This publication is protected by federal copyright law. No part of this publication
may be copied or distributed, transmitted, transcribed, stored in a retrieval system, or
transmitted into any human or computer language, in any form or by any means,
electronic, mechanical, magnetic, manual, or otherwise, or disclosed to third parties
without the express written permission of Dionex Corporation, 1228 Titan Way,
Sunnyvale, California 94088-3603 U.S.A.
DISCLAIMER OF WARRANTY AND LIMITED WARRANTY
THIS PUBLICATION IS PROVIDED “AS IS” WITHOUT WARRANTY OF
ANY KIND. DIONEX CORPORATION DOES NOT WARRANT,
GUARANTEE, OR MAKE ANY EXPRESS OR IMPLIED
REPRESENTATIONS REGARDING THE USE, OR THE RESULTS OF THE
USE, OF THIS PUBLICATION IN TERMS OF CORRECTNESS, ACCURACY,
RELIABILITY, CURRENTNESS, OR OTHERWISE. FURTHER, DIONEX
CORPORATION RESERVES THE RIGHT TO REVISE THIS PUBLICATION
AND TO MAKE CHANGES FROM TIME TO TIME IN THE CONTENT
HEREINOF WITHOUT OBLIGATION OF DIONEX CORPORATION TO
NOTIFY ANY PERSON OR ORGANIZATION OF SUCH REVISION OR
CHANGES.
TRADEMARKS
MMS™ MicroMembrane™ Suppressor is a trademark of Dionex Corporation.
AES® Atlas Electrolytic Suppressor, Chromeleon®, OnGuard®, and SRS® SelfRegenerating Suppressor are registered trademarks of Dionex Corporation.
Teflon® is a registered trademark of E.I. duPont de Nemours & Co.
Windows® is a registered trademark of Microsoft Corporation.
The Dionex ICS-1000 Ion Chromatography System (ICS-1000) performs ion
analyses using suppressed or non-suppressed conductivity detection. An ion
chromatography system typically consists of a liquid eluent, a high-pressure
pump, a sample injector, a guard and separator column, a chemical suppressor, a
conductivity cell, and a data collection system.
Before running a sample, the ion chromatography system is calibrated using a
standard solution. By comparing the data obtained from a sample to that obtained
from the known standard, sample ions can be identified and quantitated. The data
collection system, typically a computer running chromatography software,
produces a chromatogram (a plot of the detector output vs. time). The
chromatography software converts each peak in the chromatogram to a sample
concentration and produces a printout of the results.
Doc. 031879-01 3/031-1
ICS-1000 Ion Chromatography System
n
6
A typical IC analysis consists of 6 stages (see Figure 1-1).
. Data Analysis
Conductivity
Cell
Suppressor
5. Detectio
1. Eluent
Eluent
Delivery
Pump
2. Sample
Injection
Figure 1-1. Ion Analysis Process
1. Eluent Delivery
•Eluent, a liquid that helps to separate the sample ions, carries the
sample through the ion chromatography system. The ICS-1000 is an
isocratic delivery system. This means that the eluent composition and
concentration remain constant throughout the run.
2. Sample Injection
•The liquid sample is loaded into a sample loop either manually or
automatically (if an automated sampler is installed). When triggered,
the ICS-1000 injects the sample into the eluent stream.
Injection
Val ve
Guard Column
Sample Loop
Separator
Column
4. Suppression
3. Separation
Sample
•The pump pushes the eluent and sample through the guard and
separator columns (chemically-inert tubes packed with a polymeric
resin). The guard column removes contaminants that might poison the
separator column.
1-2Doc. 031879-01 3/03
3. Separation
•As the eluent and sample are pumped through the separator column,
the sample ions are separated. In the ICS-1000, the mode of
separation is called ion exchange. This is based on the premise that
different sample ions migrate through the IC column at different rates,
depending upon their interactions with the ion exchange sites.
4. Suppression
•After the eluent and sample ions leave the column, they flow through
a suppressor that selectively enhances detection of the sample ions
while suppressing the conductivity of the eluent.
