Digi-Sense 93210-50 User Manual

Page 1
Cole-Parmer Instrument Co.
625 East Bunker Court
Vernon Hills, Illinois U.S.A. 60061-1844
(847) 549-7600
(847) 247-2929 (Fax)
800-323-4340
www.coleparmer.com
A-1299-0879
Edition 01
ThermoLogR
THERMISTOR THERMOMETER
MODEL NO . 93210-50
Page 2
CERTIFICA TE OF
CONFORMANCE
This thermometer was calibrated using equipment traceable to the National Insti­tute of Standards and Technology (NIST).
This instrument conforms to ITS-90.
The accuracy of the thermometer at the time of calibration was within specifications stated in the operating manual.
Model No.:__________________________ Serial Number:______________________ Date placed in service:________________ To purchase a NIST certificate of traceabil-
ity with test data and test date for meter and probe, please contact your dealer or:
Cole-Parmer Instrument Company
625 East Bunker Court
Vernon Hills, Illinois USA 60061-1844
Toll-free: 800-323-4340
Page 3
1
INTRODUCTION
This versatile hand-held instrument provides highly accurate temperature measurements in Celsius or F ahrenheit using a wide selection of 400, 500 and 700 series thermistor probes.
This instrument covers these measuring ranges: 400 Series: –40.00°F to +302.0°F (–40.00°C to
+150.0°C)
500 Series: –40.00°F to +302.0°F (–40.00°C to
+150.0°C)
700 Series: –40.00°F to +212.0°F (–40.00°C to
+100.0°C)
The instrument is designed for easy oper ation and includes the follo wing features:
• Operator selection of temperature scale
• Resolution of 0.01° from –40.00°C to +99.99°C or –40.00°F to +99.99°F
• LCD with three four-digit displa ys
• 1/4 in diameter input
• Hold feature f or temporarily retaining a reading
Page 4
2
• Field calibration capability
• Low battery warning
• Stores or logs up to 1000 readings with real-time markers
• Scrolls through all stored readings
• Displays MIN and MAX readings
• Interfaces with optional HEWLETT PACKARD
®
infrared printer or optional RS-232-C adapter
• Prints temperature and time of reading
• Built-in tilt stand for easy hands-free operation
Page 5
3
SAFETY PRECAUTIONS
THIS INSTRUMENT IS
DESIGNED TO ACCEPT LOW LEVEL SIGNALS SUPPLIED BY STANDARD THERMISTORS . UNDER NO CIRCUMST ANCES SHOULD THE INPUT VOLTAGE EXCEED THE SPECIFIED 10 V RMS .
DO NOT USE OR STORE
THIS INSTRUMENT IN MICROWAVE OVENS OR ANY ABNORMALLY HOT OR COLD AREAS .
WEAK BA TTERIES SHOULD
NOT BE LEFT IN THE INSTRUMENT. DEAD BATTERIES CAN LEAK AND CA USE D AMAGE T O UNIT.
VOLTAGES PRESENT AT
THE THERMISTOR MAY ALSO BE PRESENT AT THE BATTERY TERMINALS. ALWAYS DISCONNECT THE THERMISTOR WHEN CHANGING BATTERIES.
W ARNING
D ANGER
CAUTION
CAUTION
Page 6
4
SPECIFICATIONS
THERMISTOR PR OBES T ype T emperature Range
400 Series: –40.00°F to +302.0°F
(–40.00°C to +150.0°C)
500 Series: –40.00°F to +302.0°F
(–40.00°C to +150.0°C)
700 Series: –40.00°F to +212.0°F
(–40.00°C to +100.0°C)
Accuracy: ±0.06°F (±0.03°C) exclusive of probe Linearization: Conforms to ITS-90. Input Protection: 10V rms Display Update: 0.6 seconds per update Input: One probe with 1/4 inch phone jack
connector.
Battery
Size: Two AA, 1.5V alkaline ANSI-L40,
IEC-LR6
Life: >500 hours continuous, typical
Display: LCD with 0.4 in high characters main
readout and 0.2 in high characters secondary displays, 4 digits each display plus various annunciators.
