High pressure liquids and gases are potentially hazardous. Energy stored in these liquids and gases
can be released unexpectedly and with extreme force. High pressure systems should be assembled
and operated only by personnel who have been instructed in proper safety practices.
This instrument is not to be operated in any other manner than that specified by the manufacturer.
Information in this document is subject to change without notice. No part of this document may be reproduced or transmitted in any
form or by any means, electronic or mechanical, for any purpose, without the express written permission of DH Instruments, a
Fluke Company 4765 East Beautiful Lane, Phoenix AZ 85044-5318, USA.
DH Instruments makes sincere efforts to ensure accuracy and quality of its’ published materials; however, no warranty, expressed
or implied, is provided. DH Instruments disclaims any responsibility or liability for any direct or indirect damages resulting from the
use of the information in this manual or products described in it. Mention of any product or brand does not constitute an
endorsement by DH Instruments of that product or brand. This manual was originally composed in English and was subsequently
translated into other languages. The fidelity of the translation cannot be guaranteed. In case of conflict between the English version
and other language versions, the English version predominates.
Products described in this manual are manufactured under international patents and one or more of the following U.S. patents:
5,142,483; 5,257,640; 5,331,838; 5,445,035. Other U.S. and international patents pending.
AutoRange, AutoZ, DH Instruments, DH, DHI, CalTool, COMPASS, RPM4, QDUT, Q-RPT, RPM4 and SDS are trademarks,
registered and otherwise, of DH Instruments, a Fluke Company.
Document No. 550129a-03
050815
Printed in the USA.
2.3.1 PREPARING FOR OPERATION ...................................................................................................................8
2.3.2 FRONT AND REAR PANELS ........................................................................................................................ 8
2.3.2.1 FRONT PANEL ..........................................................................................................................................8
2.3.3 POWER CONNECTION................................................................................................................................. 9
2.3.3.1 85 TO 264 VAC, 50/60 HZ VAC POWER ..................................................................................................9
2.3.8 TURN OFF ABSOLUTE AND NEGATIVE GAUGE MODE (AXXX RPT)....................................................12
2.3.9 SDS FULL TIME OFF ..................................................................................................................................12
2.4 POWER-UP AND VERIFICATION.........................................................................................................12
2.4.1 SWITCH POWER ON ..................................................................................................................................12
3.1 USER INTERFACE ................................................................................................................................15
3.1.1 MAIN RUN SCREEN....................................................................................................................................15
3.1.2 FUNCTION / DATA KEYPAD LAYOUT AND PROTOCOL.........................................................................17
3.3.6.3 DEV (DEVIATION) ...................................................................................................................................38
3.5.4.2 <2FULL TIME ON/OFF> ..........................................................................................................................66
5.2.2.1 GAS OPERATED Q-RPTS, A10M AND LOWER ..................................................................................117
5.2.2.2 OIL OR GAS OPERATED Q-RPTS A14M AND HIGHER .....................................................................117
5.2.3 SET-UP AND PREPARATION...................................................................................................................118
5.2.4 RECOMMENDED CALIBRATION POINT SEQUENCE............................................................................119
5.2.4.1 STANDARD CLASS Q-RPTS ................................................................................................................119
5.2.4.2 PREMIUM CLASS Q-RPTS ...................................................................................................................120
5.2.5 TURNING OFF ABSOLUTE AND NEGATIVE GAUGE M E ASU R E M E N T MODES F O R AXXX
5.2.6 Q-RPT CALIBRATION USING CALTOOL FOR RPTS SOFTWARE........................................................ 122
5.2.7 EDITING AND VIEWING Q-RPT CALIBRATION INFORMATION............................................................122
5.2.8 Q-RPT CALIBRATION/ADJUSTMENT WITHOUT CALTOOL FOR RPTS SOFTWARE.........................124
5.3 ADJUSTMENT OF ON-BOARD BAROMETER...................................................................................125
5.4 RELOADING EMBEDDED SOFTWARE INTO FLASH MEMORY......................................................126
5.5 SUBASSEMBLY DESCRIPTION AND LOCATION ............................................................................127
5.5.1 MINI MICRO BOARD................................................................................................................................. 127
5.5.2 POWER SUPPLY MODULE ......................................................................................................................127
5.5.5.1 HI Q-RPT MODULE ...............................................................................................................................128
5.5.5.2 LO Q-RPT MODULE ..............................................................................................................................128
Figure 1. Front Panel ...................................................................................................................................8
This manual is intended to provide the user with the basic information necessary to operate an RPM4
reference pressure monitor. It also includes a great deal of additional information provided to allow you to
optimize RPM4 use and take full advantage of its many features and functions.
RPM4-AD is a special air data configuration of RPM4 which has its own dedicated manual, p/n
550148. This manual is for all other configurations of RPM4. If your are using an RPM4-AD (“-AD” in
front panel configuration window) use the RPM4-AD Operation and Maintenance Manual p/n 550148.
Before using the manual, take a moment to familiarize yourself with the Table of Contents structure:
Sections 1, 2 and 3 should be read by all first time RPM4 users. Section 3 is most important for those
using the local front panel interface but should be read over by all users to familiarize themselves with
general RPM4 operating principles. Section 4 is for remote operation from an external computer. Section 5
provides maintenance and calibration information. Section 6 is a quick troubleshooting guide. Use it to
troubleshoot unexpected RPM4 behavior based on the symptom of that behavior. Certain words and
expressions have specific meaning as they pertain to RPM4. The Glossary, Section 6, is useful as a
quick reference for exact definition of specific words and expressions as they are used in the manual.
For those of you who “don’t read manuals”, go directly to Section 2.3 to set up your RPM4 and then
go to Section 2.4 for power-up and verification. This will get you up and running quickly with a minimal
risk of causing damage to yourself or your new RPM4. THEN… when you have questions or start to
wonder about all the great features you might be missing, get into the manual!
T
T
HHIISS
M
AANNUUAAL
M
L
Manual Conventions
(CAUTION) is used in throughout the manual to identify user warnings and cautions.
(NOTE) is used throughout the manual to identify operating and applications advice and
additional explanations.
[ ] indicates direct function keys (e.g., [RANGE]).
RPM4 is a stand-alone, microprocessor driven, reference pressure monitor intended to precisely
measure gas or liquid pressure in a wide variety of pressure calibration, measurement and testing
applications. It has been designed to provide very high performance and extensive features combined
with maximum versatility and ease of use.
RPM4 uses one or two quartz reference pressure transducer (Q-RPT) modules and, in some cases a
barometer, to measure pressure.
RPM4 can be controlled locally by the operator using its front panel display and keypad or remotely by a
computer using ASCII character command strings transmitted over its standard RS232 or IEEE-488.2
interface.
RPM4 models are available in ranges from as low as - 3 to 3 kPa (0.4 psi) to as high as 280 MPa (40 000 psi)
in absolute, gauge, compound gauge and differential measurement modes.
RPM4 can be integrated into a PPC3 Pressure Controller/Calibrator System to serve the PPC3’s external
pressure measuring reference (see the PPC3 Operation and Maintenance Manual).
A special configuration of RPM4 designated RPM4-AD is available for air date (altitude and
airspeed) applications. The RPM4-AD model has its own specific manual p/n 550148.
Pressure Connections Q-RPTs up to A70M Q-RPTs > A70M
Self Defense System (SDS™)
Pressure Limits
CE Mark
85 to 264 VAC, 50/60 Hz, 25 VA max consumption
and 12VDC, 1.2 A
15 to 35 °C
- 20 to 70 °C
Meets MIL-T-28800D
To prevent product overheating, provide proper ventilation. Allow 10 cm
(4 in.) clearance from rear panel cooling fan.
5 kg (11 lb) approx. (varies slightly with number and type of Q-RPT modules)
10 cm H x 22.7 cm W x 24 cm D (3.9 in. x 9.3 in. x 9.5 in.)
Motorola 68302, 16 MHz
RS232 (COM1, COM2), IEEE-488.2
1 A, 250 VAC fuse, 5 x 20 mm, time lag type fuse
Internal power supply fuse not replaceable by operator: 2.5A, 250 VAC
One or two independent quartz reference pressure transducer (Q-RPT)
modules cover vacuum to 280 MPa (40 000 psi)
Any clean, dry, non-corrosive gas or liquid. Q-RPTs less than 7 MPa full scale,
gas only
TEST(+): 1/8 in. NPT F DH500 F
TEST(-): 1/8 in. NPT F None
VENT or ATM: 10-32 UNF (VENT) 10-32 UNF (ATM)
DH500 is a gland and collar type fitting for 6mm (1/4 in.) coned and left
hand threaded tubes equivalent to AE F250C, HIP HF4, etc.
Included on all gas operated Q-RPT modules of A7M or lower. Isolates
Q-RPT module TEST(+) port and vents it to atmosphere. SDS is not included
with liquid filled Q-RPTs.
Maximum WorkingPressure: 104 % Q-RPT maximum
Maximum Pressure Without Damage: 125 % Q-RPT maximum
RPM4 can be configured with one or two quartz reference pressure transducer (Q-RPT)
modules to measure pressure. The type (Axxx, Gxxx, BGxxx, BAxxx) and range of
the Q-RPT module(s) determines the RPM4 measurement specifications.
All Q-RPTs whose maximum pressure is over 200 kPa (30 psi) are of the absolute
pressure type (Axxx) using an evacuated, permanently sealed reference.
Axxx Q-RPTs can measure absolute, gauge and negative gauge pressure. Gauge
pressure with an Axxx Q-RPT is defined by offsetting atmospheric pressure and
applying dynamic compensation for atmospheric changes using the on-board
barometer (see Section 3.2.2). Gxxx (gauge) Q-RPTs can measure positive gauge
pressure only. BGxxx (bi-directional gauge) Q-RPTs can measure gauge and
negative gauge pressure. See Section 3.3.3 for additional information on
absolute, gauge and negative gauge measurement modes.
RPM4s configured with two Q-RPT modules have independent TEST ports and can
measure separately. Two Q-RPTs can also be used together to operate in
differential mode (see Section 3.3.3.1) and parallel mode (see Section 3.2.4).
1. INTRODUCTION
Q-RPTs of A10M and lower are available with two different performance
levels, STANDARD class and PREMIUM class. See the product label on the
RPM4 rear panel, the Q-RPT module label on the rear panel and/or the
product calibration reports to determine the class of the Q-RPT(s) installed
in RPM4.
Table 1. Reference Pressure Transducer (Q-RPT) Module Designations and Ranges
30 minute temperature stabilization recommended from
cold power up.
Resolution
Compensated Temperature Range
Acceleration Affect
To 1 ppm, user adjustable
5 to 35 °C
± 0.008 % /g maximum, worst axis
Allows operation at ± 20° from reference plane without
significant effect
Predicted One Year Stability
(all classes and types)
± 0.005% of reading
1
Q-RPTs UP TO A10M (1500 psi)
Precision
2
Measurement Uncertainty3
STANDARD CLASS PREMIUM CLASS
± 0.008% of reading or
0.0024% of Q-RPT span,
whichever is greater
± 0.01% of reading or 0.0030%
of Q-RPT span, whichever is
greater
4
4
± 0.005% of reading, 0.0015%
of AutoRanged span, or
0.0005% of Q-RPT span,
whichever is greater
± 0.008 % of reading, 0.0024%
of AutoRanged span, or
0.0007% of Q-RPT span,
whichever is greater
5
5
Q-RPTs A14M TO A140M (2 000 to 20 000 psi)
2
Precision
Measurement Uncertainty3
± 0.012% of reading or 0.0036% of Q-RPT span, whichever is
4
greater
± 0.013% of reading or 0.0039% of Q-RPT span, whichever is
4
greater
Q-RPTs A200M TO A280M (30 000 to 40 000 psi)
2
Precision
Measurement Uncertainty3
1. Predicted Q-RPT measurement stability limit (k=2) over one year assuming regular use of AutoZero function. AutoZero
occurs automatically in gauge mode whenever vented, by comparison with barometric reference in absolute mode.
Absolute mode predicted one year stability without AutoZ is ± (0.005 % Q-RPT span + 0.005 % of reading).
2. Combined linearity, hysteresis, repeatability. Add + 1 Pa (0.00015 psi) in gauge mode with an Axxx (absolute) Q-RPT for
the resolution and short term stability of the on-board barometer.
3. Maximum deviation of the Q-RPT indication from the true value of applied pressure including precision, predicted one year
stability limit, temperature effect and calibration uncertainty, combined and expanded (k=2) following the ISO “Guide to the
Expression of Uncertainty in Measurement.”
4. % of reading value times measured pressure from 100 to 30 % of Q-RPT span. Under 30 % of Q-RPT span, % of reading
value times 30 % of Q-RPT span. For example, if the Q-RPT is a Standard A160K, the Measurement Uncertainty in
pressure is 0.010% times the measured pressure to 48 kPa (160 kPa span x 30%) and 0.0048 kPa (160 kPa span x 30%
x 0.01%) under 48 kPa.
5. % of reading value times measured pressure from 100 to 30 % of AutoRanged span. Under 30% of AutoRanged span, % of
reading value times 30% of AutoRanged span. If AutoRanged span is less then 30% of maximum Q-RPT span, % of reading
values times measured pressure, or % of reading times 9% of Q-RPT span, whichever is greater. For example, if the Q-RPT is a
Premium A160K and AutoRanged span is 160 kPa, the Measurement Uncertainty in pressure is measured pressure x 0.008% to
48 kPa (160 kPa AutoRanged span x 30%) and 0.0038 kPa (160 kPa span x 30% x 0.008%) under 48 kPa. If the AutoRanged
span is 100 kPa (greater than 30% of 160 kPa maximum Q-RPT span), the Measurement Uncertainty in pressure is measured
pressure x 0.008% to 30 kPa (100 kPa AutoRanged span x 30%) and 0.0025 kPa (100 kPa span x 30% x 0.008%) under 30 kPa.
If the AutoRanged span is 30 kPa (less than 30% of the 160 kPa maximum Q-RPT span), the Measurement Uncertainty in
pressure is measured pressure x 0.008% to 14.4 kPa (160 kPa maximum Q-RPT span x 9%) and 0.0012 kPa (160 kPa maximum
Q-RPT span x 9% x 0.008%) under 14.4 kPa.
± 0.015% of reading or 0.0045% of Q-RPT span, whichever is
4
greater
± 0.018% of reading or 0.0054% of Q-RPT span, whichever is
4
greater
1.2.2.2 ON-BOARD BAROMETER
The on-board barometer is used only to measure changes in atmospheric pressure
to provide dynamic compensation of the Q-RPT’s atmospheric pressure offset when
using an Axxx (absolute) Q-RPT to make gauge pressure measurements and for
line pressure compensation of G15K and BG15K Q-RPTs.
There is no on-board barometer in RPM4s that have G100K or G200K
RPM4 is delivered in a corrugated container with polyethylene inserts to hold it in place; or in
the optional molded, medium density polyethylene shipping case with a custom foam insert
for holding the RPM4.
Remove the RPM4 and its accessories from the shipping container and remove each
element from its protective plastic bag.
2.1.2 INSPECTING CONTENTS
Check that all items are present and have no visible damage.
A standard RPM4 includes all items indicated in Table 2.
Table 2. RPM4 Packing List
DESCRIPTION PART #
1 ea. RPM4 Reference Pressure Monitor FAM008
1 ea. Calibration Report 550100
ACCESSORIES: 401967 (401967-CE)
1 ea. Operation and Maintenance Manual 550129
1 ea. Power Cord (7.5 ft.) 100770 (100770-CE)
1 ea. Ferrite sleeve (snap on to COM2 cable for CE) 103303
Install RPM4 on a flat, stable surface at a convenient height. The front feet can be extended so that the
unit can be inclined for easier viewing. The RPM4 can also be mounted in a standard 19 in. rack mount
using the optional rack mount kit.
2.3 SETUP
2.3.1 PREPARING FOR OPERATION
To prepare RPM4 for check out and operation:
Remove the plastic caps from the RPM4 rear panel pressure connections.
Remove the protective plastic sheet from the front panel display.
Familiarize yourself briefly with the front and rear panels (see Section 2.3.2).
2.3.4 REMOTE [ENT] CONNECTION (FOOTSWITCH OR OTHER
SWITCH)
Connect the optional remote ENTER footswitch, if available or a user supplied switch fitted to
the optional cable (see Section 7.1). Connect the cable to the RPM4 rear panel connection
labeled REMOTE ENTER. Activating the switch is equivalent to pressing the [ENT] key on
the front panel (see Section 3.1.3).
2.3.5 CONNECTING TO MEASURE PRESSURE (TEST(+) AND
TEST(-) PORTS)
Using a pressure connecting hose or tube of appropriate pressure rating, connect the device
or system to be tested to the RPM4 TEST(+) port.
The RPM4 TEST(+) connection is 1/8 in. NPT female (Q-RPTs of A70M or lower) or DH500 F
(Q-RPTs higher than A70M).
DH500 is a gland and collar type fittings for 6.35 mm (1/4 in.) coned and left hand
threaded tubes. DH500 is equivalent to AE F250C, HIP HF4, etc.
The TEST(+) and TEST(-) ports on Q-RPT modules of A70M and lower are threaded into
the aluminum Q-RPT manifold. Take care in making and breaking fittings not to cross
thread or otherwise damage the threads. It is recommended that an adaptor be
installed in the in Q-RPT manifold and make break operations be made on the adaptor
rather than the manifold.
SDS Self Defense System: Q-RPT modules designated A7M or lower include the SDS Self
Defense System. SDS, operated properly, allows a the Q-RPT TEST(+) port to be left
connected to a pressure up to 10 MPa (1 500 psi) without damage to the Q-RPT.
Do NOT attempt to use SDS in this manner without first becoming thoroughly familiar
with its operation and limitations (see Sections 3.2.7, 3.3.8, 3.5.4).
Q-RPT MODULE TEST(+) AND TEST(-) PORTS
Q-RPT modules of A70M and lower have both TEST(+) and TEST(-) ports. Q-RPT modules
greater than A70M have a TEST(+) port only. See Section 5.6, Figure 8 for RPM4 Q-RPT
module TEST port configurations.
The TEST(+) port is connected to Axxx (absolute) Q-RPTs’ sole measurement port and to the
high side of Gxxx or BGxxx (gauge or bi-directional gauge) Q-RPTs.
The TEST(-) port of Axxx and BGxxx Q-RPTs is connected to RPM4’s internal barometer.
The TEST(-) is connected to the low side of Gxxx (gauge) or Gxxx (bi-directional gauge)
Q-RPTs.
•When operating in absolute mode: The TEST(-) port, if present, is left open to
atmosphere.
•When operating in gauge or negative gauge mode with a range greater than 50 kPa
(7.5 psi): The TEST(-) port, if present, is normally left open to atmosphere. A possible
exception is when the device or system under test is in an ambient pressure that may differ
significantly from the ambient pressure around the RPM4. For example, if the RPM4 is
measuring the pressure applied to DUTs in an environmental chamber, the pressure in the
environmental chamber may be different from ambient pressure around the RPM4. In this
case, connecting a tube from the TEST(-) port to the inside of the chamber may improve
measurement results. If the Q-RPT in use is an Axxx Q-RPT, this tube must be left open to
the environment so that the pressure inside cannot deviate too far from ambient. When using
an Axxx Q-RPT, if this tube is connected to the low or reference side of DUTs, be sure to
open it to the local environment as well.
•When operating in gauge or negative gauge mode with a range less than 50 kPa
(7.5 psi): As a general rule, it is preferable to connect the RPM4 TEST(-) port(s) directly to
the low or reference side of the device under test to assure that these are at the same
pressure. When using an Axxx Q-RPT, this connection must also be open to atmosphere.
When using a Gxxx or BGxxx Q-RPT, it is preferable that this connection not be open to
atmosphere.
Do not connect a pressure supply to the TEST(-) port. The pressure applied to this port
should be maintained at atmospheric pressure (between 70 and 110 kPa (10 and 16 psia)).
Exceeding these limits may damage a Gxxx or BGxxx Q-RPT and/or the RPM4’s on-board
barometer.
Using the RPM4 with gas operated Q-RPTs connected to a system with liquid
contaminants without taking proper precautions to purge the system and test line may
cause contamination of the RPM4 that will require non-warranty service.
2.3.6THE VENT OR ATM PORT
RPM4 Q-RPT modules of A7M and lower have a VENT port. The VENT port is connected to the
TEST(-) port when SDS is CLOSED (see Section 5.6, Figure 8). For Axxx and BGxxx Q-RPTs,
the RPM4 on-board barometer, is connected to the VENT port when SDS is CLOSED.
The VENT port should always be left completely unobstructed and open to atmosphere.
RPM4 Q-RPT modules of A10M and higher have an ATM port. The ATM port is connected
to the RPM4 on-board barometer. The ATM port should always be left completely unobstructed
and open to atmosphere.
NEVER plug, obstruct or connect a supply pressure to the RPM4 VENT or ATM port. This
may adversely affect GAUGE mode operation and AutoZeroing functions.
2.3.7 CHECK/SET SECURITY LEVEL
RPM4 has a security system based on user levels. By default, the security system is set to
“low”, which includes access restriction to internal calibration coefficients, and there is no
password required to change the security level. See Section 3.5.5.5 for information on the
security level system. As part of the RPM4 startup, determine the security level that is
appropriate for the RPM4 and set a password if desired.
RPM4 is delivered with the security level set to “low” to avoid inadvertent altering of critical
internal settings but with access to changing security levels unrestricted. It is recommended
that the low security level be maintained at all times and password protection be
implemented if control over setting of security levels is desired.
2.3.8 TURN OFF ABSOLUTE AND NEGATIVE GAUGE MODE
(AXXX RPT)
If your RPM4 has one or two Axxx (absolute) Q-RPTs, it is able to operate in
gauge, negative gauge and absolute measurement modes (see Section 3.3.3).
If the RPM4 will be used in gauge mode only, the other measurement modes can
be turned off so they are no longer accessible. Turning off modes that are not
needed can avoid confusion and/or accidental use of the wrong measurement mode.
See Section 5.2.5 for complete information on turning off absolute and negative
gauge measurement modes.
Turn off absolute and negative gauge mode if desired.
2.3.9 SDS FULL TIME OFF
Gas operated Q-RPT modules of A7M or less, include the SDS Self Defense
System (see Section 3.2.7) to shut off the Q-RPT from the TEST(+) port. If this
function is not desired, SDS can be turned full time off so that in regular
operation, SDS is not present. With SDS full time off, the SDS overpressure
protection is still active.
See Section 3.5.4.2 for instructions on turning SDS full time off if desired.
2.3.10 PARALLEL MEASUREMENT MODE
If your RPM4 has two Q-RPTs of the same type (Axxx, Gxxx, BGxxx), it supports
parallel measurement mode in which both RPTs are used simultaneously to
measure a common pressure, statistically reducing uncertainty in the
measurement. Parallel measurement creates a new Q-RPT, designated <HL>,
which is the combination of the Hi and Lo Q-RPTs. See Section 3.2.4 for
complete information on parallel measurement mode.
