Dionex GP50 Operator's Manual

GP50 GRADIENT PUMP
OPERATOR'S MANUAL
© 2001 Dionex Corporation
Document No. 031377
Revision 03
July 2001
©2001 by Dionex Corporation All rights reserved worldwide. Printed in the United States of Amer ica.
This publication is protected by federal copyright law. No part of this publication may be copied or distributed, transm itted, tr anscri bed , stored in a retrieval system, or transmitted into any human or computer language, in any form or by any means, electronic, mechanical, magnetic, manual, or otherwise, or disclosed to third parties without the express written permission of Dionex Corporation, 1228 Titan Way, Sunnyvale, California 94088-3603 U.S.A.
DISCLAIMER OF WARRANTY AND LIMITED WARRANTY
THIS PUBLICATION IS PROVIDED “AS IS” WITHOUT WARRANT Y OF ANY KIND. DIONEX CORPORATION DOES NOT WARRANT, GUARANTEE, OR MAKE ANY EXPRESS OR IMPLIED REPRESENTATI ONS RE GARD ING THE U SE, O R TH E RES ULTS OF THE USE, OF THIS PUBLICATION IN TERMS OF CORRECTNESS, ACCURACY, RELIABILITY, CURRENTNESS, OR OTHERWISE. FURTHER, DIONEX CORPORATION RESERVES THE RIGHT TO REVISE THIS PUBLICATION AND TO MAKE CHANGES FROM TIME TO TIME IN THE CONTENT HEREINOF WITHOUT OBLIGATION OF DIONEX CORPORATION TO NOTIFY ANY PERSON OR ORGANIZATION OF SUCH REVI SION OR CHANGES.
TRADEMARKS
DX-LAN™ is a trademark, and PeakNet®, Self-Regenerating Suppressor®, and SRS® are registered trademarks of Dionex Corporation Tefzel® is a registered trademark of E.I. duPont de Nemours & Co.
PRINTING HISTORY
Revision 01, May 1998 Revision 02, October 2000 Revision 03, July 2001

Contents

1 • Introduction
1.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
1.2 About This Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
1.2.1 Safety Messages and Notes . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
1.2.2 Safety Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
2 • Description
2.1 Front Control Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
2.1.1 Control Panel Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
2.1.2 Display Screens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7
2.2 Electronics Chassis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8
2.3 Mechanical Chassis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-11
2.4 Mechanical Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-12
2.5 Vacuum Degas Pump Assembly (Optional) . . . . . . . . . . . . . . . . . . . .2-16
2.6 Piston Seal Wash (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-16
2.7 Eluent Reservoirs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-17
Doc. 031377-03 7/01
2.4.1 Pump Eluent Manifold . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-13
2.4.2 Pump Heads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-14
2.4.3 Pump Mixers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-14
2.4.4 Pump Priming Block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-15
2.4.5 Pressure Transducer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-15
i
GP50 Gradient Pump
2.8 Rear Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-18
2.9 Functional Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-18
2.9.1 Operating and Control Modes . . . . . . . . . . . . . . . . . . . . . . . . 2-18
2.9.2 Local and Remote Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-19
2.9.3 Method Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-20
2.9.4 Eluent Delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-22
3 • Operation and Maintenance
3.1 Getting Ready to Run . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3.1.1 Degas Eluents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3.1.2 Filter Eluents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
3.1.3 Pressurize Eluent Reservoirs . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
3.1.4 Start-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
3.1.5 Selecting the Pressure Limits . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
3.1.6 Selecting the Operating and Control Modes . . . . . . . . . . . . . . 3-5
3.2 Running Under Direct Control (Local Mode) . . . . . . . . . . . . . . . . . . . 3-6
3.3 Running Under Method Control (Local Mode) . . . . . . . . . . . . . . . . . . 3-7
3.3.1 Creating a New Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
3.3.2 Running a Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-10
3.3.3 Editing a Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
3.3.4 Deleting a Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
3.3.5 Changing the Running Method . . . . . . . . . . . . . . . . . . . . . . . 3 -12
3.3.6 Controlling the Method Clock . . . . . . . . . . . . . . . . . . . . . . . .3-12
ii
Doc. 031377 -03 7/01
3.4 Example Methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-13
3.4.1 Isocratic Method Example . . . . . . . . . . . . . . . . . . . . . . . . . . .3-13
3.4.2 Linear Gradient Method Example . . . . . . . . . . . . . . . . . . . . .3-14
3.4.3 Curved Gradient Method Example . . . . . . . . . . . . . . . . . . . . .3-19
3.4.4 Editing a Running Method Example . . . . . . . . . . . . . . . . . . .3-23
3.5 Routine Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-25
3.5.1 Daily Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-25
3.5.2 Periodic Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-27
3.6 Shutdown . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-27
4 • Troubleshooting
4.1 Left-Right Pump Head Pressure Fluctuations . . . . . . . . . . . . . . . . . . . .4-1
Contents
4.2 Pump Will Not Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3
4.3 Pump Stops . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3
4.4 Liquid Leaks/Leak Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-7
4.5 Noisy Pump Motor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-8
4.6 Vacuum Degas Pump Does Not Run . . . . . . . . . . . . . . . . . . . . . . . . . .4-9
4.7 Vacuum Degas Pump Calibration Fails . . . . . . . . . . . . . . . . . . . . . . .4-10
4.8 Vacuum Degas Pump Low Vacuum . . . . . . . . . . . . . . . . . . . . . . . . . .4-11
4.9 Relays or TTLs Inoperative . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-11
4.10 Poor Chromatographic Reproducibility . . . . . . . . . . . . . . . . . . . . . . .4-12
Doc. 031377-03 7/01
iii
GP50 Gradient Pump
5•Service
5.1 Cleaning and Replacing the Check Valves . . . . . . . . . . . . . . . . . . . . . . 5-1
5.2 Piston Seal Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
5.3 Pump Piston Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
5.4 Pressure Transducer Waste Valve O-Ring Replacement . . . . . . . . . . 5-10
5.5 Proportioning Valve Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . .5-11
5.6 Changing Main Power Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-13
A • Specifications
A.1 Electrical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-1
A.2 Environmental . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-1
A.3 Physical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-1
A.4 Display and Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-1
A.5 Hydraulics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-2
A.6 Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-3
A.7 Vacuum Degas Pump Assembly (Optional) . . . . . . . . . . . . . . . . . . . . .A-3
B • Installation
B.1 Facility Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-1
B.2 Installation Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-2
B.2.1 Power Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-2
B.2.2 Electronics Chassis Connections . . . . . . . . . . . . . . . . . . . . . . .B-4
B.2.3 DX-L AN Interface: 10B ASE-T Connections (Optional). . . . .B-6
iv
Doc. 031377 -03 7/01
B.2.4 DX-LAN Interface: BNC Connections (Optional) . . . . . . . . . B-9
B.2.5 Waste Lines. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-13
B.2.6 Eluent Inlet Line Connections. . . . . . . . . . . . . . . . . . . . . . . . B-14
B.2.7 Eluent Outlet Line Connections . . . . . . . . . . . . . . . . . . . . . . B-14
B.2.8 Piston Seal Wash Connections (Optional) . . . . . . . . . . . . . . B-15
B.2.9 Priming the Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-17
B.3 Automatic SRS Power Control (Optional) . . . . . . . . . . . . . . . . . . . . B-20
C • Display Screens
C.1 Operational Screens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-2
C.1.1 Menu of Screens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-2
C.1.2 Main Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-3
Contents
C.2 Diagnostic Screens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-14
Doc. 031377-03 7/01
C.1.3 Detail Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-5
C.1.4 Method Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-6
C.1.5 Method Extension Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-8
C.1.6 Degas Options. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-9
C.1.7 Module Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-11
C.1.8 Pump Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-12
C.1.9 Time Function In. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-13
C.2.1 Diagnostic Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-14
C.2.2 Power-Up Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-15
C.2.3 Elapsed Time. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-16
C.2.4 DSP Status. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-17
v
GP50 Gradient Pump
C.2.5 DX-LAN Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-18
C.2.6 Keyboard Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-20
C.2.7 Diagnostic Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-21
C.2.8 Pressure Statistics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-23
C.2.9 DSP Code Version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-24
C.3 Calibration Screens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-25
C.3.1 Calibration Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-25
C.3.2 Calibration Status. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-26
C.3.3 Leak Sensor Calibration and Status . . . . . . . . . . . . . . . . . . . .C-27
C.3.4 Degas Status. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-28
C.3.5 Flow Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-29
C.3.6 Pressure Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-30
C.3.7 Degas Pump Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-34
D • TTL and Relay Control
D.1 TTL and Relay Output Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-2
D.2 TTL Input Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-3
D.2.1 TTL Input Signal Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-3
D.3 TTL and Relay Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-5
D.3.1 Example Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-6
vi
Doc. 031377 -03 7/01

1.1 Overview

The GP50 Gradient Pump is an integral part of a Dione x chr omat ogr aphy system. It is a microprocess or -based, dua l-pisto n, v ariable -speed, gr adient de li ver y system designed to blend and pump mixtures of up to four different eluents at precisely controlled flow rates. The pump can deliver the selected eluent composition isocratically, or as a multistep linear or curved gradient. A Digital Signal Processor (DSP) provides high speed control of pump flow and pressure.
The pump can operate as a stand-alone product or with other Dionex modules as part of a complete chromatography system. It can also be used with non-Dionex modules that meet interface requirements for software, TTL, or relay control.
The GP50 can be controlled locally, using the front panel keypad and display, or from a remote host computer with a Dionex DX-LAN™ interface installed and
®
PeakNet ava ilabl e from an y de vi ce capabl e of pro v iding compati ble TTL signa ls to contr ol the pump.
The pump's two basic mode s of control, Dire ct control and Method control, e nable it to operat e with or without reference to time-based events.
software installed on t he hos t comput er. Limited remote control is also
1 • Introduction
The GP50 is available in four versions. An optional vacuum degas pump is available for all versions:
GP50 Gradient Pump V e r sio n With Degas Pump Without Degas Pump
Standard bore with PEEK components
Standard bore with stainless steel components
Microbore with PEEK components P/N 054045 P/N 054429 Microbore with stainless steel
components
Doc. 031377-03 7/01
P/N 054426 P/N 054427
P/N 054419 P/N 054420
P/N 054425 P/N 054424
1-1
GP50 Gradient Pump

1.2 About This Manual

Chapter 1
Introduction
Chapter 2
Description
Chapter 3
Operation and
Maintenance
Chapter 4
Troubleshooting
Chapter 5
Service
Appendix A
Specifications
Appendix B
Installation
Appendix C
Display Screens
Appendix D
TTL and Relay
Control
Gives an overview of the GP50 Gradient Pump, and explains conventions used in this manual, including safety-related information.
Describes the GP50 front panel controls, electronic and mechanical components, and operating modes.
Provides an overview of GP50 opera tion, including how to create, edit, and run methods from the GP50 front panel. Lists routine preventive maintenance procedures.
Lists problems and presents step-by-step procedures for how to isolate and eliminat e them.
Provides step-by-step instructions for routine service and parts replacement procedures.
Lists the GP50 specifications and installation site requirements.
Describes how to install the GP50.
Illustrates and describes all of the screens that can be displayed on the GP50 front panel.
Describes relay and TTL input and output functions and provides installation instructions
.
1-2
Doc. 031377 -03 7/01

1.2.1 Saf et y Me ss a ges and Note s

This manual contains warnings and precautionary statements that, when properly followed, can prevent personal injury and/or damage to the GP50. Safety messages appear in bold type and are accompanied by icons, as shown below.
Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury.
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury.
1 • Introduction
Doc. 031377-03 7/01
Indicates that the function or process of the instrument may be impaired. Operation does not constitute a hazard.
Informational messages also appear throughout this manual. These are labeled NOTE and are in bold type:
NOTE NOTES call attention to certain information. They alert
you to an unexpected result of an action, suggest how to optimize instrument performance, etc.
1-3
GP50 Gradient Pump

1.2.2 Saf et y Label s

The TUV GS, C, US Mark safety label and the CE Mark label on the GP50 indicate that the GP50 is in compliance with the following standards: EN 61010-1:1993 (safety), CAN/CSA-C22.2 No. 1010.1-92 (safety), UL 3101-1/10.93 (safety), EN 50082-1:1992 (susceptibility), and EN 55011:1991 (emissions).
The symbols below appear on the GP50, or on GP50 labels.
Alternating current
Protective conductor terminal
Power supply is on
Power supply is off
1-4
Doc. 031377 -03 7/01
2 • Description
The GP50 Gradient Pump consists of two units (see Figure 2-1). The upper unit houses the electronics components and the lower unit houses the pump heads and other mechanical pump assemblies.
Doc. 031377-03 7/01
Pump Heads
Figure 2-1. GP50 Enclosure
2-1
GP50 Gradient Pump

2.1 Front Control Panel

The control panel on the upper door of the pump enclosure contains a liquid crystal display (LCD), a membrane keypad, and the actuator for the main power switch (see Figure 2-2). The door opens to provide access to the electronics chassis (see Section 2.2).
NOTE If no keypad buttons are pressed within a two-hour
period, the front panel backlight will automatically turn off. To restore the backlight, press any button.
Screen Contrast
Information is displayed on the LCD, also called the contrast, use the knurled knob in the recess below the keypad (see Figure 2-2).
Tilt Panel
To maximize visibility, the front control panel can be tilted to four different positions. To tilt the panel, support the door at the left side (to prevent it from opening) and lift firmly on the tab in the middle of the recess below the keypad (see Figure 2 -2). Push on the tab t o return the panel to its vertical position.
. To adjust the screen
screen
2-2
Power Switches
The main powe r swit ch is on the bulkhead behind the uppe r door (see Figur e 2-1). An actuator for the main power switch is on the outside of the front door, at the lower left corner (see Figure 2-2).
The actuator functio ns only when the door is fully clos ed. Whe n the door is open, press the main power switch on the bulkhead to turn the module off and on.
To prevent damage to the pump circuitry and components, always wait at least 15 seconds after powering down before turning on the power again.
Doc. 031377 -03 7/01
GP50 G radient Pump
100.0 %A
0.0 %B
0.0 %C
0.0 %D
Help Message
LOAD
COLUMN A
2000
2.00
LOCAL 2
10.27
METHOD
2•Description
PSI m L /MIN
MIN
Off/O n
Prime Reset
Insert
Delete
Ma in P o wer Sw itc h A c tu a to r
Hold/Run
Select
Select
Tab (for opening the door)
Help Menu
Figure 2-2. GP50 Display and Keypad Layout

