Di-LOG DLPT2+ Operating Instructions Manual

Page 1
Page 2
DLPT2+
DLPT2+
Operating Instructions
Di-LOG Test Equipment 28 Wheel Forge Way Manchester M17 1EH
Tel: 0161 877 0322
www.dilog.co.uk
sales@dilog.co.uk
403A592 Rev 2
Operating Instructions
2
Page 3
DLPT2+
Limited Warranty & Limitation of Liability
DILOG Electronic Limited guarantees this product to be free from defects in material and workmanship under normal use and service for a period of 1 year. The period of warranty will be effective from the day of delivery.
(c) Copyright 2019 All rights reserved. Nothing from this edition may be multiplied, or made public in any form or manner, either electronically, mechanically, by photocopying, recording, or in any manner, without prior written consent from DILOG Electronic Limited. This also applies to accompanying drawings and diagrams.
Due to a policy of continuous development DILOG Electronic Limited reserves the right to alter the equipment specification and description outlined in this publication without prior notice and no part of this publication shall be deemed to be part of any contract for the equipment unless specifically referred to as an inclusion within such contract.
Operating Instructions
3
Page 4
DLPT2+
Disposal of old product
Operating Instructions
This product has been designed and manufactured with high quality materials and components that can be recycled and reused.
When this symbol is attached to a product it means the product is covered by the European Directive 2012/19/EU.
Please familiarise yourself with the appropriate local separate collection system for electrical and electronic products.
Please dispose of this product according to local regulations. Do not dispose of this product along with normal waste material. The correct disposal of this product will help prevent potential negative consequences for the environment and human health.
4
Page 5
DLPT2+
1 Important Information ...................................... 6
2 Introduction .................................................... 17
3 Performing Tests ............................................ 19
4 Memory & Print Functions .............................. 35
5 Specification ................................................... 38
6 Maintenance ................................................... 40
7 Battery Check ................................................ 40
8
9 Service and Calibration .................................. 43
Contents
3.1
Checking a mains power outlet ............. 19
3.2
Testing a Class I Appliance ................... 20
3.3
Testing a Class II Appliance .................. 25
3.4
Testing a mains cord ............................. 28
3.5
Testing an extension lead...................... 29
3.6
Changing the Insulation Test Voltage ... 30
3.7
Changing the CLI Leakage limit ............ 30
3.8
Performing an RCD test on an installation
RCD ............................................................... 31
3.9
Performing an RCD test on a Plug top
RCD ............................................................... 32
3.10 Performing a 3-Phase Sequence .......... 33
4.1
Saving Test Results ............................... 35
4.2
Viewing Test Results ............................. 35
4.3
Deleting Test Results ............................. 35
4.4
Transferring Test Resuits ...................... 35
4.5
Printing Test Results ............................. 36
4.6
Setting the Real Time Clock .................. 36
7.1
Battery Replacement ............................. 41
Fuse Replacement ..................................... 42
Operating Instructions
5
Page 6
DLPT2+
Caution, risk of electric shock.
Indicates instructions must be followed to
Caution, risk of danger. The operating
If the
DLPT2+
is used in a manner not
Always ensure that the circuit or
The
DLPT2+
and all associated cables and
damage before equipment is operated. Do
Do not touch
test probes beyond the hand
Operating Instructions
1 Important Information
These operating instructions are intended for the use of adequately trained personnel.
The following symbols are used in these operating instructions and on the DLPT2+.
avoid danger to persons.
instructions must be adhered to in order to avoid danger.
Before use, ensure unit is clean and dry; visually inspect all leads, connectors, and case. Any damage or wear must be rectified prior to use.
specified by these operating instructions then the protection provided may be impaired.
appliance under test is electrically isolated.
leads must be checked for signs of
not use if there are signs of damage. Only use the correct leads supplied with the DLPT2+.
barrier on the test probe.
6
Page 7
DLPT2+
The
DLPT2+
may apply high voltage or
If the
DLPT2+
is being used to determine
Do not operate the
DLPT2+
in an
The
DLPT2+
has been designed to make
Do not open the
DLPT2+
, no user
mains power to the appliance under test. Do not touch conductive parts of the appliance while tests are active.
the presence or absence of hazardous voltages, always prove the operation of voltage measurement function before and after use by means of a known voltage source or proving unit.
explosive gas or dust environment.
measurements in a dry environment.
serviceable parts.
