
9083P
17th Edition and Part P Multifunction Tester

328A561 revision1.1 - 2 -
Figure 1 9083P Front View
Figure 2 9083P End View

328A561 revision1.1 - 3 -
Figure 3 Earth Continuity/Insulation Measurement
Figure 4 Earth Continuity/Insulation Measurement using
Cordless probe

328A561 revision1.1 - 4 -
Figure 5 Voltage Measurement using test probes
Figure 6 Earth Loop Impedance/Line Impedance/ RCD/
Mains Voltage Measurement

328A561 revision1.1 - 5 -
Figure 7 Earth Loop Impedance/Line Impedance
Figure 8 Line Impedance between phases

328A561 revision1.1 - 6 -
Limited Warranty & Limitation of Liability
DI-LOG Test Equipment guarantees this product to be
free from defects in material and workmanship under
normal use and service for a period of 2 year. The
period of warranty will be effective at the day of
delivery.
(c) Copyright 2007
All rights reserved. Nothing from this edition may be
multiplied, or made public in any form or manner, either
electronically, mechanically, by photocopying,
recording, or in any manner, without prior written
consent from DI-LOG. This also applies to
accompanying drawings and diagrams.
Due to a policy of continuous development DI-LOG
reserves the right to alter the equipment specification
and description outlined in this publication without prior
notice and no part of this publication shall be deemed
to be part of any contract for the equipment unless
specifically referred to as an inclusion within such
contract.

328A561 revision1.1 - 7 -
Table of Contents
Limited Warranty & Limitation of Liability.....................6
Table of Contents........................................................7
CERTIFICATE OF CONFORMITY.................................8
Introduction.................................................................9
1 Important Information..............................................9
2 Safety Information ..................................................10
3 Accessories............................................................11
3.1 Standard Accessories......................................11
3.2 Optional Accessories.......................................11
4 Unit Description......................................................12
4.1 Identifying parts of the unit.........................12
4.2 LCD display.....................................................13
5 Using the 9083P.....................................................15
5.1 Power On........................................................15
5.2 Battery Health Check.......................................15
5.3 Continuity Tests.........................................15
5.4 Insulation Resistance Tests........................17
5.5 Voltage Measurement......................................18
5.6 Earth Loop Impedance / Line Impedance ........18
5.7 Auto RCD Test Sequence................................19
5.8 RCD Trip Time Tests........................................21
5.9 RCD trip current (Ramp) Tests.........................23
6 Electrical Specifications..........................................25
6.1 Earth Continuity...............................................25
6.2 Insulation Resistance.......................................25
6.3 Earth Loop Impedance ....................................25
6.4 Line Impedance...............................................26
6.5 RCD ................................................................26
6.6 Voltage/Frequency Measurement....................26
7 Environmental Conditions.......................................27
8 Maintenance...........................................................28
8.1 Preparing to work on the 9083P.......................28
8.2 Securing the 9083P.........................................28
8.3 Cleaning..........................................................29
8.4 Battery Replacement.......................................29
8.5 Replacing the Fuse..........................................30
8.6 Service and Calibration....................................30
8.7 Spare Parts......................................................31

328A561 revision1.1 - 8 -
CERTIFICATE OF CONFORMITY
As the supplier of the apparatus listed, declare under
our sole responsibility that the product:
9083P Multifunction Tester
To which this declaration relates are in conformity with
the relevant clauses of the following standard:
BS EN 61010-1:2001
Safety requirements for electrical equipment for
measurement, control, and laboratory use – Part 1:
General requirements.
BS EN 61557-1,2,3,4,6,10:1997
Electrical safety in low voltage distribution systems
up to 1000V a.c. and 1500V d.c. – Equipment for
testing, measuring and monitoring of protective
measures
BS EN 61326:1998
Electrical equipment for measurement, control and
laboratory user-EMC Requirements
Performance: The instrument operates within
specification when used under the conditions in the
above standards EMC and Safety Standards.
The product identified above conforms to the
requirements of Council Directive 89/336/EEC and
73/23 EEC.