Operating:
Stated Accuracy: 0°C to 40°C
(32°F to 104°F)
Useful Range: 0°C to 40°C
(32°F to 104°F)
Storage: –40°C to 65°C
(–40°F to 149°F)
Humidity: 10% to 90%
(non-condensing)
Page 7
5
Dimensions
3 cm D x 8.4 cm W x 15.8 cm H (1.2 in x 3.3 in x 6.2 in)
Weight with batteries: 227 grams (8 ounces) Ingress protection: Meets IEC-529 IP-54 for dust
and water resistant enclosures.
Electromagnetic Compatibility (IEC-801-3)
Unit exceeds radiated susceptibility requirements of ENV -50140.
Page 8
6
BATTERY INSTALLATION AND
REPLACEMENT
WEAK BA TTERIES SHOULD
NOT BE LEFT IN THE INSTRUMENT. DEAD BATTERIES CAN LEAK AND CA USE D AMAGE T O UNIT.
VOLTAGES PRESENT AT
THE THERMISTOR INPUT MAY ALSO BE PRESENT AT THE BATTERY TERMINALS. ALWAYS DISCONNECT THE THERMISTOR WHEN CHANGING BATTERIES.
CAUTION
D ANGER
Page 9
7
If battery indicator turns on, battery life is approximately 8 to 10 hours. The battery indicator will start blinking with less than 1 hour of life remaining.
AT THIS POINT , BATTERY MUST BE CHANGED . IF BA TTER Y VOL T AGE GOES TOO LOW , DISPLA Y WILL GO BLANK.
See SPECIFICATIONS f or battery type.
1. Before changing battery , turn instrument off and disconnect probe.
2. Loosen screw and lift battery cover off back of case.
3. Remov e the two AA batteries.
4. Insert two new batteries, observing polarity .
5. Install cov er and tighten screw .
AC ADAPTER
An optional AC adapter can be used f or po w er to conserve batteries. The adapter is not a charger and will not charge batteries. The adapter connects to the bottom of the thermometer. When the adapter is connected, the batteries are disconnected. Refer to the ACCESSORIES section for adapter information.
Page 10
8
CONNECTING A THERMISTOR
Use either a 400 series or 700 series ther mistor with your instrument. Be sure your instrument setting matches the thermistor, using an incorrect thermistor type will result in erroneous readings. (Series 500 thermistor probes may be used along with the supplied correction chart to convert meter readings to actual temperature.)
T o change between 400 and 700 series thermistors, see Selecting Thermistor Series on page 14.
Insert the 1/4 in plug into the mating connector on the top of the instrument.
Page 11
9
Page 12
10
Page 13
11
Page 14
12
QUICK SETUP
Note: Revie w warnings on page 3.
1. Install batteries.
2. Connect thermistor probe.
3. Press the ON/OFF key. The thermometer performs a self-test and all display digits and indicators, as shown below, should remain on for approximately one second.
The clock will remain on f or about three seconds.
4 . Use SETUP to select the correct thermistor
ser ie s and v arious other options. Use s or t to make selections, to enter selections and go to next step .
Page 15
13
If a probe has not yet been connected to the input on the instrument, you will see this display:
This message also appears if a probe is broken. No measurements can be made while this warning is display ed.
Page 16
14
COMPLETE SETUP PROCEDURE
The setup function scrolls through a series of steps for selecting thermistor series, resolution, filter rate, setting the real-time clock, selecting the line frequency and accessing the Log sub-menu and the Print sub-menu. Either the complete setup can be run as described below, or the setup can be initiated and terminated at any step. The Log sub­menu and the Print sub-menu can be accessed without going through the complete setup. These variations are cov ered in the following procedures.
NOTE
Settings selected in the Setup procedure are stored in memory and will remain even after power is turned off or while batteries are being replaced.
Press the SETUP key. The lower left display will show either the time of day or “OFF” and it will be flashing.
SELECTING LO WER DISPLA Y MODE
Press the MAX s key or MIN t k ey to toggle the lower display betw een showing the time of day and showing minimum and maximum temperature
values. Press the HOLD key to adv ance to the next step in the setup sequence.
SELECTING THERMISTOR SERIES
Press the MAX s key or MIN t key to select either 400 or 700 series ther mistors. When the correct thermistor series is flashing, press the HOLD
key to advance to the next step in the setup
sequence.
