Dionex CD25 Operator's Manual

Page 1
CD25 CONDUCTIVITY DETECTOR
OPERATOR'S MANUAL
© 2000 Dionex Corporation
Document No. 031687
Revision 01 April 2000
Page 2
©2000 by Dionex Corporation All rights reserved worldwide. Printed in the United States of Amer ica.
This publication is protected by federal copyright law. No part of this publication may be copied or distributed, transmitted, transcribed, stored in a retrieval system, or transmitted into any human or computer language, in any form or by any means, electronic, mechanical, magnetic, manual, or otherwise, or disclosed to third parties without the express written permission of Dionex Corporation, 1228 Titan Way, Sunnyvale, California 94088-3603 U.S.A.
DISCLAIMER OF WARRANTY AND LIMITED WARRANTY
THIS PUBLICATION IS PROVIDED “AS IS” WITHOUT WARRANTY OF ANY KIND. DIONEX CORPORATION DOES NOT WARRANT, GUARANTEE, OR MAKE ANY EXPRESS OR IMPLIED REPRESENTATIONS REGARDING THE USE, OR THE RESULTS OF THE USE, OF THIS PUBLICATION IN TERMS OF CORRECTNESS, ACCURACY , RELIABILITY, CURRENTNESS, OR OTHERWISE. FURTHER, DIONEX CORPORATION RESERVES THE RIGHT TO REVISE THIS PUBLICATION AND TO MAKE CHANGES FROM TIME TO TIME IN THE CONTENT HEREINOF WITHOUT OBLIGATION OF DIONEX CORPORATION TO NOTIFY ANY PERSON OR ORGANIZATION OF SUCH REVISION OR CHANGES.
TRADEMARKS
Teflon® and Tefzel® are registered trademarks of E.I. duPont de Nemours & Co. AutoSuppression™, DX-LAN™, IonSep®, MPIC™, OmniPac®, OnGuard™, SRS®, and Self-Regenerating Suppressor® are trademarks of Dionex Corp.
PRINTING HISTORY
Revision 01, April 2000
Page 3

Contents

1 • Introduction
1.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
1.2 About This Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
1.2.1 Typefaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
1.2.2 Safety Messages and Notes . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
1.3 Safety Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
2 • Description
2.1 Front Control Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1
2.1.1 Control Panel Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1
2.1.2 Control Panel Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
2.2 Electronics Chassis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8
2.3 Conductivity Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-12
2.4 Rear Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-15
2.5 Functional Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-16
Doc. 031687-01 4/2000
2.1.3 Initial Display Screens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
2.2.1 Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-9
2.2.2 Cards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-9
2.3.1 DS3 Detection Stabilizer . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-13
2.3.2 Shielded Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-14
2.5.1 Operating and Control Modes . . . . . . . . . . . . . . . . . . . . . . . .2-16
i
Page 4
CD25 Conductivity Detector
2.5.2 Local and Remote Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-17
2.5.3 Method Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-17
3 • Operation and Maintenance
3.1 Getting Ready to Run . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3.2 Running Under Direct Control (Local Mode) . . . . . . . . . . . . . . . . . . . 3-2
3.3 Running Under Method Control (Local Mode) . . . . . . . . . . . . . . . . . . 3-3
3.3.1 Running a Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
3.3.2 Changing the Running Method . . . . . . . . . . . . . . . . . . . . . . . . 3-3
3.3.3 Creating a New Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
3.3.4 Editing an Existing Method . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
3.4 Optimizing Temperature Compensation . . . . . . . . . . . . . . . . . . . . . . . 3-7
3.4.1 With a DS3 and Conductivity Cell . . . . . . . . . . . . . . . . . . . . . 3-7
3.4.2 With a Shielded Conductivity Cell . . . . . . . . . . . . . . . . . . . . . 3-8
3.5 Routine Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
4 • Troubleshooting
4.1 No Detector Response . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
4.2 Low Detector Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
4.3 High Detector Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
4.4 Noisy or Drifting Baseline . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2
4.5 Liquid Leaks from the DS3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
4.6 DS3 Temperature Inaccurate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
4.7 Conductivity Inaccurate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
ii
Doc. 031687-01 4/2000
Page 5
4.8 Faulty DX-LAN Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-7
4.9 Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-8
5•Service
5.1 Eliminating Liquid Leaks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-2
5.2 Removing Trapped Air from the Cell . . . . . . . . . . . . . . . . . . . . . . . . . .5-2
5.3 Replacing the DS3 Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-2
5.4 Calibrating the Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-4
5.5 Replacing the Main Power Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-6
A • Specifications
Contents
5.6 Physical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
A.1 Environmental . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
A.2 Electrical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
A.3 Display and Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
A.4 Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
A.5 Conductivity Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
A.6 SRS Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3
A.7 DS3 Detection Stabilizer (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . A-3
B • Installation
B.1 Facility Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1
B.2 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1
Doc. 031687-01 4/2000
iii
Page 6
CD25 Conductivity Detector
B.2.1 Power Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-1
B.2.2 DX-L AN Interfa ce: 10BASE-T Connection s (Optional ). . . . .B-3
B.2.3 DX-LAN Interface: BNC Connections (Optional). . . . . . . . . .B-6
B.2.4 DS3 Detection Stabilizer Installation. . . . . . . . . . . . . . . . . . . .B-9
B.2.5 Shielded Conductivity Cell Installation . . . . . . . . . . . . . . . . . .B-9
B.2.6 DS3 or Shielded Cell Plumbing. . . . . . . . . . . . . . . . . . . . . . . .B-9
B.2.7 Recorder/Diagnostic Connection . . . . . . . . . . . . . . . . . . . . . .B-17
B.3 Automatic SRS Power Control (Optional) . . . . . . . . . . . . . . . . . . . . .B-18
C • User Interface
C.1 Operational Screens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-2
C.1.1 Menu of Screens. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-2
iv
C.1.2 Main Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-3
C.1.3 Detail Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-4
C.1.4 Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-5
C.1.5 Module Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-7
C.1.6 Analog Out Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-8
C.1.7 Time Function In . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-9
C.2 Diagnostic Screens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-10
C.2.1 Diagnostic Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-10
C.2.2 Power-Up Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-11
C.2.3 Elapsed Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-12
C.2.4 Analog Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-13
C.2.5 DX-LAN Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-14
Doc. 031687-01 4/2000
Page 7
C.2.6 Keyboard Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-16
C.2.7 Diagnostic Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-17
C.2.8 Leak Sensor Calibration and Status. . . . . . . . . . . . . . . . . . . . C-19
C.2.9 Signal Statistics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-20
C.2.10 Calibrate Conductivity Cell. . . . . . . . . . . . . . . . . . . . . . . . . . C-21
D • TTL and Relay Control
D.1 TTL and Relay Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-2
D.2 TTL and Relay Output Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-3
D.3 TTL Input Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-4
D.3.1 TTL Input Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-4
D.3.2 TTL Input Signal Modes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-5
Contents
E • Signal Processor Functions
F • Connector Pinouts
F.1 Recorder/Diagnostic Signal Pinouts . . . . . . . . . . . . . . . . . . . . . . . . . . F-1
F.1.1 Signal Electrical Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . F-2
F.2 TTL/Relay Pinouts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-4
F.3 DS3 Connector Pinouts—SCR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-5
F.4 SRS Connector Pinouts—SCR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-6
F.5 Conductivity Cell Connector Pinouts—SP . . . . . . . . . . . . . . . . . . . . . F-6
G • Reordering Information
Doc. 031687-01 4/2000
v
Page 8
CD25 Conductivity Detector
vi
Doc. 031687-01 4/2000
Page 9

1.1 Overview

The CD25 Conductivity Det ect or is a sensitive, accurate , and versatile ins tr ument for detecting and quant ifying ionic analytes in liq uid and ion chromatog raphy. It is especially useful for analytes that lack UV chromophores and cannot be determined with adequate sensitivity by UV absorbance. Conductivity detection, especially when combined with chemical eluent suppression, provides excellent sensitivity and selectivity for numerous ionic species, both organic and inorganic.
The major organic analytes are carboxylic, sulfonic, and phosphonic acids; and primary, secondary, tertiary, and quaternary amines. Inorganic analytes include strong acid anions such as the halid es, sulfate , nitrat e, and phos phate; alkali metal and alkaline earth cations.
The CD25 ca n be controll ed locally, from the front panel, or remotely (via th e Dionex DX-LAN™ interface) from a host computer running PeakNet 6, Release
6.1 (or later) software.
1 • Introduction
Doc. 031687-01 04/2000
1-1
Page 10
CD25 Conductivity Detector

1.2 About This Manual

Chapter 1
Introduction
Chapter 2
Description
Chapter 3
Operation and
Maintenance
Chapter 4
Troubleshooting
Chapter 5
Service
Appendix A
Specifications
Appendix B
Installation
Appendix C
Display Screens
Provides a brief overview of the CD25 Conductivity Detector. Explains the meaning of safety messages and icons in the manual and safety labels on the detector.
Describes physical aspects of the CD25, including the front panel controls, electronics, and flow cell. Explains the detecto r operating m odes.
Describes operating features and how to create, edit, and run methods from the CD25 front panel. Lists routine preventive maintenance requirements.
Lists possible causes of problems and step-by-step procedures to isolate and elimina te th em.
Contains step-by-step instructions for routine service and parts replacement procedures.
Lists the CD 25 specifica tions and installation site requirements.
Describes how to install the CD25.
Illustrates and describes all operating and diagnostic screens that can be displayed on the front panel.
1-2
Appendix D
TTL and Relay
Control
Appendix E
Signal Processor
Functions
Appendix F
Connector
Pinouts
Appendix G
Reordering
Information
Describes TTL and relay control functions. Provides connection instructions.
Lists the functions of the Signal Processor (SP) card.
Describes the pinouts for all CD25 connectors.
Lists spare parts for the detector.
Doc. 031687-01 04/2000
Page 11

1.2.1 Typefaces

Capitalized bold type indicates a front panel button:
Enter
Press
Uppercase bold type indicates the name of a menu or a screen, or an on-screen entry:
1 • Introduction
to begin running the method.
Go to the
METHOD
Move the cursor to the
screen.
EDIT
field.

1.2.2 Safety Messages and Notes

This manual contains warnings and precautionary statements that can
prevent personal injury and/or damage to the CD25 when properly
followed. Safety messages appear in bold type and are accompanied by
icons, as shown below.
Indicates an imminently hazardous situation which, if not
avoided, will result in death or serious injury.
Indicates a potentially hazardous situation which, if not
avoided, could result in death or serious injury.
Indicates a potentially hazardous situation which, if not
avoided, may result in minor or moderate injury.
Indicates that the function or process of the instrument may be
impaired. Operation does not constitute a hazard.
Doc. 031687-01 04/2000
1-3
Page 12
CD25 Conductivity Detector
Informational messages also appear throughout this manual. These are labeled NOTE and are in bold type:
NOTE NOTES call attention to certain information. They alert
you to an unexpected result of an action, suggest how to
optimize instrument performance, etc.

1.3 Safety Labels

The CE and GS safety label on the CD25 attests to compliance with the following European, EMC, and safety requirements: Council Directives 73/23/EEC and 89/ 336/EEC, EN 61010-1:1993 (safety), EN 50082-1:1992 (susceptibility), and EN 55011:1991 (emissions).
The symbols below appear on the CD25 or on CD25 labels.
Alternating current
˜
Protective conductor terminal (earth ground)
Power supply is on
Power supply is off
1-4
Doc. 031687-01 04/2000
Page 13

2.1 Front Control Panel

The control panel on the f ront door of t he CD25 contains the liquid crystal di splay (LCD), the membrane keypad, and the actuator for the main power switch. The electronics chassis, described in Section 2.1.3, is located behind the front door.
Power Switches
The main power switch is on a bulkhead inside the electronics chassis (in the front, left-hand corner). The actuator for the power switch is on the front door, below the control panel (see Figure 2.1.2). The actuator functions only when the front door is fully closed. When the door is open, press the main power switch to turn the CD25 on and off.

