Digi-Sense Dual J-T-EK 60010-40, Dual J-T-E-K 91100-40 User Manual

-
Digi
THERMOCOUPLE THERMOMETER
Dual J-T-E-
MODEL NO. 60010-40
Sense
®
K
Digi
Eutech Instruments Pte Ltd
Blk 55 Ayer Rajah Crescent
#04-16 Singapore 139949
Website: www.eutechinst.com
email: marketing@eutechinst.com
68X309907 Rev.1 12/03
-
Sense
Tel: (65) 6778 6876
Fax: (65) 6773 0836
INTRODUCTION
This versatile hand-held instrument provides highly accurate temperature measurements in Celsius, or Fahrenheit, using a wide range of ther mocouple types. The temper ature range f or each type ther­mocouple is listed in the follo wing chart.
TYPE °C °F
J –200 to 1000 –328 to 1832 K –250 to 1372 –418 to 2501 T –250 to 400 –418 to 752 E –250 to 1000 –418 to 1832
1
The instrument is designed for easy operation and includes the follo wing features:
•Operator selection of temperature scale
•Resolution of 0.1° from 150°C to 999.9°C (238°F to 999.9°F)
•LCD with three four-digit displa ys
• Two (2-blade f emale) ANSI mini-connector inputs
•Hold feature f or temporarily retaining a reading
•Two-point field calibration capability
•Low battery warning
•Stores up to 25 readings
•Scrolls through all stored readings
•Displays MIN and MAX readings
•Scrolls between T1, T2, and T1-T2 readings
•Built-in tilt stand for easy hands-free operation
2
SAFETY PRECAUTIONS
WARNING
LOW LEVEL SIGNALS SUPPLIED BY STAN­DARD THERMOCOUPLES. UNDER NO CIR­CUMSTANCES SHOULD THE INPUT VOLTAGE EXCEED THE SPECIFIED 50V RMS.
CAUTION
MICROWAVE OVENS OR ANY ABNORMALLY HOT OR COLD AREAS .
CAUTION
STR UMENT . DEAD BATTERIES CAN LEAK AND CA USE D AMAGE T O UNIT.
DANGER
MA Y ALSO BE PRESENT AT THE BA TTER Y TER­MINALS. AL W A YS DISCONNECT THE THERMO­COUPLE WHEN CHANGING BATTERIES.
THIS INSTRUMENT IS DESIGNED TO ACCEPT
DO NOT USE OR STORE THIS INSTRUMENT IN
WEAK BA TTERIES SHOULD NOT BE LEFT IN THE IN-
VOLTAGES PRESENT AT THE THERMOCOUPLES
WARNING
PHERE, BATTERIES MUST ONL Y BE CHANGED IN AN AREA KNO WN T O BE NON-HAZARDOUS.
AVERTISSEMENT
D’ATMOSPHERES DANGEREUSES, NE CHANGER LES BATTERIES QUE DANS DES EMPLACEMENTS DESIGNES NON DAN­GEREUX.
WARNING
PHERE BY ELECTROSTATIC DISCHARGE, CLEAN WITH DAMP CLOTH.
TO PREVENT IGNITION OF A HAZARDOUS ATMOS-
AFIN DE PREVENIR L’INFLAMMATION
TO PREVENT IGNITION OF A HAZARDOUS ATMOS-
3
SPECIFICATIONS
THERMOCOUPLE PROBES
T ype T emperature Range
Type J: –200°C to 1000°C (–328°F to 1832°F ) Type K: –250°C to 1372°C (–418°F to 2501°F ) Type T : –250°C to 400°C (–418°F to 752°F ) Type E: –250°C to 1000°C (–418°F to 1832°F )
Accuracy > –150°C: ±0.1% of reading, ±0.4°C
(±0.7°F)
Accuracy < –150°C: ±0.25% of reading, ±1°C
(±2°F)
Linearization: Conforms to NIST monograph 175
revised to ITS-90.
Input Protection: 50V rms Conversion Rate: Reading 0.6 seconds per
update.
Connector: T wo-b lade female ANSI mini-connector
inputs.
