
Single-Input
Instruction Manual
Single-Input
Thermocouple
Thermometers
Type J, K, T
Thermocouple
Thermometers
T
ype J
, K,
T
68X001351 Rev0 0717 1065DGMAN_91428-00,-01,-02

99 Washington Street
Melrose, MA 02176
Phone 781-665-1400
Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com

TABLE OF CONTENTS
1. INTRODUCTION .................................... 1
2. SAFETY PRECAUTIONS ....................... 2
3. SPECIFICATIONS .................................. 3
4. BATTERY INSTALLATION
AND REPLACEMENT ............................ 5
5. INSERTING AND REMOVING
RUBBER ARMOR .................................. 6
6. ASSEMBLING OPTIONAL
HANDS-FREE ACCESORIES ................ 7
7. SELF TEST ............................................ 8
8. CONNECTING A THERMOCOUPLE ...... 8
9. KEY FUNCTIONS ................................. 10
10. DISPLAY OVERVIEW .......................... 11
11. SELECTING TEMPERATURE SCALE . 12
12. HOLD FUNCTON ................................. 12
13. MIN AND MAX FUNCTION ................... 13
14. AUTO OFF FUNCTION ........................ 13
15. CALIBRATION ...................................... 13
16. CANCELING A FIELD
CALIBRATION ...................................... 14
17. FIELD CALIBRATION LOCKOUT
AND RE-ENABLE ................................ . 15
18. MAINTENANCE .................................... 16
19. CLEANING ........................................... 16
20. BATTERIES .......................................... 16
21. TROUBLESHOOTING .......................... 17
22. ACCESSORIES .................................... 18
23. WARRANTY ......................................... 20
24. PRODUCT RETURN ............................ 20
25. INNOCAL® CALIBRATION AND
REPAIR SERVICES ............................. 21

1. INTRODUCTION
This versatile handheld
instrument provides
highly accurate
temperature
measurements. The
instrument is designed
for easy operation and
includes the following
features:
• Operator selection of Celsius
or Fahrenheit scale
• Resolution of 0.1°C/F from
–199.9 to 999.9°
• Large backlit LCD with two lines
of four-digit display
• Hold feature for temporarily
retaining a reading
• Displays Min and Max readings
•
Field calibration capability
•
Disabling of Auto-Off function
• Low-battery warning
• Two-blade female ANSI mini-
connector input
• Operates with a wide selection
of probes
- 1 –

2. SAFETY PRECAUTIONS
WARNING:
1. This instrument is designed to
accept low level signals supplied
by standard thermocouples. Under
NO circumstances should the input
voltage exceed the specified 50V
RMS.
2. To prevent ignition of a hazardous
atmosphere, batteries must only be
changed in an area known to be
nonhazardous.
CAUTION:
1. Do not use or store this instrument
in microwave ovens or any
abnormally hot or cold areas.
2. Weak batteries should not be left in
the instrument. Dead batteries can
leak and cause damage to unit.
DANGER:
1. Voltages present at the
thermocouple may also be present
at the battery terminals. Always
disconnect the thermocouple
when changing batteries.
- 2 –

- 3 –
3. SPECIFICATIONS
Thermocouple Thermometers
Out-of-range display: - - -
-
Resolution:
0.1°/1° auto-ranging,
0.1°C/F from –199.9 to 999.9°,
1°C/F outside this range
Accuracy:
Below –150°C (–238°F):
±0.25% of reading ±1°C (±2 °F)
Above –150°C (–238°F):
±0.1% of reading ±0.4°C (±0.7°F)
Display:
Backlit LCD of viewing area 58 x 40mm
Min/Max function:
Yes
Auto Off (after 17.5 minutes):
Enable/disable option available
Stability criteria:
Yes, upon stability of 5 seconds

- 4 –
Display update
rate:
0.5
sec per up
date
Input:
One
thermocouple with
ANSI
conn
ector
Input protection:
50V
rms
Storage:
–40°C to 65°C (–40°F to 149°F)
Humidity:
10% to 90% RH (noncondensing)
Battery life:
Size: Three AA, 1.5V; alkaline
Life: 750 hours continuous, typical,
(without backlighting)
Dimensions:
Without armor:
175mm (L) X 97mm (W) X 42mm (H)
With armor:
180mm (L) X 102mm (W) X 52mm (H)
Weight with batteries:
Without armor: 267g
With armor: 362g
Ingress protection:
Meets IEC-529 IP-54 for dust and water
resistant enclosures (probe attached)
CE Compliance
EN61326-1/A1: 1998 (EU EMC Directive)

