digimess DS25 User Manual

Page 1
DS25 Colour Digital Storage Oscilloscope User Manual
1
digimess
®
DS25
Portable Colour Digital Storage Oscilloscope
User's Manual
Page 2
DS25 Colour Digital Storage Oscilloscope User Manual
2
www.digimessinstruments.co.uk
Page 3
DS25 Colour Digital Storage Oscilloscope User Manual
3
Copyright 2006
Änderungen und Irrtümer vorbehalten.
Nachdruck, auch auszugsweise, nur mit schriftlicher Genehmigung des Herstellers.
Alle Rechte vorbehalten.
Subject to alterations, errors excepted.
Reprints, also in extracts, are only allowed with written permission of the manufacturer.
All rights reserved.
Page 4
DS25 Colour Digital Storage Oscilloscope User Manual
4
WARRANTY
The perfect working order of the oscilloscope is guaranteed for 12 months as from delivery. There is no warranty for faults arising from improper operation or from changes made to the oscilloscope or from inappropriate application. If a fault occurs please contact or send your oscilloscope to:
The oscilloscope should be sent in appropriate packing - if possible in the original packing. Please enclose a detailed fault report (functions working incorrectly, deviating specifications and so on) including unit type and serial number. Would you also kindly verify warranty cases by enclosing your supply delivery note. Any repairs carried out without reference to a valid warranty will initially be at the owner’s expense. Should the warranty have expired, we will, of course, be glad to repair your oscilloscope as per our General Terms Of Assembly And Service.
Agents details :
Page 5
DS25 Colour Digital Storage Oscilloscope User Manual
5
Table of Contents
General Safety Requirements ______________________________________________ 7 Safety Terms and Symbols_________________________________________________ 8 General Characteristics of the DS25 Colour Digital Storage Oscilloscope __________ 9 Junior User’s' Guidebook________________________________________________ 10
Introduction to the Front Panel and the User's Interface of the DS series Oscilloscope __ 11 How to Carry out the General Inspection _______________________________________15 How to Carry out the Function Inspection ______________________________________16 How to Implement Probe Compensation________________________________________18 How to Set the Probe Attenuation Coefficient____________________________________ 19 How to Use the Probe Safely __________________________________________________ 20 How to Implement Auto-calibration____________________________________________ 21 Introduction to the Vertical System ____________________________________________22 Introduction to the Horizontal System__________________________________________ 23 Introduction to the Trigger System_____________________________________________24
Advanced User's Guidebook ______________________________________________ 25
How to Set the Vertical System ________________________________________________ 26 Implementation of Mathematical Manipulation Function__________________________ 32 Application of VERTICAL POSITION and VOLTS/DIV Knobs ____________________ 33 How to Set the Horizontal system______________________________________________ 34 Main Time Base ____________________________________________________________ 35 Window Setting_____________________________________________________________ 36 Window Expansion__________________________________________________________37 How to Set the Trigger System ________________________________________________38 Trigger Control_____________________________________________________________ 39 Edge Trigger _______________________________________________________________39 Video Trigger ______________________________________________________________ 42 How to Operate the Function Menu____________________________________________ 45 How to Implement Sampling Setup ____________________________________________ 45 How to Set the Display System ________________________________________________ 48
Page 6
DS25 Colour Digital Storage Oscilloscope User Manual
6
Persist ____________________________________________________________________ 50
XY Format ____________________________________________________________________50
How to Save and Recall a Wave Form __________________________________________52 Save and Recall the Wave Form _______________________________________________53 How to Carry out the Auxiliary System Function Setting __________________________54 Do Self Cal (Self-Calibration) _________________________________________________ 55 SYS STAT (System State)_____________________________________________________55 How to Conduct Automatic Measurement_______________________________________ 57 Measure___________________________________________________________________ 57 How to Carry out Cursor Measurement ________________________________________59 Cursor Measurement________________________________________________________ 60 How to Use Executive Buttons_________________________________________________62 AUTOSET_________________________________________________________________ 62
Examples of Application _________________________________________________ 63
Example 1: Measurement of Simple Signals _____________________________________63 Example 2: Gain of the Amplifier in the Metering Circuit__________________________ 64 Example 3: Capture a Single Signal____________________________________________ 65 Example 4: Analyze the Details of a Signal ______________________________________ 67 Example 5: Application of X-Y Function________________________________________ 69 Example 6: Video Signal Trigger ______________________________________________71
Fault Treatment________________________________________________________ 73 Appendix A: Technical Specifications ______________________________________ 74 Appendix B: Enclosure __________________________________________________ 79 Appendix C: Maintenance, Cleaning and Repair _____________________________ 79
Page 7
DS25 Colour Digital Storage Oscilloscope User Manual
7
General Safety Requirements
Before operating, please read the following safety precautions to avoid any possible injury and prevent this product or any other products connected from damage. In order to avoid any contingent danger, this product is only to be used within the range specified.
Only qualified technicians should implement maintenance procedures.
Prevention of Fire or Bodily Injury. Use the correct power line. Only use the power cord specially provided for this
product or one which has been approved. Connect or Disconnect Correctly. When the probe or testing wire is connected to the
power lead, please do not connect and disconnect the probe or testing wire freely. Product Grounding. This product is grounded through the power lead grounding
conductor. In order to prevent any electric shock, the grounding conductor must be connected to the ground. It requires guarantee that this product has been already grounded correctly before any connection with its input or output terminal.
Connect the probe correctly. The grounding end of the probe corresponds to the grounding phase. Please don't connect the grounding end to the positive phase. Pay attention to the nominal values of all terminals. In order to prevent any fire or electric shock risks, please pay attention to all the nominal values and marks of this product. Before implement any connections for this product, please read the user's manual of this product to understand the information about the rated values further.
Do not operate without the instrument cover installed. If the cover or panel has already been removed, please don't operate this product.
Use the proper fuse. Only the fuse complying with the specified type and nominal value for this product can be used.
Avoid touching any exposed circuit. When the product is on power, please don't touch the uncovered contacts and parts.
Please do not operate if a fault is suspected. If suspecting damage to this product, please contact the qualified maintenance personnel for check.
Keep a good ventilation condition. Please consult the detailed installation instruction in the user's manual so that this product can be erected correctly, keeping it under a good ventilation condition.
Do not operate in a moist environment. Do not operate in an explosive environment . Keep the product’s surface clean and dry.
Page 8
DS25 Colour Digital Storage Oscilloscope User Manual
8
Safety Terms and Symbols
Terms in this manual. The following terms may appear in this manual:
Warning. A warning statement indicates the conditions and actions which may
endanger the life safety.
Note. A note statement indicates the conditions and actions which may cause
damage to this product or other property.
Terms on the product. The following terms may appear on this product: Danger: It indicates that there may be an immediate injury to you when you encounter
this mark. Warning: It indicates that there may not be an immediate injury to you when you
encounter this mark.
Note: It indicates that there may be damage to this product or other property. Symbols on the products. The following symbol may appear on the products:
High Voltage Please Consult the Manual. Protective Ground End Grounding End for Measurement Earth End on the Shell
Page 9
DS25 Colour Digital Storage Oscilloscope User Manual
9
General Characteristics of the DS25 Colour Digital Storage Oscilloscope
Bandwidth of 25MHz; Record length of 6,000 points for each channel; Sampling rates of 100MS/s for each channel; Read-out with the cursor; Five automatic measurement functions; Color liquid crystal display with high resolution and high contrast including
adjustable back light;
Store and recall waveforms; Automatic setting function provides fast set up; Multiple-waveform calculation function; Detect average and peak values of a waveform; Digital real-time oscilloscope; Edge and video triggering function; USB communication ports; Different continuous display times; User interface in two languages.
