Dionex IonPac TCC Series, IonPac TCC-LP1, IonPac TCC-ULP1, IonPac TCC-XLP1 Product Manual

IonPac® TCC
IonPac TCC-LP1, TCC-ULP1, and TCC-XLP1 Product Manual
Page 1 of 17
PRODUCT MANUAL
FOR
TRACE CATION CONCENTRATOR
(TCC)
TCC-LP1 Column – Low Pressure, 4 x 35 mm (P/N 063079)
TCC -ULP1 Column – Ultra Low Pressure, 5 x 23 mm (P/N 06 3 78 3)
TCC -XLP1 Column – Extremely Low Pressure, 6 x 16 mm (P/N 063889)
®
Dionex
Corporation, 2005
Document No. 034973
Revision 06
September 2005
Doc. No. 034973-06 ©2005 DIONEX September 2005
IonPac TCC-LP1, TCC-ULP1, and TCC-XLP1 Product Manual
Page 2 of 17
TABLE OF CONTENTS
1.
INTRODUCTION.......................................................................................................................................................4
2.
INSTALLATION ........................................................................................................................................................6
2.1.
TCC Operation Precautions.........................................................................................................................6
2.2.
Column Description......................................................................................................................................6
3.
OPERATION...............................................................................................................................................................7
3.1.
Sample Loading.............................................................................................................................................7
3.2.
Reagent and Sample Handling.....................................................................................................................9
3.3.
Concentrator Capacity ............................................................................................................................... 10
4.
EXAMPLE APPLICATIONS..................................................................................................................................13
4.1.
Concentration in the Manual Mode...........................................................................................................13
4.2.
Concentration in the Manual Mode...........................................................................................................14
5.
TROUBLESHOOTING GUIDE..............................................................................................................................15
5.1.
High Backpressure from a Contaminated Inlet Bed Support.................................................................15
5.2.
High Background, Noise, or Baseline Instability......................................................................................16
5.3.
Poor Peak Shape..........................................................................................................................................16
APPENDIX A...........................................................................................................................................................................17
A.1 Recommended Operating Pressures.................................................................................................................17
A.2 Column Start-up.................................................................................................................................................17
A.3 Column Storage..................................................................................................................................................17
A.4 Column Cleanup of Polyvalent Cati ons and Acid-soluble Contaminan ts.....................................................17
A.5 Column Cleanup Of Organic/Cationic Contaminants....................................................................................17
Doc. No. 034973-06 ©2005 DIONEX September 2005
IonPac TCC-LP1, TCC-ULP1, and TCC-XLP1 Product Manual
Page 3 of 17
IMPORTANT INFORMATION
Several icons are used throughout this document to emphasize important points. The symbols are shown below, along with the purpose of the information.
Safety information can help prevent bodily harm.
SAFETY
WARNING
CAUTION
NOTE
Warning information can help prevent equipment harm.
Caution information can help prevent problems.
Note information can help with tips for improved use.
Doc. No. 034973-06 ©2005 DIONEX September 2005
IonPac TCC-LP1, TCC-ULP1, and TCC-XLP1 Product Manual
Page 4 of 17
1. INTRODUCTION
The IonPac® Trace Cation Concentrator Low Pressure (TCC-LP1, P/N 046027), Ultra Low Pressure (TCC-ULP1, P/N
063783), and Extremely Low Pressure (TCC-XLP1, P/N 063889) Columns are designed primarily for high purity water analysis. The function of the columns is to strip ions from a measured volume of a relatively clean aqueous sample matrix. This process “concentrates” the desired analyte species onto the columns leading to a lowering of detection limits by 2-5 orders of magnitude. The unique advantage of the columns to the analytical chemist is the capability o f perfo rming routine Low Pressure, Ultra Low Pressure, and Extremely Low Pressure Trace analyses of sample matrix ions at µg/L levels without extensive and laborious sample pretreatment.
