Dionex DX-120 Operator's Manual

DX-120 ION CHROMATOGRAPH
OPERATOR’S MANUAL
© 1998 Dionex Corporatio n
Document No . 031183
Revision 03
September 1998
©1998 by Dionex Corpor ation All rights reserved worldwide. Printed in the United States of America.
This publication is protected by federal copyright law. No part of this publication may be copied or distributed, transmitted, transcribed, stored in a retrieval system, or transmitted into any human or computer language, in any form or by any means, electronic, mechanical, magnetic, manual, or otherwise, or disclosed to third parties without the express written permission of Dionex Corporation, 1228 Titan Way, Sunnyvale, California 94088-36 03 U.S.A.
DISCLAIMER OF WARRANTY AND LIMITED WARRANTY
THIS PUBLICATION IS PROVIDED “AS IS” WITHOUT WARRANTY OF ANY KIND. DIONEX CORPORATION DOES NOT WARRANT, GUARANTEE, OR MAKE ANY EXPRESS OR IMPLIED REPRESENTATIONS REGARDING THE USE, OR THE RESULTS OF THE USE, OF THIS PUBLICATION IN TERMS OF CORRECTNESS, ACCURACY, RELIABILITY, CURRENTNESS, OR OTHERWISE. FURTHER, DIONEX CORPORATION RESERVES THE RIGHT TO REVISE THIS PUBLICATION AND TO MAKE CHANGES FROM TIME TO TIME IN THE CONTENT HEREINOF WITHOUT OBLIGATION OF DIONEX CORPORATION TO NOTIFY ANY PERSON OR ORGANIZATION OF SUCH REVISION OR CHANGES.
EMISSIONS COMPLIANCE
This equipment meets ITE standard EN 55022 part A for radiated emissions and is suitable for use in a laboratory environment. Operation of this equipment in a residential area, however, may cause interference to radio and television reception.
TRADEMARKS
Teflon® an d Tefzel® are registered t rademarks of E.I. du Pont de Ne mours & Co. AutoSuppression, DX-LAN, IonSep®, MPIC®, OmniPac®, OnGuard, Self-Regenerating Suppressor, and SRS are trademarks of Dionex Corp.
PRINTING HISTORY
Revision 01, May 1996 Revision 02, March 1997 Revision 03, September 1998

