LABORATORY SCALE
LABORWAAGE
BALANCE POUR LABORATOIRES
BILANCIA DA LABORATORIO
HLD
OPERATING MANUAL
Industrial Weighing Systems
9 Richmond St, Picton, Ontario, Canada, K0K 2T0
www.iwsystems.ca e-mail info@iwsystems.ca
HLD_V1.003_11.06_EN_DE_FR_IT
ENGLISH
INDEX
1. INTRODUCTION ................................................................................................................................................................. 4
2. MAIN TECHNICAL SPECIFICATIONS ............................................................................................................................... 4
3. INSTALLATION ................................................................................................................................................................... 5
3.1 POWER SUPPLY AND START-UP ............................................................................................................................ 5
3.2 BATTERY POWERED ............................................................................................................................................... 5
4. FRONT PANEL KEYS AND INDICATORS ......................................................................................................................... 6
5. SYMBOLS ON THE LCD DISPLAY .................................................................................................................................... 7
6. OPERATION ....................................................................................................................................................................... 7
6.1 AUTOMATIC ZERO UPON START-UP ...................................................................................................................... 7
6.2 SCALE ZERO FUNCTION ......................................................................................................................................... 7
6.3 TARE FUNCTION ...................................................................................................................................................... 7
6.4 OVERLOAD MESSAGE ............................................................................................................................................ 7
6.5 TOTALISATION OF THE WEIGHT ............................................................................................................................ 8
6.5.1 MANUAL OR AUTOMATIC TOTALIZATION .................................................................................................... 8
6.6 COUNTING PIECES .................................................................................................................................................. 8
6.7 SAMPLE WEIGHT PERCENTILE .............................................................................................................................. 9
6.8 UNIT OF MEASURE CONVERSION ......................................................................................................................... 9
6.8.1 SECONDARY UNITS OF MEASURE ............................................................................................................. 9
7. SET-UP ENVIRONMENT .................................................................................................................................................... 9
7.1 SET-UP ENVIRONMENT BLOCK DIAGRAM .......................................................................................................... 10
7.2 F1 Unt CONFIGURATION OF SECONDARY UNITS OF MEASURE ...................................................................... 10
7.3 F2 bL DISPLAY BACKLIGHTING ............................................................................................................................ 10
7.4 F3 CoM TRANSMISSION MODE ............................................................................................................................ 11
7.5 “tEcH” - SCALE CONFIGURATION ......................................................................................................................... 11
8. SERIAL OUTPUT .............................................................................................................................................................. 13
8.1 TRANSMISSION MODES ........................................................................................................................................ 13
9. ERROR MESSAGES ........................................................................................................................................................ 14
RECYCLING INSTRUCTION ................................................................................................................................................ 14
DECLARATION OF CONFORMITY ...................................................................................................................................... 15
WARRANTY .......................................................................................................................................................................... 15
1. INTRODUCTION
with fitted 240 Vac mains power adapter or with fitted battery
30 hours of continuous use.
Calibration fully digital and programmable from the keyboard
LCD backlit with 6 ½ digits 1
+/- 2% of max load capacity.
up: +/- 10% of max load capacity.
This manual was created to help you install and learn all about the functional possibilities of the HLD scale.
WARNINGS
The scales should be sited in a location that will not degrade the accuracy.
Avoid extremes of temperature. Do not place in direct sunlight or near air conditioning vents.
Avoid unsuitable tables. The tables or floor must be rigid and not vibrate. Do not place near vibrating machinery.
Avoid unstable power sources. Do not use near large users of electricity such as welding equipment or large motors.
Avoid high humidity that might cause condensation. Avoid direct contact with water. Do not spray or immerse the scales in
water.
Avoid air movement such as from fans or opening doors. Do not place near open windows.
Keep the scales clean.
Do not stack material on the scales when they are not in use.
2. MAIN TECHNICAL SPECIFICATIONS
dimensions
mm
g
Division
g (*)
Division
g (*)
g +/-
sensitivity
g
recommended
g
(*) Standard and HR division selectable in weighing phase, through the unit of measure conversion function, see section
6.8.
.
3. INSTALLATION
a) Unpack then product.
b) If present (depending on the model), it is necessary to unscrew and remove the locking screw for the transportation
placed under the scale, as shown in the figure below:
c) Level the platform by adjusting the adjustable feet until the bubble is in the centre of the level. The stability of the
platform is very important.
ALL THE CORNERS MUST REST IN A UNIFORM WAY. Carefully check that all feet fully rest on the ground and that the
platform, loaded on the corner, is not unstable (if a corner is not resting on the ground its relative foot is easier to turn).
3.1 POWER SUPPLY AND START-UP
The indicator can be powered with a fitted 240 Vac external power adapter or just with the fitted battery.
NOTE: it is advisable to completely recharge the battery (12 hours) with the first installation of the instrument.
TO POWER the instrument through the 240 Vac mains put the AC/DC power plug in the appropriate socket under the
scale and the power adapter in the 240 Vac mains socket.
TO TURN ON/TURN OFF the instrument press the ON/OFF key.
3.2 BATTERY POWERED
The weighing indicator can be operated from the battery if desired.
When the battery needs charging the
when the symbol is on. The scale will still operate for about 10 minutes after which it will automatically switch off to protect
the battery.
To charge the battery simply plug into the mains power. The scale does not need to be turned on.
The battery should be charged for 12 hours for full capacity.
You can find the
symbol on the display to indicate the status of battery charging. When the scale is plugged into
symbol on the weight display will be empty. The battery should be charged