Di-LOG Test Equipment
28 Wheel Forge Way
Manchester
M17 1EH
Tel: 0161 877 0322
www.dilog.co.uk
sales@dilog.co.uk
403A592 Rev 2
Operating Instructions
2
Page 3
DLPT2+
Limited Warranty & Limitation of Liability
DILOG Electronic Limited guarantees this product to
be free from defects in material and workmanship
under normal use and service for a period of 1 year.
The period of warranty will be effective from the day
of delivery.
(c) Copyright 2019
All rights reserved. Nothing from this edition may be
multiplied, or made public in any form or manner,
either electronically, mechanically, by photocopying,
recording, or in any manner, without prior written
consent from DILOG Electronic Limited. This also
applies to accompanying drawings and diagrams.
Due to a policy of continuous development DILOG
Electronic Limited reserves the right to alter the
equipment specification and description outlined in
this publication without prior notice and no part of
this publication shall be deemed to be part of any
contract for the equipment unless specifically
referred to as an inclusion within such contract.
Operating Instructions
3
Page 4
DLPT2+
Disposal of old product
Operating Instructions
This product has been designed and manufactured
with high quality materials and components that can
be recycled and reused.
When this symbol is attached to a product it means
the product is covered by the European Directive
2012/19/EU.
Please familiarise yourself with the appropriate local
separate collection system for electrical and
electronic products.
Please dispose of this product according to local
regulations. Do not dispose of this product along with
normal waste material. The correct disposal of this
product will help prevent potential negative
consequences for the environment and human
health.
4
Page 5
DLPT2+
1 Important Information ...................................... 6
These operating instructions are intended for the
use of adequately trained personnel.
The following symbols are used in these operating
instructions and on the DLPT2+.
avoid danger to persons.
instructions must be adhered to in order
to avoid danger.
Before use, ensure unit is clean and dry; visually
inspect all leads, connectors, and case. Any
damage or wear must be rectified prior to use.
specified by these operating instructions
then the protection provided may be
impaired.
appliance under test is electrically
isolated.
leads must be checked for signs of
not use if there are signs of damage. Only
use the correct leads supplied with the
DLPT2+.
barrier on the test probe.
6
Page 7
DLPT2+
The
DLPT2+
may apply high voltage or
If the
DLPT2+
is being used to determine
Do not operate the
DLPT2+
in an
The
DLPT2+
has been designed to make
Do not open the
DLPT2+
, no user
mains power to the appliance under test.
Do not touch conductive parts of the
appliance while tests are active.
the presence or absence of hazardous
voltages, always prove the operation of
voltage measurement function before and
after use by means of a known voltage
source or proving unit.
explosive gas or dust environment.
measurements in a dry environment.
serviceable parts.
Operating Instructions
Note: Data may be lost or altered in virtually
any electronic memory under certain
circumstances. Therefore DILOG
Electronic assumes no responsibility for
financial losses or claims due to data lost
or otherwise rendered unusable whether
as a result of abuse, improper use,
defects, disregard of operating
instructions or procedures, or any allied
causes.
PATGuard 3 software can also be trialled and
purchased online from the below web address:
www.DILOG.co.uk/pg3trial
8
Page 9
DLPT2+
Operating Instructions
Figure 1. DLPT2+ Front View
9
Page 10
DLPT2+
Operating Instructions
Figure 2. DLPT2+ End View
Figure 3. Testing CLI appliances, no mains leakage
10
Page 11
DLPT2+
Operating Instructions
Figure 4. Testing CLII appliances, no mains leakage
11
Page 12
DLPT2+
Operating Instructions
Figure 5. Testing CLI appliances, mains leakage
12
Page 13
DLPT2+
Operating Instructions
Figure 6. Testing CLII appliances, mains leakage
13
Page 14
DLPT2+
Operating Instructions
Figure 7. Testing IEC Leads
14
Page 15
DLPT2+
Operating Instructions
Figure 8. Testing Installation RCD
15
Page 16
DLPT2+
Operating Instructions
Figure 9. Testing Plug Top RCD
16
Page 17
DLPT2+
2 Introduction
The DLPT2+ is a hand held battery powered unit
suitable for carrying out electrical safety checks on:
• Class I appliances
• Class II appliances
• IEC mains leads
• Extension leads
• 30mA Plug-top RCDs
• 30mA Installation RCDs
• Mains outlet wiring
• 3-Phase appliances when used with the DILOG
Overview
With reference to Figures 1 and 2.