Battery
Size: Two AA, 1.5V alkaline ANSI-L40,
IEC-LR6
Life: 750 hours continuous, typical
Display: LCD with 0.4 in high characters main
readout and 0.2 in high characters secondary displays, 4 digits each display plus v arious an­nunciators.
4
T emperature/Humidity Range Operating:
Stated Accuracy: 18°C to 28°C
(64°F to 82°F)
Useful Range: 0°C to 40°C
(32°F to 104°F)
Storage: −40°C to 65°C
(40°F to 149°F)
Humidity: 10% to 90%
(non-condensing)
Dimensions
3 cm D x 8.4 cm W x 15.8 cm H (1.2 in x 3.3 in x 6.2 in)
Weight with batteries: 227 g rams (8 ounces) Ingress protection: Meets IEC-529 IP-54 f or dust
and water-resistant enclosures.
Intrinsic safety
This product is energy limited for intrinsically safe operation in h ydrogen atmospheres per Class I, Division 1, Groups A, B, C, and D hazardous (classified) locations for UL per UL913 and CSA per C22.2 No. 0-M1982 and No. 157-M1987. Maximum surface temperature: 135°C (T4); UL file No. E182612 (1997).
Compliance (For CE Mark)
EN61326-1/A1: 1998 (EU EMC Directiv e)
5
BATTERY INSTALLATION AND
REPLACEMENT
CAUTION
INSTRUMENT. DEAD BATTERIES CAN LEAK AND CA USE DAMAGE TO UNIT.
DANGER
MAY ALSO BE PRESENT A T THE BATTER Y TER­MINALS. AL W A YS DISCONNECT THE THERMO­COUPLE WHEN CHANGING BATTERIES.
WARNING
PHERE, BATTERIES MUST ONLY BE CHANGED IN AN AREA KNO WN T O BE NON-HAZARDOUS.
AVERTISSEMENT
D’ATMOSPHERES DANGEREUSES, NE CHANGER LES BATTERIES QUE DANS DES EMPLACEMENTS DESIGNES NON DAN­GEREUX.
WEAK BATTERIES SHOULD NOT BE LEFT IN THE
VOLTAGES PRESENT AT THE THERMOCOUPLES
TO PREVENT IGNITION OF A HAZARDOUS ATMOS-
AFIN DE PREVENIR L’INFLAMMATION
6
If battery indicator turns on, battery life is approxi­mately 8 to 10 hours. The battery indicator will start blinking with less than 1 hour of life remaining.
AT THIS POINT BATTER Y MUST BE CHANGED . IF BATTERY VOLTAGE GOES TOO LOW, DIS­PLAY WILL GO BLANK.
See SPECIFICATIONS for battery type.
1. Before changing battery , turn instrument off and disconnect thermocouple(s).
2. Loosen screw and lift battery cover off back of case.
3. Remov e the two AA batteries.
4. Insert two new batteries observing polarity .
5. Install cov er and tighten screw .
7
CONNECTING A THERMOCOUPLE
Note: Be sure your instrument setting (J, T, E or K)
matches the thermocouple you are using. Use the correct thermocouple type f or your instru-
ment. Using an incorrect ther mocouple type will result in erroneous readings. Thermocouples are color coded by type using the North American ANSI color code as follows:
TYPE COLOR
J Black K Yellow T Blue E Purple
Thermocouple connectors have one wide blade and one narrow blade.
wards.
Connect thermocouple(s) to receptacle at
Do not force connector in bac k-
top of instrument.
If only one thermocouple is being used, it can be connected to either T1 or T2. The thermometer will automatically determine which connector is being used. When tw o thermocouples are used, connect one to the T1 connector and one to the T2 connector .
Thermocouple wiring polarity must be correct. If readings decrease as the temperature increases, the thermocouple wires may be reversed. The red wire is negative f or thermocouple wires manufac­tured in North America.
8
9
10
11
QUICK SETUP
Note: Review w arnings on page 6.
1. Install batteries.
2. Connect thermocouple(s).
3. Press the ON/OFF ke y. The thermometer per­forms a self-test and all displa y digits and indi­cators, as shown below, should remain on for approximately one second.