2.1.1 Control Panel Keypad

Use the keypad to directly control pump operation, as well as to create and modify programmed series of timed events, called summary:
Press
_
_
To go from a menu to a screen, press the numeric button that corresponds to the screen’s number on the menu, or move the cursor to the desired screen name and press
to display a list of available screens.
Menu
Knob (for adjusting the contrast)
7
4
1
0
Enter
8
5
23
Tab (for tilting the panel)
9
6
Enter
methods
.
. In
Doc. 031377-03 7/01
Only fields shown in reverse video on a screen can be edited. Other
_
fields display information only. To edit a value in a reverse video field, use the four directional arrow
_
buttons to move the cursor to the field. Use the numeric buttons to
2-3
GP50 Gradient Pump
enter variable values, or use the choose from among predetermined options.
Select
and
Select
buttons to
Keypad Buttons
To confirm the selected value, press
_
some screens or screen fields, pressing
NOTE A high-pitched beep sounds when you press a
button. When an error occurs, this beep is lower in frequency. The beeps can be disabled from the
MODULE SET-UP
Function
Turns the pump motor off and on. In Direct control (see Section 2.9), turnin g on the m otor cause s it to pump isocratically using the displayed eluent percentages and flow rate. In Method control (see Section 2.9.3), turning on the moto r causes it to pump at the eluent percentages and flow rate for the elapsed time of the selected method, or at the initial conditions (when the method clock is at
).
INIT
This button is used when priming the pump heads. pump to run at maximum vo lume (2.5 mL/min, microb ore; 10.0 mL/min, standard bore). If the pump motor is off when pump automatically turns on. To exit priming and return to the normal flow rate, press See Section B.2.8 for detailed priming in str uctions.
Prime
screen (see Section C.1.7).
again or press
or an arrow button. In
Enter
is required.
Enter
Prime
is pressed, the
Prime
Off/On
to turn off the pump motor.
causes the
2-4
Inserts a new step into a method. This button functions only when the cursor is in a
1. Move the cursor to the added below the cursor position. Parameter values in the new step are blank.
2. Fill in the time value and press
Note:
time value, the inserted step is not saved because it is incomplete.
3. Insert steps in any order. When you press automatically reorganized in the correct chronological order.
TIME
field on the
If you move the cur sor out of the
Table 2-1. GP50 Front Panel Buttons
METHOD
field and press
TIME
METHOD
or
or a cursor arrow button.
Enter
TIME
Enter
extension screen.
. The new step is
Insert
field before entering a
, they will be
Doc. 031377 -03 7/01
2•Description
Keypad Buttons
Function
Removes the value from the current entry field. To restore the previous value, move the cursor from the field before entering a new value.
Pressing screen “blanks” the step parameter value. Moving the cursor to another field does not restore the previo us v alue; ins tead, the step remains blank, indicating no change from the previous step.
To delete an entire method step:
1. Position the cursor in the method’s time field and press
2. Press
Turns the method clock off ( only when the pump is under Method control (see Section 2.9.3).
When the method clock is in Hold, pressing at either the initial step of a new method or, if resuming an interrupted method, at the time at which the clock was put in Hold.
When the method clock is in Run, pressing this “holds” the method and freezes the current conditions.
Delete
time is removed and the help line prompts you to press again to delete the step.
parameters, press any button except
when the cursor is in a step entry field o n the
Delete
again. Or, to restore the original time and step
.
Delete
) and on
Hold
(Run)
Hold/Run
Hold/Run
METHOD
. The
Delete
Delete
. This button functions
starts the clock
stops the clock;
Doc. 031377-03 7/01
Changes the method clock time to specified by the method to occur. This button functions only when the GP50 is under Method control (see Section 2.9.3).
If the method is running, it continues running. If the method is in Hold, the method clock executes the initial conditions and holds.
When the cursor is positioned at a field with predetermined parameters, these buttons cycle through the options. In fields with predetermined numeric values, pressing
pressing
Select Enter
button increases (or decreases) the value continuously. Press
or a cursor arrow button to confirm the selected value.
Table 2-1. GP50 Front Panel Buttons (Continued)
decreases the value by one unit. Holding down a
Select
Select
, causing the initial conditions
INIT
increases the value by one unit;
2-5
GP50 Gradient Pump
Keypad Buttons
Function
The four cursor buttons move the cursor, in the direction of the arrow, to the next entry field. If there is no changeable field in that direction, the cursor moves diagonally or remains in its current location.
In most cases, after entering a new value in an entry field, pressing an arrow button saves and/or executes the change. The exceptions are the
field, the
METHOD SAVE TO
command, the Displays a help screen with information pertaining to the current entry
field.
Displays one of three menus, depending on the current screen:
_
From an operational screen, pressing
SCREENS
_
From a diagnostic screen, pressing
DIAGNOSTIC MENU
From a calibration screen, pressing
_
CALIBRATION MENU
Enters numeric values into the current entry field. The numeric buttons are 0 through 9 and the decimal.
DIAGNOSTIC TESTS
.
METHOD RUN
.
.
screen, and all menu screens.
Menu
Menu
Menu
any calibration
field,
displays the
displays the
displays the
MENU of
2-6
From a menu screen, press ing a nu meri c b utt on open s t he co rres pon ding screen.
Saves and/or executes changes made in entry fields. If a menu screen is displayed, pressing
Table 2-1. GP50 Front Panel Buttons (Continued)
opens the highlighed screen.
Enter
Doc. 031377 -03 7/01

2.1.2 Display Screens

When the pump has successfully powered-up and passed all diagnostic tests, the the power-up, the screen. See Section C.2.7 if this occurs.
POWER-UP
screen (see Figure 2-4). If one of the diagnostic tests fails at
MAIN
DIAGNOSTIC TEST
2•Description
screen (see Figure 2-3) displays briefly, followed by
screen displays instead of the
MAIN
NOTE
Help Message
The selecting the screen from the (see Section C.2.1).
GP50 GRADIENT PUMP
Figure 2-3. Power-Up Screen
100.0 %A
0.0 %B
0.0 %C
0.0 %D
Help Message
POWER-UP
screen can be viewed at any time by
PUM P H EA D VO LUM E 100 uL
MODULEWARE REV n.nn
B IO S R E V n.n n
LOAD
2000
COLUMN A
2.00
10.27
LOCAL 2
METHOD
DIAGNOSTIC MENU
PSI
mL /MIN MIN
Doc. 031377-03 7/01
Figure 2-4. Main Screen
The allow viewing from a d istance. Use the parameters, such as the flow rate and the percentages of eluents to run.
T o access oth er GP50 screens, pre ss the
of SCREENS
screen displays status information in enlarged char acters to
MAIN
screen to select operating
MAIN
button to displ ay th e
Menu
(see Figure 2-5).
MENU
2-7
GP50 Gradient Pump
There are two ways to select a screen from a menu:
Press the numeric button on the front panel keypad that corresponds
_
to the screen number on the menu. For example, p re ss 3 to di splay th e
METHOD
Move the cursor to the field containing the screen number and press
_
Enter
See Appendix C for a description of each screen.
screen.
.
MENU of SCREENS
1
M AIN S CR E EN
2
DETAIL SCREEN
3
METHOD
4
DEGAS OPTIONS
Help Message
Figure 2-5. Menu of Screens

2.2 Electronics Chassis

The electronics chassis is located behind the upper door of the pump enclosure. The chassis includes se v eral ele ctronics car ds (printed ci rcuit boards) that are used to control the pump. Connectors on the cards also allow communication between the pump and other Dionex chromatography modules. Figure 2-6 shows the electronics components with the upper door open. To open the door, pull on the tab located to the right of the main power actuator (see Figure 2-2).
Do not remove any of the electronics cards from the pump. There are no user-serviceable components on the cards. If servicing is required, it must be performed by qualified personnel and appropriate electrostatic discharge (ESD) handling procedures must be followed.
5
MOD ULE SET-UP PUMP OPTIONS
6 7
TIME FUNCTION IN
8
DIAGNOSTIC MENU
2-8
Ne retirez aucune des cartes électroniques de la pompe. Auc un des composants sur les cartes ne peut être réparé par l'utilisateur. Toute réparation doit être effectuée par un personnel qualifié utilisant des procédures correctes de décharge électrostatique.
Doc. 031377 -03 7/01
2•Description
PWR SPY
L C
L E A K
GP50/IP25-DSP130W
DIST
MOTOR
SLOT 2SLOT 1
BLANK
D
L C
C O M M
P6
I
L
S
C
T
A
1
I
3
R
P I N
P7
D I S T
1
2 P
I N
P8
SLOT 3 BLANK
SLOT 4
LAN-000K
RLY-1 OUT
RLY-2 OUT
+
TTL-1
-
OUT
+
TTL-2
-
OUT
+
TTL-1
-
IN
+
TTL-2
-
IN
+
TTL-3
-
IN
+
TTL-4
-
IN
POWER SUPPLY GREEN - OK RED - FAULT
SLOT 5
CPU/RLY
F R O N T
P A N E L
Figure 2-6. GP50 Electronics Chassis
(Located behind pump door)
LC LEAK
The leak control cable from the LC10 Chromatography Organizer or LC20 Chromatography Enclosure, connects t o the
LC LEAK
connector in
slot 1. When a leak occurs i n the LC10 or LC2 0 it is re ported to t he GP50.
Doc. 031377-03 7/01
NOTE The LC25 Chromatography Oven and LC30
Chromatography Oven do not connect to the GP50
LEAK
connector. These ovens contain internal leak
LC
control electronics.
LC COMM
The LC30 Chromatography Oven's RJ-11 serial cable connects to the
connector in slot 1. When connected, the LC30 can be remotely
COMM
controlled by the PeakNet workstation.
LC
2-9
GP50 Gradient Pump
LC AIR
The cable from the ai r soleno id v alv es in t he LC10, LC20, L C25, or LC3 0 chromatography module conne cts to the connected, the GP50 can electrically actuate the solenoid valves that control the position of the injection valve and the optional column switching valve in the chromatography module.
LC AIR
connector in slot 1. When
To select the valve positions, go to either the Section C.1.2) or the
TTL/RELAY
METHOD
screen (see Section C.1.4).
MAIN
screen (see
A strip of eight relay and TTL connectors is located in slot 4. These connectors interface with Dionex and non-Dionex modules for relay and TTL control of the pump. Appendix D describes the relay and TTL functions and the connections between the GP50 and other modules.
CPU
Control Moduleware for the pump resides on the CPU/Relay cards. The CPU logic and Relay I/O cards occupy slot 5. The Relay I/O card
rides piggybac k o n t he CPU card and exten ds over the f ro nt of slot 4. The card is short enough to allow the optional DX-LAN pump interface card (P/N 044195) to mount behi nd it in slot 4. A 60- pin ri bbon cab le lin ks the CPU logic to the display and keypad. The logic monitors the internal power supply outputs, and reports the status on the multicolored LED at the bottom of slot 4.
Green indicates normal operation.
_
Red indicates a power fault. The GP50 will enter its diagnostic state
_
and inhibit all other cont rols unti l the f ault is corr ected . If this occurs, turn off the power for a few seconds and then turn it back on.
2-10
Doc. 031377 -03 7/01

2.3 Mechanical Chassis

The mechanical chassis is housed in a pull-out drawer located behind the lower door of the pump enclosure. The front of the chassis contains the components described in Section2.4. Other mechanical assemblies are located inside the chassis drawer. The drawer should be pulled out only for service procedures. For routine operation, push in the drawer and tighten the lock located on the lower right corner of the chassis.
Observe the warning label on the inside of the lower door. The arrows on the label indicate moving mechanical parts that present pinch hazards when the pump is on and the mechanical drawer is open. Do not operate the pump with the mechanical chassis drawer pulled out.
Respectez l'étiquette d'avertissement apposée à l'intérieur de la porte inférieure. Les flèches sur l'étiquette indiquent des pièces mécaniques mobiles qui posent un danger de pincement lorsque le GP50 est sous tension et le tiroir mécanique est ouvert. N'ut ilisez jamais le GP50 avec le tiroir du châssis mécanique ouvert.
2•Description
Doc. 031377-03 7/01
2-11
GP50 Gradient Pump

2.4 Mechanical Components

Figure 2-7 shows the mechanical componen ts located behi nd the lo wer door of the enclosure.
Pressure
Transducer
Pressure
Transducer
Waste Valve
Pr im ing
Block
Pump Heads
2-12
Eluent
Manifold
Proportioning
Figure 2-7. GP50 Mechanical Components
Doc. 031377 -03 7/01

2.4.1 Pump Eluent Manifold

Eluent lines A through D are routed through the rear of the GP50 to the vacuum de g as pump ( if ins ta ll ed) (see Sec ti on 2.5) and then to the rear of the proportioning valve. If the vacuum degas pump is not installed, the eluent lines are connected directly to the proportioning valve. The proper proportion of eluent exits the front of the valve and is directed to the priming block. Figure 2-8 shows the eluent flow path through the system.
2•Description
Eluent Reservoirs
VAC Chamber IN
VAC Chamber OUT
ABCD
Pump H eads
OUT
D
Proportioning Valve/
A
C
Manifold Assembly
B
Pressure Transducer
Figure 2-8. Eluent Flow Schematic
Mixer To Column
Rinse Ports
Priming Block
Doc. 031377-03 7/01
2-13
GP50 Gradient Pump

2.4.2 Pump Heads

There are two types of pump heads: standard bore and microbore. The table below summarizes the features and operating conditions for each type.
Pump Head Type
Standard Bore
Microbore 25µL Isocratic and
*Flow rates are adjustable in increments of 0.01 mL/min.
NOTE Although there is some overlap in flow rates
Piston Volume
100µL Isocratic 0.04–10.0 35 MPa
between the two pump configurations, continuous operation of the microbore pump at flow rates above 2.0 mL/min will decrease seal and pump life. For the best extended operation at 2.0 mL/min or above, use a standard bore pump. To achieve optimum performance at flow rates below
0.4 mL/min, use a microbore pump.
Pump Operation
Gradient 0.4–10.0 35 MPa
Gradient
Flow Rate (mL/min)*
0.0–2.50 35 MPa
Maximum Operating Pressure
(5000 psi)
(5000 psi)
(5000 psi)

2.4.3 Pump Mixers

A pump mixer ensures complet e mixi ng of the proportioned eluents prior to injection. The mixer is installed between the pump outlet and the injection valve. Three types of mixers are available:
_
_
_
2-14
In standard bore PEEK pumps, a GM-5 Mixer (P/N 054044) is used. In microbore PEEK pumps, a GM-4 Mixer (P/N 049136) is used. In stainless ste el pumps , a Sta inles s Ste el Mi x er (P /N 0 54043) is use d.
Doc. 031377 -03 7/01

2.4.4 Pump Priming Block

The priming block “tee” directs the flow of eluent fr om the manifo ld i nto the pump heads. The priming block is also used for rapid removal of air from the system.
Refer to Section B. 2.8 for instructions on priming the pump heads.