1. Save Key
2. Recall Key
3. Print Key
4. CLI / Cord Sequence Key
5. CLII Test Sequence Key
6. Leakage Test Start Key
7. Setup Button
8. RCD Test Key
9. 3-Phase Test Sequence Key
10. LCD
11. Test/Mains outlet socket
12. USB Interface
13. IEC Test Connection
14. Mini DIN Printer Interface
15. Earth Continuity Test Terminal
16. Mains inlet socket
User Interface
The LCD display shows test progress, results for
individual tests and the overall test result for an
appliance or mains cord.
TPA range of 3-phase adaptors.
Operating Instructions
17
Page 18
DLPT2+
Operating Instructions
Power ON/OFF = press and hold CLI and CLII
simultaneously
Class I appliance test = press the CLI key
Class II appliance test = press the CLI key
Cord / extension lead test = press the CLI key
Enter Setup Mode = press the SETUP key
Once in the setup mode the CLII button can be used
to the change the Insulation voltage while the LKGE
button can be used to change the ILEAK pass/fail
limit. To leave the setup mode press the SETUP
button.
30mA RCD test = press the 30mA RCD key
3 phase appliance test = press the 3 phase key
Note: The DLPT2+ will automatically switch OFF
after approximately 1 minute if no keys are
pressed.
Note: When a key is pressed to initiate a test
sequence, the DLPT2+ will compare the type of
appliance connected with the test sequence that
has been selected, and where possible, will
prompt the user if an inappropriate test has been
selected.
For example, if an IEC lead is connected
between the front panel mains socket and the
IEC socket on the end panel and a CLI appliance
test is selected, the DLPT2+ will flash the Cord
test enunciator to indicate that the Cord test is
the most appropriate test.
18
Page 19
DLPT2+
LN
LE
NE
3 Performing Tests
Press CLI and CLII keys simultaneously to switch on
the DLPT2+. When the unit is ready the display will
be as shown below.
Operating Instructions
3.1 Checking a mains power outlet
Connect DLPT2+ mains lead to the mains inlet
socket and to the power outlet to be tested.
Each time mains voltage is detected, the lineneutral, line-earth and neutral-earth voltages will be
displayed for 2s as shown below.
If the voltages are within the acceptable limits
shown below, a tick is shown next to the enunciator.
195V-253V
195V-253V
<30V
19
Page 20
DLPT2+
After 2s the display will return to the start up screen.
If the mains voltage is incorrect then the DLPT2+ will
remain in voltage mode until the voltage is corrected
or removed.
3.2 Testing a Class I Appliance
• Visually inspect the appliance and mains cord
• If the appliance passes a visual inspection
• Plug the earth test lead into the Earth Continuity
• Plug the appliance into the DLPT2+ front panel
• Connect the earth test probe to an exposed
• If the appliance under test has an ON/OFF
• Press the CLI test key
• The DLPT2+ will now perform a measurement of
• If the measured value is higher than the factory
for signs of damage.
proceed with the electrical tests.
Test Terminal on the DLPT2+ end panel.
test/mains outlet socket.
metal part on the appliance.
switch, make sure it is in the ON position.
the protective earth continuity, reverse the test
current and repeat the measurement. The
highest measured value that occurred during the
two tests is displayed.
set pass/fail threshold, the measured value is
displayed and the unit indicates a fail result as
shown below and the test sequence is halted.
Operating Instructions
20
Page 21
DLPT2+
•If the measured value is less than the factory set
pass/fail threshold, the measured value is
displayed and the unit indicates a pass result, as
shown below.
Operating Instructions
Note: The DLPT2+ can perform a continuous Rpe
measurement (max 3 minutes). Press and
hold the CLI key for approximately 5s until
the unit emits a beep and the symbol
appears on the display. The DLPT2+
continuously displays the current
measured value on the display and stores
the maximum value in the memory. By
pressing the CLI key again, the
measurement is repeated with reversed
polarity. Press the key again to terminate
the Rpe test and display the maximum
value of RPE on the display.
