Digi-Sense 20250-22 Instruction Manual

User Manual
Data Logging Anemometer
with NIST-Traceable Calibration
Model 20250-22
THE STANDARD IN PRECISION MEASUREMENT
Introduction
1.1 to 20.0 m/s and records up to 32,000 readings. Sam­pling rate, high/low alarm levels, and start mode can be defined by the user with the Windows
®
98, 2000, XP, Vista and Win 7 compatible software provided. Logged data is downloaded to a computer via the integrated USB interface for analysis, graphing, and print out. The instrument is fully tested and calibrated to NIST-traceable standards. Careful use of this meter will provide years of reliable service.
Unpacking
Check individual parts against the list of items below. If anything is missing or damaged, please contact your instrument supplier immediately.
1. Data logging anemometer
2. Cover
3. Fan sensor
4. Metal mounting stand
5. Software CD
6. Battery
7. User manual
8. NIST-traceable calibration report with data
2
Meter Description
1. Cover
2. USB interface
3. Start button
4. Indicator lights (record and alarm)
5. Fan sensor
6. Metal stand
6
3
Key Features
• Memory for 32,000 readings
• Analyze and graph data with included software
• Selectable sampling interval
• Operating status indicated by red, yellow, or
green LED light
• User-defined high/low alarm set points
• Two modes (manual and automatic) to start data logging
• Low-battery alarm
• Mounting stand for in-place monitoring
Setup and Operation
Install the supplied data logging software as follows:
1. Insert the CD into the PC.
2. Double-click the set up.exe file to launch the software
installation program
3. Follow the installation wizard to complete the
installation.
4. Upon successful installation of the software the
“AnemometerDL” software icon shortcut will be automatically placed on your PC desktop.
To initiate the software, double click on the “AnemometerDL” software icon. An easy-to-follow user interface will display. A complete software user guide can be produced by selecting the “Help” option followed by “contents” from the user interface.
4
LED Status Guide
LEDs Indication
No LED light flash
REC ALM
One green flash every 10 seconds (20 s or 30 s)
REC ALM
One red flash every 10 seconds (20 s or 30 s)
REC ALM
Yellow double flash every 10 seconds
REC ALM
Note: 10s (20s or 30s) can be set up by PC software. When replacing the battery, you need to disconnect the sensor fan first to allow access to the battery compartment.
– logging not active – wrong installment of battery
– logging, no alarm
Green double flash every 10 seconds (20 s or 30 s) when in manual mode standby
– to start logging (press yellow button until 3 green
flashes and 1 red flash, it is now logging)
– logging, low alarm for measurement value
Red double flash every 10 seconds (20 s or 30 s)
– logging, high alarm for measurement value
– low-battery alarm
Yellow single flash every 1 second (5s,10s or 15s)
– logging finished
5
Specifications
Air velocity
M/S (meters per second)
KPH (kilometers per hour)
FPM (feet per minute) 80 to 3937 ft/min 1 ft/min ±(3% + 40 ft/m)
MPH (miles per hour) 0.9 to 44.8 mph 0.1 mph ±(3% + 0.4 MPH)
KNT (nautical MPH) 0.8 to 38.8 knots 0.1 knots ±(3% + 0.4 knot)
Memory: 32,000 readings
Selectable sampling interval: 2 s, 5 s, 10 s, 30 s, 1 m, 5 m, 10 m, 30 m, 1 hr, 2 hr, 3 hr, 6 hr, 12 hr, 24 hr
Operating temperature: 32 to 122°F (0 to 50°C)
Operating humidity: <80% RH
Weight: 1.8 oz (49 g)
Dimensions: 5
3
4" x 13⁄8" x 11⁄8" (14.5 x 3.5 x 3 cm)
Range Resolution Accuracy
1.1 to 20.00 m/s 0.01 m/s ±(3% + 0.20 m/s)
0.8 to 72.0 km/h 0.1 km/h ±(3% + 1.0 km/hr)
6
7
Maintenance, Recalibration, and Repair
It is recommended that Digi-Sense products are calibrated annually to ensure proper function and accurate measure-
ments; however, your quality system or regulatory body may require more frequent calibrations. To schedule your recalibration, please contact InnoCal, an ISO 17025 calibra­tion laboratory accredited by A2LA.
Phone: 1-866-INNOCAL (1-866-466-6225) Fax: 1-847-327-2993 E-mail: sales@innocalsolutions.com Web: InnoCalSolutions.com