5. Detection
•A conductivity cell measures the electrical conductance of the sample
ions as they emerge from the suppressor and produces a signal based
on a chemical or physical property of the analyte.
6. Data Analysis
•The conductivity cell transmits the signal to a data collection system.
1 • Introduction
•The data collection system (for the ICS-1000, this is Chromeleon®)
identifies the ions based on retention time, and quantifies each analyte
by integrating the peak area or peak height. The data is quantitated by
comparing the sample peaks in a chromatogram to those produced
from a standard solution. The results are displayed as a chromatogram
and the concentrations of ionic analytes can be automatically
determined and tabulated.
Doc. 031879-01 3/031-3
ICS-1000 Ion Chromatography System
1.2Overview of the ICS-1000
The ICS-1000 is an integrated ion chromatography system containing a pump,
injection valve, column heater, and conductivity detector. Other system
components, including a guard column, separator column, and suppressor vary,
depending on the analyses to be performed.
The ICS-1000 can optionally be configured with a column heater for temperature
control of the column.
ICS-1000 operation is controlled remotely by a personal computer running
Windows
later). Chromeleon also provides data acquisition and data processing functions.
For communication between the ICS-1000 and Chromeleon, the ICS-1000 is
connected to a USB (Universal Serial Bus) port on the computer or a USB hub.
For installation instructions, see Section B.5Chromeleon IC System (Document No. 031883).
® 2000 or Windows XP and Chromeleon software (version 6.5 SP2 or
and also refer to Installing the
1-4Doc. 031879-01 3/03
1.3About This Manual
1 • Introduction
Chapter 1
Introduction
Chapter 2
Features
Overview
Chapter 3
Operation and
Maintenance
Chapter 4
Troubleshooting
Chapter 5
Service
Appendix A
Specifications
Appendix B
Installation
Appendix C
TTL and Relay
Control
Introduces ion analysis and the ICS-1000; explains the
conventions used in this manual, including safety-related
information.
Provides an overview of ICS-1000 operating features and
system components; introduces the Chromeleon user
interface.
Provides operating instructions and describes routine
preventive maintenance procedures.
Lists problems and presents step-by-step procedures for
how to isolate and eliminate the cause of each problem.
Provides step-by-step instructions for routine service and
parts replacement procedures that the user can perform.
Lists the ICS-1000 specifications and installation site
requirements.
Describes how to install the ICS-1000.
Describes the ICS-1000 TTL and relay control features.
Appendix D
Reordering
Information
Appendix E
FAQ
Appendix F
Glossary
Doc. 031879-01 3/031-5
Lists spare parts for the ICS-1000.
Provides answers to frequently asked questions about
ICS-1000 operation.
Defines terms commonly used in ion analysis.
ICS-1000 Ion Chromatography System
1.3.1Safety Messages and Notes
This manual contains warnings and precautionary statements that can
prevent personal injury and/or damage to the ICS-1000 when properly
followed. Safety messages appear in bold type and are accompanied by
icons, as shown below.
Indicates an imminently hazardous situation which, if not avoided, will
result in death or serious injury.
Indicates a potentially hazardous situation which, if not avoided,
could result in death or serious injury.
Indicates a potentially hazardous situation which, if not avoided, may
result in minor or moderate injury. Also used to identify a situation or
practice that may seriously damage the instrument, but will not cause
injury.
Indicates that the function or process of the instrument may be
impaired. Operation does not constitute a hazard.
Messages d'avertissement en français
Signale une situation de danger immédiat qui, si elle n'est pas évitée,
entraînera des blessures graves à mortelles.
Signale une situation de danger potentiel qui, si elle n'est pas évitée,
pourrait entraîner des blessures graves à mortelles.
Signale une situation de danger potentiel qui, si elle n'est pas évitée,
pourrait entraîner des blessures mineures à modérées. Également
utilisé pour signaler une situation ou une pratique qui pourrait
gravement endommager l'instrument mais qui n'entraînera pas de
blessures.
1-6Doc. 031879-01 3/03
1 • Introduction
Warnhinweise in Deutsch
Bedeutet unmittelbare Gefahr. Mißachtung kann zum Tod oder
schwerwiegenden Verletzungen führen.
