Digi-Sense 91100-00, 91100-10, -1100-20 Instructions Manual

Thermocouple Thermometer Models
91100-00 (Type J)
91100-10 (Type K)
91100-20 (Type T)
Cole-Parmer Instrument Co.
625 East Bunker Court
V ernon Hills, Illinois U.S .A. 60061-1844
(847) 247-2929 (F ax)
800-323-4340
www.coleparmer.com
e-mail: techinfo@coleparmer.com
A1299-0628
Edition 06
®
ACCESSORY THERMOCOUPLE PROBES
Cole-Parmer offers a wide v ariety of thermocouple probes, wire, connectors, e xtension cables, s witch box es, accessories, and calibration services.
The popular probes listed below have a nylon handle, a 5-ft coiled cord and a 316 SS sheath.
General-purpose probes, 5 in long x 0.156 in dia with 0.093 in dia tip. 3 second time constant.
T ype J 08517-55 T ype K 08516-55 Type T 08500-55
Penetration probes, 4 in long x 0.156 in dia sharp
tip. 5 second time constant.
T ype J 08517-65 T ype K 08516-65 Type T 08500-65
Surface probes, 10 in long with 5/8 in dia alumi-
num and ceramic tip . 6 second time constant.
T ype J 08517-60 T ype K 08516-60 Type T 08500-60
Air/gas probes, 8 in long sheath with 1/4 in dia
radiant heat shield. 45 second time constant.
T ype J 08517-75 T ype K 08516-75 Type T 08500-75
39800-98 Carrying case. Hard plastic case with
foam insert holds one meter and up to five probes. Measures 21 in W x 12 in H x 3 in D.
08520-05 Soft carrying case. Water-resistant cloth. Extra room for batteries or small probe . Measures
3.5 in W x 10 in H x 2 in D.
1
INTRODUCTION
This versatile hand-held instrument provides highly accurate temperature measurements. The instru­ment is designed for easy operation and includes the following f eatures:
• Operator selection of Celsius or F ahrenheit scale
• Resolution of 0.1° from –99.9 to 299.9°
• Four-digit LCD
• Hold feature f or temporarily retaining a reading
• Field calibration capability
• Low battery warning
• Built-in tilt stand for easy hands-free oper ation
• Two blade female ANSI mini-connector input
• Operates with a wide selection of probes
2
SAFETY PRECAUTIONS
THIS INSTRUMENT IS
DESIGNED TO ACCEPT LOW LEVEL SIGNALS SUPPLIED BY STAN­DARD THERMOCOUPLES. UNDER NO CIR­CUMSTANCES SHOULD THE INPUT V OLTAGE EXCEED THE SPECIFIED 50V RMS.
DO NOT USE OR STORE
THIS INSTRUMENT IN MICROWAVE OVENS OR ANY ABNORMALLY HOT OR COLD AREAS .
WEAK BA TTERIES SHOULD
NOT BE LEFT IN THE IN­STRUMENT. DEAD BA TTERIES CAN LEAK AND CAUSE D AMA GE T O UNIT.
VOLTAGES PRESENT AT
THE THERMOCOUPLE MA Y ALSO BE PRESENT AT THE BATTERY TERMI­NALS. ALWAYS DISCONNECT THE THERMO­COUPLE WHEN CHANGING BATTERIES.
TO PREVENT IGNITION OF
A HAZARDOUS ATMOS­PHERE, BA TTERIES MUST ONLY BE CHANGED IN AN AREA KNOWN T O BE NON-HAZARDOUS.
AFIN DE PREVENIR
L'INFLAMMATION D'ATMOSPHERES DANGEREUSES, NE CHANGER LES BATTERIES QUE DANS DES EMPLACEMENTS DESIGNES NON DAN­GEREUX.
TO PREVENT IGNITION OF
A HAZARDOUS ATMOS­PHERE BY ELECTROSTATIC DISCHARGE, CLEAN WITH DAMP CLOTH.
WARNING
WARNING
CAUTION
CAUTION
WARNING
D ANGER
AVERTISSEMENT
3
SPECIFICATIONS
THERMOCOUPLE PROBES
T ype T emperature Range
Type J: –200°C to 1000°C (–328°F to 1832°F ) Type K: –250°C to 1372°C (–418°F to 2501°F ) T ype T: –250°C to 400°C (–418°F to 752°F )
Out of range display: - - - ­Resolution: 0.1°/1° autoranging, 0.1° from
99.9° to 299.9°, 1° outside this range.
Accuracy
> –99.9°:
+(0.2% of reading +0.5°C) +(0.2% of reading +0.9°F)
< –99.9°:
+(0.25% of reading, +1°C) +(0.25% of reading +2°F)
Display: 4-digit LCD with 0.5 in high numerals. Display update rate: 0.6 sec per update. Input: One thermocouple with ANSI connector. Input Protection: 50V rms Battery: Two AA, 1.5V alkaline ANSI-L40,
IEC-LR6.
