1. This instrument is designed to
accept low level signals supplied by
standard Thermocouples. Under NO
circumstances should the input
voltage exceed the specified 50V
RMS.
2. To prevent ignition of a hazardous
atmosphere, batteries must only be
changed in an area known to be
non-hazardous.
CAUTION:
1. Do not use or store this instrument in
microwave ovens or any abnormally
hot or cold areas.
2. Weak batteries should not be left in
the instrument. Dead batteries can
leak and cause damage to unit.
DANGER:
1. Voltages present at the
Thermocouple may also be present
at the battery terminals. Always
disconnect the Thermocouple when
changing batteries.
The total battery life without
backlighting and alarm is about 400
hours. The battery bar annunciator
represents the battery strength. An
empty battery annunciator indicates low
battery strength; a blinking battery
annunciator
should be replaced immediately
indicates that the batteries
.
Selected settings
and
will remain in memory even
power is turned
are being replac
are
stored in
off, or
ed.
while
memory
after
batteries
1. Before
2.
3.
4. Insert three
5.
changing
instrume
thermocoupl
Loosen screw and
back
Remove
1. To insert thermometer into the
optional rubber armor, slide in from
the top of meter before pushing the
bottom edges of meter down to set it
into position. Lift up the stand at the
back of meter for bench top
applications if necessary.
2. To remove thermometer from armor,
push out from the bottom edges of
meter until it is completely out of boot.
the
your instrume
incorr
erroneous readi
colour coded by type using
American
correct thermocouple type
ect th
ANSI
nt setting.
ermocouple
ngs. T
Colour Code as
Using an
type
hermocouples are
for
will result in
th
e North
follows:
TYPE COLOR
J
Black
K Yello
T Blue
E Purple
Thermocouple connectors have one wide
blade and one narrow blade. Do not force
connector in backwards. Connect
thermocouples to receptacles at top of
instrument as shown in the following
illustration.
Thermocouple wiring polarity must be
correct. If readings decrease as the
temperature increases, the thermocouple
wires may be reversed. The red wire is
negative for thermocouple wires
manufactured in North America.
If no probe is connected the display will
read “open”.
hermocouples are sensitive
T
measuring junction
measurements
to stabilize. Multip
of the
approx
The dot matrix display features a large
primary display, smaller secondary
displays for channel info or min/max/ave,
and helpful annuciators for added
measurement data
1 HOLD - Active
2 Data Logging is Active
3 Alarm Enabled – channel in alarm
indicated: T1 or T2 or T1&T2
4 MAX/MIN/AVG of Secondary
channel if MIN/MAX/AVG activated
5 MAX/MIN/AVG of Primary channel
if MIN/MAX/AVG activated
6 Min/Max hit time since Min/Max
activated. For Avg, it is continually
increments since activated
Press the hold key to retain the reading
on the display. Press the hold key again
for normal operation.
12. MIN, MAX, and AVE FUNCTION
Press the F1 key to toggle between the
minimum, maximum, and average
readings. The minimum and maximum
reading function is ideal for monitoring
unattended operations while continually
displaying every temperature change that
occurs. The minimum and maximum
values are sensed and automatically
stored.
To exit and clear this function, press the
F3 to access the Menu functions.
See the Clear Reset menu section for
more details.
13. DATA LOGGING
Press the
reading to memory. The memory
indicator M = 1234 shows the memory
location for the next stored reading.
Press the
readings.
See section on Data Logging for timed
logging, and logging to a computer (300
model only).
See section on Clear/Reset for
information on clearing stored readings.
The thermometer is factory calibrated and
does not require calibration before use.
The Calibration function allows single
point calibration of the thermometer, at
0°C (32°F) to compensate for
thermocouple off-set error. It is NOT
necessary to perform a field calibration to
obtain specified meter accuracy. Use the
field calibration feature to improve
thermometer/probe accuracy or to
compensate for thermocouple drift.
..
Before go into the calibration mode, must
enter the password. Press F2 to change to
the next digit. (Default Password is 9900)
There are three calibration options:
Offset – adjusts at a single point
Slope – adjusts at two points
Match – adjusts readings on T1 to match
those on T2. Or adjusts T2 to match T1.
Disable or enable the alarm for individual
probe by pressing recall▲ or log▼and F1
to accept. Increase or decrease individual
limit by pressing recall▲ or log▼.
Data Transfer from Meter to Computer
(Temp-300 only)
Press recall▲ or log▼to choose the
logging methods as auto or manual. If it is
auto logging, using recall▲ or log▼to set
time interval. Its range is from 0min to
60min.
Button “Page1” will appear only in Temp-
300.