Operating Instructions
Note: Data may be lost or altered in virtually
any electronic memory under certain circumstances. Therefore DILOG Electronic assumes no responsibility for financial losses or claims due to data lost or otherwise rendered unusable whether as a result of abuse, improper use, defects, disregard of operating instructions or procedures, or any allied causes.
7
Page 8
DLPT2+
Part Number
Part Number
Standard Accessories
Operating Instructions
DILOG DLPT2+ Unit DLPT2+ Operating Instructions
Optional Accessories
403A592
TPA 5/16 Phase Adaptor TPA 5/32 Phase Adaptor TPA 4/16 Phase Adaptor TPA 4/32 Phase Adaptor NiMH Batteries and charger 339A950
PATGuard 3 software can also be trialled and purchased online from the below web address:
www.DILOG.co.uk/pg3trial
8
Page 9
DLPT2+
Operating Instructions
Figure 1. DLPT2+ Front View
9
Page 10
DLPT2+
Operating Instructions
Figure 2. DLPT2+ End View
Figure 3. Testing CLI appliances, no mains leakage
10
Page 11
DLPT2+
Operating Instructions
Figure 4. Testing CLII appliances, no mains leakage
11
Page 12
DLPT2+
Operating Instructions
Figure 5. Testing CLI appliances, mains leakage
12
Page 13
DLPT2+
Operating Instructions
Figure 6. Testing CLII appliances, mains leakage
13
Page 14
DLPT2+
Operating Instructions
Figure 7. Testing IEC Leads
14
Page 15
DLPT2+
Operating Instructions
Figure 8. Testing Installation RCD
15
Page 16
DLPT2+
Operating Instructions
Figure 9. Testing Plug Top RCD
16
Page 17
DLPT2+
2 Introduction
The DLPT2+ is a hand held battery powered unit suitable for carrying out electrical safety checks on:
Class I appliances
Class II appliances
IEC mains leads
Extension leads
30mA Plug-top RCDs
30mA Installation RCDs
Mains outlet wiring
3-Phase appliances when used with the DILOG
Overview
With reference to Figures 1 and 2.
1. Save Key
2. Recall Key
3. Print Key
4. CLI / Cord Sequence Key
5. CLII Test Sequence Key
6. Leakage Test Start Key
7. Setup Button
8. RCD Test Key
9. 3-Phase Test Sequence Key
10. LCD
11. Test/Mains outlet socket
12. USB Interface
13. IEC Test Connection
14. Mini DIN Printer Interface
15. Earth Continuity Test Terminal
16. Mains inlet socket
User Interface
The LCD display shows test progress, results for individual tests and the overall test result for an appliance or mains cord.
TPA range of 3-phase adaptors.
Operating Instructions
17
Page 18
DLPT2+
Operating Instructions
Power ON/OFF = press and hold CLI and CLII simultaneously
Class I appliance test = press the CLI key
Class II appliance test = press the CLI key
Cord / extension lead test = press the CLI key
Enter Setup Mode = press the SETUP key Once in the setup mode the CLII button can be used to the change the Insulation voltage while the LKGE button can be used to change the ILEAK pass/fail limit. To leave the setup mode press the SETUP button.
30mA RCD test = press the 30mA RCD key
3 phase appliance test = press the 3 phase key
Note: The DLPT2+ will automatically switch OFF after approximately 1 minute if no keys are pressed.
Note: When a key is pressed to initiate a test sequence, the DLPT2+ will compare the type of appliance connected with the test sequence that has been selected, and where possible, will prompt the user if an inappropriate test has been selected.
For example, if an IEC lead is connected between the front panel mains socket and the IEC socket on the end panel and a CLI appliance test is selected, the DLPT2+ will flash the Cord test enunciator to indicate that the Cord test is the most appropriate test.
18
Page 19
DLPT2+
LN
LE
NE
3 Performing Tests
Press CLI and CLII keys simultaneously to switch on the DLPT2+. When the unit is ready the display will be as shown below.
Operating Instructions
3.1 Checking a mains power outlet
Connect DLPT2+ mains lead to the mains inlet socket and to the power outlet to be tested.
Each time mains voltage is detected, the line­neutral, line-earth and neutral-earth voltages will be displayed for 2s as shown below.
If the voltages are within the acceptable limits shown below, a tick is shown next to the enunciator.