328A561 revision1.1 - 9 -
Introduction
The 9083P is a hand held, battery powered, multifunction electrical installation test instrument capable of
performing a comprehensive range of tests, including:
Earth Continuity @ 200mA
Insulation Resistance at 100V, 250V and 500V
Voltage
Frequency
Auto RCD trip time sequence
RCD Trip Time at ½I∆n, I∆n and 5xI∆n
RCD Trip current
Earth Loop Impedance
Prospective Fault Current
Mains Impedance
Prospective Short-circuit Current
Power socket wiring polarity
1 Important Information
These operating instructions are intended for the use of
adequately trained personnel.
The following symbols are used in these operating
instructions and on the 9083P.
Caution, risk of electric shock.
Instructions must be followed to avoid risk of
electric shock.
Caution, risk of danger. This symbol indicates
that the operating instructions must be adhered
to in order to avoid danger.

328A561 revision1.1 - 10 -
2 Safety Information
This 9083P is fully compliant with the requirements of:
BS EN 61010-1: 2001.
BS EN 61557 part 1, 2, 3, 4, 6 and 10.
In order to ensure safe operation of this instrument, all
notes and warnings must be observed at all times.
The 9083P has been designed to make
measurements in a dry environment. Do not
operate the 9083P in an explosive gas or dust
environment.
The 9083P may be used to test circuits with a
maximum over-voltage Category III, 300 V
AC/DC with reference to earth.
Always hold test probes above the hand guards.
The 9083P and all associated cables and leads
must be checked for signs of damage before
equipment is operated.
Prior to any resistance measurement, always
ensure that the circuit under test is electrically
isolated.
Where safe operation of the 9083P is no longer possible
it should be immediately shut down and secured to
prevent accidental operation.
It must be assumed that safe operation is no longer
possible:
- if the instrument or leads show visible signs of
damage or
- the instrument does not function or
- after long periods of storage under adverse
environmental conditions.
If the 9083P is used in a manner not specified by
this document then the protection provided by
the equipment may be impaired.

328A561 revision1.1 - 11 -
3 Accessories
3.1 Standard Accessories
The 9083P is supplied with the following items:
1 off 9083P unit
1 off professional carry case
1 off Di-LOG mains test lead
1 off 1.2 M black test lead
1 off 1.2 M red test lead
1 off 1.2 M green test lead
1 off Operating Instruction Manual
1 off Leadless probe
1 off black crocodile clip
1 off red crocodile clip
1 off green crocodile clip
3.2 Optional Accessories
3.15A T 500V Fuse
Do not open unit, no other serviceable parts.

328A561 revision1.1 - 12 -
4 Unit Description
The 9083P is a hand held, multi-function electrical
installation test instrument, capable of performing all of
the required electrical tests. Tests are selected using
the colour coded rotary switch.
4.1 Identifying parts of the unit
The numbering below refers to figure 1 and figure 2.
1. LCD Display
2. Function keys F1, F2, F3 and F4
3. TEST key
4. Rotary Switch
a. Voltage/frequency
b. Insulation resistance @ 100V
c. Insulation resistance @ 250V
d. Insulation resistance @ 500V
e. Continuity @ 200mA
f. Off
g. Earth Loop / Mains Impedance
h. Auto RCD
i. RCD trip time @ ½ I∆N
j. RCD trip time @ I∆N
k. RCD trip time @ 5I∆N
l. RCD trip time (ramp test)
5. Test lead input (RED)
6. Test lead input (BLACK
7. Test lead input (GREEN)
Note: The function performed by keys F1 – F4
depends upon the rotary switch position.
For each rotary switch position, the lower
part of the LCD display indicates the
function of the key above.

