Dickson TH6P1, TH6P3, TH6P2 Getting Started

Page 1
1
DICK
DICKSON
DICKDICK
GETTING STARTED
GETTING STARTED
GETTING STARTEDGETTING STARTED 7-day chart rotation 0 to 100oF (TH6P3) +32 to 100F (TH6P1/2)
QUICK START
QUICK START
QUICK STARTQUICK START
SON
Remote Sensing
1.
1. Remove the protective pen caps.
1.1.
2.
2. Connect Sensor: TH6P1/2: Make sure the R200 replaceable sensor is connected firmly to
2.2. the recorder. TH6P3: Make sure the R250 sensor is connected firmly to the 10’ extension cable and the extension cable in plugged in firmly to the recorder.
3.
3. Insert 4AA batteries and plug in AC adapter (See Figure 3). Unit will power on.
3.3.
4.
4. The instrument will move pens to the appropriate readings.
4.4.
5.
5. Install the chart that matches selected Dip Switch settings (See p.4).
5.5.
Press the PEN automatically raised off the chart.
Remove the old chart, place the new chart on the Chart Hub being certain that the edge of the chart slides under the Chart Guide Clip located at the outside of the chart.
6.
6. Set the appropriate time. There are two ways to adjust the chart and set the appropriate
6.6. time:
PEN HOME
PEN PEN
a. Set the chart time manually by inserting a coin into the groove in the chart hub
and turning clockwise until the correct hour (and day if applicable) on the chart is referenced to the timing arrow (just to the right of the pens tip on the dial). Press Pen
Pen Home
PenPen
TTTTHHHH6P1 /
Remote Sensing Temperature
Remote SensingRemote Sensing
HOME key to move the pens to the outside of the chart. The pens are
HOMEHOME
Home to move the pens back onto the chart.
HomeHome
Temperature/Humidity
TemperatureTemperature
6P1 / TH6
TH6P2 /
6P1 / 6P1 /
/Humidity Recorder
/Humidity /Humidity
P2 / TH6P3
TH6TH6
P2 / P2 /
TH6P3
TH6P3TH6P3
Recorder
RecorderRecorder
Dickson TH6P Manual V5 052418
www. .com
information@itm.com1.800.561.8187
Page 2
2
DDDDISPLAY SYMBOLS
ISPLAY SYMBOLS TH6
ISPLAY SYMBOLS ISPLAY SYMBOLS
Figure 1
Figure 1
Figure 1Figure 1
TH6
TH6TH6
Figure 2
Figure 2 Figure 3
Figure 2Figure 2
Figure 3
Figure 3Figure 3
Keypad and Button Functions
Keypad and Button Functions
Keypad and Button FunctionsKeypad and Button Functions
On/Off
On/Off
On/OffOn/Off The On/Off key turns the unit on and off.
Pen
Pen Home
Home
PenPen
HomeHome If the pens are located on the outside edge of the chart, press Pen recording position. If the pens are located on the chart, press Pen outside edge of the chart.
Dickson TH6P Manual V5 052418
www. .com
Pen Home
Home to move pens to
PenPen
HomeHome
Pen Home
Home to move the pens to the
PenPen
HomeHome
information@itm.com1.800.561.8187
Page 3
3
AAAAlarm
#4
#4#4
#4
DDDD
own
ownown
own
#4
#4#4
#4
Down
DownDown
Down
#4
#4#4
#4
Down
DownDown
Down
#4
#4#4
#4
Up
UpUp
Up
larmssss ((((TH6P3
TH6P3))))
larmlarm
TH6P3TH6P3
Pen 1 = Red Pen
Pen 1 = Red Pen
Pen 1 = Red PenPen 1 = Red Pen Pen 2 = Blue Pen
Pen 2 = Blue Pen
Pen 2 = Blue PenPen 2 = Blue Pen
1.
1. To set the alarm, make sure unit is on and press and hold the Alarm
1.1.
red rapidly for about five seconds, then the LED will turn solid green. Release the Alarm button and the LED will turn solid red. At this point the display will show “On” or “Off”.
Pressing either the Adjust the dip switch, will toggle the alarm on or off.
2.
2. Pressing PEN
2.2.
3.
3. In order to set pen alarm minimums and maximums, pressing Adjust
3.3.
alarm value, and pressing Adjust the Adjust
Repetitively pressing the Pen button is pressed to exit from alarm adjust. Each press of the Pen store the new settings. The unit will take one minute to exit Alarm Set Mode once the Alarm button is pressed.
4.
