Operation Manual
Instruction Notes for 3622561, 3622562, 3622563, 3622564
Digital Control Module
Load Fuse
ON / OFF Switch Load Outlet Power Inlet
Grid Support
General Description Sensor Input
This temperature control displays both the setpoint and process temperature and uses the most modern control
technology for the best temperature stability. The Auto-tune feature minimizes setpoint overshoot and learns your
process. Changing your setpoint is easily done with the interface keys. The controller can also perform ramp rate
operations to allow the user to slowly raise the process temperature. There are several input types available. This unit
comes with a 6’ detachable power cord, grid support bracket, which is ideal for fume hood mounting to maximize bench
space. This control is a microprocessor-based, digital indicating, automatic temperature control with a single input and a
single output. It features an auto-tuning function that allows automatic setting of control parameters with a minimum of
user input required. This family of control accepts a type "J", “K”, “T” thermocouple, or RTD (model 3622564) input
depending on the model ordered.
The control automatically sets the PID parameters through a "learning" sequence in the auto-tuning mode. PID
parameters include proportional band, reset/integral and rate/derivative. User-friendly features include automatic LED
indicators to aid in monitoring and setup, as well as dual LED displays for process temperature and set point indication.
This control automatically stores all information in a non-volatile memory.
Control Features
1. On/Off switch.
2. 15-amp output.
3. Control PID with Auto-tune algorithm, solid state relay
4. Circuit protection (fuse)
5. Low profile housing.
6. Safety feature- Output is de-energized if sensing device is unplugged or damaged.
09/14 1 Rev.5
How to Setup and Operate
First, connect the heater load and the sensor to the load and sensor receptacles on the front panel. Next, plug the control
line cord into an appropriate 3-wire grounded power receptacle. Push the power switch to the "ON" position. Wait five
seconds for the control unit to energize.
The control is shipped from the factory with the display reading in degrees C. If a display in degrees F is desirable, press
the UP/DOWN arrow keys simultaneously for three seconds to access the Setup Page. Press the Advance key until the
Celsius_Fahrenheit parameter [C-F] is shown in the lower display. Press the UP arrow key to change from C to F.
After changing the display, press the Infinity key to display the set point temperature and the process temperature
(temperature at the sensor).
To begin the heating process, use the Up/Down arrow keys to enter the desired set point temperature. This is shown in
the lower display. Auto-tuning is recommended in all applications. Auto-tuning allows the control to set the PID
parameters without those parameters being manually entered/calculated by the user. It also minimizes the amount of
temperature overshoot of the set point and decreases the time required for the process to stabilize. See page 8 for
instruction on performing the auto-tune function.
It is recommended that the auto-tune feature be used every time the process temperature set point has been changed. If
the process is being repeated, the parameters learned in the initial setup are stored in the memory of the control and the
auto-tune process isn’t required to be repeated in most applications.
09/14 2 Rev.5
Control Sensor
Proper placement of the sensor can eliminate many problems associated in the total operation of the system. The probe
should be placed so that it can detect any temperature change with little thermal lag. In a process that requires fairly
constant heat output, the probe should be close to the heating source. In processes where heat demand is variable, the
probe should be placed close to the work area. Experimenting with probe location should be tried to provide optimum
results for your particular process.
Specifications
Control Mode
- Microprocessor-based, single input, single output.
- PID parameters
Proportional band: 1 to 999o F
Reset: 0.00 to 99.99 repeats per minute.
Integral: 0.00 to 99.99 minutes per repeat.
Rate or Derivative: 0.00 to 9.99 minutes.
Operator Interface
- Advance, Infinity, Up and Down keys, and ON/OFF switch.
- (2) 4 character LED display.
- Thermocouple receptacle and 3 wire load receptacle.
Input
- Type J, K , T thermocouple or (RTD): input grounded or ungrounded.
- Automatic cold junction compensation and break protection for sensor.
- Degrees F or degrees C display; user selectable (preset for degrees C)
Primary Output (Heating or Cooling)
- 15A @ 120VAC or 10A @ 240VAC depending on model
Accuracy
- Calibration accuracy: 0.1% of span.
- Temperature stability: 0.2 oF / oF rise in ambient maximum.
- Voltage stability: 0.01% of span / % of rated line voltage.
Power
- 50/60 Hz 5%
- Data retention upon power failure via nonvolatile memory.
Operating Environment
- 32 to 149 oF / 0 to 65 oC; 0 to 90% RH, non-condensing.
09/14 3 Rev.5