1. This instrument is designed to
accept low level signals supplied by
standard Thermistors. Under NO
circumstances should the input
voltage exceed the specified 50V
RMS.
2. To prevent ignition of a hazardous
atmosphere, batteries must only be
changed in an area known to be
non-hazardous.
CAUTION:
1. Do not use or store this instrument in
microwave ovens or any abnormally
hot or cold areas.
2. Weak batteries should not be left in
the instrument. Dead batteries can
leak and cause damage to unit.
DANGER:
1. Voltages present at the Thermistor
may also be present at the battery
terminals. Always disconnect the
Thermistor when changing batteries.
1. To insert thermometer into the
optional rubber armor, slide in from
the top of meter before pushing the
bottom edges of meter down to set it
into position. Lift up the stand at the
back of meter for bench top
applications if necessary.
2. To remove thermometer from armor,
push out from the bottom edges of
meter until it is completely out of boot.
Use the correct 400 series Thermistor for
your instrument. Using an incorrect probe
type will result in erroneous readings.
Insert the bayonet plug into the mating
connector on the top of the instrument.
If no probe is connected the display will
read “open”.
The dot matrix display features a large
primary display, smaller secondary
displays for channel info or min/max/ave,
and helpful annuciators for added
measurement data
(Battery/Main adaptor/USB)
2 Date (format of mmm – dd)
3 Time (hour:min
4 Time format (am/pm/hrs)
5 Measurement mode
6 Hold function indicator
7 Stable indicator
8 Data logging indicator
9 PC data logging indicator
10 Main reading display
11 Current reading unit indicator
12 Min/Max/Avg elapsed time
13 Current Min/Max/Avg reading
14 Meter logging memory location
15 Measurement alarm active
16 Countdown time indicator with
17 Min/Max/Avg Display
18 Min/Max/Avg unit indicator
19 Function keys
Press the hold key to retain the reading
on the display. Press the hold key again
for normal operation.
12. MIN, MAX, and AVE FUNCTION
Press the F1 key to toggle between the
minimum, maximum, and average
readings. The minimum and maximum
reading function is ideal for monitoring
unattended operations while continually
displaying every temperature change that
occurs. The minimum and maximum
values are sensed and automatically
stored.
To exit and clear this function, press the
F3 to access the Menu functions.
See the Clear Reset menu section for
more details.
13. DATA LOGGING
Press the
reading to memory. The memory
indicator M = 1234 shows the memory
location for the next stored reading.
Press the
readings.
See section on Data Logging for timed
logging, and logging to a computer.
See section on Clear/Reset for
information on clearing stored readings.
The thermometer is factory calibrated and
does not require calibration before use.
The Calibration function allows single point
calibration of the thermometer, at any
temperature point to compensate for
Thermistor off-set error. It is NOT
necessary to perform a field calibration to
obtain specified meter accuracy. Use the
field calibration feature to improve
thermometer/probe accuracy or to
compensate for Thermistor drift...
Before go into the calibration mode, must
enter the password. Press F2 to change to
the next digit. (Default Password is 9900)
Offset
calibration can be performed at any
temperature in the offset range of ±5.00
°C or 9.00 °F
Slope – adjusts at two points. The two
calibration points must be at least 20.00
°C (36.00 °F) apart. The 2nd calibration
point should be at a higher value than the
1st calibration point.
Use the ▲ or ▼ keys to adjust the value to
match known temperature standard. Press
F1 to accept.
Press F2 key in the middle of a countdown
will stop the process:
You will need to restart the countdown by
selecting the “start” key (F2).
Alternatively, you can also choose “Menu”
to go into the timer menu and disabling
the alarm.
The buzzer will sound for 30 seconds, or
until the reset (F2) is pressed, at the end
of the countdown.
Flashing
announciator
indicates that
the countdown
has expired
To reset the timer, select “restart” (F2) it.
You display should look like this:
To repeat the functions, select F2
Note:
The °C/ °F function is disabled when the
count down function is enabled. To
activate the oC/ oF function on
measurement screen, please go into the
“Alarm Setting page 2” and disable the
count down alarm function.
When activated, the countdown timer
temporarily over-rides the Auto-shutoff
until the countdown is completed or
manually stopped. If the meter is manually
or auto shutoff,
The Countdown Alarm is automatically
reset to “Disable” each time the meter is
shut off. The 'Timer' key will revert back to
the '°C/°F' key.