Contents

1 • Introduction
1.1 About This Manual . . . . . . . . . . . . . . . . . 1-2
1.1.1 Typefaces and Conventions . . . . . . . . 1-3
1.1.2 Safety Messages and Notes . . . . . . . . 1-4
1.1.3 Symbols . . . . . . . . . . . . . . . . . . . 1-5
1.2 Related Manuals . . . . . . . . . . . . . . . . . . . 1-5
2 • Description
2.1 Operating Features . . . . . . . . . . . . . . . . . 2-1
2.1.1 Front Control Panel . . . . . . . . . . . . 2-3
2.1.2 Pump . . . . . . . . . . . . . . . . . . . . 2-6
2.1.3 Configuration DIP Switches . . . . . . . 2-6
2.1.4 Eluent Reservoirs . . . . . . . . . . . . . 2-7
Doc. 031138-03 9/98
2.1.5 Component Panel . . . . . . . . . . . . . 2-8
2.1.6 Rheodyne Injection Valve . . . . . . . . . 2-11
2.1.7 Detector Cells . . . . . . . . . . . . . . . 2-12
2.1.8 DS4 Detection Stabilizer . . . . . . . . . 2-13
2.2 Dual-Column Configuration Features . . . . . . . 2-15
2.2.1 Column Select Mode . . . . . . . . . . . 2-16
2.2.2 Eluent Select Mode . . . . . . . . . . . . 2-18
2.3 Fluid Schematics . . . . . . . . . . . . . . . . . . 2-19
2.4 Control Modes . . . . . . . . . . . . . . . . . . . . 2-22
2.4.1 Local Mode . . . . . . . . . . . . . . . . . 2-22
2.4.2 Remote Mode . . . . . . . . . . . . . . . 2-22
i
DX-120 Operator’s Manual
2.5 TTL Control . . . . . . . . . . . . . . . . . . . . . 2-23
2.5.1 Injection Valve/Remote Start Control . . 2-23
2.5.2 Recorder Range Control . . . . . . . . . . 2-23
3 • Operation and Maintenance
3.1 Preparing Eluents . . . . . . . . . . . . . . . . . . 3-1
3.1.1 Degassing Eluents . . . . . . . . . . . . . 3-1
3.1.2 Filtering Eluents . . . . . . . . . . . . . . 3-2
3.1.3 Pressurizing Eluent Reservoirs . . . . . . 3-2
3.2 Preparing Samples . . . . . . . . . . . . . . . . . 3-3
3.2.1 Collecting and Storing . . . . . . . . . . . 3-3
3.2.2 Pretreating . . . . . . . . . . . . . . . . . 3-3
3.2.3 Diluting . . . . . . . . . . . . . . . . . . . 3-4
3.3 Operating . . . . . . . . . . . . . . . . . . . . . . 3-5
3.3.1 Starting Up . . . . . . . . . . . . . . . . . 3-5
3.3.2 Injecting the Sample . . . . . . . . . . . . 3-7
3.4 Using an Integrator . . . . . . . . . . . . . . . . . 3-9
3.5 Running under PeakNet Control . . . . . . . . . . 3-9
3.6 Optimizing Temperature Compensation . . . . . . 3-10
3.6.1 With a DS4 . . . . . . . . . . . . . . . . . 3-10
3.6.2 With a CDM-3 Cell . . . . . . . . . . . . 3-10
3.7 Maintenance . . . . . . . . . . . . . . . . . . . . . 3-11
4 • Troubleshooting
4.1 Alarms . . . . . . . . . . . . . . . . . . . . . . . . 4-1
4.2 Error Codes . . . . . . . . . . . . . . . . . . . . . 4-3
4.3 Liquid Leaks . . . . . . . . . . . . . . . . . . . . . 4-4
4.4 Pump Difficult to Prime . . . . . . . . . . . . . . 4-8
ii
Doc. 031138-0 3 9/98
4.5 Pump Loses Prime . . . . . . . . . . . . . . . . . 4-8
4.6 Pump Does Not Start . . . . . . . . . . . . . . . . 4-9
4.7 No Flow . . . . . . . . . . . . . . . . . . . . . . . 4-9
4.8 Excessive System Backpressure . . . . . . . . . . 4-9
4.9 Peak “Ghosting” . . . . . . . . . . . . . . . . . . . 4-10
4.10 Nonreproducible Peak Height or Retention
Time . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
4.11 Abnormal Retention Time or Selectivity . . . . . 4-11