TEMPERA TURE RESOLUTION
There are three choices for temperature resolution. The first choice is with 0.01° resolution is selected: the thermometer will automatically range to show the greatest possible resolution. The next choice is
Page 17
15
0.1° resolution: the thermometer will automatically range between 0.1° and 1° as required. The final choice is 1° resolution: the display will remain in this resolution for all displa yed temperatures.
When 0.01° or 0.1° resolution is selected, a decimal point will appear in the upper numerical display . When 1° resolution is selected, no decimal point will be present.
The upper display will show 0.01, 0.1 or 1 and it will be flashing. Press either MAX s k ey or MIN t key to change resolution, then press the HOLD
to advance to the next step in the setup
sequence.
FIL TER RA TE
The filter rate can be set to OFF, Lo or Hi. This r ate is changed to smooth out fluctuations in the readings. Normally this would be set to OFF. If readings are unstable, try changing to Lo or Hi.
The lower left display will read “FILt” and the upper display will flash either OFF, Lo or Hi. Use the MAX s key or MIN t ke ys to switch the dampening rate, the press HOLD key to store the rate and continue setup by setting the real-time clock.
SETTING REAL-TIME CLOCK
1. When setting the real-time clock, the right two digits in the lower left display will be flashing. Use the MAX s k ey or MIN t k ey to adjust the flashing digits to the desired minute setting.
2. Press HOLD key to lock in the minutes and advance to the hours setting.
Hours are set in the 24 hour time system. Theref ore, time after 12:59 pm requires 12 to be added to the time. For e xample, 3:00 PM is (3+12) = 15:00 hours.
Page 18
16
3. The left two digits in the lower left display will be flashing. Use the MAX s key or MIN t key to adjust the flashing digits to the desired hour setting.
4. Press the HOLD ke y to store the hours and advance to the month setting.
5. The MO annunciator will be on and the left two digits in the lower right display will be flashing. Use the MAX s key or MIN t key to adjust the flashing digits to the desired month setting (1 to 12).
6. Press the HOLD key to store the months and advance to the da y setting.
7. The DAY annunciator will be on and the right two digits in the lower right display will be flashing. Use the MAX s key or MIN t key to adjust the flashing digits to the desired day setting (1 to 31).
8. Press the HOLD key to store the day and advance to the year setting.
9. The YR annunciator will be on and the four digits in the lower left displa y will be flashing. Use the MAX s key or MIN t ke y to adjust the flashing digits to the desired year setting (2000 to 2063).
10 . Press the HOLD key to loc k in the year . To
set up the log or print functions proceed to the follo wing paragraphs.
LINE FREQUENCY SELECTION
For maximum noise rejection, it is possib le to select one of two line frequency settings. This selection should be based on the local mains power line frequency of either 50 Hz or 60 Hz. For example , all of North America is 60 Hz, while Europe is 50 Hz.
Page 19
17
The lower left display will show “LINE” while the upper display is flashing either 50 or 60. Use the MAX s key or MIN t key to adjust the flashing digits to the desired line frequency setting.
Press the HOLD key to lock in the line frequency . This completes the basic oper ation setup and returns to normal operation.
LOG SETUP
The Log sub-menu is used to set the time intervals between the logged readings. The time can be set from 1 second to 60 minutes. Once the log sub-menu is setup , logging can be toggled ON or OFF by pressing the LOG key. When logging is turned ON, it will continue until turned OFF or 1000 logs have occurred at the entered rate. The first 1000 logs are retained.
1. Anytime while in SETUP, press the LOG key. The LOG annunciator will be on and the presently set time interval (default is one second) will be displayed in the lower left display. The two right digits represent seconds and will be blinking.
2. Use the MAX s key or MIN t keys to adjust the “seconds” as desired. Press the HOLD
key to store setting and adv ance to “minutes”
setting.
3. Use the MAX s key or MIN t keys to adjust the “minutes” as desired. Press the HOLD key to store setting. This completes the Logging setup .
Page 20
18
PRINT SETUP
The Print sub-menu is used to set the time intervals between the readings being sent to the printer. The default rate is once e v ery three seconds. The time can be set from 3 seconds to 60 minutes. On ce the print sub-menu is setup, printing can be toggled ON or OFF by pressing the PRINT ke y . In addition, the data rate for the infrared RS-232-C output can also be set.
