Digi-Sense 91428-04, 91428-03 Instruction Manual

Dual-Input J/K/T/E Thermocouple Thermometer (Model 91428-03)
Dual-Input Data Logging Thermocouple Thermometer
(Model 91428-04)
68X001352 Rev0 0717 1065DGMAN_91428-03,-04
TABLE OF CONTENTS
1. INTRODUCTION ..................................... 1
2. SAFETY PRECAUTIONS ........................ 2
3. SPECIFICATIONS .................................. 3
4. BATTERY INSTALLATION ...................... 6
5. OPTIONAL RUBBER ARMOR .................. 7
6.
OPTIONAL HANDS-FREE KIT
.................. 8
7. CONNECTING A THERMOCOUPLE ........ 9
8. KEY FUNCTIONS ................................. 11
9. DISPLAY OVERVIEW ........................... 12
10. MEASUREMENT MODE........................ 13
11. HOLD FUNCTON .................................. 14
12. MIN, MAX, AND AVG FUNCTION ............ 14
13. DATA LOGGING ................................... 14
14. SETUP MODE ...................................... 15
15. GENERAL SETUP SCREEN ................. 16
16.
CALIBRATION SCREEN
........................ 18
17.
ALARMS SCREEN
................................ 21
18. DATA LOGGING SCREEN .................... 22
19.
CALIBRATION REPORT SCREEN
......... 23
20. CLEAR / RESET SCREEN..................... 24
21.
MAINTENANCE
.................................... 25
22.
CLEANING
........................................... 25
23.
BATTERIES
.......................................... 25
24. TROUBLESHOOTING ........................... 26
25.
ACCESSORIES
.................................... 27
26.
WARRANTY
......................................... 28
27.
PRODUCT RETURN
............................. 28
28. INNOCAL
®
CALIBRATION AND
REPAIR SERVICES .............................. 29
- 1 -
INTRODUCTION
This versatile handheld instrument provides highly accurate temperature measurements. The instrument is designed for easy operation and includes the following features:
Menu driven setup and operation
Data logging for up to 1000 points
on model 91428-03, 2000 points on model 91428-04
USB output (model 91428-04 only)
Differential temperature
measurements
Operator selection of Celsius or Fahrenheit scale
Resolution of 0.1°C/F from 199.9 to 999.9°
Large backlit LCD with two lines
of four-digit display
Hold feature for temporarily retain a reading
Displays min and max readings
Field calibration capability
Disabling of Auto-Off function
Low-battery warning
Two blade female ANSI mini-
connector input
Operates with a wide selection of probes
- 2 -
SAFETY PRECAUTIONS
WARNING:
1. This instrument is designed to accept low level signals supplied by standard Thermocouples. Under NO circumstances should the input voltage exceed the specified 50V RMS.
2. To prevent ignition of a hazardous atmosphere, batteries must only be changed in an area known to be nonhazardous.
CAUTION:
1. Do not use or store this instrument in microwave ovens or any abnormally hot or cold areas.
2. Weak batteries should not be left in the instrument. Dead batteries can leak and cause damage to unit.
DANGER:
1. Voltages present at the
thermocouple may also be present at the battery terminals. Always disconnect the thermocouple when changing batteries.
- 3 -
SPECIFICATIONS
Thermocouple Thermometers
Type
Temperature range
Type J
–210°C to 1200°C
(–346°F to 2192°F)
Type K
–250°C to 1372°C
(–418°F to 2501°F)
Type T
–250°C to 400°C
(–418°F to 752°F)
Type E
–250°C to 1000°C
(–418°F to1832°)
Type R (model 91428-04 only)
0°C to 1768°C
(32°F to 3214°F)
Type S (model 91428-04 only)
0°C to 1768°C
(32°F to 3214°F)
Type N (model 91428-04 only)
–250°C to 1300°C
(–418°F to 2372°F)
Type B (model 91428-04 only)
200°C to 1800°C
(392°F to 3272°F)
Out-of-range display: - - -
-
Reso
lution:
0.1°/1° auto-ranging,
0.1°C/F from –199.9 to 999.9°,
1° outside this range
- 4 -
Accuracy: Types J, K, T, E, and N:
Below –150°C (–238°F): ±0.25% of reading ±1°C (±0.25% ±0.7°F) Above –150°C (–238°F): ±0.1% of reading ±0.4°C (±1% ±0.7°F)
Types R, S, and B: ±0.1% of reading ±1°C (±0.1% ±2°F)
Display:
Backlit dot matrix, 50mm x 37.2mm
Data logging:
Model 91428-03: 1000 points Model 91428-04: 2000 points
Logging interval:
1 sec to 60 min
Min/Max/Avg function:
Yes
Auto Off (adjustable time):
Enable/disable option available
Stability criteria:
Yes, upon stability of 5 seconds
Display update
rate:
0.6
sec per up
date
Input:
Two thermocouples with ANSI connector
- 5 -
Input protection: 50V
rms
Storage:
40°C to 65°C (40°F to 149°F)
Humidity:
10% to 90% RH (noncondensing)
Battery life: Size: Three AA, 1.5V; alkaline Life: 400 hours continuous, typical,
(without backlighting)
Dimensions: Without armor:
175mm (L) X 97mm (W) X 42mm (H)
With armor:
180mm (L) X 102mm (W) X 52mm (H)
Weight with batteries: Without armor: 267g With armor: 362g
Ingress protection:
Meets IEC-529 IP-54 for dust and water resistant enclosures (probe attached)
CE Compliance:
EN61326-1/A1: 1998 (EU EMC Directive)
- 6 -
BATTERY INSTALLATION
The total battery life without backlighting and alarm is about 400 hours. The battery bar annunciator represents the battery strength. An empty battery annunciator indicates low battery strength; a blinking battery annunciator indicates that the batteries should be replaced immediately.
Selected settings are stored in memory and will remain in memory even after power is turned off, or while batteries are being replaced.
1. Before changing battery, turn instrument off and disconnect thermocouple.
2. Loosen screw and lift battery cover off back of case.
3. Remove the three AA batteries.
4. Insert three new batteries observing polarity.
5.
Install cover and tighten screw.
- 7 -
RUBBER ARMOR
1. To insert thermometer into the optional rubber armor, slide in from the top of meter before pushing the bottom edges of meter down to set it into position. Lift up the stand at the back of meter for bench top applications if necessary.
2. To remove thermometer from armor, push out from the bottom edges of meter until it is completely out of boot.
- 8 -
OPTIONAL HANDS-FREE KIT
You can use the optional magnets and strap in the Hands-Free Kit accessories for convenient hands-free operations.
- 9 -
CONNECTING A THERMOCOUPLE
Use the correct thermocouple type for your instrument setting. Using an incorrect thermocouple type will result in erroneous readings. Thermocouples are colour coded by type using the North American ANSI Color Code as follows:
TYPE COLOR
J Black K Yello
w T Blue E Purple
Thermocouple connectors have one wide blade and one narrow blade. Do not force connector in backwards. Connect thermocouples to receptacles at top of instrument as shown in the following illustration:
Loading...