4. Use J-T-E-K key to select the correct thermo- couple and the °C/°F key to select the desired scale.
If a thermocouple has not yet been connected to one or more inputs on the instrument, you will see this display:
This message also appears if a thermocouple is broken. No measurements can be made while this warning is displayed.
12
COMPLETE SETUP PROCEDURE
The setup procedure is used to select the tempera­ture scale, resolution, and thermocouple type.
NOTE
Selected settings are stored in memory and will remain in memory even after power is turned off, or while batteries are being replaced.
SELECTING TEMPERATURE SCALE
Select °C or °F by pressing the °C/ °F ke y. Each time the key is pressed the temper ature scale will switch. Switching between °C and °F can be done at any time during operation.
Each time you turn the instrument on, it will power up with the same settings that were set when the unit was last turned off.
SELECTING RESOLUTION
Select 0.1° or 1° resolution by pressing the
0.1°/1° key. This k ey toggles between the two reso­lutions. When measuring temperature abov e 999.9° or below 150°C the thermometer automatically autoranges to 1° resolution.
SELECTING THERMOCOUPLE TYPE
Select the thermocouple type, either J, T, E, K, by pressing the J-T-E-K key. Each push of the key will step to the next type thermocouple. The se­lected thermocouple type is indicated by the an­nunciators in the lower right corner of the display.
Continue with the OPERATING PROCEDURES section or turn the instrument OFF by pressing the ON/OFF ke y again. The displa y will b lank.
13
OPERATING PROCEDURES
The unit will always power up with the upper dis­play showing T1 unless T1 is not connected. If only T2 is connected at power up, the upper dis­play sho ws T2 and the lo wer left display indicates T1 is open. If thermocouples are not connected to T1 or T2 at power up, the upper display indicates an open T1 and the lower right display indicates an open T2.
If you are in single probe mode and the probe is mov ed to the other connector or to restore the dual probe mode, turn power OFF then back ON.
F or optimum operation, allow about one minute for ambient temperature stabilization. If the unit has been stored at an extreme ambient condition, more time may be needed.
BASIC TEMPERATURE MEASUREMENTS
Check that the thermometer is turned on, the probe(s) connected, the desired resolution 0.1° or 1° is selected, and the desired scale °C or °F, is selected.
Single Probe Measurements
The thermometer will automatically determine if T1 or T2 connection is being used.
Initially the upper display will indicate the measured temperature and the annunciator will indicate which probe is making the measurement. One of the lower displa ys will indicate open and the annunciator will indicate which probe is open. Pressing the T1/T2 ke y will cause the thermometer to go into the single probe differential mode described later .
Temperatures > 999.9° have a resolution of 1° and therefore no decimal will be present.
14
Dual Probe Measurements
The thermometer will automatically determine if one or two thermocouples are connected. When two thermocouples are connected, the upper main dis­play will initially show T1 and the lower right dis­play will sho w T2. Pressing the T1/T2 k ey once will cause the upper display to sho w T2 and the lowe r left display to show T1. Pressing the T1/T2 key again will cause the upper display to show T1-T2, the lower left displa y to show T1, and the lower right display to sho w T2.
MAXIMUM READINGS
The maximum reading function displa ys the maxi­mum reading since power up or since the last time the clear function was used. The maximum read­ing function is ideal for monitoring unattended op­erations while continually displa ying eve ry tempera­ture change that occurs. The maximum and mini­mum v alues are sensed and automatically stored until you are ready to observe the reading.
Do not turn the instrument OFF when a maximum or minimum temperature value may be needed; MAX/MIN memory contents will be lost. Factory Calibration will be maintained.
15
Single Probe Measurements
The display ed information depends on whether you are using a single probe in the dual probe mode or hav e changed to the single probe differential mode by pressing the T1/T2 key .
If you are using a single probe and in the dual probe mode, momentarily press the MAX
key. The
MAX annunciator turns on. If the probe is connected to T1 the maximum reading will be shown in the lower left display. If the probe is connected to T2 the maximum reading will be shown in the lower right display. If a higher maximum occurs while in the MAX mode, the higher reading will be displa yed. To turn off the maximum reading, press the MAX
key again.