Turn on parallel measurement mode if desired.
2.4 POWER-UP AND VERIFICATION
2.4.1 SWITCH POWER ON
Actuate the power switch on the RPM4 rear panel (if a 12 VDC power supply is already connected
to the 12 VDC connection, RPM4 power is already on). Observe the front panel display as
RPM4 initializes, error checks and goes to the MAIN RUN screen (see Section 3.1.1).
RPM4 default power-up condition is Hi Q-RPT active, SDS closed if present.
If the RPM4 fails to reach the MAIN RUN screen, service is required. Record the sequence
of operations and displays observed.
If the RPM4 Q-RPT module is equipped with SDS, SDS is CLOSED at power up and the
TEST(+) port is shut off. This causes <SDS CLOSED> to be flash periodically in the main
run screen in place of the measured pressure value.
If the RPM4 has an Axxx (absolute) Q-RPT, check that it operates properly in
absolute mode.
Make sure that the TEST(+) port is vented to atmosphere.
Use the [RANGE] function key to select the Axxx (absolute) Q-RPT DF range
(see Section 3.3.1).
Press the [MODE] function key and select <absolute> mode (see Section 3.3.3).
Use [UNIT] to change the pressure unit if desired (see Section 3.3.2).
If SDS is CLOSED (<SDS CLOSED> flashes over the display of pressure on
the top line of the display), OPEN SDS. Press [SDS], <2yes> to OPEN SDS
(see Section 3.3.8).
Do NOT OPEN SDS with a pressure higher than the maximum pressure of the
Q-RPT applied to the TEST(+) port. Damage to the Q-RPT may result.
Observe the current value of atmospheric pressure. Check that the value agrees
with the local value of atmospheric pressure within measurement tolerances (see
Section 1.2.2.1) Repeat this process for both Axxx (absolute) Q-RPTs if there
are two in the RPM4. . If a Q-RPT does not agree within tolerance, it may need
to be AutoZeroed (see Section 3.3.9), calibrated (see Section 5.2)
or repaired.
2.4.2.2 CHECKING GAUGE MODE PRESSURE MEASUREMENT
Make sure that the TEST(+) port is open to atmosphere.
Press the [MODE] function key and select <gauge> mode. Use [UNIT] to change
the pressure unit if desired (see Section 3.3.2)..
If SDS is CLOSED (<SDS CLOSED> flashes over the display of pressure on the
top line of the display), OPEN SDS. Press [SDS], <2yes> to OPEN SDS.
Do NOT OPEN SDS with a pressure higher than the maximum pressure of the
Q-RPT applied to the TEST(+) port. Damage to the Q-RPT may result.
The value indicated should be near zero. It is normal for the Q-RPT to indicate
a value other than zero when vented when gauge mode is first entered or the
range is changed. Press [AutoZ]. This runs AutoZ to zero the Q-RPT reading
(see Section 3.3.9.1). Upon return to the main run screen, observe that the
indication of measured pressure has zeroed.
If the display fails to zero properly, RPM4 may need repair.
RPM4 is designed to offer a balance between simple, straight forward operation and the availability of a
wide variety of advanced functions with a high level of operator discretion. The local operator interface is
through a 2 x 20 character alphanumeric display, a function/data keypad, a cursor control pad and a
Ready/Not Ready indicator.
Remote communications are by RS232 (COM1) and IEEE-488. See Section 4 for information on remote
communication.
3.1.1 MAIN RUN SCREEN
The RPM4 MAIN RUN screen is its home display that is reached on power-up and from
which other functions and menus are accessed. It is the very top level of all menu structures.
The MAIN RUN screen is where RPM4 is left in normal operation. It displays the current
measured pressure as well as a variety of additional information if desired.
Figure 3 and its legend summarize the RPM4 MAIN RUN screen fields and their functions.
RPM4 has a screen saver function which causes the display to dim if no key is pressed for
10 minutes. Pressing a key restores full power to the display. The screen saver time can
be changed or screen saving can be completely suppressed (see Section 3.5.5.1).
The RPM4 has a function/data keypad for local operator access to direct
functions, function menus and for data entry.
1. The Function/Data keys allow very commonly
used functions to be accessed directly by a
single keystroke when pressed from the
MAIN RUN screen (see Section 3.1.1).
The name of the function is on the bottom
half of the key. These keys enter numerical
values when editing.
2. The Editing and Execution keys are for
starting and suspending command execution,
cursor control in menus and editing entries.
3. The Menu/Data keys provide access to
function menus when pressed from the
MAIN RUN screen. The menu name is on
the bottom half of the key. The SETUP
menu is for more frequently used functions
(see Section 3.4). The SPECIAL menu is for
functions that are not generally used as a part
of day to day operation (see Section 3.5).
These keys enter numerical values when editing.
Figure 4. Keypad Layout
Pressing the [ENT] key generally causes execution or forward movement in the
menu tree.
Pressing the [ESC] key moves back in the menu tree and/or causes execution to
cease or suspend. Pressing [ESC] repeatedly eventually returns to the MAIN RUN
screen and, from there, allows momentary viewing of the RPM4 introduction
screen.
Pressing the [+/-] key changes a numerical sign when editing. It also toggles
through multiple screens when available and, from some run screens, is a
shortcut to a momentary display of active RANGE.
Pressing the [
], [], [] and [] keys allows up, down, reverse and
forward cursor movement when editing data entry or moving in menus.
Some screens go beyond the two lines provided by the display. This is
indicated by a flashing arrow in the second line of the display. Press the
cursor control keys to move the cursor to access the lines that are not
visible or directly enter the number of the hidden menu choice if you know it.
RPM4™ OPERATION AND MAINTENANCE MANUAL
↑
3.1.3 REMOTE [ENT] (ENTER) FOOTSWITCH
The optional remote ENTER function is a switch that duplicates the function of the front panel
[ENT] key. The remote ENTER function is serviced by a connector on the RPM4 rear panel.
An optional footswitch is available to activate remote entry hands free or a different switch
may be used. See Section 7.1 for information on remote ENTER switch wiring.
The remote ENTER feature can be particularly convenient when running AutoTests
(see Section 3.3.10) in which using a footswitch to [ENT] allows hands free operation. It can
also be used, with the FREEZE display function (see Section 3.3.6.6) to capture the
activation point of a pressure switch.
3.1.4 SOUNDS
RPM4 is equipped with a variable frequency tone device to provide audible feedback and
alarms. The beeper is used for the following indications.
Valid key press
Invalid key press
Leak check completed
Upper or lower limit exceeded
Pmax! (overpressure limit)
exceeded
Parallel measurement mode
possible disconnection of the
Hi and Lo Q-RPTs
AutoTest in/out of tolerance
reading
Brief beep. Choice between three frequencies or NO
sound is available (see Section 3.5.5.2).
Descending two tone “blurp”
Three two second beeps (see Section 3.3.5).
Intermittent one second beeps (see Section 3.4.4).
Eight second high frequency beep (see Section 3.4.4.1).
Rapid beeps for 8 seconds (3.2.4).
Ascending triad/descending triad (see Section 3.3.10).
3.2 GENERAL OPERATING PRINCIPLES
3.2.1 PRESSURE READY/NOT READY
There is a Ready/Not Ready indication LED on the RPM4 front panel. This indication is
intended to provide the user with a clear and objective indication of when a stable pressure
has been achieved. Ready is indicated when the current stability (rate of change) of
pressure is less than the stability limit. The user can set the stability limit (see Section 3.4.3).
The ready indication is often used when comparing the RPM4 and a test device to indicate
when a valid reading can be made.
In RPM4s with two Q-RPTs, the Ready/Not Ready indication always applies to the active
Q-RPT, whose pressure measurement is displayed on the top line of the RPM4 display.
When the inactive Q-RPT measurement is displayed on the second line using the RPT
display mode, the Ready/Not Ready indication for the inactive RPT is indicated by the
leading character on the line. <)> indicates Ready. <
The Ready/Not Ready LED indications are:
<Green >Pressure ReadyThe pressure stability is within the stability limit.
In differential measurement mode (see Section 3.3.3.1), the LED Ready/Not Ready
indication applies to the differential pressures.
In parallel measurement mode (see Section 3.2.4), the LED Ready/Not Ready indication
applies to the average pressure read by the <HL> Q-RPT.
<Red> Pressure NotReadyThe pressure stability is NOT within the stability limit.
When RPM4 is being used as an external measurement reference of a PPC3 pressure
controller, the Ready/Not Ready indication is still based on the RPM4 stability and
stability limit only. The RPM4’s Ready/Not Ready LED does not indicate whether the
PPC3 pressure control is Ready.
3.2.2 GAUGE AND NEGATIVE GAUGE MODES WITH AN AXXX
(ABSOLUTE) Q-RPT, DYNAMIC COMPENSATION FOR
ATMOSPHERIC PRESSURE
Q-RPTs with the designation Axxx are intrinsically absolute but they are also used for gauge and
negative gauge measurement modes (see Section 3.3.3, PRINCIPLE). Gauge measurement
modes are achieved by subtracting the value of atmospheric pressure, P
Q-RPT’s absolute reading using AutoZ (see Section 3.2.2). The AutoZ routine that measures
, is run by pressing [AutoZ] whenever RPM4 is in the vented condition. This assures
P
offset,G
the continuous updating of the P
pressure is the measured absolute pressure, P
value corresponding to atmospheric pressure. Gauge
offset,G
, minus the atmospheric offset.
u
P
gauge
= Pu - P
offset,G
However, atmospheric pressure may change between opportunities to run AutoZ and update
the value of P
, for example when running an extended test without venting. RPM4 uses
offset,G
dynamic compensation for atmospheric pressure to correct for changes in atmospheric
pressure between opportunities to run AutoZ and update P
P
is determined, the reading of RPM4’s on board barometer, P
offset,G
Later, when no longer vented, the change in atmospheric pressure, ΔP
updated, is the difference between the current barometer reading, P
atm
atm,0
= P
:
atm
- P
atm,0
atm
reading at the time of AutoZ execution, P
ΔP
Dynamic compensation for atmospheric pressure uses ΔP
. When AutoZ runs, and
offset,G
, is also recorded.
atm,0
, since P
atm
, and the barometer
atm
to correct the value of P
thus always compensating real time for changes in atmospheric pressure:
offset,G
, from the
was
offset,G
offset,G
,
Gauge pressure measurement on an Axxx (absolute) Q-RPT allows instantaneous switching
between gauge and absolute measurements modes. The additional uncertainty in gauge
pressure mode due to the dynamic compensation for atmospheric pressure technique is a
function of the resolution and short term stability of the on-board barometer, not its absolute
measurement uncertainty. This additional uncertainty is ± 1 Pa (0.00015 psi).
3.2.3 MULTIPE RANGES (Q-RPTS, AUTORANGE AND INFINITE
RANGING)
An RPM4 may have one or two Q-RPTs. Position indication of the currently active Q-RPT is
continuously displayed in the upper right hand corner of the MAIN RUN screen and most other
screens. See Table 3 for position designation protocol for the Q-RPTs in an RPM4.
Table 3. Position Designators of Q-RPTs in an RPM4 System
Q-RPT POSITION DISPLAY SYMBOL*
Q-RPT with highest maximum pressure,
installed in HIGH RPT position
Q-RPT with lower maximum pressure in
dual Q-RPT RPM4, installed in LOW RPT position.
Two Q-RPTs being used simultaneously
in parallel measurement mode
Hi Q-RPT when used in differential
measurement mode
* The display symbol is included in the upper, right hand corner of most RPM4
menu displays as a convenient indicator of the active Q-RPT.
Hi
Lo
HL
Hd
An RPM4 may have multiple ranges. Each RPM4 Q-RPT has a default range which is its
maximum span. Additional ranges, lower than the Q-RPT’s maximum span, may also be
created using AutoRange (see Section 3.3.4). Ranges created using are AutoRange
temporary but may be saved with all their settings for reactivation (see Section 3.4.1).
AN RPM4 range is identified by a range screen showing the Q-RPT used by the range, its current unit
of measure and its full scale pressure in gauge and absolute (if available) measurement modes.
The range screen is:
See Table 1 for a complete listing of RPM4/RPM4 Q-RPT choices and their default ranges.
1. Q-RPT designator.
2. Type of range. DF for the Q-RPT’s default range; AR for a range
created by AutoRange.
3. Q-RPT position designator.
4. Current pressure unit of measure.
5. Full scale pressure in current unit of measure in gauge (<g>) and/or absolute
(<a>) measurement mode depending on type of Q-RPT and range.
Active A700K DF Hi
psi 100g/100a
The ranges available on an RPM4 system are accessed using [RANGE] (see Section 3.3.1)
and/or created using [AutoRange] (see Section 3.3.4).
Most settings made in an RPM4 range, such as unit of measure, measurement mode, display
resolution, and stability setting are specific to the range. Settings selected while one range is
active apply to that range and not to other ranges. The range specific settings are stored with
the range and recalled whenever the range is made active. This makes setting up ranges a
convenient way to store and recall frequently used operating configurations. See Table 4 for a listing
of RPM4 adjustments and settings and whether they are range, Q-RPT or system specific.
[Display] Set bottom line display function System 3.3.6
[Head] Set fluid head correction height, fluid, unit of
[SDS] Open and close SDS Q-RPT 3.3.8
[AutoZ] Run AutoZ Q-RPT and measurement mode 3.3.9
Resolution Set pressure display resolution Range 3.4.2
Stability Set Ready/Not Ready stability test. Range 3.4.3
Upper Limit Set upper and lower pressure limit alarm Range and measurement mode 3.4.4
AutoZ AutoZ, ON/OFF, set and view values Q-RPT and measurement mode 3.5.1
Screen Saver,
Sound, Time,
ID, Level
Lo vnt
Cal Various Q-RPT and barometer calibration
gauge, negative gauge)
measure
Set system user preferences System 3.5.5
Set status of TEST(-) vent valve, if present
functions, including turning off absolute and
negative gauge modes
Range 3.3.3
System 3.3.7
Q-RPT 3.5.7.4
Q-RPT or barometer 3.5.8
3.2.4 PARALLEL MEASUREMENT MODE
RPM4s equipped with two Q-RPT modules of the same type (Axxx, Gxxx, BGxxx) can be set
to operate in parallel measurement mode.
Parallel measurement mode is intended to improve measurement reliability and reduce measurement
uncertainty by using the measurements of two Q-RPTs together. In parallel measurement
mode, the measured pressure is the average of the readings of the two Q-RPTs.
Parallel measurement mode is considered a third Q-RPT designated <HL> (combination of
Hi and Lo). Once it has been set up, the HL Q-RPT can be selected using the [RANGE] key
and operated with all the features of a Hi or Lo Q-RPT. When setting up parallel measurement,
the Hi and Lo Q-RPTs can be suppressed so that only the HL Q-RPT is active. This feature is
used when the RPM4 will always be used in HL mode.
Parallel measurement mode is set up using [SPECIAL], <7internal>, <3RPT2x>
(see Section 3.5.7.3).
3.2.5 USING RPM4 WITH A PPC3 CONTROLLER/CALIBRATOR
RPM4 can be integrated as a reference pressure measuring device in a PPC3 Pressure
Controller/Calibrator System. See the PPC3 Operation and Maintenance Manual for
information on setting up RPM4 to operate as the PPC3’s external device.
Before an RPM4 is used with a PPC3, the following settings should be made:
SDS Close SDS on all Q-RPT modules on which SDS is available (see Section 3.2.7).
Head If the RPM4 TEST port is at a different level than the PPC3’s TEST port AND the
RPM4 will be used in absolute measurement mode, set the RPM4 head height to
the PPC3’s TEST port level so that RPM4 is corrected to indicate the pressure at
the level of the PPC3 TEST port (see Section 3.3.7).
AutoZ Make sure AutoZ is ON for all RPTs and measurement modes (see Section 3.5.1).
Certain RPM4 settings may be changed when an RPM4 is operated with a PPC3. Settings
that may be changed when an RPM4 is used as a PPC3 external device include:
SDSClosed or open status may be changed and SDS full time off will be
overwritten to full time on (see Sections 3.3.8, 3.5.4)
AutoRange range May be rewritten (see Section 3.3.4).
Display When an RPM4 is initialized by PPC3 as an external device, the
second line of the RPM4 display is <PPC3 EXT DEV: IDLE>. When
the RPM4 is the active Q-RPT being used by the PPC3, the second
line of the RPM4 display is <PPC3 EXT DEV: ACTIVE>
Range Active range may be changed (see Section 3.3.1).
Unit of measure Active unit may be changed (see Section 3.3.2).
Measurement mode Active mode may be changed (see Section 3.3.3).
Lo Vent Setting may be changed (see Section 3.5.7.4).
Communications For a PPC3 to be able to identify and communicate with an RPM4,
the RPM4’s COM1 port must be set to:
Baud rate 9200 to 19200
Parity Even
Data bits 7
Stop bit 1
When RPM4 is activated by a a PPC3, the PPC3 sets the RPM4
COM1 port baud rate to 19200. The COM1 port keeps this setting
until power is cycled. Cycling power restores the previous baud rate
setting. The same is true for the COM2 of the first RPM4 in a chain
of two RPM4s used with a PPC3.
When RPM4 is initialized as an external device to a PPC3 pressure controller, the RPM4
front panel keypad is locked out and the second line of the display becomes:
<PPC3 EXT DEV: IDLE> (if RPM4 is initialized by PPC3 but not currently in use)
<PPC3 EXT DEV: ACTIVE> (if RPM4 is initialized by PPC3 and is currently in use)
After using an RPM4 as an external device for a PPC3, cycle power to return to the
previous RPM4 settings, restore front keypad capability and return to the normal main
run screen.
An optional 12VDC battery/charger pack is available to supply power to RPM4s 12VDC
power connection (see 1.2.3).
The battery/charger pack combines the battery and the battery charger.
To Charge the Battery
The battery may be charged while the battery/charge pack is connected to the RPM4.
To charge the battery, connect the battery/charger pack to a 100 to 240 V, 50/60 Hz power
supply using the supplied power cable and turn on the charger power switch. The green
CHARGE indicator light goes on to indicate that battery charging is in progress. Allow 14 to
16 hours for the battery to charge fully. When the battery is fully charged, the CHARGE
indicator light flashes.
When the battery charger power switch is in the ON, the battery is connected to the charger,
not the 12VDC output. When the battery charger switch is in the OFF position, the CHARGE
indicator light goes off, and 12VDC is supplied on the 12VDC connection.
To Use the Battery/Charger Pack to Supply RPM4
After charging the battery, connect the 12VDC connection of the battery/charger pack to the
12VDC connection on the RPM4 rear panel.
Turn the battery/charger pack power switch to OFF (when AC power is connected to the
battery/charger pack, the power switch must be OFF for 12VDC power to be output).
When RPM4 is connected to both a 12 VDC source and an AC source,if the RPM4 power
switch is OFF, the 12VDC source supplies the RPM4; if the RPM4 power switch is ON, the
12 VDC source is disconnected and the AC source supplies the RPM4.
To Check the Battery Charge Level
To check the current charge level of the battery, press the toggle switch on the
battery/charger pack front panel. Observe the BATTERY CHECK indicator lights. All three
lights are lit when battery is fully charged. If no light or one light is on, the battery should be
recharged.
Q-RPT modules designated 7M or lower that are NOT liquid filled are equipped with the SDS self
defense system. The SDS system includes hardware and embedded software to isolate Q-RPTs
from overpressure. When SDS is CLOSED for a Q-RPT module, the Q-RPT is isolated from the
modules TEST(+) port and opened to the modules VENT port (See Section 5.6, Figure 8).
SDS can be OPENED and CLOSED from the front panel using [SDS]. SDS can also be
controlled using functions under [SPECIAL], <4SDS>. The SDS “temp open/close” function
allows SDS on both Q-RPTs, whether active or inactive, to be opened and closed directly
(see Section 3.5.4.1). The SDS “full time on/off” function allows SDS to be opened permanently
so the RPM4 can be operated as if SDS were not installed (see Section 3.5.4.2).
SDS is normally used to isolate an RPM4 Q-RPT from a test system which may sometimes
be subjected to pressures higher than the Q-RPTs maximum pressure. When configuring a
multi-range system, this can eliminate the need for external valving or connecting and
disconnecting RPM4s. For example, two Q-RPTs of different ranges might be connected to
a single calibration system. When the Hi Q-RPT is in use, SDS is closed on the Lo Q-RPT to
protect it from overpressure. When used in this manner, SDS should be CLOSED prior to
applying pressure. Automatic SDS closing in response to overpressure should only be used in
an emergency overpressure situation.
SDS also attempts to protect Q-RPTs from accidental overpressure. Whenever RPM4 is
powered ON it continuously monitors the pressure read by its Q-RPT(s). If the pressure
reaches the maximum pressure limit for a Q-RPT, SDS CLOSES to isolate the Q-RPT from
the TEST port and vent it to atmosphere (see Section 3.4.4.1). This is not a fail safe
overpressure system. It does not protect against very rapid overpressure or pressure
exceeding 10 MPa (1 500 psi).
When an RPM4 is used as an external device to a PPC3 pressure controller, PPC3 controls the
RPM4’s SDS to isolate the RPM4 Q-RPTs from the test pressure when they are not in use.
IF the RPM4 SDS is “full time off” PPC3 changes it to “ON”.
The maximum pressure that should be applied to an RPM4 TEST(+) port when SDS is
CLOSED is 10 MPa (1 500 psi).
Though the SDS self defense system includes features to automatically protect Q-RPTs
against accidental overpressure, SDS should NOT be considered a fail-safe overpressure
protection system. SDS cannot guarantee that overpressure damage will NOT occur.
Conventional measures for overpressure protection should also always be used. Damage
to Q-RPTs due to overpressure is NOT covered by the RPM4 product warranty even when
SDS is present and ON.
Table 5 provides a brief summary of direct function key operation. It may be useful to keep
a copy of this summary near the RPM4, especially when first becoming acquainted with
its operation.
Local operation of RPM4 is through the front panel keypad. To minimize multi-layered menu
structures, the keypad numerical keys also provide direct access to the most commonly used
functions. The function accessed is labeled on the bottom half of the key. Direct function
keys are active whenever RPM4 is in its MAIN RUN screen. Table 5 summarizes the
operation of the direct function keys.
Table 5. Summary of RPM4 Function Key Operation
DIRECT FUNCTION KEYS ARE ACTIVE FROM THE MAIN RUN SCREEN
See corresponding manual sections for full detail.
Adjust height of fluid head
correction. Set to zero to defeat
correction.
.SDS controls the connection of
the Q-RPT to the TEST(+) port.
Close any open SDS. Open SDS
of active Q-RPT if closed.
Run the AutoZ function to
AutoZero the active Q-RPT and
measurement mode.
Set up and optimize RPM4
features for a specific operating
range and measurement mode.
