
Functional Description ....................................................................................................1-25
The Mass Analyzer and Detector .................................................................................................1-26
The Vacuum System.....................................................................................................................1-27
The Data System...........................................................................................................................1-28
Software...........................................................................................................................1-28
Raw Data .........................................................................................................................1-30
Raw Data Types...............................................................................................................1-31
____________________________MSQ Hardware Manual _____________________________ 1-i
Introducing the MSQ Introduction ____________________________________________________________________________
1-ii____________________________ MSQ Hardware Manual ____________________________
Introducing the MSQ
___________________________________________________________________________ Introduction

Introduction

The MSQ™ MS detector has been specifically designed and engineered for liquid chromatographic detection using Atmospheric Pressure Ionization (API) and Mass Spectrometry (MS) technology. These technologies can provide sensitive and selective detection of organic molecules.
Interfacing High Performance Liquid Chromatography (HPLC or LC) and MS provides the separation scientist with one of the most powerful analytical tools available. Both LC and MS have developed to a point whereby they represent two of the most important techniques in characterizing and detecting organic compounds. Although the potential benefits of interfacing LC to MS have been clearly recognized for many years, producing a truly automated “connect-and-use” interface has proven to be a challenging task.
Atmospheric Pressure Ionization (API) techniques now provide highly sensitive detection using conventional to capillary LC flow rates on bench­top MS detector systems. LC/MS works with typical solvent compositions, whether the separation is achieved by isocratic or gradient elution.
Historically, LC/MS has been compatible only with volatile buffer systems using modifiers such as trifluoroacetic acid, formic acid, and acetic acid. Phosphate buffers, although extensively used in LC separations, were not suited to LC/MS due to rapid blocking of the ion sampling region caused by the deposition of involatile phosphate salts. The self-cleaning API source allows for extended periods of operation in LC/MS with chromatographic buffers such as phosphates or ion-pairing agents and samples in dirty matrices.
API using Electrospray (ESI) or Atmospheric Pressure Chemical Ionization (APCI) interfaces has proved to be invaluable in meeting sensitivity requirements in quantitative methods. It can also provide structural information, which is complementary to techniques such as NMR and infra red spectroscopy.
This introduction focuses on the principal components of the system.
___________________________MSQ Hardware Manual ____________________________ 1-1
Introducing the MSQ System Overview ________________________________________________________________________

System Overview

The MSQ MS detector is an integral part of the LC detection system. Key points of the system are:
The sample is introduced into the ion source using an LC system,
possibly through a column.
In an API MS detector, the part of the source where ionization takes place is held at atmospheric pressure, giving rise to the term Atmospheric Pressure Ionization (API).
In ESI, the sample is ionized in the liquid phase, while in APCI, ionization occurs in the gas phase. In both cases, efficient desolvation is needed to remove the solvents from the sample.
Ions, now in the gas phase, are passed through the mass analyzer and are collected at the detector.
The detected signal is sent to the data system and stored ready for processing.
LC System
Sample
introduction
LC Column
Separation
Figure 1-1. The key components of the MSQ API LC detection system
Ion Source
Ionization &
transmission
Mass Analyzer
Sorting of ions
Turbomolecular
& rotary pumps
Molecular weight information
Structural information
Positive identification
Quantitative information
Detector
Detection
of ions
Data System
Windows NT
1-2 ___________________________ MSQ Hardware Manual ____________________________
Introducing the MSQ
_______________________________________________________________________System Overview
The main features of the MSQ MS detector are:
Dual ESI/APCI orthogonal probe
• • Self-cleaning API-LC/MS interface
M-Path™ triple orthogonal source
Quadrupole mass analyzer
Split flow turbomolecular pump
Square quadrupole RF lens
Rotary pump
Exit cone
Entrance cone
From HPLC
Cone wash
Orthogonal sample introduction probe
Figure 1-2. Schematic diagram of the MSQ API inlet, analyzer, and
detector system
The LC eluent is ionized at the API probe and the resulting ions are focused into a square quadrupole RF lens. The quadrupole mass analyzer filters the ions before detection.
___________________________MSQ Hardware Manual ____________________________ 1-3
Introducing the MSQ System Overview ________________________________________________________________________

What Is Mass Detection?

