DIGI-SENSE DIGI-SENSE 20250-12 Manual

99 Washington Street Melrose, MA 02176
User Manual
Visit us at www.TestEquipmentDepot.com
Thermohygrometer with Single Type K Input and Superheat Function
with NIST-Traceable Calibration
Model 20250-12
THE STANDARD IN PRECISION MEASUREMENT
Introduction
The Digi-Sense Thermohygrometer (Model 20250-12) with Single Type K Input and Superheat Function offers fast response and high accuracy. This meter is both a dig­ital thermohygrometer and a valuable HVAC diagnostic tool—it measures humidity, dew point, wet bulb, dry bulb, temperature, plus target superheat (TSH) and target evaporator exit temperature (TEET). Meter also features Min/Max/Hold readings and automatic power-off, and accepts a wide variety of thermocouple probes. The instrument is fully tested and calibrated to NIST-traceable standards. Careful use of this meter will provide years of reliable service.
Unpacking
Carefully unpack the instrument and accessories from shipping package. Check individual parts against the list of items below. If anything is missing or damaged, please contact Cole-Parmer immediately.
1. Meter
2. One type K flexible probe
3. Carrying pouch
4. Three AAA batteries
5. User manual
6. NIST-traceable calibration report with data
2
Meter Description
1. Thermocouple type K input
2. Thermistor temperature sensor (internal)
3. MODE button
4. ENTER button
5. OUTPUT button
6. °C/°F /
7. SENSOR button
8. MAX/MIN/HOLD button
9. Power On/Off button
10. LCD
11. Humidity sensor (internal)
12. Battery cover
Backlight button
1
3
4
5
9
2
11
6
7
8
10
12
3
Display Functions
1. Low-battery indication
2. Measurement icons (RH, DP, WB, DB)
3. °C/°F icons for upper display
4. Indoor (ID) / Outdoor (OD) icons
5. Upper measurement display of humidity, dew point, wet bulb, dry bulb, temperature
6. MIN / MAX icons
7. Humidity measurement (%RH) icon
8. HOLD icon
9. Thermocouple (T/C) and Thermistor (NTC) icons
10. MODE select icons (NORM, TEET
11. Auto power-off (APO) indication
Lower measurement display of target superheat
12.
and target evaporator exit temperature
13. °C/°F icons for lower display
, TSH)
2
1
5
9
10
12
4
3
4
6
7
8 11
13
Setup and Operation
Power
Press the Power on/off button to turn unit on.
Taking Measurements
1. Hold the sensor in the area to be tested.
2. Allow adequate time for readings to stabilize.
3. Press the MODE button until the NORM icon is displayed on the screen.
4. Press the SENSOR button and hold for two seconds to select the temperature sensor in use: either T/C for external thermocouple or NTC for internal thermistor.
While in the internal sensor
5. SENSOR button to toggle between other available mea­surements: humidity (RH), dew point (DP), wet bulb (WB), and dry bulb (DB). The icon displayed indicates the selected measurement.
6. Press the °C/°F / Backlight button to select either °C or °F temperature units.
Viewing the Min and Max Readings and Hold Function
1. Press MAX/MIN/HOLD button to step through the mini- mum (MIN), maximum (MAX), and Hold readings.
2.
Press MAX/MIN/HOLD button again to exit the MAX/MIN/HOLD mode.
(NTC) mode, use the
LCD Backlight
Press and hold the °C/°F / Backlight button for two seconds to turn backlight on or off.
5
Setup and Operation (continued)
Automatic Power-Off
To conserve battery life, the meter automatically turns off after 15 minutes. To disable this feature:
1. Turn the meter off.
2. Press and hold the MAX/MIN/HOLD button while powering on.
3. The “APO” will no longer been shown on the display, indicating that the Auto power-off feature is disabled.
Note that Auto power
4. the meter is turned on.
Using Target Superheat Mode (TSH)
Press the
1. displayed on the screen.
Blinking numbers indicate that you are in the
a.
unlocked mode and seeing real-time measurements.
b. Solid non-blinking numbers indicate that the
reading has been “locked.”
c.
Press the new reading, and then press ENTER button again to relock the measurement.
MODE button until the TSH icon is
ENTER button to unlock and take a
-off is re-enabled each time
6
Setup and Operation (continued)
While the numbers are blinking (unlocked mode),
2. press and hold the icon is displayed. Press SENSOR button again to select either the WB or DB icon.
a. When the WB icon is displayed, place the internal
sensor cap in front of the indoor air return and press the ENTER button when the reading is stable.
b. When the DB icon is displayed, place the internal
sensor cap in front of the condenser and press the ENTER button when the reading is stable.
3. Press the OUTPUT button to display the Target Superheat temperature in the lower display.
4. Compare the Target Superheat temperature to the Actual Superheat temperature. Adjust refrigerant levels accordingly for a fixed restrictor system.
a.
If Actual Superheat is higher than T Superheat, add refrigerant.
If Actual Superheat is lower than T
b.
Superheat, recover refrigerant.