2.4.5 Pressure Transducer

From the priming block, the liquid stream is directed to the inlet check valves on the pump heads, through the pump heads, and finally through the outlet check valves to the pressure transducer.
Flow from the outlet check valves on the pump heads is combined in the pressure transducer. The pressure transducer measures the system pressure at this point. The interactive constant-flow/constant-p ressure control program on the DSP precisely controls the pump motor speed to assure flow rate accuracy.
A waste line exits the bottom of the pressure transducer. Opening the valve on the pressure transducer diverts flow to the waste line and relieves system pressure, forcing air out of the system.
2•Description
Doc. 031377-03 7/01
Flow output from the pressure transducer is directed out of the pump module, through the gradient mixer, and on to the rest of the chromato graphy system (inj ection valve, column, detector).
See Section B.2.6 for eluent outlet line connections.
2-15
GP50 Gradient Pump

2.5 Vacuum Degas Pump Assembly (Optional)

The Dionex vacuum degas pump provides continuous on-line vacuum degassing of up to four eluents. The assembly, which must be installed at the factory, consists of :
A 4-channel degas chamber (with degas membranes) with 17 mL internal
_
capacity p er channel A dual-stage diaphragm vacuum pump
_
A solenoid valve
_
An on-board vacuum sensor
_
The electronics required to operate the vacuum pump
_
Fittings, tubing, and other accessories
_
By default, the de gas pump tur ns on f or 2 minu tes when th e GP50 power is turned on. Thereafter, the pump turns on for 30 seconds at 10-minute intervals. The
DEGAS OPTIONS
Section C.1.6). You can check the vacuum chamber pressure from the
STATUS
screen (see Section C.3.4).
screen allows you to change the cycle time and duration (see
DEGAS
NOTE All components of the vacuum degas assembly are made
of inert materials or corrosion-resistant materials. However, Dionex recommends thoroughly flushing any chemicals out of the tubing with deionized water after each use to avoid crystallization in the membrane pores.

2.6 Piston Seal Wash (Optional)

When using highly concentrated buffer solutions, Dionex recommends continuously rinsing the piston seal with a piston seal wash. Rinsing removes salt crystals that may abrade the piston, thereby causing the seal to wear out prematurely and allow leaks.
Dionex offers a continuous seal wash kit (P/N 059187) for the GP50. For installation instructions, see Section B.2.8 or the instructions included with the kit.
2-16
Doc. 031377 -03 7/01

2.7 Eluent Reservoirs

NOTE Dionex strongly recommends degassing all eluents and
storing them in reservoirs pressurized with helium. This helps prevent bubbles (resulting from eluent outgassing) from forming in the el uent proportionin g valves, pump heads, and the detector cell.
Degassed eluents and pressurized reservoirs are especially important when combining aqueous and non­aqueous components (e.g., water and methanol). Pressurizable reservoirs allow eluents to be stored under a specific atmosphere.
The following pressurizable reservoirs are available from Dionex:
1-liter glass reservoirs with shatterproof plastic coating (P/N 044126)
_
2-liter glass reservoirs with shatterproof plastic coating (P/N 044127)
_
1-liter plastic reservoirs (P/N 044128)
_
2•Description
2-liter plastic reservoirs (P/N 044129)
_
Do not use the 2-liter plastic reser voir (P/N 044129) for off-line vacuum degassing of eluents. The reservoir was not designed for this purpose.
N'utilisez pas le réservoir en plastique de 2 litres (N /P 044129) pour le dégazage à vide hors ligne d'éluants. Le réservoir n'a pas été conçu à cette fin.
Refer to the
Pressurizable Reservoir Installation Instructions
(Document No.
034581) for installation details.
EO1 Eluent Organizer (Optional)
The Dionex EO1 Eluent Organizer (P/ N 044 125) holds eluent reservoirs in a liner that contains spills and leaks. The EO1 can also be used to pressurize reservoirs. Up to two optional E01 Eluent Organizers can be places on top of the system
Doc. 031377-03 7/01
2-17
GP50 Gradient Pump
enclosure. Each organizer can accommodate one or two reservoirs, depending on the type of reservoir used (see the table below).
No. of Reservoirs
2 1-Liter glass or plastic 2 L 2 2-Liter plastic 4 L 1 2-Liter glass 2 L
Description Total Volume

2.8 Rear Panel

The rear panel contains the main power receptacle and fuses. It also includes a DX-LAN connector for interfacing the GP50 with the PeakNet workstation. The rear panel is illustrated in Figure B-1.

2.9 Functional Description

2.9.1 Operating and Control Modes

The operating mode determines commands:
_
In Local mode, the GP50 receives commands from the front control panel buttons and screens.
in Liters
how
the GP50 receives operating
2-18
In Remote mode, PeakNet 5 software sends commands from the host
_
computer via the DX-LAN int erfac e. Limited o perating changes fr om the front panel are allowed.
_
In Locked Remote mode, PeakNet 5 or PeakNet 6 software sends commands from the host computer via the DX-LAN interface. All operating changes from the f ront pane l are disab led. When PeakNet 6 software is controlling the GP50, the pump is always in Locked Remote mode.
The control mode determines
_
In Direct control, the GP50 exe cutes commands immediately.
when
operating commands are executed.
Because there is no time-ba sed program, th e method clock i s not used and the
Hold/Run
and
buttons do not operate.
Reset
Doc. 031377 -03 7/01
2•Description
_
In Method control, the GP50 executes commands according to the timed steps in a method. The method is programmed from the GP50 front panel or from PeakNet 5 software. See Section 2.9.3 for details about Method control from the front panel. using PeakNet 5 to program and run methods, refer to the onli ne Help or user’s guide.
The table below summarizes the various operating and control mode configurations. Select the modes from the Section C.1.2),
DETAIL
screen (see Section C.1.3), or chromatography
software.
Operating/Control Mode Pump Operat ion
Local/Direct Control Commands are entered from the GP50 front control
panel and executed immediately after being entered.
Local/Method Commands are entered from the GP50 front control
panel and executed by running a method programmed from the front panel.
Remote/Direct Control Commands are sent from PeakNet 5 and executed
immediately when received.
For information about
screen (see
MAIN
Locked Remote/Direct Control
Remote/Method Commands are sent from PeakNet 5 and executed by

2.9.2 Local and Remote Modes

Local Mode
When the GP50 power is turned on, the pump is in Local mode. In Local mode the pump accepts operating commands from two sources:
_
Direct input from the front panel keypad and screens. With direct input, all GP50 operating functions are available.
_
TTL inputs from a remote controller (for example, a Dionex detector module or autosampler). TTL signals can be used to turn the pump motor on and off, turn the method clock on and off, and increment or decrement the method number.
Doc. 031377-03 7/01
Commands are sent from PeakNet 5 or PeakNet 6 and executed immediately when received.
running a method programmed in PeakNet 5.
2-19
GP50 Gradient Pump
Remote Mode
In Remote mode, PeakNet 5 software sends operatin g commands fr om the host computer via the DX-LAN interface. In Remote mode, operating parameters can be changed from the front panel, provided they do not interfere with a running method.
Locked Remote Mode
In Locked Remote mode, PeakNet 5 or PeakNet 6 software sends operating commands from the host computer via the DX-LAN interface. In Locked Remote mode, all operating changes from the front panel are disabled.
When the GP50 is controlled by PeakNet 5, select the Locked Remote Start option from the PeakNet 5 Run program to operate in the Locked Remote mode. To return the GP50 to Local mode, clear the Start opt ion or turn the GP50 power off and then on.
When the GP50 is controlled by PeakNet 6, connecting to the PeakNet 6 timebase automatically selects the Locked Remote mode. To return the GP50 to Local mode, either clea r the Connect check box on the PeakNet 6 control panel or turn the GP50 power off and then on.

2.9.3 Method Control

In Method control, commands are executed according to the time-based steps programmed in a method. Each step specifies the eluent composition and flow rate to be delivered by the pump at a given time. The selected eluent mixture is delivered either isocratically, or as a multistep linear or curved gradi ent. As the method runs, the GP50 calculates the changes i n eluent compositio n require d to deli v er a gradient from one method step to the next or to match the selected curve.
2-20
Doc. 031377 -03 7/01
2•Description
Methods are programmed, saved, and edited from the
METHOD
(see Figure 2-9). See Section 3.3 for programming instructions.
M E TH OD E DIT
SAVE TO RUN 100605
LIMITs 0 - 5000
TIME % A
IN IT
0.00
123.45
345.67
25.0
100.0
10.0
17.2
25.0
0.0
22.2
19.6
25.0
0.0
32.3
33.2
25.0
0.0
35.5
30.0
V
C%B %C %D
L1.00
0
I L
5
FLOW
2.00
Help Message
Figure 2-9. Method Screen
NOTE For information about using PeakN et 5 to program and
run methods, refer to the online Help or user’s guide.
Here is a summary of basic information about using methods.
Each method can contain up to 50 time-based steps. Step 1 always
_
starts at The GP50 can store up to 100 separate methods (0 through 99) in
_
(initial condition). Step 2 always starts at
INIT
memory. Methods are retained in memory even after the pump is powered down.
PSI
> >
> >
TIME = 0.0
screen
.
Doc. 031377-03 7/01
The total number of methods that can be stored in memory depends on the length of each method and the amount of available memory; thus, the actual total may be less than 100.
The pump can run under method control while you are entering or
_
editing any method. When you save changes to the currently running method or switch to
_
a different method, the method clock continues running unaffected. Only those parameter changes which affect the method
after
the
current time will be implemented in the current run.
2-21
GP50 Gradient Pump

2.9.4 Eluent Delivery

Isocratic Eluent Run
The simplest use of the GP50 is for the delivery of an isocratic (unchanging) mixture of one or more eluents. If more than one eluent is selected, the pump d elivers a proport ional mixt ure of the el uents based on the percentage of each eluent selected. The combined percentages of all eluents selected must total 100% or the pump will not run.
Gradient Eluent Run
The GP50 can produce step, linear, concave, or convex curves in eluent concentration over a specified time period. The slope of the gradient is determined by the selected gradient curve (see Figure 2-10) and the time between the starting and end ing points of a grad ient step. It is important to note the following points:
The curve number p arameter determines whethe r the pump deliver s a
_
linear or curved gradient.
2-22
1.0 .9
.8
.7 .6 .5
.4
.3 .2 .1 0
1
0
1
2
3
4
5
6
7
8
9
2
45678
3
910
Figure 2-10. Eluent Composition Profile for Curves 1 - 9
Doc. 031377 -03 7/01
2•Description
Curve numbers are defined as follows:
Curve Number Gradient Type
1, 2, 3, 4 Convex 5 Linear (power-up default) 6, 7, 8, 9 Concave
NOTE A curve number in a step instructs the pump to use
the selected curve number when moving from the previous step to that step. Because there are no previous steps for cannot be entered for these steps.
The gradient slope does not change during a step if curve 5 (the
_
INIT
default setting) is selected, because curve 5 represents a linear gradient.
Convex curves cause rapid changes in eluent composition at the
_
beginning of the curve and slower changes at the end. Concave curves cause slower changes at the beginning and rapid changes at the end.
TIME=0.0
or
, curve numbers
Slope changes o ver time become mor e ex treme as c urves go from 6 t o
_
9 (more concave) and from 4 to 1 (more convex). Figure 2-10 shows the eluent composition profiles corresponding to curves 1 through 9, normalized for 0–100% for 10 minutes.
Any consecutive method steps specifying identical eluent
_
compositions will generate an isocratic segment, regardless of the curve number selecte d.
A step change is a steep linear gradient in which the eluent
_
composition changes from one eluent to another within 6 seconds.
Doc. 031377-03 7/01
2-23
GP50 Gradient Pump
2-24
Doc. 031377 -03 7/01
3 • Operation and Maintenance

3.1 Getting Ready to Run

NOTE The GP50 Gradient Pump is designed to perform ion
chromatography and BioLC applications only and should not be used for any other purpose. If there is a question regarding appropriate usage, contact Dionex.

3.1.1 Degas Eluents

Dionex strongly recommends degassing all eluents and storing them in reservoirs pressurized with filtered inert gas (see Sect ion 3.1. 3). This helps prevent bubbles caused by eluent outgassing from forming in the eluent proportioning valves, pump heads, and the detector cell. Degassed eluents and pressurized reservoirs are especially important when combining aqueous and nonaqueous components (e.g., water and acetonitrile).
The GP50 with the optional vacuum degas pump assembly provides continuous on-line vacuum degassing of up to four eluents.
Doc. 031377-03 7/01
If the GP50 is not equipped with the vacuum degas assembly, manually vacuum-degas eluents daily, as described below, and store them in pressurized reservoirs.
Degassing Eluents Manually
1. Prepare the eluent required for your applicat ion. Pour it into a v acuu m flask and attach the flask to a vacuum pump or water aspirator.
2. Vacuum degas the eluent for 5-10 minutes by shaking or sonication.
3. Remove the flask from the vacuum.
the aspirator back into the flask.
4. Pour the degassed eluent into a press uriza ble res ervoir . Be car eful not to shake the e luent.
5. Install end-line filters and press urize the re servoi rs (see Sectio ns 3.1.2 and 3.1.3).
Do not allow water to flo w fr om
3-1
GP50 Gradient Pump

3.1.2 Filter Eluents

Always f il t er e lue nt s bef or e beginning operation. Fil ter in g re moves small particulates that may contaminate the eluent proportioning valves or the pump check valves and cause erratic flow rates or loss of prime. The pressurizable reservoir Ship Kits supply end-line filters (P/N 045987) for this purpose.
Install an end-li ne f il ter on the end of each elu ent li ne insi de the re se rv oir. To prevent air from being drawn through the lines, mak e sure that the end of each filter r eaches the bot tom of the el uent reser voi r and that eac h fi lter is submerged in eluent.