Mass detection is a very powerful analytical technique used in a number of fields, including:
Identification of unknown compounds
• • Quantitation of known compounds
Determination of chemical structure
The basic function of an MS detector is to measure the mass-to-charge ratio of ions.
The unit of mass used is the Dalton (Da). One Dalton is equal to 1/12 of the mass of a single atom of carbon-12. This follows the accepted convention that an atom of carbon-12 has exactly 12 atomic mass units (amu). The MS detector does not directly measure molecular mass, but the mass-to-charge ratio of the ions. Electrical charge is a quantized property and so can exist only as an integer; that is, 1, 2, 3, and so on. The unit of charge used here (z) is that which is on an electron (negative) or a proton (positive). Therefore, the mass-to-charge ratio measured can be denoted by m/z. Most ions encountered in mass detection have just one charge. In this case, the mass­to-charge ratio is often spoken of as the “mass” of the ion.
1-4 ___________________________ MSQ Hardware Manual ____________________________
Introducing the MSQ
_______________________________________________________________________System Overview

Exterior Features of the MSQ

This section highlights the exterior features of the MSQ. The parts labeled here may be referred to in later chapters of this manual or other manuals supplied with the MSQ.
Status light
Figure 1-3. Front view of the MSQ
Figure 1-3 shows the front view of the MSQ. The main feature is the status light.
___________________________MSQ Hardware Manual ____________________________ 1-5
Introducing the MSQ System Overview ________________________________________________________________________
Table 1-1. Instrument status light
Instrument Status Light
Vented Red
Venting Red
Pumping down Flashing yellow
Under vacuum (above vacuum trip)
Under vacuum (ready for use)
Operate on (MSQ in use)
Source enclosure open Red
Red
Yellow
Green
The vacuum trip is the pressure below which it is safe to switch on the voltages in the source. When the instrument is functioning normally, the status light will go from flashing yellow to solid yellow and Operate can be switched On. If the pressure in the instrument rises above the operating pressure, the status light turns red to indicate that the pressure is above a safe level. See the chapter Shutting Down and Restarting the System for information on pumping down the MSQ.
Figure 1-4 shows the MSQ with the doors open. The source enclosure and reference inlet are now visible.
Source enclosure
Figure 1-4. The MSQ with the doors open
1-6 ___________________________ MSQ Hardware Manual ____________________________
Reference inlet
Introducing the MSQ
_______________________________________________________________________System Overview
Figure 1-5 is a schematic of the rear view of the MSQ.
PUMP RELAY
Rotary pump power
To P C
Contact closure and analog inputs
To rotary pump
USB
SOURCE
USER I/O
EXHAUST
Exhaust from API source
BACKING
To rotary pump
Figure 1-5. Rear view of the MSQ
RESET
MODEL: RATING: 220-240v
GAS IN
6 BAR MAX
50/60 Hz 1000 VA
MAINS ON/OFF
Gas inlet for nebulizer and sheath gas
Reset communications
Power switch
MAINS IN
Power supply
___________________________MSQ Hardware Manual ____________________________ 1-7
Introducing the MSQ The Source–An Introduction to API Techniques ________________________________________________

The Source–An Introduction to API Techniques

The source, or interface, performs four main functions:
Separates the analytes from the solvent and buffer systems used in LC
• • Ionizes the analyte molecules
Allows efficient transfer of ions into the mass analyzer for detection
LC eluent enters the source through the orthogonal sample introduction probe. The primary objective of an orthogonal probe is to direct any involatile components present in the LC eluent, such as those from buffers, ion-pairing agents, or matrices, away from the entrance orifice. Under operating conditions, however, both the sample ions and the charged liquid droplets (containing any involatile components, if they are present) are deflected by the electric field towards the entrance orifice. This leads to a gradual buildup of involatiles and a concomitant loss in sensitivity with time. The self-cleaning source delivers a constant, low flow of solvent (the cone wash™) to the edge of the inlet orifice, helping to prevent a buildup of involatiles during an LC/MS run.
1-8 ___________________________ MSQ Hardware Manual ____________________________
Introducing the MSQ
_______________________________________________ The Source–An Introduction to API Techniques
Rotary pump
Exit cone
To the mass analyzer
Entrance cone
From HPLC
Figure 1-7. Schematic of the MSQ source showing the cone wash
Cone wash
Orthogonal sample introduction probe
Two types of API are commonly encountered. These are Electrospray Ionization (ESI) and Atmospheric Pressure Chemical Ionization (APCI). The following sections discuss the mechanism of ion generation in each.