Bedeutet eine mögliche Gefährdung. Mißachtung kann zum Tod oder
schwerwiegenden Verletzungen führen.
Bedeutet eine mögliche Gefährdung. Mißachtung kann zu kleineren
oder mittelschweren Verletzungen führen. Wird auch verwendet, wenn
eine Situation zu schweren Schäden am Gerät führen kann, jedoch
keine Verletzungsgefahr besteht.
Notes
Informational messages also appear throughout this manual. These are
labeled NOTE and are in bold type:
NOTE NOTES call attention to certain information. They alert
you to an unexpected result of an action, suggest how to
optimize instrument performance, etc.
Doc. 031879-01 3/031-7
ICS-1000 Ion Chromatography System
1.3.2Safety Labels
The ICS-1000 meets EN 61010-1:1993 (safety), CAN/CSA-C22.2 No.
1010.1-92 (safety), UL 3101-1/10.93 (safety), EN 50082-1:1992
(susceptibility), EN 55011:1991 (emissions). The TUV GS, C, US Mark
safety labels and the CE Mark label on the ICS-1000 attest to compliance
with these standards.
The symbols below appear on the ICS-1000 or on ICS-1000 labels.
Alternating current
Protective conductor terminal (earth ground)
Power supply is on
Power supply is off
1-8Doc. 031879-01 3/03
This chapter describes key ICS-1000 features and introduces the Chromeleon user
interface.
2.1Operating Features
2.1.1Front Panel
Figure 2-1 illustrates the front panel of the ICS-1000.
Power
LED
Status
LEDs
Injection
Port
2•Features
Injection Port
The sample to be analyzed can
be injected manually into the
injection port, using a syringe.
For automated sample
injection, the ICS-1000 must be
connected to an autosampler.
For more information about
sample injection, see
Section 3.10
LEDs
The status LEDs (see
Figure 2-2
of various system functions.
See Table 2- 1
of each LEDs function. The
power LED indicates whether
the ICS-1000 power is on.
.
) indicate the status
for a description
Figure 2-1. ICS-1000 Front Panel
Doc. 031879-01 3/032-1
ICS-1000 Ion Chromatography System
INJECT VALVE
Load
Inject
ALARM
STATUS
Module Connected
Acquiring Data
Pump
Suppressor
Column Heater
Figure 2-2. Status LEDs
LED LabelIf On (Green)If Flashing
Load
Injection valve is in Load
Valve error.
position.
Inject
Injection valve is in Inject
Valve error.
position.
Alarm
No “on” (green) state.Error detected. Check the Chromeleon
Audit Trail for the cause.
Module
Connected
ICS-1000 is connected to
a Chromeleon timebase.
Does not flash.
Acquiring
Data
Pump
Sequence or manual data
Sequence has stopped due to an error.
acquisition is in progress.
Pump is on.High or low pressure limit is exceeded.
The pump is turned off
Suppressor
Column
Heater
Suppressor is on and
current is being applied to
it.
Column heater is at set
temperature.
Continuity check failed or suppressor is
over the voltage, current, or power limit.
The suppressor is turned off
Column heater is transitioning to a new
temperature.
Table 2-1. ICS-1000 Status LED States
2-2Doc. 031879-01 3/03
2.1.2Top Cover
Figure 2-3 illustrates the top cover of the ICS-1000.
Figure 2-3. ICS-1000 (Top View)
2 • Features
Reservoir Storage
The ICS-1000 top cover has room
for up to three 2-L plastic
reservoirs (P/N 044129).
Tubing Chase
The tubing chase under the top
cover routes tubing to the front of
the ICS-1000.
Doc. 031879-01 3/032-3
ICS-1000 Ion Chromatography System
2.1.3Component Panel
Figure 2-4 shows the user-accessible components installed on the
component panel behind the ICS-1000 front door.