•The insulation test voltage is shown momentarily
next to the Riso enunciator.
•The unit will proceed with the Insulation and
Leakage tests.
Note: The power switch on the appliance under
test must be in the ON position to perform
an insulation test. If no appliance is
detected the DLPT2+ will display the
following.
21
Page 22
DLPT2+
Operating Instructions
•Check that the appliance power switch is in the
ON position. The test will automatically proceed
if the appliance power switch is placed in the
ON position.
•If the LO LOAd enunciator remains on the
display, the load presented by the appliance
may be too small for the DLPT2+ to detect. In
this case, press the CLI test key to continue.
•If the Insulation Resistance is greater than the
factory set limit a tick is placed next to the Riso
enunciator. If the Insulation Resistance is less
than the set limit a cross is placed next to the
Riso enunciator, the FAIL enunciator is
displayed and the test sequence is halted.
Note: If a mains supply is detected, the DLPT2+
will perform a differential leakage test.
If the DLPT2+ does not detect a mains
supply then the DLPT2+ will perform a
substitute leakage test.
•If a substitute leakage test is performed then the
DLPT2+ will start the test without the need for
user intervention. If the substitute leakage is less
than the factory set limit a tick is placed next to
the Iea enunciator.
22
Page 23
DLPT2+
LN LE NE Mains status
Operating Instructions
•If a differential leakage test is performed then
the test DLPT2+ will prompt the user to start the
Leakage test by flashing the Ileak icon. If no key
is pressed the DLPT2+ will timeout and return to
the start screen.
Note: If the LKGE key is pressed momentarily the
differential test duration is 5s.
If the LKGE key is pressed and held for 2s
the differential test duration is 2 minutes.
When the duration is set to 2 minutes, the
test can be terminated at any time by
pressing the LKGE key.
•When the LKGE key is pressed to start the
differential leakage test the DLPT2+ will check
the incoming mains supply to ensure that the
voltages are correct. The enunciators are
displayed if there is an error with the connected
mains supply:
Flash Flash OFF No mains
OFF Flash Flash Earth fault or Live
and Neutral reverse
•If the mains voltage is correct, the DLPT2+ will
proceed with the differential leakage test.
23
Page 24
DLPT2+
Operating Instructions
•If the Differential Leakage is less than the factory
set limit a tick is placed next to the Ileak
enunciator.
•If all of the tests within the sequence have
passed then the PASS enunciator is illuminated.
24
Page 25
DLPT2+
3.3 Testing a Class II Appliance
• Visually inspect the appliance and mains cord
• If the appliance passes a visual inspection
• Plug the earth test lead into the Earth Continuity
• Plug the appliance into the DLPT2+ front panel
• Connect the earth test probe to an exposed
• If the Appliance under test has an ON/OFF
• Press the CLII test key
• The display will momentarily show the insulation
for signs of damage.
proceed with the electrical tests.
Test Terminal on the DLPT2+ end panel.
test/mains outlet socket.
metal part on the appliance.
switch, make sure it is in the ON position.
test voltage before proceeding with the test.
Operating Instructions
Note: The power switch on the appliance under
test must be in the ON position to perform
an insulation test. If no appliance is
detected the DLPT2+ will display the
following.
25
Page 26
DLPT2+
• Check that the appliance power switch is in the
• If the LO LOAd enunciator remains on the
• The DLPT2+ will perform an Insulation test. If
Note: If a mains supply is detected, the DLPT2+
•If a Substitute Leakage test is performed then
ON position. The test will automatically proceed
if the appliance power switch is placed in the
ON position.
display, the load presented by the appliance
may be too small for the DLPT2+ to detect. In
this case, press the CLII test key to continue.
the Insulation Resistance is greater than the
factory set limit a tick is placed next to the Riso
enunciator. If the Insulation Resistance is less
than the set limit a cross is placed next to the
Riso enunciator, the FAIL enunciator is
displayed and the test sequence is halted.
will perform a touch leakage test.
If the DLPT2+ does not detect a mains
supply then the DLPT2+ will perform a
substitute leakage test.
the DLPT2+ will start the test without the need
for user intervention. If the Substitute Leakage is
less than the factory set limit a tick is placed
next to the Iea enunciator.