Run automated leak checking
routine.
View active range and toggle
through available ranges. [ENT] on
a range activates it.
Select pressure unit of measure for
the active range. Choice of units
available in this menu can be
customized.
Select the pressure measurement
mode for the active range (absolute,
gauge, negative gauge, differential).
Menu of less commonly used internal
functions and settings.
Menu of commonly used setup
features including save/delete
AutoRange ranges, set display
resolution, set stability limit, set
upper/lower, view AutoTest data
Select the display function for the
bottom line of the RPM4 display.
3.3 DIRECT FUNCTION KEYS
3.3.1 [RANGE]
PURPOSE
To view and/or change the active pressure measurement range and associated settings.
PRINCIPLE
An RPM4 range is defined as a pressure span and associated settings (see Section 3.2.3).
An RPM4 may have multiple ranges. Each Q-RPT in the RPM4 system has a default range
which is its maximum span. Additional ranges, lower than the Q-RPT’s maximum span, may
also be created using AutoRange (see Section 3.3.4). Ranges created using AutoRange
may be saved, with their settings, for reactivation (see Section 3.4.1).
The [RANGE] function key allows the available RPM4 pressure ranges to be viewed and selected.
Start or set functions such as Leak
Check, AutoTest and Freeze.
ENTER values when editing.
RPM4™ OPERATION AND MAINTENANCE MANUAL
OPERATION
Pressing the [RANGE] function key activates the range viewing and selecting function.
When the [RANGE] function key is first pressed, the active range is displayed. For example:
1. Q-RPT designator.
2. Type of range. DF for the Q-RPT’s default range; AR for a range
created by AutoRange.
3. Q-RPT position designator.
4. Current pressure unit of measure.
5. Full scale pressure in current unit of measure in gauge and/or
absolute measurement mode depending on type of Q-RPT and range.
Pressing the [+/-] key or the [RANGE] key again while in the RANGE functions toggles
through displays of the other available ranges, from lowest range to highest range.
To select a range to become the active range, press [ENT] while the range is displayed.
Pressing [ESC] while in the RANGE function returns to the MAIN RUN screen without
making a range change.
Range full scale limits are given in the pressure unit that is currently active for that range.
Active A700K DF Hi
psi 100g/100a
Pressing [+/-] from the MAIN RUN screen provides a shortcut to a momentary view of the
active range.
Many RPM4 settings and functions are range specific. See Table 4 for identification of
range specific settings.
For best metrological performance, vent the RPM4 Q-RPT before changing ranges.
3.3.2 [UNIT]
PURPOSE
To select the unit of measure in which RPM4 displays pressure values.
For information on selecting the measurement mode (absolute, gauge, negative gauge,
differential), see Section 3.3.3.
PRINCIPLE
RPM4 allows the unit of measure in which pressure values are displayed to be changed.
RPM4 supports 15 standard pressure units of measure as well as user defined units.
To simplify operation, quick access to six units is made available under the [UNIT] key.
The default units available depend on whether the RPM4 was originally configured as an SI
or US version. The [UNIT] key contents can be customized by the user to any configuration
of six units (see Section 3.5.6).
Internally, RPM4 always operates in Pascal (Pa), the SI unit of pressure. Values of pressure
are represented in other units by the application of conversion factors to convert from Pa
(see Section 7.2.1).
OPERATION
To change the active pressure unit of measure for a
range, press [UNIT] from the MAIN RUN screen while
the range is active. The display is:
1kPa 2Pa 3MPa
4hPa 5bar 6mbar
The cursor is on the number corresponding to the active pressure unit of measure for the active range.
To change the pressure unit of measure, select the desired unit. Making the selection
returns to the MAIN RUN screen with the selected unit active.
The pressure unit of measure selected is range specific. When in a given range, all functions
and settings are represented in the current measurement unit for that range. However,
certain internal and/or metrological functions (e.g., Q-RPT calibration coefficients) are always
represented in Pa regardless of the current range unit. When the current unit is an altitude
unit, the range and upper limit indications are in kPa if the unit is meters (m) and psi if
the unit is feet (ft).
See Section 7.2.1 for tables of the conversion factors used by RPM4.
If the pressure unit selected is inWa (inches of water), the reference temperature for water
density must we specified in a separate menu (choices are 4ºC, 20ºC 60ºF). No reference
temperature selection is necessary for the unit mmWa as the only reference temperature
commonly used for mmWa is 4 ºC.
The default pressure units of measure available under the UNIT function depend on whether
the RPM4 has been set up as an SI or US version (indicated by SI or US at bottom right of
introduction screen). The choice of six units available under the UNIT function can be
customized from a wider selection by the user (see Section 3.5.6). The units available under
the UNIT function can be reset to default by reset (see Section 3.5.9.2).
3.3.3 [MODE]
PURPOSE
To set the measurement mode (absolute, gauge, negative gauge or differential) for the active range.
For information on selecting the unit of measure, see Section 3.3.2.
PRINCIPLE
RPM4 supports simple, one-step switching between up to four different measurement modes:
Absolute Measures pressure relative to vacuum (zero is hard vacuum). Range is
Gauge Measures pressure relative to atmosphere (zero is ambient pressure).
Range is from zero gauge to full scale, full scale must be greater than zero.
Negative Gauge Measures pressure relative to atmosphere (zero is ambient pressure).
Range is positive and negative from zero, negative to minus one
atmosphere and positive to full scale. With a BGxxx RPT, negative range
is the negative equivalent of the positive range.
DifferentialAvailable only for the Hi Q-RPT in RPM4s equipped with two Q-RPT modules
of the same type (Axxx, Gxxx, BGxxx). Measures the difference between the
Hi and the Lo Q-RPT (Differential = Hi – Lo). Differential measurement mode
is selected “on top of” the Hi Q-RPT’s base measurement mode (gauge,
absolute or differential). Differential range is positive to maximum pressure of
the Hi Q-RPT and negative to the minimum pressure of the Lo Q-RPT base
measurement mode (see Section 3.3.3.1 for specific information on operation
in differential measurement mode).
Which measurement modes are available depends on the type of Q-RPT type(s) available.
There are three types of Q-RPTs:
AxxxQ-RPTs with designators starting with “A” measure pressure relative to a sealed
vacuum reference and are intrinsically absolute. Axxx Q-RPTs support
measurement in absolute, gauge and negative gauge modes. Gauge and negative
gauge modes are accomplished by offsetting of atmospheric pressure with
dynamic compensation for atmospheric pressure changes between offsets using
an on-board barometer (see Section 3.2.2). Access to absolute and negative
gauge operation can be turned OFF (see Section 5.2.5). In RPM4s with two Axxx
Q-RPTs, differential measurement mode is supported on the Hi Q-RPT.
Gxxx Q-RPTs with designators starting with “G” measure positive pressure relative to the
TEST(-) port that is connected to atmosphere. Gxxx Q-RPTs support measurement
in gauge mode only. In RPM4s with two Gxxx Q-RPTs, differential measurement
mode is supported on the Hi Q-RPT.
BGxxx Q-RPTs with designators starting with “BG” measure positive and negative
pressure relative to the TEST(-) port that is connected to atmosphere. BGxxx Q-RPTs
support measurement in gauge and negative gauge modes. In RPM4s with two
BGxxx Q-RPTs, differential measurement mode is supported on the Hi Q-RPT.
When changing modes for a range, if the full scale of the range is 700 kPa (100 psi) or
greater, the full scale is the same in both absolute and gauge modes. If the full scale is less
than 700 kPa (100 psi), the gauge mode full scale is 100 kPa (14.5 psi) lower than the
absolute mode full scale. If the absolute mode full scale is less than 100 kPa (14.5 psi), there
is no gauge mode available. Gauge and negative gauge modes, when available, always
have the same full scale.
Measurement mode selection is range specific.
OPERATION
See Section 3.3.3.1 for information on differential mode operation. Differential mode
operation is only available for the Hi Q-RPT in RPM4s that are equipped with two Q-RPTs
of the same type (Axxx, Gxxx, BGxxx).
To change the active measurement mode for a range,
press [MODE] from the MAIN RUN screen while the
range is active. The display depends upon the type of
Q-RPT that is active.
If an Axxx Q-RPT is active, all thee measurement modes are supported: absolute, gauge
and negative gauge. Only gauge is available if absolute and negative gauge modes have
been turned OFF in calibration (see Section 5.2.5).
If a Gxxx Q-RPT is active, only gauge measurement mode is supported. When the [MODE]
key is pressed, <Gauge RPT, gauge mode only> is displayed.
If the Hi Q-RPT is active in its default range (<DF>), and the RPM4 has two Q-RPTs of the
same type (Axxx, Gxxx, BGxxx), differential measurement mode is available in addition to the
other modes (see Section 3.3.3.1).
Certain RPM4 settings, including AutoZ ON/OFF (see Section 3.5.1) are range AND measurement
mode specific. See Table 4 for a listing of settings and what they are specific to.
Absolute and negative gauge modes on an Axxx Q-RPT can be turned OFF in the
calibration function (see Section 5.2.5). When absolute and negative gauge mode are
turned OFF for an Axxx Q-RPT, only gauge mode can be activated. <Gauge mode only,
other modes OFF> is displayed when [MODE] is pressed.
3.3.3.1 DIFFERENTIAL MEASUREMENT MODE OPERATION
Differential mode operation is only available on the Hi Q-RPT, default
(<DF>) range, in RPM4s that are equipped with two Q-RPTs (Hi and Lo) of
the same type (Axxx, Gxxx, BGxxx).
To activate differential measurement mode, press [MODE] from the MAIN RUN
screen while the Hi Q-RPT is active and in its default (<DF>) range. The actual
display depends upon the measurement modes available on the Hi Q-RPT.
The cursor is on the base measurement
mode. The line pressure measurement mode
is the base pressure measurement mode.
To activate differential measurement mode, press the keypad number key for the
<dif> selection or move the cursor to the <dif> selection and press [ENT].
Operation returns to the main run screen. When differential mode is active,
the cursor is on both the base mode and differential mode.
When differential mode is active, the Q-RPT position indicator is in the upper
right hand corner of the main run screen and other displays is <Hd> instead
of <Hi> (see Section 3.2.3, Table 3).
Meas mode: 1abs
2gage 3neg gage 4dif
When differential mode is activated, AutoZ should be run at the line pressure
with zero differential pressure applied to eliminate the zero error on differential
pressure due to the disagreement of the Hi and Lo Q-RPTs (see Section 3.3.9.3).
RPM4™ OPERATION AND MAINTENANCE MANUAL
The following operating conditions are specific to differential mode:
Main run screen measured pressure display
Default DISPLAY function The default display function is RPT with the Lo Q-RPT
SDS function [SDS] key operation to OPEN SDS opens SDS of
AutoZ function AutoZ manages zeroing of differential mode and is
The display of measured pressure in the main run
screen is the difference between the Hi and the Lo
Q-RPTs (Hi – Lo) with a differential mode AutoZ
correction (see Section 3.3.9.3). The pressure
measurement mode indicator is <d>. The Q-RPT
position indicator is <Hd>.
displayed on the bottom display line (see Section
3.3.6.4). This allows the line pressure (Lo Q-RPT
indication) to be viewed simultaneously with the
differential pressure. All other DISPLAY functions
(average, rate, deviation, HiLo, freeze) apply to
the differential pressure
both the Hi and Lo Q-RPTs simultaneously. (see
Section 3.3.8). Overpressure is driven by the pressure
range of the Lo Q-RPT (see Section 3.4.4.1).
intended to “tare” the two Q-RPTs at a common
line pressure (see Section 3.3.9.3). Switch to a
mode other than differential for AutoZ functions of
an individual Q-RPT.
Leak Check function The leak check function uses the differential mode
Head function The head function applies to the differential
AutoRange function Differential measurement mode is not available in
UL function The UL limits apply to positive and negative
AutoTest function The AutoTest function (see Section 3.3.10) is not
3.3.4[AUTORANGE]
PURPOSE
measurement (see Section 3.3.5).
pressure (see Section 3.3.7).
AutoRange (see Section 3.3.4).
differential pressure (see Section 3.4.4) and the
default is the maximum value of the differential
pressure. Both the Lo and Hi RPT base
measurement mode ULs also remain active. The
Hi RPT UL provides a upper limit for the maximum
pressure on the Hi side of the differential pressure.
Both the differential and base mode Hi RPT UL
cause the display of differential pressure to flash
when exceeded. The Lo RPT UL provides an
upper limit for the line pressure.
available in differential measurement mode.
To automatically set up RPM4 to optimize its measurement characteristics and features for
specific, user specified range of operation.
PRINCIPLE
RPM4 is designed to support the calibration and test of a very wide variety of test ranges in
various measurement modes.
The AutoRange function simplifies the task of selecting the Q-RPT best suited to cover a
specific test range and setting measurement parameters appropriately for the range. These are
set automatically based on operator entry of desired measurement mode, pressure unit of measure
and range full scale. The selections and settings made by AutoRange are summarized in Table 6.
A range set up using AutoRange and all its range specific settings, may be saved for recall
using [SETUP], <1range> (see Section 3.4.1). Saved ranges are available for later selection
using [RANGE] (see Section 3.3.1). If a range created by AutoRange is not saved, it is
overwritten by the next AutoRange or erased when another range is selected.
Press [ENT] from the main run screen in RATE mode to use the [AutoTest] function to
AutoRange RPM4 based on DUT tolerance (see Section 3.3.10).
Table 6. Settings Made by AutoRange
SETTING/SELECTION PURPOSE SET TO SECTION
Unit Set pressure unit of measure Operator specified during
Mode Set pressure measurement mode
Full scale Establish span maximum Operator specified during
AutoRange span Basis of span based setting
Q-RPT Select most appropriate Q-RPT
Upper limit
Lower limit
Resolution Set display resolution 10 ppm of AutoRange span or
Stability Limit Set Ready/Not Ready stability
Q-RPT turn down
(Premium class Q-RPTs only)
(absolute, gauge, negative gauge)
calculations
to cover range and mode.
Set maximum and minimum
pressure alarm limits.
criterion.
Turn down Q-RPT
characterization to improve
measurement uncertainty when
span is reduced
AutoRange
Operator specified during
AutoRange
AutoRange
Absolute mode: Zero absolute
to AutoRange full scale
Gauge mode: Zero gauge to
AutoRange full scale
Negative gauge mode: Minus
atmosphere to AutoRange full
scale (negative equivalent of full
scale on BG15K Q-RPT).
Absolute and negative gauge
mode: Axxx Q-PRT with lowest
default full scale that is greater
than the AutoRange full scale
Gauge mode: Q-RPT with
lowest span whose full scale is
greater than AutoRange full
scale, Gxxx over Axxx if spans
are equal.
Upper limit: 105 % of operator
entered full scale pressure
Lower limit: None in absolute
and gauge modes. Minus
atmosphere in negative gauge
mode, minus equivalent of upper
limit in negative gauge mode on
a BG15K Q-RPT. Minus 66 kPa
(9.6 psi) with BA100K.
1 ppm of default Q-RPT span,
whichever is larger.
50 ppm of AutoRange span or 2
ppm of default Q-RPT span,
whichever is larger.
To use the AutoRange function, press [AutoRange] and respond to the RPM4’s prompts.
Select AutoRange measurement mode: The screen is identical in appearance and
function to the [MODE] screen (see Section 3.3.3). The choice of measurement modes
offered reflects all the Q-RPT types available (Axxx, Gxxx, BGxxx) in the RPM4 system,
not just the active Q-RPT.
Differential pressure measurement mode (see Section 3.3.3.1) is not available in
AutoRange.
Select pressure unit of measure: The screen is identical in appearance and function
to the [UNIT] screen (see Section 3.3.2).
Altitude units of measure (m, ft) are not available for AutoRange.
Enter the AutoRange full scale pressure: Enter the full scale pressure value in the
<------> entry field and press [ENT]. If the full scale value entered cannot be covered by
any of the Q-RPTs available, an error message appears. If the full scale value can be
covered, AutoRange proceeds to the range display.
View proposed AutoRange range: The proposed AutoRange range screen is:
1. Indication that this is a screen of the proposed AutoRange range.
2. Designator and position of the Q-RPT that AutoRange has selected
to cover the AutoRange full scale and measurement mode.
3. Pressure unit of measure specified in AutoRange.
4. AutoRange full scale pressure in gauge mode and absolute
mode if available. Absolute mode only if AutoRange measurement
mode is absolute and full scale is less than 100 kPa (14.5 psi).
Gauge mode only if the Q-RPT is a Gxxx or BGxxx or an Axxx
with absolute mode turned off.
AutoRange A700K Hi
kPa 300g/100a
If the RPM4 has two Q-RPTs and you would like to select the other one for the
AutoRange, the cursor control keys may be used to toggle between Q-RPTs selections if
both Q-RPTs can handle the specified AutoRange.
Accept proposed AutoRange: To accept the proposed AutoRange and go to the MAIN
RUN screen with the AutoRange range active, press [ENT] or [AutoRange]. To modify
the AutoRange full scale, measurement mode or pressure unit of measure, use [ESC] to
back through the AutoRange screens and make changes.
To verify or confirm the range from the MAIN RUN screen, use [+/-] or [RANGE] to
view the active range (see Section 3.3.1).
See Table 6 for a listing of all selections and settings affected by AutoRange and
To run an automated leak check routine that determines the leak rate of the system
connected to the RPM4 active Q-RPT TEST(+) port.
PRINCIPLE
The LEAK CHECK function is provided as a means of checking and quantifying the leaks that
may be present in the system connected to the RPM4’s active Q-RPT test port.
The principle of the LEAK CHECK function is the measurement of the natural decrease or
increase of pressure in a fixed volume over time. The LEAK CHECK function allows a leak
check time to be set. The total pressure change and the average rate of change over the
leak check time are calculated and displayed.
In differential pressure measurement mode (see Section 3.3.3.1), the leak check
measurement is based on the measurement of the differential pressure.
OPERATION
Changing the pressure in a test system causes adiabatic temperature changes in the
pressurized medium that need to have dissipated before a valid leak measurement can be
made. In general, a 0.5 to 1 minute wait before running a leak check is adequate to allow
the adiabatic temperature change to dissipate and valid leak measurements to be made.
However, stabilization time may be much longer with liquid test media, and as volume and
pressure increases.
To run a leak check, first set the pressure to the desired leak check pressure.
To access the LEAK CHECK function, press [LEAK CHECK] from the MAIN RUN screen.
Select <1run> to run a leak test. The display is:
Set leak check time:
1. Edit field for the time over which the leak rate will be
determined, in seconds.
15 s
Edit the leak check time if desired (minimum 1, maximum 999 seconds) and press [ENT].
Press [ENT] again when ready to start the leak test. The leak test display is:
1. Standard MAIN RUN screen first line showing measured
pressure.
2. Active Q-RPT position indicator
3. Indication that leak test is running and countdown of time
remaining.
[ESC] can be used to abort the running leak test. [ENT] while the leak test is running
restarts the leak check timer. When the leak check timer countdown has completed,
RPM4 beeps three times and the leak check results screen is displayed:
1. Total pressure change from start to finish of leak check time.
2. Indicator of Q-RPT used to run the leak check.
3. Average rate of change of pressure over the leak check
time period (∆P/time in seconds).
∆P 0.61 kPa a Hi
Rate 0.06 kPa/sec
From the leak check results screen, press [ENT] to repeat the leak test.
Press [ESC] to return to leak check main menu and exit to the MAIN RUN screen.
To view the results of the most recently completed leak check, press <2view>. If NO leak
check data is stored (i.e., if the RPM4 has never run a leak test or a reset has cleared
previous leak test results), the results screen displays <Data NOT available> briefly and
returns to MAIN RUN screen. Press [ENT] or [ESC] to return to the MAIN RUN screen.
Leak check is range specific in the sense that a leak check is run using the active range.
However, only one set of leak check results is maintained in memory and each leak test
completed overwrites the memory. View leak check always shows the results of the last
leak check run regardless of the range that is now active. The results screen includes the
range indicator to indicate the range in which the leak check was run.
3.3.6 [DISPLAY]
PURPOSE
To select, from a variety of choices, the information that is displayed on the second line of the
RPM4 display.
PRINCIPLE
RPM4 supports a variety of advanced pressure measurement functions that are displayed on
the second (bottom) line of the RPM4 display. In summary, the available display functions
included are:
AVERAGE Calculates the average pressure measurement over a user specified period
of time and displays the average, the standard deviation about the mean and
a countdown in seconds to the next average (see Section 3.3.6.1). This function
is often used to filter out pressure noise in an unstable system. The magnitude
of the noise is quantified by the standard deviation about the mean. A second
Avg screen allows the instantaneous pressure values to be viewed during an
averaging cycle.
RATECalculates and displays the current rate of change of pressure in current
pressure units/second (see Section 3.3.6.2). This function is a useful
indication of the stability of the pressure being measured. It is often used as
an indication of positive or negative leak rate and as a go/no go criterion of
when to take data when comparing RPM4 and a device under test, for
example in a calibration. Rate is used by the Ready/Not Ready function to
determine when a Ready condition exists (see Section 3.2.1).
DEVIATION Continuously calculates and displays the difference between the pressure
measured by RPM4 and a target pressure entered by the user (see Section
3.3.6.3). This function is useful in monitoring the evolution of pressure around
and/or away from a desired set point.
RPT If the RPM4 has two Q-PRTs, allows pressure measurement from the two
RPTs to be displayed simultaneously (see Section 3.3.6.4). This function is
particularly useful in dual Q-RPT RPM4s with independent TEST ports to
monitor two separate pressures simultaneously. It is also used to when the
Hi Q-RPT is in differential measurement mode, to observe the line pressure
measured by the Lo Q-RPT.
HI/LORecords and displays maximum and minimum pressures measured (see
Section 3.3.6.5). This function is used to keep track of the minimum and
maximum pressure observed in a system over a period of time or to monitor
if a pressure min/max limit has been exceeded..
FREEZECaptures and displays the pressure measured by the active range of RPM4
when the [ENT] key is pressed (see Section 3.3.6.6). This function is useful
to record the pressure present at the time of an operator observed trigger
event, for example when the needle of an analog gauge was on the nominal
point or when a switch activates.
CLEANBlanks out the second line of the display (see Section 3.3.6.7). This function
is used when a simple display of pressure measured by the RPM4 active
range without additional information is desired.
OPERATION
To set the DISPLAY function press [DISPLAY] from the main run screen.
The display is:
1avg 2rate 3dev 4RPT
5HiLo 6freeze 7clean
The cursor is on the active DISPLAY function. Selecting a display function returns to the
main run screen with the selected function active.
Display menu choices include:
<1avg>Display average pressure and standard deviation. Specify averaging time period
(see Section 3.3.6.1).
<2rate> Display pressure rate of change (see Section 3.3.6.2).
<3dev> Display deviation of the measured pressure from a target value and specify
the target value (see Section 3.3.6.3).