Page 10
DS25 Colour Digital Storage Oscilloscope User Manual
10
Junior User’s' Guidebook
This chapter deals with the following topics mainly:
Introduction to the front panel and the user’s interface of the DS series oscilloscope How to carry out general inspection How to carry out functional inspection How to make a probe compensation How to set the probe attenuation coefficient How to use the probe safely How to implement an auto-calibration Introduction to the vertical system Introduction to the horizontal system Introduction to the trigger system
Page 11
DS25 Colour Digital Storage Oscilloscope User Manual
11
Introduction to the Front Panel and the User's Interface of the DS series Oscilloscope
When you get a new-type oscilloscope, you should get acquainted with its front panel at first and the DS series digital storage oscilloscope is no exception. This chapter makes a simple description of the operation and function of the front panel of the DS series oscilloscope, enabling you to be familiar with the use of the DS series oscilloscope in the shortest time. The DS series oscilloscope offers a simple front panel with distinct functions for completing basic operations, in which the knobs and function pushbuttons are included. The knobs have the functions similar to other oscilloscopes. The 5 buttons in the column on the right side of the display screen are menu selection buttons (defined as F1 to F5 from top to bottom respectively), through which, you can set the different options for the current menu. The other pushbuttons are function buttons, through which, you can enter different function menus or obtain a specific function application directly.
Page 12
DS25 Colour Digital Storage Oscilloscope User Manual
12
Fig. 1 The Front Panel of a DS series Oscilloscope
Function Buttons
Fig. 2 Explanatory Drawing for Operation of the DS series Oscilloscope
Back-light
Adjustment
Menu Selection
Vertical Control
Connector
Horizontal
Control
Trigger Control
Page 13
DS25 Colour Digital Storage Oscilloscope User Manual
13
Fig. 3 Illustrative Drawing of Display Interfaces
1. The Trigger State indicates the following information:
Auto: The oscilloscope is under the Automatic mode and is
collecting the waveform under the non-trigger state.
Trig' d: The oscilloscope has already detected a trigger signal and is
collecting the after-triggering information.
Ready: All pre-triggered data have been captured and the
oscilloscope has been already ready for accepting a trigger.
Scan: The oscilloscope captures and displays the waveform data
continuously in the scan mode.
Stop: The oscilloscope has already stopped the waveform data
acquisition.
2. Waveform Viewing Area.
3. The purple pointer indicates the horizontal trigger position, which can be adjusted
by the horizontal position control knob.
4. The reading shows the time deviation between the horizontal trigger position and
the screen centre line, which equals 0 in the screen center.
5. This reading shows the time deviation between the horizontal trigger position and
the window centre line, which is regarded as 0 in the window center.
6. It indicates the current function menu.
Page 14
DS25 Colour Digital Storage Oscilloscope User Manual
14
7. It indicates the operation options for the current function menu, which changes
with the function menus.
8. The purple pointer shows the trigger level position.
9. The reading shows the trigger level value.
10. The reading shows the trigger source.
11. It shows the selected trigger type:
Rising edge triggering
Falling edge triggering
Video line synchronous triggering
Video field synchronous triggering
12. The reading shows the window time base set value.
13. The reading shows the main time base set value.
14. The two yellow dotted lines indicate the size of the viewing expanded window.
15. The icon shows the coupling mode of the CH2 channel.
“—“ indicates the direct current coupling “ ~” indicates the AC coupling
16. The reading shows the vertical scale factor (the Voltage Division) of the CH2
channel.
17. The icon indicates the coupling mode of the CH1 channel:
The icon "–" indicates the direct current coupling The icon "~" indicates the AC coupling
18. The reading indicates the vertical scale factor (the Voltage Division) of the CH1
channel.
19. The information shows the zero point positions of CH1 or CH2 channel.
20. The yellow pointer shows the grounding datum point (zero point position) of the
waveform of the CH2 channel. If the pointer is not displayed, it shows that this channel is not opened.
21. The red pointer indicates the grounding datum point (zero point position) of the
waveform of the CH1 channel. If the pointer is not displayed, it shows that the channel is not opened.
22. The positions of two purple dotted line cursors measurements.
Page 15
DS25 Colour Digital Storage Oscilloscope User Manual
15
How to Carry on the General Inspection
After you receive a new DS series oscilloscope, it is recommended that you should make a check of the instrument according to the following steps:
1. Check whether there is any damage caused by transportation.
If it is found that the packaging carton or the foamed plastic protection cushion has suffered serious damage, do not throw it away until the complete device and its accessories satisfy the electrical and mechanical property tests.
2. Check the Accessories
The supplied accessories have been already described in the Appendix B “Accessories” of this Manual. You can check whether there is any loss of accessories with reference to this description. If it is found that there is any accessory lost or damaged, please get in touch with the distributor of digimess responsible for this service.
3. Check the Complete Instrument
If it is found that there is damage to the appearance of the instrument, or the instrument does not work normally, or fails in the performance test, please get in touch with the digimess distributor responsible for this business. If there is damage to the instrument caused by the transportation, please keep the package.
Page 16
DS25 Colour Digital Storage Oscilloscope User Manual
16
How to Carry out the Function Inspection
Make a quick function check to verify the normal operation of the instrument, according to the following steps:
1. Connect the Instrument to the Power and Push down the Power Switch Button.
The instrument carries out all self-check items and shows the prompt “Press any Key Enter the Operating Mode”. Press the “UTILITY” button to get access to the “FUNCTION” menu and push down F2 the menu selection button to call out the function “Recall Factory”. The default attenuation coefficient set value of the probe in the menu is 10X, shown as Fig. 4.
Fig. 4 Power on the Instrument
2. Set the Switch in the Oscilloscope Probe as 10X and Connect the Oscilloscope with CH1 Channel.
Align the slot in the probe with the plug in the CH1 connector BNC, and then tighten the probe with rotating it to the right side.
Connect the probe tip and the ground clamp to the connector of the probe compensator, shown as Fig. 5.
Page 17
DS25 Colour Digital Storage Oscilloscope User Manual
17
Probe comp
CH1
Fig. 5 Connection of the Probe
3. Press the AUTOSET Button.
The square wave of 1 KHz frequency and 5V peak-peak value will be displayed in several seconds (see Fig. 6).
Fig. 6 Autoset
Check CH2 by repeating Step 2 and Step 3.
Page 18
DS25 Colour Digital Storage Oscilloscope User Manual
18
How to Implement the Probe Compensation
When connecting the probe with any input channel for the first time, perform this adjustment to match the probe with the input channel. A probe which is not compensated or presents a compensation deviation will result in a measuring error or mistake. For adjusting the probe compensation, please carry out the following steps:
1. Set the attenuation coefficient of the probe in the menu as 10X and that of the switch in the probe as 10X, and connect the oscilloscope probe with the CH1 channel. If a probe hook tip is used, ensure that it keeps in close touch with the probe. Connect the probe tip with the signal connector of the probe compensator and connect the reference wire clamp with the ground wire connector of the probe connector, and then press the button AUTOSET (see Fig. 5).
2. Check the displayed wave forms and regulate the probe till a correct compensation is achieved (see Fig. 7 and Fig. 8).
Fig. 7 Displayed Wave Forms of the Probe Compensation
3. Repeat the steps mentioned if necessary.
Fig. 8 Adjust Probe
Page 19
DS25 Colour Digital Storage Oscilloscope User Manual
19
How to Set the Probe Attenuation Coefficient
The probe has several attenuation coefficients, which will influence the vertical scale factor of the oscilloscope.
If it is required to change (check) the set value of the probe attenuation coefficient, press the function menu button of the channels used, then push down the selection button corresponding to the probe till the correct set value is shown.
This setting will be valid all the time before it is changed again.
Note: The attenuation coefficient of the probe in the menu is preset to 10X when
the oscilloscope is delivered from the factory.
Make sure that the set value of the attenuation switch in the T5100 probe is the same as the menu selection of the probe in the oscilloscope.
The set values of the probe switch are 1X and 10X (see Fig. 9).
Fig. 9 Attenuation Switch
Note: When the attenuation switch is set to 1X, the T5100 probe will limit the
bandwidth of the oscilloscope in 5MHz. If it is needed to use the whole bandwidth of the oscilloscope, the switch must be set to 10X.
Page 20
DS25 Colour Digital Storage Oscilloscope User Manual
20
How to Use the Probe Safely
The safety guard ring around the probe body protects your finger against the electric shock, shown as Fig. 10.
Fig. 10 Finger Guard
Warning: In order to avoid suffering from an electric shock, please keep your
finger behind the safety guard ring of the probe body during operation.
In order to protect you from suffering an electric shock during your use of the
probe, do not touch the metal part of the probe tip when the probe is connected to the signal source.
Before making any measurements, please connect the probe to the instrument and
connect the ground terminal to the earth.