2.1.1 Control Panel Display

2 • Description
and
screen
, displays CD25 status and operating
Menu
buttons (see Figure 2-1).
option on the
MODULE SETUP
screen
The LCD, also called the
information. Fields on the sc ree n tha t ar e in reverse video (blue lette rs on
a white background) can be edited, while normal video fields are simply
informational displays.
To adjust the screen contrast, rotate the knurled knob in the recess below the
To adjust the brightness of the screen backlight, select a different
DISPLAY PANEL BACKLIGHT
(see Section C.1.5).
Help
Doc. 031687-01 04/2000
2-1
Page 14
CD25 Conductivity Detector
MAIN SCREEN DETAIL SCREEN METHOD
---



E
!
"

 

Main Power Switch Actuator


Ta b (for opening the door)
Figure 2-1. CD25 Front Panel

2.1.2 Control Panel Keypad

The buttons on the front panel keypad are used to affect screen functions, to directly control CD25 operat ion, and to create and modif y programmed series of timed events, called methods. For example, pressing turns the method clock off displays a list of availa ble screens. In summary:
To go from a menu to a screen, press the number button that corresponds to the screen’s number on the menu, or move the cursor to the desired screen name and press
Hold
(

Knob (for a d ju s tin g the contrast)
) and on
Run
, while pressing
(
)
Enter
.
#

Hold/Run
Menu
2-2
To edit a reverse video entry field on a screen, move the cursor to the field using the four directional arrow buttons. Then, use the numeric
Doc. 031687-01 04/2000
Page 15
2•Description
Keypad Buttons
Offset
Mark
buttons to enter variable values or the
Select
and
Select
to choose predetermined values.
In most cases, confirm the selected value, by pressing either a cursor arrow button or must press
METHOD SAVE TO
the command, the
NOTE A beep sounds each time you press a keypad button,
and a lower frequency beep sounds whenever an error occurs. You can disable the beeps from the
MODULE SETUP
Function
Returns the analog (recorder) output to a predetermined baseline and zeros the display. The resultant value of the offset required is displayed on the
Sends a 10% positive event mark to the analog (recorder) output. A mark is typically used to indicate a sample injection. This fun ction can be programmed in a method.
Enter
. In certain fields and/or screens, however, you
Enter
to confirm the selected value. These exception s are
field, the
DIAGNOSTIC TESTS
DETAIL
screen. This function can be programmed in a method.
METHOD RUN
field, any calibration
screen, and all menu screens.
screen (see Section C.1.5).
buttons
Insert
Doc. 031687-01 04/2000
Inserts a new step into a method. T o add a new s tep, mov e the cur sor to the
A new blank step appears below the cursor position. After entering a time value, press order and they will be automatically reorganized in the correct chronological order.
Enter
or a cursor arrow button. Insert steps in any
TIME
field and pr
ess
Insert
.
2-3
Page 16
CD25 Conductivity Detector
Keypad Buttons
Delete
Hold/Run
Function
Removes the value from the curren t entry field. To restore the previous value, move the cursor from the field before entering a new value.
Pressing
METHOD
Delete
when the cursor is in a step entry field on the
screen “blanks” the step parameter value. Moving the cu rsor to another field does not restore the previous value; instead, the step remains blank, indicating no change from the previous step .
To delete an entire method step:
1. Position the cursor in the method’s time field and press
Delete
.
The time is removed and the help line prompts you to press
Delete
again to delete the step.
2. Press
Turns the method clock off (
Delete
again. Or, to restore the original time and step
parameters, press any button except
Hold
) and on
Delete
(Run)
.
. This button functions only when the detector is under Method control (see Section 2.5.3).
When the method clock is in Hold, pressi ng
Hold/Run
starts the clock at either the initial step of a new method or, if resuming an interrupted method, at the time at which the clock was put in Hold.
2-4
Rese
Select
and
Select
When the method clock is in Run, pressing
Hold/Run
stops the method clock, thereby “holding” the method and freezing the current conditions.
t
Changes the method clock time to
INIT
, causing the initial conditions specified by the method to occur. This button functions only when the detector is under Method control (see Section 2.5.3).
If the method is running, it continues running. If the method is in Hold, the method clock executes the initial conditions and holds.
When the cursor is positioned at a field with predetermined parameters, these buttons cycle through the options. In fields with
predetermined numerical values, pressing
Select
Enter
Select
button increases (or decreases) the value
by one unit; pressing Holding down a continuously. Press
decreases the value by one unit.
or a cursor arrow button to confirm the
Select
increases the value
selected value.
Doc. 031687-01 04/2000
Page 17
2•Description
Keypad Buttons
Help
Menu
Numeric Buttons
Function
The four cursor buttons move the cursor, in the direction of the arrow, to the next entry field. If there is no changeable field in that direction, the cursor moves diagonally or remains in its current location.
In most cases, after entering a new value in an entry field, pressing an arrow button saves and/or executes the change. The exceptions are the
METHOD SAVE T O
command, the
DIAGNOSTIC TESTS
field, the
METHOD RUN
screen, and all menu screens.
field,
any calibration
Displays a help screen with information pertaining to the current entry field.
Displays one of two menus, depending on the current screen:
From an operational screen, pressing
SCREENS
From a diagnostic screen, pressing
DIAGNOSTIC MENU MENU of SCREENS
.
; pressing
.
Menu
displays the
Menu
once returns you to the
Menu
again returns you to the
MENU of
Enters numeric values into the current e ntry field. The numer ic buttons are 0 through 9 and the decimal.
From a menu screen, pressing a numeric button opens the corresponding screen.
Enter
Doc. 031687-01 04/2000
Saves and/or executes changes made in entry fields. I f a menu screen is displayed, pressing
Enter
opens the highlighed screen.
2-5
Page 18
CD25 Conductivity Detector

2.1.3 Initial Display Screens

When the CD25 has successfully powered-up and passed all diagnostic tests, the the power-up, the screen. See Section C.2.7 if this occurs.
POWER-UP
MAIN
screen (see Figure 2-3). If one of the diagnostic tests fails at
DIAGNOSTIC TEST
screen (see Figure 2-2) displays briefly, followed by
screen disp lays instead of the
MAIN
NOTE The information on the
reviewed at any time by selecting the screen from
DIAGNOSTIC MENU
the
CD25 CONDUCTIVITY DETECTOR
MODULEWARE REV
Help M essag e
Figure 2-2. Power-Up Screen
+2. 34
uS
POWER-UP
.
TOTA L 138.71 uS
RANGE
screen can be
BIO S REV n.nn
SRS
n.nn nnnnnnDX L-AN ID#
50
1000 uS
2-6
LOCAL
H e lp Mess a g e
MAIN
The facilitate viewing from a d istance. Operating parameters (flow rate, method number to run, etc.) are selected here.
screen dis plays status information in enlarged characters to
METHOD 05
Figure 2-3. Main Screen
Doc. 031687-01 04/2000
Page 19
2•Description
To access other screens, press the display the
MENU of SCREENS
Menu
(see Figure 2-4). T o select an opti on, move the cursor to a screen name and press on the keypad and press
Enter
. See Appendix C for a description of each screen.
MENU of SCREENS
MAIN SCREEN
1
DETAIL SCREEN
2
METHOD
3
- - -
4
5 6 7 8
Help Message
Figure 2-4. Menu of Screens
button on the front panel to
Enter
, or enter the screen number
MODULE SETUP ANALOG OUT SETUP TIME FUNCTION IN DIAGNO STIC MENU
Doc. 031687-01 04/2000
2-7
Page 20
CD25 Conductivity Detector

2.2 Electronics Chassis

The electronics chassis is located immediately behind the front door of the CD25 . The chassis contai ns several el ectronics ca rds (printe d circuit boa rds) that ar e used to control the CD25. Various connectors on the cards allow communication between the CD25 and other system modules and accessories. Figure 2-5 identifies the cards and their connectors.
Do not remove any of the electronics card s from the detector. There are no user-serviceable components on the cards. If servicing is required, it must be performed by qualified personnel using appropriate electrostatic discharge (ESD) handling procedures.
 

  
 
 

 
Figure 2-5. CD25 Electronics Chassis Components

 #"
 !


 

 

 

 






 $

 


 !"
 
 












    


2-8
Doc. 031687-01 04/2000
Page 21

2.2.1 Connectors

Recorder (Slot 2)
This connector is typi cally used for a recorder/integrator or diagnostic instruments. For a description of the connector pinouts, see Table F-1 in Appendix F. For connection instructions, see Section B.2.7.
DS3 (Slot 2)
The control cable from the DS3 Detection Stabilizer connects here.
SRS® (Slot 2)
The control cable from the Self-Regener at ing Suppressor® conne ct s her e.
Conductivity Cell (Slot 3)
The conductivity cell cable connects here.
TTL/Relay (Slot 4)
This strip of eight connectors provides an interface with Dionex and non­Dionex modules for TTL and relay control of the detector. See Appendix D for a description of relay and TTL functions and the connections between the CD25 and other modules.
2 Description
60-pin Ribbon Connector (Slot 5)
The 60-pin ribbon cable to the CD25 front panel (display and keypad) connects here.

2.2.2 Cards

Power Supply Card
Provides 45 watts of power for the detector electronics.
Blank Card
Slot 1 contains a blank card that must be present to complete the electromag netic interference (EMI) shielding.
Doc. 031687-01 04/2000
2-9
Page 22
CD25 Conductivity Detector
SCR (Supply Control/Relay) Card
Interfaces to the CPU; th e card contains three functions:
16-bit Recorder Output Digital-to-Analog Converter
electronic switch for selection of full-scale outputs of 0.01, 0.1, and
1.0 V.
SRS Power Supply
50, 100, 300, or 500 mA to the suppressor. An over-voltage detector shuts off the power i f the vol tage ex cee ds 8.5 V. An over- temp eratur e detector shuts off the power if the SRS temperature excee ds 40 either of these events o ccurs, the S CR card sends an SRS Alarm error message to the CPU.
DS3 Power Supply
Stabilizer . While warmi ng or coolin g to a lowe r set poin t, a “BELOW TEMP or ABOVE TEMP message is displayed. Once a set point in the range of 25 to 45 maintains a constant temperature .
SP (Signal Processor) Card
Includes an
Supplies a regulated current (set by the user) of
°
C. If
Supplies heating power to the DS3 Detection
°
C is reached, proportional heat control
2-10
Contains all the analog circuitry needed by the CD25, as well as the digital circuitry required to interface to the CPU. See Appendix E for a list of SP ca rd functions.
Relay/DX-LAN and CPU Cards
The CD25 control Moduleware and BIOS reside on the CPU card. The CPU card provides control and monitoring of other modules. A 60-pin ribbon cable assembly links the logic to the CD25 front panel display a nd keypad .
The Relay I/O card provides two isolated low voltage relay outputs, two TTL outputs, and four TTL inputs.
The cards are installed in slot 5 of the card cage. The Relay card is a half­card which rides piggyback on the CPU card and sits behind slot 4.
The Relay I/O card is short enough to allow th e optional detector interface card (P/N 044196) to be mount ed behi nd it in slot 4. The interface card is required for communica tion bet ween the CD25 an d PeakNet so ft ware via the DX-LAN.
Doc. 031687-01 04/2000
Page 23
2 Description
Power Supply LED
The multicolored LED below the TTL/Relay connections indicates the power supply status:
A green LED indicates normal operation.
A red LED indicates a fault conditi on. When a fault occurs, th e CD25 enters its diagnostic state and no other control is permitted until the problem is corrected. Turning off the power for a few seconds may clear the fault; however, if the red LED is lighted when you turn on the power again, notify Dionex.
Doc. 031687-01 04/2000
2-11
Page 24
CD25 Conductivity Detector

2.3 Conductivity Cell

The flow-through conductivity cell has an active volume of about 1.0 µL. T wo 316 stainless stee l elect rodes are permanentl y sealed into the PEEK cell bo dy. The cell constant has a nominal value of 160 cm sensor located slightly downstream from the electrodes senses the temperature of liquid passing thro ugh the cell. The me asured value is used to provid e temperature compensation.
The advanced geometry of the cell provides several benefits:
Excellent accuracy and linearity over the working range
Efficient sweepout and low volume for low dispersion
Reduced sensitivity to electrode surface conditions
Low electrode mass
Effective temperature compensation
You can order the conductivity cell installed in a DS3 Detection Stabilizer (P/N 044130) for temperature control, or installed in an electrical shield (P/N 044132) which provides no temperature control. Bec ause th e DS3 maintai ns a constant temperature, thus reducing the effects of variations in laboratory temperature, the DS3 is recommended for conductivity detection.
-1
and is calibrated electronically. A
2-12
Temperature Control and Compensation
Temperature directly affects the conductivity of a solution. As conductivity increases, the effect of temperature changes becomes more pronounced. For example, building temperature control systems can cause a regular oscillation in the baseline. This, in turn, can affect the reproducibility of an analysis.
In ion chromatography, suppressing eluent conductivity minimizes the effect of temperature variation. Temperature compensation further improves baseline stability. When the conductivity cell is housed in a DS3 Detection Stabilizer, the DS3 enhances the ability of these techniques to reduce temperature effects below the detection limit.
Temperature compensation also ensures that th ere will be no major chang e in the baseline or peak heights, should it be necessary to change the DS3 operating set
°
point. Readings will be normalized to 25
C.
Doc. 031687-01 04/2000
Page 25

2.3.1 DS3 Detection Stabilizer

The DS3 is a temperature-controll ed chamber c onsisting of a cast aluminum base and cover enclosed in insulating foam. The chamber houses both the conductivity cell and the eluent heat exchanger. Figure 2-6 identifies the major components of the DS3.
The DS3 provides the following benefits:
Conductivity measurements that are nearly impervious to laboratory temperature variation
Very low dispersion in the eluent heat exchanger
Excellent peak height reproducibility
Remote mo unting after either the column or suppressor
User-settable temperature
Power input to a pair of trans istors on op posite sid es of the DS3 heat s it to
°
a set temperature from 25 outlet senses the eluent temperature. The CD25 circuitry compares this temperature with the user-selected temperature and adjusts the heat input in real time to hold the temperature within a few millidegrees.
C to 45 °C. A sensor near the heat exchanger
2 Description
The DS3 is sealed with an O-ring to trap eluent that may leak from the cell. If 5 mL of liquid acc umulates, a thermistor s ensor becomes immersed and signals a leak to the CPU. Any additional leakage will be discharged via the spill/overflow line. A second thermistor, above the discharge level, acts as a temperature re ference for the leak sensor.
Doc. 031687-01 04/2000
2-13
Page 26
CD25 Conductivity Detector
To Detector Electronic s
Conductivity Cell
Ground Wire
Ce ll Inlet
Figure 2-6. DS3 Detection Stabilizer

2.3.2 Shielded Cell

A conductivity cell installed in an electrical shield (P/N 044132) is available. The shield is a stainless steel housing that protects the cell from electromagnetic interference (EMI), but does not provide temperature control.
DS3 Inle t
DS3 Outlet
Sp ill/Overflo w Line
DS3 Enclosure
Keyhole Mounting Slots
2-14
Doc. 031687-01 04/2000
Page 27