- 5 –
4. BATTERY INSTALLATION AND
REPLACEMENT
The total battery life without
backlighting is about 750 hours.
Each bar of the battery annunciator
represents 200 hours. A blinking
empty battery annunciator indicates
150 hours remaining.
Selected settings are stored in
memory and will remain in memory
even after power is turned off, or
while batteries are being replaced.
1. Before changing battery, turn
instrument off and disconnect
thermocouple.
2. Loosen screw and lift battery
cover off back of case.
3. Remove the three AA batteries.
4. Insert three new batteries
observing polarity.
5.
Install cover and tighten screw.

5. INSERTING AND REMOVING
RUBBER ARMOR
1. To insert thermometer into the
optional rubber armor, slide in from
the top of meter before pushing the
bottom edges of meter down to set
it into position. Lift up the stand at
the back of meter for bench top
applications if necessary.
2. To remove thermometer from
armor, push out from the bottom
edges of meter until it is completely
out of boot.
- 6 –

- 7 –
6.
ASSEMBLING OPTIONAL
HANDS-FREE ACCESORIES
You can use the optional magnets and
strap in the Hands-Free Kit accessories
for hands-free operations.

- 8 –
7. SELF TEST
Press the on/off key. The meter
performs a self-test and all display
digits and indicators, should remain
on for approximately one second.
The thermometer will then display
the thermocouple type and resume
to the measurement mode.
8. CONNECTING A
THERMOCOUPLE
Use the correct thermocouple type for
your instrument. Using an incorrect
thermocouple type will result in
erroneous readings. Thermocouples
are color coded by type using the North
American ANSI Color Code as follows:
TYPE COLOR
J
Black
K Yello
w
T Blue
Thermocouple connectors have one
wide blade and one narrow blade.
Do not force connector in backwards.
Connect thermocouple to receptacle
at top of instrument as shown in the
following illustration:

- 9 –
Thermocouple wiring polarity must be
correct. If readings decrease as the
temperature increases, the thermocouple
wires may be reversed. The red wire is
negative for thermocouple wires
manufactured in North America.

9. KEY FUNCTIONS
Toggle between Celsius or
Fahrenheit.
Activates Calibration mode
for 0°C offset calibration
(after field calibration
enabled).
Activates/deactivates
freezing of the measured
reading while in
measurement mode.
Activates/deactivates display
of the minimum and
maximum readings.
Powers on and shuts off the
thermometer by holding it for
3 seconds. Otherwise, it
activates/deactivates the
backlight display. (Backlight
will automatically turn off
within 30 seconds of
activation).
- 10 –

- 11 –
10. DISPLAY OVERVIEW
Celsius or Fahrenheit indicator.
Minimum reading annunciator.
Maximum reading annunciator.
3 Bars : < 750 hours
2 Bars : < 550 hours
1 Bar : < 350 hours
Blinking: < 150 hours
Blinks during calibration mode.
Remains in display indicating
field calibration is active.
Remain in display during hold
mode.
Displayed upon recognizing
final value.
Error indicator on main digit
display Indicates calibration error
due to out of tolerance value.

- 12 –
If a thermocouple is not connected or if
the thermocouple is defective, the
display will indicate “OPEn”.
Thermocouples are sensitive at the tip
or measuring junction. When taking
measurements, allow time for the
reading to stabilize. Multiplying the time
constant of the probe by 5 will give you
the approximate time required.
11. SELECTING TEMPERATURE
SCALE
Select °C or °F by pressing the °C/°F
key. Each time the key is pressed the
temperature scale will switch. Switching
between °C and °F can be done at any
time during operation.
Each time you turn the instrument on,
it will power up with the same settings
that were set when the unit was last
turned off.
12. HOLD FUNCTON
Press
the
hold key
to
retain
the
reading on
the
displa
y.
Press
the
hold key again
for
normal ope
ration.