Page 21
DS25 Colour Digital Storage Oscilloscope User Manual
21
How to Implement Auto-calibration
The auto-calibration application can make the oscilloscope reach the optimum condition rapidly to obtain the most accurate measurement value. You can carry out this application program at any time, but when the range of variation of the ambient temperature is up to or over 5C, this program must be executed.
For the performing of the self-calibration, all probes or wires should be disconnected with the input connector first. Then, press the “UTILITY” button to call out the FUNCTION menu; push down the F3 menu selection button to choose the option “ Do Self Cal”; finally, run the program after confirming that everything is ready now.
Page 22
DS25 Colour Digital Storage Oscilloscope User Manual
22
Introduction to the Vertical System
Shown as Fig.11, there are a series of buttons and knobs in VERTICAL CONTROLS. The following practices will gradually direct you to be familiar with
the use of the vertical setting.
Fig. 11 Vertical Control Zone
1. Use the button “VERTICAL POSITION” knob to show the signal in the center of the waveform window. The “VERTICAL POSITION” knob functions the regulating of the vertical display position of the signal. Thus, when the “VERTICAL POSITION” knob is rotated, the pointer of the earth datum point of the channel is directed to move up and down following the wave form.
Measuring Skill
If the channel is in the DC coupling mode, you can rapidly measure the DC component of the signal through the observation of the difference between the wave form and the signal ground.
If the channel is in the AC mode, the DC component will be removed by filtration. This mode helps you display the AC component of the signal with a higher sensitivity.
2. Change the Vertical Setting and Observe the Consequent State Information Change.
With the information displayed in the status bar at the bottom of the waveform window, you can determine any changes in the channel vertical scale factor.
Rotate the vertical “VOLTS/DIV” knob and change the “Vertical Scale Factor
(Voltage Division) ”, it can be found that the scale factor of the channel corresponding to the status bar has been changed accordingly.
Press buttons of “CH1 MENU”, “CH2 MENU” and “MATH MENU”, the
operation menu, symbols, wave forms and scale factor status information of the corresponding channel will be displayed in the screen.
Page 23
DS25 Colour Digital Storage Oscilloscope User Manual
23
Introduction to the Horizontal System
Shown as Fig.12, there are a button and two knobs in the “HORIZONTAL CONTROLS”. The following practices will gradually direct you to be familiar with
the setting of the horizontal time base.
Fig. 12 Horizontal Control Zone
1. Use the horizontal “SEC/DIV” knob to change the horizontal time base setting and observe the consequent status information change. Rotate the horizontal “SEC/DIV” knob to change the horizontal time base, and it can be found that the “Horizontal Time Base” display in the status bar changes accordingly. The horizontal scanning speed steps from 5ns up to 5s in the sequence of 1=2=5.
2. Use the “HORIZONTAL POSITION” knob to adjust the horizontal position of the signal in the waveform window. The “HORIZONTAL POSITION” knob is used to control the triggering displacement of the signal or for other special applications. If it is applied to triggering the displacement, it can be observed that the wave form moves horizontally with the knob when you rotate the “Horizontal Position” knob.
3. With the “HORIZONTAL MENU” button pushed down, you can set and initiate the Window Expansion.
Page 24
DS25 Colour Digital Storage Oscilloscope User Manual
24
Introduction to the Trigger System
Shown as Fig.13, there are a knob and four buttons in the “TRIGGER CONTROLS”. The following practices will direct you to be familiar with the setting
of the trigger system gradually.
Fig. 13 Trigger Control Zone
1. Press the “TRIG MENU” button and call out the trigger menu. With the operations of the 5 menu selection buttons, the trigger setting can be changed.
2. Use the “LEVEL” knob to change the trigger level setting. With the rotation of the “LEVEL” knob, it can found that the trigger indicator in the screen will move up and down with the rotation of the knob. With the movement of the trigger indicator, it can be observed that the trigger level value displayed in the screen changes.
3. Press the button “SET TO% 50” to set the trigger level as the vertical mid point values of the amplitude of the trigger signal.
4. Press the “FORCE TRIG” button to force a trigger signal, which is mainly applied to the “Normal" and "Single” trigger modes.
5. The “TRIG VIEW” button is used to reset the trigger horizontal position.
Page 25
DS25 Colour Digital Storage Oscilloscope User Manual
25
Advanced User's Guidebook
You should already be familiar with the initial operations of “VERTICAL CONTROLS”, “HORIZONTAL CONTROLS” and “TRIGGER CONTROLS”,
and the functions of the function areas, buttons and knobs in the front panel of the DS series oscilloscope. Based on the introduction of the previous Chapter, the user should have an intimate knowledge of the determination of the change of the oscilloscope setting through observing the status bar. If you have not been familiar with the above-mentioned operations and methods yet, we advise you to read the section of “Chapter One Junior Users' Guidebook”.
This chapter will deal with the following topics mainly:
How to Set the Vertical System How to Set the Horizontal System How to Set the Trigger System How to Carry on the Sampling Setup How to Set the Display System How to Implement the Storage and Call-out Operations How to Carry out the Auxiliary Function Setting How to Carry out the Automatic Measurement How to Carry out the Cursor Measurement How to Use Executive Buttons
It is recommended that you read this chapter carefully to get acquainted with the various measurement functions and other operating methods of the DS series oscilloscope.
Page 26
DS25 Colour Digital Storage Oscilloscope User Manual
26
How to Set the Vertical System
The VERTICAL CONTROLS includes three menu buttons such as CH1 MENU, CH2 MENU and MATH MENU, and four knobs such as VERTICAL POSITION , VOLTS/DIV ( one group for each of the two channels ).
Setting of CH1 and CH2
Every channel has an independent vertical menu and each item is set respectively based on the channel.
With the “CH1 MENU” or “CH2 MENU” menu button pushed down, the system shows the operation menu of the corresponding channel (see Fig. 14).
Fig. 14 Channel Setting Menu
Page 27
DS25 Colour Digital Storage Oscilloscope User Manual
27
The description of the Channel Menu is shown in the following list:
Function Menu Setting Description
Coupling AC
DC
Block the AC component in the input signal. Unblock the AC and DC components in the input signal.
Channel OFF
ON
Close the measurement channel. Open the measuring channel.
Probe 1X
10X 100X 1000X
Choose one according to the probe attenuation factor to make the vertical scale reading accurate.
Inverted OFF
ON
The wave form is displayed normally. Initiate the wave form inverted function.
1. Setting Channel Coupling
Taking the Channel 1 for example, the measured signal is a square wave signal containing the direct current bias. The operation steps are shown as below:
(1). Press the CH1 MENU button and call out the CH1 SETUP menu. (2). Press the F1 menu selection button and select the Coupling item as “AC” to
set the channel coupling as ac mode, under which the direct current component in the signal will be blocked.
Then, press the F1 menu selection button again and select the Coupling item as “DC”, setting the channel coupling as dc mode, under which both dc and ac components in the signal will be unblocked.
The wave forms are shown as Fig. 15 and Fig. 16.
Page 28
DS25 Colour Digital Storage Oscilloscope User Manual
28
Fig. 15 AC Coupling
Fig. 16 DC Coupling
Page 29
DS25 Colour Digital Storage Oscilloscope User Manual
29
2. Setting the Channel “ON/OFF”
Taking the Channel 1 for example, the operation steps are shown as below: (1). Press the CH1 MENU button and call out the CH1 SETUP menu. (2). Press the F2 menu selection button and select the Channel as OFF, with
Channel 1 switched off.
(3). Press F2 menu selection button again, select the channel as ON, with
Channel 1 is switched on.
3. Regulate the Attenuation Ratio of the Probe
In order to match the attenuation coefficient of the probe, it is required to adjust the attenuation ratio coefficient of the probe through the operating menu of the Channel accordingly. If the attenuation coefficient of the probe is 1:1, that of the oscilloscope input channel should also be set to 1X to avoid any errors presented in the displayed scale factor information and the measured data.
Take the Channel 1 as an example, the attenuation coefficient of the probe is 1:1, the operation steps is shown as follows:
(1). Press the CH1 MENU button, access CH1 SETUP menu. (2). Press the F3 menu selection button and select 1X for the probe. The Figure 17 illustrates the setting and the vertical scale factor when the probe
of the attenuation coefficient of 1:1. is used.