4.12 DS4 Temperature Inaccurate . . . . . . . . . . . . 4-12
4.13 No Detector Response . . . . . . . . . . . . . . . 4-13
4.14 Low Detector Output . . . . . . . . . . . . . . . . 4-14
4.15 High Detector Output . . . . . . . . . . . . . . . . 4-14
4.16 Noisy or Drifting Baseline . . . . . . . . . . . . . 4-15
5 • Service
Contents
Doc. 031138-03 9/98
5.1 Replacing Tubing and Fittings . . . . . . . . . . . 5-2
5.2 Changing the Sample Loop . . . . . . . . . . . . 5-2
5.3 Isolating a Restriction in the Liquid Plumbing . . 5-3
5.4 Replacing the DS4 Cell . . . . . . . . . . . . . . . 5-3
5.5 Cleaning Cell Electrodes . . . . . . . . . . . . . . 5-5
5.6 Calibrating the Cell Constant . . . . . . . . . . . 5-6
5.7 Calibrating the Pump Flow Rate . . . . . . . . . . 5-8
5.8 Cleaning and Replacing Pump Check Valves . . . 5-8
5.9 Replacing a Pump Piston Seal . . . . . . . . . . . 5-9
5.10 Replacing a Pump Piston . . . . . . . . . . . . . . 5-12
5.11 Replacing the Pressure Transducer Waste
Valve O-Ring . . . . . . . . . . . . . . . . . . . . 5-14
5.12 Changing the Main Power Fuses . . . . . . . . . . 5-15
iii
DX-120 Operator’s Manual
A • Specifications
A.1 Electrical . . . . . . . . . . . . . . . . . . . . . . . A-1
A.2 Environmental/Physical . . . . . . . . . . . . . . . A-1
A.3 Control Panel . . . . . . . . . . . . . . . . . . . . A-2
A.4 Pump . . . . . . . . . . . . . . . . . . . . . . . . . A-2
A.5 Pulse Damper . . . . . . . . . . . . . . . . . . . . A-2
A.6 Detector . . . . . . . . . . . . . . . . . . . . . . . A-2
A.7 Conductivity Cell . . . . . . . . . . . . . . . . . . A-3
A.8 DS4 Detection Stabilizer (Optional) . . . . . . . . A-3
A.9 Valves . . . . . . . . . . . . . . . . . . . . . . . . A-3
A.10 Delay Volume . . . . . . . . . . . . . . . . . . . . A-4
B • Installation
B.1 Facility Requirements . . . . . . . . . . . . . . . . B-1
iv
B.2 Installation Overview . . . . . . . . . . . . . . . B-1
B.3 Rear Panel Connections . . . . . . . . . . . . . . B-2
B.3.1 Power Connection . . . . . . . . . . . . . B-2
B.3.2 Switched AC Outlet Connection
(Optional) . . . . . . . . . . . . . . . . . . B-4
B.3.3 Waste Lines . . . . . . . . . . . . . . . . . B-5
B.3.4 Gas Connection . . . . . . . . . . . . . . B-5
B.3.5 DX-LAN Cable Connection (Optional) . B-6
B.4 Eluent Reservoir Connections . . . . . . . . . . . B-8
B.5 Pump Setup . . . . . . . . . . . . . . . . . . . . . B-8
B.5.1 Priming the Pump . . . . . . . . . . . . . B-8
B.5.2 Checking the Pump Flow Rate
Calibration . . . . . . . . . . . . . . . . . B-9
Doc. 031138-0 3 9/98
Contents
B.6 Connections to Chromatography Components . . B-10
B.6.1 Backpressure Requirements . . . . . . . . B-12
B.6.2 Self-Regenerating Suppressor (SRS)
Installation . . . . . . . . . . . . . . . . . B-13
B.6.3 Column Installation . . . . . . . . . . . . B-17
B.7 TTL Control Connections (Overview) . . . . . . . B-19
B.7.1 TTL Inputs . . . . . . . . . . . . . . . . . B-19
B.7.2 TTL Outputs . . . . . . . . . . . . . . . . B-20
B.7.3 F errite Core Installation on TTL Output
Cables . . . . . . . . . . . . . . . . . . . . B-21
B.8 Autosampler Connections (Optional) . . . . . . . B-22