328A561 revision1.1 - 13 -
4.2 LCD display
1. Mains supply status icons
These icons indicate the status of the mains supply
between phase-earth (PE), phase-neutral (PN) and
phase-earth (PE) during RCD and Loop tests.
Note: Testing is inhibited if the mains supply is
incorrect.
2. Analog bargraph
The analogue bargraph provides an analogue
representation of the measurement shown by the
primary digital display.
3. Primary digital display
The primary digital display is used to indicate the test
measurement for the active test type.
4. Icons for function key F1.
These icons are used to display the available options for
the selected test. Repeatedly pressing function key F1
cycles through the available options.
5. Icons for function key F2.
These icons are used to display the available options for
the selected test. Repeatedly pressing function key F2
cycles through the available options.
6. Icons for function key F3.
These icons are used to display the available options for
the selected test. Repeatedly pressing function key F3
cycles through the available options.
7. Secondary display for function key F4.
The secondary display is used to show the measured
output voltage during insulation resistance

328A561 revision1.1 - 14 -
measurements, PFC/PSC calculations during earth loop
/ line impedance measurements and rated residual
operating current for RCD tests.
8. Battery status icon.
This icon displays the status of the battery voltage.
9. RCD test icons.
These icons display the selected RCD test function.
10. Warning Icons.
These icons are used to inform the user of the potential
of any hazard or warning which may restrict the
operation of the 9083P.
Note: The 9083P is equipped with a thermal cut-out
to protect against damage from excessive
internal temperature. If the thermal cut-out is
activated the HOT icon is illuminated and all
RCD and Loop tests are inhibited until the unit
cools and the thermal cut-out self-resets.

328A561 revision1.1 - 15 -
5 Using the 9083P
5.1 Power On
To turn the 9083P on simply rotate the rotary switch to
the required test type.
5.2 Battery Health Check
The 9083P will automatically perform battery health
checks periodically or when a new test type is selected.
Note: When the battery symbol is flashing all tests
will be inhibited and the batteries should be
replaced as described in section 8.4.
5.3 Continuity Tests
Always ensure that the circuit under test is
electrically isolated before performing continuity
tests.
If the test probes are connected across a
voltage of >30V then the 9083P will
automatically display the voltage between the
probes, the warning buzzer will sound and the
TEST key is inhibited.
Rotate the rotary switch until the Ω test is selected.
When the continuity test is selected, the 9083P will
display the available test options for 1 second; Buzzer,
Test Lock, Lead Zero. If the Buzzer was previously
enabled then the icon will remain highlighted. The Test
Lock and Lead Zero must be enabled each time the
Continuity test is selected.
Functions keys F1-F4 have are used to select the
options below:
F1 F2 F3 F4
Buzzer Test Lock ZERO Not used
Buzzer (F1)
When enabled, the Buzzer will sound when the
continuity measurement is less than 20 ohms.

328A561 revision1.1 - 16 -
Test Lock (F2)
The Test Lock is used to lock! the instrument in a
continuous measurement mode, with a single press of
the TEST key. When Test Lock is enabled the LCD
shows the padlock icon. When Test Lock is active the
TEST key is locked until the option is disabled or the
rotary switch is moved to another position.
To enable the Test Lock mode test press the TEST key
and F2 simultaneously.
To disable Test Lock, press F2 or turn the rotary switch
to another position.
ZERO (F3)
The instrument can automatically compensate for the
resistance of the test leads as follows:
Connect the two test probes firmly together, then press
and hold the ZERO key. The measured resistance of the
test leads is shown in the primary display until a beep is
heard and the ZERO icon is illuminated. All subsequent
measurements will take into account the test lead
resistance compensation until the function is disabled
by pressing function key F3.
Note: A maximum test lead resistance of 10 ohms
can be nulled out. If the test lead resistance is
great than 10 ohms an error beep will indicate
that the Lead Zero function has failed.
Note: For ease of use, the 9083P will store the lead
compensation when the instrument is
switched off and recall this value when next
switched on. The stored value is only
applicable to the test leads used when the
compensation measurement was made. If the
test leads are replaced the ZERO function
should be repeated using the replacement test
leads.
To make a continuity measurement, press and hold the
TEST key. The resistance between the test probes is
displayed until the TEST key is released. Alternatively,
use the Test Lock function to allow measurements to be
started or stopped with single press of the TEST key.
During a measurement, the measured value is shown in
the primary display and on the analogue bargraph.