If you are in the single probe differential mode, momentarily press the MAX
ke y. The MAX and
MIN annunciators turn on and both the maximum and minimum readings appear in the lower displa ys. The minimum will be on the left display and the maximum on the right display. If a higher maximum or lower minimum occurs while in the MAX mode , the higher reading will be display ed. T o turn off the maximum reading, press the MAX
key again.
16
Dual Probe Measurements Momentarily press the MAX
ke y. The MAX an-
nunciator turns on, and both maximum T1 and T2 readings appear in the lower displa y. The T1 read­ing is on the left display and the T2 reading is on the right display. Press MAX
key again to cancel.
Clearing a Maximum Reading Press the CLEAR key then the MAX
key. The
maximum memory will be cleared.
MINIMUM READINGS
The minimum reading function displays the mini­mum reading since power up or since the last time the clear function was used. While continually dis­playing e very temperature change that occurs, the maximum and minimum values are sensed and automatically stored until you are ready to observe the reading.
Do not turn the instrument OFF when a maximum or minimum temperature value may be needed; MAX/MIN memory contents will be lost.
17
Single Probe Measurements
The display ed information depends on whether you are using a single probe in the dual probe mode or hav e changed to the single probe differential mode by pressing the T1/T2 key .
If you are using a single probe and in the dual probe mode, momentarily press the MIN
ke y . The MIN
annunciator turns on. If the probe is connected to T1 the minimum reading will be shown in the lo wer left display. If the probe is connected to T2 the mini­mum reading will be shown in the lower right dis­play. If a lower minimum occurs while in the MIN mode, the low er reading will be display ed. To turn off the minimum reading, press the MIN
key
again. If you are in the single probe differential mode,
momentarily press the MIN
key. The MAX and
MIN annunciator turns on and both the maximum and minimum readings appear in the lower displa ys. The minimum will be on the left display and the maximum on the right display. If a lo wer minimum or higher maximum occurs while in the MIN mode , the new readings will be displa yed. To turn off the minimum reading, press the MIN
key again.
Dual Probe Measurements Momentarily press the MIN
key. The MIN an-
nunciator turns on, and both minimum T1 and T2 readings appear in the lower displa y. The T1 read­ing is on the left display and the T2 reading is on the right display. Press the MIN
key again to
cancel.
Clearing a Minimum Reading Press the CLEAR ke y then press the MIN
key.
The minimum memory will be cleared.
18
DIFFERENTIAL READINGS (T1-T2)
The differential function is used to compare two different measurements and displa y the difference between them. Diff erential measurements can be made using either one probe or two probes.
Single Probe Measurements
For single probe measurements , one measurement is stored as the reference measurement, then each following measurement is compared to the refer­ence. The reference measurement is viewed on the lower left displa y , the diff erential value is shown in the lower right display and the main displa y shows the present T1 or T2 reading.
1. Place probe at first measurement point.
2. Press the T1/T2 key to store this ref erence tem­perature. The ∆T and M= ann unciators will be on and the reference temperature will be dis­played on the lower left display. The T indi­cates that the differential mode is selected and the M= indicates that a reference temperature has been stored. Any time after this that the T1/ T2 ke y is pressed, the M= value will be updated to the present value shown in the main displa y.
3. Place the probe at the second measurement point. The main display will immediately show the new measured temperature and the lower right display will immediately show the diff erence between the reference temperature and the present temperature point.
19
Dual Probe Measurement
1. Connect the two probes to the two points of measurement. The lo wer left display will indicate the T1 temperature, the lower right display will indicate the T2 temperature.
2. The main display can be scrolled between dis­pla ying T1, T2 or the differential temperature T1­T2 by repeatedly pressing the T1/T2 k ey . T o dis­play the diff erential temperature press the T1/T2 key until the T1-T2 annunciator to the right of the main display is on.
HOLD FUNCTION
Press the HOLD the display. Press HOLD
key to retain the reading on
key again for normal
operation.