<4RPT>Display the pressure measured by the second Q-RPT in RPM4s with two
Q-RPTs (see Section 3.3.6.4).
<5HiLo>Display the maximum and minimum pressure observed since HiLo reset
(see Section 3.3.6.5).
<6freeze>Capture and display the value measured when [ENT] is pressed
(see Section 3.3.6.6).
<7clean> Leave the second line of the display blank (see Section 3.3.6.7).
The DISPLAY selection is NOT range specific. A DISPLAY selection made in one range
applies to all ranges.
The default DISPLAY function is Rate which causes the second line of the display to show “R”
followed by the current rate of change of pressure in current pressure units per second.
The default DISPLAY function in differential mode is RPT.
When RPM4 is initialized as an external device to a PPC3 pressure controller, the RPM4
front panel keypad is locked out and the second line of the display becomes:
<PPC3 EXT DEV: IDLE> (if RPM4 is initialized by PPC3 but not currently in use)
<PPC3 EXT DEV: ACTIVE> (if RPM4 is initialized by PPC3 and is currently in use)
After using an RPM4 as an external device for a PPC3, cycle power to return to the
previous RPM4 settings, restore front keypad capability and return to the normal main
run screen.
3.3.6.1 AVG (AVERAGE)
PURPOSE
To activate the Average DISPLAY and/or adjust the period of time over which
averaging occurs.
See Section 3.3.6. Principle.
OPERATION
To access the Average DISPLAY, press [DISPLAY], <1Avg>. The display is:
Averaging Period:
1. Edit field for averaging period in seconds.
Default is 20. Minimum 1, maximum 999.
20 s
Edit the averaging time period if desired. Pressing [ENT] returns to the main run
screen with the Average DISPLAY active.
With the Average DISPLAY active the main run screen is:
1. Average measured over last completed
averaging period.
2. Standard deviation of last completed
averaging period.
3. Countdown in seconds until completion of ongoing averaging period.
96.929 kPa a Hi
δ 0.0007 18Avg
The Average DISPLAY has a second screen that allows the instantaneous
pressure readings to be viewed while an averaging cycle is running.
The instantaneous Average screen is:
1. Instantaneous pressure values at RPM4’s
normal update rate.
2. Current rate of change of pressure in pressure
units/second.
3. Countdown in seconds until completion of ongoing averaging period.
96.929 kPa a Hi
R 0.0025 18Avg
The [+/-] key toggles between the main run Average screen and the
instantaneous values Average screen.
Pressing [ENT] while in the Average DISPLAY aborts the current averaging
period and causes a new one to begin. [ENT] can thus be use to trigger a
new averaging period on demand.
In the Average DISPLAY the Ready/Not Ready indication applies to the result
of the previous averaging period (see Section 3.2.1). <Ready> indicates that
all readings during the previous averaging period met the stability criterion.
<Not Ready> indicates that one or more readings were outside of the
stability criterion.
Changing the pressure unit of measure, measurement mode or range while
the averaging screen is active, starts a new averaging period.
To go to a DISPLAY other than Average, press [DISPLAY] and make a new
DISPLAY choice (see Section 3.3.6).
3.3.6.2 Rate
PURPOSE
To activate the Rate DISPLAY.
See Section 3.3.6 [DISPLAY], Principle.
The AutoTest function (see Section 3.3.10) to AutoRange RPM4 based on
DUT characteristics and run an automated test sequence is accessed by
pressing [ENT] from the Rate run screen.
OPERATION
To activate the Rate DISPLAY press [DISPLAY], <2Rate>. Pressing <2Rate>
returns to the main run screen with the Rate DISPLAY active.
With the Rate DISPLAY active the main run screen is:
99.1135 psi a Hi
1. Current rate of change of pressure in current
pressure units per second.
R 0.0001/sec
The Rate DISPLAY is different and separate from the stability setting which
is used to set the stability criterion on which the Ready/Not Ready
indication is based (see Sections 3.4.3 and 3.2.1). The Rate DISPLAY only
causes the current rate of change to be displayed and has NO affect on the
stability setting or the Ready/Not Ready condition.
RPM4™ OPERATION AND MAINTENANCE MANUAL
The Rate DISPLAY is the default display except in differential measurement
mode in which the default DISPLAY is RPT.
To go to a DISPLAY other than Rate, press [DISPLAY] and make a new
DISPLAY choice (see Section 3.3.6).
3.3.6.3 Dev (Deviation)
PURPOSE
See Section 3.3.6 [DISPLAY], Principle.
To activate the Deviation DISPLAY and/or set the deviation target value.
OPERATION
To activate the Deviation DISPLAY press
[DISPLAY], <3Dev>. The display is:
Target:
100.0000 psi a
Edit the desired target value. Pressing [ENT] returns to the main run screen with
the Deviation DISPLAY active using the entered target value.
The target value is the value from which deviations (D) are measured by the
Deviation DISPLAY following:
D = current pressure - target pressure
With the Deviation DISPLAY active the main run screen is:
1. Deviation of current pressure from the target
value.
2. Target value.
99.1135 psi a zH3
D -0.8865 T100.0000
Pressing [ENT] from the main run screen when the Deviation DISPLAY is
If the pressure measurement unit or mode is changed while the Deviation
To go to a DISPLAY other than Deviation, press [DISPLAY] and make a new
active goes directly to the Target editing screen. This allows the target
value to be changed without going through the DISPLAY menu.
DISPLAY is active the target value remains at the same numerical value.
It is NOT converted.
DISPLAY choice (see Section 3.3.6).
3. OPERATION
3.3.6.4RPT
PURPOSE
See Section 3.3.6, Principle.
To activate the RPT DISPLAY.
The RPT DISPLAY is only available in RPM4s with two Q-RPTs.
OPERATION
For the sake of clarity, when describing the RPT DISPLAY, the Q-RPT
displayed on the top line of the display when the RPT function is selected is
referred to as the “active” Q-RPT. The other Q-RPT to be displayed on the
second (bottom) line of the RPM4 display is referred to as the “inactive”
Q-RPT. It is “inactive” in the sense that all RPM4 functions and settings
such as UNIT and RES still apply to the “active” Q-RPT. To make changes to
the “inactive” Q-RPT, for example to change its measurement units, it must
be made the active Q-RPT by a conventional range change (see Section 3.3.1).
In differential pressure measurement mode, the Hi Q-RPT is always the
“active” Q-RPT.
To activate the RPT DISPLAY, press [DISPLAY], <4RPT>. Operation returns to
the MAIN RUN screen with the inactive Q-RPT displayed on the display’s bottom
line. The range of the inactive Q-RPT is always the default range, not an
AutoRanged range (see Section 3.3.4).
If the RPM4 has only one Q-RPT, <Not available with single RPT> displays
momentarily and operation returns to the main run screen when [DISPLAY],
<4RPT> is pressed.
If the RPM4 is in parallel (<HL>) (see Section 3.5.7.3) mode in which both
Q-RPTs are used simultaneously, when <4RPT> is selected, the Q-RPT to display
on the lower display line must be specified. The Q-RPT specific settings of the
Q-RPT displayed in the lower display line (measurement mode, unit of measure,
resolution, etc.) can only be changed by making that Q-RPT active.
With the RPT DISPLAY active the main run screen is:
1. Active RPT display.
2. Inactive RPT display with Ready/Not Ready
indicator on far left.
If the inactive RPT is equipped with SDS Self Defense System (see Section
3.2.7), SDS will almost certainly be CLOSED when the RPT DISPLAY is activated.
SDS can be OPENED on the inactive RPT using [SPECIAL], <4SDS> (see Section
3.5.4). In differential or parallel measurement mode, SDS can be opened
simultaneously on both Q-RPTs using [SDS] (see Section 3.3.8).
The RPT DISPLAY is the default display in differential measurement mode.
When the RPT display is active, executing a range change to a range on the
inactive Q-RPT makes the inactive Q-RPT the active Q-RPT. The DISPLAY
defaults back to Rate ( Section 3.3.6.2).
To go to a DISPLAY other than RPT, press [DISPLAY] and make a new
DISPLAY choice (see Section 3.3.6).
3.3.6.5 Hi/Lo
PURPOSE
To activate the Hi/Lo DISPLAY.
See Section 3.3.6, Principle..
OPERATION
To activate the Hi/Lo DISPLAY press [DISPLAY], <5Hi/Lo>. Pressing <5Hi/Lo>
resets the Hi/Lo values and returns to the main run screen with the Hi/Lo
DISPLAY active.
With the Hi/Lo DISPLAY active the main run screen is:
99.1135 psi a zHi
1. Highest pressure observed since Hi/Lo reset.
2. Lowest pressure observed since Hi/Lo reset.
H 99.2254 L98.0098
The Hi/Lo values change each time a new Hi or Lo pressure is observed.
The Hi/Lo record can be reset at any time by pressing [ENT]. This allows a
Hi/Lo reset without going back through the DISPLAY menu.
If the pressure measurement unit, mode or range is changed while the Hi/Lo
To go to a DISPLAY other than Hi/Lo, press [DISPLAY] and make a new
To activate the Clean DISPLAY press [DISPLAY], <7Clean>. Pressing <7Clean>
returns to the main run screen with the Clean DISPLAY active.
With the Clean DISPLAY active the main run screen is:
1. Conventional main run screen first line.
2. “Clean” second line.
99.135 psi a zHi
To go to a DISPLAY other than Clean, press [DISPLAY] and make a new
DISPLAY choice (see Section 3.3.6).
3.3.7 [HEAD]
PURPOSE
To cause a pressure fluid head correction to be added to or subtracted from the pressure
measured by the RPM4 reference pressure transducer in order to predict the pressure at a
height other than the RPM4’s reference level.
PRINCIPLE
RPM4 measures gauge or absolute pressure at the height of the rear panel TEST(+) port.
Frequently, when performing a calibration or test, the device or system under test is at a
different height than the RPM4’s TEST(+) port. This difference in height, frequently called
head, can cause a significant difference between the pressure measured by the RPM4 at its
TEST(+) port height and the pressure actually applied to the device under test which is at a
different height. In this case, it is useful to make a head correction to the pressure measured
by the RPM4 in order to predict the pressure actually applied at a different height.
RPM4 can accurately determine “head” pressures for gases (nitrogen, helium and air) and
liquids (oil, water) as the pressurized medium. In calculating the head value, standard gravity
(9.80665 m/s
density correcting for temperature of 20°C and the measured pressure using the gas’s
compressibility factor to 100 MPa (15 000 psi) and extrapolated above 100 MPa. Oil density
is taken to be 850 kg/m
to be 998.2321kg/m
2
) is used. Gas densities are calculated from the selected gas’s standard
3
, the density of typical calibration oils at 20°C. Water density is taken
3
(20°C). A custom liquid density may also be specified.
The [HEAD] function key is used to specify the height difference between the RPM4 TEST(+)
port and another height. Entering a height of zero turns the function off. The height unit and
the test fluid are specified by pressing [SPECIAL], <3Head> (see Section 3.5.3).
Use of the HEAD function to assure in tolerance measurements is most important in low
absolute pressure ranges. Specifying the head height within ± 5 cm (2.5 in.) is adequate
to assure that, even in the worst case, the uncertainty on the head correction will be
insignificant relative to the tolerance on the measurement.
To access the HEAD function, press [HEAD]. The display is:
Edit head height:
1. Test fluid currently specified for the head correction.
2. Edit field for head height.
+ 25 cm N2
Edit the head height to the desired value. Press [ENT] to return to the MAIN RUN screen
with the new head correction active. Press [ESC] to return with no changes.
The reference height of the RPM4
pressure measurement is the middle of
the RPM4 TEST port. The head height
should be entered as a positive value if
the device or system under test is
higher than the RPM4 and negative if it
is lower.
The HEAD function is NOT range specific. The HEAD ON or OFF status remains the same
as Q-RPTs and ranges are changed. Edits made to the head settings are independent of
active range or Q-RPT.
When a head correction is being applied, it is indicated by <h> in the top line of the MAIN
RUN screen (see Section 3.1.1). When the head correction is zero, the <h> is not shown.
To change units of head height between inches and centimeters and to change the test
fluid species, use [SPECIAL], <3head> (see Section 3.5.3).
3.3.8 [SDS] (SELF DEFENSE SYSTEM)
PURPOSE
To CLOSE SDS on all reference pressure transducers (Q-RPTs). To OPEN SDS on the
active Q-RPT.
SDS is present only on Q-RPTs designated A7M or lower.
The [SDS] key should be used with care to avoid accidentally overpressuring a Q-RPT.
Always double check to ensure that the pressure applied to the Q-RPT’s TEST port is
lower than the Q-RPT’s maximum pressure before OPENING SDS.
SDS can be set to permanently be OPEN as if no SDS were installed (see Section
SDS is a self defense system to protect RPM4 Q-RPTs from overpressure. SDS CLOSES
automatically to protect a Q-RPT that is not active. In a two Q-RPT RPM4, activating one
Q-RPT automatically CLOSES SDS on the other Q-RPT. Powering down the RPM4 or
energizing it also CLOSES SDS on all Q-RPTs.
When a Q-RPT’s module’s SDS is CLOSED, the Q-RPT is isolated from the TEST(+) port
and opened to the VENT port. SDS must be OPENED for the pressure applied to the
TEST(+) port to be measured. The [SDS] key is used to OPEN SDS.
The SDS function can also be used to CLOSE SDS as might be desired when leaving RPM4
at rest or when a possible overpressure situation is anticipated. Pressing [SDS] always
CLOSES any SDS systems that are OPEN.
See Section 3.2.7 for general information on SDS and its principles and Section 5.6,
Figure 8 for Q-RPT module schematics.
SDS closed is indicated by <SDS CLOSED> flashing in the current measured pressure
display field. When SDS is closed, the Q-RPT is isolated from the TEST(+) port and
connected to the VENT port.
Though the SDS self defense system includes features to protect Q-RPTs against
overpressure, SDS should NOT be considered a fail-safe overpressure protection system.
SDS cannot guarantee that overpressure damage will NOT occur. SDS closing automatically
may protect an RPT in case of accidental overpressure but conventional measures for
overpressure protection should always be used. Damage to Q-RPTs due to overpressure
is NOT covered by the RPM4 product warranty even when SDS is present and ON.
OPERATION
The maximum pressure that should be applied to an RPM4 TEST(+) port when SDS is
CLOSED is 10 MPa (1 500 psi).
When SDS is CLOSED with a pressure applied, a valve shuts isolating the Q-RPT from the
TEST(+) port. Another valve vents the Q-RPT to atmosphere through the VENT port.
If the pressure applied at the time SDS is CLOSED is less than 1 700 kPa (250 psi),
the VENT valve will open causing sudden release of pressure through the VENT port and
returning the Q-RPT to atmospheric pressure. If the pressure applied at the time SDS is
CLOSED is greater than 1 700 kPa, the VENT valve may NOT open. In this case, to vent
pressure from the Q-RPT module, remove the pressure from the TEST(+) port. This will
cause the SDS TEST port isolation valve to open partially, reducing pressure under
1 700 kPa at which time the SDS VENT valve will open (see Section 5.6, Figure 8)
Pressing [SDS] always CLOSES any SDS system that is OPEN. In a dual Q-RPT RPM4,
even if the active Q-RPT has NO SDS, pressing [SDS] CLOSES SDS for the inactive Q-RPT
if SDS is present.
Use [SDS] to OPEN SDS if it is CLOSED for the active Q-RPT and/or to CLOSE SDS for the
active and inactive Q-RPT.
When [SDS] is pressed, there are three possible results depending whether the active Q-RPT is
SDS equipped and ON (see Section 3.5.4.2), and whether its SDS is OPEN or CLOSED.
1. If the active Q-RPT IS NOT SDSequipped or its SDS if “full time OFF” (see Section 3.5.4.2):
<No SDS on RPT> or <SDS is OFF and OPEN> is displayed for 5 seconds and
operation returns to the main run screen. If there is a second Q-RPT with SDS, its SDS
is CLOSED.
2. If the active RPT IS SDS EQUIPPEDand SDS on the active RPT is OPEN:
SDS is CLOSED and operation returns to the main run screen.
3. If the active RPT IS SDS EQUIPPED and SDS on the active RPT is CLOSED:
The OPEN SDS routine is presented. The display is :
1. Maximum acceptable pressure of active Q-RPT (always
absolute on an Axxx Q-RPT and gauge on a Gxxx Q-RPT).
2. Position designator of Q-RPT for which SDS will be
opened.
Pmax! 350 kPa a Hi
Open SDS? 1no 2yes
Check that the pressure applied to the TEST(+) port does NOT exceed the maximum
acceptable value for the Q-RPT. Select <1no> to return to the main run screen without
OPENING SDS, select <2yes> to OPEN SDS and return to the main run screen with
SDS OPEN (the active Q-RPT open to the TEST(+) port).
Opening SDS with a pressure greater than Pmax! applied to the test port may cause
overpressure damage to the Q-RPT.
In differential and parallel measurement modes, opening SDS acts simultaneously on
BOTH THE Hi AND Lo Q-RPTs (see Section 3.3.8.1). The Pmax! pressure is the Lo RPT.
Be sure to check BOTH Q-RPT TEST(+) ports are at a safe pressure before opening SDS.
The [SDS] key is also the [2] key which is pressed to select <2Yes> to OPEN SDS.
Thus, SDS can be OPENED by two rapid presses of [SDS]. Use this feature to
conveniently OPEN SDS but always check that the pressure connected the TEST(+) port
does NOT exceed Pmax! before doing so.
When changing ranges causes a change of active Q-RPT, SDS is automatically CLOSED for
the Q-RPT that is being changed from, leaving the inactive Q-RPT with SDS CLOSED.
The Lo and Hi Q-RPTs’ SDS, if present, can be controlled directly regardless of which
Q-RPT is currently active (see Section 3.5.4).
SDS can be permanently opened so operation is as if SDS were not present
3.3.8.1 SDS IN DIFFERENTIAL AND PARALLEL MEASUREMENT
MODES
In differential and parallel measurement modes, opening SDS opens SDS for
BOTH the Hi and Lo Q-RPTs. Before opening SDS, be certain that the
pressure applied to both the Hi and Lo RPM4 TEST(+) ports is less than the
maximum pressure of the Lo Q-RPT.
Differential (see Section 3.3.3.1) and parallel measurement modes (see Section
3.5.7.3) use two Q-RPTs in one RPM4 to simultaneously. In these modes,
function of the [SDS] key is adapted to facilitate operation using both Q-RPTs
simultaneously. [SDS] can be used to OPEN SDS on both the active an inactive
Q-RPTs at the same time.
Pressing [SDS] always CLOSES any ON and OPEN SDS, as it does in single Q-RPT
mode operation. However, if SDS is CLOSED for any and all modules equipped
with SDS, pressing SDS goes to the usual OPEN SDS routine which will OPEN
SDS for both Q-RPTs.
The display is :
1. Maximum acceptable pressure of the Lo
Q-RPT (always absolute on an Axxx Q-RPT
and gauge on a Gxxx Q-RPT).
2. Position designator of active Q-RPT. <HL> in
parallel mode, <Hd> in differential mode.
Pmax! 350 kPa a HL
Open Hi&Lo? 1no 2yes
Check that the pressure applied to BOTH the TEST(+) ports does NOT exceed
the maximum acceptable value for the Lo Q-RPT. Select <1no> to return to the
main run screen without OPENING SDS, select <2yes> to OPEN SDS for both
Q-RPTs and return to the main run screen with SDS OPEN (the active Q-RPT
open to the TEST(+) port).
3.3.9 [AUTOZ]
PURPOSE
To run the AutoZ function that rezeros the active Q-RPT between full calibrations.
See Section 3.5.1, Principle for a full explanation of the AutoZ function.
PRINCIPLE
Run AutoZ is the function by which the current Q-RPT reading is compared to a reference,
P
, at atmospheric pressure to determine a new value of P
std,0
. The value of P
offset
offset
is then
used by AutoZ to automatically correct the Q-RPT for possible changes in zero over time
(see Section 3.5.1, PRINCIPLE).
Running AutoZ in parallel mode (see Section 3.5.7.3) autozeros the Hi and Lo Q-RPTs
simultaneously and independently. There is are no Zoffset values specific to the HL QRPT.
See Section 3.5.1, Principle for a full explanation of the AutoZ function.
To “rezero” the active Q-RPT in gauge measurement mode.
OPERATION
For the AutoZ function key to run AutoZ, AutoZ must be turned ON for the
active range and measurement mode. AutoZ ON is indicated by <z> to the
left of the Q-RPT designator on the first line of the main run screen. AutoZ
ON and OFF is set using [SPECIAL], <1AutoZ> (see Section
3.5.1). If AutoZ
is OFF for the active Q-RPT, and measurement mode, <AutoZ is off >
is displayed when [AutoZ] is pressed.
Running AutoZ in or parallel mode (see Section 3.5.7.3) autozeros both the
Hi and Lo Q-RPTs simultaneously.
To run AutoZ in gauge measurement mode, set gauge or negative gauge as the
measurement mode (see Section 3.3.3) and press [AutoZ] from the main run
screen. If the Q-RPT is a Gxxx or BGxxx type, the display is <Preparing to AutoZ> for five seconds to allow time for the bypass valve between the TEST(+)
and TEST(-) sides of the Q-RPT to assure zero differential pressure and for the
pressure to stabilize (see Section 5.6, Figure 8). <Running gauge AutoZ> is
displayed briefly before returning to the main run screen.
Before running AutoZ in gauge mode, ensure that the pressure applied to
the Q-RPT is truly zero gauge (atmospheric pressure). If running AutoZ in
gauge mode results in a zero offset that RPM4 considers unusually large,
<Confirm 0 gauge P!> is displayed when AutoZ is pressed. Check that zero
gauge pressure is applied to the TEST(+) port and there is no differential
pressure across the TEST(+) and TEST(-) ports and press ENTER to
continue or ESCAPE to abort.
Allow the RPM4 to stabilize at atmospheric pressure and ambient
temperature for 1 to 2 minutes before running AutoZ in gauge mode.
See Section 3.5.1, Principle for a full explanation of the AutoZ function.
To “rezero” the active Q-RPT in absolute measurement mode
OPERATION
For the AutoZ function key to run AutoZ, AutoZ must be turned ON for the
active Q-RPT and measurement mode. AutoZ ON is indicated by <z> to the
left of the range designator on the first line of the main run screen. AutoZ
ON and OFF is set using [SPECIAL], <1AutoZ> (see Section 3.5.1). If AutoZ
is OFF for the active range, and measurement mode, <AutoZ OFF> is
displayed when [AutoZ] is pressed.
Running AutoZ in or parallel mode (see Section 3.5.7.3) autozeros both the
Hi and Lo Q-RPTs simultaneously.