2.4 Rear Panel

The CD25 rear panel (see Figure B-1) contains fuses, connectors for line power, and a connection for the optional DX-LAN interface.
DX-LAN Connection (Optional)
When you order the DX-LAN network, a DX-LAN connector is factory-installed in the upper left corner of the CD25 r ear panel (see Figure B-1). See Appendix B for DX-LAN interface connection instructions.
External Connection Access
Connections to the front of the electronics chassis, such as TTL and relay cables, are usually rout ed to t he back of the d etector through t he cab le chase in the bo ttom of the electroni cs chass is. The y may also be pass ed throug h slots at the front o f the detector. Cables exit the CD25 through an opening in the rear panel (see Figure B-1).
2 Description
Doc. 031687-01 04/2000
2-15
Page 28
CD25 Conductivity Detector

2.5 Functional Description

2.5.1 Operating and Control Modes

The operating mode determines
how
the CD25 receives operating
commands:
In Local mode, the CD25 receives commands from the front control panel buttons and screens.
In Locked Remote mode, PeakNet 6 software sends commands from the host computer via the DX-LAN interface.
The control mode determines
In Direct control, the CD25 executes commands immediately.
In Method control, the CD25 executes commands according to the
when
operating commands are executed.
timed steps in a method. The method is programmed from the CD25 front panel.
The table below summarizes the various operating and control mode
MAIN
configurations. Select the modes from the Section C.1.2),
DETAIL
screen (see Section C.1.3), or chromatography
screen (see
software.
Operating/Control Mode Detector Operation
Local/Direct Control Commands are entered from the CD25 front control
panel and executed immediately after being entered.
2-16
Local/Method Commands are entered from the CD25 front control
panel and executed by running a method programmed from the front panel.
Locked Remote/Direct Control
Commands are sent from PeakNet 6 and executed immediately when received.
Doc. 031687-01 04/2000
Page 29

2.5.2 Local and Remote Modes

Local Mode
When the CD25 is powered up, it is a lways in Loc al mode. In Loc al mode the detector accepts operating commands from two sources:
Direct input from the front panel keypad and screens. All operating functions are available with direct input.
TTL inputs from a remote controller (for example, a Dionex gradient pump module or an integrator). TTL signals can be used to offset the recorder, run a method, turn the SRS off and on, send a mark to the recorder, and increase the recorder range.
Locked Remote Mode
In Locked Remote mode, the CD25 accepts operating commands from a Dionex data station via the DX-LAN.
In Locked Remote mode, all operating changes from the front panel are disabled. Selecting the Connect command from PeakNet 6 immediately selects the Locked Remote mode. To return the CD25 to Local mode, select the Disconnect command, or turn off the CD25 power.
2 Description

2.5.3 Method Control

In Method control, commands are executed according to the time-based steps specified in a metho d. Methods are crea ted, edite d, and saved on the
METHOD
instructions.
METHOD EDIT SAVE TO
TEMP COMP
TIME
INIT
0.00
2.00
Help M essag e
Doc. 031687-01 04/2000
screen (see Figure 2-7). See Section 3.3 for programming
RUN
RANGE
200 uS
v
1.6 OFFSET MARK
**
Figure 2-7. Method Screen
3333
SRS 100DS3 TEMP 40
TTL RLY
1 0
1
25
212 000
1
2-17
Page 30
CD25 Conductivity Detector
The following summarizes basic information about using methods:
The CD25 can run under method control while you are entering or editing any method (including the one that is currently running).
When saving changes to the currently running method, or switching to a different method, the method cloc k continue s running unaf fected. Only those p arameter changes which affect the method after the current time will be implemented in the current run.
The CD25 can store up to 100 separate methods (0 through 99) in memory. The actual number, which depends on the size of each method and the amount of available memory, is typically less than this.
Methods are retained in memory even after the CD25 is powered down.
Each method can have a maximum of 32 time-based steps. Step 1 always starts at
0.0.
INIT
(initial conditions). Step 2 always starts at
TIME
=
2-18
Doc. 031687-01 04/2000
Page 31

3 Operation and Maintenance

3.1 Getting Ready to Run

NOTE The CD25 is designed for use with IC (ion
chromatography) and HPLC (high-performance liquid chromatography) applications and should not be used for any other purpose. If there is a question regarding appropriate usage, contact Dionex before proceeding.
After installing the CD25 Conductivity Detector, or after the power has been off for some time, use the following check list to ready the detector for operation.
Verify that all cables are correctly connected to the CD25.
Verify that the CD25 power cord is plugged into the main power.
Press the power switch actuator on the CD25 front panel (see Figure 2-1) to turn on the power. The diagnostic tests begin running. If the tests are successfully completed, the
MAIN
screen appears after a few seconds. If one or more tests fails, the
DIAGNOSTIC TEST
POWER-UP
screen appears, inst ead.
screen displays briefly and then some
If the conductivity cell is in a DS3 Detec t ion Stabiliz er, set the DS3 temperature from the guidelines:
If the DS3 is installed in a non-temperature controlled module, for example, an LC10 Chromatography Organizer or LC20 Chromatography Enclosure, select a temperature at least 5 °C above the highest expected ambient temperature surrounding the DS3.
If the DS3 is installed in an LC30 Chromatography Oven, select a temperature at least 5 °C above th e oven te mperatu re. Do not set t he oven temperature above 40 °C (104 °F).
Turn on the DS3 power. The DS3 will begin warming up at about 1 °C/minute. Baseline conducti vit y should stabilize once the DS3 reaches the selected temperature.
Doc. 031687-01 04/2000
DETAIL
screen (see Section C.1.3), according to these
3-1
Page 32
CD25 Conductivity Detector
As soon as the proper current is determined and eluent is flowing through the Self-Regenerating Suppressor (SRS) at the correct rate, turn on the SRS. (The SRS power is always off when the CD25 is powered up.) After start-up, the usual direction of drift is downward as SRS efficiency improves.
While waiting for acceptable drift, you may want to select a lower sensitivity. Set the offset to 50%.
When you are ready to start a run, select the desired sensitivity and offset (if you have not already do ne so). Pr ess
Offset
before injection and du ri ng a run,
also, if nec essary.

3.2 Running Under Direct Control (Local Mode)

In the Direct cont rol operat ing mod e, commands a re ca rrie d out immediat ely a fter they are entered. All detector pa ramete rs remai n in ef fect unti l new comma nds are entered. Because ther e ar e no ti me- base d st eps, the method clock is not used. The
Hold/Run
and
Reset
buttons are not operable in Direct control.
Select Direct control from the
DIRECT CNTRL
If
is displayed, this mode is already selected and no action is
necessary.
METHOD
If
Select
or
is displayed, move the cursor to
button to toggle to
button to activate the selection.
MAIN
or
DIRECT CNTRL
DETAIL
screen, as follows:
METHOD
. Press
and press the
Enter
or a cursor arrow
Select
3-2
Doc. 031687-01 04/2000
Page 33
3 Operation and Maintenance

3.3 Running Under Method Contr ol (Local Mode)

In the Method control operating mode, the CD25 is controlled by a series of programmed timed events, known as a even after the detector power is turned off.

3.3.1 Running a Method

method
. Methods are retained in memory
1. Go to the
CNTRL
to
2. In the
METHOD
Enter
or a cursor arrow button.
NOTE The method can also be selected from the
MAIN
METHOD
DETAIL
or
screen. If necessary, toggle from
.
field, enter the desired method number and press
METHOD
the
screen. Enter the method number in
RUN
field and press Enter.
DIRECT
3. If the method clock is already running, the method will start as soon as you enter the method number. If the clock is in Hold, press
Hold/Run
to start the method.
4. The elapsed time on the method clock when the method begins determines at which step the method begi ns runni ng:
If the method clock is at running using the
INIT
INIT
or time zero, the method begins
condition parameters, f ollowed by the t ime zero step. The remaining steps are executed according to their programmed times.
If the method clock is greater than zero, the method begins running using the paramete rs speci fi ed in the step for the elapsed time. To start the method at the
Reset
.
INIT
conditions instead, press

3.3.2 Changing the Running Method

To switch from the method currently running to a different method, enter the new method number in the
Enter
. The new method will begin running, using the parameters specified in the step for the current elapsed time. If you prefer to start the method at the
Doc. 031687-01 04/2000
INIT
conditions, press
RUN
field on the
Reset.
METHOD
screen and press
3-3
Page 34
CD25 Conductivity Detector

3.3.3 Creating a New Method

1. Go to the
METHOD
screen. In the
number from 0 through 99 and press
EDIT
field, enter an
Enter
unused
method
or a cursor arrow button.
This displays a blank method on the screen. The first step of every meth od is an in itia l condi tions step wi th
TIME
the
TIME
field. The second step is always a time step with 0.00 in the
field. You cannot delete these steps, although you can change
their parameters.
2. Enter the parameters for the initial conditions and time 0.00 steps.
NOTE The
METHOD EDIT SAVE TO
TEMP COMP
TIME
field is the only field in each method step that must have an entered value. Leaving any other field blank indicates that the value selected for that parameter in the preceding step is still in effect.
RUN
3333
1.6
25
SRS 100DS 3 TE M P 40
TTL RLY
TIME
INIT
0.00
2.00
RANGE
200 uS
v
OFFSE T M ARK
**
1
212
0
000
1
1
Help M essage
INIT
in
3-4
Figure 3-1. Method Screen
3. To create a new method step, move the cursor to a blank
TIME
field,
enter the time (in minutes) for the action to be performed, and press
Enter
or a cursor arrow button. Enter a value for each step parameter, or leave a field blank if you want the value selected in the preceding step to remain in effect.
4. Repeat Step 3 for each additional step. Up to 30 steps can be added after the time 0.00 step.
Sometimes a method contains more steps than can be seen on the
v
screen at one time. If there is a small bottom of the screen, move the cursor
next to the time entry at the
down
to view additional steps. If there is a caret (^) next to the top tim e entry, move the cursor up to view additional steps.
Doc. 031687-01 04/2000
Page 35
3 Operation and Maintenance
5. To save the new method, move the cursor to the the number that appears in the

3.3.4 Editi n g an Ex is ting Method

NOTE Once you sav e chan ges to a m eth od, t here i s no way
to recall the original method. To make experimental changes while retainin g th e orig inal me thod , be s u r e to save the new method, or a copy of the original method, under a different method number.
Y ou can modify an existi ng method by changing, ad ding, or deletin g steps and/or parameters. If the method you are editing is currently running, the changes are stored in memory and implemented when you save the method.
To edit an existing method, go to the number in the
Follow the editing instructions in the appropriate section below. When you finish, save t he cha nges t o the curr ent me thod nu mber or s elect a ne w number.
Changing Method Parameters
EDIT
field, and press
EDIT
field, and press
METHOD
Enter
screen, enter the method
or a cursor arrow button.
SAVE TO
Enter
field, enter
.
Move the cursor to the desired field and enter a new value, using the CD25 front panel buttons. Press editing change.
Adding a Method Step
There are two ways to add a step to an existing method:
Doc. 031687-01 04/2000
Enter
or a cursor arrow button after each
Move the cursor on the
METHOD
screen to any
time and pa rameters for the new step, and press
TIME
field. Enter the
Enter
or a cursor arrow button. If neces sary , the step will automatical ly be moved to th e correct chronological point in the method.
Move the cursor on the preceding the intended location of the new step. Press
METHOD
screen to the line immediately
Insert
to insert a blank line below the cursor location . Enter the time and parameters for the new step, and the n press
Enter
or a cursor arrow button.
3-5
Page 36
CD25 Conductivity Detector
Deleting a Method Step
Delete
METHOD
twice.
METHOD
screen to t he time of the step to be
screen to the
EDIT
Move the cursor on the deleted and press
Deleting an Entire Method
Move the cursor o n the twice.
Saving a Modified Method
replace
To of the original method in the
retain
To
the original me th od wi t h a m odi fi ed version, ente r the number
SAVE TO
field and press
the original method an d sa ve t he modi fi ed version elsewhere in
memory, enter an unused method number in the
Enter
.
field and press
Enter
.
SAVE TO
field and press
Delete
3-6
Doc. 031687-01 04/2000
Page 37
3 Operation and Maintenance

3.4 Optimizing Temperature Compensation

The CD25 built-in temperature compensation stabilizes conductivity readings by correcting for changes in ambient temperature that occur during a run. For more information about temperature control and compensation, see Section 2.3.

3.4.1 With a DS3 and Conductivity Cell

If the ambient temperature exceeds 45 °C, the DS3 may be permanently damaged. If the DS3 is installed in an LC30 Chromatography Oven, do not se t the ov en temp erat ure abo v e 40 °C (104 °F).
Housing the cell in a DS3 Detection S tabilizer ensures that there is no more than a minor temperature varia tion in liquid re aching th e cell. Thus ,
TEMP COMP
the
setting on the
Many users are able to keep their systems at a single operating temperature. For optimal accuracy, calibrate the cell at this temperature, using the proper temperature coefficient setting. If you later reset the temperature, the CD25 temperature compensation will normalize conductivity measurements to 25 °C (77 °F) to prevent a major upset in system calibration. If you change the DS3 set point, recalibrate the cell.
DETAIL
screen can remain at 1.7% per ºC.
If temperature-induced baseline cycling occurs, it is probably caused by another component of the chromatography system. If the variation increases as the eluent reservoir empties, move the reservoir to a more temperature-stable environment and/or wrap the reservoir in thermal insulation.
Doc. 031687-01 04/2000
3-7
Page 38
CD25 Conductivity Detector

3.4.2 With a Shielded Conductivity Cell

When the conductivity cell is not inside a DS3 Detection Stabilizer, actual conductivity drifts up and down with fluctuations in laboratory temperature. This is especially noticeable when the thermostat cycles on and off in laboratories with very high air turnover rates and controlled temperature. Selecting the proper temperature compensation factor will minimize the effect of temperature fluctuations.
Start by setting
TEMP COMP
baseline variation of the same period as the laboratory cooling or heating occurs, increase or decrease the temperature compensation setting. Continue adjusting it until you find the optimal setting. This is typically between 1.5% to 2% per ºC, suppressed or non-suppressed.
If you notice a slowly increasing or decreasing baseline drift in a laboratory with out a te mperature control system, l ook for a cor responding temperature drift. If there is a correlation, adjust the temperature compensation setting until you find the optimal setting.