B.8.1 Autosampler Outlet Line Connection . . B-22 B.8.2 AS40 Automated Sampler Connections . B-22 B.8.3 AS3 500 Automated Sampler
Connections . . . . . . . . . . . . . . . . B-24
B.9 Recorder and Integrator Connectio ns
(Optional) . . . . . . . . . . . . . . . . . . . . . . B-27
C • Integrator Programming
Doc. 031138-03 9/98
B.9.1 Chart Recorder Connections . . . . . . . B-27
B.9.2 4400 or 4600 Integrator Connections . . B-27
B.10 Configuration DIP Switch Settings . . . . . . . . B-29
B.11 DS4 Detection Stabilizer Installation . . . . . . . B-33
B.12 CDM-3 Cell Installation . . . . . . . . . . . . . . B-35
B.13 Injection Valve Connections . . . . . . . . . . . . B-36
B.14 Column Switching Valve Connections . . . . . . . B-37
B.15 DX-LAN Card Installation (Optional) . . . . . . . B-38
C.1 Integrator Power-up Configuration . . . . . . . . C-1
C.2 Setting Offsets . . . . . . . . . . . . . . . . . . . . C-2
C.3 Area Percent Mode . . . . . . . . . . . . . . . . . C-2
v
DX-120 Operator’s Manual
C.4 Using DIALOG to Create a Method . . . . . . . . C-3
C.4.1 Time Functions . . . . . . . . . . . . . . . C-3
C.4.2 Creating a Method . . . . . . . . . . . . . C-4
C.5 BASIC Programming . . . . . . . . . . . . . . . . C-7
D • Conductivity Detection
D.1 How Conductivity Is Measured . . . . . . . . . . D-1
D.2 Conductivity of Solutions . . . . . . . . . . . . . D-2
D.2.1 Effect of Hydration Sphere and
D.2.2 Effect of Temperature on Condu ctivity . D-4
D.2.3 Species Detected by Conductivity . . . . D-5
D.2.4 Chemical Suppression . . . . . . . . . . . D-6
D.2.5 Eluents for Conductivity Detection . . . D-9
Solvent on Conductivity . . . . . . . . . . D-4
vi
E • Glossary
Doc. 031138-0 3 9/98
1 • Introduction
The Dionex DX-120 Ion Chro matograph performs isocratic ion analysis applications using conductivity detection. The DX-120 is an integrated system, which includes a pump, detector, and injection valve. The chromatography components, including the column(s), Self-Regenerating Suppressor (SRS), and conductivity cell are ordered separately. These components are moun ted on the inside of the DX-120 door.
The DX-120 can be controlled locally, using the front panel keypad and display, or remotely, from a PeakNet workstation. The PeakNet workstation consists of a computer with a Dionex DX-LAN interface card and PeakNet software (Release 4.30 or higher) installed. Limited remote control is also available using TTL signals.
The DX-120 can be ordered as a single-column system or a dual-column system. The dual-column system a llows switching between two sets of columns (column select mode) or b etween two eluents (eluent select mode). In the eluent select mode, the DX-120 can perform step gradients.
The following models are available:
Configuration Voltage/Line Frequency* Without
Single-column 115 VAC/60 Hz P/N 050100 P/N 050200
Dual-column 115 VAC/60 Hz P/N 050102 P/N 0502 02
*Must match the voltage and line frequency of the installation site’s power source.
Doc. 031183-03 9/98
With
DX-LAN
100 VAC/50 Hz P/N 050103 P/N 0502 03 230 VAC/50 Hz P/N 050106 P/N 0502 06
100 VAC/50 Hz P/N 050105 P/N 0502 05 230 VAC/50 Hz P/N 050108 P/N 0502 08
DX-LAN
1-1
DX-120 Operator’s Manual