Electrospray

Electrospray Ionization (ESI) is regarded as a soft ionization technique providing a sensitive means of analyzing a wide range of polar molecules. Since the first combined ESI LC/MS results were announced in 1984, and its first application to protein analysis four years later, the technique has become an established analytical tool in separation science.
When applied to smaller molecules up to 1000 Daltons in molecular mass, electrospray ionization results in either a protonated, [M+H] 1-8) or deprotonated, [M-H]
-
, molecule. Choice of ionization mode is governed by the functional chemistry of the molecule under investigation. In ESI, fragmentation is generally not apparent; however, increased source voltages can induce fragmentation to provide structural information.
___________________________MSQ Hardware Manual ____________________________ 1-9
+
(see Figure
Introducing the MSQ
z
The Source–An Introduction to API Techniques ________________________________________________
100
240
OH
NH
tBu
HO
%
HO
Chemical structure of salbutamol, (molecular weight 239)
0
60 80 100 120 140 160 180 200 220 240 260 280 300
241
m/
Figure 1-8. Electrospray mass spectrum of salbutamol in positive ion
mode
The base peak at m/z 240 (see Figure 1-8) corresponds to the protonated salbutamol molecule. It is notable that ESI results in a prominent base peak with minimal fragmentation, quite dissimilar from the results often achieved with GC/MS.
Mechanism of Ion Generation
Electrospray ionization operates by the process of emission of ions from a droplet into the gas phase, a process termed Ion Evaporation. A solvent is pumped through a stainless steel insert capillary that carries a high potential, typically 3 to 5 kV (see Figure 1-9). The strong electric field generated by this potential causes the solvent to be sprayed from the end of the insert capillary (hence, electrospray), producing highly charged droplets. As the solvent is removed by the desolvation process, the charge density on the surface of the droplets increases until the Rayleigh limit is exceeded; after this, a multitude of smaller droplets are formed by coulombic explosion. This process is repeated until charged sample ions remain. These ions are then available for sampling by the ion source.
-
+
Insert capillary +3-5 kV
+
+
+
+
­+
Droplet containing ions
+
+
+
+
-
+
+
+
+
-
+
+
+
­+
+
+
+
As the droplet evaporates, the electric field increases and ions move towards the surface
Figure 1-9. Positive ion electrospray mechanism
-
+
+
­+
-
+
+
+
+
+
-
+
-
+
+
+
-
+
+
++
+
+
+
Ions evaporate from the surface
1-10 __________________________ MSQ Hardware Manual ____________________________
Introducing the MSQ
_______________________________________________ The Source–An Introduction to API Techniques
Electrospray Ionization Using the MSQ Source
The sample, in solution, enters the source via a stainless steel insert capillary held at a voltage of 3 to 5 kV. The insert capillary is surrounded by a tube that directs a concentric flow of nitrogen nebulizing gas past the droplets of liquid forming at the probe tip. The action of the nebulizing gas, high voltage, and heated probe produces an aerosol of liquid droplets containing ions of the sample and solvent. The ion evaporation process is assisted by a second concentric flow of heated nitrogen gas. This is the sheath gas. This highly efficient evaporation process close to the entrance cone enables the routine use of high LC flow rates (up to 2.0 mL/min) in ESI mode.
The newly formed ions then enter the focusing region through the entrance cone. This is due to the following:
The high electric field. The insert capillary is at 3 to 5 kV with respect
to the rest of the source, which is typically at 20 to 30 V.
The gas flow into the focusing region.
Ions then exit the focusing region and pass into the RF lens. The RF lens (square quadrupole) helps to focus the ions before they enter the mass analyzer region.
Region
Intermediate
Pressure
Region
Exit cone
Entrance cone
Probe
Insert capillary
LC eluent
Nebulizing gas, N
Sheath gas, N
2
2
Insert
Cone Wash
Atmospheric
Pressure
Region
Figure 1-10. Schematic of the ESI source on the MSQ, showing the
principal components and pressure regions
Rotary pump
___________________________MSQ Hardware Manual ___________________________ 1-11
Introducing the MSQ The Source–An Introduction to API Techniques ________________________________________________
Spectral Characteristics
Polar compounds of low molecular weight (<1000 amu) typically form singly charged ions by the loss or gain of a proton. Basic compounds (for example, amines) can form a protonated molecule [M+H] analyzed in positive ion mode to give a peak at m/z M+1. Acidic compounds (for example, sulphonic acids) can form a deprotonated molecule [M-H]
-
, which can be analyzed in negative ion mode to give a peak at m/z M-1. As electrospray is a very soft ionization technique, there is usually little or no fragmentation and the spectrum contains only the protonated or deprotonated molecule.
Some compounds are susceptible to adduct formation if ionization takes place in the presence of contamination or additives such as ammonium or sodium ions. The spectra will show other ions in addition to, or instead of, the quasi-molecular ion. Common adducts are ammonium ions NH [M+18]
100
+
, sodium ions Na+ [M+23]+, and potassium ions K+ [M+39]+.
+
, which can be
+
4
[M+H]
322
+
%
+
100
103
80
99
0
60 80 100 120 140 160 180 200 220 240 260 280 300 320 340 360
141
145
181
187
241
244
261
279
282
[M+Na]
344
363
m/z
Figure 1-11. Electrospray spectrum showing a sodium adduct
The singly charged ions arising from samples of relatively low molecular masses can be interpreted directly, as they represent the protonated or deprotonated molecule. Electrospray, however, can produce multiply charged ions for analytes that contain multiple basic or acidic sites, such as proteins and peptides. As an MS detector measures mass-to-charge ratio (m/z), these ions appear at a m/z value given by the mass of their protonated molecule divided by the number of charges:
n
+
nHM
+
 