SENSOR button until the NTC
arget
arget
7
Setup and Operation (continued)
Using Target Evaporator Exit Temperature (TEET)
T
wo thermocouple probes are required for a TEET
1. measurement.
Press the
2. displayed on the screen.
a. Blinking numbers indicate that you are in the
unlocked mode and seeing real-time measurements.
b. Solid non-blinking numbers indicate that the
reading has been “locked.”
Press the
c.
reading, and then press ENTER button again to relock the measurement.
3. Insert thermocouple “A” into a sock and wet the sock.
4. Clip both thermocouples (thermocouple “A” and thermocouple “B”) in front of the evaporator.
Plug thermocouple “A” into the top of the meter
5. Note: If “OL” is displayed, then the measurements are out of range. Retake the temperature measurements and ensure the temperature inputs are set correctly.
6. Press the SENSOR button until the WB icon is dis­played. Press and hold the SENSOR button until the T/C icon is displayed. When reading is stable, press the ENTER button to lock the reading.
7. Unplug the thermocouple “A” and plug in the thermocouple “B“.
MODE button until the TEET icon is
ENTER button to unlock and take a new
.
8
Setup and Operation (continued)
8. Press the SENSOR button until the DB icon is displayed. Press and hold the SENSOR button until the T/C icon is displayed. When reading is stable, press the ENTER button to lock the reading.
9. Press OUTPUT button and read the Target Evapora- tor Exit Temperature (TEET) in the lower display. Note: If “OL” is displayed, then the measurements are out of range. Retake the temperature measure­ments and ensure the temperature inputs are set correctly.
10. Compare the Target Evaporator Exit Temperature to
the Actual Evaporator Exit Temperature. The Actual Evaporator Exit Temperature is the measured temperature of the air after it has passed through the evaporator. Adjust airflow accordingly.
a. An Actual Evaporator Exit Temperature below the
Target Evaporator Exit Temperature indicates low airflow. Increasing airflow can be accomplished by eliminating restrictions in the duct system, increasing blower speed, cleaning filters or opening registers.
b. An Actual Evaporator Exit Temperature above the
Evaporator Exit Temperature usually indicates low capacity. Occasionally airflow is higher than expected. Look for causes of low capacity such as refrigerant mischarge or a dirty condenser coil. If the airflow is high, correct it by lowering the fan speed.
9
Specifications
Range
Humidity 0 to 100% RH
Temperature, external type K thermocouple (T/C)
Temperature, internal thermistor sensor (NTC)
Dew point (calculated) –22 to 212°F (–30 to 100°C)
Wet bulb (calculated) 32 to 176°F (0 to 80°C)
Resolution
Humidity 0.1% RH
Temperature, external type K thermocouple (T/C)
Temperature, internal thermistor sensor (NTC)
Accuracy
Humidity
Temperature, external type K thermocouple (T/C)
Temperature, internal thermistor sensor (NTC)
–328 to 2501°F (–200 to 1372°C)
32 to 140°F (0 to 60°C)
0.1° (<1000°) or 1° (1000°)
0.1°
±2.5% RH from 10 to 90% RH ±5% RH from 0 to 10% RH and 90 to 100% RH
±[0.5% reading + 1.8°F (1°C)] above –148°F (–100°C) ±[0.5% reading + 3.6°F (2°C)] below –148°F (–100°C)
±1.8°F (1°C)
10
Specifications (continued)
Auto power-off Unit shuts off automatically after
Operating temperature 32 to 122°F (0 to 50°C)
Storage temperature 14 to 122°F (–10 to 50°C)
Weight 6 oz (173 g)
Dimensions 7
Power Three AAA 1.5 V batteries
15 minutes to preserve battery life
1
4" x 21⁄2" x 1" (18.5 x 6.3 x 2.6 cm)
Maintenance
Cleaning and storage
1. The meter should be cleaned with a damp cloth and mild detergent when necessary. Do not use solvents or abrasives.
2. Store the meter in an area with moderate temperature and humidity (refer to the operating and storage range in the specifications chart earlier in this manual).
Battery Replacement
When the battery power falls low, the low-battery icon will appear on the screen. Replace the three AAA batteries by removing the screw holding the rear battery compartment cover to access the battery compartment. Ensure that the compartment cover is securely fastened when finished.
11
Maintenance, Recalibration, and Repair
It is recommended that Digi-Sense products are calibrated annually to ensure proper function and accurate measure-
ments; however, your quality system or regulatory body may require more frequent calibrations. To schedule your recalibration, please contact InnoCal, an ISO 17025 calibra­tion laboratory accredited by A2LA.
Phone: 1-866-INNOCAL (1-866-466-6225) Fax: 1-847-327-2993 E-mail: sales@innocalsolutions.com Web: InnoCalSolutions.com
For Product and Ordering Information, Contact:
99 Washington Street Melrose, MA 02176
Phone 781-665-1400 Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com
1065DGMAN_20250-12 Rev.1
Manual Part No. 00100-50
Loading...