Calibration Manual
Digi-Sense® TC9000 and TC9500
Advanced Temperature Contr ollers
Models 89800-01, 89800-02, 89800-03, and 89800-04
THE STANDARD IN PRECISION MEASUREMENT
1

Introduction
Table 1. Sensor Input Range
Each temperature controller has built into the operation menus a calibration gateway. This
gateway allows the temperature controller to be calibrated to six temperature points, allowing
for consistent and accurate temperature readings. The calibration process allows for both
full-range calibration and segmented calibration. Each unit comes preloaded with tested
default calibration temperature points.
WARNING: Do not enter the calibration gateway unless you have the proper calibration
equipment to calibrate the sensor inputs associated with this controller. Changing calibration
settings can cause errors in sensor temperature readings and operation of the controller.
Full-Range Calibrati on
This process provides accurate temperature measurements along the entire range of a given
type of temperature probe. Ranges of each type of temperature probe are provided on the
controllers menu screen, as well as in Table 1 below.
SENSOR TYPE RANGE
THERMOCOUPLES
Calibration accuracy ±0.1% of span or ±1°C for types J, K, T, E, N;
±0.2% of span or ±2°C for types B, R, S. Accuracy span is1000°F (540°C) minimum.
Type J -190 to 1000°C
Type K -200 to 1372°C
Type N -200 to 1300°C
Type R 0 to 1768°C
Type S 0 to 1768°C
Type T -200 to 400°C
Type B 600 to 1800°C
Type E -200 to 1000°C
PLATINUM RTD
Calibration accuracy ±0.1% of span. Accuracy span is 1000°F (540°C) minimum.
0.003850
0.003916
-200 to 800°C
-200 to 630°C
Calibration accuracy ±0.4% of span or ±0.5°C. Accuracy span is 266°F (130°C) minimum.
TYPE 400 -30 to 100°C
TYPE 700 -15 to 100°C
2

Segmented calibration allows you to calibrate between any two temperatures along the full
Table 2. Default Calibration Points
range of the probe, providing a more accurate reading between those temperatures. For
example, placing all six calibration points between 0.0°C and 600°C will yield more accurate
temperature readings between those temperatures than the full-range calibration would.
WARNING: Measurements outside the calibrated segment may be notably skewed
Default Calibration Points
Default calibration points come preloaded in the controller (also see Table 2). These points have
been selected and tested to provide accurate temperature readings across the full range. While
these calibration points can be adjusted for segmented calibration, it is not suggested that they
be changed for full-range calibration due to nonlinearity of temperature probes.
SENSOR
TYPE
Type J
Type K
Type N
Type R
Type S
Type T
Type B
Type E
0.003850
0.003916
Type 400
Type 700
DEFAULT CALIBRATION POINTS (°C)
THERMOCOUPLES
PLATINUM RTD
THERMISTOR
3

.
ENTER
Cal i br ati on Set up...
Cal i br at i on Gat eway
Use the UP or DOWN arrows to change
values for proper calibration of sensor input type.
Set pr obe t emp t o XXX. X
Pr ess MENU t o cal i br ate...
Calibration: Poi nt X of X
the next point when complete.
Ar e you sur e you want t o
save t hese poi nt s?
NO
Cal i br at i on Conf i rmat i on
Calibration Mode Setup
Equipment requirements: Fluke 753 or equivalent calibrator for use in thermocouple and RTD
calibration, and an IET decade box for thermistor calibration.
Step 1
Step 2
Step 4
Step 5
Step 6
Connect the calibrator to the appropriate
temperature connector on the back of the
controller.
Select input sensor type and the specific
sensor type within the setup menu.
Advance the menu to the Calibration
Gateway screen.
Select ENTER Calibration Setup within
the Calibration Gateway menu.
the default temperature setting.
Recommend leaving temperature setting at default
Press MENU to calibrate at the stated
temperature set point.
NOTE: The unit will say “Calibrating…” for
approximately 5 seconds and then advance to
Repeat Steps 4 and 5 for all six (6) calibration points
To save these new calibration points, select
YES
when prompted for confirmation.
Step 7
To discard these calibration points and
restore the previous calibration points,
select NO
when prompted.