1.1 About This Manual

Chapter 1
Introduction
Chapter 2
Description
Chapter 3
Operation and
Maintenance
Chapter 4
Troubleshooting
Chapter 5
Service
Appendix A
Specifications
Appendix B
Installation
Appendix C
Integrator
Programming
Introduces the DX-120 and explains the conventions used in the manual, including safety-related information.
Describes the DX-120 operating features, the chromatographic fluid path, and the control modes.
Provides operating and routine preventive maintenance procedures.
Lists problems, with step-by-step procedures to isolate and eliminate their sources.
Provides step-by-step instructions for routine service and parts replacement procedures.
Lists the DX-120 specifications and installation site requirements.
Describes how to install the DX-120.
Describes how to program a Dionex 4400 or 4600 integrator for automated control of the DX-120.
1-2
Appendix D
Conductivity
Appendix E
Detection
Glossary
Describes conductivity detection and its applications.
Defines terms commonly used in ion chromatography.
Doc. 031183-0 3 9/98

1.1 .1 Typefaces and Conventions

Capitalized bold type indicates a front panel button. For example:
1 • Introduction
Press
Upper-case bold type indicates information displayed on
Pump
to turn on the pump.
the front panel screen. For example:
LEAK ALARM
Upper-case italic type indicates a grouping of front panel
displays when a leak occurs.
buttons. For example: Use the buttons in the DISPLAY group to select the type
of information shown on-screen.
When a function can be controlled by a DIP switch setting, the switch and position numbers are in parentheses. For example:
(SW1-3) indicates DIP switch 1, position 3. (SW2-5, 6, 7, 8) indicates DIP switch 2, positions 5, 6, 7,
and 8.
Doc. 031183-03 9/98
1-3
DX-120 Operator’s Manual

1.1 . 2 Safety Mess ages and Notes

The DX-120 meets European, EMC, and safety requirements per Council Directives 73/23/EEC and 89/336/EEC, EN 61010-1:1993 (safety), EN 5008 2-1:1992 (susceptibility), and EN 55011:1991 (emissions). The TUV/CE and GS safety label on the DX-120 attests to compliance with these standards.
The DX-120 is designed for ion chromatography applications and should not be used fo r any other purpose. If there is a question regarding appropria te usage, contact Dionex before proceeding.
This manual con tains warnings and precau tionary statements that, when properly followed, can prevent personal injury and/or damage to the instrument. Safety messages appear in bold type and are accompanied by icons, as follows:
Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury.
1-4
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury.
Indicates that the function or process of the instrument may be impaired. Operation does not constitute a hazard.
Doc. 031183-0 3 9/98
Informational messages also appear throughout this manu al. These are label ed NOTE and are in bold type:
NOTES call attention to certain information. They alert you to an unexpected result of an action, suggest how to optimize instrument performance, etc.

1.1.3 Symbols

The symbols below appear on the DX-120, or on DX-120 labels.
1 • Introduction
NOTE
~
Alternating current
Protective conductor terminal
Power supply is on
Power supply is off

1.2 Related Manuals

During installation and operation of the DX-120, you may need to refer to one or more of the following manuals for information about other components or instru ments in the system.
4440 Integrator User’s Guide (Document No. 034200) or 4600
Integrator User’s Guide (Document No. 034408) AS40 Automated Samp ler Operator’s Manual (Document
No. 034970) Installation of Dionex Ferrule Fittings (Document No. 034213)
Doc. 031183-03 9/98
1-5
DX-120 Operator’s Manual
PeakNet Software User’s Guide (Document No. 0349 14)
Installing the PeakNet System (Document No. 034941)
1-6
Doc. 031183-0 3 9/98

2 Description

2.1 Operating Features

Figures 2-1 and 2-2 illustrate the main operating features of the DX-120.
Doc. 031183-03 9/98
Figure 2-1. DX-120 Operating Features (Exterior)
2-1
DX-120 Operator’s Manual
2-2
Figure 2-2. DX-120 Operating Features (Interior)
Doc. 031183-0 3 9/98

