Min/Max
1. Min/Max readings are always displayed.
2. To reset Min/Max readings:
a. Hold down the Clear button (on the front of the unit) until CLR shows on the display. CLR will display while clearing
min/max.
b. The Min/Max readings will display the current temperature until next reading is taken.
3. Powering off the unit resets stored min/max values
Setting Alarm High/Low
1. The most common alarm settings have been preset for each channel:
a. MM120
i. Channel 1: 34°F (low) and 47°F (high) (refrigerator)
b. MM125
i. Channel 1 (top probe): -20°F (low) and 6°F (high) (freezer)
ii. Channel 2 (bottom probe): 34°F (low) and 47°F (high) (refrigerator)
2. To change the presets:
a. Press and release the Alarm set button on the back of the unit.
b. CH1 and LOW will flash on the display.
c. Using the UP and DOWN buttons on the back of the unit, set the target low temperature for channel 1.
d. Once the target temperature has been set, Press and release the Alarm Set button again to set CH1 / HIGH.
e. Repeat steps a – d above to toggle through CH1/LOW, CH1/HIGH, CH2/LOW, CH2/HIGH (CH2 MM125 only).
Note: If no buttons have been pressed after 15 s, the unit will exit Alarm Set mode and will save all new settings.
Clearing Alarm Conditions
1. When the unit enters alarm mode:
a. The Audio Alarm will sound for 30 s
b. The alarm symbol will flash, and the out of range high or low reading will flash.
c. If no action is taken to clear the alarm, the audio alarm will beep 2 times every 5 min and the display will continue to flash
until the clear button is pressed. This condition will continue even if the temperature returns to normal.
2. Clearing an Alarm
a. Press the clear button to temporarily stop the audible alarm. The display will continue flashing. If the alarm condition still
exists, the unit will sound again once the sample rate period has passed.
b. Once the unit moves out of the alarm condition, press and hold the Clear button until CLR is displayed. This will reset the
alarm and stop the flashing on the display.
3. Alarm Notification
a. If the alarm is not on (dip switch is down), the display will flash if the temperature moves into alarm range.
Note: If no alarm notification is desired, change alarm settings to high/low outer range limits (Low = -58°F, High = 158°F)
b. Once the temperature moves out of alarm range, press and hold the clear button until CLR is displayed to reset