Maximum/Minim um Hold Readings Press the MAX
key, then press the HOLD
key. To turn off the HOLD function and return to
normal operation, press the HOLD key again. To clear the maximum readings, press the CLEAR
ke y, then the MAX
key .
Minimum/Hold Press the MIN
key , then press the HOLD
key. To turn off the HOLD function and return to normal operation press the HOLD
ke y again.
To clear the minimum readings, press the CLEAR ke y, then the MIN
STORED READINGS
The store function allows you to store up to 25 read­ings. Both T1 and T2 readings are stored. To store readings proceed as follo ws:
key.
20
1. Momentarily press the STORE k ey . Both T1 and
T2 are stored. The upper main displa y will mo­mentarily show the storage location number and the STO ann unciator will be on.
After three seconds the storage number will be re­placed with the temperature reading but the STO annunciator remains on to indicate a temperature reading has been stored. The STORE key ma y be pressed as fast as once per second.
2. Repeat step 1 for all the points to be recorded
up to a maximum of 25. Each time the STORE key is pressed the new reading will be stored and the upper main display will show the stor­age location number f or about 3 seconds. After 25 stored readings, the next time the STORE key is pressed the main display will indicate FULL.
RECALL READINGS
This function allows the stored readings to be re­called. RECALL shows each stored reading indi­vidually. To recall readings proceed as follows:
1. Momentarily press the RECALL key, the first
stored reading will be display ed on the lower dis­play and the store number “1” will be displayed on the upper display. The STO, and RCL an­nunciators will be on.
2. To step through the readings, press the
MIN
key or the MAX
key. Each press of
the key will advance to the next reading in se­quence.
3. To retur n to normal operation press any other
key except ON/OFF, MAX
CALL.
Clearing Stored Readings
Press the CLEAR key, then the STORE key. The stored readings in memory will be cleared.
21
, MIN
or RE-
CALIBRATION
The calibration function allo ws both single point and dual point calibration of the thermometer. Single point calibrates the offset only. Dual point calibrates the offset first then calibrates the slope. The ther­mometer can be calibrated at any temperature. When two probes are used, a match calibration matches the T1 and T2 offsets.
It is not necessary to perform a field calibration to obtain specified accuracies. Use the calibration feature to impro ve thermometer/probe accuracy or to compensate for thermocouple probe calibration drift.
The thermometer has a memory retention capabil­ity to hold calibration values even while power is off or the battery is removed.
When you restart, there is no need to recalibrate. Each of the four thermocouple types ma y be indi­vidually calibrated and stored.
NOTE
The temperature at which water boils (100°C/ 212°F) is at sea level and standard atmospheric pressure using distilled water . Changes in altitude and barometric pressure will cause the boiling tem­perature to change.
As a general rule, the boiling temperature of water will decrease 1°C (1.8°F) f or every 1000 feet abov e sea lev el. For e xample, at an elev ation of 5300 feet (1600 meters), water will boil at approximately
94.7°C (202.5°F). Other liquids may also be used as calibration stan-
dards. Consult a chemical handbook f or their freez­ing (melting) point and boiling point properties.
22
When calibrating at freezing (0°C or 32°F) it is rec­ommended to use crushed ice made of distilled water in a dewar flask. Add crushed ice to top of flask. T op off flask with distilled w ater. Continue to add crushed ice to maintain tightly packed crushed ice/water in flask.
CALIBRA TION PR OCEDURES
Calibration Procedure (One Pr obe Detected)
This calibration function provides f or both an offset and slope field calibration of either T1 or T2. If only one probe is detected, the procedure applies to that probe, and the main displa y will indicate T1 or T2 as applicable. For proper calibration the f ollowing conditions must be observed:
• The slope point must be a higher temperature than the offset point.
•The difference must be at least 20°C .
•Use two points based on the e xpected high and low temperatures. Temperatures measured out­side of these limits may no longer meet accuracy specifications.
Proceed as follo ws to calibrate the probes.
1. Place the probe at the lower known reference
temperature.
2. Offset Calibration: Momentarily press the CAL
ke y to enter the calibration mode, the CAL an­nunciator will flash. The temperature is displayed on the main display and Lo is displayed on the lower left display. “Lo” signifies that this is the offset point.
