Digi-Sense 91428-05 Instruction Manual

Single-Input Thermistor
Thermometer
(Model 91428-05)
68X001349 Rev0 0717 1065DGMAN_91428-05
TABLE OF CONTENTS
1. QUICK-START GUIDE ........................... 1
2. INTRODUCTION .................................... 2
3. SAFETY PRECAUTIONS ....................... 3
4. SPECIFICATIONS ................................. 4
5. BATTERY INSTALLATION AND
REPLACEMENT .................................... 5
6. INSERTING AND REMOVING
OPTIONAL RUBBER ARMOR ............... 6
7. ASSEMBLING OPTIONAL HANDS-
FREE ACCESORIES ............................. 7
8. CONNECTING A THERMISTOR............ 8
9. KEY FUNCTIONS .................................. 9
10. DISPLAY OVERVIEW ........................... 10
11. MEASUREMENT MODE .......................11
12. SELECTING TEMPERATURE SCALE ..12
13. MIN AND MAX FUNCTION ...................12
14. AUTO OFF FUNCTION .........................13
15. CALIBRATION ......................................13
16. CLEAR FIELD CALIBRATION ...............14
17. FIELD CALIBRATION UNLOCK ............15
18. FIELD CALIBRATION LOCKOUT .........16
19. MAINTENANCE ....................................17
20. CLEANING ............................................17
21. BATTERIES ..........................................17
22. TROUBLESHOOTING ..........................18
23. ACCESSORIES ....................................19
24. WARRANTY .........................................20
25. PRODUCT RETURN .............................20
26. INNOCAL® CALIBRATION AND REPAIR
SERVICES (NORTH AMERICA) ..........21
- 1 -
1. QUICK-START GUIDE
To ensure best results please read the complete manual.
1. Connect probe.
2. Press the on/off key.
Press again to turn on backlight.
3. Select °C or °F display using °C/°F key.
4. Insert probe in sample.
Allow adequate time for reading to stabilize. Response time will vary depending on probe.
5. Take reading.
Press hold key to freeze display. Press min/max to display minimum or maximum temperature.
6. Press and hold the on/off key for three
seconds to turn meter off.
- 2 -
2. INTRODUCTION
Thank you for choosing the Single-Input Thermistor Thermometer. This sturdy handheld thermometer provides highly accurate temperature measurements using thermistor probes. The instrument is easy to use and includes the following features:
Operator selection of Celsius or Fahrenheit
scale
Resolution of 0.1° C/F
Large backlit LCD with two lines of
four-digit display
Hold feature for temporarily retaining
a reading
Min and max readings display
Field calibration capability
Auto power off with Enable/Disable function
Low-battery warning
Standard ¼ inch (6.3mm) phone jack
connector input
Operates with a wide selection of probes
- 3 -
3. SAFETY PRECAUTIONS
Voltages present at the thermistors may also be present at the battery terminals. Always disconnect the thermistor when changing batteries.
This instrument is designed to accept low level signals supplied by standard 400 series thermistors. Under no circumstances should the input voltage exceed the specified 10 v rms.
1. Do not use or store this instrument in
microwave ovens or any abnormally hot or cold areas.
2. Weak batteries should not be left in the
instrument. Dead batteries can leak and cause damage to unit.
WWARNING
- 4 -
4. SPECIFICATIONS
Range
–40 to 150°C (–40.0 to 302.0°F)
Resolution
0.1°C (0.1°F)
Accuracy
±0.2°C (±0.4°F) from –40 to 125°C (-40 to 257°F)
±0.5°C (±0.9°F) from 125 to 150°C (257 to 302°F)
LCD display
4+4 digit, 58 x 40mm dual-line display with backlight
Out-of-range display
“Ur” or “OPEN”
Display update rate
0.5 sec per update
Input
One probe with ¼ inch (6.3mm) phone jack connector
Input protection
10 v rms
Battery
Three AA, 1.5V alkaline
Battery life
700 hours continuous, typical
Auto shutoff
17 minutes after last key press
Stability criteria
STABLE will display upon 5 seconds of stability
Stated accuracy
18 to 28°C (64 to 82°F)
Useful range
0 to 40°C (32 to 104°F)
Storage
40 to 65°C (40 to 149°F)
Humidity
10% to 90% RH (noncondensing)
Dimensions (L x W x H)
Without armor:
175mm x 97mm x 42mm
With armor:
180mm x 102mm x 52mm
Weight with batteries
Without armor: 267g With armor: 362g
Ingress protection
Meets IEC-529 IP-54 for dust and water-resistant enclosures
Compliance (for CE mark)
EN61326-1/A1: 1998 (EU EMC Directive)
- 5 -
5. BATTERY INSTALLATION AND
REPLACEMENT
Voltages present at the thermistors may also be present at the battery terminals. Always disconnect the thermistor when changing batteries.
Weak batteries should not be left in the instrument. Dead batteries can leak and cause damage to unit.
The typical battery life is about 700 hours. Selected settings
are
stored in
memory and
will remain in memory
even after
power is
turned
off, or
while
batteries
are being
replac
ed.
1. Before
changing
battery, tur
n instrume
nt
off
and disconnect probe
.
2.
Loosen screw and
lift battery
cover
off
the
back
of case.
3.
Remove
the t
hree
AA batteries.
4. Insert three
new
batteries
observing
pol
arity.
5.
Install
cover and
tighten
screw
.
- 6 -
6. INSERTING AND REMOVING
OPTIONAL RUBBER ARMOR
1. To insert thermometer into the optional
rubber armor, slide in from the top of meter before pushing the bottom edges of meter down to set it into position. Lift up the stand at the back of meter for benchtop applications if necessary.
2. To remove thermometer from armor,
push out from the bottom edges of meter until it is completely out of boot.