To run AutoZ in absolute measurement mode, set absolute measurement mode
(see Section 3.3.3) and press [AutoZ] from the main run screen. The display is:
1. Active Q-RPT designator.
2. Current measurement mode (absolute).
3. Selection of source of P
to AutoZ.
reference to which
std,0
Selecting <1Entry> allows the value of P
Selecting <2COM2> allows the value of P
to be entered from the front panel keypad.
std,0
std,0
AutoZ Hi abs by:
1Entry 2COM2 3Lo RPT
to be read automatically from another
DHI RPMx connected to RPM4’s COM2 communications port.
Selecting <3Lo RPT> is shown only if the active Q-RPT is Hi and there is a Lo
Axxx Q-RPT in the RPM4. This selection allows the value of P
for the Hi
std,0
Q-RPT to be read automatically from the RPM4s Lo Q-RPT.
Allow the RPM4 to stabilize at atmospheric pressure and ambient
temperature for 10 to 15 minutes before running AutoZ in absolute mode.
If running AutoZ results in a value of P
that is greater than ± 0.005 % FS of
offset
the span of the Q-RPT that is being AutoZeroed, the Q-RPT and/or the reference
used as the source of P
have been faulty. Before activating a new P
may be out of tolerance or the AutoZ process may
std,0
greater than ± 0.005 % FS of
offset
the active Q-RPT, check to be sure that both the Q-RPT and the reference
were in good working order, properly vented to stable atmospheric pressure,
at the same height, and reading in the same pressure units when AutoZ was run.
When the run AutoZ selection is made, if a HEAD correction is currently active
(see Section 3.3.7) the head correction is momentarily disabled while running
AutoZ to avoid “zeroing out” the head value.
The value of P
is always displayed and entered in Pascal (Pa).
offset
Run AutoZ by Entry
AutoZ by entry allows the value of P
(see Section 3.5.1, PRINCIPLE) to be
std,0
entered directly from the RPM4 front panel. This provides a simple way of
AutoZeroing relative to an independent reference device such as a house
barometer that does not interface directly with RPM4.
To access run AutoZ by entry press [AutoZ], <1entry>. The display is:
1. Real time pressure reading (without head
correction), pressure unit of measure and
measurement mode of the active Q-RPT.
2. Active Q-RPT position designator.
3. Entry field for the value of P
pressure unit of measure.
. in the current
std,0
Enter the value of the AutoZ reference (P
std,0
96.7755 kPa a Hi
Pstd,0:0.00000
) in the same unit of measure as the
top line display and press [ENT]. RPM4 logs the reading and calculates a new
AutoZ offset value. The next display is:
Press [ENT] to activate the new value of
or [ESC] to start over with entry of a
P
offset
new AutoZ reference (P
std,0
) value.
Old P
New P
offset
offset
: 0.0 Pa
: 8.3 Pa
The value of P
is always in Pascal (Pa). The value of P
offset
is entered in
std,0
the current pressure unit of measure.
Run AutoZ by COM2
AutoZ by COM2 allows the value of P
(see Section 3.5.1 Principle) to be read
std,0
automatically from a DHI RPMx connected to the RPM4 COM2 communications
port. The RPMx is read and the new P
To access run AutoZ by COM2 press [AutoZ], <2COM2>. The display is:
1. Real time pressure reading (without head
correction), pressure unit of measure and
measurement mode of the active Q-RPT.
2. Active Q-RPT position designator.
3. Real time reading from the RPMx connected
to RPM4’s COM2 communications port.
Observe the displayed pressures and verify that they are stable. When ready,
press [ENT] to cause AutoZ to run. RPM4 logs both readings and calculates a
new AutoZ offset value. The display is:
96.7755 kPa a Hi
Pstd,0:0.00000
offset
offset
: 0.0 Pa
: 3.1 Pa
Press [ENT] to activate the new value of
or [ESC] to start over.
P
offset
Old P
New P
For RPM4 to communicate with an RPMx connected to its COM2 port, the
RPM4 and the RPMx RS-232 interfaces must be set up properly
(see Section 3.5.2). If, the RPM4 is unable to communicate with an RPMx via
COM2 when running AutoZ by COM2, it times out after 6 seconds and
displays <RPM NOT detected>.
The value of P
is always displayed and entered in Pascal (Pa).
offset
Run AutoZ by Lo RPT
AutoZ by Lo RPT is available only if the active Q-RPT is the Hi Q-RPT and there
is an Axxx Lo Q-RPT. AutoZ by Lo RPT allows the Lo Q-RPT in an RPM4 to serve
as the AutoZ reference (source of P
The Lo Q-RPT is read and the new P
) (see Section 3.5.1, PRINCIPLE).
std,0
is calculated automatically.
offset
To access run AutoZ by Lo RPT press [AutoZ], <3Lo RPT>. The display is:
1. Real time pressure reading (without head
correction), pressure unit of measure and
measurement mode of the active Q-RPT.
2. Active Q-RPT position designator.
3. Real time reading from the Lo Q-RPT.
96.7755 kPa a Hi
Pstd,0:97.7786
Observe the displayed pressures and verify that they are stable. When ready,
press [ENT] to cause AutoZ to run. RPM4 logs both readings and calculates a
new AutoZ offset value. The display is:
To compensate for the offset between the Hi Q-RPT and the Lo Q-RPT at the
operating “line pressure” in differential measurement mode.
See Section 3.5.1, Principle for a full explanation of the AutoZ function.
OPERATION
For the AutoZ function key to run AutoZ, AutoZ must be turned ON for the
active range and measurement mode. AutoZ ON is indicated by <z> to the
left of the Q-RPT designator on the first line of the main run screen. AutoZ
ON and OFF is set using [SPECIAL], <1AutoZ> (see Section
3.5.1). If AutoZ
is OFF for the active Q-RPT, and measurement mode, <AutoZ is off > is
displayed when [AutoZ] is pressed.
To run AutoZ in differential measurement mode, set differential as the
measurement mode (see Section 3.3.3) and press [AutoZ] from the main run
screen. <Running differential AutoZ> is displayed briefly before returning to the
main run screen.
Before running AutoZ in differential mode, set the “line pressure” at which
the differential measurements will be made and assure that a common
pressure is applied to both Q-RPTs. If running AutoZ in differential mode
results in a zero offset that RPM4 considers unusually large, <Confirm 0
DP> is displayed when AutoZ is pressed. Check that zero differential
pressure is applied across the Hi and Lo TEST(+) ports and press ENTER to
continue or ESCAPE to abort.
Allow the RPM4 to stabilize at atmospheric pressure and ambient
temperature for 2 to 3 minutes before running AutoZ in differential mode.
3.3.10 [ENT] (RUN AUTOTEST)
PURPOSE
To run the AutoTest function which AutoRanges the RPM4 and runs a calibration test based
on user entries of pressure unit, measurement mode, full scale, increment and DUT
tolerance.
Run AutoTest is accessed by pressing [ENT] from the Rate or Clean DISPLAY mode (see
Section 3.3.6.2). In other DISPLAY modes, the function of [ENT] is specific to the mode.
This feature not available at the time of release of this manual. Check
for new RPM4 embedded software releases and manual updates.
OPERATION
Feature not available at the time of release of this manual.
www.dhinstruments.com
3.4[SETUP]
PURPOSE
[SETUP] accesses a menu of functions and features commonly used in setting up and using RPM4.
OPERATION
To access the SETUP menu, press [SETUP] from the MAIN RUN
screen. The display is:
SETUP menu choices include:
<1range> Save and delete ranges created by AutoRange (see Section 3.4.1).
<2res> Adjust the resolution of pressure displays (see Section 3.4.2).
<3stab> Adjust the pressure stability test that is the criterion for the Ready/Not Ready indication
(see Section 3.3.3).
<4UL> Adjust upper and lower pressure limit alarms (see Section 3.4.4).
<5ATest> View data files resulting from running QDUT AutoTests (See Section 3.4.5).
1range 2res 3stab
4UL 5ATest
3.4.1 <1RANGE>
PURPOSE
Save AutoRange ranges and associated settings for recall. Delete previously saved AutoRange ranges.
PRINCIPLE
The RPM4 AutoRange function creates a range and automatically sets various operating
parameters proportionally to the range (see Section 3.3.4). Operating settings can then be
changed by the user. Many operating settings are specific to the current range (see Table 4).
[SPECIAL], <1range> allows the AutoRange range and associated settings to be saved for
recall using the [RANGE] key (see Section 3.3.1). This can avoid having to recreate
frequently used ranges and settings.
[SPECIAL], <1range> is also used to delete ranges that have been saved but are no longer needed.
3.4.1.1 SAVING AN AUTORANGE RANGE
OPERATION
To save a range created by AutoRange, the range must be the active range.
Use AutoRange to create the range (see Section 3.3.4) and then make any
desired feature and setting adjustments.
Once the AutoRange range is set up as desired, press [SETUP], <1range>,
<1save>. The display is:
1. Indication that this is a range to be saved.
2. Designator and position of the Q-RPT used by
the range to be saved.
3. Pressure unit of measure of the range to be
saved.
4. Full scale pressure of the range to be saved in
gauge mode and absolute mode if available.
Save range A700K Hi
kPa 300g/100a
Press [ENT] to save the range or [ESC] to return to the MAIN RUN screen without
saving the range.
Default (<DF>) Q-RPT ranges cannot be saved. They are permanent.
Duplicate AutoRange ranges cannot be saved. A duplicate AutoRange range
is a range that has the same measurement mode and full scale and uses the
same Q-RPT as a range that has already been saved.
3.4.1.2 DELETING AUTORANGE RANGES
OPERATION
To delete a single saved AutoRange range, press [SETUP], <1range>, <2delete>.
If there are any saved ranges available to delete, the display is:
1. Indication that the cursor control keys may be
used to toggle through the saved ranges to
select the one to be deleted.
2. Designator and position of the Q-RPT used by
the range to be deleted.
3. Pressure unit of measure of the range to be
deleted.
4. Full scale pressure of the range to be deleted
in gauge mode and absolute mode if available.
Use the cursor control keys to toggle through the ranges available to be deleted.
When the desired range to be deleted is displayed, press [ENT]. If you are sure you
want to delete this range, press [ENT] again in response to the <Delete?> query.
[ESC] returns to the <AutoRange:> screen without deleting a range.
To delete all of the saved AutoRange ranges, press [SETUP], <1range>, <3delete all>. If you are sure you want to delete all the ranges, respond <1yes>
to the <Delete all saved ranges?> query. If not, respond <2no>.
3.4.2 <2RES> (RESOLUTION)
< and > A700K Hi
kPa 300g/100a
PURPOSE
To set the resolution with which measured pressures and other pressure indications and
settings are displayed.
PRINCIPLE
The resolution with which RPM4 displays pressure values can be adjusted. This feature can
be used to reduce the resolution when lower precision measurements are being made and
excess resolution might confuse or distract the operator.
The resolution setting determines the number of digits with which pressure is displayed.
The resolution is calculated based on the span of the range and then rounded to the furthest
digit to the right. For example, resolution of 0.001 % on a range of 150 kPa is
150 X 0.001 % = 0.0015 which is rounded down to 0.001 kPa.
Default resolution is 10 ppm of active range span. Resolution is set automatically by
AutoRange (see Section 3.3.4).
OPERATION
To access the resolution function, press [SETUP],
<2res>. The display is:
Display resltn: Hi
0.0010 %FS < and >
Use the cursor control keys to set the desired level of resolution. Press [ENT] to return to the
MAIN RUN screen with the new resolution setting active or [ESC] to make no changes.
The resolution setting is range specific. A resolution setting made in one range does NOT
affect other ranges.
The measured pressure resolution is fixed for altitude units at 1 m in meters and 1 ft in feet.
The maximum resolution setting is limited in AutoRange ranges that are less than 10% of the
Q-RPT default full scale.
3.4.3 <3STAB>
PURPOSE
To view and/or adjust the stability test that is the Ready/Not Ready criterion for the active
RPT and range.
See Section 3.2.1 Pressure Ready/Not Ready Indication.
PRINCIPLE
RPM4 continuously monitors the rate of change of pressure measured by the active RPT and
range and compares this rate to the stability limit to make a Ready/Not Ready determination
(see Section 3.2.1 Pressure Ready/Not Ready Indication). The stability function allows the
stability limit to be adjusted by the user to increase or decrease the stability required for a
Ready condition to occur.
The default stability limit is + 0.005% FS of the active range.
The stability limit is separate and different from the Rate DISPLAY function (see Section
3.3.6.2) which allows the current rate of change of pressure to be displayed.
To access the stability setting adjustment, press
[SETUP], <3stab>. The display is:
Stability: Hi
0.020 Pa
Edit the desired stability limit setting if desired. [ENT] activates the stability limit for the active
range and returns to the main run screen Press [ESC] to return to the MAIN RUN screen
with no change to the stability limit.
The stability setting is range specific. A stability setting made in one range does NOT
affect other ranges.
3.4.4 <4UL> (UPPER LIMIT)
PURPOSE
To set the upper and lower pressure limit values for a pressure range and measurement mode.
PRINCIPLE
The UPPER LIMIT function provides the user with settable pressure limits at which an alarm
sounds. Absolute and gauge measurement modes (see Section 3.3.3, PRINCIPLE) have
upper limits only. Negative gauge mode and the BA100K also have a lower limit.
When the limit is reached, RPM4’s beeper sounds intermittently as long as pressure is
outside the limit.
The UPPER LIMIT function has two purposes. First, when UL is set to its default value, it serves
as a warning that the maximum pressure of the active range (and lower pressure in negative
gauge mode) is about to be exceeded. Second, UL can be set by the user to a value other
than the default value to provide an alarm that a specific pressure limit has been exceeded.
This feature is often used to help protect an external device or system on which RPM4 is
being used to measure pressure. For example, it might be set just over the full scale of a
device under test (DUT) that is being calibrated. Note that using AutoRange automatically
sets the UL to just above the AutoRange full scale (see Section 3.3.4).
The default upper limit is 105 % of AutoRange span or 102 % of Q-RPT default (maximum)
range, whichever is lower. The default lower limit (negative gauge mode only ) is – 110 kPa
(- 16 psi), except with the BG15K Q-RPT in which the default lower limit is the negative
equivalent of the upper limit. The default lower limit of the BA100K Q-RPT is 66 kPa (9.6 psi).
In parallel measurement mode (see Section 3.5.7.2), the default and maximum UL is the
UL of the Lo Q-RPT.
In differential measurement mode (see Section 3.3.3.1), UL has upper and lower limits.
The default upper differential UL is the maximum pressure of the Hi Q-RTP. The default lower
UL is the negative value of the upper limit.
In differential measurement mode, the ULs on the Lo Q-RPT and Hi Q-RPT default range are
also active so that line pressure (Lo Q-RPT) and total pressure (Hi Q-RPT) upper limit can be
set and monitored in addition to a differential limit.
To view or edit upper and lower limits press [SETUP], <4UL>.
If the current measurement mode is absolute or gauge, there is an upper limit only.
The display is:
1. Entry field for upper limit value in active pressure unit of
measure and measurement mode.
Upper limit: Hi
204.000 kPa a
Edit the upper limit value as desired. The maximum upper limit is 105 % of span in an
AutoRange range or 102 % of default Q-RPT full scale, whichever is smaller. Press [ENT] to
return to the MAIN RUN screen with the new upper limit active. Press [ESC] to return to the
MAIN RUN screen with no change to the upper limit.
If the current measurement mode is negative gauge, there is an upper limit and a lower limit.
The display is:
1. Entry field for upper limit value.
2. Entry field for lower limit value and active range pressure unit
of measure and measurement mode indication.
Upper: 104.00 Hi
Lower:-110.00 kPa g
Edit the upper and lower limit values as desired. Use the cursor control keys to move
between the two edit fields. The lower limit must be a negative value. Press [ENT] to return
to the MAIN RUN screen with the new upper and/or lower limit active. Press [ESC] to return
to the MAIN RUN screen with no change to the limits.
When the upper or lower limit has been exceeded, the display of current pressure flashes
and a buzzer sounds for 3 seconds on/2 seconds off intervals. Change the pressure applied
to RPM4 so that it is within the upper and/or lower limit to return to normal operation.
Upper limit values are specific to each range and measurement mode. Be careful not to
assume that the upper limit set in one measurement mode will apply to the other.
For example, if you change the upper limit in gauge mode, the upper limit will not be
changed in negative gauge mode.
Upper limits are always specified and displayed in the current pressure unit except for
altitude units. When in altitude units, upper limits are expressed in kPa if the altitude
unit is meters (m) and psi if the altitude unit is feet (ft).
In Hi Q-PRT differential mode (see Section 3.3.3.1) there are two UL values for the Hi Q-
RPT. One is the upper and lower limts for differential pressure, with a default value equal
to the span of the Lo RPT. The other is the UL on the base mode of the Hi Q-RPT DF
range to monitor the maximum pressure on the Hi Q-RPT. In differential mode, the Lo QRPT UL is also active to provide monitoring of the differential mode line pressure.
In addition to the UL function, RPM4 has an over pressure function.
The over pressure function executes when a Q-RPT measures a pressure that is
104 % of its default span. The G15K and BG15K Q-RPTs also have a negative
over pressure function. The negative over pressure function executes when the
Q-RPT measures a pressure less than - 17.5 kPa (- 2.3 psi).
The over pressure function causes the measured pressure display to flash.
If the Q-RPT module is equipped with SDS, SDS closes shutting off the TEST(-)
port (see Section 3.2.7). If the Q-RPT is a Gxxx or BGxxx, the TEST(+)/TEST(-)
bypass valve is opened (see Section 5.6, Figure 8). The overpressure function also
logs the time and date of the overpressure condition in both user and factory logs to
assist in incident diagnosis (see Section 3.5.7.5).
To recover from an overpressure condition, cycle RPM4 power. Be sure to
correct the condition that caused the overpressure before cycling power.
3.4.5 <5ATEST>
This feature not available at the time of this manual’s release. Check www.dhinstruments.com
for new RPM4 embedded software releases and manual updates.
3.5 [SPECIAL]
PURPOSE
[SPECIAL] accesses a menu of RPM4 functions and features that are less commonly used or not
normally used in regular operation.
OPERATION
1AutoZ 2remote 3head
Press [SPECIAL] from the MAIN RUN screen to access the
SPECIAL menu. The display is:
Some screens, such as the SPECIAL menu, go beyond the two lines provided by the display. This is
indicated by a flashing arrow in the second line of the display. Press the cursor control keys to move the
cursor to access the lines that are not visible or directly enter the number of the hidden menu choice if you
know it.
SPECIAL menu choices include:
<1AutoZ> Manage AutoZero function for the active Q-RPT (see Section 3.5.1).
<2remote> View and edit COM port (RS-232) and IEEE-488 interface settings. Select remote syntax
style (see Section 3.5.2).
<3head> Set HEAD function fluid and height unit of measure (see Section 3.5.3).
<4SDS> To view and set the ON/OFF and OPEN/CLOSED status of SDS for all Q-RPTs.
<5prefs> View and set screen saver time, keypad sound frequency, unit ID number, time/date,
security protection level (see Section 3.5.5).
<6Punit> Customize the [UNIT] key pressure unit of measure selections (See section 3.5.6).
<7intern> Access internal functions including on-board barometer viewing, read rate adjustment, parallel
measurement mode ON/OFF, direct control of Lo Vent valve(s), incident log viewing (see
Section 3.5.7).
<8cal> View and adjust RPM4, Q-RPT and barometer calibration coefficients (see Sections 3.5.8, 5.2, 5.3).
<9reset> Access RPM4’s various reset functions (see Section 3.5.9).
To offset the RPM4 Q-RPT(s) readings relative to a reference value in order to compensate
for possible changes in Q-RPT zero between full recalibrations. In differential measurement
mode, to compensate for the difference between the reading of the Hi and Lo Q-RPTs at the
“line pressure” at which differential measurements are made.
To assure operation within measurement uncertainty specifications (see Section 1.2.2), it is
recommended that AutoZ be run (the value of P
whenever the line pressure is changed in differential mode and at least every 30 days or when
RPM4 has been exposed to temperature changes exceeding 15 ºC (36 ºF) in absolute mode.
PRINCIPLE
updated) whenever vented in gauge mode,
offset
AutoZ Purpose and Principle
The main component of the change over time of the RPM4 Q-RPTs is change in zero or
offset, independent of span. Offsetting or “rezeroing” RPM4 Q-RPTs relative to a reference
between recalibrations allows measurement uncertainty specifications to be maintained with
less frequent full calibrations. AutoZ is also used to compensate for the difference between
the Hi and Lo Q-RPT at the “line pressure” at which differential measurements are made.
The RPM4 AutoZero function (AutoZ) provides full on-board support for the rezeroing
process to simplify its application by the user.
The AutoZero function uses three values:
1. P
: The pressure value indicated by the AutoZ reference, the device that is acting as
std,0
the reference relative to which to offset the RPM4 Q-RPT.
For Axxx (absolute) Q-RPTs in absolute measurement mode
AutoZ is performed is normally atmospheric pressure and the P
, the pressure at which
value can be supplied
std,0
a) by manual entry, b) automatically from a DHI RPMx connected to RPM4 COM2
communications port, or c) automatically from an Axxx Lo Q-RPT if one is present in the
RPM4.
For Gxxx or BGxxx (gauge) Q-RPTs or Axxx (absolute) Q-RPTs in gauge measurement
, P
mode
is always zero (atmospheric pressure) which is supplied by definition when
std,0
the Q-RPT is vented to atmosphere.
For RPM4s with two Q-RPTs of the same type (Axxx, Gxxx, BGxxx) with the Hi Q-RPT in
differential measurement mode, P
is the pressure indicated by the Lo Q-RPT when
std,0
the Hi and Lo Q-RPTs are connected to a common pressure.
2. P
: The pressure reading of the Q-RPT, with no AutoZ offset, at the time AutoZ is
u,0
performed.
offset,G
, P
offset,A
3. P
AutoZ offset (P
absolute(
P
represents the difference between the Q-RPT reading (P
offset
). The AutoZ function manages the determination, storage and application of P
(P
std,0
) or differential (D) measurement mode:
A
correct the zero offset in absolute, gauge and differential measurement modes.
When the Q-RPT is calibrated, P
intervals using the AutoZ function. The most recent value of P
reading to correct for change in zero.
: The difference between the reading of the Q-RPT with no
offset,D
P
= P
offset
is set to zero. P
offset
u,0
- P
std,0
u,0
is then redetermined at regular
offset
offset
) for gauge (G),
std,0
) and the AutoZ standard
to
offset
is applied to the Q-RPT
3. OPERATION
The AutoZ function can be turned ON and OFF. Table 7 summarizes the effect of AutoZ ON
and OFF for the three Q-RPT types in different measurement modes.
AutoZ in absolute measurement mode with an Axxx Q-RPT
Q-RPTs with the designation Axxx are intrinsically absolute. They have an evacuated and
sealed reference. Axxx Q-RPTs can measure absolute pressure (difference from vacuum).