Press recall▲ or log▼to choose the
logging methods as auto or manual. If it is
auto logging, using recall▲ or log▼to set
time interval. Its range is from 0min to
60min.
Press F1 to choose which data you want
to clear or reset. For calibration, logged
data and reset all, you will have to enter
the password to proceed. (Default
Password is 9900)
Properly used, the thermometer should
maintain calibration indefinitely and not
require service other than occasional
cleaning of the housing and changing of
the batteries.
22. CLEANING
WARNING:
TO PREVENT IGNITION OF A
HAZARDOUS ATMOSPHERE BY
E
LEC
TROSTATIC DISCHARGE,
CLEAN WITH DAMP CLOTH.
Do not clean with abrasives or solvents.
Use mild detergents, never immerse nor
use excessive fluid.
23. BATTERIES
If there is no display when the
thermometer is turned on, check condition
of the three AA batteries. Also check that
the battery terminals are clean and
batteries are properly installed. If
replacement is necessary, refer to the
BATTERY INSTALLATION AND
REPLACEMENT section for replacement
procedure.
The following chart lists the most probable
faults. There are no internal adjustments
or user-replaceable parts.
FAULT ACTION
Check condition of batteries.
NO
Display
Check that batteries are inserted
properly.
Display
Shows
OPEN
Display
Shows
Err
Cannot
connect
to PC
No Thermistor connected in the
Connector
If display shows this message
other than during the field
calibration mode, please return the
instrument for servicing
Ensure that the meter is powered
on and measuring prior to
connecting to USB port. Before and
after the driver is installed, your PC
may not recognize the Temp-340 if
the Temp-340 is not powered on
prior to connecting the USB to the
computer.
The Manufacturer warrants this product
to be free from significant deviations
from published specifications for a
period of three years. If repair or
adjustment is necessary within the
warranty period, the problem will be
corrected at no charge if it is not
due to misuse or abuse on your part
as determined by the Manufacturer.
Repair costs outside the warranty
period, or those resulting from product
misuse or abuse, may be invoiced to
you.
27. PRODUCT RETURN
To limit charges and delays, contact the
seller or Manufacturer for authorization
and shipping instructions before
returning the product, either within or
outside of the warranty period. When
returning the product, please state the
reason for the return. For your
protection, pack the carefully and insure
it against possible damage or loss. The
Manufacturer will not be responsible for
damage resulting from careless or
insufficient packing.
Optimum performance of your
temperature-measuring instrument is
not a timeless condition. To ensure
quality measurements, have your
instrument calibrated regularly. Trust
InnoCal® to satisfy your calibration and
equipment repair needs. With over a
decade of service, we've helped
thousands of customers meet ISO, FDA,
EPA, GLPs/cGMPs and other quality
standards.
Our substantial inventory of replacement
parts ensures a fast turnaround and
prevents costly downtime. Most
instruments serviced in five business
days!
Excellent Value
Get quality at a fair price. Our InnoCal®
NIST-traceable certificates offer
extensive test data on a broad range of
measurement parameters without
breaking the bank!
Trust in our free diagnostic support and
troubleshooting advice. Our factorytrained metrologists and technicians are
armed with years of experience and
extensive technical data.
Convenient Reminders
It’s so easy to keep your instruments
functioning properly. Based on your
requirements, InnoCal will send you a
reminder when it’s time to re-certify or
service your instrument.
We provide you with the documentation
you need to meet your most stringent
quality requirements for the control of
inspection, measuring, and test
equipment.
Certification includes certificate of
calibration with test data, including:
● description and identification of the
item certified
as ISO 9000, certification is becoming
increasingly important. Traceability is
not a timeless condition. It must be
verified and maintained over the life of
the calibration to ensure the highest
accuracy possible. When you have your
calibration done by InnoCal, we will
send you an automatic reminder when it
is time to recalibrate your instrument.
Are your calibration certificates good
enough?
InnoCal surpasses the competition by
providing the most complete certificates
as required by NIST. All of our
certificates include measured data and
point-by-point measurement uncertainty,
and by request, we’ll provide test
accuracy and test uncertainty ratios at
no extra cost. Call us today and see why
InnoCal is The Choice of Quality.
*See our Scope of Accreditation for any
limitations.