2.1 .1 Front Control Panel

The control panel liquid crystal d isplay (LCD) shows status information and alarm conditions. Press a button in the group labeled DISPLAY to determine the type of status information shown. The remaining buttons control DX-120 operation.
Figure 2-3. DX-120 Control Panel
2 • Description
Button Function
DISPLAY
Doc. 031183-03 9/98
Displays the pump flow rate setting (0.5 to 4.5 mL/min).
Displays the pump pressure transducer reading (0 to
27.6 MPa or 0 to 4000 psi)
NOTE MPa is the default pressure unit. For psi, set DIP SW1-4 on. See Section B.10 for details.
Displays the total conductivity reading (0 to 999.9 µS).
Displays the offset conductivity reading (-999.9 to
999.9 µS). Offset conductivity is the total conductivity minus the offset for the current run (see the description of the Auto Offset button).
Table 2-1. Control Panel Button Functions
2-3
DX-120 Operator’s Manual
Button Function
RECORDER
Delivers a chart mark signal to the analog output. The mark is 10% of the full-scale voltage.
Delivers a 100% signal to the analog output. Pressing the button continuously keeps the output signal at full scale. The default full-scale voltage is 1 V, but can be changed to 10 V (SW4-4).
Reduces th e analog output si gnal to zero. Pres sing the button continuously keeps the output signal at zero.
COMPONENT ON/OFF
Turns the gas pressure to the eluent reservoirs on and off. When on, the LED on the button is illuminated. This button is
disabled
Local/Remote below). Turns the pump flow on and o ff. When on, the LED on the
button is illuminated. This button functions in Local or Remote mode.
Turns the SRS power on and off, as well as the DS4 Detection Stabilizer, if installed. When on, the LED on the button is illuminated. In the dual-column configuration, this button controls the power to the SRS and DS4 of the selected column set (see Section 2.2.1). This button is
disabled
Local/Remote below).
when the DX-120 is in Remote mode (see
when the DX-120 is in Remote mode (see
2-4
SYSTEM CONTROL
Toggles between Local and Remote control modes. Local is control from the DX-120 control panel and Remote is control from a PeakNet workstation via the DX-LAN. The selected mode is shown in the lower right-hand corner of the display.
A red LED on the left side of the button indicates an alarm condition. The top line of the display in dicates the source of the alarm: cabinet or cell leak, high or low pressure, or SRS. Pressing the button clears the alarm. If the alarm condition is still present, the alarm reappears after 15 seconds.
Table 2-1. Control Panel Button Functions (continued)
Doc. 031183-0 3 9/98
Button Function
Offsets the background conductivity. After start-up, allow the system to equilibrate. The display shows the background conductivity (the eluent conductivity before sam ple injection). Press Auto Offset to offset this background reading and zero the chromatogram baseline.
Switches the injection valve between the Load and Inject positions. The current position is shown in the lower left-hand co rner of the dis play.
When switched from Load to Injec t, the DX-120 also:
- sends a chart mark signal to the analog output (SW4-7)
- performs an auto offset (SW3 -3)
- sends a TTL signal indicating that an injection has occurred
After 1 minute, the valve automatically returns to the Load position (SW1-1).
This button is mode.
COLUMN SELECTION
The COLUMN SELECTION buttons are active in the dual-column configuration only.
In column select mode, this button initiates the column switching sequence from column set B to column set A (see Section 2.2.1). In eluent select mode, this button switches to eluent delivery from line A (see Section 2.2. 2).
In column select mode, this button initiates the column switching sequence from column set A to column set B (see Section 2.2.1). In eluent select mode, this button switches to eluent delivery from line B (see Section 2.2.2).
disabled
2 • Description
when the DX-120 is in Remote
Doc. 031183-03 9/98
These buttons are mode.
disabled
when the DX-120 is in Remote
Table 2-1. Control Panel Button Functions (continued)
2-5
DX-120 Operator’s Manual

2.1.2 Pump

The pump is located on the rig ht side of the main compartment (see Figure 2-2). The knob o n the front of the pump adjusts the flow rate from 0.5 to 4.5 mL/min.
Eluent Save Mode
If the DX-120 is idle (i.e., no control panel buttons have been pressed and no PeakNet commands have been received) for 90 minutes, the pump flow automatically decreases to 1/20 th of its cur rent flow rate and the SRS cycles on and off. When this occurs, the LEDs on the Press any button to return to the last selected flow rate.
To turn off this feature, reset the Pump Time-out DIP switch (SW1-2).