In absolute measurement mode, the source of P
must be an absolute pressure, nominally
std,0
atmospheric pressure, with uncertainty significantly better than that of the Q-RPT that is
being AutoZeroed. For higher range Q-RPTs, this is easily accomplished with a variety of
digital barometers or with a Lo Q-RPT in the RPM4. For lower range Q-RPTs, a piston gauge or
other standard may be required to provide P
with low enough uncertainty.
std,0
When using an Axxx (absolute) Q-RPT in absolute measurement mode, with AutoZ ON,
absolute pressure is calculated as:
P
= Pu - P
abs
offset,A
In absolute measurement mode, running the AutoZ function to update the AutoZ value
(P
), is initiated by the operator using [AutoZ].
offset,A
AutoZ in gauge and negative gauge measurement modes with an Axxx Q-RPT,
dynamic compensation for atmospheric pressure
Q-RPTs with the designation Axxx are intrinsically absolute but they are also used in gauge
and negative gauge measurement modes (difference from atmosphere) (see Section 3.3.3,
PRINCIPLE). Gauge measurement mode is achieved by subtracting the value of
atmospheric pressure from the Q-RPT’s absolute reading using AutoZ and by dynamically
compensating for changes in atmospheric pressure between opportunities for AutoZ to
execute (see Section 3.2.2).
In gauge measurement modes on an Axxx Q-RPT, the value of P
is always zero gauge
std,0
pressure. Zero gauge pressure, by definition, is applied to the Q-RPT when it is vented to
atmosphere. In gauge measurement modes, AutoZ is run when initiated by the operator
pressing [AutoZ] when the RPM4 is vented. This assures regular updating of the P
offset,G
value corresponding to atmospheric pressure. Gauge pressure is the measured absolute
pressure, P
, minus P
u
offset,G
.
P
gauge
= Pu - P
offset,G
When AutoZ is ON, dynamic compensation for atmospheric pressure is also applied to
compensate for changes in atmospheric pressure as measured by a barometer (ΔP
AutoZ updates (see Section 3.2.2). The measured gauge pressure is calculated using ΔP
to correct the value of P
offset.G
.
P
gauge
= Pu - P
offset.G
- ΔP
atm
) between
atm
atm
In gauge measurement mode on an Axxx RPT, running the AutoZ function to update the
value of P
, is initiated by the operator using [AutoZ]. Updating of ΔP
atm,0
occurs
atm
automatically and continuously.
When gauge or negative gauge mode is first activated on an Axxx Q-RPT, the value of
P
subtracted from the absolute pressure reading to achieve gauge pressure is
offset.G
standard atmosphere of 101.325 kPa. This can cause a large zero error when gauge mode
is activated if atmospheric pressure is significantly different from standard atmospheric
pressure. When gauge mode is activated, always press [AutoZ] to run AutoZ and obtain
an actual value of P
AutoZ in gauge measurement mode with a Gxxx or BGxxx Q-RPT
Q-RPTs with the designation Gxxx or BGxxx are intrinsically gauge. They have an atmospheric
(TEST(-)) reference port and measure gauge pressure (difference from atmosphere).
In gauge measurement mode the value of P
is always zero gauge pressure. Zero gauge
std,0
pressure, by definition, is applied to the Q-RPT when it is vented to atmosphere.
When using a Gxxx Q-RPT in gauge measurement mode, with AutoZ ON, gauge pressure is
calculated as:
= Pu - P
P
gauge
offset,G
In gauge measurement modes, AutoZ is run when initiated by the operator pressing [AutoZ]
when the RPM4 is vented. This assures the regular automated updating of the P
offset,G
value.
AutoZ in differential measurement mode
RPM4s with two Q-RPTs of the same type (Axxx, Gxxx, BGxxx) support differential
measurement mode (see Section 3.3.3.1) in which the measured pressure is the difference
between the pressure measured by the Hi Q-RPT and the Lo Q-RPT (Hi – Lo).
In differential measurement mode, the purpose of AutoZ is to compensate for the offset
between the Hi and Lo Q-RPT indications at the “line pressure” at which the differential
pressure is being measured. The value of P
is the pressure indicated by the Lo Q-RPT at
std,0
the time of AutoZ.
When using an RPM4 with two Q-RPTs in differential measurement mode, with AutoZ ON,
differential pressure is calculated as:
P
differential
= PHi - PLo - P
offset,D
In differential measurement mode, AutoZ is run when initiated by the operator pressing
[AutoZ] when both Q-RPTs of the RPM4 are connected to a common pressure..
AutoZ ON/OFF
The AutoZ function can be turned ON and OFF, separately for gauge, absolute and
differential measurement modes. Table 7 summarizes the effect of AutoZ ON and OFF.
Table 7. AutoZ ON and OFF
Q-RPT
TYPE
Axxx
Gxxx
or
BGxxx
Any, Hi Q-RPT
in dual Q-RPT
RPM4
MEASUREMENT
MODE
Absolute
Gauge or
negative gauge
Gauge or
negative gauge
Differential
AutoZ
STATUS
ON YES
OFF NO
ON YES
OFF
ON YES
OFF NO
ON YES
OFF NO
P
offset
APPLIED
YES
ΔP
atm
APPLIED
Not Applicable
NO
Not Applicable
Not Applicable
AutoZ
ROUTINE
RUNS
When initiated
by operator
Not available
When initiated
by operator
When initiated
by operator
Not available
When initiated
by operator
Not available
Recommendations for the Use of the AutoZ Function
The AutoZ function provides a powerful and easy to use tool for improving the stability over
time of RPM4 Q-RPTs and maximizing the recalibration interval by compensating for change
in zero between full recalibrations. In differential measurement mode, it minimizes the
uncertainty in the differential pressure due to the offset between the Hi and Lo Q-RPTs at the
“line pressure”. The following simple recommendations will help assure that you use this
feature to best advantage.
• In gauge and differential measurement mode: Always leave AutoZ ON when
operating.
•In absolute measurement mode: Always leave AutoZ ON when operating if the AutoZ
routine has been run regularly using a valid atmospheric reference.
• To assure operation within measurement uncertainty specifications (see Section 1.2.2),
run AutoZ whenever vented in gauge mode, whenever the line pressure is changed in
differential mode and at least every 30 days or when RPM4 has been exposed to
temperature changes exceeding 15 ºC (36 ºF) in absolute mode.
• In absolute measurement mode: Run AutoZ only when a reference whose
measurement uncertainty is known to be significantly better than that of the Q-RPT to be
AutoZeroed is available. Keep range ratios in mind when comparing uncertainty.
A ± 0.01 % FS barometer is roughly 10 times better than an ± 0.01 % 1 MPa (150 psi)
Q-RPT because the Q-RPT/barometer pressure ratio is 10:1. The measurement
uncertainty comparison should be made at 30 % of span for both standard class and
premium class Q-RPTs. Though it may not be practical and generally is not necessary,
the best possible reference with which to run AutoZ in absolute measurement mode is a
gas operated piston gauge (such as a DHI PG7601) applying atmospheric pressure to
the RPM4 test port. The best day to day reference is a properly calibrated DHI RPM4
with a BA100K Q-RPT interfaced directly as an external device to the RPM4 COM2 port.
• In absolute measurement modes: Allow the RPM4 to stabilize at atmospheric pressure
and ambient temperature for 10 to 15 minutes before running AutoZ.
OPERATION
The AutoZ function and values are Q-RPT AND measurement mode (gauge/negative gauge
or absolute) specific.
To RUN the AutoZ routine that rezeroes the Q-RPT, use [AutoZ] (see Section 3.3.9).
To access the RPM4 AutoZ function press [SPECIAL], <1AutoZ>. The display is:
1. Active Q-RPT designator.
2. Indication of active measurement mode (<abs> for absolute, <gage> for gauge and negative gauge, <dif> for differential).
3. Indication of whether AutoZ is currently ON or OFF for this
Q-RPT and measurement mode.
1off 2view Hi
3edit abs ON
Select <1off> (or <1on>) to change the AutoZ status for the current Q-RPT and
measurement mode from ON to OFF or vice versa.
AutoZ ON is indicated by a <z> in the MAIN RUN screen, top line, fourth character from
the right. When AutoZ is OFF, the character is blank.
Select <2view> to view the current value of P
P
should be zero when the RPM4 is new or has just been calibrated. P
offset
roughly equal to atmospheric pressure for an Axxx (absolute) Q-RPT operating in gauge mode.
Select <3edit> to edit the value of P
(see Section 3.5.1.1).
for the active Q-RPT and measurement mode
offset
The value of P
is always displayed and entered in Pascal (Pa).
offset
3.5.1.1 EDIT AUTOZ
The edit AutoZ function should be used with great caution as entering
inappropriate values and turning ON AutoZ may result in out of tolerance
measurements. In normal operation, the value of the AutoZ offset, P
should be changed by running AutoZ using [AutoZ] (seen Section 3.3.9).
Before editing P
To edit the current P
, see Section 3.5.1, PRINCIPLE.
offset
value for the active Q-RPT and measurement mode,
offset
press [SPECIAL], <1AutoZ>, <3edit>. The display is:
1. Indication of active measurement mode
(<abs> for absolute, <gage> for gauge and
negative gauge, <dif> for differential).
2. Active Q-RPT for which P
3. Edit field for value of P
Edit P
as desired and press [ENT] to activate the new value. Press [ESC] to
offset
is being viewed.
offset
.
offset
Poffset, Abs Hi
8.4 Pa
abandon changes.
offset
,
The value of P
is always displayed and entered in Pascal (Pa).
offset
3.5.2 <2REMOTE>
PURPOSE
To configure the RPM4 COM1, COM2 and IEEE-488 (GPIB) communication ports. To test COM1
and COM2 communications. To select the remote programming communications format.
PRINCIPLE
The RPM4 has two RS-232 communications ports referred to as COM1 and COM2 and a
single IEEE-488 (GPIB) port. COM1 or the IEEE-488 port is for communicating with a host
computer (see Section 4), and COM2 is reserved for communicating with an external device
(e.g. an RPMx, a multimeter, etc.). These ports’ settings can be viewed and changed using
[SPECIAL], <2remote>.
RPM4 has two remote communications formats, classic and ENHANCED (see Section 4.3).
Which of these is active can be selected.
A self test is supplied for RS-232 communications. The self test allows verification that the
RPM4 RS232 ports (COM1 and COM2) are operating properly and that a valid interface
cable is being used.
To access the communications settings, press [SPECIAL], <2Remote>.
• Select <1COM1> to view and edit COM1 settings (see Section 3.5.2.1.).
• Select <2COM2> to view and edit COM2 settings (see Section 3.5.2.1.).
• Select <3IEEE> to view and edit IEEE-488 settings (see Section 3.5.2.2).
• Select <4format> to select the remote communications command format (see Section 3.5.2.3).
• Select <5RS232test> to run the COM1 and COM2 communications test (see Section 3.5.2.4).
3.5.2.1 <1COM1, 2COM2>
The COMx ports can be set for specific communications settings. The settings
are baud rate, parity, data bits and stop bits. The available options are:
The default is <2 400, E, 7,1> for both COM1 and COM2.
The user can also specify one or two termination characters as well as define
these characters. These are referred to as “Term1” and “Term2”. These define
the characters that mark the end of commands that are sent to the RPM4. The
RPM4 looks for an ASCII(13) (carriage return) to terminate a received command
but responds with both and ASCII(13) (carriage return) and an ASCII(10) (line
feed). There are no other options.
3.5.2.2 <3IEEE-488>
The IEEE-488 port’s primary address can be set from 1 to 31 in this screen.
The factory default value is 10. Secondary addressing is not used or supported.
This address must not conflict with the address of any other device on the same
IEEE-488 bus.
The receiving terminating character must be a line feed and EOI. Carriage returns
are ignored if received. The RPM4 sends a line feed and asserts the EOI line to
terminate a reply. These settings are fixed to agree with IEEE Std. 488.2. If you
change the address, the IEEE interface will reset (PON) and become idle.
3.5.2.3 <4FORMAT>
The RPM4 has two different syntax formats available for the remote program
commands.
The classic remote command format is highly intuitive and conforms with
previous DHI PPC and RPM products.
The enhanced remote command format generally uses the same commands as
the classic format, but in addition it follows the syntax, format, and status
reporting features of IEEE Std 488.2.
The details of each format are covered in Section 4.3.
[SPECIAL], <2remote>, <4format> allows the remote program command
syntax to be selected. The cursor is on the active format. Select <1classic> or
<2enhanced> as desired.
The RS232 self-test is provided to check the RPM4 COM ports and the interface
cable independently of an external device or computer.
If you are having difficulty communicating with RPM4 from a host computer
using RS232, the RS232 self test can help establish that the RPM4 COM1 port
you are trying to communicate with and the interface cable you are using are good.
To run a self test of the RS232 ports (COM1 and COM2), press [SPECIAL], <2remote>, <5RS232test>.
The display prompts you to connect COM1 to COM2 using a standard pin-to-pin
DB-9F to DB-9M RS232 cable (see Section 4.2.1.1, 4.2.1.3).
Once the cable has been installed, press [ENT] to run the self-test. The test is first
executed in the COM1→COM2 direction and then in the COM2→COM1 direction.
If the COM1→COM2 test passes: <PASSED> displays briefly and the test
If COM2→COM1 passes: <PASSED> is displayed briefly followed
If a test fails: Execution is suspended until [ENT] is
proceeds to COM2→COM1.
by the conclusion, <The RS232 test has PASSED>.
pressed.
The RPM4 RS232 test can fail for three reasons:
1. The RS232 cable being used is incorrect (Section 4.2.1.1, 4.2.1.3 for
information on the correct cable).
2. COM1 and COM2 do NOT have the same serial communications settings
and therefore cannot communicate together (see Section 3.5.2.1 to set
the COM ports).
3. COM1 or COM2 is defective.
The reason for failed communications is almost always a cable or incorrect
RS232 interface settings. Be sure that these are correct before concluding
that a COM port is defective.
3.5.3 <3HEAD>
PURPOSE
To view or change the properties of the HEAD function (see Section 3.3.7) including the unit of
measure of length for head height entry and the test fluid type and species for density calculations.
OPERATION
From the MAIN RUN screen, pressure [SPECIAL], <3Head>.
Select the desired height unit of measure and press [ENT].
Select <1gas> or <2liqud> depending on the pressurized medium between the RPM4 and
the height at which pressure is to be calculated. Select the gas or liquid species. The liquid
selection <3User>, is to create a custom liquid with a user entered density.
To open and close SDS temporarily for the Hi or Lo Q-RPT or to turn the SDS function ON
and OFF full time for the Hi or Lo Q-RPT.
See Section 3.2.7 for complete information on RPM4’s SDS (Self Defense System) feature.
PRINCIPLE
SDS is a self defense system to protect RPM4 Q-RPTs from overpressure (see Section 3.2.7).
Normally, opening and closing SDS is performed automatically by the RPM4 or by the
operator using the [SDS] key.
[SPECIAL], <4SDS> gives further control over the SDS function. This includes:
a) the possibility of temporarily opening or closing SDS on either the Hi or Lo Q-RPT (see
Section 3.5.4.1), and; b) turning OFF SDS so that it is open full time and RPM4 can be
operated as if SDS were not installed (see Section 3.5.4.2).
The temporary opening or closing of SDS allows the SDS open/closed state to be set
immediately to the desired state. Following the change, normal automated operation of SDS
and control using the [SDS] key is still valid. This function is often used to open SDS on the
inactive Q-RPT in PRM4s with dual Q-RPTs.
Setting the SDS function to full time OFF eliminates the SDS function completely until is set
to full time ON again.
3.5.4.1 <1TEMP OPEN/CLOSE>
OPERATION
See Section 3.2.7 for complete information on RPM4’s SDS (Self Defense
System) feature.
To temporarily open or close the SDS of the Hi or Lo Q-RPT press [SPECIAL],
<4SDS>, <1temp open/close>. If the RPM4 has two Q-RPTs and both have
SDS, the Q-RPT must then be selected. The next display is:
The cursor is on the current state of SDS for the
designated Q-RPT. Select <1close> to close
SDS or <2open> to open SDS and return to the
MAIN RUN screen. [ESC] to return without
making changes.
SDS temporary: Hi
1close 2open
If <2open> is selected and SDS for the Q-RPT is not already open, the SDS
opening screen with indication of the maximum pressure of the Q-RPT is displayed.
This is the same screen used when opening SDS using the [SDS] key (see
Section 3.3.8). Check that the pressure applied to the Q-RPT module TEST(+)
port does not exceed the Pmax! value before confirming that SDS should be opened.
Opening SDS with a pressure greater than Pmax! applied to the test port
See Section 3.2.7 for complete information on RPM4’s SDS (Self Defense
System) feature.
OPERATION
Turning SDS “full time OFF” for a Q-RPT module eliminates normal SDS
operation so that the Q-RPT module TEST(+) port remains open. However,
the SDS overpressure function (see Section 3.4.4.1) remains active.
To turn SDS full time ON or OFF for a Q-RPT module, press [SPECIAL],
<4SDS>, <2full time on/off >. If the RPM4 has two Q-RPTs and both have SDS,
the Q-RPT must then be selected. The next display is:
The cursor is on the current state of SDS for the
designated Q-RPT. Select <1on> to set SDS for
normal operation and return to the main run
screen. Select <2off> to open SDS permanently
so operation is as if SDS were not installed.
SDS full time: Hi
1on 2off
[ESC] to return without making changes.
If <2off> is selected and SDS for the Q-RPT is not already open, the SDS opening
screen with indication of the maximum pressure of the Q-RPT is displayed. This is
the same screen used when opening SDS using the [SDS] key (see Section
3.3.8). Check that the pressure applied to the Q-RPT module TEST(+) port does
not exceed the Pmax! value before confirming that SDS should be opened.
Opening SDS with a pressure greater than Pmax! applied to the test port
may cause overpressure damage to the Q-RPT.
3.5.5 <5PREFS>
PURPOSE
To access a menu of RPM4 operational preferences and functions.
OPERATION
To access the PREFS menu press [SPECIAL],
<5prefs>. The display is:
The PREFS menu includes:
<1ScrSvr> View and change the screen saver activation time (see Section 3.5.5.1).
<2sound> View and change the keypress sounds (see Section 3.5.5.2).
<3time> View and edit the internal time and date settings (see Section 3.5.5.3).
<4ID> View the RPM4 serial number (SN) and view or edit the ID number
(see Section 3.5.5.4).
<4level>View and set user security level and/or password (see Section 3.5.5.5).
To adjust the idle time after which RPM4’s SCREEN SAVER activates.
PRINCIPLE
RPM4 has a SCREEN SAVER function which causes the display to dim after a
front panel key is NOT pressed for a certain amount of time. The default is for the
screen saver to activate after 10 idle minutes. The screen saver activation time
can be adjusted by the user or screen saving can be completely eliminated.
Setting screen saver time to zero eliminates the SCREEN SAVER function so
that the display permanently remains at full brightness.
OPERATION
To access the SCREEN SAVER function, press [SPECIAL], <5prefs>,
<1ScrSav>. Edit, in minutes, the idle time after which screen saver will activate
to dim the screen. Set the time to zero to eliminate the SCREEN SAVER function.
3.5.5.2 <2SOUND>
PURPOSE
To adjust or suppress the RPM4 valid key press sound.
PRINCIPLE
RPM4 provides audible feedback by a brief “beep” when a valid key press is
made. The tone frequency of this beep may be selected from three choices or it
may be completely suppressed. Invalid key presses are indicated by a
descending two tone “blurp” which cannot be suppressed.
OPERATION
To access the keypad sound adjustment function press [SPECIAL], <5prefs>,
<2sound>.
Select between <2lo>, <3mid> or <4hi> to adjust the valid key press tone frequency.
Select <1none> to suppress the valid key press tone.
The sound function only affects the valid key press tone. The invalid key
press tone and other RPM4 sounds cannot be adjusted or suppressed.
3.5.5.3 <3TIME>
PURPOSE
To view and edit the RPM4 internal time and date settings.
OPERATION
To access the TIME function
press [SPECIAL], <5prefs>, <3time>.
The display is:
Select <1time> to edit the time. Edit hours, then minutes, then am/pm by
pressing [ENT] after each entry. Seconds go to zero when minutes are entered.
This can be used to synchronize the time with a time standard.
The RPM4 date and time are set to United States Mountain Standard Time
in the final test and inspection process at the factory. If desired, use the
TIME and DATE function to set your local time and date.
3.5.5.4 <4ID>
PURPOSE
To view or edit the RPM4 user ID and to view the RPM4 serial number.
OPERATION
To access the ID function press [SPECIAL] and select <5prefs>, <2ID>.
Select <1view> to view the current ID.
Select <2edit> to edit the ID.
The ID has twelve characters. When the edit screen is opened, the cursor is on
the first character. Numerical values can be entered directly from the keypad. In
addition, the cursor control keys can be used to toggle through a list of available
alphanumeric characters. Holding the key slews through the characters.
Character order going up is: blank space, symbols, lower case letters, upper
case letters, numbers. After selecting a character, press [ENT] to activate it and
move to the next character field.
Select <2date> to edit the date. The date must be specified in YYYYMMDD format.
When a character is selected the cursor moves to the next character. To leave a
blank character, press [ENT] with the field for that character blank. Use this for
the trailing characters if the ID being entered is less than twelve characters.
After the last of the twelve characters has been entered, the <Save ID?> option is offered.
Select <1no> to return to the ID edit screen. Select <2yes> to save the edited ID.
The ID can be set remotely from a computer which is quite a bit more
convenient than entering characters from the keyboard (see Section 4.4.4,
ID command). The ID is not cleared or reset by any RESET function
(see Section 3.5.9).
3.5.5.5 <5LEVEL> (SECURITY)
PURPOSE
To set user protection levels to restrict access to certain functions and to edit the
password required for changing user levels.
PRINCIPLE
RPM4’s front panel user interface provides the means to access all RPM4 user
defined data, settings and functions including calibration data. Inadvertent,
uninformed or unauthorized altering or deleting of data, settings and functions
could require extensive reconfiguration by the user and might cause invalid
readings and behavior. For these reasons, depending upon the application in
which RPM4 is being used, it may be desirable to restrict access to some
functions. The user level function provides a means of restricting access to certain
functions. Four different levels of security are available.
Access to changing security levels can be left open, or be protected by a password.
Security Levels
The security levels are structured to support typical operating environments as follows:
NoneThis level is intended for use only by the system manager and/or
calibration facility. It allows access and editing in all areas including
critical metrological information.
LowLow security is designed to protect the specific metrological
information and SYSTEM DIAGNOSTIC AND MAINTENANCE
functions of the system against accidental alteration. It is intended for
an advanced operator performing many different tasks. Low security
is the default user level setting.
Medium Medium security is designed to protect specific metrological
information in the system and to assure that the RPM4 is operated
using consistent operational parameters.
HighHigh security is designed to protect all operating parameters. It is
intended to minimize operator choices, for example to perform
repeated identical calibrations under consistent conditions.