 
n
zm
=
Where, M = actual mass, n = number of charges, and H = mass of a proton.
Electrospray allows molecules with molecular weights greater than the mass range of the MS detector to be analyzed. This is a unique feature of electrospray.
1-12 __________________________ MSQ Hardware Manual ____________________________
Introducing the MSQ
_______________________________________________ The Source–An Introduction to API Techniques
Flow Rate
The electrospray source can be used with flow rates from 5.0 µL/min to
2.0 mL/min.

Atmospheric Pressure Chemical Ionization

Atmospheric Pressure Chemical Ionization (APCI) is also a very soft ionization technique and has many similarities to electrospray ionization. Ionization takes place at atmospheric pressure and the ions are extracted into the MS detector in the same way as in electrospray.
Similarly, as observed in ESI, [M+H] providing molecular weight information. Fragmentation can be induced in the source by increasing the source voltage to give structural information.
Mechanism of Ion Generation
+
and [M-H]- ions are usually formed
In APCI, the liquid elutes from an insert capillary, surrounded by a coaxial flow of nitrogen nebulizing gas into a heated region. The combination of nebulizing gas and heat form an aerosol that evaporates quickly to yield desolvated neutral molecules (see Figure 1-12).
At the end of the probe is a corona pin held at a high potential (typically 2.0 to 3.5 kV). This produces a high-field corona discharge that causes solvent molecules eluting into the source to be ionized. In the atmospheric pressure region surrounding the corona pin, a series of reactions occur that give rise to charged reagent ions. Any sample molecules, which elute and pass through this region of reagent ions, can be ionized by the transfer of a proton to form [M+H]
+
or [M-H]-. This is a form of chemical ionization; hence the
name of the technique, Atmospheric Pressure Chemical Ionization.
Heated nebuliser
N
2
Liquid
N
2
Solvent molecules
An aerosol is formed
Sample molecules
Solvent and
sample molecules
are desolvated
Corona pin
Collisions and
+
+
+
proton transfer
+
+
+
+
Solvent molecules
are ionized
+
+
Sample
+
[M+H]
ions formed
Figure 1-12. Positive ion APCI mechanism
___________________________MSQ Hardware Manual ___________________________ 1-13
Introducing the MSQ The Source–An Introduction to API Techniques ________________________________________________
APCI Using the MSQ Source
The sample is carried to a spray region via a stainless steel insert capillary. The action of both the nebulizing gas and the heated probe leads to the formation of an aerosol. The desolvation process is assisted by a second concentric flow of nitrogen gas, the sheath gas.
In contrast to electrospray, APCI is a gas phase ionization technique. Ionization occurs as the aerosol leaves the heated nebulizer region. A corona pin, mounted between the heated region and the entrance cone, ionizes the sample molecules with a discharge voltage of approximately 3.0 to 3.5 kV in positive ion mode and 2.0 to 3.0 kV in negative ion mode.
Region
Intermediate
Pressure
Region
Exit cone
Rotary pump
Entrance cone
Pressure
Region
Cone Wash
LC eluent
Nebulizing gas, N
Sheath gas, N
Probe
2
2
Atmospheric
Corona pin
Figure 1-13. Schematic of the APCI source on the MSQ, showing the
principal components and pressure regions
The newly formed ions then enter the focusing region through the entrance orifice and pass into the RF lens region. The RF lens (square quadrupole) helps to focus the ions before they enter the mass analyzer region.
1-14 __________________________ MSQ Hardware Manual ____________________________
Introducing the MSQ
_______________________________________________ The Source–An Introduction to API Techniques
Spectral Characteristics
Like electrospray, APCI is a soft ionization technique and forms singly charged ions–either the protonated, [M+H]
+