Please refer to softcopy of the driver
manual in the CD for installation and
datalogging instructions.
***NOTE: METER MUST BE TURNED
ON PRIOR TO CONNECTING USB
CABLE OR COMPUTER MAY NOT
RECOGNIZE THE INSTRUMENT***
Properly us
maintain calibr
require service
cleaning
ed, the th
ermome
ation indefinitely
other than
of the
housing and changing
ter
should
and
occasional
not
of
the batteries.
22.
CLEANING
WARNING:
TO PREVENT IGNITION OF A
HAZARDOUS ATMOSPHERE BY
E
LEC
TROSTATIC DISCHARGE,
CLEAN WITH DAMP CLOTH.
Do
not
clean wi
Use mild
use excessive
BATTERIES
23.
If there is no display when the
thermometer is turned on, check condition
of the three AA batteries. Also check that
the battery terminals are clean and
batteries are properly installed. If
replacement is necessary, refer to the
BATTERY INSTALLATION AND
REPLACEMENT section for replacement
procedure.
The following chart lists the most probable
faults. There are no internal adjustments
or user-replaceable parts.
FAULT ACTION
NO
Display
Display
shows
- - - -
Display
Shows
OPEN
Display
Shows
Err
Can not
connect
to PC
(Temp-
300 only)
Check condition of batteries.
Check that batteries are inserted
properly.
Out of range indication
No thermocouple connected in the
Connector
If display shows this message
other than during the field
calibration mode, please return the
instrument for servicing
Ensure that the meter is powered
on and measuring prior to
connecting to USB port. Before and
after the driver is installed, your PC
may not recognize the Temp-300 if
the Temp-300 is not powered on
prior to connecting the USB to the
computer.
Manufac
be free from significant deviations from
published
three years. If
necessary within
problem will be corrected
if it is not
your
part as deter
Manufacture
warr
anty
prod
uct
invoiced
turer
warra
specifications
repair or adjus
the
warranty period
due
to
misuse or abuse on
mined by
r.
Repair
period, or
misuse or abus
to
you
costs
those
.
nts this product to
for a
period
of
tment
is
at
outside
resulting
e,
may be
, the
no charge
the
the
from
27.
PRODUCT RETURN
To limit
seller or Manu
and shipping in
the
warranty peri
produc
return
carefully and insure
damage or
be responsible for damage resulting from
careless or insufficient packing.
Optimum performance of your
temperature-measuring instrument is not a
timeless condition. To ensure quality
measurements, have your instrument
calibrated regularly. Trust InnoCal® to
satisfy your calibration and equipment
repair needs. With over a decade of
service, we've helped thousands of
customers meet ISO, FDA, EPA,
GLPs/cGMPs and other quality standards.
Our substantial inventory of replacement
parts ensures a fast turnaround and
prevents costly downtime. Most
instruments serviced in five business days!
Excellent Value
Get quality at a fair price. Our InnoCal®
NIST-traceable certificates offer extensive
test data on a broad range of
measurement parameters without breaking
the bank!
Trust in our free diagnostic support and
troubleshooting advice. Our factory-trained
metrologists and technicians are armed
with years of experience and extensive
technical data.
Convenient Reminders
It’s so easy to keep your instruments
functioning properly. Based on your
requirements, InnoCal will send you a
reminder when it’s time to re-certify or
service your instrument.
We provide you with the documentation
you need to meet your most stringent
quality requirements for the control of
inspection, measuring, and test equipment.
Certification includes certificate of
calibration with test data, including:
● description and identification of the item
certified
● condition of the item
● issue date
● identification of calibration procedure
● calibration date
● as found/as left test data (where
applicable)
● signature of technician
● statement of estimated uncertainty
● list of equipment used to perform
calibration (including their calibration
dates)
With today's high quality standards such
as ISO 9000, certification is becoming
increasingly important. Traceability is not a
timeless condition. It must be verified and
maintained over the life of the calibration to
ensure the highest accuracy possible.
When you have your calibration done by
InnoCal, we will send you an automatic
reminder when it is time to recalibrate your
instrument.
Are your calibration certificates good
enough?
InnoCal surpasses the competition by
providing the most complete certificates as
required by NIST. All of our certificates
include measured data and point-by-point
measurement uncertainty, and by request,
we’ll provide test accuracy and test
uncertainty ratios at no extra cost. Call us
today and see why InnoCal is The Choice
of Quality.
*See our Scope of Accreditation for any
limitations.
Calibration against NIST-traceable
standards
Four test points across range of instrument.
Meters: –270 to 2316°C (–454 to 4200°F)
Probes & Systems: –80 to 1000°C (–112 to 1832°F) **Actual range is dependent on type of probe