T
yp e J
, K ,
T
Thermometers
Type J, K, T
68X001351 Rev0 0717 1065DGMAN_91428-00,-01,-02
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T
A
BLE OF CONTENTS
1. INTRODUCTION .................................... 1
2. SAFETY PRECAUTIONS ....................... 2
3. SPECIFICATIONS .................................. 3
4. BATTERY INSTALLATION
AND REPLACEMENT ............................ 5
5. INSERTING AND REMOVING
RUBBER ARMOR .................................. 6
6. ASSEMBLING OPTIONAL
HANDS-F R EE ACCES O R IES ................ 7
7. SELF TEST ............................................ 8
8. CONNECTING A THERMOCOUPLE ...... 8
9. KEY FUNCTIONS ................................. 10
10. DISPLAY OVERVIEW .......................... 11
11. SELECTING TEMPERATURE SCALE . 12
12. HOLD FUNCTON ................................. 12
13. MIN AND MAX FUNCTION ................... 13
14. AUTO OFF FUNCTION ........................ 13
15. CALIBRATION ...................................... 13
16. CANCELIN G A FIELD
CALIBRATION ...................................... 14
17. FIELD CALIBRATION LOCKOUT
AND RE-ENABLE ................................. 15
18. MAINTENANCE .................................... 16
19. CLEANING ........................................... 16
20. BATTERIES .......................................... 16
21. TROUBLESHOOTING .......................... 17
22. ACCESSORIES .................................... 18
23. WARRANTY ......................................... 20
24. PRODUCT RETURN ............................ 20
25. INNOCAL® CALIBRATION AND
REPAIR SERVICES ............................. 21
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
1. INTRODUCTION
This versatile handheld
instrument provides
highly accurate
temperature
measurements. The
instrument is designed
for easy operation and
includes the following
features:
x Operator select ion of Celsiu s
or Fahrenheit scale
x Resolution of 0.1°C/F from
–199.9 to 999.9°
x Large backlit LCD with two lines
of four-digit display
x Hold feature for temporarily
retaining a reading
x Displays Min and Max readings
x
Field calibration capability
x
Disabling of Auto-Off function
x Low-battery warning
x Two-blade female ANSI mini-
connector input
x Operates with a wide selection
of probes
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
- 1 –
Weak batteries should not be left in
the instrument. Dead batteries can
2. SAFETY PRECAUTIONS
WARNING:
1. This instrument is designed to
accept low level signals sup pli ed
by standard thermocouples. Under
NO circumstances should the input
voltage exceed the speci fied 50V
RMS.
2. To prevent ignition of a hazardous
atmosphere, batterie s must only be
changed in an area known to be
nonhazardous.
CAUTION:
1. Do not use or store this instrument
in microwave ovens or any
abnormally hot or cold areas.
2.
leak and cause damage to unit.
DANGER:
1. Voltages present at the
thermocouple may also be present
at the battery terminals. Always
disconnect the thermocouple
when changing batteries.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
- 2 –
3. SPECIFICATIONS
Thermocouple Thermometers
ype J
ype K
ype T
ra
Out-of-
nge display: - - -
Resolution:
0.1°/1° auto-ranging,
0.1°C/F from –199.9 to 999.9°,
1°C/F outside this range
Accuracy:
Below –150°C (–238°F):
-
±0.25% of reading ±1°C (±2 °F)
Above –150°C (–238°F):
±0.1% of reading ±0.4°C (±0.7°F)
Display:
Backlit LCD of viewing area 58 x 40mm
Min/Max function:
Yes
Auto Off (after 17.5 minutes):
Enable/disable option available
Stability criteria:
Yes, upon stability of 5 seconds
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
- 3 –
1/A1: 1998 (EU EMC Directive)
Display update
0.5
sec per up
rate:
date
t:
Inpu
O
ne thermocouple with
ANSI
conn
Inpu
50V
t
protection:
rms
Storage:
–40°C to 65°C ( –40°F to 149°F)
Humidity:
10% to 90% RH (noncondensing)
Battery life:
ector
Size: Three AA, 1.5V; alkaline
Life: 750 hours continuous, typical,
(without backlighting)
Dimensions:
Without armor:
175mm (L) X 97mm (W) X 42mm (H)
With armor:
180mm (L) X 102mm (W) X 52mm (H)
Weight with batteries:
Without armor: 267g
With armor: 362g
Ingress protection:
Meets IEC-529 IP-54 for dust and water
resistant enclosures (probe attached)
CE Compliance
Find Quality Products Online at: sales@GlobalTestSupply.com
EN61326-
www.GlobalTestSupply.com
- 4 –
4. BATTERY INSTALLATION AND
REPLACEMENT
The total battery life without
backlighting is about 750 hours.
Each bar of the battery annunciator
represents 200 hours. A blinking
empty battery annunciator indicates
150 hours remaining.
Selected settings are stored in
memory and will remain in memory
even after power is turned off, or
while batteries are being replaced.
1. Before changing ba ttery, turn
instrument off and disconnect
thermocouple.
2. Loosen screw and lift battery
cover off back of case.
3. Remove the three AA batteries.
4. Insert three new batterie s
observing polarity.
5.
Install cover and tighten screw.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
- 5 –
5. INSERTING AND REMOVING
RUBBER ARMO R
1. To insert thermometer into the
optional rubber armor, slide in from
the top of meter before pushing the
bottom edges of meter down to set
it into position. Lift up the stand at
the back of meter for bench top
applications if necessary.
2. To remove thermometer from
armor, push out from the bottom
edges of meter until it is completely
out of boot.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
- 6 –