4. If the alarm is triggered, the LED will show as solid red and the alarm will sound. Press the
4.4.
Alarm
Alarm button to silence the audible alarm.
AlarmAlarm
DIP SWITCH SETTINGS
DIP SWITCH SETTINGS
DIP SWITCH SETTINGSDIP SWITCH SETTINGS
PEN HOME
PENPEN a. Alarm on or off b. Pen 1 alarm minimum c. Pen 1 alarm maximum d. Pen 2 alarm minimum e. Pen 2 alarm maximum
Adjust----Up
AdjustAdjust
Up button is held down.
UpUp
Adjust----Up
AdjustAdjust
HOME will scroll to the next alarm option. The options are as follows:
HOMEHOME
Up or Adjust
Adjust----Down
UpUp
AdjustAdjust
Adjust----Dow
AdjustAdjust
Pen Home
Home button will scroll through the five options until the AAAAlarm
PenPen
HomeHome
Down buttons, located on the back of the unit next to
DownDown
Downnnn will decrease the alarm value. There is acceleration if
DowDow
Alarm button. The LED will blink
AlarmAlarm
Adjust----Up
AdjustAdjust
Pen Home
PenPen
Up will increase the
UpUp
Home or Alarm
HomeHome
Alarm button will
AlarmAlarm
Alarm
AlarmAlarm
Alarm
AlarmAlarm
larm
larmlarm
Recording Time
Recording Time:
Recording TimeRecording Time
7 Day
7 Day #2 Down
7 Day7 Day
24 Hour
24 Hour #2 Up
24 Hour24 Hour
Temperature Range
Temperature Range:
Temperature RangeTemperature Range
0 to +100 F (TH6P3)
0 to +100 F (TH6P3) #3
0 to +100 F (TH6P3)0 to +100 F (TH6P3)
32
32 to
to ++++10
3232
0 to +185 F (TH6P3)
0 to +185 F (TH6P3) #3
0 to +185 F (TH6P3)0 to +185 F (TH6P3)
----0 to +50C (All Models)
0 to +85C (TH6P3)
0 to +85C (TH6P3) #3 Up
0 to +85C (TH6P3)0 to +85C (TH6P3)
100 F
to to
1010
0 to +50C (All Models) ####3333 Up
0 to +50C (All Models)0 to +50C (All Models)
Dickson TH6P Manual V5 052418
: The TH6 recorder has two different recording time options
: :
: Select the temperature ranges that matches the chart paper
: :
0 F (TH6P1/2)
(TH6P1/2) ####3333 Down
0 F0 F
(TH6P1/2)(TH6P1/2)
#2 Down
#2 Down#2 Down
#2 Up
#2 Up#2 Up
#3 Down
Down
#3#3
DownDown
Down
DownDown
#3 Down
Down
#3#3
DownDown
#4
#4 Up
Up
#4#4
UpUp Up
UpUp
#3 Up
#3 Up#3 Up
www. .com
information@itm.com1.800.561.8187
Page 4
4
PENS
PENS
PENSPENS The blue pen has a longer pen arm and records humidity. The red pen has a shorter arm and records temperature. The pens are offset to allow the red pen to glide under the blue pen. The blue pen indicates the correct time and the red pen precedes it by 3/16 of an inch.
POWER
POWER
POWERPOWER The TH6 operates on AC power with an optional 4AA battery backup. Depending on chart rotation selected and model; battery backup will last approximately 3 days.
Note: Alarm will not operate under battery only power.
LED Indicators
LED Indicators
LED IndicatorsLED Indicators
• AC Power with Battery Backup - Solid Green
• AC Power with Low Battery or No Battery - Blinks Red
• Battery Only - Blinks Green
• Battery Only (Low Battery) - Solid Red
PEN
PEN HOME ADJUST
HOME ADJUST
PENPEN
HOME ADJUSTHOME ADJUST If over time the pen location does not match the display, you may need to adjust the pens location on the chart. A Pen Home Adjustment will not affect calibration of the sensor, but rather corrects for the natural drift that occurs with mechanical moving parts over time.
1. While the unit is on, press and hold both Pen Home solid green. Release the Pen Home green for one second then the LED will turn off. Note:
Note: Press Pen Home
Note:Note:
2. Both pens will move to the outer edge of the chart, and then the shorter red move to the outer chart ring (maximum temperature/RH line).
3. If the red pen tip does not line up with the outer chart ring line, use the Adjust Adjust
Adjust---- Down
AdjustAdjust on top of the outer chart ring.