Operating Instructions
•If a touch leakage test is performed then the
DLPT2+ will prompt the user to start the
Leakage test by flashing the Ileak icon. If no key
26
Page 27
DLPT2+
LN LE NE Mains status
Note: If the LKGE key is pressed momentarily the
• When the LKGE key is pressed to start the
• If the mains voltage is correct, the DLPT2+ will
• If the touch leakage is less than the factory set
• If all of the tests within the sequence have
is pressed the DLPT2+ will timeout and return to
the start screen.
touch leakage test duration is 5s
If the LKGE key is pressed and held for 2s
the touch leakage test duration is 2
minutes.
When the duration is set to 2 minutes, the
test can be terminated at any time by
pressing the LKGE key.
touch leakage test the DLPT2+ will check the
incoming mains supply to ensure that the
voltages are correct. The enunciators are
displayed if there is an error with the connected
mains supply:
Flash Flash OFF No mains
OFF Flash Flash Earth fault or Live
proceed with the touch leakage test.
limit a tick is placed next to the Ileak enunciator.
passed then the PASS enunciator is illuminated.
Operating Instructions
and Neutral reverse
27
Page 28
DLPT2+
Operating Instructions
3.4 Testing a mains cord
•Visually inspect the mains cord and plug for
signs of damage.
• Check that the correct fuse is fitted.
• If the cord passes a visual inspection proceed
with the electrical tests.
•Plug the mains cord under test into the IEC
socket and the front panel test/mains socket on
the DLPT2+.
• Press the CLI test key
• The DLPT2+ will first test the continuity of the
protective earth.
•If the measured value is greater than the factory
set limit a cross is placed next to the RPE
enunciator, a FAIL is indicated and the test
sequence will be halted.
•If the measured value is less than the factory set
limit a tick is placed next the RPE enunciator.
• The unit will proceed with the Insulation test.
• If the Insulation Resistance is lower than the
factory set limit a cross is placed next to the
Riso enunciator and the test sequence is halted.
•If the Insulation Resistance is greater than the
factory set limit a tick is placed next to the Riso
enunciator. If the Insulation Resistance is less
than the set limit a cross is placed next to the
Riso enunciator, the FAIL enunciator is
displayed and the test sequence is halted.
•The unit will proceed with the wiring test,
checking the live and neutral conductors for
short or open circuit and reverse connections
•If the wiring is correct a tick is placed next to the
cord enunciator, the GOOd enunciator is
illuminated and a PASS is indicated for the
sequence.
28
Page 29
DLPT2+
Operating Instructions
Note: If the tested cord has a wiring fault, one of
the following enunciators will be illuminated in
place of GOOd
•OPEn indicates that either the live or
neutral conductor is broken (open circuit)
•Shor indicates that the live and neutral
conductors are shorted together
•CrOS indicates that the live and neutral
conductors are reversed
3.5 Testing an extension lead
•Visually inspect the mains cord and plug for
signs of damage.
• Check that the correct fuse is fitted.
• If the cord passes a visual inspection proceed
with the electrical tests.
•Plug the supplied 0.5m IEC lead into the IEC
socket and into a mains outlet on the extension
lead. Plug the mains plug of the extension lead
into the front panel mains socket on the
DLPT2+.
•The extension lead can now be tested in the
same manner as an IEC as described above.
29
Page 30
DLPT2+
Operating Instructions
3.6 Changing the Insulation Test Voltage
The Insulation test performed in the CLI and CLII
test sequences can be changed to 250V for
appliances with MOVs which would fail an Insulation
test performed at 500V.
In order to change the Insulation voltage press the
SETUP button to enter setup mode. Once in setup
mode press the CLII button, the selected insulation
voltage will be displayed next to the Riso
enunciator.
To leave setup mode press the SETUP button.
3.7 Changing the CLI Leakage limit
The CLI Leakage pass/fail limit can be changed
between 0.75mA and 3.5mA.
In order to change the Leakage limit press the
SETUP button to enter setup mode. Once in setup
mode press the LKGE button, the selected pass/fail
limit will be displayed next to the Ileak enunciator.