3.5 Routine Maintenance

This section describes routine maintenance procedures that can be performed by the user. Any other maintenance procedures must be performed by qualified Dionex personnel.
NOTE The CD25 electronic components are not customer-
serviceable. Repair of electronic components must be performed by Dionex.
Periodically check liquid line conne ctions to the cell (inside the chromatography module) for leaks, and clean up any spills.
on the
DETAIL
screen to 1.7%. If a sinusoidal
3-8
Doc. 031687-01 04/2000
Page 39
This chapter is a guide to troubleshooting problems that may occur while operating the CD25 Conduc ti vi ty De te ctor. Turn to the section that b est des cr ibe s the operating problem. There, the possible causes of the problem are listed in order of probability, along with the recommended courses of action. For additional help, refer to Appendix C for instructions on running the CD25 diagnostics program.
If you are unable to eliminate a problem, contact Dionex. In the U.S., call 1-800­346-6390 and select the Technical Support option. Outside the U.S., call the nearest Dionex office.

4.1 No Detector Response

Cell is off

4 Troubleshooting

Turn on the cell (from the
Analog output range set too high; although the display indicates a
response, no recorder response observed
Select a more sensitive analog output range.
Wrong full-scale output (or no full-scal e output) selected
Select 0.01, 0.10, or 1 volt full-scale.
No flow from pump
Check the pressure reading on the pump to verify that the pump is on.
Detector o f fset out of range
Offset
Press
on the CD25 front panel.
MAIN
or
DETAIL
screen).
Doc. 031687-01 04/2000
4-1
Page 40
CD25 Conductivity Detector

4.2 Low Detector Output

Analog output range set too high; although the display indicates a
response, no recorder response observed
Select a more sensitive analog output range.
Insufficient sample injected
Increase the injection size or concentration.
Cell out of ca libration
Recalibrate the cell (see Section 5.4).

4.3 High Detector Output

Auto offset not activated recently
Offset
Press
Background not suppressed by SRS
Check the SRS regenerant out line for bubbles; if there are no bubbles, the suppressor may be contaminated. Refer to the SRS manual for troubleshooting guidanc e.
on the CD25 front panel before making an injection.

4.4 Noisy or Drifting Baseline

Flow system leak ahead of cell; erratic baseline
Check all fittings and liquid lines for leaks. Tighten or, if necessary, replace all liquid line connections. If the connections are made with ferrule fittings,
Installation of Dionex Ferrule Fittings
4-2
first refer to requirements.
Pump not properly primed
Prime the pump as instructed in the pump manual.
Rapid changes in ambient temperature
Redirect heating and air conditioning vents away from the cell. Purchase a cell in a DS3 Detection Stabilizer (P/N 044130). When using a shielded cell, install the cell in a chromatography oven.
for tightening
Doc. 031687-01 04/2000
Page 41
4 Troubleshooting
Insufficient system equilibration following any changes to operating
parameters; especially apparent when operating at high sensitivities
Allow longer system equilibration before beginning operation.
Air trapped in cell; excessive regular pulses in baseline
Remove the trapped air (see Section 5.2). To prevent air fro m becoming trapped in the cell in the future, increase backpressure on the cell by reducing the inner diameter of the tubing installed after the cell and before the SRS.
Inappropriate SRS operating conditions
Refer to the SRS manual for the correct operating conditions.
Temperature compensation setting not optimized
Optimize the selected setting (see Section 3.4).
DS3 above or below set point
See Section 4.6.
Doc. 031687-01 04/2000
4-3
Page 42
CD25 Conductivity Detector

4.5 Liquid Leaks from the DS3

When a leak is detected, the DS3 leak sensor signals the CPU. If more than 5 mL of liquid accumu lat es i n t h e DS3, the excess will be dr ained via the spill overflow line (see Figure 2-6).
Loose or defective fittings, or overpressurization caused by a restriction
Check the waste line for blockage; a trapped particle can plug the line, causing a restri ction and/or lea k. If necessar y , c lear the waste line by rever sing the direction of flow.
Make sure the plumbing downst r eam f rom t he DS3 is clear; a blockage in the plumbing may overpressurize the DS3, causing it to leak.
Make sure the downstream backpressure coils are appropriate for the operating flow rate (see Section B.2.4).
Follow the steps below to disassemble the DS3 and inspect it for the source of the leak. Test and dry the DS3 before reassembly.
1. Turn of f the CD25 power.
2. Disconnect the DS3 cables.
3. Disconnect the DS3 inlet and outlet lines. Do not misplace the ferrule fittings at the end of the tubing (see Figure 4-1).
4. Remove the DS3 by lifting it upward and then pulling it away from its mounting location. Place the DS3 on the workbench.
5. Open both latches on the DS3 and carefully lift off the top half of the cover, along with its insulating foam insert, exposing an aluminum box (see Figure 4-1).
6. Remove the box from the bottom half of the cover.
7. Remove the four Phillips screws securing the top of the box to its bottom plate. Carefully separ at e t he t wo par ts , expos in g the cel l ( see Figure 4-2). Lay the top aside, being careful not to pull or stress the group of wires that connect the two parts.
8. Check the inlet and outlet cell fittings for leaks. Tighten or replace if necessary.
9. Dry the DS3, test for leaks, and reassemble. Make sure the foam insert is adjusted evenly aroun d the vari ous DS3 component s, with no pin ching or
4-4
Doc. 031687-01 04/2000
Page 43
Bo tto m Cover
4 Troubleshooting
folding. Check that the top and bottom inserts meet on all sides before latching the cover.
Ins u la t in g
Cables
Foam Ins e rt
Aluminum Box
Inle t
Outlet
Ph illips Screws (4)
Sp ill Ov e rf lo w Line
Figure 4-1. DS3 Exploded View
Inle t
Grounding Strap
Outlet
Cell
Outlet Fitting
Inle t Fitting
Figure 4-2. DS3 Interior Components
Top Cover
Ph illi p s Screws
Doc. 031687-01 04/2000
4-5
Page 44
CD25 Conductivity Detector
Liquid seeping from around cell cables
The cell is inoperable; return it to Dionex for repair or exchange.

4.6 DS3 Temperature Inaccurate

DS3 SET POINT XX C READY message displays continuously
The control sensor may be faulty. Remove the DS3 cover and measure the temperature of the housing, using any surface thermometer. The cell sensor can be read on the detecto r and should be wi thin 1 or 2 de gre es of the DS3 set point tempe r ature.
DS3 SET POINT XX C UNDER TMP or DS3 SET POINT XX C
OVER TMP message displays continuously
Verify that the set temperature is at least 5 °C above the external DS3 temperature. Allow 30 to 60 minutes for the DS3 to heat or cool.
At high flow rates and temperature settings far above ambient, the DS3 requires more time to heat. In extreme cases, such as a very cold room and a high DS3 set point, the DS3 may not be able to reach the set point temperature.
4-6
DS3 does not heat
Remove the cover and inspect the DS3 for broken or shorted wires or for moisture bridging the control thermistor. If a wire is broken or shorted, replace the wire or call Dionex for assistance. If a leak has caused a short, fix the leak and dry the control sensor.
DS3 SET POINT XX C READY message displays intermittently
Make sure that the heater transistors are snug and the control sensor has not been pulled out of the heat exchanger plate.
Doc. 031687-01 04/2000
Page 45

4.7 Conductivity Inaccurate

4 Troubleshooting
Cell constant reported on
calibration is not 130 to 190 µS
The DS3 has not reached its set point temperature. Set the intended operating temperature and wait until 5 minutes after the NOT READY message is displayed, or until the reading is stable.
Make sure the selected temperature compensation setting is 1.7%/°C. Except at 25 °C, an incorrect temperature co mpensation set ting can caus e an incorrec t reading.
The 1 mM KCl calibration solution is spoiled, or there was a mixing error. Remake the solution.
The electrodes are fouled by grease, precipitate, etc. Flush with a suitable solvent, such as acid, and then flush with 1 mM KCl until the reading is stable.
The cause is internal leaks, broken or shorted cell or sensor wires. Check the sensor by comparing the cell temperature readout with the actual DS3 temperature. If necessary, return the cell to Dionex for repair.
disassemble the cell; this will break the seal and v oid your warranty. Cell temperature readout deviates by more than 2°C from DS3 set
temperature
This indicates a serious cell or DS3 sensor problem. Call Dionex for assistance.
CONDUCTIVITY CALIBRATION
screen after cell
Do not

4.8 Faulty DX-LAN Communication

DX-LAN interface incorrectly install ed
See Section B.2.2 for 10BASE-T connection inst ructions or Section B.2.3 for BNC connection instructions.
Doc. 031687-01 04/2000
4-7
Page 46
CD25 Conductivity Detector

4.9 Diagnostics

NOTE The CD25 electronic components are not customer-
serviceable. Before running a diagnostic test, use the troubleshooting information in earlier sections of Chapter 4 to isolate problems that are unrelated to the CD25 electronics.
The CD25 Moduleware includes several diagnostic tests of the electronics. To access these, select the diagnostic screens are illustrated and described in Appendix C.
DIAGNOSTIC MENU
from the
DIAG NOSTIC MENU
MENU of SCREENS
. The
POWER-UP SCREEN
1
ELAPSED TIME
2
ANALOG STATUS
3
-
4
DX LAN STATUS
5
KEYBOARD TEST
Help M essag e
Figure 4-3. Diagnostic Menu Screen
6
DIAG NOSTIC TESTS LEAK CAL & STATUS
7 8
SIGNAL STATISTICS
9
CALIBRATE CD CELL
10
4-8
Doc. 031687-01 04/2000
Page 47

5 Service

This chapter describes routine service procedures that the user can perform. All other procedures must be performed by Dionex personnel.
NOTE The CD25 electronics components are not customer-
serviceable. Repair of electronics components must be performed by Dionex.
The CPU card contains a lithium battery. If the CPU card is replaced, dispose of the used battery according to the manufacturer's instructions.
Before replacing any parts, refer to th e troubleshooting information in Ch apter 4 to isolate the cause of the problem. To contact Dionex in the U.S., call 1-800-346­6390 and select the Technical Support option. Outside the U.S., call the nearest Dionex office.
Substituting non-Dionex parts may impair CD25 performance, thereby voiding the product warranty. Refer to the warranty statement in the Dionex Terms and Conditions for more information.
Doc. 031687-01 04/2000
5-1
Page 48
CD25 Conductivity Detector

5.1 Eliminating Liquid Leaks

The PEEK version of the CD25 is plumbed with 1.60-mm (1/16-in) OD PEEK tubing, Dionex ferrule fittings (P/N 043276), and 10-32 fitting bolts (P/N
043275). For tightening requirements, refer to
Fittings
.
Installation of Dionex Ferrule

5.2 Removing Trapped Air from the Cell

Air bubbles in the cell can cause regular pulsations of the baseline, random noise, and low r eadings. Ai r may have bee n intr oduced in the columns d uring installation, or ma y result from outgassing of the mobi le phase. Connecting a backpressure line to the cell applies enough backpressure to shrink bubbles, thereby allow ing them to pass more easily through the cell.
1. Connect a piece of 0.25-mm (0.010-in) ID tubing (P/N 042690), with fittings on both ends, to the cell outlet. Make the line 1 m long for a flow rate of
1.0 mL/min, 2 m for 0.5 mL/min, etc.
2. Use tubing with an inn er diameter of at least 1mm (0.04 in) t he rest of the way to the waste containe r. Use a union (P/N 042627) to connect the backpres sur e line to the w aste line.
Make sure the a dded backpres sure does not exceed 0.20 MPa (30 psi). Avoid large increases in flow rate tha t would result in more than 1.4 MPa (200 psi) being applied to the cell.