Battery Life: 750 hours continuous, typical. Low battery indication
Battery Symbol on when 8 to 10 hours of battery life remains. Battery symbol blinking: replace battery
Operating Conditions
Stated accuracy: 18°C to 28°C (64°F to 82°F) Useful range:C to 40°C (32°F to 104°F) Storage: –40°C to 65°C (–40°F to 149°F) Humidity (non-condensing): 10% to 90%
4
Dimensions
3 cm D x 8.4 cm W x 15.8 cm H (1.2 in x 3.3 in x 6.2 in)
Weight with batteries: 227 grams (8 ounces) Ingress protection: Meets IEC-529 IP-54 for dust
and water-resistant enclosures.
Intrinsic safety
This product is energy limited for intrinsically saf e operation in hydrogen atmospheres per Class I, Division 1, Groups A, B, C and D hazardous (classified) locations for UL per UL913 and CSA per C22.2 No. 0-M1982 and No. 157-M1987. Maximum surf ace temperature: 135°C (T4); UL file No. E182612 (1997).
Compliance: (For CE Mark)
EN61326-1/A1: 1998 (EU EMC Directive)
5
BATTERY INSTALLATION AND
REPLACEMENT
WEAK BA TTERIES SHOULD
NOT BE LEFT IN THE INSTRUMENT. DEAD BATTERIES CAN LEAK AND CA USE D AMA GE T O UNIT.
VOLTAGES PRESENT AT
THE THERMOCOUPLE MA Y ALSO BE PRESENT AT THE BATTERY TERMI­NALS. ALWAYS DISCONNECT THE THERMO­COUPLE WHEN CHANGING BATTERIES.
TO PREVENT IGNITION OF
A HAZARDOUS ATMOS­PHERE, BA TTERIES MUST ONLY BE CHANGED IN AN AREA KNOWN T O BE NON-HAZARDOUS.
AFIN DE PREVENIR
L'INFLAMMATION D'ATMOSPHERES DANGEREUSES, NE CHANGER LES BATTERIES QUE DANS DES EMPLACEMENTS DESIGNES NON DAN­GEREUX.
CAUTION
D ANGER
WARNING
AVERTISSEMENT
6
If battery indicator turns on, battery life is appro xi­mately 8 to 10 hours. The battery indicator will start blinking with less than 1 hour of life remaining.
AT THIS POINT , BATTER Y MUST BE CHANGED . IF BATTERY VOLTAGE GOES TOO LOW, DIS­PLAY WILL GO BLANK.
See SPECIFICATIONS for battery type.
1. Before changing battery , turn instrument off and disconnect thermocouple.
2. Loosen screw and lift battery cover off back of case.
3. Remov e the two AA batteries.
4. Insert two new batteries observing polarity.
5. Install cov er and tighten screw .
7
CONNECTING A THERMOCOUPLE
Use the correct thermocouple type for your instru­ment. Using an incorrect thermocouple type will result in erroneous readings. Thermocouples are color coded by type using the North American ANSI Color Code as follows:
TYPE COLOR
JBlack K Yellow T Blue
Thermocouple connectors have one wide blade and one narrow blade .
Do not force connector in bac k-
wards.
Connect thermocouple to receptacle at top
of instrument as shown in the follo wing illustration.
Thermocouple wiring polarity must be correct. If readings decrease as the temperature increases, the thermocouple wires may be rev ersed. The red wire is negative f or thermocouple wires manufac­tured in North America.
8
9
10
2. If a thermocouple is not connected or if the ther­mocouple is defective, the display will indicate OPEn.
3. Select °C or °F by pressing the °C/°F key. Tem- perature scale can be switched at an y time. The thermometer remembers the last selected scale the next time y ou turn the power on.
4. For optimum instrument accuracy, allow one minute for ambient temper ature stabilization. If the unit has been stored at an extreme ambient condition, more time will be needed.
Thermocouples are sensitive at the tip or measur­ing junction. When taking temperature measure­ments, allow time f or the reading to stabilize. Mul­tiplying the time constant of the probe by 5 will give you the appro ximate time required.
Note that the readings will automatically change to 1° resolution below –99.9° or abov e 299.9°.
If desired, press the HOLD key to retain the read­ing on the display. Press HOLD ke y again for nor­mal operation.
If the display sho ws "- - - -" the temperature read­ing is out of range of the instrument.
OPERATION
Note: Be sure the thermocouple you are using
matches the instrument setting indicated on the unit.
1. Press the ON/OFF key. The thermometer per­forms a self-test and all display digits and indi­cators, as shown below, should remain on for approximately two seconds .
11
FIELD CALIBRATION
The thermometer is factory calibrated and does not require calibration before use. The CAL function allows single point calibration of the thermometer , at 0°C (32°F) to compensate for thermocouple off­set error . It is not necessary to perform a field cali­bration to obtain specified meter accuracy . Use the field calibration feature to improve thermometer/ probe accuracy or to compensate for thermocouple drift.