2.1 .3 Configuration DIP Switches

Pump
and
SRS
buttons flash.
The DIP switches on the left side of the main compartment control system parameters. The factory-set defaults can be changed to meet specific system and application requirements (see Section B.10).
NOTE In this manual, when a function is controlled by a DIP switch setting, the switch and position numbers are shown in parentheses. For example: (SW1-3, 4) indicates DIP switch 1, positions 3 and 4.
2-6
Doc. 031183-0 3 9/98

2.1.4 Eluent Reservoirs

Dionex strongly recommends degassing all elue nts and storing them in reservoirs pressurized with helium. This helps prevent bubbles (resulting from eluent outgassing) from forming in the pump head and the detector cell. Degassed eluents and pressu rized reservoirs are especially impo rtant when combining aqueo us and non-aqueous components (e. g., water and methanol). With non-aqueous components, glass reservoirs are recommended.
The single-column DX-120 in cludes one 2-liter plastic reservoir (P/N 0441 29). The dual-colu mn DX-120 includes two 2-liter plastic reservoirs.
The following additional reservoirs are available from Dionex:
1-liter plastic reservoir (P/N 044128)
1-liter glass reservoir with shatterproof plastic coating (P/N 044126)
2 • Description
Doc. 031183-03 9/98
2-liter glass reservoir with shatterproof plastic coating (P/N 044127)
The 2-liter plastic reservoir is not designed for vacuum degassing. Do not use it for this purpose.
2-7
DX-120 Operator’s Manual

2.1 .5 Component Panel

Chromatography components are mounted on the inside fron t door. Figure 2-4 shows the single-column component panel layout. Figure 2-5 shows the dual-column layout.
The DX-120 is equipped with a Rheodyne injection valve (see Section 2.1.6). The following additional components must be ordered separately:
Self-Regenerating Suppressor(s) (SRS): The SRS neutralizes the eluent and enha nces analyte conductivity. For a dual-column system, order two suppressors.
Column(s): One or two analytical columns and one or two guard columns can be installed on colu mn clips. The clips have larger clasps on one side for supporting 4 mm columns and smaller clasps on the other side for 2 mm columns.
Column switching valve: The column switching valve is installed only in a dual-column system. Th e valve controls liquid flow to the selected column (in column select mode) or from the selected eluent (in eluent sele ct mode). See Section 2.2 for details abou t dual-column systems.
2-8
Detector cell: Only one flow-through cell is required in either the single- or the dual-column system. Two cell models are available: a DX-120 standard cell (the CDM-3) and a DX-120 high-performance cell with heater (the DS4 Detection Stabilizer). The DS4 is recommended for applications requiring enhanced thermal stability. See sections 2.1.7 and 2.1.8 for details about the cells and the DS4.
Doc. 031183-0 3 9/98
SRS
2 • Description
CDM-3 Cell
Injec tion Va lv e
Doc. 031183-03 9/98
Guard Column
Analytical Colum n
Figure 2-4. Single-Column Component Panel Layout
NOTE A DS4 Detection Stabilizer can be installed instead of the CDM-3 cell.
2-9
DX-120 Operator’s Manual
2-10
Figure 2-5. Dual-Column Component Panel Layout
NOTE A CDM-3 cell can be installed instead of the DS4 Detection Stabilizer.
Doc. 031183-0 3 9/98

2.1 .6 Rheodyne Injection Va lve

The Rheodyne injection valve has two operating positions: Load and Inject. In the Load position, sample is loaded into the sample loop, where it is held until injection. In the Inject position, sample is swept to the column for analysis. Eluent flows through one of two paths, depending on the valve position:
In the Load position, eluent flows from the pump, through the valve, and to the column, bypassing the sample loo p. Sample flows from the syringe or autosampler line, through the valve, and into the sample loop; excess sample flows out to waste.
In the Inject position, eluent flows from the pump, through the sample loop, and on to the column, carrying the contents of the sample loop with it.
2 • Description
Doc. 031183-03 9/98
Figure 2-6. Rheodyne Injection Valve Flow Schematics
2-11
DX-120 Operator’s Manual