Fig. 17 Regulation of the Attenuation Ratio of the Probe
Page 30
DS25 Colour Digital Storage Oscilloscope User Manual
30
A List of the Attenuation Coefficients of Probes and the Corresponding Menu Settings.
Attenuation Coefficient of the Probe
Corresponding Menu Setting
1:1 1X 10:1 10X 100:1 100X 1000:1 1000X
5. Setting of Wave Form Inverted
Wave form inverted: the displayed signal is turned 180 degrees against the phase of the earth potential.
Taking the Channel 1 for example, the operation steps are shown as follows: (1). Press the CH1 MENU button and get access to the CH1 SETUP menu. (2). Press the F4 menu selection button and select ON in the Inverted. The wave
form inverted function is initiated.
(3). Press the F4 menu selection button again and select OFF for Inverted item.
The function of wave form inverted is closed off.
For the screen display, see Fig. 18 and Fig. 19.
Page 31
DS25 Colour Digital Storage Oscilloscope User Manual
31
Fig. 18 Wave Form not Inverted
Fig. 19 Wave Form Inverted
Page 32
DS25 Colour Digital Storage Oscilloscope User Manual
32
Implementation of Mathematical Manipulation Function
The Mathematical Manipulation function is used to show the results of the additive and subtraction operations between Channel 1 and Channel 2.
Taking the additive operation between Channel 1 and Channels 2 for example, the operation steps are as follows:
1. Press the MATH MENU button and call out the WAVE MATH menu.
2. Press the F3 menu selection button and choose CH1+CH2. The green calculated wave form M is displayed in the screen; press the F3 menu selection button again, the wave form M is closed off (see Fig. 20).
Fig. 20 Wave Form resulted from CH1 +CH2 Mathematical Manipulation
The corresponding FCL (Functional Capabilities List) of the Wave Form
Calculation
Setting Description
CH1-CH2 Subtract the Channel 2 wave form from the Channel 1
wave form.
CH2-CH1 Subtract the Channel 1 wave form from the Channel 2
wave form.
CH1+CH2 Add the Channel 1 wave form to the Channel 2.
Page 33
DS25 Colour Digital Storage Oscilloscope User Manual
33
Application of VERTICAL POSITION and VOLTS/DIV Knobs
1. The .VERTICAL POSITION knob is used to adjust the vertical positions of the wave forms of all Channels (including those resulted from the mathematical operation). The analytic resolution of this control knob changes with the vertical division.
2. The VOLTS/DIV knob is used to regulate the vertical resolution of the wave forms of all channels (including those obtained from the mathematical manipulation), which can determine the sensitivity of the vertical division with the sequence of 1-2-5. The vertical sensitivity goes up when the knob is rotated clockwise and goes down when the knob is rotated anticlockwise.
3. When the vertical position of the channel wave form is adjusted, the screen shows the information concerning the vertical position at the lower left corner (see Fig.21).
Fig. 21 Information about Vertical Position
Page 34
DS25 Colour Digital Storage Oscilloscope User Manual
34
How to Set the Horizontal system
The HORIZONTAL CONTROLS includes the HORIZONTAL MENU button and such knobs as HORIZONTAL POSITION and SEC/DIV.
1. HORIZONTAL POSITION knob: this knob is used to adjust the horizontal positions of all channels (include those obtained from the mathematical manipulation), the analytic resolution of which changes with the time base.
2. SEC/DIV knob: it is used to set the horizontal scale factor for setting the main time base or the window.
3. HORIZONTAL MENU button: with this button pushed down, the screen shows the operating menu (see Fig. 22).
Fig. 22 Time Base Mode Menu
The description of the Horizontal Menu is as follows:
Function Menu Setting Description
Main Time Base The setting of the horizontal main time base is
used to display the wave form. Window Setup A window area is defined by two cursors. Window Expansion
The defined window area for display is
expanded to the full screen.
Page 35
DS25 Colour Digital Storage Oscilloscope User Manual
35
Main Time Base
Press the F1 menu selection button and choose the Main Time Base. In this case, the HORIZONTAL POSITION and SEC/DIV knobs are used to adjust the main
window. The display in the screen is shown as Fig. 23.
Fig. 23 Main Time Base
Page 36
DS25 Colour Digital Storage Oscilloscope User Manual
36
Window Setting
Press the F2 menu selection button and choose Set Window. The screen will show a window area defined by two cursors. In this case, the HORIZONTAL POSITION and SEC/DIV knobs can be used to adjust the horizontal position and size of this window area (see Fig. 24).
Fig. 24 Window Setting
Page 37
DS25 Colour Digital Storage Oscilloscope User Manual
37
Window Expansion
Press the F3 menu selection button and choose Zone Window. As a result, the window area defined by two cursors will be expanded to the full screen size (see Fig.
25).
Fig. 25 Zone Window
Page 38
DS25 Colour Digital Storage Oscilloscope User Manual
38
How to Set the Trigger System
When the oscilloscope begins to collect the data and display the wave form depends on a trigger. Once it is set correctly, the trigger can transfer the unstable display into a meaningful wave form.
When beginning to collect data, the oscilloscope will collect adequate data to draw the wave form at the left side of the trigger point at first. It will continuously perform the data acquisition while waiting for the trigger condition. After a trigger is detected, the oscilloscope will continuously collect data enough to draw the wave form at the right side of the trigger point.
One knob and four function menu buttons are included in the trigger control zone. LEVEL: Trigger the level control knob and set the signal voltage corresponding to
the trigger point. SET TO %50: Set the trigger level as the vertical mid point value of the amplitude of
the trigger signal. FORCE TRIG: It is a force trigger button for the generation of a trigger signal,
which is mainly used in the “Normal” and "Single” triggering modes.
TRIG VIEW: Trigger the resetting of the horizontal position. TRIG MENU: It is a trigger menu button. When it is pressed, an operation menu will
be presented in the screen, shown as Fig. 26.
Fig. 26 Trigger Mode Menu
Page 39
DS25 Colour Digital Storage Oscilloscope User Manual
39
Trigger Control
There are two trigger modes: Edge Trigger and Video Trigger. Each trigger mode uses different function menu. The switching between two modes can be implemented by pressing the F1 menu selection button.
Edge Trigger: It happens when the trigger input passes through a given level along the set direction.
Video Trigger: Carry out the field or line video trigger of the standard video signal. The Edge and Video Trigger menus are described respectively as below:
Edge Trigger
Under the Edge Trigger mode, a trigger happens in the trigger threshold value of the input signal edge. When the Edge Trigger is selected, a trigger will occur in the rising or falling edge of the input signal.
The Edge Trigger Menu is shown as Fig. 27.
Fig. 27 Edge Trigger Menu
Page 40
DS25 Colour Digital Storage Oscilloscope User Manual
40
The Edge Trigger Menu is described as below:
Function Menu Setting Description
Slope Rising
Falling
Set a trigger on the rising edge of the signal. Set a trigger on the falling edge of the signal.
Source CH1
CH2 EXT
EXT/5
Set Channel 1 as the trigger signal of the source. Set Channel 2 as the trigger signal of the source. Set the external trigger channel as the trigger signal of the source. Divide the External Trigger Source by 5 to expand the external trigger level range.
Mode Auto
Normal
Single
Collect the wave form even if there is not trigger condition is detected. Collect the wave form only when the trigger condition is fulfilled. Collect a wave form when a trigger is detected then stop sampling.
Coupling DC
AC HF Rjc
LF Rjc
Block the direct current component. Unblock all components. Block the high-frequency signal and only unblock the high-frequency component. Block the low-frequency signal and only unblock the low frequency component.
Make such settings in Channel 1 as Rising for Slop, Auto for Mode and DC for Coupling. The operation steps are as follows:
1. Press the TRIG MENU button and call out the trigger menu.
2. Press the F1 menu selection button and select Edge for Type.
3. Press the F2 menu selection button and select Rising for Slope.
4. Press the F3 menu selection button and choose CH1 for Source.
5. Press the F4 menu selection button and choose Auto for Mode.
6. Press the F5 menu selection button and choose AC for Coupling. The display
in the screen is as Fig. 28.
7. Press the F2 menu selection button again and choose Falling for Slope. For
the screen display, see Fig. 29.