The TCC-LP1, TCC-ULP1, and TCC-XLP1 columns are packed with a 20 µm styrene/divinylbenzene copolymer that is grafted to yield a resin with carboxylate functional groups. Due to its highly cross-linked structure, the resin is solvent compatible. The capacity of the TCC-LP1, TCC-ULP1, and TCC-XLP1 is 260 µeq/column with a void volume of approximately 250 µL. The physical rigidity of this resin allows the TCC-LP1, TCC-ULP1, and TCC-XLP1 to be used at pressures up to 3,000 psi (20.68 MPa). The TCC-LP1, TCC-ULP1, and TCC-XLP1 can be readily converted between the acid and the salt form without significant changes in the operating pressure.
The recommended maximum flow rate is 5 mL/min. The back pressure generated by the TCC-LP1 is less than 70 psi (0.48 MPa) at 1 mL/min, the TCC-ULP1 is less than 45 psi (0.31 MPa) at 1 mL/min, and the TCC-XLP1 is less than 30 psi (0.21 MPa) at 1 mL/min. The large resin particle size (20 µm) in the TCC-LP1, TCC-ULP1, and TCC-XLP1 makes it possible to do manual injections onto the concentrator. Syringes with up to 3 mL capacities can be used to manually push samples through the TCC-LP1. It takes approximately 1 minute to manually push 3 mL of sample through the TCC-LP1, TCC-ULP1, and TCC-XLP1 columns.
The TCC-LP1, TCC-ULP1, and TCC-XLP1 columns can be used with methanesulfonic acid, sulfuric acid, and hydrochloric acid eluents, with or without solvent, to concentrate samples on 5-mm, 4-mm, 3-mm, or 2-mm analytical systems.
This manual assumes that you are familiar with the installation and operation of the DIONEX Ion Chromatograph (IC) and the suppressor. If you do not understand the operation of the system, take the time to familiarize yourself with the various system components before beginning an analysis. The TCC-LP1, TCC-ULP1, and TCC-XLP1 have 10-32 threaded PEEK end fittings for use with ferrule/bolt liquid line fittings. If you find it necessary to install a component with 1/4-28 ports and therefore need to obtain or make one or more transition lines b e tween 10 -32 and 1/4-28 threaded ThermoFlare ports, DIONEX recommends the use of Tefzel® liquid lines with a PEEK ferrule/bolt fitting on one end and a 1/4-28 ThermoFlare fitting on the other end. See, “DIONEX Liquid Line Fittings,” for detailed instructions on purchasing or making these lines.
Assistance is available for any problem during the shipment or operation of Dionex instrumentation, columns, and consumables through the Dionex North America Technical Call Center at 1-800-DIONEX-0 (1-800-346-6390) or through any of the Dionex Offices listed in “Dionex Worldwide Offices” on the Dionex Reference Library CD-ROM.
Doc. No. 034973-06 ©2005 DIONEX September 2005
IonPac TCC-LP1, TCC-ULP1, and TCC-XLP1
Page 5 of 17
Product Manual
TABLE 1
IonPac TCC-LP1, TCC - ULP1, and TCC-XLP1 Concentrator Column Packing Specifications
Column
TCC-LP1
4-mm
TCC-ULP1
5-mm
TCC-XLP1
6-mm
Particle
Diameter
20 55 260 µeq/col
20 55 260 µeq/col
20 55 260 µeq/col
Substrate
X-Linking
Column
Capacity
Functional
Group
Carboxylic
Acid
Carboxylic
Acid
Carboxylic
Acid
Hydro-
phobicity
Very low
Very low
Very low
TABLE 2
IonPac TCC-LP1, TCC-ULP1, and TCC-XLP1 Concentrator Column Operating Parameters
Column
TCC-LP1
4-mm
TCC-ULP1
5-mm
TCC-XLP1
6-mm
Typical Backpressure
MPa at 30 ºC
< 70 (0.48) 0.50 5.0
< 45 (0.31) 0.50 5.0
< 30 (0.21) 0.50 5.0
Standard Flow Rate
mL/min
Maximum Flow Rate
mL/min
Doc. No. 034973-06 ©2005 DIONEX September 2005