3.1.3 Pressurize Eluent Reservoirs

The GP50 Gradient Pump is capable of operation with or without head pressure on the eluent. Pressurization of the eluent reservoirs, if used, should be with filtered inert gas (preferably helium). Refer to the
Pressurizable Reservoir Installation Instructions
1. V erify that a re gu lator (P/ N 046594) is install ed on the g as supp ly line to the reservoirs.
for details.
3-2
2. Turn on the gas supply and adjust the pressure to 55 KPa (8 psi).
Never pressurize the reservoirs above 69 KPa (10 psi).
Ne mettez jamais les réservoirs d'éluants sous une pression supérieure à 69 kPa (10 lb/po²) .
Doc. 031377 -03 7/01

3.1.4 Start-Up

1. Turn on the main pump switch.
3 • Operation and Maintenance
The
POWER-UP
screen displays (see Figure 2-4). A series of diagnostics tests is
MAIN
run at power -up. If one of the test s fail s, the
screen displays briefly (see Figure 2-3) and then the
DIAGNOSTIC TEST
displays instead. See Section C .2.7 if this occurs. When the GP50 powe r is turned on, the injection v alv e is initia lized to
the Load position.
2. Press
Off/On
3. Check the pressure reading on the
to start the pump flow.
screen. The GP50 display
MAIN
updates the pressure readout once pe r piston str oke. The rea ding from one stroke to the next should be within 3%.
A variation of more than 3% indicate s that the pump is out of prime. Refer to Section B.2.8 for priming instructions, or see Section 4.1 for other conditions which can cause the pump to lose prime.
NOTE After starting the pump or changing the flow
rate, wait at least 5 minutes (up to 20 minutes for low flow rates in a standard bore pump) before beginning an analysis. This allows the pump's real-time electronic pulse damping circuitry to stabilize the flow rate.
screen
Doc. 031377-03 7/01
3-3
GP50 Gradient Pump

3.1.5 Selecting the Pressure Limits

The high and low pres sur e li mit s aut omat ic all y st op the pump if a sy stem malfunction occurs, such as overpressurization caused by a blockage or low pressure caused by a leak downstream from the pump.
When PeakNet is controlling the pump, select the pressure limits
_
from the software. When the pump is running under Local mode, Direct control, enter
_
the pressur e limits on the When the pump is running under Local mode, Method control, enter
_
the pressure limits on the are set in the When a limit trip stops t he pump, the method c lock immediate ly stops and goes to Hold. The current status of the program that was running at the time is displayed on the front panel.
To select the pressure limits from the front panel:
step and remain unch anged t hro ughout the ana lysis .
INIT
DETAIL
METHOD
screen (see Figure 3-1).
screen (see Figure 3-5). The limits
1. Go to the
DETAIL
or
METHOD
screen and move the cursor to the
field.
2. Enter a low pressure limit between 1.4 and 2.8 MPa (200-400 psi).
DETAIL SCREEN
100.0 %A
0.0 %B
0.0 %C
0.0 %D
LOCAL
Help Message
Figure 3-1. Detail Screen: Setting Pressure Limits
2125
2.00
IN J E C T
COLUMN
LIMIT
DIRECT CNTRL
PSI m L/M IN SAMPLE A
300-2500 PSI
TTL1 TTL2 RLY1 RLY200
1 1
The setting may vary, depending on the system operating pressure. The low pressure limit is activated after 13 pump piston strokes (i.e., after 1.3 mL (standard bore) or 0.325 mL (microbore) of fluid is pumped through).
3. Enter a high pressure limit that is 2.8 to 3.4 MPa (400-50 0 psi ) above the normal system operating pressure. The pump is equipped with a pressure limit that prevents operation above 35 MPa (5076 psi).
LIMIT
3-4
Doc. 031377 -03 7/01
3 • Operation and Maintenance

3.1.6 Selecting the Operating and Control Modes

1. Go to either the displays displays
2. To select the operating m ode:
_
_
3. To select the control mode, move the cursor to the field, press
Select
a cursor arrow button.
LOCAL, REMOTE DIRECT CNTRL
GP50 is controlled by PeakNet 6
If the
PeakNet 6 timebase a utomatica lly select s the GP50 to PeakNet 6 control panel or turn the GP50 power off and then on.
GP50 is controlled by PeakNet 5
If the
on the
Select Enter
To select Remote Start option from t he Run program. To return the GP50 to
LOCAL
on
MAIN
or
or a cursor arrow button.
LOCKED RMT
, clear the Start op tion o r turn the GP50 po wer of f and then
.
or
Select
or
MAIN
LOCAL
or
Select
DETAIL
or
, either clear the Connect check box on the
DETAIL
to toggle to the desired mode, and press
to toggle to the desired mode, and press
screen. The operating mode field
, or
LOCKED RMT
METHOD
screen. Move the cursor to the field, press
with PeakNet 5, select the Locked
. See the example in Figure 3-2.
. The control mode field
, connecting to the
LOCKED RMT
select
,
LOCAL
. To return
or
REMOTE
or
Enter
25.0 %B
15.0 %C 0 .0 %D
Doc. 031377-03 7/01
60.0 % A
LOAD
COLUMN A
LOCA L
2000
2.00
DIRECT CNTRL
PSI
m L/MIN
Help Message
Control ModeOperating Mode
Figure 3-2. Main Screen: Operating and Control Mode Fields
3-5
GP50 Gradient Pump

3.2 Running Under Direct Control (Local Mode)

Direct Control, Local mode is use d most often when a PeakNet workst atio n is not configured.
Direct Control Example
Specify an isocrati c mi xture of 60% eluent A, 25% eluent B, and 15% eluent C to be pumped at 2.0 mL/min (see Figure 3-3).
3-6
1. Go to the control modes to
Figure 3-4 illustrates the up.
MAIN
60.0 % A
25.0 %B
15.0 %C 0 .0 %D
Help Message
Figure 3-4. Main Screen: Running Under Direct Control
Figure 3-3. Isocratic Run Profile
or
LOCAL
DETAIL
screen and if necessary, change the operating and
and
DIRECT CNTRL
screen as it wil l appear when the e xa mple is set
MAIN
LOAD
2000
COLUMN A
LOCA L
DIRECT CNTRL
(see Section 3.1.6).
PSI
2.00
m L/MIN
Doc. 031377 -03 7/01
3 • Operation and Maintenance
2. Move to the %A field and enter 60; move to the %B field and enter 25; move to the eluent percentages equa l 100).
field and press
%C
(15 is automatically filled in to ma ke the
Enter
3. Move to the
mL/MIN
4. If the pump is currently off, press
field and enter 2.
Off/On
to turn on the motor and begin the
isocratic delivery.
NOTE After starting the pump or changing the flow rate, wait
at least 5 minutes (up to 20 minutes for low flow rates in a standard bore pump) before beginning an analysis. This allows the flow rate to stabilize.

3.3 Running Under Method Control (Local Mode)

This section pro vi des general instructions on how to create, edit , and run methods from the GP50 front panel. For step-by-step examples, see Section 3.4.
When entering parameters on the
After starting the pump or changing the flow rate, wait at least 5 minutes (up
_
METHOD
to 20 minutes for low flow r ates in a sta ndard bore p ump) befor e beginning an analysis. T his allows the flow rate to stabilize.
In the %A, %B, %C, and %D columns, enter decimal percentage values from
_
0.1% through 100% for the eluent compositions. The combined percentages for all eluents must total 100% or the pump will not run.
In the V column, select the injection valve posi tion (L for load or I for inject).
_
screen, observe these guidelines:
In the
_
increments of 0.01 mL/min. See Section 2.4.2 for the available flow rate ranges. They vary, depending on the size of the pump head and whether the run is isocratic or gradient.
For steps other than
_
The curve num ber determines whether the pump delivers a lin ear or curved
Doc. 031377-03 7/01
column, ent er the pump flow rate. Fl ow rates are adjustable in
FLOW
Continuous operation of the microbore pump at flow rates above 2.0 mL/min will decrease seal and pump life. For optimum performance above 2.0 mL/min, a standard bore pump should be used.
INIT
and
TIME=0.0
, enter a curv e number in the C column.
3-7
GP50 Gradient Pump
gradient when moving to the step from the one preceding it (see Section 2.9.4). The default is curve 5 (linear).
NOTE Because there are no previous steps f or
curve numbers are not entered for them.
If a step field is “blank” (has no entry), the last selected value for the field
_
INIT
TIME=0.0
or
remains in effect. The symbol
_
next to the bottom time ent ry indicates there are addit ional steps
below. Move the cursor to the bottom time entry and press the down arrow to see the additional step(s).
The symbol ^ next to the top time entry indicates that it is preceded by at least
_
one more step. Move the cursor to the entry and press the up arrow to see the additional step(s).
The symbol > at the right edge of each line indicates a lateral extension to the
_
line. Move the cursor to the end of a line and press the right arrow to display the
METHOD
extension screen (see Section C.1.5).

3.3.1 Creating a New Method

You can create a new method when the method clock is in either
.
Run
1. Go to the
MAIN
or
DETAIL
operating and control modes to Section 3.1.6).
screen and if necessary, change the
LOCAL
and
METHOD
(see
,
Hold
or
3-8
2. Go to the
3. In the
EDIT
METHOD
field, enter the number of the method to be created. This
screen.
can be the number of an unused method or of an existi ng met hod that you plan to edit and sa v e as a new method. If you enter the number of an unused method, the screen will look similar to th e example screen in Figure 3-5.
Doc. 031377 -03 7/01
3 • Operation and Maintenance
M ETHO D E DIT
TIME %A
100.0
0.00
Help Message
Figure 3-5. Method Screen: Creating a New Method
4. In the
5
%B %C %D V
field, set the low and high pressure limi ts (see
LIMITs
SAVE TO RUN
LIMITs
5
0 - 5000
C FLOW
_
L1.00IN IT
_
0
PSI
> >
> >
Section 3.1.5).
5. Each method starts out with two timed steps (see Figure 3-5): an initial conditions st ep (contai ning zero step (containing 0.00 in the
in the
INIT
column). The parameters in
TIME
column) and a time
TIME
each of these first two steps can be changed but the steps cannot be deleted. Enter the parameters for both steps.
6. Enter a new step using one of the following methods:
Move the cursor to the empty
_
field below the last step and
TIME
enter the elapsed time at whic h to start the new step. Press or a cursor arrow button.
Enter
Doc. 031377-03 7/01
Move the cursor to any of the
_
fields and press
TIME
Insert
. This adds a ne w step af ter th e curs or posit ion. Ente r the elapsed time at which to start the new step and pres s
or a cursor arrow
Enter
button.
After entering a new step, all timed steps are automatically organized in chronological order.
7. Continue entering parameters for the new step. When you finish, move the cursor to the
If you are editing an exist in g metho d, ent er a new number for the
_
method and press If you are editing an unused method, press
_
SAVE TO
Enter
field and do one of the following:
.
.
Enter
3-9
GP50 Gradient Pump

3.3.2 Running a Method

1. If the pump motor is off, press
2. Go to the
MAIN
or
DETAIL
screen and if necessary change the
operating and control modes to
to turn on the motor.
Off/On
LOCAL
and
METHOD
(see
Section 3.1.6).
3. In the
METHOD
# field, enter the desired method number. If the method clock is already runn ing, the method starts immediatel y . If the clock is in Hold, press
NOTE You can also select the method number in the
METHOD
and enter the desired method number.
Hold/Run
screen. Move the cursor to the
to start the method.
RUN
fiel d
4. The elapsed time on the method clock when the method begins determines where (at what step and parameters) the method begins running:
If the method clock is at
_
running using the If the method clock is greater than zero, the method begins
_
ial condition parameters.
INIT
or time zero, the method begins
INIT
running using the par ameters s pec if ie d in the s tep f or t hat e lap sed time. To start the method at the
.
Reset
ial conditions instead, press
INIT
3-10
Doc. 031377 -03 7/01

3.3.3 Editing a Method

Existing methods can be modified by changing, adding, or deleting steps and parameters. Changes can be made when the method clock is stopped or running. If the meth od you ar e editin g is curre ntly runn ing, the cha nges are stored in memory and implemented when you save the method.
NOTE After saving changes, there is no way to recall the
original method. If you plan to make experimental changes to a method but also want to retain the original method, save the modified method to a new number.
To edit a method:
3 • Operation and Maintenance
1. Go to the
METHOD
screen. In the
field, enter the number of the
EDIT
method to be modified.
2. Make the required changes:
To change a parameter
_
, position the cursor in the field and enter
the new value. The previo us value is automatically deleted.
To add a method step
_
, or move the cursor to the empty
Insert
, move the cur sor to an y
TIME
field and press
TIME
field below the last step and enter the elapsed time at which to start the new step. When you press
or a cursor arrow button, the new step is
Enter
automatically moved to the correct chronological position. Continue entering parameters for the new step.
To delete a method step
_
and press
Delete
3. When changes are complete, move the cursor to the Press
to save the changes to the current method, or enter a new
Enter
method number and press
, move the cur so r to th e step to be deleted
twice.
SAVE TO
Enter.
If you save changes to the currently running method, they are immediately incorporated in the run and execut ed at the prog rammed time, the method at the method, press
the modified event has already been executed. To restart
unless
ial conditions and run all steps of the updated
INIT
.
Reset
field.
Doc. 031377-03 7/01
3-11
GP50 Gradient Pump

3.3.4 Deleting a Method

To delete an entire me th od, move the cursor on the
step and press
INIT
Delete
twice.

3.3.5 Changing the Running Method

To change from t he method currently running t o a di fferent method, g o to the field, and press
MAIN
or
DETAIL
Enter
screen, enter the new method number in the
.
The new met hod will begin runnin g, u si ng the parameters speci fied in the step for the current elapsed time. To start the method at the conditions, press
Reset
.

3.3.6 Controlling the Method Clock

The
Hold/Run
and
DETAIL
To start and stop the method clock, press
_
To reset the clock to
_
To set the clock to a specific elapsed time, enter the time in the
_
field on the continue) running, using the method parameters specified for that time.
button, the
button, and the
Reset
screens control the method clock:
ial conditions, press
INIT
MAIN
or
DETAIL
screen. The method will start (or
METHOD
field s in the
MIN
Hold/Run
Reset
screen to the
INIT
.
.
METHOD
ial
MAIN
MIN
3-12
Doc. 031377 -03 7/01

3.4 Example Methods

The examples in this section provide step-by-step instructions for creating three types of methods: iso crati c, linea r gra dient, and cur v ed g radie nt. The l ast ex ample demonstrates how to edit a running method.
3 • Operation and Maintenance
For all of the examples, set the pump to
LOCAL
Section 3.1.6).

3.4.1 Isocratic Method Example

Specify an isocratic mixture of 60% eluent A, 25% eluent B, and 15% eluent C to be pumped at 2.0 mL/min. Figure 3-3 illu strates the isocratic profile for this example. Figure 3-6 illus trates the appears when the example is set up .
M ETHO D E DIT
TIME %A
60 .0_ 25.0 15.0
0.00
Help Message
Figure 3-6. Method Screen: Isocratic Run Example
1. Go to the
%B %C %D V
METHOD
screen and enter a method number in the field (1, for example). The screen auto matically changes the number in the
SAVE TO
field to the number of the method being edited.
SAVE TO
LIMITs 0 - 5000
mode,
METHOD
C FLOW
_
METHOD
RUN
L2.00IN IT
control (see
screen as it
011
PSI
> >
> >
EDIT
Doc. 031377-03 7/01
If Method 1 currently e xis ts and you w ant to ret ain it, e nter a ne w,
_
unused, method number in the If Method 1 currently exists and you want to delete it, move the
_
cursor to
TIME=INIT
2. Move the cursor to the and enter 25; move to the
and press
field and enter 60; move to the %B field
%A
field and press
%C
EDIT
Delete
field.
twice.
Enter
(15 is automatically filled in to make the eluent percentages equal 100). Ignore the
(Curve) an d V (Valve) fields. Mov e to the
C
FLOW
enter a flow rate of 2.00.
field and
3-13
GP50 Gradient Pump
3. Move the cursor to
4. Move the cursor to press off, press
to select the programmed method. If the pump motor is
Enter
Off/On
SAVE TO
Run
to have the pump st ar t del ivering the eluen t mixt ur e.
5. If the method clock is in hold, press
and press
, enter the method number (1, in this case) and
Hold/Run
method.