RPM4 is delivered with the security level set at low to avoid inadvertent
altering of critical internal settings but with access to changing security
levels unrestricted. It is recommended that the low security level be
maintained at all times. If there is a risk of unauthorized changing of the
security level, changing authority should be password protected
(see OPERATION of this section).
The security levels are structured to support typical levels of operation as shown
in Table 8. Specifically, the security levels prevent execution of the functions
accessed by the key strokes marked by “•”:
RPM4 is delivered with no active password and access to the User Level menu is
open. The user level is set to <1Low>. User levels can be changed freely until a
password has been created.
To access the LEVEL function
press [SPECIAL], <5prefs>, <5level>.
The display is:
Selecting <1change user level> brings
up the restriction menu:
1change user level
2edit password
Restrictions: 1none
2low 3medium 4high
The cursor is on the current restriction level. Select a different level or
or [ESC] back to the MAIN RUN screen.
If no password is active, selecting
<2edit password> displays the user
password and allows it to be edited.
Password: pppppp
0 disables password
Once a password has been entered, the user level cannot be changed without
reentering the password.
Passwords can be up to six numbers in length and cannot start with a zero.
If <0> is entered, the password is made inactive and the user will not be required
to enter a password to access the user level menu. This condition, with a
security level of <2low>, is the factory default.
If there is an active password, the
RPM4 password entry screen appears.
The user must enter the user defined
password or the factory secondary
RPM4 SNnnnn-xx
Password: pppppp
password to proceed further:
The first field, <nnnn>, is the serial number of the RPM4, followed by a second field,
<xx>, that represents the number of times that a secondary password has been
used. The second field, <xx>, increments each time a secondary password is
used. The third field, <pppppp> is for user entry of the normal password.
The factory secondary password is available in case the user’s password has
been misplaced or forgotten. It can be obtained by contacting a DHI
Authorized Service Center (see Table 29). The factory secondary
password
is different for all RPM4’s and changes each time it is used.
3.5.6 <6PUNIT>
PURPOSE
To customize the selection of pressure units of measure that are available in the [UNIT] key menu.
The [UNIT] function key makes available a choice of six default pressure units (US or SI units
depending on whether the RPM4 has been factory set as US or SI) (see Section 3.3.2).
RPM4 also supports many commonly used units other than those included in the default set up.
These units can be made available for active selection by customizing the UNIT function
using [SPECIAL], <6PresU>. This allows RPM4 to offer a very wide selection of units while
simplifying day to day operation. The typical user customizes the [UNIT] function key to
support his/her six most commonly used units.
OPERATION
To customize the [UNIT] function key, from the MAIN RUN screen press [SPECIAL],
<6PresU>. The display is:
1. Entry field to specify which unit position (1 to 6) of the [UNIT]
function key menu is to be changed.
Set up unit #6
Enter the number of the unit position that you would
like to change. The display becomes:
Unit#6 1SI 2other
3altitude 4user
Select the desired pressure unit category (SI units include units based on SI units such as
mmHg), then select the desired unit from the unit menu.
The units of measure available are listed in Table 9.
Table 9. UNIT Function - Available Units of Measure
<1SI> <2OTHER> <3ALTITUDE> <4USER>
<1Pa>
<2hPa>
<3kPa>
<4MPa>
<5mbar>
<6bar>
<7mmHg>
<8mmWa>
<1psi>
<2psf>
<3inHg>
<4inWa>
<5kcm2>
<6Torr>
<7mTor>
<1feet>
<2meters>
If <4user> is selected, the user unit must be defined. The display is:
Define user unit:
1. Entry field.
1.000000 unit/Pa
Enter the number of user units per Pascal (Pa) in the entry field. Pressing [ENT] defines the
user unit and returns to the <Set up unit #n> screen.
The user defined unit label can be customized to any alphanumeric, four character label
using the remote command “UDU” (see Section 4.4.4).
See Section 7.2.1 for the pressure unit conversion factors used by RPM4.
<5log>View the RPM4 incident log (see Section 3.5.7.5).
3.5.7.1 <1BARO>
PURPOSE
To view the value of atmospheric pressure as measured by the RPM4 on-board
barometer.
There is no on-board barometer in RPM4s that have G100K or G200K
Q-RPTs only”.
PRINCIPLE
RPM4’s that include an Axxx (absolute) and/or BGxxx (bi-directional gauge)
Q-RPT are equipped with a separate, on-board barometer. The atmospheric
pressure measurements made by the on-board barometer are used for dynamic
compensation of atmospheric pressure when using an Axxx Q-RPT to make gauge
pressure measurements (see Sections 3.3.3, PRINCIPLE and 3.2.2) and for static
pressure compensation when using a G15K or BG15K Q-RPT.
See Section 5.5, Figure 7 and Section 5.6, Figure 8 for a photo and
The on-board barometer is a low accuracy sensor used only for measuring
schematics showing the position of the on-board barometer in RPM4.
small changes in atmospheric pressure over short periods of time
(see Section 3.2.2) and for line pressure compensation when using a G15K or
BG15K Q-RPT. RPM4 measurement uncertainty does not depend on the
measurement uncertainty of the on-board barometer.
RPM4™ OPERATION AND MAINTENANCE MANUAL
RPM4’s that have only G100K or G200K Q-RPTs (no Axxx, G15K or BG15K)
are NOT equipped with an on-board barometer.
OPERATION
To view the current reading of the on-board barometer press [SPECIAL],
<7internal>, <1baro>. The display is in the active pressure unit of measure
(see Section 3.3.2). The display resolution is fixed to 0.01 Pa, or its equivalent.
3.5.7.2 <2READRT>
PURPOSE
To turn ON and OFF RPM4’s automated, rate of change dependent, reading
integration rate feature.
PRINCIPLE
To obtain maximum resolution from RPM4 Q-RPT pressure measurements, an
integration time of about 1.2 second per reading is used. In most situations,
maximum precision is needed when pressures are stable so a relatively slow
display update rate presents no disadvantage. However, when pressure is
changing quickly, more rapid pressure updates are usually more important than
obtaining maximum precision on individual readings. The RPM4 read rate
function automatically adjusts pressure measurement integration time depending
on the rate of change of pressure. When pressure is changing rapidly, reading
rate is increased. When pressure is evolving slowly, reading rate is decreased
and maximum precision is obtained.
When the automated read rate function is ON, three pressure rate of change
dependent read rates are used. The result is three display update rates:
Table 10. READRT - Display Update Rates
PRESSURE
RATE OF CHANGE
> 3 % of range span/s
> 0.5 and < 3 % of range span/s
< 0.5 % of range span/s
DISPLAY
UPDATE
≈ 0.2 s
≈ 0.6 s
≈ 1.2 s
For situations in which maximum reading precision is desired regardless of
pressure rate of change, the RPM4 automated read rate function can be turned
OFF. In this case, the reading rate is always the high resolution rate of about 1.2
seconds per reading.
OPERATION
To turn the automated read rate function ON or OFF or check its current status,
press [SPECIAL], <7internal>, <3ReadRt>.
Selecting <1on> activates the automated reading rate and returns to the MAIN
RUN screen. Selecting <2off> turns OFF the automated reading rate and
returns to the MAIN RUN screen.
The default RPM4 condition is auto read rate ON. The read rate can be set to a
fixed value between 200 ms and 20000 ms using the remote command
“READRATEn” (See Section 4.4.4).
Auto read rate ON/OFF is NOT Q-RPT or range specific. Turning auto read
rate ON or OFF in one range turns it ON or OFF for all RPM4 ranges.
3.5.7.3 <3RPT2X>
PURPOSE
To enable and disable the RPM4 parallel measurement mode in which two
Q-RPT modules are used to measure in parallel and their readings are averaged.
PRINCIPLE
RPM4s that have two Q-RPT modules of the same type (Axxx, Gxxx or BGxxx)
support parallel measurement mode. In this mode, both Q-RPTs are used to
measure the same pressure in parallel and their readings are averaged.
Parallel measurement mode sets up a third Q-RPT designated HL (combination
of Hi and Lo). The HL RPT can be selected by the [RANGE] function and
supports other RPM4 functions such as UNIT, MODE and DISPLAY.
When parallel measurement mode is set up, the resulting HL Q-RPT can be a
third Q-RPT in addition to the Hi and Lo RPTs. It is also possible to make the HL
Q-RPT the sole Q-RPT, suppressing the Hi and Lo Q-RPTs. This feature is used
when an RPM4 will always be used in parallel measurement mode and the
availability of the individual Hi and Lo Q-RPTs is not desired.
[SPECIAL], <7internal>, <3RPT2x> is used to set up or suppress parallel
measurement mode in RPM4s that support it.
In parallel measurement mode, the maximum pressure is the maximum
pressure of the Lo Q-RPT.
OPERATION
Parallel measurement mode is only available on RPM4s with two Q-RPTs of
the same type (Axxx, Gxxx, BGxxx).
To access the parallel measurement mode enable/disable function press
[SPECIAL], <7internal>, <3RPT2x>.
The cursor is on the number corresponding to the current setting.
Select <1off> to disable parallel measurement mode. This will eliminate the HL
Q-RPT.
Select <2on> to enable parallel measurement mode. This creates the HL
parallel mode Q-RPT. The Hi and Lo Q-RPTs are also still available.
Select <3on, w / Hi & Lo RPT off> to enable parallel measurement mode and
disable access to the Hi and Lo Q-RPTs individually. This creates the HL
parallel mode Q-RPT and makes it the sole Q-RPT available for selection.
Operation returns to normal when the disagreement between the two Q-RPTs
returns to normal.
When using parallel measurement mode, the Hi and Lo Q-RPTs should be
connected IN PARALLEL to the pressure to be measured (both Q-RPTs
connected to the same pressure). The measured pressure indicated in
parallel measurement mode is the average of the Hi and Lo Q-RPTs.
3.5.7.4 <4LO VNT>
PURPOSE
To cause an RPM4’s Gxxx or BGxxx Q-RPT TEST(-) vent valve to stay open or
closed rather than open and close automatically.
This feature is active only when the active Q-RPT is a Gxxx (gauge) or
BGxxx (bi-directional gauge) Q-RPT.
PRINCIPLE
An RPM4 equipped with a Gxxx (gauge) or BGxxx (bi-directional gauge) Q-RPT has
a TEST(-) vent valve in its Q-RPT module (see Figure 8, Gxxx, BGxxx, Valve C).
The TEST(-) vent valve is used to open the TEST(-) port to atmosphere which
also connects the Q-RPT and the device or system under test low side to atmosphere.
In normal RPM4 operation, the TEST(-) valve is closed to isolate the low pressure
circuit from fluctuations in ambient pressure when the Q-RPT is being used to
measure pressure. It is open only when SDS is closed (see Section 3.2.7).
However, it may be desirable to open the TEST(-) port to atmosphere while
measuring pressure and/or without closing SDS. For example, this may be
necessary to assure that the pressure on the low side of the Q-RPT and the
system under test does not move too far away from atmosphere. [SPECIAL], <7internal>, <4lo vnt> allows the TEST(-) vent valve to be set to be
permanently open, permanently closed or to operate in its default, automated
condition. It also allows the current status of the valve to be viewed.
OPERATION
The normal, default setting for the TEST(-) vent valve is <1auto> in which RPM4
automatically controls the valve based on current Q-RPT module operation.
Setting other than <1auto> should only be selected by advanced users.
To access the LO VNT function press [SPECIAL], <7internal>, <4lo vent>. If the
RPM4 does not have a Gxxx (gauge) or BGxxx (bi-directional gauge) Q-RPT, the
display is: <Available with Gxxx or BGxxx Q-RPT only>. If the RPM4 has
more than one Gxxx or BGxxx Q-RPTs, the Hi or Lo Q-RPT must be selected.
The next display is:
Test(-) vent:1auto
2open 3close 4view
Select <1auto> for the TEST(-) vent valve operation to be controlled automatically
by RPM4 based on current operation. This is the default and recommended setting.
Select <2open> to cause the TEST(-) valve to open and remain open regardless
of RPM4 operation until the setting is changed. If RPM4 detects a condition
in which it determines there could be a risk of overpressure to the Q-RPT
by opening the TEST(-) vent valve, the message <Cannot open lo vent, DP overpressure risk> is displayed. If this message is displayed, return to
normal operation, vent the RPM4 Q-RPT module and try again.
Select <3close> to cause the TEST(-) vent valve to close and remain closed
regardless of RPM4 operation until the setting is changed.
Select <4view> to view the current condition of the TEST(-) valve. The display
indicates whether current valve operation is controlled by the RPM4
(<Auto TEST(-) vent>) or has been set manually (<Manual TEST(-) vent>)
followed by the current valve status (<Open> or <Close>).
If the RPM4 is being used as an external device to a PPC3 pressure controller
(see Section 3.5.7.4), the Lo Vent setting may be changed from the PPC3
(see the PPC3 Operation and Maintenance Manual).
3.5.7.5 <5LOG>
PURPOSE
To view and/or clear the RPM4 event log.
PRINCIPLE
RPM4 records to a log each time one of the following events occurs:
•Pmax! of an internal RPM4 Q-RPT or utility sensor is exceeded (see Section
3.4.4.1).
•A memory fault occurs.
OPERATION
To view the event log press [SPECIAL], <9Log>.
Use [
] and [] to move back and forth between older and newer log entries.
Each log entry has two screens, one with the event description and one with the
even time and date. Use [
oldest logged event appears.
After the last log has been viewed, the option to clear the log, <1no>, <2yes> is
presented. Use <2yes> to remove all entries from the log. Use <1no> to
continue without altering the log.
To leave the log, use [ESC].
3.5.8 <8CAL>
PURPOSE
] and [] to toggle between the two screens. The
To calibrate the RPM4 Hi and Lo Q-RPTs and adjust the on-board barometer. These
functions are considered part of RPM4 maintenance and are covered in the maintenance
section of this manual (see Sections 5.2, 5.3).
To reset various RPM4 settings to default or factory values.
PRINCIPLE
RPM4 stores its user definable settings in non-volatile memory. The reset menu allows the
user to selectively or completely reset these settings to factory defaults. This clears any
settings that the user has made, and should be used only to restore the RPM4 to a known
state. RPM4 goes through its power up sequence after any type of reset is executed.
RPM4 reset functions will change current settings to factory defaults. These may include
settings vital to RPM4 operation and affecting the calibration of the quartz reference
pressure transducers (Q-RPTs). Reset functions should only be used by qualified personnel
with knowledge of reset consequences. Reset functions should never be used “experimentally”.
OPERATION
To access the RESET menu, press [SPECIAL],
<9reset>. The display is:
1sets 2units 3ATest
4cal 5all
RESET menu choices include:
<1set> to reset general system operating parameters (see Section 3.5.9.1).
<2units> to reset unit of measure functions (see Section 3.5.9.2).
<3ATest> to reset AutoTest parameters and clear the AutoTest data log (see Section 3.5.9.3).
<4cal> to reset internal calibration coefficients and modes (see Section 3.5.9.4).
<5all> to reset all settings except ID and security password to factory default values
(see Section 3.5.9.5).
3.5.9.1 <1SETS>
PURPOSE
Sets most general operating parameters back to default values. Does not affect
calibration coefficients, remote interfaces or AutoRange ranges. The Reset – Sets
resets are itemized in Table 11.
Stability Limit 0.005 % of Hi Q-RPT default range 3.4.3
Upper Limit Hi Q-RPT default range default value 3.4.4
Resolution 0.001 % FS of Hi Q-RPT default range 3.4.2
AutoZ AutoZ ON for all Q-RPTs and measurement modes.
AutoZ
Leak Check 15 second run time. Clear logged results 3.3.5
Screen Saver 10 minutes to activation 3.5.5.1
Key Sounds Medium tone valid key press sound 3.5.5.2
Lo Vnt Automatic 3.5.7.4
ReadRate Automatic 3.5.7.2
Pressure unit of measure to first of the six available 3.3.2
Measurement mode to the native mode of the Q-RPT 3.3.3
Hi Q-RPT default range 3.3.1
0 cm height and Nitrogen medium 3.3
Z
value not affected
offset
P
set to 101.325 kPa a
atm,0
3.5.1
3.5.1, 3.2.2
3.5.9.2 <2 UNITS>
PURPOSE
Sets the six pressure units available under the UNIT function to the SI or US
default selections depending on whether the RPM4 has been factory set for SI or US
(see Section 3.3.2).
Sets the user defined unit to 1.000/Pa (see Section 3.5.6).
Sets the reference temperature for inWa unit to 20°C.
3.5.9.3 <3ATEST>
PURPOSE
• Resets Auto Tests to default characteristics (see Section 3.3.10).
• Clears AutoTest data log (see Section 3.4.5).
3.5.9.4 <4 CAL>
PURPOSE
The Reset - Cal function will reset Q-RPT, utility sensor and barometer
calibration coefficients and settings and reset AutoZ values to zero. This
will change the RPM4 calibration and could cause it to make out of tolerance
measurements.
Clears all user values affecting the calibration of Q-RPTs and the on-board
barometer. The Reset – Cal resets are itemized in Table 12.
Calibration Date Set all dates to 19800101 5.2.7, 5.3
AutoZ values Axxx Q-RPTs: All P
AutoZ function ON, all Q-RPTs, all measurement modes 3.5.1
3.5.9.5<5 ALL>
Table 12. Reset – Cal
PA to zero, PM to 1 5.2.1.1, 5.2.7
ON 5.2.5
PA to zero, PM to 1 5.3
values to zero for
absolute mode and 101325 Pa for gauge and
negative gauge modes
Gxxx, BGxxx Q-RPTs: All P
offset
values to zero
offset
3.5.1
The reset - all function clears and deletes large amounts of user defined
information including critical calibration data.
PURPOSE
Combines all resets in one global reset command that clears the entire user
section of non-volatile memory except the ID function (see Section 3.5.5.4) and
the security level password (see Section 3.5.5.5) returning RPM4 to the
“as delivered” condition. The Reset – All resets are itemized in Table 13.
Table 13. Reset – All
RESET RESULT SEE SECTION
Reset – Sets All the resets of Reset - Sets 3.5.9.1
Reset – Units All the resets of Reset - Units 3.5.9.2
Reset – Atest All the resets of Reset - ATest 3.5.9.3
Reset – Cal All the resets of Reset - Cal 3.5.9.4
Remote Interfaces COM1, COM2 and IEEE-R88 interfaces to
default settings.
Remote Communications Remote command format to Classic 3.5.2.3
Level (Security) Reset security level to low 3.5.5.5
Most of the RPM4 front panel functions can also be executed by commands from a remote computer.
The host computer can communicate to the RPM4 using the RPM4’s COM1 RS232 port or it’s IEEE-488 port.
Before writing test code using RPM4 remote commands, familiarize yourself with its operating principles
by reading Section 3 of this manual.
4.2 INTERFACING
Sending a program message to the RPM4 places it into remote mode. The remote indicator to the right
of the display window lights when the RPM4 is in remote mode. It also flickers when a program message
is received. The menus usually accessed from the front panel are locked out while in remote. The [ESC]
key returns the RPM4 to local operation unless the <REMOTE> program message, which locks out all
keypad operation, was sent to the unit.
4.2.1 RS232 INTERFACE
4.2.1.1 COM1
The RPM4 COM1 RS232 interface is located on the back of the unit. It is a 9-pin
male DB-9F connector configured as a DCE device. Data is transmitted out of
the unit using pin 2, and is received on pin 3. This allows a normal pin-to-pin
DB-9M to DB-9F RS232 cable to be used to connect to a DTE host.
Handshaking is not required or supported. The COM1 receive buffer is
80 bytes deep. If you overflow the buffer by sending too much data, the data will
be lost. Because of this, you must send a single program message at a time
and you must wait for the RPM4 to reply from the previous command before
issuing another command.
Table 14. COM1 Pin Designations and Connections
RPM4 COM1 DB-9F PIN DESIGNATIONS
PIN # FUNCTION DESCRIPTION
2 TxD This pin transmits serial data from the RPM4 to the host.
3 RxD This pin accepts serial data from the host computer.
5 Grn This pin is the common return for the TxD and RxD signals.
The RPM4 IEEE-488 interface is located on the back of the unit. The physical
and electrical interface conforms to IEEE Std 488.1-1987 Subset E2 and IEEE
Std. 488.2-1992. You should not attempt to communicate with the IEEE-488
interface while using the COM1 interface. The IEEE-488 receive buffer is 250
bytes deep. If you attempt to overflow the buffer, the RPM4 will hold off release
of the NRFD handshake line until it can service and empty the receive buffer.
This keeps the buffer from overflowing. It is recommended that you check for
errors using the “ERR?” query after sending a group of non-query program
messages. When using queries, ensure that you wait for a reply to each query to
ensue proper operation and order of command execution. Replies to queries
remain in the reply queue until the host gets them, so they can “stack up”,
causing replies to appear out of sequence.
4.2.1.3 COM2
The RPM4 COM2 RS232 interface is located on the back of the unit. It can be used
to allow the host computer to communicate with another device through the RPM4.
This allows the user to use one host COM port to communicate with the RPM4 and
an additional RS232 device. Refer to the “#” remote program command for
details. When using the RPM4 as a reference for a PPC3, another RPM4 can be
connected to the first RPM4’s COM2 port, allowing the PPC3 to communicate
with both RPM4s (see Section 3.2.5).
COM2 is a 9-pin female DB-9F connector configured as a DTE device. Data is
transmitted out of the RPM4 using pin 3, and is received on pin 2. This allows a normal
pin-to-pin DB-9M to DB-9F RS232 cable to be used to connect to a DCE device.
Handshaking is not required or supported.
Table 15. COM2 DB-9F Pin Designations
PIN # FUNCTION DESCRIPTION
2 RxD This pin transmits serial data from the RPM4 to a device.
3 TxD This pin accepts serial data from the external device.
4 DTR This pin is Data Terminal Ready (DTR) (held at + 5 V).
5 Grn This pin is the common return for the TxD and RxD signals.
4.3 PROGRAMMING FORMATS
RPM4 supports two program message formats, the “classic”, and the “enhanced” formats. The user
must select which format to use. Selection can be accomplished from the front panel (see Section
3.5.2.3) or remotely using the “L2” or “L3” program message (see Section 4.4.4). The ”MSGFMT”
command can also be used to select the format, but is not recommended for new designs.
The main difference between the “classic” and “enhanced” formats is that when using the IEEE-488 interface,
a query operator “?” must be included in an enhanced command to yield a reply from the RPM4. When
using the COM1 port in classic or enhanced mode or using the IEEE-488 port in classic mode, every
command has a reply which the host must wait for before continuing. In addition, the enhanced message
format supports IEEE Std 488.2 syntax, format and status reporting. The default is the classic format.
In either format, it is recommended that you start out a command sequence with the “*CLS” command,
which clears all of the communication and error queues. The basic commands are similar for both the
classic and enhanced formats, but the usage, syntax, format and status reporting are different.
Many RPM4 classic and enhanced commands are common with PPC2+, PPCK+ and PPC3
Pressure/Controller Calibrators.