23
3. Allow the reading to stabilize. If the displayed
temperature is higher or lower than the ref erence temperature, use the MAX the display ed reading or the MIN
key to increase
key to de-
crease the display ed reading until the reference temperature is display ed. The MIN
or MAX
key must be pressed at least once. The CAL annunciator should be blinking during this procedure.
4. Press the HOLD
key to lock the offset cali­bration in and advance to the slope calibration. To return to normal operation press any k ey ex­cept CAL or HOLD
.
5. Slope Calibration: Place the probe at the higher reference temperature .
6. Allow the reading to stabilize. If the displayed temperature is higher or lower than the ref erence temperature, use the MAX the displayed reading or the MIN
key to increase
key to de-
crease the display ed reading until the reference temperature is displa yed. The MIN
or MAX
key must be pressed at least once. The CAL annunciator should be blinking during this pro­cedure.
7. Press the HOLD
key to lock the calibration
in.
24
Calibration Procedure (T wo Pr obes Detected)
This calibration function provides f or both an offset and slope field calibration of either or both T1 and T2. For proper calibration the f ollowing conditions must be observed:
• The slope point must be a higher temperature than
the offset point.
•The difference must be at least 20°C .
•Use two points based on the e xpected high and
low temperatures. Temperatures measured out­side of these limits may no longer meet accuracy specifications.
Proceed as follo ws to calibrate the probes.
1. Place the probe at the lower known reference temperature.
2. Offset Calibration: Momentarily press the CAL ke y to enter the calibration mode, the CAL an­nunciator will flash. The temperature is displayed on the main display and Lo is displayed on the lower left display. “Lo” signifies that this is the offset point.
3. Allow the reading to stabilize. If the displayed temperature is higher or lower than the ref erence temperature, use the MAX the displayed reading or the MIN
key to increase
key to de-
crease the display ed reading until the reference temperature is displa yed. The MIN
or MAX
key must be pressed at least once. The CAL annunciator should be blinking during this procedure.
4. Press the HOLD ke y to loc k the T1 calibra­tion in and advance to T2 offset calibration, or press the CAL k ey to skip T1 calibr ation. In either case you will now be in T2 offset calibration.
25
5. Allow the reading to stabilize . If the displa y ed
temperature is higher or lower than the refer­ence temperature, use the MAX
key to in-
crease the displayed reading or the
MIN
key to decrease the displa yed reading
until the reference temperature is displayed.
The MIN
or MAX
key must be pressed
at least once. The CAL annunciator should remain blinking during this procedure.
6. Press the HOLD
key to lock the T2 offset calibration in, or press the CAL ke y to skip T2 offset calibration and proceed with T1 slope.
If T1 or T2 offset calibration is skipped, then the slope calibration f or that input will also be skipped.
7. Slope Calibration: Place the probe(s) at the higher reference temperature .
8. The temperature is display ed on the main dis­play and "Hi" is displa yed on the lo wer left dis­play.
9. Allow the reading to stabilize . If the displa y ed temperature is higher or lower than the refer­ence temperature, use the MAX
key to
increase the displayed reading or the
MIN
key to decrease the displa yed reading
until the reference temperature is displayed.
The MIN
or MAX
key must be pressed
at least once. The CAL annunciator should remain blinking during this procedure.
10. Press the HOLD
key to lock the T1 slope calibration in and adv ance to T2 slope calibr a­tion, or press the CAL key to skip T1 slope calibration. In either case you will now be in T2 slope calibration.
26
11. Press the HOLD
key to lock the T2 slope
calibration in, or press the CAL ke y to skip T2 slope calibration. In either case you will now be in T1 = T2 offset calibration. To retur n to normal operation press any key except CAL key or HOLD key.
If either T1 or T2 offset or slope w ere modified, CAL 1 or 2 will be lit.
12. The MATCH annunciator should be blinking. When both readings (T1 and T2) are stable, press the HOLD
key to lock the T1 = T2 offset calibration in. The MATCH annunciator will remain lit.