Page 41
DS25 Colour Digital Storage Oscilloscope User Manual
41
Fig. 28 Wave Form Triggered on the Rising Edge
Fig. 29 Wave Form Triggered on the Falling Edge
Page 42
DS25 Colour Digital Storage Oscilloscope User Manual
42
Video Trigger
After choosing “Video Trigger”, a trigger is possible in field or line of NTSC, PAL or SECAM standard video signal.
The operation menu of Video Trigger is shown as Fig. 30.
Fig. 30 Video Trigger Menu
The description of the Video Trigger menu is shown the following table:
Function Menu Setting Description
Polarity Normal
Inverted
Applicable to the video signal of low black level. Applicable to the video signal of high black level.
Source CH1
CH2 EXT
EXT/5
Set Channel 1 as the trigger signal of the source. Set Channel 2 as the trigger signal of the source. Set the external input channel as the trigger signal of the source. Divide the external trigger source by 5 and expand the external trigger level range.
Sync Line
Field
Set a trigger synchronization in the vide line. Set a trigger synchronization in the video field.
The operation steps for setting Channel 1 in Video Trigger Mode are as below:
Page 43
DS25 Colour Digital Storage Oscilloscope User Manual
43
1. Press the TRIG MENU button and get access to the trigger menu.
2. Press the F1 menu selection button and choose Video for Type.
3. Press the F2 menu selection button and choose Normal for Polarity.
4. Press the F3 menu selection button and choose CHI1 for Source.
5. Press the F4 menu selection button and choose Field for Sync. The screen
display is shown as Fig. 31.
6. Press the F4 menu selection button again and choose Line for Slope. The
screen display is shown as Fig. 32.
Page 44
DS25 Colour Digital Storage Oscilloscope User Manual
44
Fig. 31 Wave Form Triggered in the Video Field
Fig. 32 Wave Form Triggered in the Video Line
Page 45
DS25 Colour Digital Storage Oscilloscope User Manual
45
How to Operate the Function Menu
The function menu control zone includes 6 function menu buttons and 3 immediate-execution buttons: SAVE/REL, MEASURE, ACQUIRE, UTILITY,
CURSOR, DISPLAY, AUTOSET, RUN/STOP and HARDCOPY.
How to Implement Sampling Setup
Press the ACQUIRE button and the menu is displayed in the screen, shown as Fig.
33.
Fig. 33 ACQU MODE Menu
The description of the Sampling Setup Menu is shown as follows:
Function Menu Setting Description
Sampling General sampling mode. Peak Detection It is used for the detection of the jamming
burr and the possibility of reducing the confusion.
Average value It is used to reduce the random and
don’t-care noises, with the optional number of averages.
Averages 4, 16, 64,
128
Choose the number of averages.
Page 46
DS25 Colour Digital Storage Oscilloscope User Manual
46
Change the ACQU Mode settings and observe the consequent variation of the wave form displayed.
Fig. 34 Peak Detect mode, under which the burrs on the falling edge of the square wave
can be detected and the noise is heavy.
Page 47
DS25 Colour Digital Storage Oscilloscope User Manual
47
Fig. 35 Common ACQU Mode display, in which no burr can be detected.
Fig. 36 The displayed wave form after the noise is removed under the Average Mode, in
which the average number of 64 is set.
Page 48
DS25 Colour Digital Storage Oscilloscope User Manual
48
How to Set the Display System
Push down the DISPLAY button and the menu displayed in the screen is shown as Fig. 37.
Fig. 37 Display Set Menu
The description of the Display Set Menu is shown as follows:
Function Menu Setting Description
Type Vectors
Dots
The space between the adjacent sampling points in
the display is filled with the vector form. Only the sampling points are displayed.
Persist OFF
1sec 2sec 5sec Infinite
Set the persistence time for each sampling point.
Format YT
XY
Show the relative relation between the vertical voltage and the horizontal time. Channel 1 is displayed on the horizontal axis and Channel 2 on the vertical axis.
Carry Bitmap
Vectors
The data transmitted in communication are in the bitmap form. The data transmitted in communication are in the vector form.
Page 49
DS25 Colour Digital Storage Oscilloscope User Manual
49
Display Type: With the F1 menu selection button pushed down, you can shift between Vectors and Dots types. The differences between the two display types can be observed through the comparison between Fig. 38 and Fig.39.
Fig. 38 Display in the Vector Form
Fig. 39 Display in Dots form
Page 50
DS25 Colour Digital Storage Oscilloscope User Manual
50
Persist
When the Persist function is used, the persistence display effect of the picture tube oscilloscope can be simulated: the reserved original data is displayed in fade color and the new data is in bright color. With the F2 menu selection button, different persistence time can be chosen: 1sec, 2sec, 5sec, Infinite and Closed. When the “Infinite” option is set for Persist time, the measuring points will be stored until the controlling value is changed (see Fig. 40).
Fig. 40 Infinite Persistence Display
XY Format
This format is only applicable to Channel 1 and Channel 2. After the XY display format is selected, Channel 1 is displayed in the horizontal axis and Channel 2 in the vertical axis; the oscilloscope is set in the un-triggered sample mode: the data are displayed as bright spots and the sampling rate is 1MS/s and can not be changed.
The operations of all control knobs are as follows:
The Vertical VOLTS/DIV and the Vertical POSITION knobs of Channel 1 are
used to set the horizontal scale and position.
The Vertical VOLTS/DIV and the Vertical POSITION knobs of Channel 2
are used to set the vertical scale and position continuously.
Page 51
DS25 Colour Digital Storage Oscilloscope User Manual
51
The following functions can not work in the XY Format:
Reference or digital wave form Cursor Auto Set Time base control Trigger control
Operation steps:
1. Press the DISPLAY button and call out the Display Set Menu.
2. Press the F3 menu selection button and choose the form as XY. The display
format is changed to be XY mode (see Fig. 41).
Fig. 41 XY Display Mode
Page 52
DS25 Colour Digital Storage Oscilloscope User Manual
52
How to Save and Recall a Wave Form
Press the SAVE/REL button, you can save and call out the wave forms in the instrument. The menu displayed in the screen is shown as Fig. 42.
Fig. 42 Wave Form Save Menu
The description of the Wave Form Save Menu is shown as the following table:
Function Menu Setting Description
Source CH1
CH2 MATH
Choose the wave form to be saved.
WAVE A , B
C , D
Choose the address in or from which the
waveform is saved or can be get access to..
Save Save the wave form of the source in the
selected address.
CHA OFF
ON
Switch on or off the display of the stored
wave form .
Page 53
DS25 Colour Digital Storage Oscilloscope User Manual
53
Save and Recall the Wave Form
The DS series oscilloscope can store four wave forms, which can be displayed with the current wave form at the same time. The stored wave form called out can not be adjusted.
In order to save the wave form of the channel CH1 into the address A, the operation steps should be followed:
1. Press the F1 menu selection button and choose CH1 for Source.
2. Press the F2 menu selection button and choose A for Wave.
3. Press the F3 menu selection button and save the wave form.
4. Press the F4 menu selection button and choose ON for CHA. The stored wave
form A will be displayed in the screen. The voltage level and time base level will also be shown at the upper left corner of the display area at the same time (see Fig. 43).
Fig. 43 Wave Saving
Page 54
DS25 Colour Digital Storage Oscilloscope User Manual
54
How to Carry on the Auxiliary System Function Setting
Press the UTILITY button and the menu is displayed in the screen as Fig. 45.
Fig. 45 Function Menu
The description of the Auxiliary Function Menu is shown as the following table.
Function Menu Setting Description
System Status Display the system function menu. Recall Factory Call out the factory settings. Do Self Cal Carry out the self-calibration procedure. Language Chinese
ENGLISH
Choose the display language of the operating system.
Page 55
DS25 Colour Digital Storage Oscilloscope User Manual
55
Do Self Cal (Self-Calibration)
The self-calibration procedure can improve the accuracy of the oscilloscope under the ambient temperature to the greatest extent. If the change of the ambient temperature is up to or exceeds 5C, the self-calibration procedure should be executed to obtain the highest level of accuracy.
Before executing the self-calibration procedure, disconnect the probe or wire and the input connector. Then press the F3 menu selection button and choose “Do Self Cal”. After confirming it is all set, push down F3 button and choose “Do Self Cal”, entering the self-calibration procedure of the instrument.