195V-253V 195V-253V
<30V
19
Page 20
DLPT2+
After 2s the display will return to the start up screen.
If the mains voltage is incorrect then the DLPT2+ will remain in voltage mode until the voltage is corrected or removed.
3.2 Testing a Class I Appliance
Visually inspect the appliance and mains cord
If the appliance passes a visual inspection
Plug the earth test lead into the Earth Continuity
Plug the appliance into the DLPT2+ front panel
Connect the earth test probe to an exposed
If the appliance under test has an ON/OFF
Press the CLI test key
The DLPT2+ will now perform a measurement of
If the measured value is higher than the factory
for signs of damage.
proceed with the electrical tests.
Test Terminal on the DLPT2+ end panel.
test/mains outlet socket.
metal part on the appliance.
switch, make sure it is in the ON position.
the protective earth continuity, reverse the test current and repeat the measurement. The highest measured value that occurred during the two tests is displayed.
set pass/fail threshold, the measured value is displayed and the unit indicates a fail result as shown below and the test sequence is halted.
Operating Instructions
20
Page 21
DLPT2+
If the measured value is less than the factory set
pass/fail threshold, the measured value is displayed and the unit indicates a pass result, as shown below.
Operating Instructions
Note: The DLPT2+ can perform a continuous Rpe
measurement (max 3 minutes). Press and hold the CLI key for approximately 5s until the unit emits a beep and the symbol appears on the display. The DLPT2+ continuously displays the current measured value on the display and stores the maximum value in the memory. By pressing the CLI key again, the measurement is repeated with reversed polarity. Press the key again to terminate the Rpe test and display the maximum value of RPE on the display.
The insulation test voltage is shown momentarily
next to the Riso enunciator.
The unit will proceed with the Insulation and
Leakage tests.
Note: The power switch on the appliance under
test must be in the ON position to perform an insulation test. If no appliance is detected the DLPT2+ will display the following.
21
Page 22
DLPT2+
Operating Instructions
Check that the appliance power switch is in the
ON position. The test will automatically proceed if the appliance power switch is placed in the ON position.
If the LO LOAd enunciator remains on the
display, the load presented by the appliance may be too small for the DLPT2+ to detect. In this case, press the CLI test key to continue.
If the Insulation Resistance is greater than the
factory set limit a tick is placed next to the Riso enunciator. If the Insulation Resistance is less than the set limit a cross is placed next to the Riso enunciator, the FAIL enunciator is displayed and the test sequence is halted.
Note: If a mains supply is detected, the DLPT2+
will perform a differential leakage test. If the DLPT2+ does not detect a mains supply then the DLPT2+ will perform a substitute leakage test.
If a substitute leakage test is performed then the
DLPT2+ will start the test without the need for user intervention. If the substitute leakage is less than the factory set limit a tick is placed next to the Iea enunciator.
22
Page 23
DLPT2+
LN LE NE Mains status
Operating Instructions
If a differential leakage test is performed then
the test DLPT2+ will prompt the user to start the Leakage test by flashing the Ileak icon. If no key is pressed the DLPT2+ will timeout and return to the start screen.
Note: If the LKGE key is pressed momentarily the
differential test duration is 5s. If the LKGE key is pressed and held for 2s the differential test duration is 2 minutes.
When the duration is set to 2 minutes, the test can be terminated at any time by pressing the LKGE key.
When the LKGE key is pressed to start the
differential leakage test the DLPT2+ will check the incoming mains supply to ensure that the voltages are correct. The enunciators are displayed if there is an error with the connected mains supply:
Flash Flash OFF No mains OFF Flash Flash Earth fault or Live
and Neutral reverse
If the mains voltage is correct, the DLPT2+ will
proceed with the differential leakage test.
23
Page 24
DLPT2+
Operating Instructions
If the Differential Leakage is less than the factory
set limit a tick is placed next to the Ileak enunciator.
If all of the tests within the sequence have
passed then the PASS enunciator is illuminated.
24
Page 25
DLPT2+
3.3 Testing a Class II Appliance
Visually inspect the appliance and mains cord
If the appliance passes a visual inspection
Plug the earth test lead into the Earth Continuity
Plug the appliance into the DLPT2+ front panel
Connect the earth test probe to an exposed
If the Appliance under test has an ON/OFF
Press the CLII test key
The display will momentarily show the insulation
for signs of damage.
proceed with the electrical tests.
