Digi-Sense 91428-07 Instruction Manual

Instruction Manual
Single-Input
(Model 91428-07)
68X001350 Rev0 0717 065DGMAN_91428-07
TABLE OF CONTENTS
QUICK-START GUIDE ............................. 1
1. INTRODUCTION ................................ 2
2. SAFETY PRECAUTIONS ................... 3
3. SPECIFICATIONS .............................. 4
4. BATTERY INSTALLATION AND
REPLACEMENT ................................. 6
5. INSERTING AND REMOVING
OPTIONAL RUBBER ARMOR ............ 7
6. ASSEMBLING OPTIONAL HANDS-
FREE ACCESORIES .......................... 8
7. CONNECTING AN RTD PROBE ........ 9
8. KEY FUNCTIONS ..............................10
9. DISPLAY OVERVIEW .......................11
10. MEASUREMENT MODE ...................12
11. SELECTING TEMPERATURE
SCALE ..............................................13
12. MIN AND MAX FUNCTION ................14
13. AUTO OFF FUNCTION .....................14
14. CALIBRATION ...................................15
15. CLEAR FIELD CALIBRATION ...........16
16. FIELD CALIBRATION UNLOCK ........16
17. FIELD CALIBRATION LOCKOUT ......17
18. MAINTENANCE .................................18
19. CLEANING ........................................18
20. TROUBLESHOOTING .......................19
21. SERVICE ...........................................20
22. REPLACEMENT METER &
ACCESSORIES .................................20
23. WARRANTY ......................................21
24. PRODUCT RETURN .........................21
25. INNOCAL® CALIBRATION AND
REPAIR SERVICES ..........................22
- 1 -
QUICK-
START
GUIDE
To ensure best results please read the complete manual.
1. Connect probe.
2. Press the on/off key.
Press again to turn on backlight.
3. Select °C or °F display using °C/°F key.
4. Insert probe in sample.
Allow adequate time for reading to stabilize. Response time will vary depending on probe.
5. Take reading.
Press hold key to freeze display. Press min/max to display minimum or maximum temperature.
6. Press and hold the on/off key for three
seconds to turn meter off.
- 2 -
1. INTRODUCTION
This versatile handheld instrument provides highly accurate temperature measurements and is designed for easy operation, which includes the following features:
Operator selection of Celsius or
Fahrenheit scale
Resolution of 0.1° (auto ranging)
Four-digit backlit LCD
Hold feature to retain a reading
temporarily
Field calibration capability
Low-battery warning
Auto power off with Enable/Disable
function
Min and Max readings display
3-pin circular connector input
Operates with a wide selection of
probes
- 3 -
2. SAFETY PRECAUTIONS
Voltages present at the RTD may also be present at the battery terminals. Always disconnect the RTD when changing batteries.
This instrument is designed to accept low level signals supplied by standard 100 Ω RTDs. Under no circumstances should the input voltage exceed the specified 10V rms.
1. Do not use or store this instrument in microwave ovens or any abnormally hot or cold areas.
2. Weak batteries should not be left in the instrument. Dead batteries can leak and cause damage to unit.
WWARNING
- 4 -
3. SPECIFICATIONS
Platinum RTD Probes
100 ohm platinum: alpha = 0.003850 –200 to 850°C (–392 to 1562°F)
Out-of-range display: “Ur”, “Or”, “OPEN
Range
Accuracy
–200.0 to –100.0°C
(–392 to –148°F)
±2.0°C
(±4.0°F)
–99.9 to 199.9°C
(–148 to 392°F)
±0.2°C
(±0.4°F)
200.0 to 850.0°C (392 to 1562°F)
±2.0°C
(±4.0°F)
Resolution
0.1° (–200.0 to 850.0°C);
0.1° (–392.0 to 999.9°F); 1° (1000 to 1562°F)
Display
4-digit LCD (¼ x ½" digits); 2 1/3 x 1 ½" backlit viewing
area
Display update rate
0.5 sec per update
Input
One probe with 3 pin circular connector (Switchcraft TA3F)
Input protection
10V rms
Battery
Three AA, 1.5V alkaline
Battery life
700 hours continuous
Auto shutoff
17 minutes after last keypress
Stated accuracy
18 to 28°C (64 to 82°F)
Useful range
0 to 40°C (32 to 104°F)
Storage
–40 to 65°C (–40 to 149°F)
Humidity
10% to 90% RH (noncondensing)
Dimensions
Without armor With armor
(L x W x H ) 175 mm x 97 mm x 42 mm
180 mm x 102 mm x 52 mm
- 5 -
Weight
Without armor With armor
267 g 362 g
Ingress protection (with probe attached)
Meets IEC-529 IP54 for dust and water-resistant enclosures
CE compliance
EN61326-1/A1: 1998 (EU EMC Directive)
- 6 -
4. BATTERY INSTALLATION AND REPLACEMENT
Voltages present at the RTD may also be present at the battery terminals. Always disconnect the RTD when changing batteries.
