OPERATION MANUAL
1
1. ATIONSAFETY INFORM SAFETY SYMBOLS
Warning! Dangerous Voltage (Risk of electric shock). Caution! Refer to the user’s manual before using this Meter.
Double Insulation.
Alternating Current (AC). Direct Current (DC). Either DC or AC.
Ground (maximum permitted voltage between terminal and ground).
US Safety Standard.
European Safety Standard
The symbol indicating separate collection for electrical and electronic equipment.
The RESPONSIBLE BODY shall be made aware that, if the instrument
is used in a manner not specified by the manufacturer, the protection provided by the instrument may be impaired.
The finger or any part of your body shall not be beyond the barrier of
the test probe when measuring.
Individual protective equipment must be used if HAZARDOUS LIVE parts in the
installation where measurement is to be carried out could be ACCESSIBLE.
The following safety information must be observed to insure maximum personal safety during the operation at this meter.
1.1 Do not operate the meter if the body of meter or the test lead look broken.
1.2 Check the main function dial and make sure it is at the correct position before each
measurement.
1.3 When making current measurements ensure that the circuit not “live” before opening it in order to connect the test leads.
1.4 Do not perform resistance, capacitance, temperature, diode and continuity test on a live power system.
1.5 Do not apply voltage between the test terminals and test terminal to ground that exceed the maximum limit record in this manual.
1.6 Exercise extreme caution when measuring live system with voltage greater than 60V DC or 30V AC.
1.7 Change the battery when the “
” symbol appears to avoid incorrect data.
1.8 Use the DMM indoor, altitude up to 2000m and temperature 5°C to 40°C.
Maximum relative humidity 80% for temperatures up to 31°C, decreasing linearly to 50% relative humidity at 40°C.
1.9 In locations subject to radio frequency interference, the product may malfunction and it
resets automatically when leaving this environment.
2. SPECIFICATIONS
2.1 GENERAL SPECIFICATIONS
Display: LCD with a max. reading of 4000. Range control: Auto range & Manual range control
2
Polarity: Automatic negative polarity indication. Zero adjustment: Automatic. Overrange indication: The “OL” or “-OL” display.
Low battery indication: Display "
" sign.
Data hold: Display “DH” sign. Relative measurement: Display “” sign. Auto Power Off: Display “APO” sign. When measurement exceeds 10 minutes without
switching mode and pressing key, the meter will switch to standby mode. Press any key to exit standby mode. When restart the system, press and hold SELECT key to disable auto power off.
Safety standards: ETL,
EMC/LVD. CAT ⅢⅢ 600V.
The meter is up to the standards of IEC1010 Double Insulation Pollution Degree 2, Overvoltage Category .
Operating environment: Temperature 32 to104°F (0°C to 40°C), Humidity 80% RH. Storage environment: Temperature -4 to140°F (-20°C to 60°C), Humidity 90% RH. Power supply: One standard 9 volt battery (NEDA 1604, IEC 6F22 or equivalent).
Dimension: 165(H) x 83(W) x 36(D) mm Weight: Approx. 360g (including battery).
2.2 ELECTRICAL SPECIFICATIONS
Accuracies are ± ( % of reading + number in last digit ), at 23 ± 5°C ,75% RH.
2.2.1 DC Voltage
Overload protection: 600V DC or AC rms Impedance: 10MΩMore than 100MΩ on 400mV range.
2.2.2 AC Voltage (
AC+DC True RMS
)
[1]
Range Accuracy ± ([% of Reading] + [Counts]) DC, 45Hz to 500Hz 500Hz to 1kHz 1kHz to 2kHz
0.4V,4V 1.0%+3 2.0%+3
40V,400V,600V 1.0%+3 2.0%+3 Impedance : 10MΩ. More than 100MΩ on 400mV range. Overload protect:600V DC or AC RMS.
[1] All ac ranges are specified from 1 % to 100 % of range. Below 5 % of range, add 10 counts. In the range 400mV, an additional 2mV.. Because inputs below 1 % of range are not specified, it is normal for this and other true-rms meters
to display nonzero readings when the test leads are disconnected from a circuit or are shorted together. For volts and amps, crest factor of ≤3 at 3000 counts, decreasing linearly to 2 at full scale. AC volts and AC amps are dc-coupled. If the measurement of AC voltage contains more than DCV 250V, please choose manual range 600V.
2.2.3 DC Current
Range Accuracy ± ([% of Reading] + [Counts])
0.4V,4V,40V,400V 0.5%+2 600V 0.8%+3
Range Accuracy ± ([% of Reading] + [Counts]) 400µA, 4000µA 1.5%+2 40mA, 400mA 1.5%+2 4A,10A
[1]
2.0%+3
Loading...