Test Terminal on the DLPT2+ end panel.
test/mains outlet socket.
metal part on the appliance.
switch, make sure it is in the ON position.
test voltage before proceeding with the test.
Operating Instructions
Note: The power switch on the appliance under
test must be in the ON position to perform an insulation test. If no appliance is detected the DLPT2+ will display the following.
25
Page 26
DLPT2+
Check that the appliance power switch is in the
If the LO LOAd enunciator remains on the
The DLPT2+ will perform an Insulation test. If
Note: If a mains supply is detected, the DLPT2+
If a Substitute Leakage test is performed then
ON position. The test will automatically proceed if the appliance power switch is placed in the ON position.
display, the load presented by the appliance may be too small for the DLPT2+ to detect. In this case, press the CLII test key to continue.
the Insulation Resistance is greater than the factory set limit a tick is placed next to the Riso enunciator. If the Insulation Resistance is less than the set limit a cross is placed next to the Riso enunciator, the FAIL enunciator is displayed and the test sequence is halted.
will perform a touch leakage test. If the DLPT2+ does not detect a mains supply then the DLPT2+ will perform a substitute leakage test.
the DLPT2+ will start the test without the need for user intervention. If the Substitute Leakage is less than the factory set limit a tick is placed next to the Iea enunciator.
Operating Instructions
If a touch leakage test is performed then the
DLPT2+ will prompt the user to start the Leakage test by flashing the Ileak icon. If no key
26
Page 27
DLPT2+
LN LE NE Mains status
Note: If the LKGE key is pressed momentarily the
When the LKGE key is pressed to start the
If the mains voltage is correct, the DLPT2+ will
If the touch leakage is less than the factory set
If all of the tests within the sequence have
is pressed the DLPT2+ will timeout and return to the start screen.
touch leakage test duration is 5s If the LKGE key is pressed and held for 2s the touch leakage test duration is 2 minutes.
When the duration is set to 2 minutes, the test can be terminated at any time by pressing the LKGE key.
touch leakage test the DLPT2+ will check the incoming mains supply to ensure that the voltages are correct. The enunciators are displayed if there is an error with the connected mains supply:
Flash Flash OFF No mains OFF Flash Flash Earth fault or Live
proceed with the touch leakage test.
limit a tick is placed next to the Ileak enunciator.
passed then the PASS enunciator is illuminated.
Operating Instructions
and Neutral reverse
27
Page 28
DLPT2+
Operating Instructions
3.4 Testing a mains cord
Visually inspect the mains cord and plug for
signs of damage.
Check that the correct fuse is fitted.
If the cord passes a visual inspection proceed
with the electrical tests.
Plug the mains cord under test into the IEC
socket and the front panel test/mains socket on the DLPT2+.
Press the CLI test key
The DLPT2+ will first test the continuity of the
protective earth.
If the measured value is greater than the factory
set limit a cross is placed next to the RPE enunciator, a FAIL is indicated and the test sequence will be halted.
If the measured value is less than the factory set
limit a tick is placed next the RPE enunciator.
The unit will proceed with the Insulation test.
If the Insulation Resistance is lower than the
factory set limit a cross is placed next to the Riso enunciator and the test sequence is halted.
If the Insulation Resistance is greater than the
factory set limit a tick is placed next to the Riso enunciator. If the Insulation Resistance is less than the set limit a cross is placed next to the Riso enunciator, the FAIL enunciator is displayed and the test sequence is halted.
The unit will proceed with the wiring test,
checking the live and neutral conductors for short or open circuit and reverse connections
If the wiring is correct a tick is placed next to the
cord enunciator, the GOOd enunciator is illuminated and a PASS is indicated for the sequence.
28
Page 29
DLPT2+
Operating Instructions
Note: If the tested cord has a wiring fault, one of the following enunciators will be illuminated in place of GOOd
OPEn indicates that either the live or
neutral conductor is broken (open circuit)
Shor indicates that the live and neutral
conductors are shorted together
CrOS indicates that the live and neutral
conductors are reversed
3.5 Testing an extension lead
Visually inspect the mains cord and plug for
signs of damage.
Check that the correct fuse is fitted.
If the cord passes a visual inspection proceed
with the electrical tests.