Weak batteries should not be left in the instrument. Dead batteries can leak and cause damage to unit.
The typical battery life is about 700 hours. Selected settings are stored in memory and will remain in memory even after power is turned off, or while batteries are being replaced.
1. Before
changing
battery, tur
n
instrume
nt off
and disconnect probe
.
2.
Loosen screw and
lift battery
cover
off
the
back
of case.
3.
Remove
the t
hree
AA batteries.
4. Insert three
new
batteries
observing
pol
arity.
5.
Install
cover and
tighten
screw
.
- 7 -
5. INSERTING AND REMOVING OPTIONAL RUBBER ARMOR
Figure 1: Removing Optional Rubber Armor
1. To insert thermometer into the optional rubber armor, slide in from the top of meter before pushing the bottom edges of meter down to set it into position. Lift up the stand at the back of meter for bench top applications if necessary.
2. To remove thermometer from armor, push out from the bottom edges of meter until it is completely out of boot.
- 8 -
6. ASSEMBLING OPTIONAL HANDS-FREE
ACCESORIES
You can use the optional magnets and strap in the Hands-Free Kit accessories for hands-free operations.
Figure 2: Hands-Free Kit
- 9 -
7.
CONNECTING
AN RTD PROBE
Use the correct 100 Ω RTD (alpha =
0.003850) for your instrument. Using an
incorrect probe type will result in erroneous readings. Insert the 3-pin plug into the mating connector on the top of the instrument.
Figure 3: Top view of RTD thermometer
RTD Input
- 10 -
8. KEY FUNCTIONS
Figure 4: Keypad
Toggles between Celsius or Fahrenheit in measurement mode and user calibration mode.
Press and hold this key for 3 seconds to enter Calibration mode if field calibration is unprotected.
Activates/deactivates freezing of the measured reading while in measurement mode.
Toggles between minimum and maximum readings (MIN and MAX readings are calculated from the last min/max keypress).
Press and hold this key for 3 seconds to clear the MIN/MAX reading stored.
Powers on and shuts off the thermometer by holding it for 3 seconds.
Press this key to activate/ deactivate the backlight display. (Backlight will automatically turn off within 30 seconds of activation).
°C/°F
cal
hold
min
max
on/off
light
- 11 -
9. DISPLAY OVERVIEW
Figure 5: Meter Display
°C/°F
Celsius or Fahrenheit indicator.
MIN
Minimum reading annunciator.
MAX
Maximum reading annunciator.
3 Bars: 700 – 550 hours 2 Bars: 350 – 550 hours 1 Bar: 150 – 350 hours Blinking: < 150 hours
CAL
Blinks during calibration mode. Remains in display indicating field calibration is active.
HOLD
Remain in display during hold mode.
STABLE
Displayed upon recognizing final value.
- 12 -
10. MEASUREMENT MODE
Press the on/off key. The thermometer performs a self-test and all display digits and indicators should remain on for approximately two seconds before the meter enters measurement mode.
If a probe is not connected or if the probe is defective, the display will indicate “OPEN”.
Figure 6: Display when no probe is detected
For optimum instrument accuracy, allow one minute for ambient temperature stabilization. If the unit has been stored at an extreme ambient condition, more time will be needed.
A STABLE indicator appears at the top right corner of the display when reading is continuously stable for 5 seconds.
To freeze reading, press hold. Press hold key again to release the reading and resume normal operation.
Figure 7: HOLD and STABLE function
The display will show “Ur” (under range) or “Or” (over range) if the temperature reading
is out of range of the instrument.
°c
°c
HOLD STABLE
- 13 -
11. SELECTING TEMPERATURE SCALE
Figure 8: Press °C/°F key to toggle
between °C and °F
Select °C or °F by pressing the °C/°F key. Each time the key is pressed, the temperature scale will switch. Switching between °C and °F can be done at any time during operation.
Each time you turn the instrument on, it will power up with the same settings that were set when the unit was last turned off.
