Instruction Manual Type K Thermocouple Thermometer
5
3. CALIBRATION
3.1 Temperature Calibration
Your meter is factory calibrated. The meter
allows you to enter a 1-point calibration as fine
adjustment by changing its offset value. This
is useful when using a new probe for the first
time.
1. Connect the temperature probe and
switch on the meter.
2. Dip the probe in a constant
temperature bath, or in liquid whose
temperature can be checked
against another accurate
thermometer. For best accuracy,
place your thermocouple probe and
the reference thermometer in a
constant temperature bath.
3. Note the reading. Power off the
meter once the reading has
stabilized.
4. With the meter powered off, hold
down °C/°F key and press ON/OFF
key to power up the meter.
5. Continue to hold down °C/°F key
until display shows “CA”
momentarily. This is followed by a
blinking value which is with respect
to the factory default calibration.
6. To adjust the display until it matches
the correct temperature:
- Press °C/°F key to decrease
- Press HOLD key to increase
With meter off, Hold
down key and
press to
power on the meter.
Continue hold down
until display
shows CA
momentarily.
°C/°F
ON/OFF
°C/°F
C
°
C°
°C
Press HOLD key to
increase value
To co n fi rm va lu e , wa i t 5
seconds.
CO = confirms
temperature calibration
Back to
measurement
Press °C/°F key
to d ecrease
value
°C°F
°C°F
ON
OFF
HOLD
Figure 9: Calibration
Sequence