1. Anytime the PRINT key is pressed while in SETUP, the PRINT annunciator will be on and the presently set time interval (default is three seconds) will be displayed in the lower left display. The two right digits represent seconds and will be blinking.
2. Use the MAX s key and/or MIN t keys to adjust the “seconds” as desired. Press the HOLD
key to store setting and advance to the
“minutes” setting.
3. Use the MAX s key and/or MIN t keys to adjust the “minutes” as desired. Press the HOLD
key to store setting and proceed to the data
output rate.
4. Use the MAX s key and/or MIN t keys to scroll through the available data output rates of HP, 300, 600, 1200 or 2400 (use HP for the HEWLETT-PACKARD infrared printer). Press the HOLD key to store setting and end the setup. The data output rates of 300, 600, 1200 and 2400 are for use with the optional infrared to RS-232-C adapter.
SELECTING INDIVIDU AL PARAMETERS
The RTD sensitivity, resolution, filter update rate, real-time clock and line frequency can be set individually without performing the complete setup . First press the SETUP key, then repeatedly press the HOLD key until the desired function is display ed. Use the s t keys to change the function and press the HOLD key to store setting.
Page 21
19
OPERATING PROCEDURES
The unit will always power up with the upper display showing temperature. If a probe is not connected at power up , the upper display indicates OPEN.
For optimum oper ation, allow about one minute for ambient temperature stabilization. If the unit has been stored at an extreme ambient condition, more time may be needed.
BASIC TEMPERATURE MEASUREMENTS
Check that the thermometer is turned on, the probe connected, the desired resolution 0.01° or 0.1° or 1° is selected, and the desired scale °F or °C is selected.
SELECTING THE TEMPERATURE SCALE
Use the °C/°F key for selection of the Fahrenheit or Celsius scale. The last selection is retained in memory even if power is s witched off .
MAXIMUM READINGS
The maximum reading function displays the maximum reading since power up or since the last time the clear function was used. The maximum reading function is ideal for monitoring unattended operations while continually displaying every temperature change that occurs. The maximum and minimum values are sensed and automatically stored until you are ready to observe the reading.
Do not turn the instrument OFF when a maximum or minimum temperature value may be needed; MAX/MIN memory contents will be lost.
Page 22
20
Clearing a Maximum Reading
Press the CLEAR key then press the MAX s key . The maximum memory will be cleared.
MINIMUM READINGS
The minimum reading function displays the minimum reading since power up or since the last time the clear function was used. While continually displaying e very temperature change that occurs, the maximum and minimum values are sensed and automatically stored until you are ready to observe the reading.
Do not turn the instrument OFF when a maximum or minimum temperature value may be needed; MAX/MIN memory contents will be lost.
Clearing a Minimum Reading
Press the CLEAR key then the MIN t key. The minimum memory will be cleared.
HOLD FUNCTION
Press the HOLD key to retain the reading on the display. Press HOLD key again for normal operation.
Maximum Hold Press the MAX s key, then press the
HOLD key. To turn off the HOLD function and
return to normal operation, press the HOLD key again.
T o clear the maximum readings , press the CLEAR key, then the MAX s key.
Page 23
21
Minimum Hold
Press the MIN t key, then press the HOLD key. To tur n off the HOLD function and return to normal operation press the HOLD key again.
T o clear the minim um readings, press the CLEAR key, then the MIN t key.
STORED READINGS
The store function allows you to store up to 1000 readings. Each reading is logged with a storage location number and the time of occurrence. To store readings, proceed as follo ws:
1. Momentarily press the STORE k ey . The present temperature is stored. The upper main display will momentarily show the storage location number and the ST ORE annunciator will be on.
After three seconds, the storage number will be replaced with a temperature reading but the ST ORE annunciator remains on to indicate a temperature reading has been stored. The ST ORE key ma y be pressed as fast as once per second.
2. Repeat step 1 for all the points to be recorded, up to a maximum of 1000. Each time the STORE key is pressed, the new reading will be stored and the upper main display will show the stor age location number f or about 3 seconds. After 1000 stored readings, the ne xt time the STORE key is pressed the main display will indicate FULL.
Page 24
22
RECALL READINGS
This function allows the stored readings, the reading sequence number and the Time/Date of the readings to be recalled. RECALL shows each stored reading individually . When individual readings are recalled, you can toggle between the reading and the sequence number of the reading. To recall readings, proceed as follows:
1. Momentarily press the RECALL key, the time and date will be display ed on the lower displays and the stored sequence number will be display ed on the upper main display . The ST ORE and RECALL annunciators will be on.