4. Once the shorter red pen is set, press Pen Home longer blue pen to the out chart ring line.
5. If the blue pen tip does not line up with the outer chart ring line, use the Adjust Adjust
Adjust----Down
AdjustAdjust line.
6. Once the blue pen is set, pressing the Pen Home the outer edge of the chart and the shorter red pen will move to the outer chart ring line.
7. Each time PPPPen recently adjusted position) until On/Off
Note: Pressing the On/Off button will exit the adjustment and return the unit to normal operation. Each time the Pens Home or On/Off button is pressed, the adjustment currently in progress will be stored. The unit will take one minute to exit Pens Adjust Mode once the On/Off button is pressed.
Pen Home first so the unit will not turn off
Pen HomePen Home
Down buttons (on the back of the unit) to move the red pen so that the pen tip is
DownDown
Down buttons to move the blue pen so that the pen tip is on top of the outer ring
DownDown
en HHHHome
ome is pressed the two pens will switch positions (moving the most
en en
omeome
Pen Home and On/Off
Pen HomePen Home
On/Off is pressed to exit pen adjust mode.
On/OffOn/Off
Pen Home and On/Off
Pen HomePen Home
On/Off buttons. The LED will flash amber and
On/OffOn/Off
Pen Home to switch pen positions moving the
Pen HomePen Home
Pen Home button will move the longer blue pen to
Pen HomePen Home
On/Off buttons until the LED is
On/OffOn/Off
red pen will
redred
Adjust----Up
AdjustAdjust
Up and
UpUp
Adjust----Up
AdjustAdjust
Up and
UpUp
Dickson TH6P Manual V5 052418
www. .com
information@itm.com1.800.561.8187
Page 5
5
CALIBRATION
CALIBRATION
CALIBRATIONCALIBRATION
The R200/R250 Replaceable Sensor was carefully tested and calibrated before being shipped from the factory. For greatest accuracy, we recommend replacing the sensor every 6-12 months (does not include N400 and N450).
When your new Replaceable Sensor arrives, simply turn off the recorder, remove and discard the old sensor, plug in the new one and power the recorder back on. Your recorder will continue to record temperature and humidity without interruption.
USER CALIBRATION
USER CALIBRATION
USER CALIBRATIONUSER CALIBRATION If you have an accurate standard to compare against, the Replaceable Sensor calibration can be adjusted at one point. This will not adjust the span and is not as accurate as replacing the Replaceable Sensor with a newly calibrated one.
1. To activate Calibration mode, turn the unit on and press and hold both the On/Off Adjust
Adjust----Down
AdjustAdjust point only the pen being adjusted will show on the display.
Note:
Note: Make sure to press the Adjust-Down button first so the unit will not turn off
Note:Note:
Down button until the LED is solid green. The LED will then blink amber and at which
DownDown
On/Off and the
On/OffOn/Off
2. To raise the unit of measurement, press the Adjust measurement, press the Adjust and blue pens and stores the current adjustment value.
Adjust----Up
AdjustAdjust
Up button. Pressing Pen Home
UpUp
Adjust----Down
AdjustAdjust
Down button. To lower the unit of
DownDown
Pen Home switches between the red
Pen HomePen Home
3. When Calibration is complete, press the On/Off even after you turn the unit off or if AC power fails.
On/Off button. The adjustment is stored in memory
On/OffOn/Off
Note: After two hours, if no buttons are pressed, the unit will time out of Calibration mode
Note:
ote:Note:
N and resume normal operation. If you wish to cancel the Calibration, simply enter Calibration mode and toggle through the steps without adjusting displayed readings. Exit by pressing the On/Off
On/Off button. You have now restored factory calibration settings.
On/OffOn/Off
Note:
Note: It is recommended that you use a controlled chamber when determining if Calibration
Note: Note: is necessary. Adjusting the pens in an opens room is not recommended as air flow can vary greatly within a very small area.
TROUBLE SHOOTING
TROUBLE SHOOTING
TROUBLE SHOOTINGTROUBLE SHOOTING Why isn't the chart keeping time or r
Why isn't the chart keeping time or running slow?
Why isn't the chart keeping time or rWhy isn't the chart keeping time or r
The chart may be hung up or restricted, possibly caused by a rip on the outer edge of the
chart or the chart hub, or the chart may be caught between the arm and pens(s) arm platform.
Incorrect chart installed for the selected chart speed.
unning slow?
unning slow?unning slow?
Why did the chart stop turning?
Why did the chart stop turning?
Why did the chart stop turning?Why did the chart stop turning?