2.1.7 Detector Cells

The DX-120 accommodates two detector cell models. The DX-120 standard cell (CDM-3; P/N 050776) is used for applications that do not require th e enhanced baseline stability gained through thermal stabilization. For increased thermal stabilization in high-sensitivity applications, use a DX-120 high-performance cell with heater (DS4 Detection Stabilizer; P/N 050218).
If you change the cell model, the Cell Type DIP switches (SW4-1, 2) must be reset to sel ect the new cell type.
Detector Cell Features
Both cells are flow-through conductivity cells with polymeric bodies. Two 316 stainless steel electrodes are permanently sealed into the cell bodies.
NOTE
A sensor (thermistor) located slightly downstream from the electrodes senses the temperature of the liquid as it exits the cell. The measured value is used for temperature compensation.
The active volume is nominally 1.25 µL for the CDM-3 cell and 1.0 µL for the DX -120 high-perfo rmance cell.
The detector cell constant for both cells has a nominal value of 160 cm
-1
.
The advanced geometry of the cells provide several benefits:
Excellent accuracy and linearity over a broad working range
Efficient sweepout and low volume for low dispersion
Reduced sensitivity to electrode surface conditions
Low electrode mass
Effective temperature comp ensation
2-12
Doc. 031183-0 3 9/98
2 • Description
Temperature Control and Compensation
Temperature directly affects the conductivity of a solution. For example, laboratory heating and air conditioning systems can cause a regular slow cycling in the baseline. This, in turn , can affect the reproducibility of an analysis. The higher the conductivity, the more pronounced the effect.
In ion chromatography, suppressing eluent conductivity minimizes the effect of temperature variation. Temperature compensation further improves baseline stability. When the conductivity cell is housed in a DS4 Detection Stabilizer, the heater enhances the ability of these techniques to reduce temperature effects on conductivity.
Temperature compensation also ensures that there is no major change in the baseline or peak heights, should it be necessary to change the DS4 operating set point. Readings will be normalized to 25 °C.

2.1.8 DS4 Detection Stabilizer

Doc. 031183-03 9/98
The DS4 is a temperature-controlled chamber consisting of a cast aluminum base and cover enclosed in insulating foam. The chamber houses both the cond uctivity cell and the eluent heat exchanger. Figure 2-7 identifies the major components of the DS4.
The DS4 provides the following benefits:
Conductivity measurements nearly impervious to laboratory temperature variation
Very low disp ersion in the eluent heat exchan ger
Excellent peak height reproducibility
User-selected temperature set point
2-13
DX-120 Operator’s Manual
2-14
Figure 2-7. DS4 Detection Stabilizer Features
Power input to a pair of transistors on opposite sid es of the DS4 heats it to a user-selected temperature from 30 °C to 45 °C (SW3-4, 5, 6, 7). A sensor near the heat exchang er outlet senses the eluent temperature. The DX-120 circuitry compares this temperature with the selected temperature and adjusts the heat input in real time to hold the temperature within a few millidegrees.
The DS4 is sealed with an O-ring to trap eluent that may leak from the cell. If 5 mL of liquid accumulates, a thermistor sensor signals a leak to the CPU. Any additional leakage will be discharged via the spill/overflow line. A second thermistor, above the discharge level, acts as a temperature reference for the leak sensor.
Doc. 031183-0 3 9/98

2.2 Dual-Column Configuration Features

The dual-column configuration has two operating modes:
Column select mode allows switching of flow from one column
set to the other. Eluent select mode allows switching of flow from one eluent to
the other (the column set is not switched).
The Column Select DIP switch (SW3-1) selects the mode: on=column; off=eluent.
The dual-column system option adds the f ollowing features:
2 • Description
Column A
The
and
Column B
buttons on the front contro l panel
are enabled . An eluent selection valve selects which eluent reservoir is used.
A column switching valve directs flow to the selected column set
(column select mode only).
Doc. 031183-03 9/98
2-15
DX-120 Operator’s Manual

2.2 .1 Column Select Mode

In column select mode, you can switc h the flow path from one column set to the other by pressing
B
or by sending a command from PeakNet.
Figure 2-8 illustrates the sequence of events when you switch from column set A to column set B.
Column A
or
Column
2-16
Figure 2-8. Column Selection Sequence
Doc. 031183-0 3 9/98
Loading...
+ 132 hidden pages