Di-log SL500 User guide

Page 1
DL6508 User's Guide
1500Amp True RMS AC/DC Clamp Meter
Page 2
2
WARNING
CAUTION
Congratulations on your purchase of this DL6508 True RMS Clamp Meter. This meter measures AC Current, DC Current, AC/DC Voltage, Resistance, Capacitance, Frequency, Diode Test, Duty Cycle and Continuity. The molded case is designed for heavy duty use. This meter is shipped fully tested and calibrated and, with proper use, will provide years of reliable service.
Safety
International Safety Symbols
This symbol, adjacent to another symbol or terminal, indicates the user must refer to the manual for further information.
This symbol, adjacent to a terminal, indicates that, under normal use, hazardous voltages may be present
Double insulation
This WARNING symbol indicates a potentially hazardous situation, which if not avoided, could result in death or serious injury.
This CAUTION symbol indicates a potentially hazardous situation, which if
PER IEC1010 OVERVOLTAGE INSTALLATION CATEGORY
OVERVOLTAGE CATEGORY I
Equipment of OVERVOLTAGE CATEGORY I is equipment for connection to circuits in which measures are taken to limit the transient overvoltages to an appropriate low level.
Note – Examples include protected electronic circuits.
OVERVOLTAGE CATEGORY II
Equipment of OVERVOLTAGE CATEGORY II is energy-consuming equipment to be supplied from the fixed installation.
Note – Examples include household, office, and laboratory appliances.
OVERVOLTAGE CATEGORY III
Equipment of OVERVOLTAGE CATEGORY III is equipment in fixed installations. Note – Examples include switches in the fixed installation and some equipment for industrial use with permanent connection to the fixed installation.
OVERVOLTAGE CATEGORY IV
Equipment of OVERVOLTAGE CATEGORY IV is for use at the origin of the installation. Note – Examples include electricity meters and primary over-current protection equipment
not avoided, may result damage to the product.
Page 3
3
SAFETY NOTES
Function
Maximum Input
A AC,
1500A DC/AC
A DC
1500A DC/AC
V DC, V AC
1000V DC/AC
Resistance, Capacitance, Frequency, Diode Test
1000V DC/AC
Type K Temperature
30V DC, 24V AC
Do not exceed the maximum allowable input range of any function. Do not apply voltage to meter when resistance function is selected. Set the function switch OFF when the meter is not in use. Remove the battery if meter is to be stored for longer than 60 days.
WARNINGS
Set function switch to the appropriate position before measuring. When measuring volts do not switch to current/resistance modes. Do not measure current on a circuit whose voltage exceeds 600V. When changing ranges, always disconnect the test leads from the circuit under test.
CAUTIONS
Improper use of this meter can cause damage, shock, injury or death. Read and
understand this user manual before operating the meter.
Always remove the test leads before replacing the battery or fuses. Inspect the condition of the test leads and the meter itself for any damage before
operating the meter. Repair or replace any damage before use.
Use great care when making measurements if the voltages are greater than 25VAC rms
or 35VDC. These voltages are considered a shock hazard.
Always discharge capacitors and remove power from the device under test before
performing Diode, Resistance or Continuity tests.
Voltage checks on electrical outlets can be difficult and misleading because of the
uncertainty of connection to the recessed electrical contacts. Other means should be used to ensure that the terminals are not "live".
If the equipment is used in a manner not specified by the manufacturer, the protection
provided by the equipment may be impaired.
This device is not a toy and must not reach children‟s hands. It contains hazardous
objects as well as small parts that the children could swallow. In case a child swallows any of them, please contact a physician immediately
Do not leave batteries and packing material lying around unattended; they can be
dangerous for children if they use them as toys
In case the device is going to be unused for an extended period of time, remove the
batteries to prevent them from draining
Expired or damaged batteries can cause cauterization on contact with the skin. Always,
therefore, use suitable hand gloves in such cases
See that the batteries are not short-circuited. Do not throw batteries into the fire.