To leave setup mode press the SETUP button.
30
Page 31
DLPT2+
When performing an RCD test ensure that
Equipment which is connected
Voltages between the protective conductor
Leakage currents in the circuit may
Potential fields of other earthing
LN LE NE Mains status
3.8 Performing an RCD test on an installation
RCD
Operating Instructions
The DLPT2+ can be used to measure the operating
time of an installation RCD.
Connect the DLPT2+ IEC lead to the IEC socket of
the DLPT2+ and a mains outlet on the RCD
protected circuit.
To test the RCD operating time press the 30mA
RCD test key.
The mains supply status is checked and the status
is indicated using the LN, LE and NE enunciators of
the LCD.
the Continuity Test probe is disconnected
from the DLPT2+
downstream of the RCD may cause
measurement inaccuracies.
and earth may influence measurements.
influence measurements.
installations may influence the
measurement.
Flash Flash OFF No mains
OFF Flash Flash Earth fault
31
Page 32
DLPT2+
When plugging the mains lead into the
When performing an RCD test ensure that
Whenever reSet is displayed on the LCD press the
reset button on the RCD.
The DLPT2+ will perform a IΔn (30mA) 0o test, the
operating time of the RCD will be shown on the
DLPT2+ LCD. If the test passes then the DLPT2+
will perform a IΔn (30mA) 180o test. As long as the
tests continue to pass the DLPT2+ will continue to
test both 0o and 180o at IΔn (30mA).
Operating Instructions
3.9 Performing an RCD test on a Plug top RCD
The DLPT2+ can be used to measure the operating
time of a plug top RCD.
DLPT2+ plus ensure that the polarity
orientation of the connector is correct. DO
NOT FORCE THE PLUG INTO THE
CONNECTOR. Doing so may damage the
DLPT2+.
the Continuity Test probe is disconnected
from the DLPT2+
Connect the DLPT2+ mains lead to the mains inlet
socket of the DLPT2+ and a mains outlet.
Plug the RCD into the DLPT2+. Plug the supplied
IEC lead into the RCD and the DLPT2+ IEC socket.
To test the RCD operating time press the 30mA
RCD test key.
The mains supply status is checked and the status
is indicated using the LN, LE and NE enunciators of
the LCD.
32
Page 33
DLPT2+
LN LE NE Mains status
Flash Flash OFF No mains
OFF Flash Flash Earth fault
Whenever reSet is displayed on the LCD press the
reset button on the RCD.
The DLPT2+ will perform a IΔn (30mA) 0o test, the
operating time of the RCD will be shown on the
DLPT2+ LCD. If the test passes then the DLPT2+
will perform a IΔn (30mA) 180o test. As long as the
tests continue to pass the DLPT2+ will continue to
test both 0o and 180o at IΔn (30mA).
3.10 Performing a 3-Phase Sequence
The DLPT2+ can be used to measure Rpe and earth
leakage current on 3 phase equipment when used
with the optional 3 phase adaptor accessory.
• Visually inspect the appliance and mains cord
• If the appliance passes a visual inspection
• Refer to the documentation supplied with the 3
• Connect the 3 phase adaptor test lead to the
• Connect the Rpe test probe to an exposed
• Press the 3 phase key on the DLPT2+ to being
• If there is a contact voltage present on the
• Otherwise, the DLPT2+ will proceed with the
for signs of damage.
proceed with the electrical tests.
phase adaptor and connect the adaptor in line
with the 3 phase equipment under test.
mains input socket 9 (refer to figure 2) on the
DLPT2+.
conductive part on the 3 phase equipment.
the test.
conductive parts of the equipment under test,
the test is terminated and a Volt on test Probe
warning message is shown on the display.
Rpe measurement, reverse the test current and
then repeat the measurement. The highest
Operating Instructions
33
Page 34
DLPT2+
Note: There is no factory programmed test limit
•The DLPT2+ will then display the current flowing
measured value that occurred during the two
tests is displayed for approximately 2s.
for the 3 phase Rpe measurement and so
no tick or cross is displayed.
in the protective conductor of the 3 phase
supply.
Operating Instructions
•The 3-Phase leakage test will run for 30 seconds
but can be stopped earlier by pressing the 3Phase button.