5.3 Replacing the DS3 Cell

Follow the steps below to disassemble the DS3 and replace the cell. After replacing the cell you must recalibrate it (see Section 5.4).
1. Turn of f the CD25 power.
2. Disconnect the DS3 cables.
3. Disconnect the DS3 inlet and outlet lines. Do not misplace the ferrule fittings at the end of the tubing (see Figure 5-1).
4. Remove the DS3 by lifting it upward and then pulling it away from its mounting location. Place the DS3 on the workbench.
5-2
Doc. 031687-01 04/2000
Page 49
5 Service
5. Open both latches on the DS3 and carefully lift off the top half of the cover, along with its insulating foam insert, exposing an aluminum box (see Figure 4-1).
6. Remove the box from the bottom half of the cover.
7. Remove the four Phil lips screws securing the top of the box to its bottom plate. Carefully sep arate the two par ts, exp osing t he cel l (see Figure 5-1). Lay the top aside, bein g careful not to pull or stres s the grou p of wires t hat connec t the two parts .
8. Disconnect the grounding strap. Remove the two cell mounting screws.
9. Rotate the cell inlet tube fitting counterclockwise. Let the cell body back away from the fitting until the fitting threads are fully disengaged.
the ferrule
.
10. Check that the end of the inner Tefzel tubing is flush with the end of the plastic sleeve. If nec essary, trim the sleeve slightly to prevent dis persion. Push the tubing into the new cell until it bottoms out in the hole; then, hold the tubing in place while tightening the nut.
11. After testing for liquid leaks, dry the DS3 and reassemble. Make sure the foam insert is adjusted evenly around the various components, with no pinching or folding. Before latching the cover, make sure the top and bottom inserts meet on all sides.
Do not lose
12. Reconnect the DS3 cables and turn on the CD25 power. Calibrate the cell constant (s ee Section 5.4).
Doc. 031687-01 04/2000
Inle t
Grounding Strap
Figure 5-1. DS3 Interior Components
Outlet
Cell
Outlet Fitting
Inle t Fitting
Ph illi p s Screws
5-3
Page 50
CD25 Conductivity Detector

5.4 Calibrating the Cell

Every conductivity cell is calibr ated be fore it is shi pped from the fa ctory. The cell calibration constant is recorded on a tag attached to the cell cable, and is also stored in permanent detector memory for use when calculating the measured conductivity.
The calibration const ant normal ly rema ins uncha nged, unl ess the cell i s damaged. To check the value entered in memory, select
DIAGNOSTIC MENU
cell calibration constant shown on the screen does not match the value recorded on the tag, do
Enter a new value, regardless of the composition of the solution in the cell.
Calibrate the cell at 147.00 µS/cm. This automatically enters a new value for the cell calibration constant into CD25 memory.
To calibrate the cell, follow the steps below.
1. Disconnect the pump output line from the injection valve.
2. Connect the pump output directly to the inlet of the DS3 or the shielded cell.
to display the
one
of the following:
CALIBRATE CD CELL
CALIBRATE CONDUCTIVITY CELL
from the
screen. If the
3. Pump 0.001 M KCl calibration solution through the cell. Conductivity is slightly flow-rate sensitive , so select t he flow rate used in the majority of your applications.
4. Set
TEMP COMP
on the
5. If using a DS3, set
DETAIL
DS3 SETPOINT
screen to 1.7%.
on the
DETAIL
screen to the intended
operating point. Wait until the DS3 READY message appears.
6. Wait until the conductivity reading st abili zes ( within 0 .1 µS), and then, on the
CALIBRATE CONDUCTIVITY CELL
screen, select
CAL
and press
Enter
to
calibrate the cell.
7. After calibration, t he conducti vity rea ding should b e exactly 14 7.00 µS/cm. A new value for the cell constant will be shown on the
CONDUCTIVITY CELL
screen; thi s value will al so be entered in memory.
CALIBRATE
8. Flush the KCl solution from the system by pumping deionized water through the DS3 or cell lines. When the conductivity reading drops to near zero, stop the pump.
9. Disconnect the pump from the DS3 or shielded cell.
5-4
Doc. 031687-01 04/2000
Page 51
5 Service
10. Reconnect the pump to the chromatography module.
11. Reconnect the liquid line from the suppressor outlet to the cell inlet.
12. Reset
TEMP COMP
to the optima l value for the eluent.
13. S et the pump to a flow rate that is safe for the system in use.
Doc. 031687-01 04/2000
5-5
Page 52
CD25 Conductivity Detector

5.5 Replacing the Main Power Fuses

The fuse holder is in the main power receptacle on the CD25 rear panel.
1. Turn of f the main power.
HIGH VOLTAGEDisconnect the main power cord from its source and also from the rear panel of the CD25.
2. A recessed lock is located on each side of the fuse holder (see Figure 5-2). Using a small screwdriver, push each lock toward the center to rel eas e it . The fuse holder pops out approximately slightly when the locks release. When both locks are released, pull the fuse holder straight out of its compartment.
3. The holder contains two fuses. Replace both with new 3.15 amp fast-blow IEC127 fuses (P/N 954745). Dionex recommends replacing both fuses even though only one is open.
4. Reinsert the fuse holder into its compartment. The fuse holder is keyed to fit only in its proper orientation. Apply sufficient pressure evenly against the holder to engage the two locks. The holder is flush against the panel when both locks are engaged.
5-6
5. Reconnect the main power cord and turn on the power.
Locking Spring
Fuse Holder
Insert screwdriver
and twist to release
(each side)
M a in Po we r Receptacle
Figure 5-2. Main Power Fuse Holder
Fuses (2)
Key
Fuse Holder (side vie w )
Doc. 031687-01 04/2000
Page 53

5.6 Physical

A Specifications
Dimensions
Weight
Decibel Level

A.1 Environmental

Operating
Temperature
Operating
Humidity

A.2 Electrical

Main Power
Fuses
22.5 cm W x 17.0 cm H x 42.0 cm D (8.8 in W x 6.6 in H x 16.4 in D) 6 cm (2.4 in) clearance required behind the detector
8.2 kg (18 lb) 50 db (A WEIGHING setting)
4 ºC to 40 ºC (40 ºF to 104 ºF)
5 to 95% relative humidity, non-condensing
85 to 270 Vac, 47/63 Hz; 40 W Max, 25 W typical. The CD25 power supply is auto-sensing and requires no voltage adjustment.
Two 3.15 amp fast-blow IEC127 fuses (P/N 954745)
Analog Ou tput
Doc. 031687-01 04/2000
User-selectable full-scale output of 10, 100, or 1000 mV
A-1
Page 54
CD25 Conductivity Detector

A.3 Display and Keypad

Display
Keypad
Liquid crystal display with adjustable backlighting 26-button keypad for entering commands and numerical values
for screen parameters

A.4 Detector

Range
Temperature
Compensation
Cell Drive
Local Operation
Remote
Operation
DX-LAN
Operation
(Optional)
0.01 µS to 3000 µS, full-scale
0.0 to 3.0% per ºC
Variable 8 kHz square wave Front panel controls and display status of all functions Control of four functions via TTL or Relay contacts
All functions can be controlled by PeakNet software on a PC connected to the CD25 via the DX-LAN interface

A.5 Conductivity Cell

A-2
Cell Body
Active Volume
Maximum
Pressure
Electrodes
PEEK
1.0 µL
2.0 MPa (300 psi)
316 stainless steel
Doc. 031687-01 04/2000
Page 55

A.6 SRS Power Supply

A Specifications
Supply Current
Over-Voltage
Alarm
Over-
Temperature
Alarm
50, 100, 300, 500 mA at 1.5 to 7.5 V
8.5 V
40 ºC (104 ºF)

A.7 DS3 Detection Stabilizer (Optional)

Operating
Temperature
Warm-up Time
Temperature
Alarms
25 ºC to 40 ºC (50 ºF to 104 ºF)
10 minutes (typical)
Not Ready
temperature
message displayed when under or over the set
Doc. 031687-01 04/2000
A-3
Page 56
CD25 Conductivity Detector
A-4
Doc. 031687-01 04/2000
Page 57

B.1 Facility Requirements

Make sure the CD25 installation site meets the el ectrical and environmental specifications listed in Appendix A.
Install the CD25 on a sturdy table or workbench, at a height that ensures convenient viewing of the front panel display.
Allow at least 6 cm (2.4 in) clear ance behind the CD25 for power con nections and ventilation.

B.2 Installation

During installation, or whenever it is necessary to move the detector, lift it o nly from the bottom or side. Lift ing the CD25 by the front panel door will damage the door hinges.

B.2.1 Power Connection

The power to the CD25 can be controlled from the main power switch on either the CD25 or the LC30 Chromatography Oven (i f prese nt). In eith er case, no adjustment is required to select the line voltage.
B Installation
For power on/of f contro l from the CD25, conn ect the modu lar power cord (IEC 320 C13), from the main power receptacle on the rear panel (see Figure B-1) to a grounded, single-phase, power source.
The power supply cord is used as the main disconnect device. Ensure that the socket-outlet is located/installed near the module and is easily accessible.
For power on/off control from the LC30 Chromatography Oven, connect the IEC jumper power cord (P/N 960748) provided in the LC30 Ship Kit from the main power recepta cle on the CD25 rear pane l to one of the four IEC auxiliary receptacles on the rear panel of the LC30. Le ave the CD25 power switch on continuously and use the LC30 main power switch to
Doc. 031687-01 04/2000
B-1
Page 58
CD25 Conductivity Detector
turn the CD25 on and off. Other modules in the system can be connected to the LC30 in the same way, allowing the LC30 to function as the main power source for the system.
SHOCK HAZARDTo avoid electrical shock, a grounded receptacle must be used. Do not operate or connect to AC power mains without an earthed ground connection.
Operation at AC input levels outside of the specified operating voltage range may da mage the CD25.
DX-LAN
Connector
SAFETY WARNING
SERIAL NUMBER
DX-LAN
AND
LABEL
B-2
Chase
(TTL & Relay
Lines)
Fuse Holder
M a in P ow e r Receptacle
Figure B-1. CD25 Rear Panel
Doc. 031687-01 04/2000
Page 59
B Installation

B.2.2 DX-LAN Interface: 10BASE-T Connections (Optional)

NOTE There are two types of DX-LAN connectors. Before
proceeding, check the CD25 rear panel. If a DX­LAN 10BASE-T RJ-45 (telephone-style) connector is installed, follow the interface instructions in Section B.2.2. However, if a DX-LAN 10BASE-2 BNC connector is installed, follow the interface instructions in Section B.2.3.
In order to communicate with a host computer running Dionex chromatography software, the CD25 must contain a detector interface card (P/N 056798) and an unshielded twisted-pair DX-LAN cable (P/N 960281) must be connected from the RJ-45 (10BASE-T) connector on the rear panel to a “combo” 10BASE-T Ethernet hub (P/N 056909).
Installing or Replacing the Detector Interface Card
NOTE If the DX-LAN option was not installed at the
factory, order the detector interface card kit (P/N 057007). The kit includes all the components required for DX-LAN communication.
STATICThe CD25 electronics cannot be serviced by the user. The detector interface card must be installed by qualified personnel. Standard anti-static procedures must be observed when installing the interface card or handling the CPU card.
To prevent damage to the CD25, turn off the main power before installing the detector interface card. After confirming that the LED on the CPU card is off (not green or red), unplug the power. Do not rely on the front panel power switch.
1. To replace an existing detector interface card, remove the DX-LAN
2. Remove any TTL/Relay plugs from the connectors at slot 4 of the
Doc. 031687-01 04/2000
cable from the 10BASE-T connector on the CD25 rear panel (see Figure B-1).
electronics chassis (see Figure 2-5).
B-3
Page 60
CD25 Conductivity Detector
3. Disconnect the 60-pin ribbon cable from the inside of the front panel by opening the ejector latches on the connector.
4. Using a screwdriver as a lever, open the white ejector latch at the bottom of the CPU card. Remov e the CPU card, cable, and Rel ay card as a single unit.
5. Remove the detector interface card (if present) from slot 4 of the electronics chassis.
6. Insert the new detector interface card (P/N 056798) into slot 4: a. Slide the card to the rear. b. Verify that the 10BASE-T connector is aligned with the opening
at the rear and that the card is aligned with the card connector.
c. Press firmly on the card until it mat es f ul ly wi th the connector on
the rear panel. Pull on the card to verify that it cannot move.
7. Reinstall the CPU/Relay card. Press firmly until the card is inserted into the connector on the rear panel.
8. Reconnect the ribbon cable to the 60-pin connector on the inside of the front panel. The header and connector are key-polarized near the center . Th e ejector latches should be pa rtiall y open to ac cept the c able connector.
B-4
9. Install the 10BASE-T DX-LA N cable (see the following sec t ion).
Installing the 10BASE-T DX-LAN Cable
1. Install the “combo” 10BASE-T Ethernet hub (P/N 056909) on a workbench or on the wall. For installation instructions and site requirements, refer to the installation guide shipped with the hu b.
2. Plug the 10BASE-T DX-LAN cable (P/N 960281) into a 10BASE-T port on the front panel of the hub.
The 10BASE-T DX-LAN cable is a Category 5 unshielded twisted-pair cable. Do not substitute a cable of an inferior grade. Failure to use the correct cable will cause the detector to lose communication with the host computer.
3. Connect the other end of the cable into the 10BASE-T DX-LAN connector on the CD25 rear panel (see Figure B-1).
Doc. 031687-01 04/2000
Page 61
B Installation
4. Connect a 10BASE-T cable (P/N 960281) from a 10BASE-T port on the hub to the 10BASE-T port on the host computer’s in ternal DX- LAN card. If the connection is via port 8 on the hub, set the Normal/Uplink push button to Normal.
NOTE Installation instructions for the host computer’s
internal DX-LAN card are in
Peak N et 6 System
When using 10BASE-T cabling, you must use a hub. If you simply plug the 10BASE-T cable from the detector into the connector on the host computer’s DX-LAN card, the connection will either not work or will be unreliable.
Network Upgrades
(Document No. 031631).
If you upgrade the network from a BNC connector to a 10BASE-T connector, turn of f the computer po wer briefly and then turn it on aga in to allow the host computers internal Ethernet DX-LAN card to detect the 10BASE-T con nec tor. The Ethernet card searches for th e t ype of i nstal led cabling only at power-up.
Installing the Dionex
Cascading Hubs
Cascading, or connecting hubs together through their 10BASE-T ports, increases the number of ports or the number of users supported on the network. For instructions, refer to the installation guide for the hub.
Doc. 031687-01 04/2000
B-5
Page 62
CD25 Conductivity Detector