11
11
10
8
9
5
7
6
Eluent Valve
1
Leak Sensor
2
Pump Heads
3
Pressure Transducer
4
Injection Valve
5
6
Columns
Connector for
7
optional column
heater
4
Suppressor
8
DS6 Conductivity
9
3
Cell
Mounting brackets
10
for a second
suppressor
2
Tubing Chase (2)
1
11
Figure 2-4. ICS-1000 Component Panel
2-4Doc. 031879-01 3/03
2 • Features
Eluent Valve
The eluent valve controls the flow from the eluent reservoir. It opens
automatically when the pump is running and closes when the pump is off.
Leak Sensor
The leak sensor is installed in the drip tray at the bottom of the component
panel. If liquid accumulates in the tray, an error message is reported in
Chromeleon and the Alarm LED flashes.
Pump Heads
The ICS-1000 includes a serial isocratic dual-head pump. The flow rate
can be set to 0.00 and to between 0.05 and 5.0 mL/min. However, for
optimum performance, set the flow rate to between 0.4 and 2.0 mL/min.
Setting the flow rate to 0.00 turns off the pump. See Section 2.4.3
details about the pump.
Pressure Transducer
The pressure transducer measures the system backpressure.
for
Injection Valve
The injection valve is a six-port, electrically-activated valve. A 25-µL
sample loop (P/N 042857) is installed on the valve at the factory. See
Section 2.4.4
Separator and Guard Columns
for details about valve operation.
Both the separator and guard columns are packed with resin and perform
the separation of the sample ions. The main function of the guard column
is to trap contaminants and remove particulates that might damage the
separator column.
Column Heater (Optional—Not Pictured)
The column heater controls the temperature of the separator and guard
columns. The temperature can be set to between 30 °C and 60 °C and it
must be set at least 5 °C above the ambient temperature. See Section 2.4.5
for details about the column heater.
Doc. 031879-01 3/032-5
ICS-1000 Ion Chromatography System
Suppressor
The suppressor reduces the eluent conductivity and enhances the
conductivity of the sample ions, thereby increasing detection sensitivity.
Either an AES® Atlas Electrolytic Suppressor, SRS® Self-Regenerating
Suppressor, or MMS™ MicroMembrane™ Suppressor can be used with
the ICS-1000. See Section 2.4.6
DS6 Heated Conductivity Cell
The flow-through conductivity cell measures the electrical conductance
of analyte ions as they pass through the cell. A heat exchanger inside the
cell regulates the temperature, which can be set to between 30 °C and
55 °C. For optimum performance, set the temperature to at least 7 °C
above the ambient temperature. See Section 2.4.7
cell.
Tubing Chases
The upper tubing chase routes tubing from the top cover to the component
panel. The lower tubing chase routes tubing from the component panel
through the ICS-1000 interior to the rear panel.
for details.
for details about the
2-6Doc. 031879-01 3/03
2.1.4Rear Panel
Figure 2-5 illustrates the ICS-1000 rear panel.
Tubing Clips
Power
Switch
2 • Features
Analog Output
Connector
USB
Connections
Main Power
Receptacle
Analog Output Connector
The analog output connector outputs conductivity data (scaled to 1 volt)
to an integrator or recording device. See Section B.19
setup information.
USB Connections
A USB receptacle is provided to allow connection to the Chromeleon
computer. Two USB ports are provided for connecting to other USB
devices. See Section B.5
TTL and Relay
Connector Strip
Waste Lines
Figure 2-5. ICS-1000 Rear Panel
for connection and
for connection instructions.
Doc. 031879-01 3/032-7
ICS-1000 Ion Chromatography System
TTL and Relay Connector
The TTL and Relay connector strip provides two TTL outputs, two relay
outputs, and four TTL inputs. The outputs can be used to control
functions in other TTL or relay controllable devices. The inputs can be
used to switch the injection valve position, turn on the pump, perform an
autozero command, and send an event mark to the analog output. See
Section C.1
Tubing Clips
The tubing clips hold tubing in place when routing it from the top cover.
Power Switch
The power switch provides on/off control of power to the ICS-1000.
Main Power Receptacle
The power supply cord plugs into the AC power receptacle.