, or deprotonated, [M-H]-, molecule–depending on the selected ionization mode. Unlike electrospray, however, APCI does not produce multiply charged ions and so is unsuitable for the analysis of high molecular weight compounds such as proteins or peptides.
Although a high temperature is applied to the probe, most of the heat is used in evaporating the solvent, so the thermal effect on the sample is minimal. In certain circumstances (for example, with very thermally labile (unstable) compounds), the heated probe may cause some thermal fragmentation.
Flow Rate
Flow rates of 0.2 to 2.0 mL/min can be used with APCI.
___________________________MSQ Hardware Manual ___________________________ 1-15
Introducing the MSQ The Source–An Introduction to API Techniques ________________________________________________

Source Fragmentation

Both electrospray and APCI are regarded as soft ionization techniques. Ionization generally results in spectra dominated by either the protonated molecule [M+H] (negative ion mode), depending on whether positive or negative ionization mode has been selected. Choice of ionization mode is governed by the functional chemistry of the molecule under investigation.
Source fragmentation can be induced to give additional information on a compound, such as diagnostic fragment ions for structural determination or an increased response on a particular confirmatory ion for peak targeting.
Formation of Diagnostic Fragment Ions
The MSQ allows the simultaneous acquisition of MS data at a number of different source voltages. For example, the MSQ can be programmed to acquire data at source voltages of 20, 40, and 60 V on an alternating scan basis within a single acquisition. The benefits of setting up acquisitions in this way are:
+
(positive ion mode) or deprotonated molecule [M-H]-
The optimum source voltage for a particular ion can be determined in
one acquisition for compounds where sample volume is at a premium.
The intensity of fragment ions can be maximized to gain structural
information.
Fragmentation at increased source voltages is useful for most compounds. For example, using source fragmentation of salbutamol in electrospray ionization, a number of confirmatory fragment ions can be generated and their intensity maximized (see Figure 1-14).
The mechanism for the formation of the fragment ions is characteristic for not only salbutamol, but also for related β-agonists such as clenbuterol,
terbutaline, and metaproterenol. It involves loss of water (-18 amu; resulting in the fragment ion at m/z 222 (middle trace)) and an additional loss of the tert-butyl group (-56 amu; resulting in the fragment at m/z 166 (lower trace)).
Note. The MSQ uses the term “cone voltage” to represent source voltage.
1-16 __________________________ MSQ Hardware Manual ____________________________
Introducing the MSQ
+
+
+
u
_______________________________________________ The Source–An Introduction to API Techniques
100
Source voltage 10V
%
0
50 75 100 125 150 175 200 225 250 275 300
100
Source voltage 25V
%
0
50 75 100 125 150 175 200 225 250 275 300
100
Source voltage 35V
%
0
50 75 100 125 150 175 200 225 250 275 300
148
166
240
241
240
222
241
240
222
241
m/z
m/z
m/z
Protonated molecule at m/z 240
OH
+
tBu
NH
HO
HO
2
Fragment ion at m/z 222
NH
tBu
HO
HO
2
Fragment ion at m/z 166
NH
HO
HO
3
[M+H]
[M+H]+-H2O
[M+H]+-H2O-tB
Figure 1-14. Source fragmentation of salbutamol in electrospray
ionization
Optimum Response for Confirmatory Ions
When acquiring fragment ions for confirmation purposes, the applied source voltage per compound would require some optimization to maximize the intensity of these ions. It is generally observed that small changes to the source voltage result in only small intensity changes; thus, fine-tuning of this voltage is usually not critical (typically +/- 5 V is adequate).
___________________________MSQ Hardware Manual ___________________________ 1-17
Introducing the MSQ The Source–An Introduction to API Techniques ________________________________________________