328A561 revision1.1 - 17 -
5.4 Insulation Resistance Tests
Always ensure that the circuit under test is
electrically isolated.
If the test probes are connected across a
voltage of >30V then the 9083P will
automatically display the voltage between the
probes, the warning buzzer will sound and the
TEST key is inhibited.
Use the rotary switch to select either the 100V, 250V or
500V M" test.
The 9083P will display the Test Lock and battery
symbol for 1 second. If the Test Lock feature is
required, it should be activated as described below.
Functions keys F1-F4 have are used to select the
options below:
F1 F2 F3 F4
Not used
Test Lock
Not used Not used
Test Lock (F2)
The Test Lock is used to lock! the instrument in a
continuous measurement mode, with a single press of
the TEST key. When Test Lock is enabled the LCD
shows the padlock icon. When Test Lock is active the
TEST key is locked until the option is disabled or the
rotary switch is moved to another position.
To enable the Test Lock mode test press the TEST key
and F2 simultaneously.
To disable Test Lock, press F2 or turn the rotary switch
to another position.
To make an insulation resistance measurement, press
and hold the TEST key. The resistance between the test
probes is displayed until the TEST key is released.
Alternatively, use the Test Lock function to allow
measurements to be started or stopped with single
press of the TEST key. During a measurement, the
measured value is shown in the primary display and
analogue bargraph and the measured test voltage is
shown in the secondary display.

328A561 revision1.1 - 18 -
5.5 Voltage Measurement
Rotate the rotary switch until the VOLTS test is
selected. The 9083P will start measuring voltage
immediately and use of the TEST key is not required.
The function keys do not perform any operations while
in Voltage mode.
For AC voltage measurements, the frequency of the
measured voltage is shown in the secondary display.
5.6 Earth Loop Impedance / Line Impedance
The 9083P will only allow the Earth Loop
Impedance test to be performed if the correct
voltages are detected between phase-neutral
(PN illuminated), phase-earth (PE illuminated)
and neutral-earth (NE not illuminated).
Rotate the rotary switch until the Zs test is selected.
Press the function key F1 to select either the Earth
Loop Impedance test (L-PE) or the Line Impedance test
(L-N).
F1 F2 F3 F4
Line
Impedance
Not used Not used Not used
The secondary display will automatically show the
appropriate icon: PFC when L-PE is selected and PSC
when L-N is selected.
Note: When the Zs switch position is selected, the
9083P will default to the last used setting,
even if it has been switched off.
To begin the test, press and release the TEST key.
Note: The 9083P will determine the fault voltage that
may appear on the protective conductor
during the test. If the fault voltage is greater
than 25V the 9083P will indicate >25V on the
LCD, but the user may proceed with the test. If
the fault voltage is great than 50V, this is
indicated on the LCD and the test is inhibited.

328A561 revision1.1 - 19 -
During the test, progress is shown by the analogue
bargraph. When the test is complete, the Earth Loop
Impedance measurement is shown in the primary
display and the calculated Prospective Fault Current
(PFC) is shown in the secondary display.
Note: A Line Impedance measurement is
automatically made as part of the Earth Loop
Impedance test. The Line Impedance
measurement (L-N) and Prospective Shortcircuit Current (PSC) can be viewed by simply
pressing function key F1, without the need to
repeat the test.
5.7 Auto RCD Test Sequence
The 9083P will only allow the Earth Loop
Impedance test to be performed if the correct
voltages are detected between phase-neutral
(PN illuminated), phase-earth (PE illuminated)
and neutral-earth (NE not illuminated).
Leakage currents or voltage on the protective
conductor may influence the measurement.
The Auto RCD test is used to automatically perform a
sequence of 6 RCD trip time tests with a single press of
the TEST key. Each time the RCD trips, the sequence
will automatically continue once the RCD is reset. The
sequence comprises of test at:
½I∆n / 0°
½I∆n / 180°
Ι∆n / 0°
Ι∆n / 180°
5I∆n / 0°
5I∆n / 180°
Functions keys F1-F4 have are used to select the
options below:
F1 F2 F3 F4
Not used AC/DC/
Selective
RCL
I∆n