3.4.2 Linear Gradient Method Example

The following summarizes the linear gradient method steps:
Create Method 2 to be gin under i socratic co nditions with 10 0% eluent
_
A at 2.0 mL/min. After 5 minutes, begin adding eluent B and decreasi ng eluen t A until,
_
at 10 minutes, the mixture is 65% eluent A and 35% eluent B. Begin adding eluent C to the mixture while continuing to decrease
_
eluent A and increase eluent B until, at 15 minutes, the eluent composition is 0% eluent A, 50% eluent B, and 50% eluent C.
Continue increasing eluent C and begin decreasing eluent B until, at
_
25 minutes, the eluent composition is 100% eluent C. Make a step change to 100% eluent D at 25.01 minutes. Pump 100%
_
eluent D for 4.99 minutes.
to save the method.
Enter
to begin running the
3-14
At 30.01 minutes, return to 100% eluent A and re-equilibrate your
_
system for the next analysis.
Figure 3-7 illustrates the gradient profile for this method.
Doc. 031377 -03 7/01
3 • Operation and Maintenance
Figure 3-7. Linear Gradient Method Profile
1. Go to the
METHOD
screen and enter a method number in the
field (2, for example).
If Method 2 currently e xis ts and you w ant to ret ain it, e nter a ne w,
_
unused, method number in the If Method 2 currently exists and you want to delete it, move the
_
cursor to
2. Move to the
TIME = INIT
FLOW
Figure 3-8 illustrates the
and press
field of the
METHOD
INIT
field.
EDIT
Delete
twice.
step and enter a flow rate of 2.00.
screen as it app ear s so far. You can now begin entering the method steps that will generate the gradient profile.
M ETHO D E DIT
2
SAVE TO RUN 0
2
LIMITs 0 - 5000
TIME %A
100.0
%B %C %D V
C FLOW
_
L2.00IN IT
_
0.00
Help Message
Figure 3-8. Linear Gradient Method Example (After Step 2)
PSI
> >
> >
EDIT
Doc. 031377-03 7/01
3-15
GP50 Gradient Pump
3. Move the cursor to the %A field of the
TIME = 0
step and press
100% of eluent A is automatically filled in.
4. Move the cursor to the line below at
TIME = 5.0
minutes. Then mo v e the cu rsor to %A and press
TIME = 0,
and enter 5 to store a step
enter 100% and defi ne a step with the same elu ent compos ition as the previous step. Although there is no change in eluent parameters, the 100% of %A must be entered at 5.00 minutes to establish 5.00 as the gradient start point.
This marks the end of the isocratic section of the run and the beginning of t he el uent B concentration ramp. From this point on, t he concentration of eluent A decreases from 100% as the concentration of eluent B begins to increase from 0%.
5. Move the cursor to the next line. Ente r 10 in the cursor to the
field and enter 65, followed by 35 in the %B field.
%A
field . Move the
TIME
After a total of 10 minutes (5 minutes of isocratic conditions plus 5 minutes to gradually decrease the amount of eluent A in the mixture while increasing the amount of eluent B), the eluent composition is 65% eluent A and 35% eluent B.
This step marks the end of the second segment and the beginning of the eluent C concentration ramp. Figure 3-9 illustrates the
METHOD
screen as it appears after Step 5.
Enter
Enter
.
to
3-16
M ETHO D E DIT
2
SAVE TO RUN 0
2
LIMITs 0 - 5000
TIME %A
100.0
0.00
5.00
10.00
100.0
100.0
65.0 35.0
%B %C %D V
C FLOW
_
L2.00IN IT
_
H e lp Mes s a ge
Figure 3-9. Linear Gradient Method Example (After Step 5)
PSI
> >
> >
Doc. 031377 -03 7/01
3 • Operation and Maintenance
6. Move the cursor to the next line. Ente r 15 in the cursor to the
field and enter 50, followed by 50 in the %C field.
%B
field. Move the
TIME
Beginning with the method step immediately preceding this one (
TIME=10
), the pump begins adding el uent C, starting with 0%. When the method reaches this step (after 15 minutes), eluent C is at 50%, eluent B at 50%, and eluent A at 0%.
7. Move the cursor to the next line. Ente r 25 in the cursor to
and enter 100.
%C
field. Move the
TIME
After a total of 25 minutes, the concentration of e luent B drops to 0% and the concentration of eluent C increases to 100%.
8. Move the cursor to the next line. Enter 25.01 in the the cursor to
and enter 100.
%D
field. Move
TIME
The concentration of eluent C drops to zero and the concentration of eluent D, which the pump began adding 0.6 seconds (0.01 min) earlier, reaches 100%. This is a
step change
in eluent composition to 100% eluent D. A step change is a v ery st eep lin ear gra dient in which the eluent composition changes from one eluent to another in 6 seconds. Figure 3-10 illustrates the
METHOD
screen as it appears
after Step 8.
M ETHO D E DIT
Doc. 031377-03 7/01
SAVE TO RUN 022
TIME %A
10.00
>
65.0 35.0
15.00
25.00
25.01
LIMITs 0 - 5000
%B %C %D V
C FLOW
50 .0 50 .0
100.0
100.0
PSI
2.00
Help Message
Figure 3-10. Linear Gradient Method Example (After Step 8)
9. Move the cursor to the next line. Ente r 30 in the cursor to the
field and ente r 100. The el uent co mpositi on remai ns
%D
field. Move the
TIME
unchanged at 100% eluent D for 4.99 minutes.
10. Move the cursor to the next line and enter 30.01 in the Move the cursor to the
field and enter 100. This causes another
%A
step gradient from 100% D to 100% A.
> >
> >
TIME
field .
3-17
GP50 Gradient Pump
11. Move the cursor to the
SAVE TO
field and press
method to memory. Figure 3-11 illustrates the completed method.
M ETHO D E DIT
SAVE TO RU N 022
LIMITs 0 - 5000
TIME %A
25.00
>
25.01
30.00
30.01
100.0
%B %C %D V
100.0
100.0
100.0
C FLOW
Help Message
Figure 3-11. Linear Gradient Method Example (Complete)
12. Press
METHOD
and
MENU
field. Press
Enter
Reset
to go to the
screen. Enter 2 in the
MAIN
to reset the method to the
necessary).
13. If the pump motor is off, press
14. If the method clock is in hold, press
to start the pump.
Off/On
Hold/Run
to start the method running. When the method reaches the last step in the method (
TIME = 30.01
), the pump will continue to pump isocratically until the
clock is reset.
Enter
2.00
to save the
PSI
> >
> >
step (if
INIT
3-18
Doc. 031377 -03 7/01
3 • Operation and Maintenance

3.4.3 Curved Gradient Method Example

When attempting to optimize the separation of a single component in a complex mixture, it is often helpful to employ paired segments of a curved gradient. For example, if the analyte is an oligonucleotide that elutes with 0.32 M NaCl in a linear gradient run, you can move potentially interfering components of the sample away from the target oligonucleotide b y fir st programming a se gment from 0% to 32% of a 1 M NaCl solution using curve 2. Then, program a segment from 32% to 100% of 1 M NaCl using curve 8. The resulting gradient profile is shown in Figure 3-12.
Doc. 031377-03 7/01
Figure 3-12. Gradient Curve Profile
In binary curve gradients, two elements follow curves which are mirror images (e.g., E1 = Curve 8 and E2 = Curve 1 in Figure 3-13). See Section 2.9.4 for details about gradient curves.
3-19
GP50 Gradient Pump
The following summarizes the curved gradient method example steps:
Create Method 3 to be gin under i socratic co nditions with 10 0% eluent
_
A (25 mM Tris buffer, pH 8.0) at 1.5 mL/min. After 1 minute, begin a convex addition of eluent B (25 mM Tris
_
buffer, pH 8.0, containing 1.0 M NaCl) while decreasing eluent A until, after 2.5 minutes (elapsed time = 3.5 minutes), the mixture is 68% eluent A and 32% eluent B.
Figure 3-13. E1 and E2 Curves
3-20
Begin a 2.5-minute con cav e se gment in which t he amount of eluen t A
_
is decreased and the amount of eluent B is increase d until the mixtu re is 100% eluent B (elapsed time = 6 minutes).
At 7 minutes, begin a 3-minute conca ve segment, r educing elue nt B to
_
0% and increasing eluent A to 100% (total elapsed time = 10 minutes).
1. Go to the
METHOD
screen and enter a method number in the
field (3, for example).
If Method 3 currently e xis ts and you w ant to ret ain it, e nter a ne w,
_
unused method number in the If Method 3 currently exists and you want to delete it, move the
_
cursor to
2. Move to the
TIME = INIT
field of the
FLOW
and press
INIT
field.
EDIT
Delete
twice.
step and enter a flow rate of 1.50.
Doc. 031377 -03 7/01
EDIT
3 • Operation and Maintenance
Figure 3-14 illustrates the
METHOD
screen as it appears so far. You can now be gi n ent ering th e method step s that will gene rate the curv ed gradient profile (see Figure 3-13).
M ETHO D E DIT
SAVE TO RUN 033
LIMITs 0 - 5000
TIME %A
100.0
%B %C %D V
C FLOW
_
L1.50IN IT
_
0.00
Help Message
Figure 3-14. Curved Gradient Method Example (After Step 2)
3. Move the cursor to the
field of the
%A
TIME = 0
step and press
100% of eluent A is automatically filled in.
4. Move the cursor down to the next line and enter 1 in the and then move to the
field and enter 100.
%A
This marks the end of the isocratic section of the run and the beginning of the eluent B concentration ramp. At this point, the concentration of eluent A begins to decrease from 100% as the concentration of eluent B begins increasing from 0%.
PSI
> >
> >
TIME
Enter
field,
.
Doc. 031377-03 7/01
5. Move the cursor down to the next line and enter 3.5 in the Move the curso r to
and enter 68. Move the cursor to %B and enter
%A
TIME
32.
6. Move the cursor to the
field and enter 3 to run gradient curve 3.
C
After a total of 3.5 minutes (1 minute of isocratic conditions plus 2.5 minutes to gradually decrease the amount of eluent A in the mixture while increasing the amount of eluent B), the eluent composition is 68% eluent A and 32% eluent B. Figure 3-15 illustrates the screen as it appears after Step 6.
field.
METHOD
3-21
GP50 Gradient Pump
M ETHO D E DIT
TIME %A
100.0
0.00
100.0
%B %C %D V
SAVE TO RUN 033
LIMITs 0 - 5000
C FLOW
_
L1.50IN IT
_
PSI
1.00 100.0
3.50 68.0 32.0 3
Help Message
Figure 3-15. Curved Gradient Method Example (After Step 6)
7. Move the cursor to the next line and enter 6 in the the cursor to
8. Move the cursor to the
and enter 100.
%B
field and enter 8.
C
TIME
After a total of 6 minutes, the eluent composition is 0% eluent A and 100% eluent B.
9. Move the cursor to the next line and enter 7 in the the cursor to
and enter 100. The eluent composition remains
%B
TIME
unchanged at 100% eluent B for 1 minute.
10. Move the cursor to the next line and enter 10 in the the cursor to
and enter 100. After a total of 10 minutes, the
%A
TIME
concentration of eluent B drops to zero and the concentration of eluent A increases to 100%. Figure 3-16 illustrates the completed
METHOD
screen.
> >
> >
field . Move
field . Move
field. Move
3-22
M ETHO D E DIT
SAVE TO RUN 033
LIMITs 0 - 5000
TIME %A
3.50 68.0 32.0 3
>
%B %C %D V
6.00 100.0
C FLOW
1.50
8
7.00 100.0
10.00 100.0
Help Message
Figure 3-16. Curved Gradient Method Example (Complete)
11. Move the cursor to the
SAVE TO
field and press
Enter
method to memory.
PSI
> >
> >
to save the
Doc. 031377 -03 7/01
3 • Operation and Maintenance

3.4.4 Editing a Running Method Example

After entering a method, you can modify it by changing, adding, or deleting steps.
The example describes how to make the following changes to Method 2, the linear gradient example (see Section 3.4.2):
Change the eluent composition at
_
TIME = 15
from 50% eluent B and
50% eluent C to 45% eluent B and 55% eluent C. Add a step to Method 2 at
_
TIME = 20.0
to make the eluent composition
40% eluent B and 60% eluent C.
Figure 3-17 illustrates the
METHOD
screen as it will appear when editing
is complete. Figure 3-18 illustrates the edited grad ient profile.
M ETHO D E DIT
TIME %A
10.00
15.00
20.00 40.0 60.0
25.00
Help Message
>
65.0 35.0
>
Figure 3-17. Edited Linear Gradient Method Example
%B %C %D V
45.0 55.0
SAVE TO RUN 222
LIMITs 0 - 5000
C FLOW
2.00
100.0
PSI
> >
> >
Doc. 031377-03 7/01
Figure 3-18. Gradient Profile After Editing
3-23
GP50 Gradient Pump
This example assumes that the example Method 2 is currently running.
1. Go to the
METHOD
screen and enter 2 in the
EDIT
field.
2. Move the cursor down through Method 2 until you reach the
TIME = 15
to
%C
3. Move the cursor to the time field. Move the cursor to
%C
4. Move the cursor to the
5. Press
step. Move the cursor to %B and enter 45. Move the cursor
and enter 55.
field and press
TIME
%B
and enter 4 0. Move the cursor to
. Enter 20 in the
Insert
and enter 60.
and select either the
Menu
SAVE TO
field and press
screen or
MAIN
.
Enter
DETAIL
screen. Check
the status of the method clock: If the elapsed time is less than 15 minutes (the time for the first
_
change made to the method), the changes will be incorporated into this run and executed at the programmed time.
If the elapsed time is gr ea te r th an 15 minutes, the changes will not be
_
incorpora ted into this run. To put the changes into effect, either press
to set the method clock to the
Reset
elapsed time in the
MIN
field that is less than 15 (10, for example).
ial conditions, or enter an
INIT
The method will restart, using the parameters programmed for
TIME = 10
, and the method changes will be incorporated at the
programmed time.
3-24
Doc. 031377 -03 7/01

3.5 Routine Maintenance

This section describes routine maintenance procedures to be performed by the user. Any other maintenance procedures must be performed by qualified Dionex personnel.