Chart hung up or restricted, (ripped chart)
Unit may be "locked up", this can be confirmed by pressing any of the buttons on the
keypad, if the unit is locked up there will be no response to button presses and the Chart Recorder may appear to be working, but the readings won't change, also the chart will not rotate. Remove power and battery, and then re-power.
Dickson TH6P Manual V5 052418
www. .com
information@itm.com1.800.561.8187
Page 6
6
Why don't the display and chart match?
Why don't the display and chart match?
Why don't the display and chart match?Why don't the display and chart match?
Dip switches set for a specific range, but using chart for another range, or vice versa.
Pens(s) not inserted on pens arm(s) all the way
To adjust pens(s) to match chart see “Pens Home Adjust” in the “Need to Know” section
of this manual.
Why does the display read PROB?
Why does the display read PROB?
Why does the display read PROB?Why does the display read PROB?
Check the sensor pod. Make sure it is properly plugged into the unit.
Is the unit locked up? This can be confirmed by pressing any of the buttons on the
keypad, if the unit is locked up there will be no response to button presses and the unit may appear to be working, but the readings won't change, also the chart will not rotate. Remove power and battery, and then re-power.
Why does calibration seem to be off?
Why does calibration seem to be off?
Why does calibration seem to be off?Why does calibration seem to be off?
What is the tolerance of the unit it's being compared to?
It's ok if the unit is within the sum of the two tolerances.
Has an outside calibration house attempted calibration? It may not have been adjusted
properly.
Walk through calibration adjustment, found in the manual
Why won't the battery back
Why won't the battery back----up work?
Why won't the battery backWhy won't the battery back
Are there good batteries in the Chart Recorder?
Keep in mind that battery backup varies significantly depending on temperature, pens
movement and chart rotation speed.
If there was a momentary power outage (Brown out), the unit may not have had enough
time to recognize this and switch to battery mode. (The Chart Recorder may lock up or shut off in this situation) This situation may also occur if the unit is plugged into an outlet that shares the circuit with other machinery that has phase motors or compressors that cycle periodically. As these other machines cycle, they momentarily have a high current draw, therefore drawing from the unit.
up work?
up work?up work?
Why won't the chart recorder respond to changes?
Why won't the chart recorder respond to changes?
Why won't the chart recorder respond to changes?Why won't the chart recorder respond to changes?
Is the unit locked up? This can be confirmed by pressing any of the buttons on the
keypad, if the unit is locked up there will be no response to button presses and the unit may appear to be working, but the readings won't change, also the chart will not rotate. Remove power and batteries, and then re-power.
Why won't the unit power up?
Why won't the unit power up?
Why won't the unit power up?Why won't the unit power up?
Remove the batteries and power adapter for a minute or two, this will reset the unit. The
unit should respond when the adapter is plugged in.
Dickson TH6P Manual V5 052418
www. .com
information@itm.com1.800.561.8187
Page 7
7
WARRANTY
WARRANTY
WARRANTYWARRANTY
Dickson warrants that this line of instruments will be free from defects in material and workmanship under normal use and service for a period of twelve months after delivery.
This warranty does not cover routine calibration and battery replacement.
Replaceable Sensor Certificate of Validation
Replaceable Sensor Certificate of Validation
eplaceable Sensor Certificate of ValidationReplaceable Sensor Certificate of Validation
R
Dickson ensures that the Replaceable Sensors / Instruments listed below were developed, tested and validated together.
Replaceable Sensor Model: R200 / R250
Instrument Model: TH6P1 / TH6P2 / TH6P3
Operation
Operation: The Instruments and Replaceable Sensor models listed above were specifically
OperationOperation designed to work together. No other sensors will work with the Instrument models listed. No other Instrument will work with the Replaceable Sensors listed. The Replaceable Sensor must be connected to the Instrument for the Instrument to operate properly.
Calibration
Calibration: Only the Replaceable Sensor is calibrated. The sensor, and all calibration defaults
CalibrationCalibration and adjustments are stored on the Replaceable Sensor. Accuracy is strictly controlled by the sensor – no adjustments are made by the Instrument. Readings are sent from the Replaceable Sensor to the Instrument for storage purposes only. Certificates of Calibration / NISTs are only supplied for the Replaceable Sensor for this reason.
Recalibration
Recalibration: When a Replaceable Sensor is due for recalibration the existing Replaceable Sensor
RecalibrationRecalibration can be replaced with a new Replaceable Sensor thus eliminating the need to return the entire instrument for recalibration. Replaceable Sensors can be returned for recalibration as well.
Dickson TH6P Manual V5 052418
www. .com
information@itm.com1.800.561.8187
Loading...