Plug the supplied 0.5m IEC lead into the IEC
socket and into a mains outlet on the extension lead. Plug the mains plug of the extension lead into the front panel mains socket on the DLPT2+.
The extension lead can now be tested in the
same manner as an IEC as described above.
29
Page 30
DLPT2+
Operating Instructions
3.6 Changing the Insulation Test Voltage
The Insulation test performed in the CLI and CLII test sequences can be changed to 250V for appliances with MOVs which would fail an Insulation test performed at 500V. In order to change the Insulation voltage press the SETUP button to enter setup mode. Once in setup mode press the CLII button, the selected insulation voltage will be displayed next to the Riso enunciator. To leave setup mode press the SETUP button.
3.7 Changing the CLI Leakage limit
The CLI Leakage pass/fail limit can be changed between 0.75mA and 3.5mA. In order to change the Leakage limit press the SETUP button to enter setup mode. Once in setup mode press the LKGE button, the selected pass/fail limit will be displayed next to the Ileak enunciator. To leave setup mode press the SETUP button.
30
Page 31
DLPT2+
When performing an RCD test ensure that
Equipment which is connected
Voltages between the protective conductor
Leakage currents in the circuit may
Potential fields of other earthing
LN LE NE Mains status
3.8 Performing an RCD test on an installation RCD
Operating Instructions
The DLPT2+ can be used to measure the operating time of an installation RCD.
Connect the DLPT2+ IEC lead to the IEC socket of the DLPT2+ and a mains outlet on the RCD protected circuit.
To test the RCD operating time press the 30mA RCD test key.
The mains supply status is checked and the status is indicated using the LN, LE and NE enunciators of the LCD.
the Continuity Test probe is disconnected from the DLPT2+
downstream of the RCD may cause measurement inaccuracies.
and earth may influence measurements.
influence measurements.
installations may influence the measurement.
Flash Flash OFF No mains OFF Flash Flash Earth fault
31
Page 32
DLPT2+
When plugging the mains lead into the
When performing an RCD test ensure that
Whenever reSet is displayed on the LCD press the reset button on the RCD.
The DLPT2+ will perform a IΔn (30mA) 0o test, the operating time of the RCD will be shown on the DLPT2+ LCD. If the test passes then the DLPT2+ will perform a IΔn (30mA) 180o test. As long as the tests continue to pass the DLPT2+ will continue to test both 0o and 180o at IΔn (30mA).
Operating Instructions
3.9 Performing an RCD test on a Plug top RCD
The DLPT2+ can be used to measure the operating time of a plug top RCD.
DLPT2+ plus ensure that the polarity orientation of the connector is correct. DO NOT FORCE THE PLUG INTO THE CONNECTOR. Doing so may damage the DLPT2+.
the Continuity Test probe is disconnected from the DLPT2+
Connect the DLPT2+ mains lead to the mains inlet socket of the DLPT2+ and a mains outlet.
Plug the RCD into the DLPT2+. Plug the supplied IEC lead into the RCD and the DLPT2+ IEC socket.
To test the RCD operating time press the 30mA RCD test key.
The mains supply status is checked and the status is indicated using the LN, LE and NE enunciators of the LCD.
32
Page 33
DLPT2+
LN LE NE Mains status
Flash Flash OFF No mains OFF Flash Flash Earth fault
Whenever reSet is displayed on the LCD press the reset button on the RCD.
The DLPT2+ will perform a IΔn (30mA) 0o test, the operating time of the RCD will be shown on the DLPT2+ LCD. If the test passes then the DLPT2+ will perform a IΔn (30mA) 180o test. As long as the tests continue to pass the DLPT2+ will continue to test both 0o and 180o at IΔn (30mA).
3.10 Performing a 3-Phase Sequence
The DLPT2+ can be used to measure Rpe and earth leakage current on 3 phase equipment when used with the optional 3 phase adaptor accessory.
Visually inspect the appliance and mains cord
If the appliance passes a visual inspection
Refer to the documentation supplied with the 3
Connect the 3 phase adaptor test lead to the
Connect the Rpe test probe to an exposed
Press the 3 phase key on the DLPT2+ to being
If there is a contact voltage present on the
Otherwise, the DLPT2+ will proceed with the
for signs of damage.
proceed with the electrical tests.
phase adaptor and connect the adaptor in line with the 3 phase equipment under test.
mains input socket 9 (refer to figure 2) on the DLPT2+.
conductive part on the 3 phase equipment.
the test.
conductive parts of the equipment under test, the test is terminated and a Volt on test Probe warning message is shown on the display.