3.5.1 Daily Maintenance

When using a combination of eluents which contain both salt or base
_
and solvent, rinse the piston seals frequently or continuously. Eluent tends to crystallize as the solvent e vaporates; these crystals can abrade the pistons and cause the piston seals to leak. Rinse the piston seals before and after operation each day as described in the following steps, or install a continuous seal wash kit (P/N 059187), to rinse the piston seals continuously (see Section B.2.8).
1. Open the lower pump door and locate the two rinse ports on the front of each of t he pump heads. Figure 3-19 shows the rinse port connections.
2. Install a rinse waste tube (P/N 054418), provided in the GP50 Ship Kit, onto each head (see Figure 3-19).
3 • Operation and Maintenance
Rinse Waste
Doc. 031377-03 7/01
Tu bes
Figure 3-19. Rinsing the Piston Seals
Rinse Ports wi th F ema le Luer Adapters
Syringe (P/N 054578)
3-25
GP50 Gradient Pump
3. Place the end of each rinse waste tube into a waste receptacle. Attach a small syringe (P/N 054578) containing 5 to 10 mL of deionized water to the rinse inle t female luer adapt er on one of the pump heads.
4. Inject deionized water into the fitting to rinse the pump heads.
5. Repeat Steps 3–4 for the other pump head.
6. Dispose of the waste water and close the door to the mechanical chassis.
Check the entire mechanical chassis for leaks from the rinse ports, the
_
eluent manifold connectio ns and valves, the vac uum deg a s chambers , and the eluent reservoirs (see Figure 3-20). Tighten or replace any leaking fittings. Wipe up liquid spills and rinse dried reagents off the pump components with deionized water.
Dionex recommends th oro ughly flu shing t he vacuum degas assembl y
_
with deionized water after each use. Flushing chemicals out of the degas chambers and tubing avoids crystallization in the membrane pores.
3-26
Eluent Reservoirs
VAC Chamber IN
VAC Chamber OUT
ABCD
D
Proportioning Valve/
A
C
Manifold Assembly
B
Pressure Transducer
Pump Heads
OUT
Figure 3-20. Eluent Flow Schematic
Mixer To Column
Rinse Ports
Priming Block
Doc. 031377 -03 7/01

3.5.2 Periodic Maintenance

Replace both the primary and rinse seals in each pump head
_
approximately every 6 months. The seals may need to be replaced more often if you rout inely run high pr essur es or h igh fl ow rates, or if you operate the pump continuously.

3.6 Shutdown

Rinse the pump piston seals before and after daily operation to prevent build-
_
up of salt crystals or oth er cont aminants that can damage the piston seal s (s ee Section 3.5.1).
Before a shutdown of three days or more, flush the system with deionized
_
water to prevent contaminants from building up. Or, if this is not possible, maintain a continuous rinse through the system until you resume normal operation. Select a fl ow rate of 0.04 mL/min for stand ar d bor e pump he ads or
0.01 mL/min for microbore pump heads, and set all four valves in the eluent
manifold to 25% so that the valves are also flushed.
Flushing the eluent manifold is extremely important if your eluents have a combination of salt or base and solvent. If salt precipitates in the valves, the valve diaphragms may be seriously damaged. If this happens, you will have to replace the entire valve assembly.
3 • Operation and Maintenance
Before a shutdown of more than three days, reduce the pressure on the eluent
_
reservoir(s) to approximately 21 KPa (3 psi). Shut down the pump by turning off the main power on the GP50.
_
Doc. 031377-03 7/01
3-27
GP50 Gradient Pump
3-28
Doc. 031377 -03 7/01
4 • Troubleshooting
This chapter is a guide to troubleshooting minor problems that may occur while operating the GP50 Gradient Pump. To use this gu ide , t urn to the section that b est describes the operating problem. There, possible causes of the problem are listed in order of probability.
If you are unable to resolve a problem, contact Dionex Technical Support. In the U.S., call 1-800-346-6390. Outside the U.S., call the nearest Dionex office.

4.1 Left-Right Pump Head Pressure Fluctuations

The GP50 display updates the pressure readout once per piston stroke. A variation of more than 3% from one stroke to the next indicates a problem.
_
Pump out of prime; there is no eluent
1. Refill the eluent reservoirs. Also make sure that each eluent line extends to the botto m of the reservoir.
2. Reprime the pump (see Section B.2.8).
_
Pump out of prime; eluents are improperly degassed
1. If the GP50 is not equipped with the optional vacuum degas pump
2. If the GP50 is equipped with the optional vacuum degas pump assembly,
_
Pump out of prime; end-line filter is dirty or clogged
1. Replace the filter (P/N 045987).
2. Reprime the pump (see Section B.2.8).
Doc. 031377-03 7/01
assembly, degas the eluents manually (see Section 3.1 .1). Reprime the pump (see Section B.2.8).
test the degas p ump: a. Open the
Select
b. The degas pump should turn on and run for about 2 minutes. If it does
not run (see Section 4 .6).
DEGAS STATUS
button to toggle the
screen (press
field to
TEST
Menu, 8, 8
RUN
, and 3). Press a
and press
Enter
.
4-1
GP50 Gradient Pump
_
Pump out of prime; blockages in inlet tubing
Kinked or clogge d tubin g cause s the pu mp to be “st arv ed ” for e luent . Replac e the tubing and fittings. Reprime the pump (see Section B.2.8).
_
If priming the pump does not eliminate excessive pressure fluctuations, the piston seals or check valves may be dirty or defective
1. Follow these steps to isolate the cause: a. If leaks ar e seen from the piston rinse tubes, replace the piston seals
(see Section 5.2).
b. If no leaks are seen, replace the check valves (see Section 5.1).
Impurities in the eluents can cause dirty or defective check valves. Install end line filters (P/N 045987) to help prevent this (see Section 3.1.2).
c. Using a 7-mm open-end wrench or your fingers, loosen the lock on
the mechanical chassis drawer. The lock is on the lower right side of the chassis, between valves 3 and 4 (see the label on the inside of the lower door). Pull out the mechanical chassis drawer a few inches.
d. If a piston does not move when there is pump flow, examine it for
breakage and replace if necessary. If a piston moves, examine the pump head for scratches and replace if necessary (see Section 5.3). If a piston moves slightly and then breaks contact with the rocker arm follower, replace the piston seal (see Section 5.2).
e. Push the mechanical chassis drawer back in place, making sure the
cables are not pinched. Tighten the drawer lock.
Observe the warning label on the inside of the lower door. The arrows on the label indicate moving mechanical parts that present pinch hazards when the pump is on and the mechanical drawer is open. Do not touch any parts within the mechanical chassis while the pump is on.
Respectez l'étiquette d'avertissement apposée à l'intérieur de la porte inférieure. Les flèches sur l'étiquette indiquent des pièces mécaniques mobiles qui posent un danger de pincement lorsque le GP50 est sous tension et le tiroir mécanique est ouvert. N'ut ilisez jamais le GP50 avec le tiroir du châssis mécanique ouvert.
4-2
Doc. 031377 -03 7/01

4.2 Pump Will Not Start

_
Flow rate is set to zero
Reset the flow rate (see Section 2.4.2).
_
Pump starts briefly during priming, but then stops because of high pressur e li mi t
Open the pressure transduce r waste valve (see Figure B-5) by turning the knob counterclockwise two turns.

4.3 Pump Stops

_
Method or other remote input instructed the pump to stop
Check the display screen for error messages. If none are displayed, the pump was probably instructed to stop by the method, computer, or other remote signal source.
_
Electrical cables improperly installed
4 • Troubleshooting
1. Set the pump to the pump.
2. If a non-zero flow rate is displayed and the keypad LED is on, verify that the electrical cables in the mechanical chassis are properly installed.
a. Turn off the GP50 power. b. Using a 7-mm open-end wrench or your fingers, loosen the lock on
the mechanical chassis drawer. The lock is on the lower right side of the chassis, between valves 3 and 4 (see the label on the inside of the
lower door). c. Pull out t he mechanical cha ssis drawer a few inches. d. Locate the distribution card on the top of the mechanical chassis.
Check that all cables are seated correctly in the connectors on the
card. e. Push the mechanical chassis drawer back in place, making sure the
cables are not pinched. Retighten the drawer lock. Turn on the power.
LOCAL
mode,
DIRECT CONTROL
. Press
Off/On
to start
Doc. 031377-03 7/01
4-3
GP50 Gradient Pump
_
Low pressure limit was tripped. The following message is displayed:
1. Verify that eluent is present in the channel selected. If the eluent reservoir is empty, refill it or select a ch annel which does have eluent. Prime the pump (see Section B.2.8) before resuming operation.
2. Make sure the waste valve on the pressure transducer (see Figure B-5) is closed by turning the knob on the pressure transducer housing clockwise.
Overtightening the pressure transducer waste valve may damage the valve and the pressure transducer housing.
3. Make sure there are no liquid leaks in the flow system.
Low Pressure Limit Violation
4. Set the pump to
LOCAL
mode,
DIRECT CONTROL
. Press
Off/On
to start the
pump. Verify that the pistons are moving and that you can hear the pump. If there is no sound from the pump, check the LED on the CPU card
inside the door to t h e ele ct ronics chassis. A red LED ind icates a defective power supply. To have the power supply (P/N 046440) replaced, contact Dionex Technical Support.
With the pump running, open the and
). and note whether the left-right pressure varies by more than 3%
3
DSP STATUS
screen (press
Menu, 8, 8
between strokes. If it does, refer to Section 4.1. If it does not, either increase the flo w rate or reduce the lo w pre ssure limit s etting and cont inue operation.
,
4-4
Doc. 031377 -03 7/01
4 • Troubleshooting
_
High pressure limit was tripped. The following message is displayed:
High Pressure Limit Violation
1. Isolate segments of the flow path to determine the source of the high backpressure.
a. Remove the pump inlet tubing from the injection valve. b. Turn on the pump and record the backpressure. c. One at a time, add each segment of the remainder of the flow path to
determine the source of the backpressure. If reconnecting a component causes a sharp increase in backpressure, preplace the component. Replace tubing, fittings, or components as necessary to resume operation at the standard operating backpressure.
d. If the source of the high backpressure is the column, refer to the
column manual for cleanup procedures. The column may need replacement.
2. Verify that the pressure transducer is calibrated correctly. a. Open the pressur e tr ansducer waste v al ve (see Figure B-5) by turning
the knob counterclockwise about two turns.
b. Check the pressure reading; if it is above 97 KPa (14 psi), recalibrate
the pressu re transducer (see Section C.3.6).
_
A DSP-related error message is displayed
Several error messages are related to Digital Signal Processor (DSP) errors:
DSP communication fails, DSP does not acknowledge
, etc. If one of these
messages is displayed, follow the procedure below:
1. Turn off the GP50 power.
2. Verify that the DSP card is present and correctly installed in slot 1 of the electronics chassis (see Figure 2-6).
3. Turn on the GP50 power.
Doc. 031377-03 7/01
4-5
GP50 Gradient Pump
4. If the DSP error message reappears, notify Dionex Technical Support. The power supply (P/N 046440), DSP card (P/N 045369), or CPU card (P/N 046340) may need to be replaced.
Do not remove any of the electronic cards from the pump. There are no user-serviceable components on the cards. If servicing is required, it must be performed by qualified personnel and appropriate electrostatic discharge (ESD) handling procedures must be followed.
Ne retirez aucune des cartes électroniques de la pompe. Auc un des composants sur les cartes ne peut être réparé par l'utilisateur. Toute réparation doit être effectuée par un personnel qualifié utilisant des procédures correctes de décharge électrostatique.
_
The following error message displays:
Motor Drive Fails
4-6
If the pump motor is in a runaway condition, the motor automatically shuts off and the above error message is displayed. Contact Dionex Technical Support.
_
The following error message displays:
Encoder index not found
1. Turn off the GP50 power.
2. V eri fy tha t th e cable s connect ed to the DSP car d in the elect ronic s chassi s (see Figure 2-6) are seated properly.
3. Turn on the GP50 power. If the error message reappears, notify Dionex Technical Support.
Doc. 031377 -03 7/01

4.4 Liquid Leaks/Leak Alarm

_
Leaks from the fr ont rinse po rts or r e ar of the pump hea d may indi cate a defective piston seal
1. Replace the piston seal and the rinse seal (see Section 5.2).
2. Check all connections between the eluent reservoirs and the pump heads. Tighten the fitting connections just enough to stop the leak.
_
Proportioni ng valve leaks
Tighten loose fittings. If there are no loose fittings, replace the valve (see Section 5.5).
Overtightening the fitting connections may strip the threads in the valve block. If this happens, replace the entire manifold assembly (P/N 046203).
_
Pressure transducer leaks
Inspect the pressure transducer. If the waste valve is the source of the leak, replace the waste valve O-ring (see Section 5.4). If the leak is from the rear of the transducer, contact Dionex Technical Support.
4 • Troubleshooting
_
Priming valve
Tighten any leaking fittings just enough to stop the leak. If this does not stop the leak, replace th e fittings and/or tubi ng maki ng the connection. If this does not stop the leak, replace the priming block assembly (P/N 054086). Contact Dionex Technical Support.
Doc. 031377-03 7/01
4-7
GP50 Gradient Pump
_
Interior mechanical chassis leaks
Inspect the chassis for leaks. Tighten any leaking fittings. Replace any damaged parts.