2. To step through the readings, press the MAX s key or the MIN t key. Each press of the k ey will advance to the ne xt reading in sequence. Hold the MAX s key or MIN t key do wn to advance continuously through the readings at an increasing rate. To increment or decrement by 100, proceed as follo ws:
To increment, hold down the MAX s key then press the MIN t ke y.
To decrement, hold down the MIN t key then press the MAX s key.
3. To toggle betw een the readings and the reading sequence number , press the RECALL key again.
4. To return to normal operation, press any other key e xcept ON/OFF , MAX s, MIN t or RECALL.
CLEARING STORED OR LOGGED READINGS Press the CLEAR key, then the STORE key or
LOG key. Regardless of which key is pressed, STORE or LOG, the stored and logged readings
in memory will be cleared.
Page 25
23
LOGGING READINGS
The Logging function is controlled by the LOG key . Prior to logging readings, it is necessary to set up the time interval between readings (see LOG SETUP) in the COMPLETE SETUP PROCEDURE section. The time betw een readings can be set to any value from 1 second to 60 minutes. Logging is initiated by pressing the LOG key. Logging will continue at the programmed rate until either the LOG key is pressed again or 1000 logs have occurred. The display will then show “FULL”.
The STORE function can be used during logging to insert additional measurements by pressing the
STORE key.
PRINTING
Printing can output real-time readings or stored readings. The output is sent to the infrared (IR) printer output located at the top of the thermometer. The printer output default time period is once ev ery 3 seconds. This time can be set to any value between 3 seconds and 60 minutes (see PRINT SETUP) in the COMPLETE SETUP PROCEDURE section. The following sample printouts show the RS-232-C format and the HEWLETT-PACKARD infrared printer in normal and log dump print modes. Note that in log dump print mode the first line specifies the total number of readings.
The print function is controlled by the PRINT key .
Page 26
24
HP Printer Format for Normal Printing
New heading will appear when date changes or new print is started.
---- = over r ange
06-01-00 °F 13:20:54 30.56 13:20:57 32.22 13:21:00 32.44 13:21:03 164.2 13:21:06 -103.1 13:21:09 -58.87 13:21:12 -58.92 13:21:15 510.4 13:21:18 1034 13:21:21 ---­13:21:24 1034 06/01/00 °C 13:21:27 556.3 13:21:30 556.8 13:21:33 0.01 13:21:36 0.04 06/01/00 °F 13:21:39 32.05
Page 27
25
HP Printer Format for Log Dump
New heading will appear when date changes.
---- = over r ange
TOTAL READINGS:1000 02-02-00 °F 19:01:00 ---­19:01:02 -321.6 19:01:00 -279.5 19:01:07 -236.7 19:02:19 -192.9 19:01:01 -149.0 19:01:01 -104.2 19:01:03 -58.95 19:01:00 -13.51
TOTAL READINGS:1000 02-02-00 °C 19:01:00 ---­19:01:02 -196.4 19:01:00 -173.0 19:01:07 -149.3 19:02:19 -124.9 19:01:01 -100.5 19:01:01 -75.69 19:01:03 -50.53 19:01:00 -25.29 19:01:01 -0.06 19:01:00 25.52
Note that printing 1000 stored sets of readings could take nearly one hour as limited by the printer .
Page 28
26
RS-232-C Format for Normal Printing and Log Dump
---- = over range
“DATE”, “TIME”, “DEGREES F” 07-13-00,16:30:20, 72.56 ,16:30:30, 72.55 ,16:30:40, “OPEN” ,16:30:50,"----"
The print function is controlled by the PRINT key . Proceed as follows:
1. P erf orm the PRINT Setup procedure to set the desired print rate.
2. Check that the IR printer input is properly aligned with the infrared output at the top of the thermometer.
Page 29
27
Printer output will occur in real-time when the thermometer is operating in the normal mode. When the thermometer is in the RECALL mode, printing of the stored information will occur .
3. Press the PRINT key to star t pr inting. Printing will continue at the programmed print rate until the PRINT k ey is pressed again, or if stored data is being printed and all data has been transferred to the printer. When printing stored data, the temperature and times of occurrence will be output.