B.2.3 DX-LAN Inte rface: BNC Connections (Optional)

NOTE There are two types of DX-LAN connectors. Before
proceeding, check the CD25 rear panel. If a DX­LAN 10BASE-2 BNC connector is installed, follow the interface instructions in Section B.2.3. However, if a DX-LAN 10BASE-T RJ-45 (telephone-style) connector is installed, go on to Section B.2.2.
In order to communicate with a host computer running Dionex chromatography software, the CD25 must contain a detector interface card (P/N 046341) and a coaxial DX-LAN cable (P/N 960405) must be connected to the BNC tee connector (P/N 921914) on the rear panel. From the CD25, connect the DX-LAN cable to one of the following:
Another instrument equipped with a BNC connector,
A combo 10BASE-T Ethernet hub (P/N 056909). The hub has one
BNC port and eight 10BASE-T (RJ-45) ports.
Installing or Replacing the Detector Interface Card
or
B-6
NOTE If the DX-LAN option was not installed at the
factory, order the detector interface card kit (P/N 044195). The kit includes the components required for DX-LAN communication.
STATICThe CD25 electronics cannot be serviced by the user. The detector interface card must be installed by qualified personnel. Standard anti-static procedures must be observed when installing the interface card or handling the CPU card.
To prevent damage to the CD25, turn off the main power befor e installing the detector interface card. After confirming that the LED on the CPU card is off (not green or red), unplug the power. Do not rely on the front panel power switch.
1. To replace an existing detector interface card, remove the hex nut securing the DX-LAN BNC connector to the CD25 rear panel (see Figure B-1).
Doc. 031687-01 04/2000
Page 63
B Installation
2. Remove any TTL/Relay plugs from the connectors at slot 4 of the electronics chassis (see Figure 2-5).
3. Disconnect the 60-pin ribbon cable from the inside of the front panel by opening the ejector latches on the connector.
4. Using a screwdriver as a lever, open the white ejector latch at the bottom of the CPU card. Remov e the CPU card, cable, and Rel ay card as a single unit.
5. Remove the detector interface card (if present) from slot 4 of the electronics chassis.
6. Insert the new detector interface card (P/N 046341) into slot 4: a. Slide the card to the rear. b. Verify that the BNC connector is aligned with the opening at the
rear and that the card is aligned with the card connector.
c. Press firmly on the card until it mates fully with the connector on
the rear panel. Pull on the card to verify that it cannot move.
7. Reinstall the CPU/Relay card. Press firmly until the card is inserted into the connector on the rear panel.
8. Reconnect the ribbon cable to the 60-pin connector on the inside of
9. Install the hex nut, BNC tee connector, and coaxial DX-LAN cable
Installing the BNC Tee Connector and Coaxial DX-LAN Cable
1. Install the BNC tee connector (P/N 921914):
2. Install the coaxial DX-LAN cable (P/N 960405):
Doc. 031687-01 04/2000
the front panel. The header and connector are key-polarized near the center. The ejector latches must be partially open to accept the cable connector.
(see the following section).
a. Note the two small locking pins on either side of the DX-LAN
BNC connector on the rear panel (see Figure B-1).
b. Push the tee connector onto the BNC connector on the rear panel
and twist until the locking pins are fully engaged in the slots on the tee connector.
c. Pull firmly on the tee connector to verify that it cannot move.
B-7
Page 64
CD25 Conductivity Detector
a. Push the metal sleeve on the end of the DX-LAN cable onto one
port of the BNC tee connector.
b. Twist the metal sleeve until the locking pins on the tee are fully
engaged in the slots on the cable's metal sleeve.
c. Pull the end of the cable to verify that it cannot move.
The DX-LAN cable (P/N 960405) is a 50 ohm coaxial impedance cable. (Fifty ohm cables are imprinted with “RG-58U.”) Do not substitute cables. Failure to use the correct cable or to lock it into place on the BNC connector will cause the CD25 to lose communication with the host computer.
3. If the CD25 is the last module in the network to be connected, install a terminator resistor plug (P/N 921034) on the remaining port of the BNC tee connector. (These terminator resistor p lug s a re shipped with Dionex chromatography software.)
If the CD25 is not the last module in the network, connect the cable from the next module to the BNC tee.
4. Plug the free end of the coaxial DX-LAN cable into one of these:
The BNC connector on another instrument,
The BNC port on the rear panel of the “combo” 10BASE-T
or
Ethernet hub (P/N 056909).
If the connection terminates at the hub, install a BNC terminator (included with the hub) on the BNC tee connector.
50 Ohm terminator resistor plugs must be installed at each end of the DX-LAN network. Before beginning operation, verify that both ends of the DX-LAN have terminator resistor plugs installed.
B-8
Doc. 031687-01 04/2000
Page 65

B.2.4 DS3 Detection Stabilizer Installation

Install the DS3 with the conductivity cell (P/N 044130) in a chromatography module—for example, an LC20 Chromatography Enclosureor on a laboratory workbench. The instructions for installing the DS3 vary somewhat, depending on the type of enclosure. See Section B.2.6 for in stallation and plumbing instructions.
NOTE To install a DS3 or shielded cell in an AS50
autosampler chromatography compartment or oven, refer to the AS50 operators manual.

B.2.5 Shielded Conductivity Cell Installation

The installation instructions for the shielded conductivity cell (P/N
044132) are the same as for the DS3, exc ept that the shie lded cell does not have a heater power cord.

B.2.6 DS3 or Shielded Cell Plumbing

B Installation
The DS3 or shielded cell can be plumbed for three operating modes:
AutoSuppression Recycle mode (SRS required), shown in Figure B-2.
AutoSuppr ession with External Regen erant mode (SRS required ), shown in Figure B-3. Also plumb the system in this way for the AutoSuppression Chemical Suppressor and MMS Pressurized Regen System configurations.
Nonsuppressed mode, shown in Figure B-4. Note that conductivity measurements made when no suppressor is in-line may show noticeable baseline drift.
Doc. 031687-01 04/2000
B-9
Page 66
CD25 Conductivity Detector
The CD25 Ship Kit includes certain items required when installing the SRS:
SRS gas separator waste tube (P/N 045460)
Backpressure coil for 4 mm SRS (P/N 045877)
SRS-to-CD25/ED50 control cable (P/N 046346)
NOTE The instructions here do not replace the inst ructions
in your SRS manual. For detailed SRS installatio n instructions, and for the initial start-up procedure that must be performed before you begin routine operation, please refer to the SRS manual.
Backpressure Requirements
Detector cells req uire enoug h backpr essure t o prevent mobile p hase in th e cell from degassing due to abrupt volume changes between the small inner diameter of the connecting tubing and the larger volume of the cell. Degassing creates bubbles in the cell and disrupts detector responsiveness.
Before plumbing a system f or th e Aut oSupp ression mode, check the SRS manual for backpressure requirements.
NOTE The backpressure generated by the DS3, cell, and
backpressure tubing is applied to the SRS. For example, with 1 meter of 0.25-mm (0.01-in) ID tubing at a flow rate of 1.0 mL/min, the backpressure is about 200 kPa (30 psi).
B-10
Doc. 031687-01 04/2000
Page 67
B Installation
For backpressure requirements for the Nonsuppressed mode, see the following table. The minimum backpressure should be 0.21 to 0.34 MPa (30 to 50 psi) per minute.
Flow Rate (mL/min) Tubing ID Coil Length Number of
Coils
0.5 to 1.5 0.25-cm (0.010-in)
1.5 to 3.0 0.25-cm (0.010-in)
0.12 to 0.37 0.12-cm (0.005-in)
0.37 to 0.75 0.12-cm (0.005-in)
Table B-1. Coils for Backpressure Requirements (Nonsuppressed Mode)
0.76 M (2.5 feet)
0.76 M (2.5 feet)
0.30 M (1.0 feet)
0.30 M (1.0 feet)
2
1
2
1
Doc. 031687-01 04/2000
B-11
Page 68
CD25 Conductivity Detector
Injection Valve
Guard Column
Analytical Column
2 mm ID, 10-32, 0.005" ID PEEK Liquid Line 4 mm ID, 10-32, 0.010" ID PEEK Liquid Line 4 mm ID, 1/4-28, 0.012" ID Tefzel Liquid Line
REGEN OUT
ELUENT IN
10-32
1/4-28
Waste
ANION SELF-RE GENER ATING SU P PR E S S OR 4 mm P/N 043189 S /N XXX X
1/4-28
1/4-28
ELUENT OUT
10-32
REGE N IN
1/4-28
Gas
Separator
Waste
Tube
1/4-28
10-32
10-32 or 1/4-28
10-32
10-32
10-32
Coil
#2
10-32 or
1/4-28
Coil
#1
Cell
10-32 or 1/4-28
Signal to Detector
B-12
Figure B-2. AutoSuppression Recycle Mode
Doc. 031687-01 04/2000
Page 69
Inje ctio n Valv e
Guard Colum n
Analytical Column
2 m m ID, 10-32, 0.005" ID PEE K L iquid Line 4 mm ID , 10-32, 0.010" ID PEE K Liquid Line 4 mm ID , 1/4-28, 0.012" ID Tefzel Liquid Line
0.02 0" I D Tefzel
REGEN OUT
ELUENT IN
10-32
1/4-28
1/4-28
1/4-28
Waste
ANION SELF-REGENERATING SUPPRESSOR 4 mm P/N XXXXXX S/N XXXX
Separator
ELUENT OUT
10-32
REGEN IN
1/4-28
Gas
Waste
Tube
D eionized Water Source
10-32 or 1/4-28
B Installation
To Waste
1/4-28
10-32
Coil
#2
10-32
10-32
Coil
#1
10-32
10-32 or
Cell
S ig n al to C o n du c tiv ity D etector
10-32 or 1/4-28
1/4-2 8
Figure B-3. AutoSuppression Mode, External Regenerant
Doc. 031687-01 04/2000
B-13
Page 70
CD25 Conductivity Detector
Injection Valve
Analytical Column
2 mm ID, 10-32, 0.005" ID P E EK L iquid Line 4 mm ID, 10-32, 0.010" ID PEEK Liquid Line 4 mm ID, 1/4-28, 0.012" ID Tefzel Liquid Line
Guard Column
DS3
Figure B-4. Nonsuppressed Mode
Backpressure Coils
Waste
To
B-14
Doc. 031687-01 04/2000
Page 71
DS3 Detection Stabilizer (or Shielded Conductivity Cell)
Alternate DS3 or Shielded Conductivity Cell Location
B Installation
SRS
Doc. 031687-01 04/2000
FRONT
Figure B-5. Typical DS3 Installation in LC20 or LC30
B-15
Page 72
CD25 Conductivity Detector
Installing a DS3 or Shielded Cell in an LC20, LC25, or LC30
1. Position the keyhole slots on the rear of the DS3 or cell over the shoulder washers on the LC20, LC25, or LC30 component mounting panel, then pull the DS3 or cell down into position.
2. Route the DS3 or cell cables out the hole located at the back of the LC20, LC25, or LC30 and into the cable chase at the base of the detector. Next, route the cables through the cable chas e to the fr ont of the detector. (You can also route the cables into the detector through the slots on the sides of the detecto r.)
3. Plug the connectors on the cables into the appropriate jacks on the SCR card. (To identify the card location and connector positions, see the label on the inside of the CD25 front door or to Figure 2-5 in this manual.)
4. Connect the eluent outlet of the column or suppressor to the DS3 or cell inlet, using 0.25-mm (0.010-in) ID tubing (P/N 042690), 10-32 fittings (P/N 043275), and ferrules (P/N 043276).
5. Connect the outlet of the DS3 or cell to either the SRS REGEN IN port (for AutoSuppression Recycle mode) or to the backpressure line leading to waste (for AutoSuppression with External Regenerant mode or Nonsuppressed mode).
B-16
6. Direct the spill/overflow tubing to the spill tray at the bottom front of the LC20, LC25, or LC30.
Installing a DS3 or Shielded Cell in an LC10
1. If installing a DS3, cut the spill/overflow tubing within 2.0 cm (3/4 in) of the DS3 enclosure.
2. Position the keyhole slots on the rear of the DS3 or cell over the shoulder washers located on the LC10. Push the DS3 or cell down to secure it in place.
3. Route both DS3 or cell cables through the hole located at the base of the detector cable chase. Next, route the cables through the cable chase to the front of the detector. (You can also route the cables into the detector using the slots on the sides of the detector.)
Doc. 031687-01 04/2000
Page 73
4. Plug the connectors into the appropriate jacks on the SCR card. (To identify the card l ocation and co nnector pos itions, se e the label on the inside of the CD25 front door or to Figure 2-5 in this manual.)
5. Connect the eluent outlet of the column or suppressor to the DS3 or cell inlet, using 0.25-mm (0.010-in) ID PEEK tubing (P/N 042690), 10-32 fittings (P/N 043275), and ferrules (P/N 043276). See Table B­1 for backpressure requirements.
6. Connect the outlet of the DS3 or cell (the 10-32 coupler) to the SRS REGEN IN port (for AutoSuppression Recycle mode) or to the backpressure line leading to waste (for AutoSuppression with External Regenerant mode or Nonsuppressed mode).