Page 4
4
Description
Meter Description
1. Current clamp
2. Clamp opening trigger
3. RANGE Button
4. MODE button
5. MAX/MIN button
6. INRUSH button
7. Backlit LCD Display
8. Non-Contact Voltage Detector
9. NCV LED indicator
10. Function switch
11. oC / oF select button
12. Back light button
13. ZERO button
14. PEAK button
15. Multimeter input jacks
Display icons Description
HOLD Data Hold
APO Auto Power Off
AUTO Autoranging
P Peak Hold
DC Direct Current
AC Alternating Current
MAX Max reading
MIN Min reading
Low battery
ZERO DCA or CAP zero
mV or V Milli-volts or Volts (Voltage)
Ohms (Resistance)
A Amperes (Current)
F Farad (Capacitance)
Hz Hertz (Frequency)
% Duty Ratio
o
F and oC Fahrenheit and Celsius units (Temperature)
n, m, , M, k Unit of measure prefixes: nano, milli, micro, mega, and kilo
)
)
) Continuity test
Diode test
Page 5
5
Operation
NOTES: Read and understand all Warning and Caution statements in this operation
manual prior to using this meter. Set the function select switch to the OFF position when the
meter is not in use.
Non-Contact Voltage Detector
WARNING: Risk of Electrocution. Before use, always test the Voltage Detector on a known
1. Rotate the Function switch to any measurement
2. Place the detector probe tip on the conductor to be
3. If AC voltage is present, the NCV detector will turn on
NOTE: The conductors in electrical cord sets are often
NOTE: The detector is designed with high sensitivity.
live circuit to verify proper operation.
position.
tested.
with a steady red light.
twisted. For best results, move the probe tip along a length of the cord to assure placing the tip close to the live conductor.
Static electricity or other sources of energy may randomly trip the sensor. This is normal operation.
Page 6
6
AC/DC Current Measurements
WARNING: Disconnect the test leads before measuring with clamp.
1. Rotate the Function switch to the 1500AAC/DC position.
2. Press the MODE button to select AC or DC.
3. Press the trigger to open jaw. Fully enclose only one conductor.
4. Read the current value in the display.
5. If the value is less than 400A, rotate the function switch to the 400AAC/DC position to improve resolution.
DCA Zero
The Zero feature removes offset values and improves accuracy for DC current measurements. To perform a zero, select ADC, and, with no conductor in the jaw, press the ZERO button. The display will show zero. The offset value is now stored and removed
from all measurements. Frequency When ACV is selected, the measured frequency can be viewed in the lower display.
AC/DC Voltage Measurements
CAUTION: Do not measure voltages if a motor on the circuit is being switched ON or OFF.
Large voltage surges may occur that can damage the meter.
1. Rotate the function switch to the V position.
2. Press the MODE button to select AC or DC Voltage.
3. Insert the black test lead banana plug into the negative COM jack.
Insert the red test lead banana plug into the positive V jack.
4. Touch the black test probe tip to the negative side of the circuit.
Touch the red test probe tip to the positive side of the circuit.
5. Read the voltage value in the display. Frequency
When ACA is selected, the measured frequency can be viewed in the lower display.
Page 7
7
Resistance Measurements
Note: Remove power from the device under test before measuring resistance.
1. Set the function switch to the position.
2. Insert the black test lead banana plug into the negative COM jack.
Insert the red test lead banana plug into the positive V jack.
3. Touch the black test probe tip to one side of the device to be measured.
Touch the red test probe tip to the other side of the device to be measured.
4. Read the resistance value in the display.
Continuity Test
1. Connect as described for resistance measurements.
)
2. Press the MODE button to select continuity
)
).
3. Touch the test probe tips across the circuit or component under test.
4. If the resistance is < 50, a tone will sound.
Diode Test
1. Connect as described for resistance measurements
2. Press the MODE button to select diode test .
3. Touch the test probe tips to the diode or semiconductor junction under test. Note the
meter reading.
4. Reverse the test lead polarity by reversing the red and black leads. Note this reading.
5. The diode or junction can be evaluated as follows:
If one reading displays a value (typically 0.400V to 01.800V) and the other reading
displays OL, the diode is good.
 If both readings display OL the device is open.  If both readings are very small or „0‟, the device is shorted.
Page 8
8
Capacitance Measurements
WARNING: To avoid electric shock, discharge the capacitor before measuring.
1. Press the MODE button to select capacitance measurements.
2. Insert the black test lead banana plug into the negative COM jack.
Insert the red test lead banana plug into the positive jack.