328A561 revision1.1 - 20 -
AC/DC/Selective (F2)
Function key F2 is used to select the required RCD
type: AC or DC sensitive combined with standard or
selective RCD types. Each time the F2 key is pressed
the next option is selected
During selective tests the 9083P will display a delay
timer which counts down from 30s to 0s. Pressing the
Test key or turning the rotary switch while the 9083P is
counting will terminate the count.
ac current / standard RCD
ac current / selective RCD
dc current / standard RCD
dc current / selective RCD
RCL (F3)
The RCL (Recall) key is used to recall the results of the
last automatic RCD test performed by the 9083P. The
LCD will update to show all of the relevant parameters
for the result displayed. Continue to press the F3 key to
rotate through the results.
½ I#n test current / 0o
½ I$n test current / 180o
I%n test current / 0o
I&n test current / 180o
5x I'n test current / 0o
5x I(n test current / 180o
normal pre-test screen
Pressing any key while displaying a recalled
measurement will return the 9083P to the normal pretest screen.
Rated residual operating current I∆n (F4)
The rated residual operating current is selected by
pressing the F4 key.

328A561 revision1.1 - 21 -
10mA
30mA
100mA
When the required settings have been selected, press
the TEST key to begin the sequence.
Note: The 9083P will determine the fault voltage that
may appear on the protective conductor
during the test. If the fault voltage is greater
than 25V the 9083P will indicate >25V on the
LCD, but the user may proceed with the test. If
the fault voltage is great than 50V, this is
indicated on the LCD and the test is inhibited.
If the fault voltage is less than 50V the test sequence
will proceed and the trip times are shown in the primary
display.
When the sequence is completed, the RCL key (F3) is
used to recall the measurements.
5.8 RCD Trip Time Tests
The 9083P will only allow the Earth Loop
Impedance test to be performed if the correct
voltages are detected between phase-neutral
(PN illuminated), phase-earth (PE illuminated)
and neutral-earth (NE not illuminated).
Leakage currents or voltage on the protective
conductor may influence the measurement.
Note: The 9083P will determine the fault voltage that
may appear on the protective conductor
during the test. If the fault voltage is greater
than 25V the 9083P will indicate >25V on the
LCD, but the user may proceed with the test. If
the fault voltage is great than 50V, this is
indicated on the LCD and the test is inhibited.
Use the rotary switch to select the ½ I∆n, I∆n or 5 I∆n
test position.

328A561 revision1.1 - 22 -
The ½ I∆n test will perform the RCD test with a current
of 50% of that indicated on the LCD.
The I∆n test will perform the RCD with the test current
indicated on the LCD.
The 5 I∆n test will perform the RCD with 500% that
which is indicated on the LCD.
F1 F2 F3 F4
0o / 180o AC/DC/Selective Not used
I∆n
0o / 180
o
Use the F1 key to alternate between the starting angle
of the current. All RCD tests will start on zero crossing.
AC/DC/Selective
Function key F2 is used to select the required RCD
type: AC or DC sensitive combined with standard or
selective RCD types. Each time the F2 key is pressed
the next option is selected
During selective tests the 9083P will display a delay
timer which counts down from 30s to 0s. Pressing the
Test key or turning the rotary switch while the 9083P is
counting will terminate the count.
ac current / standard RCD
ac current / selective RCD
dc current / standard RCD
dc current / selective RCD
Rated residual operating current I∆n (F4)
The rated residual operating current is selected by
pressing the F4 key.

