Instruction Notes for 3622590
Figure 1 Figure 2
General Description
Operation is stable in ambient temperature from 30 to 130 degrees F. 0 to 90% RH, non-condensing
Control Features
1. On/Off switch.
2. 15-amp output.
3. Non-Linear control (scale is for reference only)
4. Circuit protection
5. Low profile housing.
How to Setup and Operate
First connect the heater load (sized as shown below) and the thermocouple plug to the load and thermocouple receptacles on the rear of the control. Next plug the control line cord into the appropriate 3-wire grounded power receptacle. Push the power switch to the “ON” position to apply power to the control. Turn the control knob to the desired temperature setting. The “LOAD” pilot light on the front panel indicates where there is power to the load. The control is accurate within + or – 1% of full scale.
3/10 1 Rev.0
Digi-Sense 625 East Bunker Court, MS 18 Vernon Hills, Illinois 60061-1844, U.S.A Phone: 847-549-7600 Fax: 847-549-7676 ( 800) 323-4340
Operator Interface
- Control knob
- Digital setpoint display
- ON/OFF switch
- Output pilot light
- Type J thermocouple receptacle
- 3-wire load receptacle
Accuracy
- Calibration accuracy: + or - 1% of span. +/- 11.0°C
- Set point accuracy: + or - 3% of span maximum.
Control Sensor
This control requires a type “J” an Iron-Constantan thermocouple. The control may be used with any type “J “thermocouple which has a lead resistance less than 100 ohms.
Proper placement of the thermocouple can eliminate many problems in the total system. The probe should be placed so that it can detect any temperature change with little thermal lag. In a process that requires fairly constant heat output, the probe should be close to the heater. In processes where heat demand is variable, the probe should be close to the work area. Some experimenting with probe location can be tried to provide optimum results.
Specifications
Control Mode
- Single input, single output
- Range 0 – 750 degrees C
Input
-Type J thermocouple input; grounded or ungrounded
- Automatic cold junction compensation and break protection for sensor.
Output
-15 amps (1800 watts), 120 Volt (Fused)
Maintenance
Simple preventative maintenance steps include keeping the controller clean. Protect it from overload, excessive dirt, oil and corrosion.
Warranty
Digi-Sense, for itself does hereby offer a warranty for products from the date of receipt by the user, under normal and proper usage, against defects in workmanship and materials for 12 months, and will repair or replace any defective part(s) without charge when same is shipped Prepaid to Digi-Sense from which the product was originally purchased.
Should the nature of any defect require that the product, or any constituent portion thereof, be returned to Digi-Sense, Vernon Hill, Illinois, prepaid for service, a condition precedent to any return shall be the procurement of authorization from Digi-Sense assigning a Return Goods Number to the product or part requiring service.
Parts and accessories manufactured by others are warranted only to the extent of the regular warranty of the manufacturer or supplier of such materials and only insofar as Digi-Sense is able to transfer the benefits of warranty coverage, if any, to the user. Any adequately warranted defective part or accessory manufactured or supplied by others may be exchanged through Digi-Sense for a replacement part is shipped prepaid and received at Digi-Sense within 30 days from the date any replacement part is obtained by the user.
This warranty supersedes and is given in lieu of all implied warranties, and is void if the user causes damages from improper usage of product under normal operating conditions.
12 MONTHS LIMITED WARRANTY ON ALL PARTS AND LABOR IS GIVEN BY DIGI-SENSE. CATALOG NUMBER 3622590 SERIAL NUMBER _________
DATE OF PURCHASE _______________
3/10 2 Rev.0
Loading...