3. Touch the black test probe tip to one side of the device.
Touch the red test probe tip to the other side of the device.
4. Read the capacitance value in the display.
Note: For very large values of capacitance measurement time can be several seconds before
the final reading stabilizes.
Note: The Zero feature removes stray test lead capacitance to improve the accuracy of low
value capacitance measurements. To perform a zero, Press and hold the MODE ZERO button for two beeps. The display will zero. The offset value is now stored and is removed from all measurements.
.
Frequency and Duty Ratio Measurements
1. Rotate the function switch to the Hz % Position.
2. Insert the black test lead banana plug into the negative COM jack.
Insert the red test lead banana plug into the positive Hz jack.
3. Touch the black test probe tip to one side of the device.
Touch the red test probe tip to the other side of the device.
4. Read the Frequency value on the upper large display.
Read the Duty Ratio on the lower small display.
5. Press the MODE button to display the Duty Ratio on the large
display.
Type K Temperature Measurements
1. Rotate the function switch to the Tempposition.
2. Press the MODE button to select °F or °C.
3. Plug the transfer socket into the negative COM jack and the positive V jack, then insert
the TYPE K into the transfer socket, make sure that the positive and negative feet are connected correctly.
4. Connect the temperature probe tip(s) to the object which will be tested.
5. Read the temperature on the display. Note: In case of an open input or a temperature overrange, the meter will display “- - - - -” .
Page 9
9
Data Hold
To freeze the LCD reading, press the HOLD button. While data hold is active, the HOLD icon appears on the LCD. Press the HOLD button again to return to normal operation.
MAX/MIN
1. Press the MAX/MIN button to activate the MAX/MIN recording mode. The display icon
"MAX" will appear. The meter will begins recording and displaying the maximum value measured.
2. Press the MAX/MIN button and “MIN” will appear. The meter will display the minimum
value measured during the recording session.
3. Press the MAX/MIN button and “MAX MIN” will appear. The meter will display the present
reading, but will continue to update and store the max and min readings.
4. To exit MAX/MIN mode press and hold the MAX/MIN button for 2 seconds.
Peak Hold
When ACA or ACV is selected, pressing the PEAK button enables the peak capture circuit. The meter will now capture and display the maximum and minimum peaks of the waveform.
INRUSH
When ACA is selected, press the INRUSH button to activate the inrush capture circuit. A transient condition, generally lasting 110-120 milliseconds that occurs during motor
start-up.
RANGE
In the Voltage, Resistance, Capacitance, Frequency or uA function the meter automatically selects the best range for the measurements being made. For measurement situations requiring that a range be manually selected, perform the following:
1. Press the RANGE button. The “AUTO” display icon will turn off.
2. Press the RANGE key to step through the available ranges. Observe the decimal point and units displayed until the preferred range is located.
3. To exit the Manual Ranging mode and return to Autoranging, press and hold the
RANGE key for 2 seconds.
LCD Backlight
The LCD is equipped with backlighting for easier viewing, especially in dimly lit areas. Press
to turn the backlight on. The backlight will automatically turn off after 30 seconds.
Automatic Power OFF with Disable
In order to conserve battery life, the meter will automatically turn off after approximately 30 minutes. To turn the meter on again, turn the function switch to the OFF position and then turn to the desired function position.
To disable APO:
1. From the OFF position, hold the MODE button and rotate the FUNCTION switch to a measurement function.
2. will appear in the display
3. Release the MODE button
4. APO is now disabled (APO icon is off) and will be reset when the Function switch is returned to the OFF position.
Low battery indication
When the icon and -----appears in the display, the battery should be replaced. Refer to the battery replacement procedure in the maintenance section.
Page 10
10
Maintenance
WARNING: To avoid electrical shock, disconnect the meter from any circuit, remove the test leads from the input terminals, and turn OFF the meter before opening the case. Do not operate the meter with an open case.
Cleaning and Storage
Periodically wipe the case with a damp cloth and mild detergent; do not use abrasives or solvents. If the meter is not to be used for 60 days or more, remove the battery and store it separately.
Battery Replacement
1. Remove the Phillips head screw secures the rear battery door.
2. Open the battery compartment
3. Replace the 9V battery
4. Secure the battery compartment door with the screw
You, as the end user, are legally bound (Battery ordinance) to return all used batteries and accumulators; disposal in the household garbage is prohibited!