B.2.7 Recorder/Diagnostic Connection

Connecting a strip chart recorder or integrator to the CD25 lets you record or monitor several parameters besides the cell analog output. For the recorder/diagnostic signal cable pinouts, see Table E-1 in Appendix E.
Use the twisted black and red wires (P/N 043598) provided in the CD25 Ship Kit to connect a recorder or integrator to the SCR card in the electronics chassis. Strip the ends of the wires and insert into the plugs; the signal wire (red) goes on the top of each plug and the ground wire (black) goes on the bottom. Tighten the locking screws with the screwdriver (P/N 046985) provided in the CD25 Ship Kit.
B Installation
Doc. 031687-01 04/2000
B-17
Page 74
CD25 Conductivity Detector

B.3 Automatic SRS Power Control (Optional)

TTL connections can be used to automatically switch off the power to a Self­Regenerating Suppressor (SRS) when the flow from the pump stops.
1. On the
PUMP OPTIONS
screen, set
TTL2 OUTPUT USAGE
to
0 Flow
. (For more details, see the operator's manual for the GP50 Gradient Pump or IP25 Isocratic Pump.)
2. On the CD25
TTL FUNCTION IN
screen, assign
TTL3
SRS OFF/ON
to
.
3. Locate the twisted black and red wires (P/N 043598) and two green TTL connector plugs (P/N 921019) provided in the CD25 Ship Kit.
4. Strip the ends of the wires and insert them into the plu gs, wit h t he si gna l wir e (red) on top of each plug and the ground wire (black) on the bottom. Tighten the locking screws with the screwdriver (P/N 046985) provided in the CD25 Ship Kit.
5. Plug one end of the cable into TTL2 OUT on the pump (see Figure B-6).
6. Route the cable through the upper chase of the pump and the CD25, and then plug the free end of the cable into TTL3 IN on the CD25.
7. Turn of f the flow (f rom the pump) and confirm tha t the detect or automat ically turns off the SRS. First, however, expect a 5-minute delay from the ti me the CD25 receives the signal until the power to the SR S is turned off; this delay allows momentary flow interruptions without disturbing the SRS.
B-18
Doc. 031687-01 04/2000
Page 75
B Installation
Doc. 031687-01 04/2000
Figure B-6. TTL Connections for SRS Power Control
B-19
Page 76
CD25 Conductivity Detector
B-20
Doc. 031687-01 04/2000
Page 77
C User Interface
This appendix illustrates and describes all of the screens available for display on the front panel of the CD25 Conductivity Detector. There are two categories of screens: operational and diagnostic.
Operational
CD25 operation, and to select default parameters for the detector.
Diagnostic
including some screens that enable you to calibrate various CD25 functions.
Operational Screens
123 5678
MAIN
SCREEN
screens enable you to create, edit, and run methods that control
screens provide access to diagnostic information and tests,
MENU of
SCREENS
DETAIL
SCREEN
METHOD
MODULE
SETUP
ANALOG OUT SETUP
FUNCTION IN
TIME
DIAGNOSTIC
MENU
Diagnostic Screens
123 5678
POWER-UP
SCREEN
ELAPSED
Doc. 031687-01 04/2000
TIME
METHOD
EXTENSION
4
ANALOG STATUS
DX LAN
STATUS
KEYBOARD
TEST
DIAGNOSTIC TESTS
Figure C-1. CD25 Menu Structure
LEAK CAL & STATUS
SIGNAL STATISTICS
9
CALIBRATE CD CELL
C-1
Page 78
CD25 Conductivity Detector

C.1 Operational Screens

C.1.1 Menu of Screens

MENU of SCREENS
The
MENU of SCREENS
1
MAIN SCREEN
2
DETAIL SCREEN
3
METHOD
4
- - -
Help Message
Figure C-2. Menu of Screens
There are two ways to select a screen from this men u:
provides top-level access to the menu structure.
MOD ULE SETUP
5
ANALOG OUT SETUP
6 7
TIM E FUNCTION IN
8
DIA GNOSTIC MENU
Use the cursor arrow buttons to move the cursor to the field containing the screen number and press
Press the number b utton on t he front pa nel keypa d that co rresp onds to
Enter
.
the screen.
To display a brief description of each menu option, press the
Help
button.
C-2
Doc. 031687-01 04/2000
Page 79

C.1.2 Main Screen

MAIN
The primary functions in large characters to make viewing easier from a distance.
screen displays the measured conductivity (siemens) and other
C User Interface
TOTA L 13 8 .7 1 u S
+2. 34
LOCAL
Help M essage
Screen Field Description TOTAL
SRS
RANGE
LOCAL/ REMOTE
Displays total conductivity (without an offset). Selects the amount of current sent to the suppressor . Current selection
depends on the eluent concentration, flow rate, etc. The discrete settings are Off, 50, 100, 300, and 500 mA.
Sets the analog output scale factor. Select a range between 0.01 and 3000 µS.
Local mode is used for front panel control. The CD25 will automatically switch to Locked Remote mode when a Connect command is given from PeakNet 6. Locked Remote mode prevents operation from the keypad. The Locked Remote mode can be canceled only by a Disconnect command from PeakNet 6 or by turning off the CD25 and then turning it back on. does not use the regular (unlocked) Remote mode.
METHOD 05
Figure C-3. Main Screen
uS
SRS
RANGE
50
1000 uS
Note:
PeakNet 6
DIRECT CONTROL/ METHOD
Doc. 031687-01 04/2000
Select
Press either Method operating modes.
button to toggle between the Direct Control and
C-3
Page 80
CD25 Conductivity Detector

C.1.3 Detail Screen

DETAIL
The
screen includes all the f ields cont ained on the
well as the TTL and Relay fields and fields to control detection. Refer to
MAIN
the
screen for a description of fields that are common to both
screens.
D E TA IL SCREEN OUTPUT
128.2
O F F SET 58 .7nAnA
TOTAL
184.9 nA
pH 12.2
POTENTIAL
Help Mess a g e
Figure C-4. Detail Screen
Screen Field Description TEMP COMP
DS3 SETPOINT TTL1
TTL2
Sets the temperature compensation factor. The range is 0 to 3%.
1.7% is appropriate for most eluents. Sets the temperature of the DS3 Detection Stabilizer. Provides TTL control of accessories. In a method, these values
occur according to the method timing. In Direct control, select off (0) or on (1).
DC AMPERO METRY
CELL
RANGE
REF
ON 200 nA Ag
+0.80V
123.45 MINLOC AL ME THOD 05
MAIN
TTL1 TTL2
RLY1 RLY200
screen, as
1 1
C-4
RLY1 RLY2
Provides relay contact closure control of accessories. In a method, these values occur according to the method timing. In Direct co ntrol, select off (0) or on (1).
Doc. 031687-01 04/2000
Page 81

C.1.4 Method

A method consists of a series of timed steps. Each step has a set of parameters associated with it.
Initial conditions are applied when a method is invoked. A lower case next to the la st step on the d isplay i ndicates that t he method contains more steps.
Blank fields denote no change from the previous step. The blanks the fie ld.
C User Interface
v
Delete
button
NOTE
TEMP COMP, DS3 TEMP
time-programmable.
METHOD EDIT SAVE TO
TEMP COMP
TIME
RANGE
INIT
1.6
200 uS
0.00 v
2.00
Help M essag e
Figure C-5. Method Screen
Screen Field Description EDIT
SAVE TO
RUN
Specifies the method number (1 through 99) to edit.
Note:
Method number 0 is reserved for methods downloaded from
PeakNet 5.2 (or later), for remote control of the CD25. Specifies the method number (1 through 99) to save the current
method to. Specifies the method number (1 through 99) to run. The
button controls the running of the method.
SRS CURRENT
, and
3333
TTL RLY
OFFSET MARK
1 0
**
1
are not
RUN
25
SRS 100DS3 TEMP 40
212 000
1
Hold/Run
TEMP COMP SET TEMP SRS
Doc. 031687-01 04/2000
Sets the Temperature Compensation factor. Sets the temperature for the DS3 Detection Stabilizer. Sets the Self-Regenerating Suppressor current.
C-5
Page 82
CD25 Conductivity Detector
Screen Field Description TIME
RANGE OFFSET
MARK
TTL RLY
Specifies the start time for each step. Times are displayed chronologically.
Sets the analog output range. Stores the offset value. The baseline is set by subtracting the offset
measured when this step is executed from all subsequent measurements. An asterisk (*) indicates that OFFSET will occur at this time.
Sends a positive pulse to the analog output (recorder) as an event marker. An asterisk (*) indicates that a MARK will occur at this time.
Sets TTL1 and TTL2 to off (0) or on (1). Sets RLY1 and RLY2 to off (0) or on (1).
C-6
Doc. 031687-01 04/2000
Page 83

C.1.5 Module Setup

The backlight intensity, key beep, and error beep are configured from the
MODULE SETUP
C User Interface
screen.
MODULE SETUP
DISPLAY PANEL BACKLIGHT:
KEY ACTUATION SOUND:
ENTRY ERROR SOUND:
Help Message
Figure C-6. Module Setup Screen
Screen Field Description DISPLAY PANEL
BACKLIGHT KEY ACTUATION
SOUND ENTRY ERROR
SOUND
Sets the display panel backlight to
OFF
.
Toggles the keypad touch sound. When this option is selected , the detector sounds a beep when a button is pressed.
Toggles the error sound. When this option is selected, the detector sounds a beep when an invalid entry is made.
MEDIUM ON ON
LOW, MEDIUM, HIGH
, or
Doc. 031687-01 04/2000
C-7
Page 84
CD25 Conductivity Detector

C.1.6 Analog Out Setup

ANALOG OUT SETUP
The analog output, such as for a recorder or oscilloscope.
ANALOG O UT SETUP
OUTPUT:
ZERO POSITION :
VOLTS FULL SCALE:
RISE TIME:
POLARITY:
Help Message
Figure C-7. Analog Out Setup Screen
Screen Field Description
screen contains parameters for setting the
OFFSET
10
% FULL SCALE V
1.0
1.0
sec
+
OUTPUT
ZERO POSITION VOLTS FULL
SCALE RISE TIME
POLARITY
Sets the analog output to one of the following:
OFFSET TOTAL
uses the offset level value.
sets the detector output to the total cell conductivity, charge
or current, disabling the offset function.
ZERO
sets the detector output to zero volts.
FULL SCALE
sets the output to the full-scale setting. Sets the analog (recorder) offset level (0 to 100%). Sets the full-scale voltage for a signal equal to the range setting. The
discrete settings are 1.0, 0.1, and 0.01 volts. Sets the ou tput filter r ise time to 0.05, 0.1, 0.2, 0.5, 1.0, 2. 0, 5.0, or
10.0 seconds. Rise time is a measure of how quickly the detector goes from 10% to 90% of step input changes. A digital filter is used with zero undershoot or overshoot for minimal peak distortion.
Determines the output voltage polarity. The output is usually set to positive polarity. Negative polarity is used to reverse peaks from indirect detection.
C-8
Doc. 031687-01 04/2000
Page 85

C.1.7 Time Function In

This screen displays the detector functions that can be controlled with TTL input from another device and lets you assign any four of the functions to the fo ur TTL inputs. Th e signal the detector will respond to. The signal type can be changed to match the signal of the controlling device. See Section D.3 for details.
Select
Use the TTL1 through TTL4 to any four of the functions.
buttons to select the TTL input signal mode and to assign
C User Interface
MODE
field selects th e type of input
TIME FUN CTION IN
OFFSET:
HOLD/RUN:
SRS
METHOD NUMBER INCR:
Help M essage
ON/OFF:
RANGEx10:
Figure C-8. Time Function In Screen
MODE: NORMAL EDGE
MARK:
TTL TTL TTL TTL TTL TTL
1 2 3
DECR: -
4
-
-
TTL
Doc. 031687-01 04/2000
C-9
Page 86
CD25 Conductivity Detector

C.2 Diagnostic Screens

To access the CD25 diagnostics, select the
SCREENS
returns you to the
Menu
. While a diagnostic screen is displayed, pressing the
DIAGNOSTIC MENU
returns you to the
MENU of SCREENS
. From the

C.2.1 Diagnostic Menu

Figure C-9 shows the explained in detail in the following sections.
POWER-UP SCREEN
1
ELAPSED TIME
2
ANALOG STATUS
3
-
DX LAN STATUS
4 5
KEYBOARD TEST
Help M essag e
Figure C-9. Diagnostic Menu Screen
There are two ways to select a screen from this men u:
DIAGNOSTIC MENU
DIAG NOSTIC MENU
6 7 8 9
10
DIAGNOSTIC MENU
DIAGNOSTIC MENU
.
. The nine menu options are
DIAG NOSTIC TESTS LEAK CAL & STATUS
SIGNAL STATISTIC S CALIBRATE CD CELL
from the
Menu
button
, pressing
MENU of
C-10
Use the cursor arrow buttons to move the cursor to the field containing the screen number and press
Press the number b utton on t he front pa nel keypa d that co rresp onds to
Enter
.
the screen.
To display a brief description of each menu option, press the
Help
Doc. 031687-01 04/2000
button.
Page 87

C.2.2 Power-Up Screen

POWER-UP
The and BIOS code. It also displays the identification number of the optional DX-LAN interface, if co nnected.
screen displays the revi sion nu mbers fo r the Module war e
CD25 CONDUCTIVITY DETECTOR
C User Interface
Help M essag e
MODULEWARE REV
BIOS REV n.nn
Figure C-10. Power-Up Screen
n.nn nnnnnnDX L-AN ID#
Doc. 031687-01 04/2000
C-11
Page 88
CD25 Conductivity Detector