Note: If the 3-Phase key is pressed while no
adaptor is connected to the DLPT2+ then a
warning message PLUG IN 3 PH ADPTR
will be displayed.
34
Page 35
DLPT2+
Operating Instructions
4 Memory & Print Functions
The DLPT2+ can store the results for a maximum of
999 appliances.
4.1 Saving Test Results
In order to store a test result press the STORE Key
when the test sequence has been completed. The
results displayed on the LCD will be recorded in
memory. Once the results for a test sequence have
been recorded the STORE Key is disabled until
another test sequence has been performed.
4.2 Viewing Test Results
When the DLPT2+ is waiting to start a new test
sequence press the RECALL Key, this will display
the last recorded results. Other results can be
displayed by scrolling through the memory. To scroll
through the memory use the RECALL and STORE
Keys. Pressing any other Key or waiting for 10
seconds will return the DLPT2+ into test mode.
4.3 Deleting Test Results
To delete all of the results in memory press and hold
the STORE and RECALL keys together while
displaying the recalled results until the counter
reaches zero.
4.4 Transferring Test Results
Test data can be transferred to a PC using the
Datalogger on the provided DLPT2+ Support CD.
Insert the CD into your disc drive and then view the
InstallationInstructions.txt file for information on how
to install.
Once the USB drivers and Datalogger are installed
you can open the Datalogger program from your
desktop.
Connect your device to the PC using a USB
connector. (The very first time you do this, your PC
will automatically install a COM port driver). You can
check which COM port is being used by looking in
the Device manager.
35
Page 36
DLPT2+
E.g.: Silicon Labs CP210x USB to UART
Bridge(COM14)
Select the correct COM port.
Click on Download. A progress bar will appear.
On the DLPT2+ in order to transfer the entire test
storage area to PC press and hold the RECALL key
for 5 seconds. The data will be transmitted using the
USB interface.
When all data has been captured and stored the
progress bar will disappear and you will be asked
where to save the data.
Once the file is saved you may click the "Open"
button to open the document with default
application.
For example, if you saved the file in .csv format (the
default) then it may open using a spreadsheet
application.
4.5 Printing Test Results
In order to print a test label press the PRINT Key
when the test sequence has been completed or
when the test result is being reviewed. The printer
must be attached before the PRINT Key is pressed.
In order to save power the DLPT2+ will turn the
printer on and off when required.
4.6 Setting the Real Time Clock
Setting time and date:
Ensure that the DLPT2+ is powered off. Press and
hold the RECALL key and then press simultaneously
the class I and class II keys to power on the tester.
The date format and time format is shown as
follows:
MM.DD = month (1-12). Day (1-31)
YYYY = year
HH.mm = hours (0-23).minutes (0-59)
SS = seconds (0-59)
Press the class I key to select a date field and time
field.
A flashing digit shows that this field can be set.
Operating Instructions
36
Page 37
DLPT2+
The save key and recall key are used to increase or
decrease the value in each field, the second field is
set to zero.
Turn off the device to save the setting.
Operating Instructions
37
Page 38
DLPT2+
5 Specification
Earth Continuity
Insulation resistance
Substitute Leakage Current
Differential Leakage Current
Touch Leakage Current
3-Phase Leakage Current
Display Range 0.01ohms to 19.99ohms
Measuring Range 0.05ohms to 19.99ohms
Accuracy ± (5% + 2 digits)
Test current 200mA minimum
Test voltage 6V nominal
Display Range 0.01Mohms to 19.99Mohms
Measuring Range 0.10Mohms to 19.99Mohms
Accuracy ± (5% + 2 digits)
Test voltage 250V / 500V
Test current >1mA
Test current <2mA into 2k
Display Range 0.10mA to 19.99mA
Measuring Range 0.25mA to 19mA
Accuracy ± (5% + 2 digits)
Test voltage 40V rms, 50Hz AC
Test current <10mA into 2k
Display Range 0.15mA to 19.99mA
Measuring Range 0.25mA to 19mA
Accuracy ± (5% + 2 digits)
Test voltage mains voltage, 50Hz AC
Display Range 0.10mA to 3.5mA
Measuring Range 0.10mA to 1.99mA
Accuracy ± (5% + 2 digits)
Test voltage mains voltage, 50Hz AC
Display Range 0.10mA to 9.99mA
Measuring Range 0.25mA to 9.99mA
Accuracy ± (5% + 2 digits)
Test voltage mains voltage, 50Hz AC
Operating Instructions
38
Page 39
DLPT2+
Earth
Insulation
Leakage
Class I
Class II
Cord
3-Phase
Cord Test
Earth continuity, insulation resistance as above.