SYS STAT (System State)
Press the F1 menu selection button and choose “SYS STAT” item. The menu pops up in the screen as Fig. 46.
Fig. 46 SYS STAT Menu
Page 56
DS25 Colour Digital Storage Oscilloscope User Manual
56
The “SYS STAT” menu is described as the following table:
Function Menu Setting Description
Horizontal Show the horizontal parameter of the channel. Vertical Show the vertical parameter of the channel. Trigger Show the parameters of the trigger system. Misc Adjust the Date and Time.
After entering into the SYS STAT menu, choose the corresponding function, with the corresponding parameters shown in the screen. If press the F1 menu selection button and choose the function item “Horizontal”, the Horizontal System State will be displayed in the screen. Press any other function button and exit from the SYS STAT menu (see Fig. 47).
Fig. 47 Horizontal System State
Page 57
DS25 Colour Digital Storage Oscilloscope User Manual
57
How to Conduct the Automatic Measurement
With the Measure button pressed down, an automatic measurement can be implemented. There are 5 types of measurements and 4 measurement results can be displayed simultaneously.
Press the F1 menu selection button to choose Source or Type menu. You can choose the channel to be measured from the Source menu and choose the measurement Type (Freq, Cycle, Mean, PK–PK, RMS and None). The menu is displayed as Fig. 48.
Fig. 48 Measure Menu
Measure
Four automatic measured values can be shown one time at maximum for the wave form of each channel. Only if the wave form channel is in the ON state can the measurement be carried out. No automatic measurement can be made for the saved or the mathematically manipulated wave form, or in the XY or Scan format.
Measure the frequency, the peak-to-peak value of the Channel CH1 and the mean, the
Page 58
DS25 Colour Digital Storage Oscilloscope User Manual
58
RMS of the Channel CH2, according the following steps:
1. Press the F1 menu selection button and choose Source.
2. Press the F2 menu selection button and choose CH1.
3. Press the F3 menu selection button and choose CH1.
4. Press the F4 menu selection button and choose CH2.
5. Press the F5 menu selection button and choose CH2.
6. Press the F1 menu selection button and choose Type.
7. Press the F2 menu selection button and choose Freq.
8. Press the F3 menu selection button and choose Pk-Pk.
9. Press the F4 menu selection button and choose Mean.
10. Press the F5 menu selection button and choose Cyc RMS. The measured value will be displayed in the reading window automatically (see Fig.
49).
Fig. 49 Automatic Measurement
Page 59
DS25 Colour Digital Storage Oscilloscope User Manual
59
How to Carry on the Cursor Measurement
Press the CURSOR button to display the cursor measurement function menu (CURS MEAS) in the screen, which includes Voltage Measurement and Time Measurement, shown as Fig. 50.
Fig. 50 CURS MEAS Menu
The description of the cursor measurement menu is shown as the following table:
Function Menu Setting Description
Type OFF
Voltage Time
Switch off the cursor measurement. Display the voltage measurement cursor and menu. Show time measures the cursor and menu.
Source CH1, CH2 Choose the channel generating the wave form to
which the cursor measurement will be applied. Delta Read the difference between cursors. Cursor 1 Read the position of Cursor 1 (the Time is read
with reference to the horizontal trigger position
and the Voltage is to the ground point). Cursor 2 Read the position of Cursor 1 (the Time is read
with reference to the horizontal trigger position
and the Voltage is to the ground point).
Page 60
DS25 Colour Digital Storage Oscilloscope User Manual
60
Cursor Measurement
When carrying out the cursor measurement, the position of Cursor 1 can be adjusted with the CURSOR 1 (VERTICAL POSITION) knob of Channel 1, and that of Cursor 2 can be adjusted with the CURSOR 2 (VERTICAL POSITION) knob of Channel 2.
Perform the following operation steps for the voltage cursor measurement of the channel CH1:
1. Press CURSOR and recall the Curs Meas menu.
2. Press the F1 menu selection button to choose Voltage for Type, with two purple
dotted lines displayed along the horizontal direction of the screen, which indicating CURSOR 1 and CURSOR 2.
3. Press the F2 menu selection button and choose CH1 for Source.
4. Adjust the positions of CURSOR 1 and CURSOR 2 according to the measured
wave form, with the absolute value of the voltage amplitude difference between Cursor 1 and Cursor 2 displayed in the increment window. The current position of Cursor 1 is displayed in the Cursor 1 window and that of Cursor 2 is displayed in the Cursor 2 window (see Fig. 51).
Fig. 51 Wave Form of Voltage Cursor Measurement
Page 61
DS25 Colour Digital Storage Oscilloscope User Manual
61
Carry out the following operation steps for the time cursor measurement of the channel CH1:
1. Press “CURSOR” and recall the CURS MEAS menu.
2. Press the F1 menu selection button and choose Time for Type, with two purple
dotted lines displayed along the vertical direction of the screen, which indicating Cursor 1 and Cursor 2.
3. Press the F2 menu selection button and choose CH1 for Source.
4. Adjust the positions of CURSOR 1 and CURSOR 2 according to the measured
wave form, with the cycle and frequency of Cursor 1 and Cursor 2 displayed in the increment window. The current position of Cursor 1 is displayed in the Cursor 1 window and that of Cursor 2 is displayed in the Cursor 2 window (see Fig. 52).
Fig. 52 Wave Form of Cursor Measurement
Page 62
DS25 Colour Digital Storage Oscilloscope User Manual
62
How to Use Executive Buttons
The executive buttons include AUTOSET, RUN/STOP and HARDCOPY.
AUTOSET
This button is used for the automatic setting of all control values of the instrument to generate the wave form suitable for observation. Press the AUTOSET button and the oscilloscope will perform the fast automatic measurement of the signal.
The function items of AUTOSET are shown as the following table:
Function Items Setting
Acquisition Mode Current Vertical Coupling DC Vertical Scale Adjust to the proper division. Bandwidth Full Horizontal Level Middle Horizontal Sale Adjust to the proper division Trigger Type Current Trigger Source Show the minimum number of channels. Trigger Coupling Current Trigger Slope Current Trigger Level Mid-point Setting Trigger Mode Auto Display Format YT
RUN/STOP:
Enable or disable the waveform sampling.
Notice:
Under the Stop state, the vertical division and the horizontal time base of the
wave form can be adjusted within a certain range, in other words, the signal can be expanded in the horizontal or vertical direction. When the horizontal time base equal to or is less than 50ms, the horizontal time base can be expanded for 4 divisions downwards.
HARDCOPY:
Press the HARDCOPY button and the oscilloscope will transmit the
current screen display to the computer connected with it. The receiving software should be installed in the superior computer, which should be connected the through the RS232 interface.
Page 63
DS25 Colour Digital Storage Oscilloscope User Manual
63
Examples of Application
Example 1: Measurement of Simple Signals
Observe an unknown signal in the circuit, and display and measure rapidly the frequency and peak-to-peak value of the signal.
1. Carry out the following operation steps for the rapid display of this signal:
1. Set the probe menu attenuation coefficient as 10X and that of the switch in the
probe switch as 10X.
2. Connect the probe of Channel 1 to the measured point of the circuit.
3. Push down the AUTOSET button. The oscilloscope will implement the AUTOSET to make the wave form optimized,
based on which, you can further regulate the vertical and horizontal divisions till the waveform meets your requirement.
2. Perform Automatic Measurement
The oscilloscope can measure most displayed signals automatically. Complete the following operations for the measurement of frequency, cycle, mean and peak-to-peak value:
1. Press the MEASURE button to show the automatic measurement function menu.
2. Press the F1 menu selection button and choose Source, with the Source menu
displayed.
3. Press the F2, F3, F4 and F5 menu selection buttons and choose CH1.
4. Press the F1 menu selection button again and choose Type, with the Type menu
displayed.
5. Press the F2 menu selection button and choose Freq.
6. Press the F3 menu selection button and choose Period.
7. Press the F4 menu selection button and choose Mean.
8. Press the F5 menu selection button and choose Pk-Pk. Then, the frequency, cycle, mean and peak-to-peak value will present in the menu and
change periodically (see Fig. 53).
Page 64
DS25 Colour Digital Storage Oscilloscope User Manual
64
Fig. 53 Waveform of Automation Measurement
Example 2: Gain of the Amplifier in the Metering Circuit
Set the probe menu attenuation coefficient as 10X and that of the switch in the probe as 10X.
