Digi-Sense 20250-60 Instruction Manual

THE STANDARD IN PRECISION MEASUREMENT
User Manual
Insulation Tester
Model 20250-60
1065DGMAN_20250-60 DS Insulation tester manual.indd 1 8/7/2017 2:09:49 PM
The Digi-Sense Insulation Tester (Model 20250-60) is versatile and easy to use. The instrument is a must-have on your electrical tool belt. It safely measures insulation resistance of electrical devices such as cables or motor coils up to 4000 M, as well as providing a means of measuring AC/DC voltages and circuit continuity. Careful use of this meter will provide years of reliable service.
Safety Precautions
Read the following safety information carefully before
attempting to operate or service the meter.
To avoid damages to the instrument do not apply the
signals which exceed the maximum limits shown in the technical specifications tables.
Do not use the meter or test leads if they look damaged.
Use extreme caution when working around bare conductors or bus bars.
Accidental contact with the conductor could result in
electric shock.
Use the meter only as specified in this manual; otherwise,
the protection provided by the meter may be impaired.
Read the operating instructions before use and follow
all safety Information.
Caution when working with voltages above 60 VDC or
30 VAC RMS. Such voltages pose a shock hazard.
Before taking resistance measurements or testing
acoustic continuity, disconnect circuit from main power supply and all loads from the circuit.
1065DGMAN_20250-60 DS Insulation tester manual.indd 2 8/7/2017 2:09:49 PM
2
Safety Symbols
Caution refer to this manual before using the meter.
Dangerous voltages.
Meter is protected throughout by double insulation or reinforced insulation.
CE Compliance with EN-61010-1 Note: When servicing, use only specified replacement
parts.
Unpacking
Check individual parts against the list of items below. If anything is missing or damaged, please contact your instrument supplier immediately.
1. Instrument
2. Test probes
3. Carrying case
4. Six AA batteries
5. User manual
1065DGMAN_20250-60 DS Insulation tester manual.indd 3 8/7/2017 2:09:49 PM
3
Meter Description
750V
1000V
Insulation Tester
400
125V
250V
500V
1000V
TEST
LOBAT
HV
(1)
POL1
(2)
(1)
OHM
BZ
(2)
(2)
POL2
4
(1)
BAR
1
8
5
1
2
3 4
6
7
8
9
10
CEM
DT-5505
5
1. Digital display
2. Data Hold button; Max./Min.
3. Lock button
4. Backlight button; Zero
5. Test button
6. Rotary function switch
7. V Jack
8. COM input jack
9. Pothook
10. Battery cover / flip stand
How It Works
How to Connect Test Leads
When the rotary function switch is on M range, 400 /BZ, ACV, or DCV, connect the red test lead into the V terminal and the black lead into the COM terminal.
Test Leads Check
Set the range select switch to the 400 range. With the tip and alligator clip of the test leads connected, the indicator should read 00.0 . When the leads are not connected, the display will read infinity indicated by “OL“. This will ensure that test leads are in good working condition.
4
1065DGMAN_20250-60 DS Insulation tester manual.indd 4 8/7/2017 2:09:49 PM
Rotary Function Switch Positions
Turn the tester on by selecting any measurement setting.
HOLD Max/Min Button
Instant-pressing the HOLD button the first time will hold the values at that instant in the primary display. Pressing it a second time will return it to current reading. Pressing and holding the button for 2 seconds will allow reading the MAX value, and an additional instant-pressing will switch it to display the MIN measured value. These can be toggled by instant-pressing the button. To go back to reading current values, press and hold the button again for 2 seconds.
Lock Button Used when testing insulation resistance. Press the
LOCK button and then push down the TEST button.
This will apply the high-voltage source and display the test result. Pressing the TEST button again will shut off the high-voltage and exit from the insulation resistance testing.
Test Button
Used when testing insulation resistance. Pressing the TEST button will apply the high-voltage source and display the test result. Pressing the TEST button again will shut off the high-voltage and exit from the insulation resistance testing.
1065DGMAN_20250-60 DS Insulation tester manual.indd 5 8/7/2017 2:09:49 PM