Di-LOG DL6780 Combi Volt 1, DL6790 Combi Volt 2 Instruction Manual

Di-LOG
Instruction Manual
DL6780 Combi Volt 1™ DL6790 Combi Volt 2™
Contents
1) Introduction
2) Safety notices
3) Instrument layout
Carrying out Measurements
4) Voltage test AC/DC
5) Single pole voltage indication
6) Continuity test
7) Phase rotation test
8) Maintenance
9) Specification
1) Introduction
Thank you for purchasing a Di-Log voltage continuity tester. This tester has been designed in accordance with the latest international safety standards. The combivolt testers are fully automatic voltage indicators capable of measuring AC/DC voltage up to 690 V. Both units have visual and acoustic continuity indication
Constructed in accordance with IEC 61010 and IEC 61243-3.
• Single pole phase indication
• 2 pole phase rotation indication
• LED& LCD display (DL6790)
2) Safety notices
This manual contains information that must be followed for operating the meter safely and maintaining the meter in a safe operating condition. If this meter is not used in the manner specified, the protection provided may be impaired.
Warning! Warns of potential danger, refer to the instruction manual to avoid personal injury or damage to the meter.
Caution! Dangerous voltage. Danger of electrical shock
Continuous double or reinforced insulation
complies with IEC536, class 11
CE Symbol of conformity, confirms conformity with relevant EU directives. The meter complies with EMC directives (89/336/EEC). Specifically standards EN 50081-1 and EN 50082-1 as well as the Low Voltage Directive (73/23/EEC) described in the standard EN 61010-1.
The meter has been designed in accordance with the safety regulations for electronic measuring instruments, EN 61010-1, IEC 61010
Voltages above 75V DC or 50V AC may constitute a serious shock hazard.
Before using the meter check for physical damage to the casing in particular around the connectors. If the case is damaged do not use the meter.
Check the test probes for damaged insulation or exposed metal. Check the leads for continuity.
Do not apply more than the rated voltage, as marked on the meter between the terminals or between any terminal and ground.
Do not use or store the meter in an environment of high temperature, humidity, fumes, vapour, gaseous, inflammable and strong magnetic field. The performance and safety of the instrument and the user may be compromised in such circumstances.
Disconnect circuit power and discharge all high voltage capacitors before testing resistance, continuity and diodes.
Remove the batteries if the meter is not in use for a long period. Constantly check the battery as it may have leaked. A leaking battery will damage the meter.
The meter may only be opened by a qualified service technician for calibration and repair.
3) Instrument Layout
1) Test Probe (- ) L1
2) Test Probe (+) L2
3) LED’s for voltage indication
4) LED for single - pole test
5) Right & Left LED, phase rotation indication
6) LED for continuity
7) LCD for voltage display (only DL6790)
8) Contact electrode for double-pole test of phase
rotation and single-pole test
9) Torch button on the back
10) Positive LED
11) Negative LED
12) Battery Compartment
Carrying out measurements
Perform a self test of the unit. Connect the two test probes L1 and L2. The continuity LED (6) will be lit and an audible tone should be heard.
Before any test check the unit on a known voltage source.
If the unit is defective it should be put out of service and returned to Di-Log for repair.
4) Voltage test
Always hold the test probes by the handles behind the finger guards.. Observe the safety notices at all times.
An audible tone is present when an AC voltage and a negative DC voltage are indicated.
The maximum switch on time is 30 s. When this time has elapsed you must wait 10 minutes before retesting.
Connect probes to voltage source observing polarity of the test probes L2 is positive probe, L1 is the negative probe.
For AC voltage the value is indicated on the LEDs (3) and on the LCD display (DL6790 only). The + and – LEDs are illuminated and buzzer is audible.
For DC voltage connect probe L2 to the positive terminal and L1 to the negative terminal. The voltage is displayed on the LEDs and the LCD display (DL6790 only). The positive LED (10) is illuminated. If the polarity is reversed the buzzer will sound. The negative LED (11) will be illuminated.
5) Single pole voltage detection
Perform a function test prior to this test.
This unit can be used as a single pole voltage detector when batteries are inserted.