CALIBRATION
The calibration function allows both single point and dual point calibration of the thermometer. Single point calibrates the offset only . Dual point calibrates the offset first then calibrates the slope. The thermometer can be calibrated at any temperature.
It is not necessary to perform a field calibration to obtain specified accuracies. Use the field calibration feature to impro ve thermometer/probe accuracy or to compensate for probe calibr ation drift.
The thermometer has a memory retention capability to hold calibration values e ven while power is off or the battery is removed.
When you restart, there is no need to recalibrate.
NOTE
The temperature at which water boils (100°C/ 212°F) is at sea level and standard atmospheric pressure using distilled water . Changes in altitude and barometric pressure will cause the boiling temperature to change.
As a rule of thumb, the boiling temperature of water will decrease 1°C (1.8°F) for every 1000 feet abo ve sea lev el. For e xample, at an elev ation of 5300 feet (1600 meters), water will boil at approximately 94.7°C (202.5°F).
Page 30
28
Other liquids may also be used as calibration standards. Consult a chemical handbook for their freezing (melting) point and boiling point properties.
When calibrating at freezing (0°C or 32°F), it is recommended to use crushed ice made of distilled water in a dewar flask. Add crushed ice to top of flask. Top off flask with distilled water. Continue to add crushed ice to maintain tightly packed crushed ice/water in flask.
CALIBRA TION PROCEDURES
Calibration Procedure
This calibration function provides for both an offset and slope field calibration. For proper calibr ation, the following conditions m ust be observed:
• The slope point must be a higher temperature than the offset point.
• The difference must be at least 20°C (68°F).
•Use two points based on the expected high and low temperatures. Temperatures measured outside of these limits may no longer meet specifications.
• Resolution is based on the selected display resolution.
Proceed as follows to calibrate the probe .
1. Place the probe at the lower known reference
temperature.
2. Offset Calibration: Momentarily press the CAL
key to enter the calibration mode, the CAL annunciator will flash. The temperature is displayed on the main display and Lo is display ed on the lower left display. “Lo” signifies that this is the offset point.
Page 31
29
3. Allow the reading to stabilize. If the displayed temperature is higher or lower than the ref erence temperature, use the MAX s key to increase the displayed reading or the MIN t key to decrease the displayed reading until the ref erence temperature is display ed. The MIN t or MAX s must be pressed at least once. The CAL annunciator should be blinking during this procedure.
4. Press the HOLD key to lock the offset calibration in and advance to the slope calibration. T o return to normal operation, press any ke y e xcept CAL or HOLD .
5. Slope Calibration: Place the probe at the higher reference temperature.
6. Allow the reading to stabilize. If the displayed temperature is higher or lower than the ref erence temperature, use the MAX s key to increase the displayed reading or the MIN t key to decrease the displayed reading until the reference temperature is displa yed. The MIN t
or MAX s key must be pressed at least once.
The CAL annunciator should be blinking during this procedure.
7. Press the HOLD key to loc k the calibration in.
Clearing Cal Point: Press CLEAR, then CAL. Unit rev erts to factory calibration with no offset or slope compensation.
FIELD CALIBRA TION LOCKOUT AND RE-ENABLE
The calibration lockout feature prevents any field calibration changes. The lock out remains in effect until a lockout re-enab le has been performed. Use the following procedures to loc kout or re-enable the field calibration operation.
Page 32
30
Lockout Pr ocedure
1. Turn the thermometer off.
2. Simultaneously press and hold the CAL and the
CLEAR keys down and momentarily press the ON/OFF key. Continue to hold the CAL and CLEAR keys f or at least 5 seconds.
Re-Enable Pr ocedure
1. Turn the thermometer off.
2. Simultaneously press and hold the HOLD and the CAL keys down and momentarily
press the ON/OFF key. Continue to hold the
HOLD and CAL k eys until the display blanks .
MAINTENANCE AND
TROUBLESHOOTING
Properly used, the thermometer should maintain calibration indefinitely and not require service other than occasional cleaning of the housing and changing of the batteries.
Do not clean with abrasives or solv ents. Use mild detergents, nev er immerse nor use excessive fluid.