Cone Voltage Ramping

Source voltage ramping can be used in Full Scan operation (see page 1-32) in electrospray when the compound of interest forms multiply charged ions. A spectrum similar to that in Figure 1-15 will be produced, showing the multiply charged envelope.
771
808
848
893
942
944
998
999
1060
1131
1212
1305
m/z
100
%
738
707
694
0
700 800 900 1000 1100 1200 1300
Figure 1-15. Electrospray spectrum of horse heart myoglobin
This envelope is represented in diagrammatic form in Figure 1-16. The first diagram shows the envelope at a source voltage of 30 V. Next, the envelope is shown at 60 V. There is no ramping applied here. The envelope has the same shape but has moved to a higher m/z value. This is due to the charge stripping that occurs at high source voltages, which lowers the charge state and produces an apparent higher mass distribution. If the source voltage is ramped between 30 and 60 V, the charge distribution envelope is extended and it resembles the lower diagram.
1-18 __________________________ MSQ Hardware Manual ____________________________
Introducing the MSQ
f
_______________________________________________ The Source–An Introduction to API Techniques
Intensity
Without s ource voltage ramping
Source voltage 30V
Increasing number of charges
Intensity
Decreasing
number o
charges
With source voltage ramping 30-60V
m/z
Intensity
Without s ource voltage ramping
Source voltage 60V
m/z
Figure 1-16. Cone voltage ramping
Cone voltage ramping is often used for proteins and peptides during:
Calibration (to extend the calibration range)
Analysis (to increase the number of charge states leading to greater
accuracy of molecular weight determination)
m/z
Note. Cone voltage ramping will not perform source fragmentation because the same cone voltage corresponds to the same m/z value each time.

Polarity Switching

Switching between positive and negative ionization modes in a single analytical run is supported by the MSQ. Rapid polarity switching is a technique that is applied to several important areas of MS analysis; for example:
Quantitation of different chemistries within the same run. In drug
metabolism studies, certain compounds preferentially ionize in positive ion mode because they may contain a primary amino group. Other metabolites, such as glucuronide metabolites, are likely to lose a proton and respond in the negative ion mode.
Rapid screening of unknown analytes; for example, in combinatorial
chemistry. If the compound has a carboxylic acid group that is sterically unhindered, it is likely that the compound will lose a proton in negative ion mode and not respond in positive ion mode.
___________________________MSQ Hardware Manual ___________________________ 1-19
Introducing the MSQ
A
The Source–An Introduction to API Techniques ________________________________________________