Each program message sent is also a query. You can only send one program message to
the RPM4 at time. After sending any program message, you must wait for the RPM4 to reply
before sending another program message. This reply will contain data, or a numeric error
message if the program message was invalid. You must wait for this reply before issuing
another program message to the RPM4. This insures that the RPM4 has completed the
program message. Most remote program messages will return a reply within 500 ms except:
“PR?”, “PRR?”, “SR?”, “ATM?”, “RATE?”: Up to 2 seconds.
“RPT”, “ARANGE”, “AUTOZERO=RUN”: Up to 3 seconds
The syntax and format used for each program message in the classic mode is listed next to
the keyword “Classic”’ in each program message summary in Section 4.4.4.
4.3.2 ENHANCED PROGRAM MESSAGE FORMAT
The enhanced program message format uses the IEEE Std. 488.2 format, syntax and status
reporting. Errors are reported using the IEEE Std. 488.2 status reporting model. If an error is
reported, the error is put into an Error Queue and the “ERR?” query program message can
be used to get a text description of the most recent error. If you are using the IEEE-488 port,
the service request line can be setup to be asserted if this occurs (see Section 4.5.2). In the
enhanced format, there are two possible program message types for every program
message. Each of these two types starts with the same basic text referred to as the program
message header. The two types are COMMAND type and QUERY type commands.
4.3.2.1 USING COMMAND TYPE COMMANDS
Enhanced format commands DO NOT reply when using the IEEE-488
interface unless a “?” is included in the command. Do not expect a response
from the RPM4’s IEEE-488 interface to non-query (no “?”) commands as
there is none. Remote software will time-out waiting for a response from
RPM4. However, in RS232 communications, there is always a response and
the response MUST be read prior to issuing another command.
The COMMAND type of program message executes a process and can
additionally send data to the RPM4 in the form of comma delimited arguments.
This data is usually a setting of some sort that is stored in the RPM4. If data is
specified, it must be preceded by at least one white space from the program
message header and be within the range and format described in the program
message description. The keyword “Command:” appears to the left of the
required syntax in each program message description in Section 4.4.4.
If you are using the IEEE-488 port, the Command type does not generate a reply
unless you place a query operator “?” immediately after the command. You also
may send multiple program messages at once by separating each program
message with a semicolon. The commands are queued and executed in as
received order after the entire message stream has been received, so care in
determining order of execution is needed.
If you are using the RS232 port COM1, the Command type will always generate
a reply so you must wait for a reply before issuing another program message.
Because of this, you can only send one Command program message at a time
while using the COM1 port.
•IEEE-488 enhanced mode command series using query operator:
“*CLS?” (Clear the error queue. Wait for reply)
“UNIT? KPA” (Generates a reply. User must wait for reply before continuing)
“MMODE? A” (Generates a reply. User must wait for reply before continuing)
“PR?” (Generates a reply. User must wait for reply before continuing)
•IEEE-488 enhanced mode commands without query operator:
“*CLS” (Clear the error queue. No reply)
“UNIT KPA” (No reply)
“MMODE A” (No reply)
“ERR?” (Wait for reply. User should use “ERR?” query following a
series of non query commands to check for errors that may
have occurred)
•IEEE-488 enhanced mode multiple commands without query operator:
“*CLS” (Clear the error queue. No reply)
“UNIT KPA;MMODE A” (Two command at once. No reply)
“ERR?” (Wait for reply. User should use “ERR?”
query following a series of non query
commands to check for errors that may
have occurred)
•COM1 enhanced mode command:
“*CLS” or “*CLS?” (Clear the error queue. Wait for reply)
“UNIT KPA” or “UNIT? KPA” (Generates a reply. User must wait for
reply before continuing)
“MMODE A” or “MMODE? A” (Generates a reply. User must wait for
reply before continuing)
“PR” or “PR?”(Generates a reply. User must wait for
reply before continuing)
4.3.2.2 USING QUERY TYPE COMMANDS
The QUERY type of program message just requests data from the RPM4.
Placing the query operator “?” immediately after the command creates a query.
You must wait for a reply with a query. If you send any type of program
message to the RPM4 after a query before receiving a reply, the program
message is discarded and an error is generated. Errors are reported using the
IEEE Std. 488.2 status reporting model. A Query program message always ends
with a question mark. Most queries return a reply within 200 ms except:
“PR?”, “PRR?”, “SR?”, “ATM?”, “RATE?”: Up to 2 seconds.
“RPT”, “ARANGE”, “AUTOZERO RUN”: Up to 3 seconds
The syntax for using a QUERY program message is listed next to the keyword
“Query:” in each program message summary in Section 4.4.4. Please note that
queries in enhanced mode via the IEEE-488 port that results in an error will not
result in a reply. You must check the error queue (use the “ERR” querie) to see
if an error has occurred.
Some commands accept an optional suffix. This suffix can be used to specify the explicit Q-RPT
to address. In most cases, each Q-RPT in the RPM4 operates as an independent
measurement device. If the suffix is not specified, then the “Active” Q-RPT is assumed.
The “RANGE” or “ARANGE” command can be used to select the “Active” Q-RPT. There are
some limitations to use of the suffix when the RPM4 is in differential mode (see Section
3.3.3.1) or the “HL” Q-RPT is active (see Section 3.2.4). See the program message
descriptions for suffix use with a specific command.
Table 16. Program Message List
COMMAND
‘n’ indicates an
optional suffix
#
ABORT
ARANGEn
ATM
AUTOZEROn
AUTOZEROn=RUN
CALAMB
COM1
COM2
CONTn
DATE
ERR
GPIB
HEAD
ID
L2
L3
LLn
LOCAL
MEM
MMODEn
MSGFMT
NVENTn
PCALn
PCAL:XX
PRn
PRRn
QPRRn
RANGE
RATEn
READRATEn
READYCKn
REMOTE
RESn
RESET
RPTn
SCRSAV
SDSn
SDSACT
SDSAUTOn
SN
SRn
SSn
SS%n
TIME
UCOEFn
UDU
Send a command string out of the RPM4 COM2 port.
Stop pending operations
Read or set a new AutoRange to use.
Read the current atmospheric pressure (on-board barometer).
Read or set the status of the AutoZ automatic zeroing function.
Run AutoZero.
Read or set the on-board barometer calibration
Read or set the configuration of the COM1 port.
Read or set the configuration of the COM2 port.
Enable continuous measurements
Read or set the current date.
Read the last error message.
Read or set the GPIB interface address.
Read or set the fluid head settings.
Read or set the RPM4 alphanumeric asset ID tag.
Selects “classic” program message format
Selects “enhanced” program message format
Read or set the lower limit for the current range (negative gauge and differential
modes only).
Return control to the RPM4 front panel.
Read the power-up memory test status.
Read or change the active measurement mode.
Read or set the type of program message format to use.
Read or set the status of low vent valve (Gxxx or BGxxx Q-RPT only)
Read or set the user Lo Q-RPT calibration information.
Read or set the user Lo Q-RPT calibration information (old command).
Read the next RPM4 pressure.
Read the next RPM4 pressure, rate, and ATM.
Read the last RPM4 pressure, rate and ATM.
Select a Q-RPT to be “active” in it’s default full scale range
Read the next available rate of change of pressure.
Read or set the Q-RPT measurement read rate and mode
Read or set a flag that is cleared by a Not Ready condition.
Enable remote local lockout operation.
Read or set the pressure display resolution for the current Q-RPT and range.
Reset the RPM4 to default user parameters.
Read the available Q-RPT data.
Read or set the front panel screen saver period.
Read or set the SDS state for a specific Q-RPT
Read or set the active Q-RPT SDS state
Read or set the SDS mode for a specific Q-RPT
Read the serial number of the RPM4.
Read the next available pressure status (Ready/Not Ready).
Read or set the stability required for a Ready condition.
Read or set the stability required for a Ready condition (% span/ s).
Read or set the current time of day.
Convert a pressure in Pascal to pressure in the current units.
Read or set the user defined pressure unit.
Read or set the upper limit for the current range.
Read or set the pressure unit of measure for the current range.
Read the RPM4 software version.
Read or set the AutoZero Z
Read or set the AutoZ Z
4.4.2ERROR MESSAGESS
Table 17. Error #s and Descriptions
REPLY DESCRIPTION
ERR# 0 “OK”
ERR# 2 “Text argument is too long”
ERR# 3 “Arguments cannot be 0”
ERR# 4 “External device not detected”
ERR# 5 Not used
ERR# 6 “Numeric argument missing or out of range”
ERR# 7 “Missing or improper command argument(s)”
ERR# 8 “External device time-out error”
ERR# 9 “Unknown command”
ERR# 10 “Missing or invalid command suffix”
ERR# 11 “Command missing argument”
ERR# 12 “System overpressured” or “overpressure may result”
ERR# 13 “Text queue overflow”
ERR# 14 “User unit not defined”
ERR# 17 Not used
ERR# 18 “Command not yet available”
ERR# 19 “Not available with absolute units”
ERR# 20 “Not available with gauge device”
ERR# 21 Not used
ERR# 22 “Pressure is not stable”
ERR# 23 “Option not available or installed”
ERR# 25 Not used
ERR# 26 “COM port failed to initialize”
ERR# 27 “Internal device failure”
ERR# 28 “Device failure”
ERR# 29 “Device not available”
ERR# 30 “Must be on range HI”
ERR# 31 “Exceeds upper or lower limit”
ERR# 32 “Not stable enough”
ERR# 37 “Data table is full”
ERR# 38 “Selected range is not available”
ERR# 39 “Data verify error”
ERR# 45 “Argument not allowed”
ERR #46 “Argument cannot be negative”
ERR #52 “Command obsolete”
ERR# 53 “Not Available”
Purpose A brief description of the programs message’s function.
Command This is the Enhanced program message syntax to send data to the RPM4 or to execute an
Query This is the Enhanced program message syntax to request data from the RPM4. The RPM4
Classic This is the Classic program message syntax to send data to the RPM4, to execute an
Suffix Some commands support an optional suffix. This suffix is used to specify a Q-RPT to which
Arguments If the program message can be used to set data inside the RPM4, then this section describes
Default If the program message can be used to set data inside the RPM4, then this line shows
Remarks This field has the details and remarks about the command.
Example Examples are given for the enhanced and classic methods.
Errors If the program message can report an argument error, the types of errors are listed.
See Also
RPM4 function. The RPM4 must be set to use the enhanced format (see Section 3.5.2.3)
to use the syntax and style shown. It may be sent alone, or followed by at least one white
space and additional argument(s) to show that arguments can be passed. If there are
multiple arguments, then commas must separate them. If you are using the IEEE-488 port,
multiple command type program messages can be sent in one message if you separate
them with a semicolon. There will be no reply from the RPM4 using the IEEE-488 port
unless the command is immediately followed by the query operator “?”. If you are using
the COM1 port, the PPC1 will reply and you must wait for this reply. If this field is not
listed in the program message description, then the Command type is not supported when
using the Enhanced format.
must be set to use the enhanced format (see Section 3.5.2.3). The RPM4 will always
reply to a query. You must wait for this reply before issuing another program message. If
this field (Query) is not listed in the program message description, then the Query type for
the program message is not supported when using the Enhanced format.
RPM4 function, or to query for data. The RPM4 must be set to use the classic format
(see Section 3.5.2.3). The command may be followed by a ‘(=)’ and additional argument
characters to show that argument(s) can be passed. If there are multiple arguments, then
commas must separate them. The RPM4 will always reply to a Classic program message.
You must wait for this reply before issuing another program message. If this field is not listed
in the program message description, then it is not supported when using the classic
format.
the command applies, since many settings are Q-RPT specific:
Suffix of ‘1’ for Hi Q-RPT access
Suffix of ‘2’ for Lo Q-RPT access
Suffix of ‘3’ for HL Q-RPT access.
If an optional suffix is allowed but not given, then the “Active Q-RPT” will be addressed. The
“Active Q-RPT” is the Q-RPT that is currently displayed on the top line of the front panel of the
RPM4, and is identified by the text label in the upper right of the pressure display screen. The
last Q-RPT ranged or selected using the “RANGE” or “ARANGE” commands is the “Active” QRPT. The “ARANGE”, “RANGE” or “RPT” command query can be used to determine which
Q-RPT is currently active. With the Hi Q-RPT In differential mode of operation, there are some
limitations on access to the Lo Q-RPT, and with the HL Q-RPT there are limitations on access
to both the Hi and the Lo Q-RPTs independently.
the arguments and their limits.
(using the enhanced format) the default setting from the factory.
Enhanced:An example of the use of an enhanced format program message to be
sent to the RPM4 is shown. The message sent to the RPM4 appears after
the “Cmd sent:” label. If only a Query type exists, the “Query sent:”
label is shown instead. Directly under this label, “Query reply” shows a
typical reply to a query type. “Reply:” shows that a query format does
not exist. It may have a short description next to it.
Classic:An example of the use of a classic program message to be sent to the
RPM4 is shown. The command sent to the RPM4 appears after the
“Cmd sent:” label. The “Reply” label shows a typical reply to the
“Sent” example. It may have a short description next to it.
If using the classic format or the COM1 port, the error message is replied after receiving
the program message. If using the enhanced format via the IEEE-488 port, the error
condition is handled by the status reporting model which stores the errors in an Error
Queue and can be programmed to assert the IEEE-488 SRQ line to signal an error has
occurred. In either case, the “ERR” or “ERR?” program message can be used to
retrieve a text description of the error.
Indicates related commands (“----“) and refers to manual sections giving detail on RPM4
operation corresponding to the program message.
Each program message description is separated into the following sections:
Purpose To allow the host PC to communicate with a device connected to the RPM4 COM2 port.
Classic “#xx”
Arguments xx: The string to send out of the RPM4 COM2 port. It must be less than
Remarks The RPM4 COM2 port can be used to communicate to another RS232 device (such as
Example
(classic)
This example assumes that a second RPM4’s COM1 port is connected to the RPM4
See Also “COM2”
ABORT
Purpose Cancels any pending data requests by the “PR”, “PRR” “RATE”, “SR” or “CONT”
Command
Classic
Remarks This program message is usually used to ensure the RPM4 is in an idle condition.
Example
(enhanced)
Example
(classic)
See Also “PR”, “PRR” “RATE”, “SR” or “CONT”
another RPM4). This allows the user to use a single COM port or IEEE-488 port on the
host computer to communicate with the RPM4 and another RS232 device. A carriage
return and a line feed (<CR><LF>) are added to the string.
After this program message is issued, the RPM4 will reply back every string received by
the RPM4 COM2 port that is terminated with a carriage return. Line feeds are discarded.
This will discontinue when the next query is sent to the RPM4.
There is no other reply from this program message. Prior to using this program message,
you must ensure that the RPM4 COM2 port is correctly set up to communicate with the
device on COM2. Refer to the “COM2=“ program message.
Sent:
Reply:
COM2 port. This example gets the version of the second RPM4.
3.5.2
commands.
“ABORT”
“ABORT”
Cmd sent:
Reply:
Sent:
Reply:
3.2.1
40 characters long.
“#VER”
“DH INSTRUMENTS, INC RPM4 us A1000/A0015 Ver2.00 ”
“ABORT”
“ABORT” (no reply if IEEE-488)
“ABORT”
“ABORT”
Purpose Read or set a new AutoRange range to use.
Command
Query
Classic
Query
Optional Suffix “n”A suffix may be used instead of “RptLabel”
Arguments
Remarks The ARANGE command is used to define an RPM4 range by specifying a desired unit of
Example
(enhanced)
Example
(classic)
Example
(enhanced)
Example
(classic)
Example
(enhanced)
Example
(enhanced)
Errors ERR# 6:
See Also “RPT”, “RANGE”, “MMODE”, “UNIT”
“ARANGEn Range, unit, mode (,RptLabel)” (set to user defined range)
“ARANGE Q-RPTLabel” (set to Q-RPT default range)
“ARANGEn?”
“ARANGEn= Range, units, mode (,RptLabel)” (set to user defined range)
“ARANGE= Q-RPTLabel” (set to Q-RPT default range)
“ARANGEn”
‘1’ Specify the Hi Q-RPT if the Hi and Lo Q-RPTs are NOT OFF
Specify the HL Q-RPT if the Hi and Lo Q-RPTs are OFF
‘2’ Specify the Lo Q-RPT
‘3’ Specify the HL Q-RPT
Range: The AutoRange maximum pressure. Cannot be negative.
Unit: The unit of measure of the Range argument.
Mode: The measurement mode of the Range argument:
“A” for absolute
“G” for gauge
“N” for negative gauge
Measured pressure values are returned with “g” to identify measurement
mode for both gauge and negative gauge measurement modes.
RptLabel:
(optional)
One of the Q-RPTs that is available.
“IH”: Hi Q-RPT if the Hi and Lo Q-RPTs are NOT OFF. HL Q-RPT if
the Hi and Lo Q-RPTs are OFF.
“IL”: Lo Q-RPT
“HL”: HL Q-RPT.
measure, measurement mode and maximum pressure, and have the RPM4 pick the best
available Q-RPT and make adjustments to optimize operation in that range.
The optional suffix or the “RptLabel” argument can be used to specify which Q-RPT you
wish to use for the AutoRange range instead of allowing the RPM4 to pick the best QRPT. By using this option you can override the internal logic that picks the most suitable QQ-RPT. This Q-RPT specified must be valid for the maximum pressure and measurement
mode specified.
The reply indicates the current range data, including the Q-RPTLabel used for the range.
If “inWa” is specified for the unit, the temperature reference can be given after the unit
text (“inWa4”, “inWa20” or “inWa60” corresponding to inWa at 4 °C, 20 °C or 60 °F). If no
number is given, a default of 20 °C is assumed. There is no indication of the temperature
reference in the reply.
If you just specify the Q-RPTLabel without any other arguments, the Q-RPT specified will
become active in it’s last range and settings. The previously set unit and measurement
mode for the Q-RPT specified will become active.
Cmd sent:
Query reply:
Cmd sent:
Query reply:
Cmd sent:
Query reply:
Cmd sent:
Query reply:
Cmd sent:
Query reply:
Cmd sent:
“ARANGE?” (read the current range in the current units)
“100.00, psi, A, IH”
“ARANGE” (read the current range in the current units)
“100.00, psi, A, IH”
“ARANGE? 250, inWa4, G (range of 250 inWa @ 4 °C in gauge mode)
“250.00 inWa, G, IH” (Lo Q-RPT is used)
“ARANGE=500, kPa, G (range of 500 kPa in gauge mode)
“500.00 kPa, G, IL” (Lo Q-RPT is used)
“ARANGE2 50, psi, A (range of 50 psi in absolute mode on
the Lo Q-RPT)
“50.000 psi, A, IL” (No reply if GPIB-488)
“ARANGE IH (Hi Q-RPT’s prev range made active)
“500.000 kPa, A, IH” (No reply if GPIB-488)
Query reply:
“Range“ maximum pressure exceeds available Q-RPTs or is negative.
ERR#10:
ERR# 19:
ERR# 20:
ERR# 29:
ERR# 53:
RPT specified by the optional suffix is not valid for the current operating
conditions
Cannot Range to ‘0’ with absolute units.
Cannot Range to ‘0’ with gauge units.
Correct type of Q-RPT for the selected “Mode” is not available.
RPT specified is not available.
Purpose Reads the next measured pressure from RPM4’s on-board barometer (if present).
Query
Classic
Remarks The atmospheric pressure as measured by the RPM4 on-board barometer is returned in
Example
(enhanced)
Example
(classic)
Errors ERR# 23: RPM4 is not equipped with a barometer.
See Also 3.5.7.1
AUTOZEROn
Purpose Read or set the status of the AutoZ function.
Command
Query
Classic “AUTOZEROn= state”
Default “AUTOZERO1”
Optional Suffix “n”
Arguments ‘state’: ’0’ Autozero OFF
Remarks The RPM4 AutoZ function can be turned ON and OFF. There is a separate AutoZ flag for
Example
(enhanced)
Example
(classic)
Errors ERR# 7:
See Also “ZOFFSET”, “MMODE”
AUTOZEROn=RUN
Purpose Run the AutoZero routine.
Command
Query
Classic “AUTOZEROn= RUN (,Pref)”
Optional Suffix “n”
Arguments ‘Pref’: The AutoZ reference pressure in Pa. Optional when the RPM4 is not in
Remarks The RPM4 AutoZ function can be remotely run with this command only if AutoZero is
Example
(enhanced)
Example
(enhanced)
Example
(classic)
Errors ERR# 6:
See Also “ZOFFSET”, “MMODE”, “AUTOZERO”
“ATM?”
“ATM”
the current pressure units of the active Q-RPT (always absolute). This measurement is
followed by the units text. Not all RPM4s are equipped with an on-board barometer.
Query sent:
Query reply:
Sent:
Reply:
“ATM?”
“97.12348 kPa a”
“ATM”
“97.12384 kPa a”
“AUTOZEROnstate”
“AUTOZEROn?”
“AUTOZEROn”
The active Q-RPT is assumed if no suffix is given.
‘1’ Specify the Hi Q-RPT if the Hi and Lo Q-RPTs are NOT OFF
Specify the HL Q-RPT if the Hi and Lo Q-RPTs are OFF
‘2’ Specify the Lo Q-RPT
‘3’ Specify the HL Q-RPT if the HL Q-RPT is active
the gauge, absolute, and differential measurement modes for each Q-RPT. This
command sets the AutoZ status for the current mode of the specified or the active Q-RPT
only. To set the sate of AutoZ for a measurement mode, you must be in that mode.
Cmd sent:
Query reply:
Sent:
Query reply:
“AUTOZERO 0”
“0” (No reply if GPIB-488)
“AUTOZERO=0”
“AUTOZERO=0”
The argument was other than ‘0’ or ‘1’.
ERR# 10:
The suffix is invalid.
3.5.1
“AUTOZEROn RUN (,Pref)”
“AUTOZEROn? RUN (,Pref)”
The active Q-RPT is assumed if no suffix is given.
‘1’ Specify the Hi Q-RPT if the Hi and Lo Q-RPTs are NOT OFF
Specify the HL Q-RPT if the Hi and Lo Q-RPTs are OFF
‘2’ Specify the Lo Q-RPT
‘3’ Specify the HL Q-RPT if the HL Q-RPT is active
absolute measurement mode, as the reference is usually “0”.
ON. You should ensure that the pressure is stable for valid determination of Z
differential mode, the previous 5 measurements of the Q-RPTs are averaged for Z
calculation so you should ensure that the RPM4 has been stable for at least 5
measurement cycles before running a remote AutoZ in differential mode.
Cmd sent:
Query reply:
Cmd sent:
Query reply:
Sent:
Query reply:
“AUTOZERO RUN”
“OK” (No reply if GPIB-488)
“AUTOZERO? RUN, 97283” (abs AutoZero to 97283 Paa)
“OK”
“AUTOZERO=RUN”
“OK”
An argument was not given for the AutoZ reference in absolute mode.
ERR# 10:
ERR# 53