- 13 –
13. MIN AND MAX FUNCTION
Press the min/max key to toggle
between the minimum and maximum
readings. This function is ideal for
monitoring unattended operations while
continually displaying every temperature
change that occurs. The minimum and
maximum values are sensed and
automatically stored.
To exit and clear this function, press
the min/max key for 3 seconds.
14. AUTO OFF FUNCTION
The thermometer has a default auto off
function of 17.5 minutes.
To deactivate this function, press and
hold the min/max key while turning on
the meter using the on/off key.
“A.OFF nO” will appear indicating that
the auto off function is disabled.
15. CALIBRATION
The thermometer is factory calibrated
and does not require calibration before
use. The Calibration function allows
single-point calibration of the meter at
0°C (32°F) to compensate for
thermocouple offset error. It is NOT
necessary to perform a field calibration
to obtain specified meter accuracy.
Use the field calibration feature to

- 14 –
improve thermometer/probe accuracy or
to compensate for thermocouple drift.
1. Pack sensing end of probe in a
container tightly packed with
crushed ice and filled with distilled
water. Allow temperature to
stabilize.
2. Press and hold the cal key for
3 seconds to enter the calibration
mode, the CAL annunciator on the
display starts blinking. Release
the cal key.
3. If the measured temperature is from
–
10 to 10°C (14 to 50°F), when
the
temperature reading is stable,
press the cal key.
The CAL indicator will stop blinking
and the reading will be set to 0°C
(32°F). The CAL indicator will remain
turned on, indicating a field calibration
is active.
If “Err CAL” is displayed, either the
displayed reading is outside the above
limits or the batteries are weak.
16. CANCELING A FIELD
CALIBRATION
1. Turn the thermometer off.
2. Hold the cal key down while
pressing the on/off key. The field
calibration is cancelled and the
thermometer reverts to the default
factory calibration. The CAL
annunciator is now turned off.

- 15 –
17. FIELD CALIBRATION LOCKOUT
AND RE-ENABLE
The calibration lockout feature prevents
any field calibration changes. The
lockout remains in effect until a lockout
re-enable has been performed. Use the
following procedures to lockout or reenable the field calibration operation.
LOCKOUT PROCEDURE
1. Turn the thermometer off.
2. Simultaneously press and hold
the cal and the °C/°F keys down and
momentarily press the on/off key.
Continue to hold the cal and °C/°F
keys
until the measurement mode is
displayed.
RE-ENABLE PROCEDURE
1. Turn the thermometer off.
2. Simultaneously press and hold
the hold and the cal keys down and
momentarily press the on/off key.
Continue to press the hold and cal
keys until the measurement mode is
displayed.

- 16 –
18.
MAINTENANCE
Properly used, the thermometer should
maintain calibration indefinitely and not
require service other than occasional
cleaning of the housing and changing
of the batteries.
19. CLEANING
WARNING:
TO PREVENT IGNITION OF A
HAZARDOUS ATMOSPHERE BY
ELECTROSTATIC DISCHARGE,
CLEAN WITH DAMP CLOTH.
Do not clean with abrasives or solvents.
Use mild detergents, never immerse
nor use excessive fluid.
20. BATTERIES
If there is no display when the
thermometer is turned on, check
condition of the three AA batteries.
Also check that the battery terminals
are clean and batteries are properly
installed. If replacement is necessary,
refer to the BATTERY INSTALLATION
AND REPLACEMENT section for
replacement procedure.

- 17 –
21. TROUBLESHOOTING
The following chart lists the most
probable faults. There are no internal
adjustments or user-replaceable parts.
Check condition of batteries.
Check that batteries are
inserted properly.
No thermocouple connected in
the connector.
If display shows this message
other than during the field
calibration mode, please return
the instrument for servicing.

- 18 –
22.
ACCESSORIES
Digi-Sense
Replacem
ent Me
ters
Type J 91428-00
Type K 91428-01
Type T 91428-02
Accessory Thermocouple Probes
Cole-Parmer offers a wide variety of
thermocouple probes, wire, connectors,
extension cables, switch boxes,
accessories, and calibration services.
General-purpose probes, 5 in. long x
0.156 in. dia with 0.093 in. dia tip.
3-second time constant.
Type J 08517-55
Type K 08516-55
Type T 08500-55
Penetration probes, 4 in. long x
0.156 in. dia sharp tip.
5-second time constant.
Type J 08517-65
Type K 08516-65
Type T 08500-65