4.5 Noisy Pump Motor

_
DSP (digital signal processing) card curr ent limit ha s been exc eeded. The card includes a built-in current limiter to protect the motor and motor drive.
Check the three small LEDs in the upper left corner of the DSP card bulkhead, which is in the electronics chassis behind the pump upper door. If the LEDs are flashing in time with the pump strokes, the current limiter is being activated. As the pump motor ages, it becomes less efficient and the current limit is acti v at ed more frequent ly. Activating the current limit is har mle ss, b ut if it occurs frequently, even at low speeds and/or pressures, the bottom plate assembly (P/N 045670) needs to be replaced. Call Dionex Technical Support for assistance.
_
Pressure servo oscillation
Check the correct pump head volume and head material are selected. If the settings are correct but the problem persists, notify Dionex.
_
Out of prim e
Reprime the pump (see Section B.2.8).
DSP STATUS
screen (pre ss
Menu, 8, 8
, and 3) to verify that the
4-8
Doc. 031377 -03 7/01

4.6 Vacuum Degas Pump Does Not Run

4 • Troubleshooting
_ DEGAS OPTIONS
Go to the field is set to
DEGAS OPTIONS
screen settings incorrect
screen (pre ss
ALWAYS OFF
, select
Menu
BY SETTING
and 4). If the
DEGAS PUMP
and then enter the cycle
duration and frequency times (see SectionC.1.6).
_
Electrical cables improperly installed
Follow the steps below to manually test the degas pump.
1. Go to the
DEGAS STATUS
button to toggle the
screen (press field to
TEST
RUN
Menu, 8, 8
and press
, and 3). Press a
.
Enter
2. The pump should turn on and run for the cycle duration time specified in the
DEGAS OPTIONS
screen (2 minutes by default). If it does not run, verify tha t the ca bles conne cted to the pump in the e lectron ics chas sis (see Figure 2-6) and in the mechanical chassis are properly connected.
To check the mechanical chassis connections: a. Turn off the GP50 power. b. Using a 7-mm open-end wrench or your fingers, loosen the lock on
the mechanical chassis drawer. The lock is on the lower right side of the chassis, between valves 3 and 4 (see the label on the inside of the lower door). Pull the drawer out a few inches.
c. Locate the distribution card on the top of the mechanical chassis.
Check that all cables are seated correctly in the connectors on the card.
Select
Doc. 031377-03 7/01
d. If the connections are correct, the distribution card may need to be
replaced. Call Dionex Technical Support.
e. Push the mechanical chassis drawer back in place, making sure the
cables are not pinched. Retighten the drawer lock.
4-9
GP50 Gradient Pump

4.7 Vacuum Degas Pump Calibration Fails

At the end of the degas calibration, the
DEGAS READING
value is less than
13000 counts and one of the following error messages appears:
Degas vacuum pump is not present or degas
circuitry is malfunctioning.
Vacuum Degas Fails
Verify that the cable to the vacuum degas pump is connected to the distribution card in the mechanical chassis.
1. Turn off the GP50 power.
2. Using a 7-mm open-end wrench or your fingers, loosen the lock on the mechanical chassis drawer. The lock is on the lower right side of the chassis, between valves 3 and 4. Pull out the drawer a few inches.
3. The distribution card is on the top of the mechanical chassis. Labels printed on the card identify the various cables plugged into it. Locate the connector for the vacuum degas pump (labeled
VAC PUMP
) near the right rear corner of the card. Make sure th e cable is fully seated in the connector.
4. Push the mechanical chassis drawer bac k in pla ce, mak ing sur e the ca bles are not pinched. Retighten the drawer lock.
4-10
5. Turn on the power.
6. Retry the calibration. If the message reappears, contact Dionex Technical Support.
Doc. 031377 -03 7/01

4.8 Vacuum Degas Pump Low Vacuum

The GP50 monitors the degas vacuum reading every 1 min. If the degas vacuum is lower than the monitoring value, the degas pump turns on. When the pump turns off, if the vacuum reading is 2000 counts or more lower than the monitor ing value, the following message displays:
LOW VACUUM ALARM!!
Check DEGAS OPTIONS settings or refer to
service manual
4 • Troubleshooting
Open the
DURATION
DEGAS OPTIONS
time and/or decreasing the
screen (press
TIME BETWEEN CYCLES
resolve the problem, contact Dionex Technical Support.

4.9 Relays or TTLs Inoperative

_
Incorrectly installed cables
Make sure the cable(s) between the GP50 relay or TTL connector(s) and the connector(s) on the other instrument(s) are connected properly (see Section D.3).
_
Method programming error
For Local control, make sure the parameter to turn on or off the relay/TTL is set correctly on the Remote control, make sure the command for turning the relay/TTL on or off is entered correctly into the PeakNet 6 program file or the PeakNet 5 Method.
For TTL inputs, make sure the controlling device is programmed correctly.
_
When attempting to set TTL2, the following message displays:
TTL2 is set to indicate FLOW/NO FLOW.
METHOD EXTENSION
and 4). Try increasing the
Menu
. If this does not
screen (see Section C.1.5). For
CYCLE
The to signal when pump flow stops ( power to a Self-Regenerating Suppressor function, open the
OUTPUT USAGE
Doc. 031377-03 7/01
TTL2 OUTPUT USAGE
PUMP OPTIONS
field to
field on the
0 FLOW
NORMAL
.
PUMP OPTIONS
screen is currently set
). This setting is used to control the
®
(SRS®). To use TTL2 for another
screen (press
and 6) and set the
Menu
TTL2
4-11
GP50 Gradient Pump

4.10 Poor Chromatographic Reproducibility

_
Liquid lines incompletely flushed after an eluent change
Attach a syringe to the prim ing block v al v e (see Fig ure B-5) and draw at least
2.5 mL (20 mL if the vacuum degas pump is installed) of the new eluent
through the liquid lines before beginning operation.
_
Leaking piston seal
Check for liquid leaks at the rinse ports in the front of the pump heads. Replace the piston seal on any head with a leak (see Section 5.2).
_
Malfunctioning proportioning valve
Test the valves (see Section C.2.7). If a test fails, the prop ortioning valve assembly may need to be replaced (see Section 5.5).
4-12
Doc. 031377 -03 7/01
5•Service
This chapter describes service and repair procedures that the user may perform. Any procedures not included here, including electronics-related service procedures, must be performed by Dionex personnel. For assistance, contact Dionex Technical Support. In the U.S., call 1-800-346-63 90. Outside the U.S., call the nearest Dionex office.
Before replacing an y par t, ref er to t he tr oublesh ooting infor mation i n Chapte r 4 to isolate the cause of the problem.
Substituting non-Dionex parts may impair performance, thereby voiding the product warranty. Refer to the warranty statement in the Dionex Terms and Conditions for more information.
The CPU card contains a lithium battery. If the CPU card is displaced, dispose of the used battery according to the manufacturer's instructions.

5.1 Cleaning and Replacing the Check Valves

A dirty check valve causes erratic flow rates and pressures and may cause the pump to lose prime and/or be difficult to reprime.
1. Turn off the main power switch, to ensure you do not unintentionall y st ar t t he
pump.
2. Disconnect the tube fittings from the inlet and outlet check valve housings
(see Figures 5-1 and 5-2).
3. Use a 1/2-inch wrench to loosen both check valve housings. Remove the
check valve housings and cartridges from the pump head. Carefully remove the check valve cartridges from the housings.
4. Place the check valve housings and cartridges in a beaker with methanol, and
sonicate or agitate for several minutes.
Doc. 031377-03 7/01
5-1
GP50 Gradient Pump
Outlet Check
Pressure Transducer
Valv e
To Column
Pump Head
Inle t
Check
Valv e
Figure 5-1. Pump Heads and Liquid Lines (PEEK)
Pressure Transducer
To C o lu mn
To Waste
5-2
Outlet Check
Valve
Inle t
Check
Valve
Figure 5-2. Pump Heads and Liquid Lines (SST)
Pump Head
To Waste
To Eluent Manifold
Doc. 031377 -03 7/01
5•Service
5. Rinse each check valve housing and cartridge thoroughly with filtered
deionized water.
6. The
check val v e assembly housing has a 1/4-28 port (PEEK pumps) or a
inlet
10-32 port (stainless steel pumps). Replace the cartridge in the inlet check valve housing, making sure the double-hole end of the cartridge is visible.
The
check valve assembly housing has a 10-32 port (PEEK and
outlet
stainless steel pumps ). Replace the ca rtridge in th e outlet check va lve housi ng, making sure the single-hole end of the cartridge is visible. Liquid flows through the check valve in the large single hole and out the small double holes.
NOTE The pump will not operate properly unless the cartridge
is installed in the housing in the correct orientation.
7. Reinstall the check valves. Be sure to install the inlet check valve on the
bottom of the head and the outlet check valve on the top of the head. Tighten only enough to seat (25 in-lb. torque). Tighten a little more only if it leaks.
Overtightening may damage the pump head and check valve housing and crush the check valve seats.
8. Reconnect the liquid lines.
9. Turn on the GP50 power.
10. Prime the pump (see Section B.2.8). If the pump will not prime and you have eliminated all other possible causes of the problem, replace the check valve cartridge.
After replacing the check valve cartridge, go to the (press
Enter
Doc. 031377-03 7/01
Check Valve Cartridge Type Part Number
Standard Bore PEEK 047747 Standard Bore Stainless Steel 047755 Microbore PEEK 047748 Microbore Stainless Steel 048279
Menu, 8
, and 2). Move the cursor to the
VALVES IN USE
to reset the field to 0 cycles.
ELAPSED TIME
field and press
screen
5-3
GP50 Gradient Pump

5.2 Piston Seal Replacement

A damaged seal allows leakage past the piston and then through the rinse ports in the front of the pump heads. The pump may be difficult to prime, flow rates will be unstable, and baseline noise may be observed.
1. Turn off the main power switch, to ensure you do not unintentionall y st ar t t he GP50.
2. Disconnect the tube fittings from the pressure transducer and the inlet check valve (see Figures 5-1 and 5-2).
3. Remove the two nuts from the pump head.
4. Carefully disengage the head from the piston by pulling the head straight off and away from its mounting guides.
Lateral motion while disengaging the head fr om the piston ma y break the piston.
Un mouvement latéral pendant la séparation de la tête et du piston peut casser le piston.
5. Place the head (front end down) on a clean work surface and lift off the backup washer to expose the piston guide (see Figures 5-3 and 5-4).
6. The pistons are captured by a magnetic retention system and do not come off as part of the pump head assembly. After removing the pump head, hold the shaft of the piston and apply just enough lateral force to overcome the magnetic field and remove the piston.
5-4
Doc. 031377 -03 7/01
Piston Seal
Outlet Check
Valve Assembly
Head
5•Service
Rinse Seal
O-Ring
Support Washer
Backup Was her
Piston
Piston Guide
Inlet Check Valve Assembly
Figure 5-3. Pump Head Assembly
Component Standard
Bore/PEEK
Standard Bore/SST
Microbore/ PEEK
Microbore/ SST
Pump Head Assembly 054087 054096 054094 054095
Pump Head 054051 054050 054098 054099 Outlet Check Valve 047661 047665 047657 047663 Inlet Check Valve 047660 047665 047656 047663 Piston Seal 054400 054402 054401 054401 Piston Guide 045633 045633 045632 045632 Rinse Seal 048722 048722 048721 048721 Support Washer 050745 050745 050744 050744 O-Ring 014895 014895 014895 014895 Backup Washer 045630 045630 045631 045631 Piston 052840 052840 053584 053584
Doc. 031377-03 7/01
Table 5-1. Pump Head Assembly Part Numbers
5-5
GP50 Gradient Pump
7. To remove the piston guide and seal: a. Fill the head cavity with deionized water by injecting through either the
piston opening or the inlet check valve. b. Reinsert the piston approximately 1/8-in into the seal (see Figure 5-4). c. Install a 10-32 fitting plug (P/N 042772) on the outlet check valve.
Tighten the plug. d. Push the piston into the head. This should hydraulically unseat the seal
and piston guide from the head. Remove the piston and pull off the guide
and seal.
NOTE If the piston guide an d seal do no t come out, m ake sure
the 10-32 plugs are tight. Then, add more water and repeat Steps b and d.
e. Remove the 10-32 fitting plug.
10-32 Fitting Plug
Outlet
Check
Valv e
Pump
Head
(P/N 042772)
Piston
Push in to unseat the seal and piston guide.
Piston
Seal
Remove the pisto n from the h e a d and pu ll off the seal and piston guide.
Guide
5-6
Inlet Ch e c k
Valv e
Fitting P lug (Optional)
Backup Washer (Remove)
Figure 5-4. Removing the Piston Seal
Doc. 031377 -03 7/01
5•Service
8. To install the new seal and reinstall the piston guide: a. Push the pisto n through the pist on guide an d the new seal. Then, inser t the
piston, piston guide, and seal into the pump head until the seal contacts the bottom of the counterbore (see Figure 5-5, View A).
b. Hold the piston guide and seal in place and remove the piston from the
head (see Figure 5-5, View B).
c. Seat the seal by pu shing the pist on guide i nto the he ad unt il it is flush wit h
the head.
View A
Seal
Piston Guide
Piston
Pump Head
Push the piston through the seal an d g u ide and pa r tially ins e r t in to the head, just until the seal contacts the coun terbore.
Counterbore
View B
Remove the piston and push the pis to n guid e in to the h ead t o fin ish seating the seal.
Figure 5-5. Installing the Piston Seal
Doc. 031377-03 7/01
5-7
GP50 Gradient Pump
9. Dionex recommends reinstalling the head and piston as a single assembly so that the piston centers itself. To do this:
a. Press the backup washer into the head until it is flush with the indented
surface of the head.
b. Insert the piston
halfway
into the head. This ensures that the magnet in th e follower picks up the piston. (The follower is the cylinder that holds the piston in place as it moves in and out of the pump head assembly.)
c. Reinstall the he ad and p iston assembl y, using a wrench to tighten the nuts
evenly (12 in-lb torque).
10. Reconnect the liquid line to the inlet check valve.
11. Reconnect the tube fittings to the pressure transducer.
12. Turn on the GP50 power.
13. Go to the the
SEALS IN USE
ELAPSED TIME
field to 0 cycles. The pump is ready for normal operation.
screen (press

5.3 Pump Piston Replacement

Continued leaking through the rinse ports after replacing the piston seal (assuming the head is tight) indicates a dirty, scratched, or broken piston.
1. Turn off the main power switch, to ensure you do not unintentionall y st ar t t he GP50.
2. Disconnect the tube fittings from the pressure transducer and the inlet check valve (see Figures 5-1 and 5-2).
Menu, 8
, and 2). Press
Enter
to reset
5-8
3. Remove the two acorn nuts from the pump head.
Lateral motion while disengaging the head fr om the piston ma y break the piston.
Un mouvement latéral pendant la séparation de la tête et du piston peut casser le piston.
Doc. 031377 -03 7/01
5•Service
4. Slowly pull the head and allow it to separate from the housing. Carefully disengage the head from the piston by pulling the head straight off and away from the mounting guides.
5. The pistons are captured by a magnetic retention system and do not come off as part of the pump head assembly. After removing the pump head, hold the shaft of the piston and apply just enough lateral force to overcome the magnetic field and remove the piston.
6. Replace the piston and the piston seal. Broken or scratched pistons can damage the piston seal.
7. Dionex recommends reinstalling the head and piston as a single assembly so that the piston centers itself. To do this:
a. Press the backup washer into the head until it is flush with the indented
surface of the head.
b. Insert the piston
halfway
into the head. This ensures that the magnet in th e follower picks up the piston. (The follower is the cylinder that holds the piston in place as it moves in and out of the pump head assembly.)
c. Reinstall the he ad and p iston assembl y, using a wrench to tighten the nuts
evenly (12 in-lb torque).
8. Reconnect the liquid line to the inlet chec k valve.
9. Reconnect the tube fittings to the pressure transducer.
10. Turn on the main power switch and prime the pump (see Section B.2.8).
Doc. 031377-03 7/01
5-9
GP50 Gradient Pump

5.4 Pressure Transducer Waste Valve O-Ring Replacement

A damaged O-ring causes leakage aro und the bas e of the pre ssur e transdu cer waste valve knob.
1. Turn off the main power switch, to ensure you do not unintentionall y st ar t t he
GP50.
2. Remove the valve from the pressure transducer housing by turning the knob
counterclockwise until it comes loose from the housing.
3. Remove the O-ring (P/N 046434) (see Figure 5-6).
4. Carefully slide a new O-ring (P/ N 046434) over the end of the valv e and push
it into the groove.
5. Reinstall the valve in the housing, turning the knob clockwise until the valve
is seated.
Do not overtighten the waste valve.
5-10
O-Ring
(P/N 046434)
Needle Valve (P/N 042625)
Screw (4)
(P/N 053988)
Transducer Cable Assembly
Gasket Support
(P/N 054065)
Transducer Housing (P/N 054052)
Washer (4) (P/N 045689)
(P/N 054064)
Gasket (P/N 054417)
Figure 5-6. Pressure Transducer Assembly
Transducer Sleeve (P/N 054053)
Doc. 031377 -03 7/01