Application of API Techniques

Both electrospray and APCI are ideal for online liquid chromatography detection, providing an additional dimension of information. With many compounds, it is possible to analyze them by both APCI and electrospray. It may be difficult to decide which is the more appropriate technique, especially when the compounds of interest lack polar functionalities.
Points to note:
Electrospray is one of the softest ionization methods available, whereas
APCI, although also a soft ionization technique, may not be suitable for some very thermally labile compounds as there may be thermal fragmentation (see Figure 1-17).
214
231
231
PCI mass spectrum,
source voltage 25V.
Electrospray mass spectrum, source voltage 25V.
100
%
0
130 140 150 160 170 180 190 200 210 220 230 240
100
%
0
130 140 150 160 170 180 190 200 210 220 230 240
156
173
Figure 1-17. Thermal fragmentation of the herbicide asulam in APCI
APCI does not yield multiply charged ions like electrospray, and so is
unsuitable for the analysis of high molecular weight compounds such as proteins.
m/z
m/z
Both APCI and electrospray generally provide data from which it is
simple to infer molecular weight values. In many cases, with the correct conditions, only one major peak is observed in the spectrum: either the protonated molecule [M+H] ion) or deprotonated molecule [M-H]
-
(-ve ion). However, some
+
(+ve
compounds are more susceptible to fragmentation than others, so different degrees of fragmentation may be seen from compound to compound. When determining molecular weights, always take account of possible adduct ions. Common adducts are [M+18]
+
NH
+
4
(ammonium adducts seen in the presence of buffers such as ammonium acetate), [M+23]+ Na+ (sodium adducts), and [M+39]+ K+ (potassium adducts).
1-20 __________________________ MSQ Hardware Manual ____________________________
Introducing the MSQ
_______________________________________________ The Source–An Introduction to API Techniques
Source fragmentation is used in both APCI and electrospray to give
structural information. In general, increasing the voltage applied to the source block (the cone voltage) yields increasing amounts of fragmentation, depending on the nature of the compound. The optimum source voltage required to give the maximum intensity of the protonated or deprotonated molecule is compound-dependent, as is the source voltage required for fragmentation. The energies involved in source fragmentation are low, so usually only weaker bonds such as C-N and C-O are broken.
Since there are many similarities between electrospray and APCI, there are many applications common to both.
Compounds suitable for analysis by electrospray are polar and of molecular weight less than 100,000 amu. The higher molecular weight compounds, such as proteins, can produce multiply charged ions. As it is the mass-to­charge ratio (m/z) that is measured by the MS detector, these can often be seen at lower masses. For example, if the molecular weight is 10,000, a doubly charged ion (2+ in +ve ion) would be seen at m/z 5001, 10+ at m/z 1001, etc.
Typical electrospray applications are: peptides, proteins, oligonucleotides, sugars, drugs, steroids, and pesticides.
Compounds suitable for analysis by APCI are generally polar (although less polar than electrospray) and of molecular weight <1000 amu.
Typical APCI applications are: pesticides, drugs, azo dyes, and steroids.
A summary comparing electrospray and APCI is shown in Table 1-2.
___________________________MSQ Hardware Manual ___________________________ 1-21
Introducing the MSQ The Source–An Introduction to API Techniques ________________________________________________
Table 1-2. Comparison of ESI and APCI
LC/MS technique Electrospray (ESI) Atmospheric Pressure Chemical
Ionization (APCI)
Compound polarity Polar Polar, some non-polar
Examples Drugs, proteins, biopolymers,
oligonucleotides, steroids, and pesticides
Pesticides, azo dyes, drugs, metabolites, agrochemicals, and steroids
Sensitivity fg to pg (compound-dependent) fg to pg (compound-dependent)
Type of spectra [M+H]+ for +ve ion mode, [M-H]- for
-ve ion mode, fragmentation via source voltage
+
[M+H]
for +ve ion mode, [M-H]- for
-ve ion mode, fragmentation via source voltage
Flow rates 2.0 µL/min to 2.0 mL/min 0.2 to 2.0 mL/min
LC columns Capillary to 4.6-mm ID columns 2.1- to 4.6-mm ID columns
Mobile phases H2O, CH3CN, CH3OH are most
frequently used.
H2O, CH3CN, CH3OH are most frequently used. Non-polar solvents can be used.
Typical mass range <100,000 <1000
1-22 __________________________ MSQ Hardware Manual ____________________________
Introducing the MSQ
______________________________________________________The Self-Cleaning Source: Cone Wash

The Self-Cleaning Source: Cone Wash

This section introduces the cone wash. Information on how and when to use it is provided in the chapter LC/MS and the Cone Wash.