13. This completes the two-point calibration.
14. To retur n to normal operation press any key except ON/OFF k e y or the CAL key.
Matching T1 and T2 Procedure
This procedure requires both probes to be at the same temperature.
1. Place both probes at the same temperature and
let stabilize.
2. Momentarily press the CAL key until the MATCH
annunciator is blinking. When both readings (T1 and T2) are stable, press the HOLD lock in the calibration.
Clearing an Individual Cal P oint: Press CAL key until the desired point is displayed, then press
CLEAR key. Clearing all Cal Points: Press CLEAR key, then
CALkey. Unit rev erts to factory calibration with no
offset or slope compensation.
key to
27
FIELD CALIBRA TION LOCK OUT AND RE-ENABLE
The calibration lock out f eature , pre v ents an y field calibration changes. The loc kout remains in eff ect until a lock out re-enable has been performed. Use the follo wing procedures to lockout or re-enable the field calibration operation.
Lock out Procedure
1. T urn the thermometer off.
2. Simultaneously press and hold the CAL and the
CLEAR ke ys down and momentarily press the ON/OFF key. Continue to hold the CAL and CLEAR ke ys f or at least 5 seconds.
Re-Enable Pr ocedure
1. T urn the thermometer off.
2. Simultaneously press and hold the HOLD the CAL keys down and momentarily press the
ON/OFFk ey . Continue to hold the HOLD CAL keys until the displa y b lanks.
and and
28
MAINTENANCE AND
TROUBLESHOOTING
Properly used, the thermometer should maintain calibration indefinitely and not require service other than occasional cleaning of the housing and chang­ing of the batteries.
TO PREVENT IGNITION OF
WARNING
PHERE BY ELECTROSTATIC DISCHARGE, CLEAN WITH DAMP CLOTH.
Do not clean with abrasives or solv ents . Use mild detergents, ne ver immerse nor use excessiv e fluid.
BATTERIES
If there is no display when the thermometer is
A HAZARDOUS ATMOS-
turned on, check condition of the two AA batteries. Also check that the battery terminals are clean and batteries are properly installed. If replacement is necessary , ref er to the BATTERY INSTALLATION AND REPLACEMENT section for replacement procedure.
All stored readings are retained until cleared, ev en if the batteries are removed for long periods.
29
SERVICE
WARNING
SUBSTITUTION OF COM­PONENTS MAY IMPAIR
INTRINSIC SAFETY.
AVERTISSEMENT
LA SUBSTITUTION DE COMPOSANTS PEUT COM-
PROMETTRE LA SECURITE INTRINSEQUE.
There are no internal adjustments or user replace­able parts.
Note: Serial number label is located inside battery compartment.
30
TROUBLESHOOTING
The following chart lists the most probable faults . There are no internal adjustments or user-replace­able parts. If this does not solve the prob lem, refer service to your dealer.
FAULT ACTION
No display when Check condition of turned on. batteries.
Check that batteries are inserted properly.
Display shows Out of range indication.
- - - ­Display shows Open thermocouple
OPEn connection. Display shows If displa yed at any time
Err other than during field
calibration, return instrument for service.
31
W ARRANTY
The Manufacturer warrants this product to be free from significant deviations from published specifications. If repair or adjustment is neces­sary within the warranty period, the problem will be corrected at no charge if it is not due to mis­use or abuse on your part as determined by the Manufacturer. Repair costs outside the warranty period, or those resulting from product misuse or abuse, may be invoiced to you.
This product comes with a warranty of 3 years.
PRODUCT RETURN
To limit charges and delays, contact the seller or Manufacturer for authorization and shipping in­structions before returning the product, either within or outside of the warranty period. When returning the product, please state the reason for the return. For your protection, pack the product carefully and insure it against possible damage or loss. Any damages resulting from improper packaging are your responsibility.
TECHNICAL ASSISTANCE
If you have any questions about the use of this product, contact the Manufacturer or authorized seller.
Trademarks bearing the ® symbol in this publication are reg­istered in the U.S.
and in other countries.
Rev.1
Printed in Singapore
68X309907 12/03
Loading...