Dilog Test Equipment DL6799 User Manual

DL6799 - CombiVolt™5
200A TrueRMS Current, Voltage & Continuity Tester
Instruction Manual
Di-LOG
...measurably better
DL6799 OPERATING MANUAL
Certicate of conformity
As per the supplier of the apparatus listed, declare under our sole responsibility.
Di-LOG DL6799 - CombiVolt™ 5
Declaration of Conformity:
To which this declaration relates are in conformity with the relevant clauses of the following standard.
Standards:
EN 61010-1 EN 61010-2-032 EN 61010-2-33 EN 61010-031 EN 61326-1 EN 61326-2-2
Performance:
The instrument operates within specication
when used under the conditions in the above
standards and EMC safety standards.
The product identied above conforms to the requirements of council directive 89/336/EEC and 73/23 EEC.
Di-LOG
...measurably better
DL6799 OPERATING MANUAL
Contents
I. Overview ------------------------------------------- 1 II. DL6799 Box Contents ------------------------- 1 III. Safety Notices ----------------------------------- 2 IV. Electrical Symbols ----------------------------- 4 V. Appearance Overview ------------------------- 5 VI. Function Keys ------------------------------------ 6 VII. Technical Index ---------------------------------- 9
1. General Specication ---------------------------- 9
2. Environmental Limits ---------------------------- 10
3. Electrical Specication -------------------------- 10
VIII. Measurement Operation ---------------------- 13
1. Measuring of AC Current (ACI) --------------- 13
2. Measuring of DC Current (DCI) --------------- 15
3. Measuring of AC Voltage (ACV) -------------- 17
4. Measuring of DC Voltage (DCV) -------------- 19
5. Measurement of Resistance (OHM) --------- 21
6. Continuity Testing (CON) ----------------------- 23
7. Diode Measurement (DIO) --------------------- 25
8. Measuring of Capacitance (CAP) ------------ 27
9. Non-Contact Voltage Detection (NCV) ------ 29
10. Single Pole Detection (SPD) ------------------ 31
IX. Maintenance, Repair & Calibration -------- 33
1. General Maintenance --------------------------- 33
2. The installation or replacement of batteries 33
Di-LOG
...measurably better
DL6799 OPERATING MANUAL
l. Overview
Thank you for purchasing a Di-Log DL6799 current, voltage and continuity tester. The tester has been designed in accordance with the latest international safety standards.
200A AC/DC current 1000 V DC, 750 V AC Continuity test with audible indication.
Warning:
The instruction manual includes relevant safety information and warnings. Please read all relevant contents carefully and strictly observe all warnings and safety notices.
If the equipment is used in a manner not specied by the manufacturer, the protection provided by the equipment may be impaired.
ll. Contents
1. DL6799 CombiVolt 5
2. Carry Case
3. Instruction Manual Please contact your supplier if any of the contents are missing or damaged.
1
Di-LOG
...measurably better
DL6799 OPERATING MANUAL
Ill. Safety Notices
This instrument is to be used only by trained
and competent persons.
The instrument is designed and produced in strict accordance with the safety requirements for Electronic Measuring Apparatus and EN61010-1, EN61010-2- 03 2,
EN61010-2-033, EN61010-031, EN61326-1 and EN61326-2-2 standards, and complies with the safety
standard of double insulation, over-voltage CATII 1000V, CAT III 600V and pollution level 2. If you fail to use the instrument according to relevant operation instructions, the protection provided by the CombiVolt 5 instrument is likely to be weakened or undermined.
CAT II: Applicable to test and measuring circuits
connected directly to live voltage (socket outlets or similar points) of the low voltage mains installation.
CAT Ill: Applicable to test and measuring circuits
connected to the distribution part of the building's low-voltage MAINS installation.
1. Prior to using, please check the CombiVolt 5 and integrated test lead & probe set for any damage or abnormality. If any of the probes insulating layer or lead are obviously damaged, the display screen fails to display or you consider there is a fault with the instrument, please do not use and contact Di-log for further assistance.
2
Di-LOG
...measurably better
DL6799 OPERATING MANUAL
2. It is strictly prohibited to use the CombiVolt 5 without the rear cover or battery cover, otherwise a potential shock hazard may occur.
3. When carrying out any measurement, please
keep ngers behind the nger guards on both the
instrument and the test probes.
4. Before measurement, the instrument must be switched to the correct functional position. It is prohib ited to switch the rotary dial to an alternative position during measurement, so as to prevent the instrument being damaged.