Rpe measurement, reverse the test current and then repeat the measurement. The highest
Operating Instructions
33
Page 34
DLPT2+
Note: There is no factory programmed test limit
The DLPT2+ will then display the current flowing
measured value that occurred during the two tests is displayed for approximately 2s.
for the 3 phase Rpe measurement and so no tick or cross is displayed.
in the protective conductor of the 3 phase supply.
Operating Instructions
The 3-Phase leakage test will run for 30 seconds
but can be stopped earlier by pressing the 3­Phase button.
Note: If the 3-Phase key is pressed while no
adaptor is connected to the DLPT2+ then a warning message PLUG IN 3 PH ADPTR will be displayed.
34
Page 35
DLPT2+
Operating Instructions
4 Memory & Print Functions
The DLPT2+ can store the results for a maximum of 999 appliances.
4.1 Saving Test Results
In order to store a test result press the STORE Key when the test sequence has been completed. The results displayed on the LCD will be recorded in memory. Once the results for a test sequence have been recorded the STORE Key is disabled until another test sequence has been performed.
4.2 Viewing Test Results
When the DLPT2+ is waiting to start a new test sequence press the RECALL Key, this will display the last recorded results. Other results can be displayed by scrolling through the memory. To scroll through the memory use the RECALL and STORE Keys. Pressing any other Key or waiting for 10 seconds will return the DLPT2+ into test mode.
4.3 Deleting Test Results
To delete all of the results in memory press and hold the STORE and RECALL keys together while displaying the recalled results until the counter reaches zero.
4.4 Transferring Test Results
Test data can be transferred to a PC using the Datalogger on the provided DLPT2+ Support CD. Insert the CD into your disc drive and then view the InstallationInstructions.txt file for information on how to install. Once the USB drivers and Datalogger are installed you can open the Datalogger program from your desktop. Connect your device to the PC using a USB connector. (The very first time you do this, your PC will automatically install a COM port driver). You can check which COM port is being used by looking in the Device manager.
35
Page 36
DLPT2+
E.g.: Silicon Labs CP210x USB to UART Bridge(COM14) Select the correct COM port. Click on Download. A progress bar will appear. On the DLPT2+ in order to transfer the entire test storage area to PC press and hold the RECALL key for 5 seconds. The data will be transmitted using the USB interface. When all data has been captured and stored the progress bar will disappear and you will be asked where to save the data. Once the file is saved you may click the "Open" button to open the document with default application. For example, if you saved the file in .csv format (the default) then it may open using a spreadsheet application.
4.5 Printing Test Results
In order to print a test label press the PRINT Key when the test sequence has been completed or when the test result is being reviewed. The printer must be attached before the PRINT Key is pressed. In order to save power the DLPT2+ will turn the printer on and off when required.
4.6 Setting the Real Time Clock
Setting time and date: Ensure that the DLPT2+ is powered off. Press and hold the RECALL key and then press simultaneously the class I and class II keys to power on the tester. The date format and time format is shown as follows: MM.DD = month (1-12). Day (1-31) YYYY = year HH.mm = hours (0-23).minutes (0-59) SS = seconds (0-59) Press the class I key to select a date field and time field. A flashing digit shows that this field can be set.
Operating Instructions
36
Page 37
DLPT2+
The save key and recall key are used to increase or decrease the value in each field, the second field is set to zero. Turn off the device to save the setting.