You can hand over your used batteries / accumulators, gratuitously, at the collection points for our branches in your community or wherever batteries / accumulators are sold!
Disposal
Follow the valid legal stipulations in respect of the disposal of the device at the end of its lifecycle
Page 11
11
Specifications
Function
Range& Resolution
Accuracy (% of reading)
DC Current
400.00 ADC
± (2.0% +30digits)
1500.0 ADC
± (2.5% +30digits)
AC Current
True RMS (50Hz to 60
Hz)
400.00 AAC
± (2.8% +30digits)
1500.0 AAC
± (2.8% +30digits)
All AC voltage ranges are specified from 5% of range to 100% of range
DC Voltage
400.00 mVDC
± (0.1% + 5 digits)
4.0000VDC
± (0.1% + 4 digits)
40.000 VDC
400. 00 VDC
1000.0 VDC
± (0.5% + 4 digits)
AC Voltage
True RMS (50 Hz to
1000 Hz)
400.00 mVAC
± (0.8% + 40 digits)(50/60Hz)
4.0000 VAC
± (1.0% + 30 digits)
40.000 VAC
400. 00 VAC
0750.0 VAC
All AC voltage ranges are specified from 5% of range to 100% of range
Resistance
400.00 Ω
± (0.5% + 9 digits)
4.0000KΩ
± (1.0% + 4 digits)
40.000KΩ
400.00KΩ
4.0000MΩ
± (2.0% + 10digits)
40.000MΩ
± (3.0% + 10 digits)
Capacitance
400.00nF
(3.5% reading + 40digits)
4000.0nF
(3.5% reading + 10digits)
40.00.μF
400.0.μF
4.000mF
(5% reading + 10 digits)
20.00mF
40.00mF
Not specified
Frequency
40.000 Hz
(0.3% reading + 2 digits)
400.00 Hz
4.0000KHz
40.000KHz
400.00KHz
4.0000MHz
40.000MHz
Sensitivity: 0.8V rms min. @ 20% to 80% duty cycle and <100kHz; 5Vrms min @ 20%
to 80% duty cycle and > 100kHz.
Duty Cycle
10.0 to 95.0%
(1.0% reading + 2 digits)
Pulse width: 100µs - 100ms, Frequency: 10Hz to 100kHz
Temp
(type-K)
(probe accuracy not
included)
-100.0 to 1000.0C
(1.0% reading + 2.5 C)
-148.0 to 1832.0F
(1.0% reading + 4.5F)
(probe accuracy not included)
Page 12
12
General Specifications
Clamp jaw opening 2.0" (52mm) approx. Display Dual 40,000/4,000 count backlit LCD Continuity check Threshold 50; Test current < 0.5mA Diode test Test current of 0.3mA typical; Open circuit voltage 2.8VDC typical Low Battery indication Battery symbol is displayed Over-range indication „OL‟ display Measurement rate 2 readings per second, nominal Peak detector >1ms Thermocouple sensor Type K thermocouple required Input Impedance 10M (VDC and VAC) AC bandwidth 50 to 400Hz (AAC and VAC) AC response True rms (AAC and VAC) Crest Factor 3.0 in 40A and 400A ranges, 1.4 in 1000A range (50/60Hz and
Operating Temperature 5C to 40C (41F to 104F)
Storage Temperature -20C to 60C (-4F to 140F) Operating Humidity Max 80% up to 31C (87F) decreasing linearly to 50% at
Storage Humidity <80% Operating Altitude 7000ft. (2000meters) maximum. Battery One (1) 9V Battery (NEDA 1604) Auto power OFF After approx. 30 minutes Dimensions & Weight 11.57x4.13x1.85” (294x105x47mm); 18.9 oz. (536g) Safety For indoor use and in accordance with the requirements for
Approvals CE
Warranty 2 Years from date of purchase
5% to 100% of range)
40C (104F)
double insulation to IEC1010-1 (2001): EN61010-1 (2001) Overvoltage Category III 600V and Category II 1000V, Pollution Degree 2.
Di-log Test Equipment Unit 28 Wheel Forge Way Trafford Park Manchester M17 1EH
Tel: 0161 877 0322 Email: sales@dilog.co.uk Web: www.dilog.co.uk
Page 13