5. Do not apply in excess of 1000V DC/750V AC between the CombiVolt test probes or between line and ground so as to prevent electric shock and instrument damage.
6. When the instrument is used to measure a DC voltage (in excess of) >42V or AC TRMS voltage (in excess of) >30V, it is necessary to operate the instrument carefully as there may be danger of electric shock.
7. Do not measure voltage or current higher than the stated input limits. Before measuring resistance, diode or continuity, it is essential to isolate all power supplies in the circuit and discharge all capacitors, otherwise, it may lead to incorrect measurement results.
3
Di-LOG
...measurably better
DL6799 OPERATING MANUAL
8. When the “ “ symbol is displayed on the LCD, it is necessary to replace the batteries.
10. Do not store or use the clamp meter in an environment with high-temperature, high-humidity,
inammable, explosive or with a strong electromagnetic eld.
11. During maintenance, please clean the case of the instrument with a soft cloth soaked with neutral detergent, and not use abrasive solvents.
12. Instrument must be proved working before
use on a known voltage source to conrm instrument is operating correctly.
IV. Electrical Symbols
4
Di-LOG
...measurably better
V. Appearance Overview
1. Clamp head
2. Clamp body
3. Rotary dial
4. Function key
5. Display screen
6. Test lead input terminal
DL6799 OPERATING MANUAL
Figure 1
5
Di-LOG
...measurably better
DL6799 OPERATING MANUAL
VI. Function Keys
Key: ACI/DCI - AC/CD Current ACV/DCV - AC/DC Voltage
RES - Resistance CON - Continuity
DIO - Diode CAP - Capacitance
SELECT: Function selection button, which works
on ACI/DCI, ACV/DCV, and RES/CON/DIO/CAP measurement.
REL: Measurement of relative values, enabling
quick and easy di󰀨erential measurement on: ACV,
DCV, ACI and RES functions.
A short press of the REL button whilst performing a measurement will allow the instrument to zero the
displayed value. The instrument will display ∆ to
indicate you have entered relative mode. The next
measured value will display the di󰀨erential value in
relation to the previously measured value.
To exit REL mode, a short press of the REL button, will enable the instrument to return to normal operation.
ZERO: The DCI Zero button: this function is used
with REL and works on the DCI measurement.
A short press of the ZERO button before perform­ing a DCI measurement will allow the instrument to zero any displayed values caused by terrestrial magnetism . The instrument will display ZERO in the secondary display to indicate you have entered zero mode.
6
Di-LOG
...measurably better
DL6799 OPERATING MANUAL
To exit REL mode, a short press of the REL button, will enable the instrument to return to normal operation.
HOLD: the data hold button works in all measurement positions. A short press of the “HOLD” button under normal measurement will enter the data-hold mode, the instrument will not update the measurement and the displayed value will remain unchanged. The “H” symbol will be displayed on the top left of the display screen, to indicate the instrument has entered the data—hold mode. The data-hold feature can be cancelled, with either a short press of the “HOLD” button, a turn the rotary dial or a press of the “SELECT” button.
SCREEN BACKLIGHT :
A long press of the “ “ button will turn on the backlight . A long press of the “ “ button again will increase the brightness. A long press of the“ “
button again will turn the backlight o󰀨.
FLASHLIGHT :
A short press of the LED ashlight button will switch
the light on, and another short press of the LED
ashlight button will switch the light o󰀨.
7
DL6799 OPERATING MANUAL
Page left blank for notes:
Di-LOG
...measurably better
8
Di-LOG
...measurably better
DL6799 OPERATING MANUAL
Vll. Technical Index
1. General specications
Liquid crystal display: Maximum display up to 5999; Polarity display: Automatic positive and negative polarity display; Over-load display: Display “OL” or “–OL”; Battery voltage display: When the battery voltage is
below 2.4V the instrument will display the low battery
symbol; Sampling rate; Approximately 3 times/second; Sensor Types: Hall-e󰀨ect sensor for AC/DC measurement; Measured position error: When measuring the
current, ±1.0% additional error may occur if the source to be measured is not centred at the base of
the clamp jaw; Impact-resistant Strength: 1m height impact can be borne; Maximum opening dimension of clamp head: Diameter of conductor 14.7mm; Electromagnetic eld e󰀨ect: When any electromagnetic eld interference exists, the instrument may displays unstable or incorrect readings;
Power supply: 2 x AA 1.5 V batteries; Auto OFF function: 15 min; Dimensions: 210mm x 53mm x35mm; Weight approximately 163.7g (battery included);