°c
°F
°C/°F
- 14 -
12. MIN AND MAX FUNCTION
Press the min/max key to toggle between the minimum and maximum readings. The minimum and maximum reading function is ideal for monitoring unattended operations while continually displaying every temperature change that occurs. The minimum and maximum values are sensed and automatically stored. To exit and clear the MIN/MAX reading stored, press and hold the min/max key for 3 seconds.
13. AUTO OFF FUNCTION
The thermometer has a default auto off function of 17 minutes.
To deactivate this function, press the min/max key while turning the thermometer on with the on/off key. “A.OFF no” flashes to show indicate that the auto off function has been disabled for that session.
Figure 9: Disabling the auto off function
- 15 -
14. CALIBRATION
The CAL function allows single point calibration of the thermometer at 0°C (32°F) to compensate for probe offset error. Use the field calibration feature to improve thermometer/probe accuracy.
The thermometer comes with a calibration lock function to prevent accidental changes to calibration settings. A field calibration can only be performed when the function has been unlocked. To unlock field calibration, follow steps 1 through 4 in FIELD
CALIBRATION UNLOCK section. To calibrate:
1. Pack sensing end of probe in a
container tightly packed with crushed ice and filled with distilled water. Allow temperature to stabilize.
2. Press and release the cal key for
3 seconds to enter the calibration mode. The CAL indicator starts blinking. Release the cal key.
3. If the measured temperature is from
–5°C to 5°C (23 to 41°F) when the temperature reading is stable, press the cal key.
The CAL indicator will stop blinking, and the reading will be set to 0°C (32°F). The CAL indicator will remain in display, indicating that field calibration is now active.
If “Err” is displayed, the displayed reading is outside the above limits.
NOTE: Press any key (except the cal key) while in calibration mode to abort calibration.
- 16 -
15. CLEAR FIELD CALIBRATION
To clear user calibration:
1. Turn the thermometer off.
2. Hold the cal key down while pressing
the on/off key.
3. The measurement mode window
appears without the CAL indicator to indicate that the user calibration has been successfully cleared.
16. FIELD CALIBRATION UNLOCK
The calibration unlock feature enables field calibration operation. To unlock field calibration:
1. Turn the thermometer off.
2. Hold the cal key and hold key down
simultaneously, then press the on/off key.
3. The “CAL” indicator appears
momentarily with the version-model window to indicate that the user­calibration function has been unlocked successfully (Figure 10).
Figure 10: Successful field calibration unlock
CAL
- 17 -
4. When the cal and hold keys are
released, the unit will go to measurement mode (Figure 11).
Figure 11: Unit goes to measurement mode
17.
FIELD CALIBRATION LOCKOUT
The calibration lockout feature prevents any field calibration changes. The lockout remains in effect until an unlock has been performed. To lock the field calibration operation:
1. Turn the thermometer off.
2. Hold the cal key and °C/°F key down
simultaneously, then press the on/off key. Do not release cal and °C/°F keys until the measurement mode is on display.
3. If field calibration lockout is successful,
the version-model window appears without the CAL indicator (Figure 12).
Figure 12: Successful field calibration lockout
°c
STABLE
- 18 -
4. When the cal and °C/°F keys are
released, the unit will return to measurement mode (Figure 13).
Figure 13: Unit returns to measurement mode
18. MAINTENANCE
Properly used, the thermometer should maintain calibration indefinitely and not require service other than occasional cleaning of the housing and changing of the batteries. Do not clean with abrasives or solvents. Use mild detergents; never immerse nor use excessive fluid.
19. CLEANING
WARNING:
TO PREVENT IGNITION OF A HAZARDOUS ATMOSPHERE BY ELECTROSTATIC DISCHARGE, CLEAN WITH DAMP CLOTH.
Do not clean with abrasives or solvents. Use mild detergents, never immerse nor use excessive fluid.
°c
STABLE
- 19 -
20. TROUBLESHOOTING
Problem
Cause
Solution
Power on but no display
Batteries not in place or wrong polarity.
Insert batteries. or re-insert batteries in correct polarity.
OPEN display on LCD
Probe not connected
Make sure probe is firmly connected.
“Ur” or “Or”
display on LCD
Measurement over (Or) or Under (Ur) range
Ensure temperature taken is within
meter’s
specification.
Unstable reading
1. Probe not deep enough in sample
2. Broken probe
1. Place probe deeper in sample.
2. Replace probe.
Slow response
Dirty probe
Clean probe
- 20 -
21. SERVICE
There are no internal adjustments or user replaceable parts.
Note: Serial number label is located behind the meter.