Connect the oscilloscope CH1 channel with the circuit signal input end and the CH2 channel to the output end.
Operation Steps
1. Push down the AUTOSET button and the oscilloscope will automatically adjust
the waveforms of the two channels into the proper display state.
2. Push down the MEASURE button to show the MEASURE menu.
3. Press the F1 menu selection button and choose Source.
4. Press the F2 menu selection button and choose CH1.
5. Press the F3 menu selection button and choose CH2.
6. Press the F1 menu selection button again and choose Type.
7. Press the F2 menu selection button and choose Pk-Pk.
Page 65
DS25 Colour Digital Storage Oscilloscope User Manual
65
8. Press the F3 menu selection button and choose Pk-Pk.
10. Read the peak-to-peak values of Channel 1 and Channel 2 from the displayed
menu (see Fig. 54).
11. Calculate the amplifier gain with the following formulas.
Gain = Output Signal / Input signal
Gain (db) = 20×log (gain)
Fig. 54 Wave Form of Gain Measurement
Example 3: Capture the Single Signal
The digital storage oscilloscope takes the lead in providing the convenience capturing of such non-periodic signals as pulse and burr, etc.. If you intent to capture a single signal, you can not set the trigger level and the trigger edge unless you have a particular priori knowledge of this signal. For example, if the pulse is the logic signal of a TTL level, the trigger level should be set to 2 volts and the trigger edge be set as the rising edge trigger. If it is uncertain as to the signal, you can make an observation
Page 66
DS25 Colour Digital Storage Oscilloscope User Manual
66
of it in advance under the automatic or ordinary mode to determine the trigger level and the trigger edge.
The operation steps are as follows:
1. Set the probe menu attenuation coefficient to 10X and that of the switch in the
probe to 10X.
2. Adjust the VOLTS/DIV and SEC/DIV knobs to set up a proper vertical and
horizontal ranges for the signal to be observed.
3. Press the button ACQUIRE to display the ACQUIRE Mode menu.
4. Press the F2 menu selection button and choose Peak Detect.
5. Press the TRIG MENU button to display the Trigger Mode menu.
6. Press the F1 menu selection button and choose Edge as the trigger type.
7. Press the F4 menu selection button and choose Single as the trigger mode.
8. Press the F2 menu selection button and choose Rising as the slope.
9. Rotate the LEVEL knob and adjust the trigger level to the mid-value of the
signal to be measured.
10. If the Trigger State Indicator at the top of the screen does not indicate Ready,
push down the RUN/STOP button and start Acquire, waiting the emergence of the signal in conformity with the trigger conditions. If a signal reaches to the set trigger level, one sampling will be made and then displayed in the screen. With this function, any random occurrence can be captured easily. Taking the burst burr of larger amplitude for example, set the trigger level to the value just greater than the normal signal level, and then push down the RUN/STOP button and wait. When there is a burr occurring, the instrument will trigger automatically and record the wave form generated during the period around the trigger time. With the HORIZONTAL POSITION knob in the horizontal control area in the panel rotated, you can change the horizontal position of the trigger position to obtain the negative delay, making an easy observation of the waveform before the burr occurs (see Fig. 55).
Page 67
DS25 Colour Digital Storage Oscilloscope User Manual
67
Fig. 55 Capture the Single Signal
Example 4: Analyze the Details of a Signal
Observe the Signal Containing Noises
If the signal is interfered by the noise, the noise may cause a failure in the circuit. For the analyzing of the noise in detail, please operate the instrument according to the following steps:
1. Press the ACQUIRE button to display the ACQU MODE menu.
2. Press the F2 menu selection button and choose Peak Detect. In this case, the screen display contains the wave form of a random noise. Especially
when the time base is set as Low Speed, then noise peak and burr contained in the signal can be observed with the peak detection (See Fig. 56).
Page 68
DS25 Colour Digital Storage Oscilloscope User Manual
68
Fig. 56 Wave Form of the Signal Containing Noises
Separate Noises from the Signal
When analyze the wave form of a signal, you should remove the noise contained in it. For the reduction of the random noise in the oscilloscope display, please operate the instrument according to the following step:
1. Press the ACQUIRE button to show the ACQU MODE menu.
2. Press the F3 menu selection button and choose Average.
3. Press the F4 menu selection button and observe the wave form obtained from
averaging the wave forms of different average number.
After the averaging, the random noise is reduced and the detail of the signal is easier to be observed. Shown as follows, after the noise is removed, the burrs on the rising and falling edges appears (see Fig. 57).
Page 69
DS25 Colour Digital Storage Oscilloscope User Manual
69
]
Fig. 57 Wave Form of the Noise-Removed Signal
Example 5: Application of X-Y Function
Examine the Phase Difference between Signals of two Channels
Example: Test the phase change of the signal after it passes through a circuit network. Connect the oscilloscope with the circuit and monitor the input and output signals of
the circuit. For the examination of the input and output of the circuit in the form of X-Y
coordinate graph, please operate according to the following steps:
1. Set the probe menu attenuation coefficient for 10X and that of the switch in the
probe for 10X.
2. Connect the probe of channel 1 to the input of the network and that of Channel 2
to the output of the network.
3. Push down the AUTOSET button, with the oscilloscope turning on the signals of
the two channels and displaying them in the screen.
5. Adjust the VOLTS/DIV knob, making the amplitudes of two signals equal in the
rough..
Page 70
DS25 Colour Digital Storage Oscilloscope User Manual
70
6. Push down the DISPLAY button and recall the DISP SET menu.
7. Press the F3 menu selection button and choose XY for Format.
The oscilloscope will display the input and terminal characteristics of the network in the Lissajous graph form.
8. Adjust the VOLTS/DIV and VERTICAL POSITION knobs, optimizing the
wave form.
9. With the elliptical oscillogram method adopted, observe and calculate the phase
difference (see Fig. 58).
Fig. 58 Lissajous Graph
Based on the expression sinq =A/B or C/D, where, q is the phase difference angle, and the definitions of A, B, C, and D are shown as the graph above. As a result, the phase difference angle can be obtained, namely, q =± arcsin (A/B ) or ± arcsin (C/D). If the principal axis of the ellipse is in the I and III quadrants, the determined phase difference angel should be in the I and IV quadrants, that is, in the range of (0- π /2) or (3π / 2- 2π). If the principal axis of the ellipse is in the II and IV quadrants, the determined phase difference angle is in the II and III quadrants, that is, within the range of (π / 2 - π) or (π- 3π /2).
The signal must be
centered and kept in the
horizontal direction.
Page 71
DS25 Colour Digital Storage Oscilloscope User Manual
71
Example 6: Video Signal Trigger
Observe the video circuit of a television, apply the video trigger and obtain the stable video output signal display.
Video Field Trigger
For the trigger in the video field, carry out operations according to the following steps:
1. Press the TRIG MENU button to display the trigger menu.
2. Press the F1 menu selection button and choose Video for Type.
3. Press the F2 menu selection button and choose Normal for Polarity.
4. Press the F3 menu selection button and choose CH1 for Source.
5. Press the F4 menu selection button and choose Field for Sync.
6. Adjust the VOLTS/DIV, VERTICAL POSITION and SEC/DIV knobs to
obtain a proper wave form display (see Fig. 59).
Fig. 59 Wave Form Obtained from Video Field Trigger
Video Line Trigger
For the trigger in the video line, please operate according to the following steps:
Page 72
DS25 Colour Digital Storage Oscilloscope User Manual
72
1. Push down the TRIG MENU button to display the trigger menu.
2. Press the F1 menu selection button and choose Video.
3. Press the F2 menu selection button and choose Normal for Polarity.
4. Press the F3 menu selection button and choose CH1 for Source.
5. Press the F4 menu selection button and choose Line for Sync.
6. Adjust the VOLTS/DIV, VERTICAL POSITION and SEC/DIV knobs to
obtain the proper wave form display (see Fig. 60).
Fig. 60 Wave Form Obtained from the Video Line Trigger
Page 73
DS25 Colour Digital Storage Oscilloscope User Manual
73
Fault Treatment
1. In the case of that the oscilloscope is still in the black-screen state without any
display after the power is switch on, implement the following fault treatment procedure.