The power supply cord is used as the main disconnect device. Make
sure the socket-outlet is located near the ICS-1000 and is easily
accessible.
for connection instructions.
Le cordon d'alimentation principal est utilisé comme dispositif
principal de débranchement. Veillez à ce que la prise de base soit
située/installée près du module et facilement accessible.
Das Netzkabel ist das wichtigste Mittel zur Stromunterbrechung.
Stellen Sie sicher, daß sich die Steckdose nahe am Gerät befindet und
leicht zugänglich ist.
2-8Doc. 031879-01 3/03
2.2Flow Schematic
Figure 2-6 illustrates the liquid flow path through an ICS-1000 when using
suppression in autorecycle mode.
2 • Features
PUMP
DEGAS
(Optional)
7
1
ELUENT
OUT
11
IN
COND
OUT
CELL
PULSE
DAMPER
SAMPLE
5
L
S
W
WASTE
C
6
P
L
VALV E
OUT
PUMPPUMP
HEAD
OUT
4
PRESS.
XDUCER
HEAD
WASTE
OUT
12
8
10
SUPPRESSOR
HEAT EXCHANGER
IN
SEPARATOR COLUMN
COLUMN HEATER (optional)
9
GUARD
3
WASTE
INOUT
ELUENT
VALV E
2
IN
Figure 2-6. ICS-1000 Flow Schematic
Doc. 031879-01 3/032-9
ICS-1000 Ion Chromatography System
•Eluent from the eluent reservoirn flows first through the pump degas
assembly (if it is installed) and then through the eluent valve
pump
p. It is then pushed through the pressure transducerq, which measures
the system pressure.
•From there, it is pushed through a pulse damperr, which smooths minor
pressure variations from the pump to minimize baseline noise. The eluent then
flows into the injection valve.
s
•After sample is loaded into the sample loopt and the injection valve is
toggled to the Inject position, eluent passes through the sample loop.
•If the optional column heater is installed, the eluent/sample mixture is
pumped through the heat exchanger
heater temperature.
•The mixture then goes to the guard and separator columnsv and through the
suppressor.
w
o to the
u, which heats the mixture to the column
•From the suppressor, the mixture flows through the cell, where the analytes
are detected. A digital signal is sent to Chromeleon software. Analog output
can be collected simultaneously.
11
•Finally, the mixture flows out of the cell and is recycled back into the
suppressor, where it is used as the water source for the regenerant chamber.
Flow is then routed to waste.
12
2-10Doc. 031879-01 3/03
2.3The Chromeleon Interface
NOTE This section provides a brief overview of the
Chromeleon interface. For details, refer to the
Chromeleon Help or user’s manual.
Chromeleon software is used for remote control of ICS-1000 operation. Two
modes of software control are available: direct control and automated control.
•With direct control, you select operating parameters and commands from the
Chromeleon menu bar, toolbars, and Control Panels. Direct control
commands are executed as soon as they are entered. Section 2.3.2
the commands available from the ICS-1000 Control Panels.
•With automated control, you create a list of samples (a sequence) to be
processed automatically. The sequence includes programs with commands
and parameters for controlling the ICS-1000, acquiring sample data, and
producing reports. Section 3.11.2
describes automatic sample processing.
2 • Features
describes
Doc. 031879-01 3/032-11
ICS-1000 Ion Chromatography System
2.3.1The Chromeleon Main Window and Browser
When you start Chromeleon, the main window and the Chromeleon
Browser appear (see Figure 2-7
).
Figure 2-7. Chromeleon Main Window and Browser
NOTE The Chromeleon Browser looks and functions like
Windows Explorer, but do not confuse the two. If you
copy files from the Chromeleon Browser into Explorer,
the files will be corrupted.
The Browser displays the directories in which chromatographic data is
located, and allows you to open, move, and delete data.
•The left pane of the Browser shows the directory tree structure. By
clicking the ‘+’ signs next to the folders and expanding the directory
tree, you can access sequences, System Control Panels, Wellness
Panels, and reports.
•The right pane displays information about the item selected in the left
pane.
2-12Doc. 031879-01 3/03
Loading...
+ 180 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.