Check for Live and Neutral open circuit, short
RCD Test
Factory Set Pass/Fail limits
circuit or reversed polarity.
Display Range IΔn - 0ms to 500ms
5IΔn – 0ms to 40ms
Measuring Range IΔn - 10ms to 500ms
5IΔn – 10ms to 40ms
Accuracy ± 2ms
Test Current IΔn – 30mA
5IΔn – 150mA
Current Accuracy -0, +10%
Operating Instructions
Continuity
0.2 ohms 1.0Mohm 0.75 / 3.5
n/a 2.0Mohm 0.25mA
0.2 ohms 1.0Mohm n/a
0.2 ohms 1.0Mohm 3.5mA
Environmental rating
IP Rating IP40
Operating temperature range 0°C to 40°C, without
moisture condensation.
Storage temperature range –25° to 65°.
Note: Batteries should be removed prior to
storage.
Overvoltage category 300V CAT II
Resistance
39
Page 40
DLPT2+
6 Maintenance
Operating Instructions
Clean only with a dry cloth; do not use solvents.
Before use, ensure unit is clean and dry; visually
inspect all leads, connectors, and case. Any
damage or wear must be rectified to preserve user
safety.
Check the battery contacts and compartment are
free of electrolytic contamination.
Any contamination of the battery contacts or
compartment should be cleaned with a dry cloth.
Note: The DLPT2+ contains no user serviceable
parts. If an Error warning should appear on the
display please contact the manufacturer or an
authorised DILOG Service Agent for advise.
7 Battery Check
The DLPT2+ is powered from 6 AA cells which are
checked before a test is performed. When the
battery voltage is low the enunciator is
illuminated. The unit will continue to perform within
specification for a limited number of tests,
dependent upon the type of the batteries fitted.
When the battery voltage reaches a level where the
performance is affected the enunciator will
flash and all test keys are disabled. The batteries
must be replaced.
40
Page 41
DLPT2+
7.1 Battery Replacement
Before opening the DLPT2+ ensure that all
test leads are disconnected.
•Switch off the unit by pressing and holding
the CLI and CLII keys.
• Disconnect all leads from the DLPT2+.
• Place the DLPT2+ face down and release
the captive screw in the battery
compartment cover.
•Remove the battery compartment cover and
remove the discharged batteries.
•Insert the replacement batteries into the
battery compartment ensuring that the
battery polarity matches the marking on the
inside of the battery compartment.
•Relocate the battery cover over the battery
compartment and fasten in position with the
battery cover captive screw.
Operating Instructions
41
Page 42
DLPT2+
Part Number
8 Fuse Replacement
Replacement F16A 250V Fuse 27B114
Before opening the DLPT2+ ensure that all
test leads are disconnected.
•Switch off the unit by pressing and holding
the CLI and CLII keys.
• Disconnect all leads from the DLPT2+.
• Place the DLPT2+ face down and release
the captive screw in the battery
compartment cover.
•Remove the battery compartment cover and
remove the discharged batteries.
•Insert the replacement fuse into the fuse
compartment ensuring that the fuse snaps
into place.
•Relocate the battery cover over the battery
compartment and fasten in position with the
battery cover captive screw.
Operating Instructions
42
Page 43
DLPT2+
9 Service and Calibration
Operating Instructions
To maintain the specified accuracy of the
measurement results, the instrument must be
recalibrated at regular intervals by either the
manufacturer or an authorised service agent. We
recommend a recalibration period of one year.
For help or advice on Service and Calibration
contact:
Di-LOG Test Equipment
28 Wheel Forge Way
Manchester
M17 1EH
Tel: 0161 877 0322
www.dilog.co.uk
sales@dilog.co.uk
43
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.