Introduction

The API source on the MSQ includes a self-cleaning solvent delivery system (the cone wash). This makes the source extremely robust and productive and greatly increases the number of samples that can be analyzed before maintenance is required.
The orthogonal API probe serves to direct the LC eluent away from the inlet orifice. However, under typical LC/MS conditions, both the ions and the charged liquid droplets (containing involatile components) are deflected by the electric field towards the inlet orifice. This effect leads to a gradual buildup of involatile components and an associated loss in sensitivity with time.
The self-cleaning API source delivers a constant, low flow of solvent to the edge of the inlet orifice (see Figure 1-18). This prevents the buildup of involatile components during LC/MS analysis with typical chromatographic buffers (for example, phosphates and ion-pairing agents). This greatly improves the quantitation precision of analysis without the need to compromise the LC method, and more importantly, dramatically extends the length of time possible for analysis.
Figure 1-18. Dispersion of involatile components from the inlet orifice
___________________________MSQ Hardware Manual ___________________________ 1-23
Introducing the MSQ The Self-Cleaning Source: Cone Wash _______________________________________________________

Functional Description

As shown in Figure 1-18, the cone wash nozzle consists of a stainless steel capillary, which is fed to the edge of the inlet orifice of the entrance cone. The capillary is attached to PEEK tubing, internal to the instrument, which when connected to an LC pump delivers the solvent (typically HPLC-grade water or organic solvents such as methanol) at a controlled flow rate.
Note. It is necessary to use the cone wash only for compounds in dirty matrices or when involatile buffers are used. Choose the cone wash solvent to give the most effective solubility for the expected contaminants.
1-24 __________________________ MSQ Hardware Manual ____________________________
Introducing the MSQ
______________________________________________________________ The Reference Inlet System

The Reference Inlet System

This section introduces the reference inlet system. Information on how to set up the system is provided in the chapter Reference Inlet System.

Introduction

The recommended way to introduce a sample for tuning and mass calibration in electrospray is to use the reference inlet system.

Functional Description

As shown in Figure 1-20, the reference inlet system consists of a pressurized reservoir containing the reference sample and a PEEK delivery tube. One end of the PEEK delivery tube is inserted into the reference inlet reservoir, while the other end is attached to the switching valve. When the pressure in the reservoir is increased, the sample is forced through the PEEK tube and infused directly into the 100 µL sample loop.
Figure 1-20. The reference inlet system
___________________________MSQ Hardware Manual ___________________________ 1-25
Introducing the MSQ The Mass Analyzer and Detector____________________________________________________________

The Mass Analyzer and Detector

The mass analysis and detection system comprises two main components:
A quadrupole mass analyzer
A detector whose main component is a channeltron electron multiplier
The square quadrupole RF lens helps to focus the ions before they are filtered, according to their mass-to-charge ratio in the mass analyzer.
The analyzer in the MSQ is a quadrupole. This is one of the most widely used types of analyzer and can be easily interfaced to various inlet systems. By applying carefully controlled voltages to the four rods in the quadrupole, only ions of a specific mass-to-charge ratio are allowed to pass through at any one time.
The ions then reach the detector, whose main component is a channeltron electron multiplier. In positive ion mode, the ions exit the analyzer and strike a conversion dynode, which results in the emission of electrons. The electrons are accelerated towards the channeltron electron multiplier, which then creates an electron cascade. The current is then converted and amplified into a voltage signal that is analyzed and processed by the MSQ’s on-board data acquisition system. The resultant peak information is sent to the data system.
Ion successfully transmitted
by the mass analyzer
Quadrupole
mass analyzer
Square quadrupole RF lens
Source exit cone
Figure 1-21. Schematic of the MSQ analyzer and detector
Detector
Ion trajectory incorrect
for transmission trough
the mass analyzer
Split flow turbomolecular pump
1-26 __________________________ MSQ Hardware Manual ____________________________
Loading...
+ 102 hidden pages