Introduction
1065DGMAN_20250-91-93-95.indd 3
The Digi-Sense Professional Single-Input Thermometers (20250-91, -93, -95) offer fast response and laboratory accuracy. Features include automatic backlight, Max/Min readings, and automatic power-off. The instrument is fully tested and calibrated to NIST-traceable standards. Careful use of this meter will provide years of reliable service.
Unpacking
Carefully unpack the instrument and accessories from ship­ping package. Check individual parts against the list of items below. If anything is missing or damaged, please contact your instrument supplier immediately.
1. Meter
2. Soft carrying case
3. Lanyard
4. One 9 volt battery
5. User manual
6. NIST-traceable calibration report with data
3
3/2/2016 3:16:06 PM
Key Features
1065DGMAN_20250-91-93-95.indd 4
• Large backlit display
• Max and Min readings
• Data Hold
Electronic Offset function allows compensation
of errors to maximize overall accuracy
• User-selectable readout in °C or °F
• Automatic power-off (sleep) mode to conserve
battery life
• Accept a wide range of probes
4
3/2/2016 3:16:06 PM
Meter Description
1065DGMAN_20250-91-93-95.indd 5
1. LCD
2. Down () button
3. Data HOLD button
4. MIN button
5. User-selectable temperature units (all models); backlight button (20250-91 only)
6. Power on/off button
20250-91 20250-93 and -95
7. CAL (offset/calibration)
button
8. MAX button
9. Up (
10. T
) button
emperature sensor
input
11. Type JKT button (20250-91 only); backlight button (20250-93 and -95 only)
5
3/2/2016 3:16:07 PM
Display Layout
1065DGMAN_20250-91-93-95.indd 6
1. Thermocouple type JKT icon (20250-91 only)
2. HOLD icon
3. Low-battery indicator
4. User-selectable temperature units
(°C or °F)
Primary display (current
5.
or Hold readings)
6. MAX icon
7. MIN icon
8. Secondary display (Max or Min readings)
6
3/2/2016 3:16:07 PM
Setup and Operation
1065DGMAN_20250-91-93-95.indd 7
Measuring Temperatures
1. Plug the probe into the T1 input connector. Make
sure that the polarity is correct.
2. Press power on/off button to turn on the thermom-
eter. After one second the thermometer displays
the first reading. The screen will display the current measured value and the maximum temperature.
If no probe is plugged into the selected input or the
3. probe is not connected properly "- - - -". If the temperature being measured is out­side the meter’s valid range, the display shows “OL” (overload).
Thermocouple Input Function (20250-91 only)
1.
Press JKT button to set the correct thermocouple type as J, K, or T.
Backlight Function
1. Press Backlight/°C/°F button for 2 seconds to turn on or off the backlight (20250-91 only).
2. Press Backlight button to turn on or off the backlight (20250-93 and -95).
Temperature Units Function
1.
Press °C/°F button to toggle between temperature
scales.
, the display shows
7
3/2/2016 3:16:07 PM
Setup and Operation (continued)
1065DGMAN_20250-91-93-95.indd 8
Hold Function
1. Press the HOLD button to freeze the readings on
the display. The display shows the HOLD icon.
2. Press the HOLD button again to turn off the
Hold function.
3.
The Hold function does not freeze the Max or Min
function or stop the meter from updating.
Max/Min Function
Press MAX or MIN button to display the maxi-
1. mum (Max) or minimum (Min) readings on the secondary display.
2. The display shows the related MAX or MIN icon.
3.
Simultaneously press the buttons to recalculate the stored maximum and minimum values in the meter.
Auto Power-Off (APO) Mode
1.
The meter’
s default mode is to automatically shut
off after 15 minutes of non-use.
Up () and Down ()
8
3/2/2016 3:16:07 PM
Offset/Calibration Function
1065DGMAN_20250-91-93-95.indd 9
Using the Offset to Adjust for Probe Errors
Use the offset/calibration function to adjust the meter’s readings to compensate for the errors of a specific
sensor. The allowable adjustment range is ±9.0°F or ±5.0°C.
1. Plug the probe into the meter’s input connector.
2.
Place the probe in a known, stable temperature environment (such as an ice bath).
3. Allow the readings to stabilize.
4. Adjust the offset until the primary reading matches
the calibration temperature.
Adjusting the Offset
The primary display shows the temperature plus the offset and the secondary display shows the offset.
1. Press the CAL
calibration mode.
2. Press the Up
offset by 0.1°. The max offset value is 5°.
3.
Press the CAL to normal measurement mode.
button for 3 seconds to enter user
() or Down () buttons to adjust the
button to save the setting and return
9
3/2/2016 3:16:07 PM
Specifications
1065DGMAN_20250-91-93-95.indd 10
Range Resolution Accuracy
Thermocouple meter 20250-91
Type J: –328 to 1832°F
(–200 to 1000°C)
Type K: –418 to 2501°F
(–250 to 1372°C)
Type T: –418 to 752°F
(–250 to 400°C)
Thermistor meter 20250-93
–40 to 257°F
(–40 to 125°C)
RTD meter 20250-95
–320 to 1562°F
(–200 to 850°C)
0.1°F (0.1°C) from –148 to 1832°F (–99.9 to 999.9°C);
1°F (1°C) outside range
0.1°F (0.1°C)
0.1°F (0.1°C) from –148 to 1562°F (–99.9 to 850°C); 1°F (1°C) outside range
Sampling rate: Three times every two seconds Auto power-off:
Unit shuts off automatically after 15 minutes
to preserve battery life Operating temperature: 32 to 122°F (0 to 50°C) Storage temperature: 14 to 122°F (–10 to 50°C) Weight: 6.2 oz (177 g)
1
Dimensions: 7" x 2 Power:
4" x 1" (17.8 x 5.7 x 2.6 cm)
One 9 volt battery
±1 of reading + 1.8°F (1°C)
below –148°F (–99.9°C);
±1 of reading + 0.9°F (0.5°C)
above –148°F (–99.9°C)
±0.9°F (±0.5°C) from
32 to 158°F (0 to 70°C);
±2°F (1°C) outside range
±1.5°F (0.8°C) from –148
to 392°F (–99.9 to 199.9°C);
±2°F (1°C) + 0.5% of reading
outside range
10
3/2/2016 3:16:07 PM
Maintenance, Recalibration, and Repair
1065DGMAN_20250-91-93-95.indd 11
Cleaning and Storage
The meter should be cleaned with a damp cloth and
mild detergent when necessary. Do not use solvents or abrasives.
Store the meter in an area with moderate temperature
and humidity (refer to the operating and storage ranges in the specifications chart earlier in this manual).
Battery Replacement
When the battery power falls low, the low-battery icon will appear on the screen. Replace the one 9 volt battery by removing the screw holding the rear battery compartment cover to access the battery compartment. Ensure that the compartment cover is securely fastened when finished. Note: If you do not intend to use thermometer for a long time, remove the batteries.
It is recommended that Digi-Sense products are calibrated annually to ensure proper function and accurate measure-
ments; however, your quality system or regulatory body may require more frequent calibrations. To schedule your recalibration, please contact InnoCal, an ISO 17025 calibra-
tion laboratory accredited by A2LA.
Phone: 1-866-INNOCAL (1-866-466-6225) Fax: 1-847-327-2993 E-mail: sales@innocalsolutions.com Web: InnoCalSolutions.com