328A561 revision1.1 - 23 -
10mA
30mA
100mA
300mA
500mA
Note: The available settings for rated residual
operating current are restricted for the 5I∆n
test as shown below.
10mA 30mA 100mA 300mA 500mA
½ I N
a a a a a
1 I N
a a a a a
5 I N
a a a
5.9 RCD trip current (Ramp) Tests
The 9083P will only allow the Earth Loop
Impedance test to be performed if the correct
voltages are detected between phase-neutral
(PN illuminated), phase-earth (PE illuminated)
and neutral-earth (NE not illuminated).
Leakage currents or voltage on the protective
conductor may influence the measurement.
Note: The 9083P will first perform a short pre-test to
determine the fault voltage that may appear
on the protective conductor during the test. If
the fault voltage is greater than 50V then the
9083P will indicate >50V on the LCD, the test
will be disabled.
F1 F2 F3 F4
0o/180o AC/DC/Selective Not used Test

328A561 revision1.1 - 24 -
AC/DC/Selective
Function key F2 is used to select the required RCD
type: AC or DC sensitive combined with standard or
selective RCD types. Each time the F2 key is pressed
the next option is selected
During selective tests the 9083P will display a delay
timer which counts down from 30s to 0s. Pressing the
Test key or turning the rotary switch while the 9083P is
counting will terminate the count.
Rated residual operating current I∆n (F4)
The rated residual operating current is selected by
pressing the F4 key.
10mA
30mA
100mA
300mA
500mA
0o / 180o (F1)
Use the F1 key to alternate between the starting angle
of the current. All RCD tests will start on zero crossing.

328A561 revision1.1 - 25 -
6 Electrical Specifications
6.1 Earth Continuity
Test Voltage Open Circuit >4V
Test Current
>200mA into 2Ω
Display Range
0.00Ω - 199Ω
Measuring Range
0.01Ω ) 1.99Ω
2.0Ω ) 19.9Ω
20Ω ) 199Ω
Resolution
0.01Ω maximum
Accuracy ±2% ±5 digits
Number of repeat tests Approx 3000
6.2 Insulation Resistance
Test Voltage Accuracy -0% +20% (open circuit)
Test current
>1mA into UN x (1000Ω/V)
Test Current Short Circuit <2mA
Display Range
0.05MΩ ) 199MΩ
Measuring Range
0.05MΩ ) 1.99MΩ
2.0MΩ ) 19.9MΩ
20MΩ ) 100MΩ
Resolution
0.01MΩ maximum
Accuracy ±5% ±5 digits
Voltage Indication
Accuracy
±5%
Number of repeat tests Approx 3000
6.3 Earth Loop Impedance
Supply Voltage 195V ) 253V, 45Hz ) 65Hz
Nominal Test Current 15mA
Display Range
0.01Ω - 2000Ω
Measuring Range
1.00Ω - 1.99Ω
2.0Ω ) 19.9Ω
20Ω - 2000 Ω
Resolution
0.01Ω maximum
Accuracy
±5%±12d (1.00Ω - 1.99Ω)
±5%±5d (2.0Ω - 19.9Ω)
PFC Range 0kA ) 26kA

328A561 revision1.1 - 26 -
6.4 Line Impedance
Supply Voltage 195V ) 440V, 45Hz ) 65Hz
Nominal Test Current 15mA
Display Range
0.01Ω - 2000Ω
Measuring Range
0.05Ω - 1.99Ω
2.0Ω ) 19.9Ω
20Ω - 2000Ω
Resolution
0.01Ω maximum
Accuracy ±5% ±2 digits
PFC Range 0kA ) 26kA
6.5 RCD
Supply Voltage 195V ) 253V, 45Hz ) 65Hz
Nominal Test Currents 10mA,30mA,100mA,300mA,
500mA
Test Current Accuracy
-0% +10% at I∆n and 5I∆n
-10% +0% at ½ I∆n
Trip Time Ranges
0ms ) 2000ms, ½ I∆n
0ms ) 300ms, I∆n General
0ms ) 500ms, I∆n Selective
0ms ) 40ms, 5I∆n
Trip Time Accuracy ±5% ±2 digits
Ramp Current Range
½ I∆n to 1.1 I∆n
Trip Current
Measurement Accuracy
10%
6.6 Voltage/Frequency Measurement
Display Range 0V ) 440V
Voltage Measuring Range 0V ) 440V
Resolution 1V
Accuracy ±5% ±2 digits
Frequency Range 45Hz ) 65Hz
Frequency Accuracy Indication only

