1. This instrument is designed to
accept low level signals supplied by
standard RTDs. Under NO
circumstances should the input
voltage exceed the specified 50V
RMS.
2. To prevent ignition of a hazardous
atmosphere, batteries must only be
changed in an area known to be
non-hazardous.
CAUTION:
1. Do not use or store this instrument in
microwave ovens or any abnormally
hot or cold areas.
2. Weak batteries should not be left in
the instrument. Dead batteries can
leak and cause damage to unit.
DANGER:
1. Voltages present at the RTD may
also be present at the battery
terminals. Always disconnect the
RTD when changing batteries.
1. To insert thermometer into the
optional rubber armor, slide in from
the top of meter before pushing the
bottom edges of meter down to set it
into position. Lift up the stand at the
back of meter for bench top
applications if necessary.
2. To remove thermometer from armor,
push out from the bottom edges of
meter until it is completely out of boot.
0.003850) for your instrument. Using an
incorrect probe type will result in
erroneous readings. Insert the 3-pin plug
into the mating connector on the top of
the instrument.
If no probe is connected the display will
read “open”.
s are sensitive
RTD
element
allow
Mu
probe by 5 will give you
time
The dot matrix display features a large
primary display, smaller secondary
displays for channel info or min/max/ave,
and helpful annuciators for added
measurement data
(Battery/Main adaptor/USB)
2 Date (format of mmm – dd)
3 Time (hour:min
4 Time format (am/pm/hrs)
5 Measurement mode
6 Hold function indicator
7 Stable indicator
8 Data logging indicator
9 PC data logging indicator
10 Main reading display
11 Current reading unit indicator
12 Min/Max/Avg elapsed time
13 Current Min/Max/Avg reading
14 Meter logging memory location
15 Measurement alarm active
16 Countdown time indicator with
17 Min/Max/Avg Display
18 Min/Max/Avg unit indicator
19 Function keys
Press the hold key to retain the reading
on the display. Press the hold key again
for normal operation.
12. MIN, MAX, and AVE FUNCTION
Press the F1 key to toggle between the
minimum, maximum, and average
readings. The minimum and maximum
reading function is ideal for monitoring
unattended operations while continually
displaying every temperature change that
occurs. The minimum and maximum
values are sensed and automatically
stored.
To exit and clear this function, press the
F3 to access the Menu functions.
See the Clear Reset menu section for
more details.
13. DATA LOGGING
Press the
reading to memory. The memory
indicator M = 1234 shows the memory
location for the next stored reading.
Press the
readings.
See section on Data Logging for timed
logging, and logging to a computer.
See section on Clear/Reset for
information on clearing stored readings.
The thermometer is factory calibrated
and does not require calibration before
use. The Calibration function allows
single point calibration of the
thermometer, at any temperature point
to compensate for RTD off-set error. It
is NOT necessary to perform a field
calibration to obtain specified meter
accuracy. Use the field calibration
feature to improve thermometer/probe
accuracy or to compensate for RTD
drift.
..
Before go into the calibration mode, must
enter the password. Press F2 to change to
the next digit. (Default Password is 9900)
Offset – adjusts at a single point. Offset
calibration can be performed at any
temperature in the offset range of ±10.00
°C or 18.00 °F
Slope – adjusts at two points. The two
calibration points must be at least 40.00 °C
(72.00 °F) apart. The 2nd calibration point
should be at a higher value than the 1st
calibration point.
Use the ▲ or ▼ keys to adjust the value to
match known temperature standard. Press
F1 to accept.
“Alarm Setting page 2” and disable the
count down alarm function.
When activated, the countdown timer
temporarily over-rides the Auto-shutoff
until the countdown is completed or
manually stopped. If the meter is manually
or auto shutoff,
The Countdown Alarm is automatically
reset to “Disable” each time the meter is
shut off. The 'Timer' key will revert back to
the '°C/°F' key.
Press recall▲ or log▼to choose the
logging methods as auto or manual. If it is
auto logging, using recall▲ or log▼to set
time interval. Its range is from 2min to
60min.
Data Transfer from Meter to Computer
Once the USB connection is establish with
PC, press the Select button to download
data from Meter to PC using
HyperTerminal.
Press F1 to choose which data you want
to clear or reset. For calibration, logged
data and reset all, you will have to enter
the password to proceed. (Default
Password is 9900)
Manufac
to be free from significant deviations
from published
period
adjus
tment
warranty period
corrected
due
to
as
deter
Repair
period, or
misuse or abus
you
.
turer
warra
nts this product
of
costs
specifications
three years. If
is necessary within
, the
at
no charge
misuse or abuse on your
mined by
outside
those
resulting
e,
repair or
problem will be
if it is not
the
Manufacture
the
warr
from
may be invoiced
for a
the
anty
prod
part
r.
uct
to
26. PRODUCT RETURN
To limit
seller or Manu
and shipping in
returning
outsi
retur
reason
protectio
it
Manufacturer will not be responsible for
damage resulting from careless or
insufficient packing.
Optimum performance of your
temperature-measuring instrument is not
a timeless condition. To ensure quality
measurements, have your instrument
calibrated regularly. Trust InnoCal® to
satisfy your calibration and equipment
repair needs. With over a decade of
service, we've helped thousands of
customers meet ISO, FDA, EPA,
GLPs/cGMPs and other quality
standards.
Our substantial inventory of replacement
parts ensures a fast turnaround and
prevents costly downtime. Most
instruments serviced in five business
days!
Excellent Value
Get quality at a fair price. Our InnoCal®
NIST-traceable certificates offer
extensive test data on a broad range of
measurement parameters without
breaking the bank!
Trust in our free diagnostic support and
troubleshooting advice. Our factorytrained metrologists and technicians are
armed with years of experience and
extensive technical data.
Convenient Reminders
It’s so easy to keep your instruments
functioning properly. Based on your
requirements, InnoCal will send you a
reminder when it’s time to re-certify or
service your instrument.
We provide you with the documentation
you need to meet your most stringent
quality requirements for the control of
inspection, measuring, and test
equipment.
Certification includes certificate of
calibration with test data, including:
With today's high quality standards such
as ISO 9000, certification is becoming
increasingly important. Traceability is not
a timeless condition. It must be verified
and maintained over the life of the
calibration to ensure the highest
accuracy possible. When you have your
calibration done by InnoCal, we will send
you an automatic reminder when it is
time to recalibrate your instrument.
Are your calibration certificates good
enough?
InnoCal surpasses the competition by
providing the most complete certificates
as required by NIST. All of our
certificates include measured data and
point-by-point measurement uncertainty,
and by request, we’ll provide test
accuracy and test uncertainty ratios at
no extra cost. Call us today and see why
InnoCal is The Choice of Quality.
*See our Scope of Accreditation for any
limitations.