- 19 –
Surface probes, 10 in. long with
5/8 in. dia aluminium and ceramic tip.
6-second time constant.
Type J 08517-60
Type K 08516-60
Type T 08500-60
Air/gas probes, 8.5 in. long sheath
with 1/4 in. dia radiant heat shield.
45-second time constant.
Type J 08517-75
Type K 08516-75
Type T 08500-75
Accessory Hands-Free Kit
Cole-Parmer offers the Hands-Free Kit
accessory that includes two magnets
and a strap for hands-free operations.
Hang your meter using the magnets
behind the armor, or attach onto a pipe
using the strap that can also be used
for handgrip.
Hands-Free Kit 35427-85

- 20 –
23.
WARRANTY
The Manufacturer warrants this
product to be free from significant
deviations from published
specifications for a period of three
years. If repair or adjustment is
necessary within the warranty period,
the problem will be corrected at no
charge if it is not due to misuse or
abuse on your part as determined
by the Manufacturer. Repair costs
outside the warranty period, or those
resulting from product misuse or
abuse, may be invoiced to you.
24. PRODUCT RETURN
To limit charges and delays, contact
the seller or Manufacturer for
authorization and shipping instructions
before returning the product, either
within or outside of the warranty period.
When returning the product, please
state the reason for the return. For
your protection, pack the carefully and
insure it against possible damage or
loss. The Manufacturer will not be
responsible for damage resulting from
careless or insufficient packing.

- 21 –
25. INNOCAL® CALIBRATION AND
REPAIR SERVICES
Optimum performance of your
temperature-measuring instrument is
not a timeless condition. To ensure
quality measurements, have your
instrument calibrated regularly. Trust
InnoCal® to satisfy your calibration and
equipment repair needs. With over a
decade of service, we've helped
thousands of customers meet ISO,
FDA, EPA, GLPs/cGMPs and other
quality standards.
Conformity*
ISO/IEC 17025:2005 accredited
NIST Handbook 150, 2000 Edition
ANSI/NCSL Z540-2-1997
NIST Technical Note 1297
ISO 9000:2000
Fast Service
Our substantial inventory of replacement
parts ensures a fast turnaround and
prevents costly downtime. Most
instruments serviced in five business
days!
Excellent Value
Get quality at a fair price. Our InnoCal
NIST-traceable certificates offer
extensive test data on a broad range of
measurement parameters without
breaking the bank!

- 22 –
Reliable Support
Trust in our free diagnostic support and
troubleshooting advice. Our factorytrained metrologists and technicians
are armed with years of experience and
extensive technical data.
Convenient Reminders
It’s so easy to keep your instruments
functioning properly. Based on your
requirements, InnoCal will send you
a reminder when it’s time to re-certify
or service your instrument.
We provide you with the documentation
you need to meet your most stringent
quality requirements for the control of
inspection, measuring, and test
equipment.
Certification includes certificate of
calibration with test data, including:
● Description and identification of
the item certified
● Condition of the item
● Issue date
● Identification of calibration procedure
● Calibration date
● As found/as left test data (where
applicable)
● Signature of technician
● Statement of estimated uncertainty
● List of equipment used to perform
calibration (including their calibration
dates)

- 23 –
With today's high quality standards such
as ISO 9000, certification is becoming
increasingly important. Traceability is not
a timeless condition. It must be verified
and maintained over the life of the
calibration to ensure the highest
accuracy possible. When you have your
calibration done by InnoCal, we will send
you an automatic reminder when it is
time to recalibrate your instrument.
Are your calibration certificates
good enough?
InnoCal surpasses the competition by
providing the most complete certificates
as required by NIST. All of our
certificates include measured data and
point-by-point measurement uncertainty,
and by request, we’ll provide test
accuracy and test uncertainty ratios at
no extra cost. Call us today and see
why InnoCal is The Choice of Quality.
*See our Scope of Accreditation for
any limitations.

Calibration
test points
against
NISTtraceable
standards
Four test points
across range of
instrument.
Meters:
–270 to
2316°C
(–454 to
4200°F);
Probes &
Systems:
–80 to 1000°C
(–112 to
1832°F).
Actual range is
dependent on
type of probe
InnoCal—The Choice of Quality
866-INNOCAL (866-466-6225)
InnoCalSolutions.com
- 24 –

TECHNICAL ASSISTANCE
If you have any questions about the
use of this product, please contact the
Manufacturer or authorized seller.
99 Washington Street
Melrose, MA 02176
Phone 781-665-1400
Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com
1065DGMAN_91428-00,-01,-02
- 25 –