22. REPLACEMENT METER & ACCESSORIES
Item
Catalog number
Single-input RTD thermometer
91428-07
Rubber armor with stand
35427-80
Hands-free kit (two magnets and a strap)
35427-85
General-purpose probe, 10 in. long x 0.188 in. dia, 10-second time constant.
08117-70
PTFE-coated general­purpose probe,
10 in. long x 0.145 in. dia, 15-second time constant.
08117-87
Penetration probe, 4 in. long x 0.188 in. dia, sharp tip, 10-second time constant.
08117-85
Surface probe, 8 in. long with 0.25 in. dia, aluminum and ceramic tip, 24-second time constant.
08117-75
Air/gas probe, 10 in. long sheath with 0.25 in. dia radiant heat shield. 4-second time constant.
08117-90
- 21 -
23. WARRANTY
The Manufacturer warrants this product to be free from significant deviations from published specifications for a period of three years. If repair or adjustment is necessary within the warranty period, the problem will be corrected at no charge if it is not due to misuse or abuse on your part as determined by the Manufacturer. Repair costs outside the warranty period, or those resulting from product misuse or abuse, may be invoiced to you.
24.
PRODUCT RETURN
To limit charges and delays, contact the seller or Manufacturer for authorization and shipping instructions before returning the product, either within or outside of the warranty period. When returning the product, please state the reason for the return. For your protection, pack the carefully and insure it against possible damage or loss. The Manufacturer will not be responsible for damage resulting from careless or insufficient packing.
- 22 -
25. INNOCAL® CALIBRATION AND REPAIR SERVICES
Optimum performance of your temperature­measuring instrument is not a timeless condition. To ensure quality measurements, have your instrument calibrated regularly. Trust InnoCal to satisfy your calibration and equipment repair needs. With over a decade of service, we've helped thousands of customers meet ISO, FDA, EPA, GLPs/cGMPs and other quality standards.
Conformity*
ISO/IEC 17025:2005 accredited NIST Handbook 150, 2000 Edition ANSI/NCSL Z540-2-1997 NIST Technical Note 1297 ISO 9000:2000
Fast Service
Our substantial inventory of replacement parts ensures a fast turnaround and prevents costly downtime. Most instruments serviced in five business days!
Excellent Value
Get quality at a fair price. Our InnoCal NIST-traceable certificates offer extensive test data on a broad range of measurement parameters without breaking the bank!
Reliable Support
Trust in our free diagnostic support and troubleshooting advice. Our factory-trained metrologists and technicians are armed with years of experience and extensive technical data.
- 23 -
Convenient Reminders
It’s so easy to keep your instruments
functioning properly. Based on your requirements, InnoCal will send you a reminder when it’s time to re-certify or service your instrument.
We provide you with the documentation you need to meet your most stringent quality requirements for the control of inspection, measuring, and test equipment.
Certification includes certificate of calibration with test data, including:
Description and identification of the item certified
Condition of the item
Issue date
Identification of calibration procedure
Calibration date
As found/as left test data (where
applicable)
Signature of technician
Statement of estimated uncertainty
List of equipment used to perform
calibration (including their calibration dates)
With today's high quality standards such as ISO 9000, certification is becoming increasingly important. Traceability is not a timeless condition. It must be verified and maintained over the life of the calibration to ensure the highest accuracy possible. When you have your calibration done by InnoCal, we will send you an automatic reminder when it is time to recalibrate your instrument.
- 24 -
Are your calibration certificates good enough?
InnoCal surpasses the competition by providing the most complete certificates as required by NIST. All of our certificates include measured data and point-by-point measurement uncertainty, and by request, we will provide test accuracy and test uncertainty ratios at no extra cost. Call us today and see why InnoCal is The Choice of Quality.
*See our Scope of Accreditation for any limitations.
Calibration
test points
against NIST-
traceable
standards
Meter
only
Probe
only
System
(meter +
probe)*
Four test
points across
range of
instrument.
17000-04
17001-04
17002-04
InnoCalThe Choice of Quality
866-INNOCAL (866-466-6225)
InnoCalSolutions.com
For calibration services outside of North
America, please contact your Local
Distributors or Local Certification Body.
- 25 -
TECHNICAL ASSISTANCE If you have any questions about the
use of this product, please contact the Manufacturer or authorized seller.
For more information, contact:
Cole-Parmer
Toll-Free: 1-800-323-4340 Phone: 1-847-549-7600 Fax: 1-847-247-2929 ColeParmer.com/Digi-Sense
1065DGMAN_91428-07
Loading...