C.2.3 Elapsed Time

This screen reports for how long various CD25 parameters have been in use. The status of each parameter updates in real time.
MODULE ON:
BACKLIGHT:
SRS ON:
CD CELL ON:
Help M essag e
Figure C-11. Elapsed Time Screen
Screen Field Description
ELAPSED TIME
nnnnnn hours nnnnnn hours nnnnnn hours nnnnnn hours
RESET
* *
MODULE ON
BACKLIGHT
SRS ON
CD CELL ON
RESET
Reports the total time the detector has been powered up in its lifetime.
Reports the total time the LCD backlight has been on in its lifetime.
Reports the total time the SRS has run. Reset this field to zero after changing the SRS.
Reports the total time the conductivity cell has run. Reset thi s field to zero after replacing the cell.
Allows a timer to be reset to zero to restart the counter. Reset the appropriate timer after installing a new component.
C-12
Doc. 031687-01 04/2000
Page 89

C.2.4 Analog Status

This screen reports the status of several analog test points. If a run is in progress, the status updates are from the last idle state. If no run is in progress, the values are in real time.
SRS CONNECTED: SRS OVER TEMP: SRS OVER VOLT:
THERMISTO R TEM PERATURE: nn.n Degrees
Help M essage
Figure C-12. Analog Status Screen
Screen Field Description
ANALOG STATUS
Y N N
DS3 CONNECTED: DS3 OVER TEMP: DS3 UNDER TEMP:
CD CELL CONNECTED:
C User Interface
Y N N N
SRS CONNECTED
SRS OVER TEMP SRS OVER VOLT DS3 CONNECTED
DS3 OVER TEMP DS3 UNDER
TEMP CD CELL
CONNECTED THERMISTOR
TEMPERATURE
Indicates whether a Self-Regenerating Suppressor is connected to the detector.
Indicates whether the SRS is over the temperature specified. Indicates whether the SRS is over the voltage range specified. Indicates whether a DS3 Detection Stabilizer is co nnected to the
detector. Indicates whether the DS3 is over the temperature specified. Indicates whether the DS3 is under the temperature specified.
Indicates whether a conductivity cell is connected to the detector.
Indicates the temperature recorded from the conductivity cell thermistor.
Doc. 031687-01 04/2000
C-13
Page 90
CD25 Conductivity Detector

C.2.5 DX-LAN Sta tus

The DX-LAN driver monitors several types of erro rs that may occur on the network. The detector reads the DX-LAN error counts and displays them on this screen, alo ng with the DX-LAN addresses. Most errors are caused by a defective cable, missing or incorrec tly terminated cable. The coaxial DX-LAN cable must be RG58U or RG58AU.
UNIQUE DX- LAN ID:
-
DX LAN ADDRESS:
-
DX LAN
STATUS FF FF FF FF FF FF FF FF FF FF
COLLISION:
B U S WRIT E :
UNDERFLOW:
nnn nnn nnn
ALIGNMENT:
OVER FLOW:
Help Message
Figure C-13. DX-LAN Status Screen
Screen Field Description UNIQUE DX-LAN
ID
DX-LAN ADDRESS
COLLISION
BUS WRITE
UNDERFLOW
Displays the three-byte DX-LAN ID programmed into the detector Moduleware (in HEX). The ID is unique to the detector and never changes.
Displays the six-byte DX-LAN address assigned by the PC (in HEX). The first three bytes are the system assignment and the last three bytes are the assignment within the system.
Indicates that 16 unsuccessful transmissions of the same packet occurred, due to collisions.
Indicates that a ready response could not be issued within 2.4 microseconds after the WR signal was asserted. This occurs when the transmit buffer memory is full.
Indicates that data from the transmit section of the hardware buffer memory is not available for serial transmission. The DX­LAN will continue to send out this data frame.
BUS READ:
nnn nnn nnn
CRC:
RUNT:
RETRY:
nnn nnn nnn
C-14
Doc. 031687-01 04/2000
Page 91
Screen Field Description
C User Interface
ALIGNMENT
BUS READ
OVERFLOW
CRC
RUNT
RETRY
Indicates that a packet was received with an alignment error, meaning that there were one to seven extra bits at the end of the packet. This is usually caused by a collision or a faulty transceiver.
Indicates that a ready response could not be issued within 2.4 microseconds after the ready signal was asserted. This occurs when reading an empty buffer.
Indicates that the DX-LAN hardware receive buffer became full and had to reject a packet for lack of space.
Indicates that a packet was received with a CRC error. This usually means that a collision has corrupted the packet.
Indicates that a “runt” packet (one less than 15 bytes in length) was received. This usually occurs after a collision has truncated the original length.
Indicates the number of retries required to transmit the last packet.
Doc. 031687-01 04/2000
C-15
Page 92
CD25 Conductivity Detector

C.2.6 Keyboard Test

You can conduct an interactive test of the CD25 front panel keypad from this screen. Pressing any front panel button changes the corresponding button indicator on the screen to reverse video. This confirms proper operation of that button. When you release the button, the display returns to normal video.
OFFSET MARK INSERT DELETE
Help Message
RUN/HOLD RESET SEL UP SEL DOWN
Figure C-14. Keyboard Test Screen
UP 7 8 9
LEFT RIG HT
DOW N
HELP MENU
456 123
0.E
C-16
Doc. 031687-01 04/2000
Page 93

C.2.7 Diagnostic Tests

These tests verify the integrity of the detector electronics and functions.
DIAG NOSTIC TESTS
CE LL: INT NO ISE TE S T:
C User Interface
RDY
DAC RAMP: SCOPE TEST: RDY
POW ER CPU
PP
RDY
P
Help M essag e
Figure C-15. Diagnostic Tests Screen
Screen Field Description CELL
NOISE TEST
DAC RAMP
Specifies whether the cell to be tested is internal (INT) or external (EXT).
Selects one of the following test modes: RDYThe system is idle and ready to begin a test. CDSelects the conductivity cell for testing. Selects the DAC test mode. The test linear ramps the output of the
selected DA on the SP or SCR card from 0 to full-scale and can be plotted from the recorder output or from PeakNet. The ramp test modes are:
RDYThe system is idle and ready to begin a test. CDOTests the CD Offset DAC (eight iterations, one for each gain;
the eighth iteration is flat on top). Disconnect the conductivity cell before running this test.
PSRS
LAN LEAK
DS3
CELL DRIVE
* *>
SCOPE TEST
Doc. 031687-01 04/2000
RECTests the Recorder Output DAC. Generates test waveforms that can be monitored on test points on the
SCR and SP cards, using an oscilloscope or chart recorder. The test modes are:
RDYThe system is idle and ready to begin a test.
C-17
Page 94
CD25 Conductivity Detector
Screen Field Description
CDFTests the CD Offset DAC output (TP9 on the SP card) and the CD fine filter circuit (TP19 on the SP card). TP9 outputs a square wave with a frequency of 200 ms and an amplitude from 0 to -0.1 V. TP19 outputs a sine wave with a frequency of 200 ms. The peak-to­peak amplitude is 4.40 V.
CDTTests the CD Temperature Compensation DAC (TP26 on the SP card). The output is a square wave in which the bottom and top amplitudes increase and decrease from 11 V to 12 V. Before running this test, connect the DS3 and set the temperature to 45 °C, or connect a dummy thermistor value of 42 kΩ to pins 2 and 3 of the conductivity cell connector.
SCRTests the SCR DAC for the DS3 set point temperature (TP13 on the SCR card) and SRS current set point (TP17 on the SCR card). The output on each test point is a linear ramp that goes from 0 to 10 V in 2.5 seconds. Disconnect the DS3 and SRS before running this test.
The following tests can be run from the
DIAGNOSTIC TESTS
screen. To
begin a test, select the asterisk (*) in the field beside the test name and
Enter
press (pass) or
NOTE Running the following tests disturbs the
POWER CPU
DX-LAN SRS DS3 CELL DRIVE
LEAK
. When the test is complete, the entry field displays either P
F
(fail).
equilibration of the detector as services cycle on and off and the system runs through its test modes. After testing, allow the system to equilibrate before resuming operation.
Checks the +5, ±15, and +24 volt monitor on the Relay card. Checks the CPU internal configuration and the Moduleware
checksum. Checks the DX-LAN hardware configuration and loop-back. Varies the SRS power supply and checks the flags. Varies the DS3 temperature set point and checks the flags. Switches in a resistor in place of the cell to provide a calibrated
signal to read. Checks the detector's leak sensor for a correct, open circuit, or
short circuit condition.
C-18
Doc. 031687-01 04/2000
Page 95

C.2.8 Leak Sensor Calibration and Status

This screen reports the status of leak sensor parameters.
LEAK SENSOR CALIBRATION AND STATUS
C User Interface
MEASURED VALUE:
CURRENT CONDITION:
CALIBRATION VALUE:
LOW LEAK THRE SHOLD :
Help M essage
Figure C-16. Leak Sensor Calibration and Status Screen
Screen Field Description MEASURED
VALUE CURRENT
CONDITION
Reports the measured voltage from the leak sensor.
Reports the current (error) condition of the leak sensor: WET, DRY, or ERR. Error indicates an open or short circuit. To calibrate a leak sensor, select calibration, the field will revert to DRY, unless an error condition exists.
CALIBRATION VALUE
LOW LEAK THRESHOLD
Reports the value saved when the leak sensor was last calibrated.
Reports the threshold value below which a leak is indicated. This is based on the calibration value.
2.48
DRY
2.50
2.70
CAL
and press
Enter
. After
Doc. 031687-01 04/2000
C-19
Page 96
CD25 Conductivity Detector

C.2.9 Signal Statistics

This screen enab les y ou to monit or t he sel ected input to the A/D ci rcuit ry. When the screen is first displayed, the equal to the reported in A/D volts and are updated dynamically.
CELL: NORM
INPUT:
Help Message
Screen Field Description
INPUT
SIGNAL STATISTICS
C FINE
MAX:
MIN:
DURATION:
Figure C-17. Signal Statistics Screen
value, and
n.nnn volts n.nnn volts n.nnn volts
nnn.nn Minutes
DURATION
MAX
and
MIN
status values are
reads 0. Status values are
CELL
INPUT
MAX MIN DURATION
Selects the cell test mode: NORMUses input data from the cell. FIXEDUses a fixed dummy cell for input. The measured data from the A/D circuit: C FINEConductivity cell fine input C COARSE—Conductivity cell coarse input C THERConductivity cell thermistor input LEAKLeak detector input DRIVE—Cell drive circuit input The maximum input data value during the duration of the test. The minimum input data value during the duration of the test. The duration (in minutes) of the test. The test starts when you enter
this screen and terminates when you exit.
C-20
Doc. 031687-01 04/2000
Page 97

C.2.10 Calibrate Conductivity Cell

This scree n allows you t o calibrate the conducti vity cell with 1 mM KCl. See Section 5.4 for detailed instructions.
CALIBRATE CONDUCTIVITY CELL
MEASURED CONDUCTIVITY:
CELL CALIBRATION CONSTANT:
PUMP 1 mM KCl THROUGH CELL AT 1 mL/MIN.
SELECT 'C AL' AND PRESS ENTER RDY
WARNING : THIS W ILL ERASE THE PRIO R
CELL CALIBRATIO N AND STORE A NEW VALUE
Help M essage
Figure C-18. Calibrate Conductivity Cell Screen
Screen Field Description
146.54 u S
C User Interface
160.0
MEASURED CONDUCTIVITY
Calibration
Reports the measured conductivity from the conductivity cell.
CAL
Select the value displayed in the conductivity field. The calibration assumes that uncontaminated 1 mM KCl is flowing through the cell at normal backpressure. The new value replaces the previous calibration value.
and press
Enter
to calibrate the conductivity cell to
Doc. 031687-01 04/2000
C-21
Page 98
CD25 Conductivity Detector
C-22
Doc. 031687-01 04/2000
Page 99
D TTL and Relay Control
The strip of eight 2-pin connectors on the CD25 electronics chassis provides two relay outputs, two TTL outputs, and four TTL inputs (see Figure D-1).
Connect the outputs to the TTL or relay inputs of a Dionex or non-Dionex device to control functions in the connected device. The devices must be compatible with the CD25’s TTL and relay signals.
Connect the inputs to a Dionex or non-Dionex device to control four detector functions.
Connection instructions are in Section D.1.
Doc. 031687-01 04/2000
Figure D-1. TTL Connector Strip
(CD25 Front View without Upper Door)
D-1
Page 100
CD25 Conductivity Detector

D.1 TTL and Relay Connections

The TTL/relay connector strip is located on the electronic chassis behind the upper door (see FigureD-1). Each 2-pin connector includes a signal pin and a ground pin.
1. Twisted pairs of wires (P/N 043598) and two-pin connector plugs (P/N 921019) are provided in the CD25 Ship Kit. Attach a two-pin plug to each end of the twi st ed pai r of wires to be connected. The signal wir e goes on top and the ground wire goes on the bottom of each plug.
2. Connect these plug s to the TTL or relay connec tors on t he CD25 and th e other module(s) as needed for your application. Check the polarity of each connection. Connect signal wires to signal (+) pins and ground wires to ground (-) pins. If necessary, remove wires from the two-pin plugs and reinsert them in the correct positions.
3. Route the wires from the CD25 elec tr onics ch assis through the upper chase to the rear pan el.
NOTE Relay outputs 1 and 2 can be programmed to switch
any low-voltage control (see Section D.2). Switched current must be less than 200 mA and 42 V peak
.
D-2
The relay outputs, TTL inputs, and TTL outputs all use a connector of the same style. Do not inadvertently connect relay loads that include power supplies to the TTL inputs or TTL outputs. Relay loads in excess of 200 mA or with included power supplies over 42 V may damage the rela y driv ers on the CPU module.
Doc. 031687-01 04/2000
Loading...