Operating Instructions
37
Page 38
DLPT2+
5 Specification
Earth Continuity
Insulation resistance
Substitute Leakage Current
Differential Leakage Current
Touch Leakage Current
3-Phase Leakage Current
Display Range 0.01ohms to 19.99ohms Measuring Range 0.05ohms to 19.99ohms Accuracy ± (5% + 2 digits) Test current 200mA minimum Test voltage 6V nominal
Display Range 0.01Mohms to 19.99Mohms Measuring Range 0.10Mohms to 19.99Mohms Accuracy ± (5% + 2 digits) Test voltage 250V / 500V Test current >1mA Test current <2mA into 2k
Display Range 0.10mA to 19.99mA Measuring Range 0.25mA to 19mA Accuracy ± (5% + 2 digits) Test voltage 40V rms, 50Hz AC Test current <10mA into 2k
Display Range 0.15mA to 19.99mA Measuring Range 0.25mA to 19mA Accuracy ± (5% + 2 digits) Test voltage mains voltage, 50Hz AC
Display Range 0.10mA to 3.5mA Measuring Range 0.10mA to 1.99mA Accuracy ± (5% + 2 digits) Test voltage mains voltage, 50Hz AC
Display Range 0.10mA to 9.99mA Measuring Range 0.25mA to 9.99mA Accuracy ± (5% + 2 digits) Test voltage mains voltage, 50Hz AC
Operating Instructions
38
Page 39
DLPT2+
Earth
Insulation
Leakage
Class I
Class II
Cord
3-Phase
Cord Test Earth continuity, insulation resistance as above. Check for Live and Neutral open circuit, short
RCD Test
Factory Set Pass/Fail limits
circuit or reversed polarity.
Display Range IΔn - 0ms to 500ms 5IΔn – 0ms to 40ms Measuring Range IΔn - 10ms to 500ms 5IΔn – 10ms to 40ms Accuracy ± 2ms Test Current IΔn – 30mA 5IΔn – 150mA Current Accuracy -0, +10%
Operating Instructions
Continuity
0.2 ohms 1.0Mohm 0.75 / 3.5 n/a 2.0Mohm 0.25mA
0.2 ohms 1.0Mohm n/a
0.2 ohms 1.0Mohm 3.5mA
Environmental rating IP Rating IP40 Operating temperature range 0°C to 40°C, without moisture condensation. Storage temperature range –25° to 65°.
Note: Batteries should be removed prior to storage.
Overvoltage category 300V CAT II
Resistance
39
Page 40
DLPT2+
6 Maintenance
Operating Instructions
Clean only with a dry cloth; do not use solvents. Before use, ensure unit is clean and dry; visually inspect all leads, connectors, and case. Any damage or wear must be rectified to preserve user safety.
Check the battery contacts and compartment are free of electrolytic contamination. Any contamination of the battery contacts or compartment should be cleaned with a dry cloth.
Note: The DLPT2+ contains no user serviceable parts. If an Error warning should appear on the display please contact the manufacturer or an
authorised DILOG Service Agent for advise.
7 Battery Check
The DLPT2+ is powered from 6 AA cells which are checked before a test is performed. When the battery voltage is low the enunciator is illuminated. The unit will continue to perform within specification for a limited number of tests, dependent upon the type of the batteries fitted.
When the battery voltage reaches a level where the performance is affected the enunciator will flash and all test keys are disabled. The batteries must be replaced.
40
Page 41
DLPT2+
7.1 Battery Replacement
Before opening the DLPT2+ ensure that all test leads are disconnected.
Switch off the unit by pressing and holding
the CLI and CLII keys.
Disconnect all leads from the DLPT2+.
Place the DLPT2+ face down and release
the captive screw in the battery compartment cover.
Remove the battery compartment cover and
remove the discharged batteries.
Insert the replacement batteries into the
battery compartment ensuring that the battery polarity matches the marking on the inside of the battery compartment.
Relocate the battery cover over the battery
compartment and fasten in position with the battery cover captive screw.
Operating Instructions
41
Page 42
DLPT2+
Part Number
8 Fuse Replacement
Replacement F16A 250V Fuse 27B114
Before opening the DLPT2+ ensure that all test leads are disconnected.
Switch off the unit by pressing and holding
the CLI and CLII keys.
Disconnect all leads from the DLPT2+.
Place the DLPT2+ face down and release
the captive screw in the battery compartment cover.
Remove the battery compartment cover and
remove the discharged batteries.
Insert the replacement fuse into the fuse
compartment ensuring that the fuse snaps into place.
Relocate the battery cover over the battery
compartment and fasten in position with the battery cover captive screw.
Operating Instructions
42
Page 43
DLPT2+
9 Service and Calibration
Operating Instructions
To maintain the specified accuracy of the measurement results, the instrument must be recalibrated at regular intervals by either the manufacturer or an authorised service agent. We recommend a recalibration period of one year. For help or advice on Service and Calibration contact:
Di-LOG Test Equipment 28 Wheel Forge Way Manchester M17 1EH
Tel: 0161 877 0322
www.dilog.co.uk
sales@dilog.co.uk
43
Loading...