Check whether the power connection is connected properly. Check whether the power switch is pushed down to the designated position. Restart the instrument after complete the checks above. If the product still does not work normally, please contact your digimess
representative.
2. After acquiring the signal, carry out the following operations if the wave form of
the signal is not displayed in the screen.
Check whether the probe is properly connected to the signal connecting wire. Check whether the signal connecting wire is correctly connected to the BNC
(namely, the channel connector).
Check whether the probe is properly connected with the object to be measured. Check whether there is any signal generated from the object to be measured ( the
trouble can be shot by the connection of the channel from which there is a signal generated with the channel in fault).
Make the signal acquisition operation again.
3. The measured voltage amplitude value is 10 times greater or smaller than the
actual value.
Check whether the channel attenuation coefficient conforms with the attenuation ration of the probe used in practical application.
4. There is wave form displayed, but it is not stable.
Check whether the Source item in the TRIG MODE menu is in conformity with
the signal channel used in the practical application.
Check on the trigger Type item: The common signal chooses the Edge trigger
mode for Type and the video signal the Video. Only if a proper trigger mode is applied, the wave form can be displayed steadily.
Try to change the trigger coupling into the high frequency suppress and the low
frequency suppress to smooth the high frequency or low frequency noise triggered by the interference.
5. No Display Responses to the Push-down of RUN/STOP.
Check whether Normal or Signal is chosen for Polarity in the TRIG MODE menu and the trigger level exceeds the wave form range. If it is, make the trigger level is centered in the screen or set the trigger mode as Auto. In addition, with the AUTOSET button pressed, the setting above can be completed
Page 74
DS25 Colour Digital Storage Oscilloscope User Manual
74
automatically.
6. After the AVERAGE value sampling is set in the ACQU MODE or the a longer
duration is set in the DISP MODE, the display rate is slowed down.
It is a normal phenomenon.
Appendix A: Technical Specifications
Unless otherwise specified, the technical specifications applied are applicable to the probe with the attenuation switch setting 10X and the DS series digital oscilloscope. Only if the oscilloscope fulfills the following two conditions at first, can these specification standards be reached.
This instrument should run for more than 30 minutes continuously under the
specified operating temperature.
If the change range of the operating temperature is up to or exceeds 5C, open the
system function menu and execute the “Auto-calibration” procedure.
All specification standards can be fulfilled, except the one(s) marked with the word “Typical”.
Sampling
ACQU Mode Sample
Peak Detection Average
Sampling Rate 100 MS/s
Input
Input Coupling DC, AC Input Impedance 1MΩ± 2%, connected with 20pF± 3pF in parallel Probe attenuation c
oefficient
1X, 10X, 100X, 1000X
Max. Input Voltage 300V, Peak Value
Level
Sampling Rate Range 10S/s to 100MS/s Waveform interpolation (sinx) /x Record Length 6000 sampling points per channel Scanning Speed Range (S/div )
5ns/div to 5s/div, according to the stepping mode of
1- 2.5- 5. Accuracy of Sampling Rate and Delay Time
± 100ppm ( for any time interval equal to or greater
than 1ms) Measuring Accuracy of Time Interval (∆T) (Full Bandwidth)
Single: ± (One Sampling Interval Time +100ppm ×
Reading +0.6ns)
>Averages of 16: ± (One Sampling Interval Time
+100ppm× Reading +0.4ns)
Page 75
DS25 Colour Digital Storage Oscilloscope User Manual
75
Page 76
DS25 Colour Digital Storage Oscilloscope User Manual
76
Vertical
A/D Converter Resolution of 8 bits, with the sampling carried out
in two channels at the same time. Sensitivity (Volt/division) Range (V/div )
5mV/div- 5V/div (at the input BNC)
Displacement Range ± 10div (5mV- 5V ) Simulative Bandwidth 25MHz Single Bandwidth Full Bandwidth Low frequency respond (AC Coupling, - 3dB)
≥
5Hz (at BNC)
Rising Time (Typical, at BNC )
≤17.5ns
DC Gain Accuracy ± 5% DC Measuring Accuracy ( Average for ACQU MODE)
The voltage difference (∆V) between any two
points on the wave form after averaging the
captured wave forms of more than 16: ± (5%
Reading + 0.05 divisions).
Page 77
DS25 Colour Digital Storage Oscilloscope User Manual
77
Trigger
DC Coupling Channel 1 and Channel 2: 1div (DC to Full
Bandwidth ) EXT 100mV(DC to Full Bandwidth) EXT/5 500mV(DC to Full Bandwidth)
Trigger Sensitivity (Edge Trigger)
AC Coupling It is the same as the DC coupling in case of
50Hz or more. Internal ± 6 divisions from the screen center.
EXT ± 600mV
Trigger Level Range
EXT/5 ± 3V Inside ± 0.3 div
EXT ± (40mV + 6% of Set Value)
Trigger Level Accuracy (Typical): It is applicable to the signal with the rising and falling time ≥ 20ns
EXT/5 ± (200mV +6% of Set Value)
Trigger Displacement Preliminary Trigger: 655 divisions
Post Trigger: 4 divisions.
Set Level to 50% (Typical)
Operation under the condition of input signal frequency ≥ 50Hz. Internal Peak-to-peak value of 2 divisions EXT 400mV
Trigger Sensitivity (Video trigger, typical)
EXT/5 2V
Signal System and Line/Field Frequency (Video Trigger Type)
Supports the NTSC, PAL and SECAM broadcast systems for any field frequency of line frequency.
Measurement
Cursor Measurement Voltage difference (∆V) and time difference (∆T)
between cursors. Automatic Measurement
Peak-to-Peak Value, Average , RMS , Frequency and
Cycle
Probe
1X Position 10X Position Bandwidth DC: up to 4 MHz DC: up to full bandwidth Attenuation Ratio 1: 1 10: 1 Compensation Range 10pf to 35pf Input Resistance 1MΩ± 2% 10MΩ± 2% Input Electric Capacity 85pf- 115pf 14.5pf- 17.5pf Input Voltage 150 V DC 300 V DC
Page 78
DS25 Colour Digital Storage Oscilloscope User Manual
78
General Technical Specifications Display
Display Type 7.8 inch Color LCD (Liquid Crystal Display) Display Resolution 640 (Horizontal) × 480 (Vertical) Pixels Display Colors 256 Colors
Output of the Probe Compensator
Output Voltage (Typical )
Approx 5V, with the Peak-to-Peak value equal to or greater than 1MΩ of load.
Frequency (Typical ) Square wave of 1KHz
Power
Mains Voltage 100-240 VAC RMS, 50Hz, CAT II Power Consumption < 15W Fuse 1A, T grade, 250V
Environment
Temperature Working temperature: 0 to +40 C
Storage temperature: -20 to +60 C Relative Humidity ≤ 90% Height
Operating: 3,000 m
Non-operating: 15,000 m Cooling Method Natural convection
Mechanical Specifications
Dimension 350mm× 157mm×120mm Weight 1 kilogram (without battery option)
Interval Period of Adjustment:
One year is recommended for the calibration interval period.
Page 79
DS25 Colour Digital Storage Oscilloscope User Manual
79
Appendix B: Enclosure
Standard Accessories:
Passive probe: 2 pcs, 1.2 m, 1:1 (10:1) CDROM software USB cable Mains power cord. User's manual
Options:
DS25-BATT – Lithium battery
Appendix C: Maintenance, Cleaning and Repair
General Maintenance
Please don't store or put the instrument in the place where the liquid crystal display will be directly exposed to the sunlight for a long time.
Caution: The instrument or probe should not be stained with the spraying agent,
liquid and solvent to avoid any damage to it.
Cleaning
Check the probe and instrument regularly according to their operating state. Clean the external surface of the instrument following the steps shown as below:
1. Please wipe the dust from the instrument and probe surface with a soft cloth. Do
not make any scuffing on the transparent LCD protection screen when clean the LCD screen.
2. Clean the instrument with a wet soft cloth not dripping water, during the period
of which please pay attention to the disconnection of power. It is recommended to scrub with soft detergent or fresh water. Please don't apply any corrosive chemical cleaning agent to prevent the instrument or probe from damage.
Warning : Before power on again for operation, it is required to confirm that
the instrument has already been dried completely, avoiding any electrical short circuit or bodily injury resulting form the moisture.
Loading...