Digi-Sense TC9600, 89800-11, TC9100, 89800-12, 89800-13 Instruction Manual

...
1
Calibration Manual
Digi-Sense® TC9100 and TC9600
Models: 89800-11, 89800-12, 89800-13 & 89800-14
THE STANDARD IN PRECISION MEASUREMENT
2
Introduction
Each temperature controller has built into the operation menus a calibration gateway. This gateway allows the temperature controller to be calibrated to six temperature points, allowing for consistent and accurate temperature readings. The calibration process allows for both full-range calibration and segmented calibration. Each unit comes preloaded with tested default calibration temperature points.
WARNING: Do not enter the calibration gateway unless you have the proper calibration equipment to calibrate the sensor inputs associated with the controller. Changing calibration settings can cause errors in sensor temperature
readings and operation of the controller.
Full-Range Calibration
This process provides accurate temperature measurements along the entire range of a given type of temperature probe. Ranges of each type of temperature probe are provided on the controllers menu screen, as well as in Table 1, below.
SENSOR TYPE RANGE
THERMOCOUPLES
Calibration accuracy +/-0.1% of span or +/-oC for types J, K, T, E, N;
+/- 0.2% of span or +/-2oC for types B,R,S. Accuracy span is 1000oF (540oC) minimum.
Type J -190 to 1000 °C
Type K -200 to 1372 °C
Type N -200 to 1300 °C
Type R 0 to 1768 °C
Type S 0 to 1768 °C
Type T -200 to 400 °C
Type B 600 to 1800 °C
Type E -200 to 1000 °C
PLATINUM RTD
Calibration accuracy +/-0.1% of span. Accuracy span is 1000oF (540oC) minimum.
0.00385
-200 to 800 °C
-200 to 630 °C
0.003916
THERMISTOR
Calibration accuracy +/-0.4% of span or +/-0.5oC. Accuracy span is 1000oF (540oC) minimum.
YSI 400 -30 to 100 °C
YSI 700 -15 to 100 °C
Table 1. Sensor Input Range