328A561 revision1.1 - 27 -
7 Environmental Conditions
The 9083P has been designed to perform tests and
measurements in a dry environment.
Maximum barometric elevation for making
measurements is 2000M.
Overvoltage category IEC 60664/IEC 61010, 300V
Category III.
Pollution degree 2 according to IEC 61010-1.
Protective system IP40 according to IEC 60529.
Electromagnetic compatibility (EMC). Interference
immunity and emitted interference conforming to IEC
61326-1.
Operating temperature range of 0°C to 40°C, without
moisture condensation.
The 9083P can be stored at any temperature in the
range -25°C to +65°C (relative humidity up to 90%). The
batteries should be taken out of the instrument for
storage.
Operating Altitude 0 to 2000 meters

328A561 revision1.1 - 28 -
8 Maintenance
Before opening the 9083P ensure that it is
disconnected from all voltage! Electric shock
danger!
8.1 Preparing to work on the 9083P.
Ensure that the 9083P is voltage free as follows, before
opening the instrument;
Power the unit off using the rotary switch by selecting
the Off position on the rotary switch.
Disconnect all of the test leads from the unit
8.2 Securing the 9083P
Under certain conditions safe operation of the 9083P
can no longer be assumed:
Visible damage of the instrument case.
Incorrect measurement results.
Recognisable abuse to the instrument due to prolonged
storage under improper conditions.
Recognisable abuse to the instrument due to
extraordinary transportation stress.
Check the battery compartment for signs of battery
electrolyte leakage.
In these cases, the 9083P should be immediately
switched off, disconnected from any test or
measurement function and secured to prevent any
further use.

328A561 revision1.1 - 29 -
8.3 Cleaning
Clean the external case of the 9083P with a clean dry
cloth.
Avoid using solvents and abrasive scouring agents to
clean the external case of the 9083P.
Check the battery contacts and compartment are free
of electrolytic contamination.
Any contamination of the battery contacts or
compartment should be cleaned with a dry cloth.
8.4 Battery Replacement
Before removing the battery cover, ensure that
the 9083P is electrically isolated from any
potential voltage source.
Power the unit off by selecting the Off position on the
rotary switch.
Disconnect all the test leads from the unit
Position the 9083P face down and release the captive
screw in the battery compartment cover.
Remove the battery compartment cover.
Remove the discharged batteries from the
compartment.
Fit a new set of alkaline batteries.
Relocate the battery cover over the battery
compartment and fasten in position with the battery
cover captive screw.

328A561 revision1.1 - 30 -
8.5 Replacing the Fuse.
Before removing the battery cover, ensure that
the 9083P is electrically isolated from any
potential voltage source.
All replacement fuse types are specified for
ratings and size on the battery compartment
cover on the rear of the 9083P.
Power the unit off by selecting the Off position on the
rotary switch.
Disconnect all the test leads from the unit.
Position the 9083P face down and release the captive
screw in the battery compartment cover.
Remove the battery compartment cover.
Lift one end of the fuse out of the fuse holder with the
help of a flat bladed screwdriver.
Lift the defective fuse completely out of the fuse holder.
Insert a new fuse of the correct type as described and
specified by the text on the battery compartment cover.
Ensure that the new fuse is seated and centred in the
fuse holder.
Relocate the battery cover over the battery
compartment and fasten in position with the battery
cover captive screw.
8.6 Service and Calibration.
To maintain the specified accuracy of the measurement
results, the instrument must be recalibrated at regular
intervals by either the manufacturer or an authorised DiLOG Service Agent. We recommend a recalibration
period of one year.

328A561 revision1.1 - 31 -
8.7 Spare Parts.
Part No.
Red, Black and Green lead set
(includes crocodile clips)
44B151
Mains test lead 44B152
Cordless Probe 43B721
Carrying Case 328A953
3.15A T 500V 1¼* 27B083
For help or advice on Service and Calibration contact:
Service Department
Di-LOG Test Equipment
28 Wheel Forge Way
Trafford Park
Manchester
M17 1EH
Freefone: 0800 018 9122 Freefax: 0800 018 6711
Email: sales@dilog.co.uk web: dilog.co.uk