5.5 Proportioning Valve Replacement

A defective propor tioning valve can cause the following problems: leak s, nonrepr oducible eluent co mpositions (whic h may cause r etention time shi fts), and flow restrictions (which may cause high backpressure).
1. Turn off the main power switch, to ensure you do not unintentionall y st ar t t he GP50.
2. Turn off the pressure on the eluent reservoirs and allow them to vent.
3. Follow these steps to disconnect the proportioning valve electrical connector from the distribution card.
a. Using a 7-mm open-end wrench or your fingers, loosen the lock on the
mechanical chassis drawer. The lock is on the lower right side of the chassis, between valves 3 and 4 (see the label on the inside of the lower door). Pull out the drawer a few inches.
b. The distribution card is on the top of the mechanical chassis. Labels
printed on the card identify the various cables plugged into it. Locate the proportioning valve electrical connector, labeled the card. Dis connect it from the card.
VALVES
5•Service
, at the front of
4. Disconnect the liquid line from the manifold outlet (see Figure 5-7).
Figure 5-7. Location of Proportioning Valve/Manifold Assembly
Doc. 031377-03 7/01
Manifold
Outlet
1
Proportioning Valve/
4
2
3
M a n ifold A s sem bly (P/N 046203)
5-11
GP50 Gradient Pump
5. Loosen the screws securing the proporti oning valve/ manif ol d assembly to the bulkhead (see Figure 5-7). Pull the valve/manifold assembly forward and disconnect the eluent lines on the manifold inlet. Remove the valve/manifold assembly completely from the bulkhead.
6. Thread the bundle of electrical lines from the new valve/manifold assembly through the bulkhead and up to the distribution card. Connect the valve electrical connector to the
7. Connect the eluent lines to the new valve/manifold assembly. Make sure eluent lines A through D are in the appropriate valve ports. Connect eluent line A to the valve marked #1, eluent B to the valve marked #2, and so on.
Tighten liquid connections to the valve no more than fingertight plus one-quarter turn. Overtightening or crossthreading the valve fittings may strip the threads in the manifold block. If this happens, replace the valve/manifold assembly.
8. Align the new assembly as shown in Figure 5-7 and mount it to the bulkhead. Tighten the screws.
VALVES
connector on the distribution card.
9. Push the mechanical chassis drawer back in plac e, maki ng sur e th e cabl es are not pinched. Retighten the drawer lock.
10. Attach the liquid line from the priming block to the manifold outlet.
11. To verify that each valve is functioning properly, connect a syringe to the priming valve. One at a time, select 100% of each eluent valve and draw liquid through the valve. After successfully drawing liquid through each valve, turn off the pump.
12. Next, veri fy that no liquid passes through the valve when th e pump is off. Do not turn on the pump. One at a time, select 100% of each eluent. No liquid should flow through the valve.
13. Turn on the main power switch. The pump is ready for normal operation.
5-12
Doc. 031377 -03 7/01

5.6 Changing Main Power Fuses

1. Turn off the main power.
HIGH VOLTAGE—Disconnect the main power cord from its source and also from the rear panel of the GP50.
HAUTE TENSION—Débranchez le cordon d'alimentation principal de sa source et du panneau arrière du GP50.
2. The fuse holder is part of the main power receptacle on the rear panel (see Figure 5-8). Note the recessed lock located on each side of the fuse holder. Using a small screwdriver, push each lock toward the center to release it. When both locks are released and the fuse holder pops out slightly, pull the fuse holder straight out of its compartment.
3. Replace the two fuses in the holder with new IEC127 fast-blow fuses rated
3.15 amps (P/N 054745). Dionex recommends always replacing both fuses.
5•Service
4. Reinsert the fuse holder into its compartment. The fuse holder is keyed to fit only in its proper orientation. Apply enough pressure evenly against the holder to engage the two locks. When both locks are engaged, the holder is flush against the panel.
5. Reconnect the main power cord and turn on the power.
Figure 5-8. Main Power Fuse Holder
Doc. 031377-03 7/01
5-13
GP50 Gradient Pump
5-14
Doc. 031377 -03 7/01

A.1 Electrical

A • Specifications
Main Power
Requirements
Fuse
Requirements

A.2 Environmental

Operating
Temperature
Operating
Humidity

A.3 Physical

Dimensions
Weight
Decibel Level
100 Vac to 240 Vac, 50/60 Hz; 2.5 amps. The GP50 power supply is main voltage auto-sensing and requires no manual adjustment.
Two 3.15 amp fast-blow IEC127 fuses (P/N 954745)
10 °C to 50 °C (50 °F to 104 °F)
5 to 95% relative humidity (non-condensing)
33.5 cm high x 22.5 cm wide x 42 cm deep (13.1 in x 8.9 in x 16.8 in) 6 cm (2.5 in) clearance required behind the module
19 kg (42 lb) 60 db (“A WEIGHING” setting)

A.4 Display and Keypad

Display
Keypad
Doc. 031377-03 7/01
Liquid crystal with adjustable backlighting. 26-button pad for entering commands and numeric values for
screen parameters.
A-1
GP50 Gradient Pump

A.5 Hydraulics

Eluent
Selection
Pump
Flow Rate
Operating
Pressure Pressure
Resolution
High Pressure
Limit
Low Pressure
Limit
Delay Volume
(No Mixer)
Four differ ent eluent componen ts; each can be pr oportioned from 0 to 100%
Dual-piston, variable speed, 100 µL (standar d bore), 25 µL (microbore) stroke, user-selectable constant pressure or constant flow feedback control
Standard bore pump head: 0.04 to 10.0 mL/min linearly variable in increments of 0.01 mL/min
Microbore pump head: 0.01 to 2.50 mL/min linearly variable in increments of 0.01 mL/min
35 MPa (5000 psi) maximum
0.07 MPa (10 psi)
0 to 35 MPa (0 to 5000 psi) in increments of 0.05 MPa (7.2 5 psi); trips instantaneously
0. to 35 MPa (0 to 5000 psi) in increments of 0.05 MPa (7.25 psi); trips after a time -ou t of 0.4 mL fo r the stan dard bore GP50 or 0.1 mL for the microbore GP50
Standard bore pump head: <700 µL Microbore pump head: <300 µL
A-2
Doc. 031377 -03 7/01

A.6 Control

A • Specifications
Remote
Local (Front Panel)
Limited remote operation via TTL-input logic level, and TTL­output and relay contact closures, or full remote programming and control via the DX-LAN interface.
Methods:
The actual number of stored methods depends on available memory. Each method may contain up to 50 separate steps.
Storage:
programs during power-down or in the event of a power failure.
Stores up to 100 separate methods (00 through 99),
Nonvolatile memory protects against the loss of

A.7 Vacuum Degas Pump Assembly (Optional)

Channels
Materials
4-channel membrane vacuum degas Wetted materials, PEEK, PTFE
Doc. 031377-03 7/01
A-3
GP50 Gradient Pump
A-4
Doc. 031377 -03 7/01

B.1 Facility Requirements

Make sure the GP50 installation site meets the electrical and environmental
_
requirements in Appendix A. Install the GP50 on a sturdy workbench at a height that ensures convenient
_
viewing of the front panel display and access to the interior.
Lift the GP50 only from the bottom or side surfaces of the module. Lifting by the front doors will dama ge the door hinges. Use caution when lifting the module, which weighs 19 kg (42 lb).
Ne soulevez le GP50 que par le fond ou les côtés. Son soulèvement par la porte du panneau avant endommagera les charnières de la porte. Soyez prudent lorsque vous soulevez le GP50: il pèse 19 kg (42 lb).
Allow at least 6 cm (2.5 in) free space behind the GP50 for connections and
_
ventilation.
B • Installation
The GP50 Gradient Pump is capable of operation with or without head
_
pressure on the eluent . If pre ssurization is used, provide a source of helium to pressurize the eluent and regenerant reservoirs (if used).
House eluents at least 3 cm (8 in) above the pump i n an EO1 Eluent Org anizer
_
(P/N 044125) or in built-in eluent containment (such as the LC25 Chromatography Oven reservoir container).
Always filter eluents to remove small particulates that may contaminate the
_
pump. Install an end-line filter (P/N 045987) on the end of each eluent reservoir line. Filters are supplied in the pressurizable reservoir Ship Kits. Refer to the
Dionex strongly recommends degassing eluents. If the vacuum degas
_
assembly is not installed, re fer to Section 3.1.1 for manual degassing instructions.
Doc. 031377-03 7/01
Pressurizable Reservoir Installation Instructions
for details.
B-1
GP50 Gradient Pump

B.2 Installation Instructions

The GP50 Ship Kit contains items necessary for completing the installation.
GP50 Gradient Pump Version Ship Kit
Standard bore/microbore with PEEK components P/N 054621 Standard bore/microbore with stainless steel components P/N 054627

B.2.1 Power Connection

SHOCK HAZARD—To avoid electrical shock, a grounded receptacle must be used. Do not operate or connect to AC power mains without an earthed ground connection.
DANGER D'ÉLECTROCUTION - Pour éviter toute électrocution, il faut utiliser une prise de courant avec prise de terre. Ne l'utilisez pas et ne le branchez pas au secteur C.A. sans utiliser de branchement mis à la terre.
The power supply cord is used as the main disconnect device. Make sure the socket-outlet is located near the GP50 and is easily accessible.
Le cordon d'alimentation principal est utilisé comme dispositif principal de débranchement. Veillez à ce que la prise de base soit située/installée près du module et facilement accessible.
Operation at AC input levels outside of the specified operating voltage range may damage the GP50.
B-2
Doc. 031377 -03 7/01
B • Installation
The GP50 power supply is auto-sensing, so no adjustment is required to select the line voltage. There are two ways to control power to the GP50.
Control from the GP50
_
To implement:
Connect a modular power cord (IEC 320 C13) from the GP50 main power receptacle (see Figure B-1) to a grounded, single-phase power source. Use the GP50 power switch to turn the pump power on and off.
Control from the LC30 Chromatography Oven
_
To implement:
Locate one of the IEC jumper cables (P/N 960748) provided in the LC30 Ship Kit. Connect the jumper cable from the GP50 main power receptacle to an IEC auxiliary receptacle on the LC30 rear panel. Leave the GP50 power switch on continuously and use the LC30 main power switch to turn the pump power on and off.
Doc. 031377-03 7/01
Figure B-1. GP50 Rear Panel
B-3
GP50 Gradient Pump

B.2.2 Electronics Chassis Connections

The electronics chassis connections vary, depending on which chromatography module is installed with the pump. Refer to the appropriate section below.
LC10 or LC20 Connections
1. Route the leak sensor cable from the rear of the LC10 Chromatography Organizer or the LC20 Chromatography Enclosure through the GP50 upper chase (see Figure B-1). Connect the cable to the
LC LEAK
Figure B-2).
2. Route the solenoid valve cable from the rear of the LC10 or LC20 through the GP50 upper chase. Connect the cable to the connector in slot 1 of the pump electronics chassis (see Figure B-2).
NOTE Refer to Appendix D for TTL and relay installation
instructions.
connector in slot 1 of the pump electronics chassis (see
LC AIR
Connect the leak sensor cable from the LC10 or LC20 here
Connect the serial interface cable from the LC30 here
Connect the solenoid valve cable from the LC10, LC20, LC25, or LC30 here
B-4
PWR SPY
L C
L E A K
GP50/IP25-DSP130W
MOTOR
L C
C O M M
DIST
P6
L C
A I R
SLOT 2SLOT 1 BLANK
D I S T
1 3
P I N
P7
D I S T
1 2
P I N
P8
SLOT 3
Figure B-2. GP50 Electronics Chassis
(Located behind pump upper door)
BLANK
SLOT 4
LAN-000K
RLY-1 OUT
RLY-2 OUT
+
TTL-1
-
OUT
+
TTL-2
-
OUT
+
TTL-1
-
IN
+
TTL-2
-
IN
+
TTL-3
-
IN
+
TTL-4
-
IN
POWER SUPPLY GREEN - OK RED - FAULT
Doc. 031377 -03 7/01
SLOT 5
CPU/RLY
F R O N T
P A N E L
B • Installation
LC25 Connections
Route the electronics injection valve cable from the rear of the LC25 Chromatography Oven through the GP50 upper chase(see Figure B-1). Connect the cable to the electronics chassis (see Figure B-2).
connector in slot 1 of the pump
LC AIR
NOTE Do not connect the LC25 to the
in slot 1; the oven contains internal leak control electronics.
LC30 Connections
LC LEAK
connector
1. Route the RJ-11 serial cable from the rear of the LC30 Chromatography Oven through the GP50 upper chase (see Figure B-1). Connect the cable to the
LC COMM
connector in slot 1 of
the pump electronics chassis (see Figure B-2).
2. Route the solenoid valve cable from the rear of the LC30 through the GP50 upper chase. Conne ct the c able to t he
connector in slot 1
LC AIR
(see Figure B-2).
NOTE Do not connect the LC30 to the
in slot 1; the oven contains internal leak control electronics.
LC LEAK
connector
Doc. 031377-03 7/01
B-5
GP50 Gradient Pump

B.2.3 DX-LAN Interface: 10BASE-T Connections (Optional)

NOTE Check the DX-LAN connector on the GP50 rear
panel. If a 10BASE-T RJ-45 (telephone-style) connector is installed, follow the instructions in Section B.2.3. However, if a BNC connector is installed, follow the instructions in Section B.2.4.
In order to communicate with a host computer running Dionex chromatography software, the GP50 must contain a pump interface card (P/N 056800) and an unshielded twisted-pair 10BASE-T DX-LAN cable (P/N 960281) must be connected from the 10BASE- T RJ-45 connector on the rear panel to a “combo” 10BASE-T Ethernet hub (P/N 056909).
Installing or Replacing the Pump Interface Card
NOTE If the DX-LAN option was not installed at the
factory, order the pump interface card kit (P/N 057005). The kit includes all the components required for DX-LAN communication.
B-6
STATIC— The GP50 electronics cannot be serviced by the user. The pump interface card must be installed by qualified personnel. Standard anti-static procedures must be observed when installing the interface card or handling the CPU card.
To prevent damage to the GP50, turn off the main power before installing the pump interface card. After confirming that the LED on the CPU card is off (not green or red), unplug the power. Do not rely on the front panel power switch.
Pour éviter d'endommager le GP50, coupez l'alimentation électrique principale avant d'installer la carte interface du pompe. Après avoir confirmé que la DEL de la carte d'unité central est étein te (ni ver te ni rouge), débranchez le courant. Ne vous fiez pas à la position de l'interrupteur d'alimentation du panneau avant.
Doc. 031377 -03 7/01
Loading...