The single pole test is intended only as a quick check. The circuit must be checked again for the presence of voltage using the two pole method.
Connect test probe L2 to the voltage source and keep finger on the contact electrode (8). If an AC voltage above 100 V is present the LED (4) is illuminated and the buzzer sounds.
The single pole test can be negatively affected by unfavourable conditions such as electrostatic field, good insulation etc.
6) Continuity test
The continuity test is only possible when batteries are inserted and in good condition.
Ensure the circuit under test is not live.
Connect test probes L1 and L2 to the circuit. The continuity LED (6) will illuminate and the buzzer will sound.
The unit will indicate continuity below 400 Kohm
7) Phase rotation test
Perform a function test prior to this test.
This unit can determine the phase rotation within a three phase supply.
Connect test probe L2 to the supposed phase 2 and the test probe L1 to the supposed phase 1. If the R LED illuminates the phases are in the correct sequence 1 to 2.
Connect test probe L2 to the supposed phase 3 and test probe L1 to the supposed phase 2. If the R LED illuminates the phases are in the correct sequence 2 to 3. Connect test probe L2 to the supposed phase 1 and the test probe L1 to the supposed phase 3. If the R LED illuminates the phases are in the correct sequence 3 to 1.
During phase rotation test touch the contact electrode.
If the L LED illuminates then the phase sequence is anti clockwise.
8) Maintenance
Do not attempt to repair this unit . There are no user serviceable items in this unit. Never attempt to open the casing apart from the battery cover.
Do not use the instrument if there is any physical damage to the case or test leads.
The outside of the unit can be cleaned with a soft damp cloth only. Do not use abrasive or chemical cleaning agents.
Changing batteries
Turn the battery cover by 90 degrees anti clockwise. Remove the cover and take out spent batteries. Replace with 2 off 1.5 V AAA (LR03) batteries, checking the correct polarity.
Replace the battery cover and turn by 90
0
clockwise.
Spent batteries should be disposed of responsibly and with compliance with current recycling and disposal regulations.
Calibration
The recommended calibration interval for the DL6780/DL6790 is 12 months.
9) Specifications
24 Month Warranty
Di-Log instruments are subject to stringent quality controls. If in the course of normal daily use a fault occurs we will provide a 24 month warranty (only valid with invoice). Faults in manufacture and materials defect will be rectified by us free of charge, provided the instrument has not been tampered with and returned to us unopened. Damage due to dropping, abuse or misuse is not covered by the warranty.
Outside the warranty period we offer a full repair and re-calibration service.
Di-Log Test Equipment 28 Wheel Forge Way Traf ford Park Manchester M17 1EH UK Tel: +44 161 877 0322 Fax +44 161 877 1614 Website: www.dilog.co.uk Email: sales@dilog.co.uk
DL6780V1
Voltage DL6780 DL6790
Voltage display 12 – 690 V AC/DC 6 – 690 V AC/DC LED Resolution 12, 24, 50, 120, 230, 12, 24, 50, 120, 230,
400, 690 400, 690 LCD Resolution 1 V +/- 3 % + 8 digits Voltage detection automatic automatic Acoustic signal AC voltage – DC AC voltage – DC voltage
voltage Polarity detection Full range Full range Response time LED < 0.1s LED <0. 1s, LCD <2s Frequency range 0 – 400 Hz 0 – 400 Hz Peak Current Is<0.3A/Is(after 5s) Is<0.3A/Is(after 5s) <3.5 mA
<3.5 mA Operation time 30s 30s Recovery time 10 min 10 min Auto power on <12 V AC/DC <12 V AC/DC
Single pole voltage indication
Voltage range 100 – 690 V AC 100 – 690 V AC Frequency range 50 – 400 Hz 50 – 400 Hz
Continuity
Measuring range 0 - 400 Kohm 0 - 400 Kohm
Phase rotation test
Voltage range 100 – 690 V 100 – 690 V Frequency 45 – 65 Hz 45 – 65 Hz Over voltage 690 V AC/DC 690 V AC/DC protection Power supply 2 x 1.5 LR03 battery 2 x 1.5 LR03 battery Dimensions 255 x 70 x 28 mm 255 x 70 x 28 mm Weight 200g 200g