1. Follow steps 1 through 4 in OPERATION sec- tion.
2. Pac k sensing end of probe in a container tightly packed with crushed ice and filled with distilled water . Allow temperature to stabilize.
3. Press and hold the CAL key for 5 seconds to enter the calibration mode, the CAL annunciator on the display starts blinking. Release the CAL key.
4. If the measured temperature is from −10°C to 10°C (14 to 50°F), when the temperature read­ing is stable, press the CAL key.
The CAL indicator will stop blinking and the read- ing will be set to 0°C (32°F). The CAL indicator will remain turned on, indicating a field calibra­tion is active.
If “Err” is display ed, either the displayed reading is outside the above limits or the batteries are weak.
CANCELING A FIELD CALIBRA TION
1. Turn the thermometer off .
2. Hold the CAL key down while pressing the ON/OFF key.
The field calibration is canceled and the thermom­eter rev erts to the default f actory calibration. The CAL annunciator is now turned off.
12
FIELD CALIBRA TION LOCKOUT AND RE-ENABLE
The calibration lockout f eature, pre v ents any field calibration changes. The lockout remains in eff ect until a lockout re-enab le has been performed. Use the following procedures to loc kout or re-enable the field calibration operation.
Lockout Procedure
1. Turn the thermometer off .
2. Simultaneously press and hold the CAL and the
°C/°F keys do wn and momentarily press the ON/ OFF key. Continue to hold the CAL and °C/°F
keys f or at least 5 seconds.
Re-Enable Procedure
1. Turn the thermometer off .
2. Simultaneously press and hold the HOLD and the CAL keys do wn and momentarily press the
ON/OFF key. Continue to hold the HOLD and CAL keys until the displa y blanks.
13
MAINTENANCE
Properly used, the thermometer should maintain calibration indefinitely and not require service other than occasional cleaning of the housing and chang­ing of the batteries.
CLEANING
TO PREVENT IGNITION OF
A HAZARDOUS ATMO­SPHERE BY ELECTROSTATIC DISCHARGE, CLEAN WITH DAMP CLOTH.
Do not clean with abrasives or solv ents. Use mild detergents, nev er immerse nor use excessive fluid.
BATTERIES
If there is no display when the thermometer is turned on, check condition of the two AA batteries. Also check that the battery terminals are clean and batteries are properly installed. If replacement is necessary , refer to the BATTERY INSTALLATION AND REPLACEMENT section for replacement procedure.
WARNING
14
SERVICE
SUBSTITUTION OF COM­PONENTS MAY IMPAIR
INTRINSIC SAFETY.
LA SUBSTITUTION DE
COMPOSANTS PEUT COM-
PROMETTRE LA SECURITE INTRINSEQUE.
There are no internal adjustments or user replaceable parts.
If Err followed by the numbers 1 through 9 is displayed (see example below) return unit for service. Note that “Err” alone ma y be displa yed during improper field calibration.
A-1925-32
Rev. F
Printed in U.S.A.
060900
Note: Serial number label is located inside battery compartment.
WARNING
AVERTISSEMENT
15
EU DECLARATION
OF CONFORMITY
Name of Apparatus:
DIGI-SENSE
®
Thermocouple Thermometer
Model Numbers: 91100-00, -10 and -20 Description of Apparatus:Electronic Thermometer
using a Thermocouple Probe 91100-00 (Type J),
-10 (Type K) and -20 (Type T)
Barnant Company declares that the above models are in conformity to the following harmonized standard and directives:
Applicable Applicable Manufacturer’s Directives Specifications Report Number
89/336/EEC EN61326-1/A1: TR9648 92/31/EEC 1998 93/68/EEC
Manufacturer: Barnant Company Division
Cole-Parmer Instrument Company 28W092 Commercial Av enue Barrington, IL 60010 USA T el.: 847-381-7050
Manufacturer’ s Signature:
James W . Doll
Vice President, Engineering
5 October, 2000
Date
WARRANTY
The Manufacturer w arrants this product to be free from significant deviations from pub lished specifi­cations. If repair or adjustment is necessary within the warranty period, the problem will be corrected at no charge if it is not due to misuse or abuse on your part as determined by the Man ufacturer . Re­pair costs outside the warranty period, or those resulting from product misuse or abuse, may be invoiced to y ou.
The warranty period for this product is noted on the W arranty Card.
PRODUCT RETURN
To limit charges and delays, contact the seller or Manufacturer f or authorization and shipping instruc­tions before returning the product, either within or outside of the warranty period. When returning the product, please state the reason for the return. For your protection, pack the product carefully and in­sure it against possible damage or loss. Any dam­ages resulting from improper packaging are your responsibility.
TECHNICAL ASSISTANCE
If you hav e any questions about the use of this prod­uct, contact the Manufacturer or authorized seller .
Trademarks bearing the ® symbol in this publication are registered in the U.S. and in other countries.
A-3477-51
Rev. F
Printed in U.S.A.
060900
Loading...