BA TTERIES
If there is no display when the thermometer is turned on, check condition of the two AA batteries. Also check that the battery terminals are clean and batteries are properly installed. If replacement is necessary , refer to the BATTERY INSTALLA TION AND REPLACEMENT section for replacement procedure.
The real-time clock will keep time f or up to 1 minute with the batteries removed. To minimize the need for resetting the cloc k, either remove and replace one battery at a time, or connect the AC adapter
Page 33
31
while changing the batteries. All stored/logged readings are retained until cleared, even if the batteries are removed f or long periods.
SER VICE
There are no internal adjustments or user replaceable parts.
If “Err” followed by the numbers 1 through 9 is displayed (see example below), return unit for service. Note that “Err” alone may be displayed during improper field calibration.
Note: Serial number label is located inside battery compartment.
Page 34
32
A-1925-41
Rev. A
Printed in U.S.A.
010600
TROUBLESHOOTING
The following chart lists the most probable faults . There are no internal adjustments or user replaceable parts. If this does not solve the problem, refer service to your dealer .
F A ULT ACTION
No display when Check condition of turned on batteries.
Check that batteries are inserted properly . Check operation using A C adapter.
Display shows Out of range indication.
- - - ­Display shows Open probe connection.
OPEn Display sho ws If displayed at any time
Err other than during field
calibration, return instrument for service.
If “Err 1” through “Err 9” remains on the display, return instrument for service.
Page 35
ACCESSORY THERMISTOR PROBES
COLE-PARMER offers a wide variety of 400, 500 and 700 series thermistor probes, connectors, ex­tension cables, switch boxes, accessories, and cali­bration services.
The general-purpose and penetration probes listed below have a nylon handle, a 5-ft. coiled cord, a 316 SS sheath and a 1/4 in phone plug. The sur­face and air/gas probes feature a vinyl-covered 10-ft. lead with a 1/4 in phone plug. All probes listed are 400 series thermistors.
General-purpose probe, 10 in long x 0.188 in dia. 4 second time constant.
93824-00 TEFLON
®
-coated general-purpose probe, 10 in
long x 0.200 in dia. 7 second time constant.
93824-12 Penetration probe, 4 in long x 0.188 in dia sharp
tip. 6 second time constant.
93824-30 Surface probe, 0.3750 in dia disk. 1.1 second
time constant.
08442-15 Air/gas probe, 1.875 in long with 0.5 in dia stainless
steel radiant heat shield. 10 second time constant.
08438-00 37000-98 Carrying case. Hard plastic case with
foam insert holds one meter and up to five probes. Measures 21 in W x 12 in H x 3 in D.
8520-05 Soft carrying case. Water-resistant cloth. Extra room for batteries or small probe. Measures
3.5 in W x 10 in H x 2 in D. 91100-52-115V AC adapter. For 93210-50
thermometer. 91100-55-230V AC adapter. For 93210-50
thermometer. European plug. 37455-70-Infrared printer includes one roll of pa-
per and four AA batteries. 37455-80-Adapter, 110–240V AC, 50/60 Hz, 6-ft
(1.8 m) cord. 91100-85-RS-232-C interface connects to your
computer 9-pin serial port for easy uploading of data. Use with most popular terminal programs.
Page 36
W ARRANTY
The Manufacturer warrants this product to be free from significant deviations from published specifications. If repair or adjustment is neces­sary within the warranty period, the problem will be corrected at no charge if it is not due to mis­use or abuse on your part as determined by the Manufacturer. Repair costs outside the warranty period, or those resulting from product misuse or abuse, may be invoiced to you.
The warranty period for this product is noted on the Warranty Card.
PRODUCT RETURN
To limit charges and delays, contact the seller or Manufacturer for authorization and shipping in­structions before returning the product, either within or outside of the warranty period. When returning the product, please state the reason for the return. For your protection, pack the prod­uct carefully and insure it against possible dam­age or loss. Any damages resulting from im­proper packaging are your responsibility.
TECHNICAL ASSIST ANCE
If you have any questions about the use of this product, contact the Manufacturer or authorized seller.
HEWLETT-PACKARD - Reg TM Hewlett-Packard Corp. TEFLON - Reg TM E.I. DuPont de Nemours & Co. Trademarks bearing the ® symbol in this publication are registered in the U.S. and in other countries.
A-3477-79
Rev. A
Printed in U.S.A.
010600